Erschienen in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 1/2016

01.01.2016 | Original Article

Radiation dose of cone-beam computed tomography compared to conventional radiographs in orthodontics

verfasst von: Luca Signorelli, Raphael Patcas, Timo Peltomäki, Marc Schätzle

Erschienen in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie | Ausgabe 1/2016

Einloggen, um Zugang zu erhalten

Abstract

Objectives

The aim of this study was to determine radiation doses of different cone-beam computed tomography (CBCT) scan modes in comparison to a conventional set of orthodontic radiographs (COR) by means of phantom dosimetry.

Materials and methods

Thermoluminescent dosimeter (TLD) chips (3 × 1 × 1 mm) were used on an adult male tissue-equivalent phantom to record the distribution of the absorbed radiation dose. Three different scanning modes (i.e., portrait, normal landscape, and fast scan landscape) were compared to CORs [i.e., conventional lateral (LC) and posteroanterior (PA) cephalograms and digital panoramic radiograph (OPG)].

Results

The following radiation levels were measured: 131.7, 91, and 77 μSv in the portrait, normal landscape, and fast landscape modes, respectively. The overall effective dose for a COR was 35.81 μSv (PA: 8.90 μSv; OPG: 21.87 μSv; LC: 5.03 μSv).

Discussion

Although one CBCT scan may replace all CORs, one set of CORs still entails 2–4 times less radiation than one CBCT. Depending on the scan mode, the radiation dose of a CBCT is about 3–6 times an OPG, 8–14 times a PA, and 15–26 times a lateral LC. Finally, in order to fully reconstruct cephalograms including the cranial base and other important structures, the CBCT portrait mode must be chosen, rendering the difference in radiation exposure even clearer (131.7 vs. 35.81 μSv). Shielding radiation-sensitive organs can reduce the effective dose considerably.

Conclusion

CBCT should not be recommended for use in all orthodontic patients as a substitute for a conventional set of radiographs. In CBCT, reducing the height of the field of view and shielding the thyroid are advisable methods and must be implemented to lower the exposure dose.
Literatur
1.
Zurück zum Zitat Al Najjar A, Colosi D, Dauer LT et al (2013) Comparison of adult and child radiation equivalent doses from 2 dental cone-beam computed tomography units. Am J Orthod Dentofac Orthop 143:784–792CrossRef Al Najjar A, Colosi D, Dauer LT et al (2013) Comparison of adult and child radiation equivalent doses from 2 dental cone-beam computed tomography units. Am J Orthod Dentofac Orthop 143:784–792CrossRef
2.
Zurück zum Zitat Alqerban A, Jacobs R, Fieuws S et al (2011) Comparison of two cone beam computed tomographic systems versus panoramic imaging for localization of impacted maxillary canines and detection of root resorption. Eur J Orthod 33:93–102CrossRefPubMed Alqerban A, Jacobs R, Fieuws S et al (2011) Comparison of two cone beam computed tomographic systems versus panoramic imaging for localization of impacted maxillary canines and detection of root resorption. Eur J Orthod 33:93–102CrossRefPubMed
3.
Zurück zum Zitat American Association of Orthodontists (2010) Statement on the role of CBCT in orthodontics (26-10 H) American Association of Orthodontists (2010) Statement on the role of CBCT in orthodontics (26-10 H)
4.
Zurück zum Zitat Baumgaertel S, Palomo JM, Palomo L et al (2009) Reliability and accuracy of cone-beam computed tomography dental measurements. Am J Orthod Dentofac Orthop 136:19–25 discussion 25–8 CrossRef Baumgaertel S, Palomo JM, Palomo L et al (2009) Reliability and accuracy of cone-beam computed tomography dental measurements. Am J Orthod Dentofac Orthop 136:19–25 discussion 25–8 CrossRef
5.
Zurück zum Zitat Baumrind S, Frantz RC (1971) The reliability of head film measurements. 2. Conventional angular and linear measures. Am J Orthod 60:505–517CrossRefPubMed Baumrind S, Frantz RC (1971) The reliability of head film measurements. 2. Conventional angular and linear measures. Am J Orthod 60:505–517CrossRefPubMed
6.
Zurück zum Zitat Baumrind S, Frantz RC (1971) The reliability of head film measurements. 1. Landmark identification. Am J Orthod 60:111–127CrossRefPubMed Baumrind S, Frantz RC (1971) The reliability of head film measurements. 1. Landmark identification. Am J Orthod 60:111–127CrossRefPubMed
7.
Zurück zum Zitat Becker A, Chaushu S, Casap-Caspi N (2010) Cone-beam computed tomography and the orthosurgical management of impacted teeth. J Am Dent Assoc 141(Suppl 3):14S–18SCrossRefPubMed Becker A, Chaushu S, Casap-Caspi N (2010) Cone-beam computed tomography and the orthosurgical management of impacted teeth. J Am Dent Assoc 141(Suppl 3):14S–18SCrossRefPubMed
8.
Zurück zum Zitat Botticelli S, Verna C, Cattaneo PM et al (2011) Two- versus three-dimensional imaging in subjects with unerupted maxillary canines. Eur J Orthod 33:344–349CrossRefPubMed Botticelli S, Verna C, Cattaneo PM et al (2011) Two- versus three-dimensional imaging in subjects with unerupted maxillary canines. Eur J Orthod 33:344–349CrossRefPubMed
9.
Zurück zum Zitat Brodie AG (1949) Cephalometric roentgenology: History, technique and uses. J Oral Surg 7:185–198PubMed Brodie AG (1949) Cephalometric roentgenology: History, technique and uses. J Oral Surg 7:185–198PubMed
10.
Zurück zum Zitat Broadbent BH (1931) A new X-ray technique and its application to orthodontia. Angle Orthod 1:45–60 Broadbent BH (1931) A new X-ray technique and its application to orthodontia. Angle Orthod 1:45–60
12.
Zurück zum Zitat Cohen MD (2009) Pediatric CT radiation dose: how low can you go? AJR Am J Roentgenol 192:1292–1303CrossRefPubMed Cohen MD (2009) Pediatric CT radiation dose: how low can you go? AJR Am J Roentgenol 192:1292–1303CrossRefPubMed
13.
Zurück zum Zitat Damstra J, Fourie Z, Ren Y (2013) Evaluation and comparison of postero-anterior cephalograms and cone-beam computed tomography images for the detection of mandibular asymmetry. Eur J Orthod 35:45–50CrossRefPubMed Damstra J, Fourie Z, Ren Y (2013) Evaluation and comparison of postero-anterior cephalograms and cone-beam computed tomography images for the detection of mandibular asymmetry. Eur J Orthod 35:45–50CrossRefPubMed
14.
Zurück zum Zitat Farman AG, Scarfe WC (2009) The basics of maxillofacial cone beam computed tomography. Semin Orthod 15:2–13CrossRef Farman AG, Scarfe WC (2009) The basics of maxillofacial cone beam computed tomography. Semin Orthod 15:2–13CrossRef
15.
Zurück zum Zitat Grunheid T, Kolbeck Schieck JR, Pliska BT et al (2012) Dosimetry of a cone-beam computed tomography machine compared with a digital X-ray machine in orthodontic imaging. Am J Orthod Dentofac Orthop 141:436–443CrossRef Grunheid T, Kolbeck Schieck JR, Pliska BT et al (2012) Dosimetry of a cone-beam computed tomography machine compared with a digital X-ray machine in orthodontic imaging. Am J Orthod Dentofac Orthop 141:436–443CrossRef
16.
Zurück zum Zitat Halazonetis DJ (2012) Cone-beam computed tomography is not the imaging technique of choice for comprehensive orthodontic assessment. Am J Orthod Dentofac Orthop 141:403, 405, 407 passim Halazonetis DJ (2012) Cone-beam computed tomography is not the imaging technique of choice for comprehensive orthodontic assessment. Am J Orthod Dentofac Orthop 141:403, 405, 407 passim
17.
Zurück zum Zitat Haney E, Gansky SA, Lee JS et al (2010) Comparative analysis of traditional radiographs and cone-beam computed tomography volumetric images in the diagnosis and treatment planning of maxillary impacted canines. Am J Orthod Dentofac Orthop 137:590–597CrossRef Haney E, Gansky SA, Lee JS et al (2010) Comparative analysis of traditional radiographs and cone-beam computed tomography volumetric images in the diagnosis and treatment planning of maxillary impacted canines. Am J Orthod Dentofac Orthop 137:590–597CrossRef
18.
Zurück zum Zitat Hofrath H (1931) Die Bdeutung der Roentgenfern und Abstandsaufnahme für die Diagnostik der Kieferanomalien. J Orofac Orthop 1:232–248CrossRef Hofrath H (1931) Die Bdeutung der Roentgenfern und Abstandsaufnahme für die Diagnostik der Kieferanomalien. J Orofac Orthop 1:232–248CrossRef
19.
Zurück zum Zitat Isaacson KG, Thom AR, Horner K et al (2008) Orthodontic radiographs—guidelines for the use of radiographs in clinical orthodontics, 3rd edn. British Orthodontic Society, London Isaacson KG, Thom AR, Horner K et al (2008) Orthodontic radiographs—guidelines for the use of radiographs in clinical orthodontics, 3rd edn. British Orthodontic Society, London
20.
Zurück zum Zitat Kaeppler G (2010) Applications of cone beam computed tomography in dental and oral medicine. Int J Comput Dent 13:203–219PubMed Kaeppler G (2010) Applications of cone beam computed tomography in dental and oral medicine. Int J Comput Dent 13:203–219PubMed
21.
Zurück zum Zitat Kokich VG (2010) Cone-beam computed tomography: have we identified the orthodontic benefits? Am J Orthod Dentofac Orthop 137:S16CrossRef Kokich VG (2010) Cone-beam computed tomography: have we identified the orthodontic benefits? Am J Orthod Dentofac Orthop 137:S16CrossRef
22.
Zurück zum Zitat Lamichane M, Anderson NK, Rigali PH et al (2009) Accuracy of reconstructed images from cone-beam computed tomography scans. Am J Orthod Dentofac Orthop 136:156e1–6; (discussion 156–7) Lamichane M, Anderson NK, Rigali PH et al (2009) Accuracy of reconstructed images from cone-beam computed tomography scans. Am J Orthod Dentofac Orthop 136:156e1–6; (discussion 156–7)
23.
Zurück zum Zitat Larson BE (2012) Cone-beam computed tomography is the imaging technique of choice for comprehensive orthodontic assessment. Am J Orthod Dentofac Orthop Off Publ Am Assoc Orthod Const Soc Am Board Orthod 141:402, 404, 406 passim Larson BE (2012) Cone-beam computed tomography is the imaging technique of choice for comprehensive orthodontic assessment. Am J Orthod Dentofac Orthop Off Publ Am Assoc Orthod Const Soc Am Board Orthod 141:402, 404, 406 passim
24.
Zurück zum Zitat Lorenzoni DC, Bolognese AM, Garib DG et al (2012) Cone-beam computed tomography and radiographs in dentistry: aspects related to radiation dose. Int J Dent 2012:813768PubMedCentralCrossRefPubMed Lorenzoni DC, Bolognese AM, Garib DG et al (2012) Cone-beam computed tomography and radiographs in dentistry: aspects related to radiation dose. Int J Dent 2012:813768PubMedCentralCrossRefPubMed
25.
Zurück zum Zitat Ludlow JB, Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:106–114CrossRefPubMed Ludlow JB, Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:106–114CrossRefPubMed
26.
Zurück zum Zitat Ludlow JB (2011) A manufacturer’s role in reducing the dose of cone beam computed tomography examinations: effect of beam filtration. Dentomaxillofac Radiol 40:115–122PubMedCentralCrossRefPubMed Ludlow JB (2011) A manufacturer’s role in reducing the dose of cone beam computed tomography examinations: effect of beam filtration. Dentomaxillofac Radiol 40:115–122PubMedCentralCrossRefPubMed
27.
Zurück zum Zitat Mah J, Hatcher DC (2005) Craniofacial imaging in orthodontics. Orthodontics: current principles and techniques. In: Graber TM, Vanarsdall RL, Vig KWL. Elsevier, St. Louis, pp 71–100 Mah J, Hatcher DC (2005) Craniofacial imaging in orthodontics. Orthodontics: current principles and techniques. In: Graber TM, Vanarsdall RL, Vig KWL. Elsevier, St. Louis, pp 71–100
28.
Zurück zum Zitat Misch KA, Yi ES, Sarment DP (2006) Accuracy of cone beam computed tomography for periodontal defect measurements. J Periodontol 77:1261–1266CrossRefPubMed Misch KA, Yi ES, Sarment DP (2006) Accuracy of cone beam computed tomography for periodontal defect measurements. J Periodontol 77:1261–1266CrossRefPubMed
29.
Zurück zum Zitat Patcas R, Muller L, Ullrich O et al (2012) Accuracy of cone-beam computed tomography at different resolutions assessed on the bony covering of the mandibular anterior teeth. Am J Orthod Dentofac Orthop 141:41–50CrossRef Patcas R, Muller L, Ullrich O et al (2012) Accuracy of cone-beam computed tomography at different resolutions assessed on the bony covering of the mandibular anterior teeth. Am J Orthod Dentofac Orthop 141:41–50CrossRef
30.
Zurück zum Zitat Patcas R, Signorelli L, Peltomaki T et al (2012) Is the use of the cervical vertebrae maturation method justified to determine skeletal age? A comparison of radiation dose of two strategies for skeletal age estimation. Eur J Orthod 35:604–609CrossRefPubMed Patcas R, Signorelli L, Peltomaki T et al (2012) Is the use of the cervical vertebrae maturation method justified to determine skeletal age? A comparison of radiation dose of two strategies for skeletal age estimation. Eur J Orthod 35:604–609CrossRefPubMed
31.
Zurück zum Zitat Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499–505PubMedCentralCrossRefPubMed Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499–505PubMedCentralCrossRefPubMed
32.
Zurück zum Zitat Rickets RM (1981) The golden divider. J Clin Orthod 15:725–759 Rickets RM (1981) The golden divider. J Clin Orthod 15:725–759
33.
Zurück zum Zitat Roberts JA, Drage NA, Davies J et al (2009) Effective dose from cone beam CT examinations in dentistry. Br J Radiol 82:35–40CrossRefPubMed Roberts JA, Drage NA, Davies J et al (2009) Effective dose from cone beam CT examinations in dentistry. Br J Radiol 82:35–40CrossRefPubMed
34.
Zurück zum Zitat Silva MA, Wolf U, Heinicke F et al (2008) Cone-beam computed tomography for routine orthodontic treatment planning: a radiation dose evaluation. Am J Orthod Dentofac Orthop 133(640):e1–e5 Silva MA, Wolf U, Heinicke F et al (2008) Cone-beam computed tomography for routine orthodontic treatment planning: a radiation dose evaluation. Am J Orthod Dentofac Orthop 133(640):e1–e5
35.
Zurück zum Zitat Smith BR, Park JH, Cederberg RA (2011) An evaluation of cone-beam computed tomography use in postgraduate orthodontic programs in the United States and Canada. J Dent Educ 75:98–106PubMed Smith BR, Park JH, Cederberg RA (2011) An evaluation of cone-beam computed tomography use in postgraduate orthodontic programs in the United States and Canada. J Dent Educ 75:98–106PubMed
36.
Zurück zum Zitat Timock AM, Cook V, McDonald T et al (2011) Accuracy and reliability of buccal bone height and thickness measurements from cone-beam computed tomography imaging. Am J Orthod Dentofac Orthop 140:734–744CrossRef Timock AM, Cook V, McDonald T et al (2011) Accuracy and reliability of buccal bone height and thickness measurements from cone-beam computed tomography imaging. Am J Orthod Dentofac Orthop 140:734–744CrossRef
37.
Zurück zum Zitat Valentin J (2007) Managing patient dose in multi-detector computed tomography (MDCT). ICRP Publication 102. Ann ICRP 37:1–79; (iii) Valentin J (2007) Managing patient dose in multi-detector computed tomography (MDCT). ICRP Publication 102. Ann ICRP 37:1–79; (iii)
Metadaten
Titel
Radiation dose of cone-beam computed tomography compared to conventional radiographs in orthodontics
verfasst von
Luca Signorelli
Raphael Patcas
Timo Peltomäki
Marc Schätzle
Publikationsdatum
01.01.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie / Ausgabe 1/2016
Print ISSN: 1434-5293
Elektronische ISSN: 1615-6714
DOI
https://doi.org/10.1007/s00056-015-0002-4

Weitere Artikel der Ausgabe 1/2016

Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 1/2016 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.