肝癌に対する呼吸停止下体幹部定位照射における横隔膜面を基準とした照合法の有用性
河原 大輔小澤 修一中島 健雄相田 雅道河合 信太朗越智 悠介奥村 拓朗増田 弘和大野 吉美木村 智樹永田 靖
著者情報
ジャーナル フリー

2014 年 70 巻 1 号 p. 51-56

詳細
抄録

Purpose: Liver image guided radiation therapy (IGRT) based on bone matching risks generating serious target positioning errors for reasons of lack of reproducibility of expiration breath hold. We therefore investigated the feasibility of 3D image matching between planning CT images and pretreatment cone-beam computed tomography (CBCT) images based on diaphragm surface matching. Method: 27 liver stereotactic body radiotherapy (SBRT) cases in whom trancecatheter arterial chemoembolization (TACE) had been performed in advance of radiotherapy were manually image-matched based on contrast, Lipiodol used in the TACE as the marker of the tumor, and the relative coordinates of the isocenter obtained by contrast matching, defined as the reference coordinate. The target positioning difference between diaphragm matching and bone matching were evaluated by using relative coordinates of the isocenter from the reference obtained for each matching technique. Results: The target positioning error using diaphragm matching and bone matching was 1.31±0.83 and 3.10±2.80 mm in the cranial-caudal (C-C) direction, 1.04±0.95 and 1.62±1.02 mm in the anterior-posterior (A-P) direction, and 0.93±1.19 and 1.12±0.94 mm in the left-right (L-R) direction, respectively. The positioning error due to diaphragm matching was significantly smaller than for bone matching in the C-C direction (p<0.05). Conclusion: IGRT based on diaphragm matching has potential as an alternative image matching technique for the positioning of liver patients.

著者関連情報
© 2014 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top