Biodegradable polymeric stents for vascular application in a porcine carotid artery model

English version

verfasst von: Dr. S. Kischkel, N. Grabow, A. Püschel, B. Erdle, M. Kabelitz, D.P. Martin, S.F. Williams, I. Bombor, K. Sternberg, K.-P. Schmitz, W. Schareck, C.M. Bünger

Erschienen in: Gefässchirurgie | Sonderheft 1/2016

Abstract

Over the past years the development of biodegradable polymeric stents has made great progress; nevertheless, essential problems must still be solved. Modifications in design and chemical composition should optimize the quality of biodegradable stents and remove the weaknesses. New biodegradable poly-L-lactide/poly-4-hydroxybutyrate (PLLA/P4HB) stents and permanent 316L stents were implantedendovascularly into both common carotid arteries of 10 domestic pigs. At 4 weeks following implantation, computed tomography (CT) angiography was carried out to identify the distal degree of stenosis. The PLLA/P4HB group showed a considerably lower distal degree of stenosis by additional oral application of atorvastatin (mean 39.81 ± 8.57 %) compared to the untreated PLLA/P4HB group without atorvastatin (mean 52.05 ± 5.80 %). The 316L stents showed no differences in the degree of distal stenosis between the group treated with atorvastatin (mean 44.21 ± 2.34 %) and the untreated group (mean 35.65 ± 3.72 %). Biodegradable PLLA/P4HB stents generally represent a promising approach to resolving the existing problems in the use of permanent stents. Restitutio ad integrum is only achievable if a stent is completely degraded.
Hinweise
The german version of this article is published in Gefässchirurgie (2014) 19:358–363.
Dr. S. Kischkel received the 2014 Springer Award for Vascular Medicine for this article.

Introduction

The ability to treat vascular disease using interventional approaches, as well as and the development of endovascular treatment techniques have made the use of stents routine. Compared with single-vessel dilatation, as in percutaneous transluminal angioplasty (PTA), they prevent in particular acute vascular recoil by providing scaffolding, thereby effectively reducing the restenosis rate [1]. Alongside permanent bare metal stents (BMS) of various designs, alloys and applications, biodegradable polymer stents are playing an ever more important role. Compared with conventional permanent BMS, they have the advantage of preventing chronic foreign-body reactions and enabling secondary vascular surgery in previously stented areas.
The aim of this article is to present the results from an animal model following endovascular implantation of an advanced and improved biodegradable PLLA/P4HB stent compared with a permanent 316L metal stent for use in the peripheral vasculature. The effect of an orally administered statin is also investigated.

Materials and methods

PLLA/P4HB stents

The biodegradable polymer stents used in this study were developed, produced and tested in vitro at the Institute for Biomedical Technology at the University Medicine Rostock [2].
In brief, the stents were produced in an automated, PC-assisted immersion process and cut from tiny tubes (Di = 2.2 mm, Da = 2.7 mm) using a CO2 laser [2]. The stents were made from a blend of poly(l-lactide) and poly(4-hydroxybutyrate) (PLLA/P4HB). The chemical composition of their basic substance consisted of 78 % PLLA (Resomer L214, Boehringer, Ingelheim, Germany) and 22 % P4HB (TephaFLEX, Tepha, Inc., Lexington, MA, US). Fig. 1 shows the PLLA/P4HB stent with its nominal dimension of 5.0 × 20 mm (wall thickness: 250 μm).

Metal stents

Non-biodegradable stents made from medical-grade stainless steel (316L) and coated with amorphous silicon carbide (Dynamic, Biotronik SE & Co. KG, Berlin, Germany) with a nominal dimension of 5.0 × 25 mm (wall thickness: 160 μm) were used as controls.

Animal preparation

The experiments were approved by the Mecklenburg-Vorpommern Regional Office for Agriculture, Food Safety and Fisheries in accordance with the regulations on animal experiments set out in §8 par. 1 of the German Animal Welfare Act. The study used 10 healthy female domestic swine weighing 26.1 ± 1.7 kg.
Food was withheld for at least 12 h preoperatively, beginning on the day prior to surgery; drinking water remained available at all times. The experimental animals were premedicated with azaperone [Stresnil®, 5 mg/kg body weight (BW); Janssen-Cilag, Neuss, Germany), ketamine (40 mg/kg BW; Bela-Pharm, Vechta, Germany), midazolam (Dormicum®, 0.2 mg/kg BW; Hoffmann La Roche, Grenzach/Wyhlen, Germany) and atropine sulfate (0.01 mg/kg BW; Germany)] via intramuscular injection in the neck area. The animals were intubated and received volume-controlled mechanical ventilation (Ventilog 2, Dräger, Lübeck, Germany) with a 40 % oxygen–1.2 Vol% isoflurane (Abbott GmbH & Co KG, Wiesbaden, Germany) mixture. The animals were anesthetized during surgery with fentanyl (0.25 mg/h i.v., Fentanyl®-Janssen, Janssen-Cilag, Neuss, Germany) and pancuronium (10 mg/h i.v., Pancuronium Hikma®, Hikma Pharma, Klein-Winternheim, Germany) administered to the ear via a 20-G indwelling catheter (Vasofix®, Braun-Melsungen AG, Germany). Fluid substitution was performed with Ringer’s lactate (Sterofundin®, Braun-Melsungen AG, Germany). Ampicillin/sulbactam (1.5 g; Unacid®, Pfizer, Karlsruhe, Germany) was administered preoperatively as a single-shot antibiotic prophylaxis.
One group of five animals also received 60 mg/day atorvastatin (Sortis, Pfizer, Germany) orally in addition to food, beginning 5 days prior to surgery and lasting until the study ended.

Stent implantation

Following surgical incision in the lower left quadrant of the abdomen, the left common iliac artery was exposed. A 6-F sheath (Radiofocus® Introducer, Terumo, Germany) was introduced by puncture. After probing the target vessel, an 8-F introducer system (Flexor® Tuohy-Borst Sidearm Introducer, Cook, Germany) was inserted over a guidewire (Rosen Wire, Cook, Germany). The stents were placed in the carotid artery under X-ray fluoroscopy and then balloon-expanded.
As a commercial stent-catheter system (Dynamic 5/25/130, Biotronik SE & Co. KG, Berlin, Germany), the metal stents were rapidly inflated with a pressure of 9 bar. The biodegradable PLLA/P4HB stents were implanted using a 5.0 × 40 mm PTA balloon catheter (Passeo-35: 5/40/130, Biotronik SE & Co. KG, Berlin, Germany). To this end, the PLLA/P4HB stents were mounted manually on the balloon catheter, predilated extracorporeally with 1 bar (Fig. 2) and dilated in situ at the target site with a pressure of 8 bar for 1 min. The side of the common carotid artery (CCA) on which a PLLA/P4HB stent or a 316L stent was implanted was left to chance. Following stent implantation, the balloon catheters were deflated and withdrawn. Once implantation was complete, layer-by-layer wound closure was performed.

Antithrombotic medication

The animals received continuous intravenous heparin (50 IU/kg per hour) during the entire intervention. Oral dual antiplatelet drugs comprising 250 mg acetylsalicylic acid (ASA; Aspirin®, Bayer, Leverkusen, Germany) and 75 mg clopidogrel bisulfate (Plavix®; Sanofi-Synthelabo, Berlin, Germany) were administered daily from postoperative day 3 until the end of the experiment.

Investigation methods

Basic parameters (BW, heart rate, arterial RR, pO2, pCO2, hemoglobin, hematocrit, partial thromboplastin time, C-reactive protein) were routinely monitored in all animals pre-, intra- and postoperatively.
Invasive intra-arterial angiography (digital subtraction angiography, DSA) was performed using a high-pressure injector (Medrad, Avida Injector, Germany) to assess vessel size before, and vessel reaction after, stent dilatation.
Contrast-enhanced computed tomography (CT) examinations were performed on the previously premedicated animals at the Radiological Unit of the University Medicine Rostock 4 weeks following stent implantation to assess stent patency and for the purposes of flow measurement.

Statistical analysis

Expansion length as well as stent and vessel diameter were measured from the CT data using the Leonardo Workstation (Siemens Medical Solution, Erlangen, Germany). The distal degree of stenosis was calculated according to the North American Symptomatic Carotid Endarterectomy Trial (NASCET), whereby the vessel inner diameter in the region of plaque-induced vessel stenosis was substituted by the stent inner diameter (IDS):
$$ {\rm{NASCET}}{\mkern 1mu} [\% ] = ({\rm{I}}{{\rm{D}}_{\rm{G}}} -{\rm{I}}{{\rm{D}}_{\rm{S}}})/{\rm{I}}{{\rm{D}}_{\rm{G}}}\cdot100 $$
(1)
Using statistical methods (SigmaPlot Version 12.0), the data were given as a mean value ± standard deviation(SD). Differences between two groups were compared using Student’s t-test for unpaired samples. The Friedmann test was used for more than two related samples, whilst the Kruskal-Wallis one-way analysis of variance was used for more than two unrelated samples. A p-value of less than 0.05 was considered statistically significant.

Results

All stents could be successfully implanted in the CCA in experimental animals. All animals survived until the end of the study period. No wound healing disorders were observed during this time. The animals exhibited a weight gain of 12.1 ± 2.3 kg within the 4-week period.
Following successful implantation, all stents showed good dilatation and patency on final angiography. None of the stents had collapsed or showed signs of relevant recoil. All stents had remained at the target site and patent following deflation and withdrawal of the balloon catheter. Figure 3 shows an X-ray demonstrating early dilatation in a polymer stent compared with a metal stent.
At 4 weeks following implantation, all carotid stents were open and permeable, irrespective of the side of the CCA on which stents had been placed. Figure 4 shows representative CT images of the CCA fitted with both types of stent in sagittal, coronal and axial planes.
A significantly higher degree of stenosis according to NASCET was seen in the CCA in the group with PLLA/P4HB stents compared with the metal stent group (52.05 ± 5.80 % vs. 35.65 ± 3.72 %; p < 0.002) 4 weeks following endovascular implantation. The distal degree of stenosis in the PLLA/P4HB stent group could be reduced to the level of the metal stent group by means of oral application of atorvastatin (39.81 ± 8.57 % vs. 44.21 % ± 2.34; p > 0.234). No differences were seen between animals receiving and those not receiving atorvastatin in the metal stent group (p > 0.07). In contrast, in the PLLA/P4HB stent group, a significantly lower degree of distal stenosis was seen in animals receiving atorvastatin compared with those not (p < 0.009) (Fig. 5).
The vessel lumen distal and proximal to the stent showed no differences between the two stent groups, irrespective of whether or not the animals had received atorvastatin (Tab. 1). No differences between nominal length (316L: 25 mm; PLLA/P4HB: 20 mm) and expanded length (316L: 25.27 ± 0.80 mm; PLLA/P4HB: 19.50 ± 1.13 mm) were observed in either stent group.
Table 1
Luminal patency of the vessel at the distal and proximal ends of the stent (inner vessel diameter, IDvessel). Data are presented as mean value±SD. The Kruskal-Wallis test was used to compare groups
  
316L
316L + atorvastatin
PLLA/P4HB
PLLA/P4HB + atorvastatin
p-Value
IDvessel (mm)
Distal
5.20 ± 0.35
5.42 ± 0.38
5.38 ± 0.63
5.60 ± 0.52
0.638
 
Proximal
5.22 ± 0.38
5.22 ± 0.72
4.74 ± 0.45
4.86 ± 0.25
0.302

Discussion

In this study, biodegradable PLLA/P4HB stents were successfully implanted in the CCA of domestic swine using an endovascular approach. It was possible to show the functionality of the PLLA/P4HB stent by direct comparison with a standard permanent 316L BMS.
The advantages of endovascular management of vascular disease compared with open surgery include reduced mortality and a lower complication rate [3]. The finality of using permanent stents has hitherto represented a problem in interventional treatment with stent placement. Once implanted, they are difficult to remove and, in contrast to biodegradable stents, make surgical intervention in the relevant vessel segment challenging. As lifelong foreign bodies, they continuously cause local immunological irritation in the vessel wall and the associated onset of in-stent restenosis (ISR) [4]. It is known that this causal inflammatory reaction and its effect on proliferation processes can be reduced by removing the foreign body, whereby the use of biodegradable stents represents an advantage.
There has been significantly less investigation into the clinical use of biodegradable stents in the peripheral vascular system compared with the current status of research relating to the coronary vasculature [5]. To date, they have only been used in infrapopliteal interventions in the peripheral vascular system, since smaller stents in this location need to withstand less stress [6]. Femoropopliteal arteries (superficial femoral and popliteal artery), on the other hand, are exposed to considerable mechanical stresses such as compression, bending and torsion. For this reason, there are varying stent fracture rates in vascular segments depending on stent design, with the result that biodegradable stents have only rarely been used in clinical applications in this region [7].
The present study used direct and indirect imaging methods to visualize the implanted stents. Direct methods permitted visualization of the stents themselves, whilst indirect methods demonstrated surrounding changes caused by the stent, without visualizing the stent itself. The permanent metal stents could be visualized, in both dilated and undilated states, at all times under X-ray fluoroscopy, making it possible to assess them directly and at any time. However, it was not possible to directly visualize the biodegradable polymer stents without X-ray markers due to lack of X-ray density during implantation. Only by inflating the balloon catheter with contrast medium was it possible to indirectly observe deployment behaviour. In this regard, biodegradable polymer stents rely on the use of X-ray markers, e.g. gold (Igaki-Tamai stent) [8] or platinum (bioresorbable vascular scaffold stent) [9]. Due to poor differentiability between polymer and vessel wall, it was not possible to measure the outer diameter of the PLLA/P4HB stents or stented vessels during contrast-enhanced CT. However, it was possible to define the inner diameter from the contrast medium column, and thus determine the distal degree of stenosis according to the NASCET definition [10]. The inner diameter of stent-induced stenosis in the distal (and proximal) vessels was comparable in this study, making the two stent groups equal in terms of distal degree of stenosis. At the same time, it should also be borne in mind that stent inner diameter was used to calculate the degree of stenosis, whereby the wall thickness of the PLLA/P4HB stents is proportionally 1.5 times greater compared with the 316L stents (316L: 160 μm vs. PLLA/P4HB: 250 μm).
The present study showed a significantly lower degree of distal stenosis in animals treated with oral atorvastatin compared with untreated animals in the PLLA/P4HB group at contrast-enhanced CT 4 weeks following implantation. Atorvastatin treatment was able to reduce the distal degree of stenosis in the PLLA/P4HB stent group to that of the metal stent group. Given the high prevalence and incidence of cardiovascular disease, statins play an important role in clinical routine. In addition to their main action, low-density lipoprotein (LDL) cholesterol reduction, statins also stabilize plaque [11] and demonstrably reduce restenosis rates [12]. In an animal study, statin-eluting stents were shown to have a significant effect on neointimal hyperplasia in stented coronary vessels in a swine model. An initial clinical study, however, showed statin-eluting stents to cause higher than expected neointimal proliferation [14]. According to the results of this study, biodegradable statin-eluting stents represent a promising opportunity in terms of further experimental investigations to increase drug concentrations at the implantation site and further minimize adverse side effects.
All experimental animals received dual oral anticoagulation with ASA and clopidogrel prior to surgery. The literature reports that this approach resulted in a reduction in occlusion rates in percutaneous transluminal coronary angioplasty (PTCA) with BMS [15], leading to dual antiplatelet therapy becoming established as the standard procedure in coronary interventions with stents [15]. Although studies to assess anticoagulation in stent implantation in the peripheral vasculature are scant, they suggest success similar to that seen in the coronary system [16]. Indeed, our model also showed no occlusion in stented vessels at 4 weeks under this form of dual anticoagulation therapy, as did earlier experiments conducted by our working group [17, 18].
To date, the polymeric stent used here is not sterilized, but rather only disinfected according to the study by Tamai et al. [8]. Various sterilization procedures can have a critical effect on the mechanical properties of the stent. Grabow et al. [19] discovered that sterilization methods using beta and gamma radiation cause a marked reduction in molecular weight and mechanical properties, while increasing crystallinity.
No specially designed balloon catheters to dilate the biodegradable PLLA/P4HB stents were available for the purposes of this study, whereas the 316L stents were available as commercial stent-balloon catheter systems. Therefore, balloon catheters for PLLA/P4HB stent implantation needed to be predilated extracorporeally after the stents had been mounted. Despite the demonstrated feasibility of advancing the predilated balloon catheter over an 8-F sheath with a rotating hemostasis valve for manual rotation (Tuohy-Borst) without removing the mounted stent, the development of a dedicated stent delivery system would make predilatation obsolete and contribute significantly to reducing the sheath diameter required.

Limitations

All materials used for stent implantation were used in the same way as in humans. This is possible due to the comparable anatomical proportions and physiology of the cardiovascular system in humans and swine [20]. A limitation of the study lies in the fact that stents were implanted in healthy, non-arteriosclerotic vessels; therefore, it is possible that the conclusions drawn here are only valid under the conditions presented.

Conclusion

  • The new biodegradable PLLA/P4HB stent demonstrated that it can be used in endovascular applications and that it is suitable for long-term testing in the established porcine model.
  • The future incorporation of biologically active substances aimed at suppressing neointimal hyperplasia could further reduce restenosis rates.
  • Overall, the PLLA/P4HB stent represents a promising approach to solving existing problems in the use of permanent stents. Restitutio ad integrum is only achievable if a stent is completely degraded.

Compliance with ethical guidelines

Conflict of interest

S. Kischkel, N. Grabow, A. Püschel, B. Erdle, M. Kabelitz, D.P. Martin, S.F. Williams, I. Bombor, K. Sternberg, K.-P. Schmitz, W. Schareck and C.M. Bünger state that there are no conflicts of interest.
All national guidelines on the care and use of laboratory animals have been followed and the necessary approval was obtained from the relevant authorities.
Open Access. This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Gefässchirurgie

Print-Titel

Themenschwerpunkte zu aktuellen Entwicklungen in der vaskulären und endovaskulären Gefäßmedizin. Vermittlung von relevantem Hintergrundwissen und Bewertung wissenschaftlicher Ergebnisse. Konkrete Handlungsempfehlungen.

Literatur
1.
Zurück zum Zitat Serruys PW, De Jaegere P, Kiemeneij F et al (1994) A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. Benestent Study Group. N Engl J Med 331:489–495CrossRefPubMed Serruys PW, De Jaegere P, Kiemeneij F et al (1994) A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. Benestent Study Group. N Engl J Med 331:489–495CrossRefPubMed
2.
Zurück zum Zitat Grabow N, Bunger CM, Kischkel S et al (2013) Development of a sirolimus-eluting poly (l-lactide)/poly(4-hydroxybutyrate) absorbable stent for peripheral vascular intervention. Biomed Tech (Berl) 1–9. doi:10.1515/bmt-2012-0050/j/bmt.ahead-of-print/bmt-2012-0050/bmt-2012-0050.xml [pii] Grabow N, Bunger CM, Kischkel S et al (2013) Development of a sirolimus-eluting poly (l-lactide)/poly(4-hydroxybutyrate) absorbable stent for peripheral vascular intervention. Biomed Tech (Berl) 1–9. doi:10.1515/bmt-2012-0050/j/bmt.ahead-of-print/bmt-2012-0050/bmt-2012-0050.xml [pii]
3.
Zurück zum Zitat Nowygrod R, Egorova N, Greco G et al (2006) Trends, complications, and mortality in peripheral vascular surgery. J Vasc Surg 43:205–216CrossRefPubMed Nowygrod R, Egorova N, Greco G et al (2006) Trends, complications, and mortality in peripheral vascular surgery. J Vasc Surg 43:205–216CrossRefPubMed
4.
Zurück zum Zitat Hoffmann R, Mintz GS, Haager PK et al (2002) Relation of stent design and stent surface material to subsequent in-stent intimal hyperplasia in coronary arteries determined by intravascular ultrasound. Am J Cardiol 89:1360–1364CrossRefPubMed Hoffmann R, Mintz GS, Haager PK et al (2002) Relation of stent design and stent surface material to subsequent in-stent intimal hyperplasia in coronary arteries determined by intravascular ultrasound. Am J Cardiol 89:1360–1364CrossRefPubMed
5.
Zurück zum Zitat Hart RG, Oczkowski WJ (2011) What’s new in stroke? The top 10 studies of 2009–2011: part II. Pol Arch Med Wewn 121:200–207PubMed Hart RG, Oczkowski WJ (2011) What’s new in stroke? The top 10 studies of 2009–2011: part II. Pol Arch Med Wewn 121:200–207PubMed
6.
Zurück zum Zitat Barras CD, Myers KA (2000) Nitinol—its use in vascular surgery and other applications. Eur J Vasc Endovasc Surg 19:564–569CrossRefPubMed Barras CD, Myers KA (2000) Nitinol—its use in vascular surgery and other applications. Eur J Vasc Endovasc Surg 19:564–569CrossRefPubMed
7.
Zurück zum Zitat Biamino G, Schmidt A, Scheinert D (2005) Treatment of SFA lesions with PLLA biodegradable stents: results of the PERSEUS study. J Endovasc Ther 12:5–5 Biamino G, Schmidt A, Scheinert D (2005) Treatment of SFA lesions with PLLA biodegradable stents: results of the PERSEUS study. J Endovasc Ther 12:5–5
8.
Zurück zum Zitat Tamai H, Igaki K, Kyo E et al (2000) Initial and 6-month results of biodegradable poly-l-lactic acid coronary stents in humans. Circulation 102:399–404CrossRefPubMed Tamai H, Igaki K, Kyo E et al (2000) Initial and 6-month results of biodegradable poly-l-lactic acid coronary stents in humans. Circulation 102:399–404CrossRefPubMed
9.
Zurück zum Zitat Serruys PW, Onuma Y, Dudek D et al (2011) Evaluation of the second generation of a bioresorbable everolimus-eluting vascular scaffold for the treatment of de novo coronary artery stenosis: 12-month clinical and imaging outcomes. J Am Coll Cardiol 58:1578–1588CrossRefPubMed Serruys PW, Onuma Y, Dudek D et al (2011) Evaluation of the second generation of a bioresorbable everolimus-eluting vascular scaffold for the treatment of de novo coronary artery stenosis: 12-month clinical and imaging outcomes. J Am Coll Cardiol 58:1578–1588CrossRefPubMed
10.
Zurück zum Zitat Anonymous (1991) Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. North American symptomatic carotid endarterectomy trial collaborators. N Engl J Med 325:445–453CrossRef Anonymous (1991) Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. North American symptomatic carotid endarterectomy trial collaborators. N Engl J Med 325:445–453CrossRef
11.
Zurück zum Zitat Brown BG, Hillger L, Zhao XQ et al (1995) Types of change in coronary stenosis severity and their relative importance in overall progression and regression of coronary disease. Observations from the FATS Trial. Familial Atherosclerosis Treatment Study. Ann N Y Acad Sci 748:407–417. (discussion 417–408)CrossRefPubMed Brown BG, Hillger L, Zhao XQ et al (1995) Types of change in coronary stenosis severity and their relative importance in overall progression and regression of coronary disease. Observations from the FATS Trial. Familial Atherosclerosis Treatment Study. Ann N Y Acad Sci 748:407–417. (discussion 417–408)CrossRefPubMed
12.
Zurück zum Zitat Schwartz GG, Olsson AG, Ezekowitz MD et al (2001) Effects of atorvastatin on early recurrent ischemic events in acute coronary syndromes: the MIRACL study: a randomized controlled trial. JAMA 285:1711–1718CrossRefPubMed Schwartz GG, Olsson AG, Ezekowitz MD et al (2001) Effects of atorvastatin on early recurrent ischemic events in acute coronary syndromes: the MIRACL study: a randomized controlled trial. JAMA 285:1711–1718CrossRefPubMed
13.
Zurück zum Zitat Miyauchi K, Kasai T, Yokayama T et al (2008) Effectiveness of statin-eluting stent on early inflammatory response and neointimal thickness in a porcine coronary model. Circ J 72:832–838CrossRefPubMed Miyauchi K, Kasai T, Yokayama T et al (2008) Effectiveness of statin-eluting stent on early inflammatory response and neointimal thickness in a porcine coronary model. Circ J 72:832–838CrossRefPubMed
14.
Zurück zum Zitat Zago AC, Matte BS, Reginato L et al (2012) First-in-man study of simvastatin-eluting stent in de novo coronary lesions: the SIMVASTENT study. Circ J 76:1109–1114CrossRefPubMed Zago AC, Matte BS, Reginato L et al (2012) First-in-man study of simvastatin-eluting stent in de novo coronary lesions: the SIMVASTENT study. Circ J 76:1109–1114CrossRefPubMed
15.
Zurück zum Zitat Furman MI, Frelinger IA, Michelson AD (2000) Antithrombotic therapy in the cardiac catheterization laboratory: focus on antiplatelet agents. Curr Cardiol Rep 2:386–394CrossRefPubMed Furman MI, Frelinger IA, Michelson AD (2000) Antithrombotic therapy in the cardiac catheterization laboratory: focus on antiplatelet agents. Curr Cardiol Rep 2:386–394CrossRefPubMed
16.
Zurück zum Zitat Tan JY, Shi WH, He J et al (2008) A clinical trial of using antiplatelet therapy to prevent restenosis following peripheral artery angioplasty and stenting. Zhonghua Yi Xue Za Zhi 88:812–815PubMed Tan JY, Shi WH, He J et al (2008) A clinical trial of using antiplatelet therapy to prevent restenosis following peripheral artery angioplasty and stenting. Zhonghua Yi Xue Za Zhi 88:812–815PubMed
17.
Zurück zum Zitat Bünger CM, Grabow N, Kroger C et al (2006) Iliac anastomotic stenting with a sirolimus-eluting biodegradable poly-L-lactide stent: a preliminary study after 6 weeks. J Endovasc Ther 13:630–639CrossRefPubMed Bünger CM, Grabow N, Kroger C et al (2006) Iliac anastomotic stenting with a sirolimus-eluting biodegradable poly-L-lactide stent: a preliminary study after 6 weeks. J Endovasc Ther 13:630–639CrossRefPubMed
18.
Zurück zum Zitat Bünger CM, Grabow N, Sternberg K et al (2007) Sirolimus-eluting biodegradable poly-L-lactide stent for peripheral vascular application: a preliminary study in porcine carotid arteries. J Surg Res 139:77–82CrossRefPubMed Bünger CM, Grabow N, Sternberg K et al (2007) Sirolimus-eluting biodegradable poly-L-lactide stent for peripheral vascular application: a preliminary study in porcine carotid arteries. J Surg Res 139:77–82CrossRefPubMed
19.
Zurück zum Zitat Grabow N, Schlun M, Sternberg K et al (2005) Mechanical properties of laser cut poly(L-lactide) micro-specimens: implications for stent design, manufacture, and sterilization. J Biomech Eng 127:25–31CrossRefPubMed Grabow N, Schlun M, Sternberg K et al (2005) Mechanical properties of laser cut poly(L-lactide) micro-specimens: implications for stent design, manufacture, and sterilization. J Biomech Eng 127:25–31CrossRefPubMed
20.
Zurück zum Zitat Kantor B, Ashai K, Holmes DR Jr et al (1999) The experimental animal models for assessing treatment of restenosis. Cardiovasc Radiat Med 1:48–54CrossRefPubMed Kantor B, Ashai K, Holmes DR Jr et al (1999) The experimental animal models for assessing treatment of restenosis. Cardiovasc Radiat Med 1:48–54CrossRefPubMed
Metadaten
Titel
Biodegradable polymeric stents for vascular application in a porcine carotid artery model
English version
verfasst von
Dr. S. Kischkel
N. Grabow
A. Püschel
B. Erdle
M. Kabelitz
D.P. Martin
S.F. Williams
I. Bombor
K. Sternberg
K.-P. Schmitz
W. Schareck
C.M. Bünger
Publikationsdatum
01.02.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Gefässchirurgie / Ausgabe Sonderheft 1/2016
Print ISSN: 0948-7034
Elektronische ISSN: 1434-3932
DOI
https://doi.org/10.1007/s00772-015-0011-z

Weitere Artikel der Sonderheft 1/2016

Gefässchirurgie 1/2016 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.