Roll and pitch set-up errors during volumetric modulated arc delivery

Can adapting gantry and collimator angles compensate?

verfasst von: Nienke A. Hoffmans-Holtzer, M.Sc., Daan Hoffmans, B.Eng., Max Dahele, MBChB, M.Sc., Ph.D. FRCR, FRCP, Ben J. Slotman, M.D., Ph.D., Wilko F.A.R. Verbakel, Ph.D.

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 3/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purpose of this work was to investigate whether adapting gantry and collimator angles can compensate for roll and pitch setup errors during volumetric modulated arc therapy (VMAT) delivery.

Methods

Previously delivered clinical plans for locally advanced head-and-neck (H&N) cancer (n = 5), localized prostate cancer (n = 2), and whole brain with simultaneous integrated boost to 5 metastases (WB + 5M, n = 1) were used for this study. Known rigid rotations were introduced in the planning CT scans. To compensate for these, in-house software was used to adapt gantry and collimator angles in the plan. Doses to planning target volumes (PTV) and critical organs at risk (OAR) were calculated with and without compensation and compared with the original clinical plan. Measurements in the sagittal plane in a polystyrene phantom using radiochromic film were compared by gamma (γ) evaluation for 2 H&N cancer patients.

Results

For H&N plans, the introduction of 2°-roll and 3°-pitch rotations reduced mean PTV coverage from 98.7 to 96.3 %. This improved to 98.1 % with gantry and collimator compensation. For prostate plans respective figures were 98.4, 97.5, and 98.4 %. For WB + 5M, compensation worked less well, especially for smaller volumes and volumes farther from the isocenter. Mean comparative γ evaluation (3 %, 1 mm) between original and pitched plans resulted in 86 % γ < 1. The corrected plan restored the mean comparison to 96 % γ < 1.

Conclusion

Preliminary data suggest that adapting gantry and collimator angles is a promising way to correct roll and pitch set-up errors of < 3° during VMAT for H&N and prostate cancer.
Literatur
1.
2.
Zurück zum Zitat Guckenberger M, Meyer J, Wilbert J et al (2007) Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice. Strahlenther Onkol 183:307–313CrossRefPubMed Guckenberger M, Meyer J, Wilbert J et al (2007) Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice. Strahlenther Onkol 183:307–313CrossRefPubMed
3.
Zurück zum Zitat Linthout N, Verellen D, Tournel K et al (2007) Assessment of secondary patient motion induced by automated couch movement during on-line 6 dimensional repositioning in prostate cancer treatment. Radiother Oncol 83:168–174CrossRefPubMed Linthout N, Verellen D, Tournel K et al (2007) Assessment of secondary patient motion induced by automated couch movement during on-line 6 dimensional repositioning in prostate cancer treatment. Radiother Oncol 83:168–174CrossRefPubMed
4.
Zurück zum Zitat Graf R, Boehmer D, Budach V, Wust P (2010) Residual translational and rotational errors after kV X-ray image-guided correction of prostate location using implanted fiducials. Strahlenther Onkol 186:544–550CrossRefPubMed Graf R, Boehmer D, Budach V, Wust P (2010) Residual translational and rotational errors after kV X-ray image-guided correction of prostate location using implanted fiducials. Strahlenther Onkol 186:544–550CrossRefPubMed
5.
Zurück zum Zitat Hüttenrauch P, Witt M, Wolff D et al (2014) Target volume coverage and dose to organs at risk in prostate cancer patients. Strahlenther Onkol 190:310–6CrossRefPubMed Hüttenrauch P, Witt M, Wolff D et al (2014) Target volume coverage and dose to organs at risk in prostate cancer patients. Strahlenther Onkol 190:310–6CrossRefPubMed
6.
Zurück zum Zitat Vineberg KA, Hunter KU, Schipper M et al (2012) The influence of rotation on delivered dose in oropharyngeal cancer. Int J Radiat Oncol Biol Phys 84:81CrossRef Vineberg KA, Hunter KU, Schipper M et al (2012) The influence of rotation on delivered dose in oropharyngeal cancer. Int J Radiat Oncol Biol Phys 84:81CrossRef
8.
Zurück zum Zitat Wu Q, Ivaldi G, Liang J et al (2006) Geometric and dosimetric evaluations of an online image-guidance strategy for 3D-CRT of prostate cancer. Int J Radiat Oncol Biol Phys 64:1596–1609CrossRefPubMed Wu Q, Ivaldi G, Liang J et al (2006) Geometric and dosimetric evaluations of an online image-guidance strategy for 3D-CRT of prostate cancer. Int J Radiat Oncol Biol Phys 64:1596–1609CrossRefPubMed
9.
Zurück zum Zitat Yue NJ, Knisely JPS, Song H et al (2006) A method to implement full six-degree target shift corrections for rigid body in image-guided radiotherapy. Med Phys 33:21CrossRefPubMed Yue NJ, Knisely JPS, Song H et al (2006) A method to implement full six-degree target shift corrections for rigid body in image-guided radiotherapy. Med Phys 33:21CrossRefPubMed
10.
Zurück zum Zitat Bose S, Shukla H, Maltz J (2010) Beam-centric algorithm for pretreatment patient position correction in external beam radiation therapy. Med Phys 37:2004CrossRefPubMed Bose S, Shukla H, Maltz J (2010) Beam-centric algorithm for pretreatment patient position correction in external beam radiation therapy. Med Phys 37:2004CrossRefPubMed
11.
Zurück zum Zitat Rijkhorst EJ, Van Herk M, Lebesque JV et al (2007) Strategy for online correction of rotational organ motion for intensity-modulated radiotherapy of prostate cancer. Int J Radiat Oncol Biol Phys 69:1608–1617CrossRefPubMed Rijkhorst EJ, Van Herk M, Lebesque JV et al (2007) Strategy for online correction of rotational organ motion for intensity-modulated radiotherapy of prostate cancer. Int J Radiat Oncol Biol Phys 69:1608–1617CrossRefPubMed
12.
Zurück zum Zitat Boswell SA, Jeraj R, Ruchala KJ et al (2005) A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy. Med Phys 32:1630CrossRefPubMed Boswell SA, Jeraj R, Ruchala KJ et al (2005) A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy. Med Phys 32:1630CrossRefPubMed
13.
Zurück zum Zitat Sarkar V, Wang B, Hinkle J et al (2012) Dosimetric evaluation of a “virtual” image guidance alternative to explicit 6 degree of freedom robotic couch correction. Pract Radiat Oncol 2:122–137CrossRefPubMed Sarkar V, Wang B, Hinkle J et al (2012) Dosimetric evaluation of a “virtual” image guidance alternative to explicit 6 degree of freedom robotic couch correction. Pract Radiat Oncol 2:122–137CrossRefPubMed
14.
Zurück zum Zitat Siddon R (1981) Solution to treatment planning problems using coordinate transformations. Med Phys 8:766–774CrossRefPubMed Siddon R (1981) Solution to treatment planning problems using coordinate transformations. Med Phys 8:766–774CrossRefPubMed
15.
Zurück zum Zitat Doornaert P, Verbakel WFAR, Bieker M et al (2011) RapidArc planning and delivery in patients with locally advanced head-and-neck cancer undergoing chemo radiotherapy. Int J Radiat Oncol Biol Phys 79:429–435CrossRefPubMed Doornaert P, Verbakel WFAR, Bieker M et al (2011) RapidArc planning and delivery in patients with locally advanced head-and-neck cancer undergoing chemo radiotherapy. Int J Radiat Oncol Biol Phys 79:429–435CrossRefPubMed
16.
Zurück zum Zitat Lagerwaard FJ, Van der Hoorn EAP, Verbakel WFAR et al (2009) Whole-brain radiotherapy with simultaneous integrated boost to multiple brain metastases using volumetric modulated arc therapy. Int J Radiat Oncol BiolPhys 75:253–259CrossRef Lagerwaard FJ, Van der Hoorn EAP, Verbakel WFAR et al (2009) Whole-brain radiotherapy with simultaneous integrated boost to multiple brain metastases using volumetric modulated arc therapy. Int J Radiat Oncol BiolPhys 75:253–259CrossRef
17.
Zurück zum Zitat Kaiser A, Schultheiss TE, Wong JYC et al (2006) Pitch, roll, and yaw variations inpatient positioning. Int J Radiat Oncol Biol Phys 66:949–955CrossRefPubMed Kaiser A, Schultheiss TE, Wong JYC et al (2006) Pitch, roll, and yaw variations inpatient positioning. Int J Radiat Oncol Biol Phys 66:949–955CrossRefPubMed
18.
Zurück zum Zitat Stieler F1, Wenz F, Scherrer D et al (2012) Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy. Strahlenther Onkol 188:1080–1084CrossRefPubMed Stieler F1, Wenz F, Scherrer D et al (2012) Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy. Strahlenther Onkol 188:1080–1084CrossRefPubMed
19.
Zurück zum Zitat Verbakel WFAR, Cuijpers JP, Hoffmans D et al (2009) Volumetric intensity modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys 74:252–259CrossRefPubMed Verbakel WFAR, Cuijpers JP, Hoffmans D et al (2009) Volumetric intensity modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys 74:252–259CrossRefPubMed
20.
Zurück zum Zitat Sterzing F, Schubert K, Sroka-Perez G et al (2008) Helical tomotherapy. Experiences of the first 150 patients in Heidelberg. Strahlenther Onkol 184:8–14CrossRefPubMed Sterzing F, Schubert K, Sroka-Perez G et al (2008) Helical tomotherapy. Experiences of the first 150 patients in Heidelberg. Strahlenther Onkol 184:8–14CrossRefPubMed
21.
Zurück zum Zitat Rijkhorst EJ, Lakeman A, Nijkamp J et al (2009) Strategies for online organ motion correction for intensity-modulated radiotherapy of prostate cancer: prostate, rectum, and bladder dose effects. Int J Radiat Oncol Biol Phys 75:1254–60CrossRefPubMed Rijkhorst EJ, Lakeman A, Nijkamp J et al (2009) Strategies for online organ motion correction for intensity-modulated radiotherapy of prostate cancer: prostate, rectum, and bladder dose effects. Int J Radiat Oncol Biol Phys 75:1254–60CrossRefPubMed
Metadaten
Titel
Roll and pitch set-up errors during volumetric modulated arc delivery
Can adapting gantry and collimator angles compensate?
verfasst von
Nienke A. Hoffmans-Holtzer, M.Sc.
Daan Hoffmans, B.Eng.
Max Dahele, MBChB, M.Sc., Ph.D. FRCR, FRCP
Ben J. Slotman, M.D., Ph.D.
Wilko F.A.R. Verbakel, Ph.D.
Publikationsdatum
01.03.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 3/2015
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-014-0766-2

Weitere Artikel der Ausgabe 3/2015

Strahlentherapie und Onkologie 3/2015 Zur Ausgabe

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.