Open Access 08.03.2018 | Knee

3D patient imaging and retrieval analysis help understand the clinical importance of rotation in knee replacements

verfasst von: Arianna Cerquiglini, Johann Henckel, Harry Hothi, Niccoló Rotigliano, Michael T. Hirschmann, Alister J. Hart

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 11/2018

Abstract

Purpose

The purpose of the present study was to correlate highly accurate CT measurements of pre-revision total knee arthroplasty (TKA) implant position with findings of retrieval analysis post-revision, to understand the clinical relevance of TKA orientation.

Methods

This study involved 53 retrieved TKA implants with pre-revision 3D-CT scans used to determine coronal (varus–valgus), sagittal (tibial slope) and rotational (internal rotation–external rotation) TKA orientation as well as tibiofemoral leg axis. Differences between femoral and tibial angles to describe the "relative rotational mismatch" were also calculated. All tibial inserts were forensically analyzed using the Hood score. Statistical analysis was performed to investigate correlations between TKA component orientation and surface damage (p < 0.05).

Results

Femoral components were found to have axial rotations mainly within ± 3° (68%), whilst 45% of the tibial components and 66% of the relative rotational mismatches were > 3° and < − 3°, respectively. The majority of femoral and tibial components (87% in both cases), as well as the femorotibial angle (70%), showed coronal orientations within ± 3°. The 64% of the tibial components showed posterior tibial slopes out of both the 0°–3° and 5°–7° ranges. There was a significant correlation between tibial slope and damage score on polyethylene tibial inserts (r = 0.2856; p = 0.0382) as well as a significant correlation between implants’ position in the axial plane and damage score on polyethylene tibial inserts (r = 0.6537, p = 0.0240).

Conclusions

This is the first study to use accurate measurements from pre-revision 3DCT to compare tibial and femoral orientation in all three planes with retrieval findings in total knee replacements. A significant correlation between implant position and polyethylene surface damage was found. These results showed the importance of optimizing component position to minimize polyethylene damage. Further analysis involving more accurate polyethylene wear measurements are fundamental to fully understand the role of components’ orientation in TKAs.

Introduction

Total knee arthroplasty (TKA) is a common intervention with an excellent survival rate; however, up to 20% of patients have reported poor outcomes, leading to revision [1]. This phenomenon has global dimensions: the demand for revision of TKAs is projected to grow by 601% between 2005 and 2030 in the United States [2] and the same trend is also expected in Europe and Asia Pacific [3, 4].
The reasons for TKA failure are multifactorial and influenced by surgical, implant and patient factors. The most commonly reported reasons for revision are aseptic loosening, pain, infection, instability, stiffness, polyethylene wear, malposition, patellofemoral problems and dislocation/subluxation [1, 59].
A suboptimal TKA position of femoral and tibial components contributes to poor outcomes, premature polyethylene wear and “unexplained” painful TKA [7, 8, 1015]. It is well established that excessive internal rotation of the femoral component in the axial plane leads to patellar maltracking, anterior knee pain and flexion instability [7, 1012]. Some studies also found a significant correlation between malposition in the coronal plane and aseptic loosening of the implant, due to a higher amount of polyethylene wear caused by abnormal force distributions [1618]. Suboptimal flexion was found in cruciate retaining TKA with reduced posterior tibial slope [16].
Findings from retrieval studies of failed TKA implants provide valuable information on the location and potential mechanisms for TKA component damage in vivo [17, 19]. Several papers have suggested that the wear pattern is associated with different clinical and mechanical factors, such as component position, orientation and alignment [1315, 18, 2025]. For example, it has been shown that there is often more medial wear on tibial polyethylene bearings in well-positioned knees [7], and changes in orientation are known to influence this (increased medial wear in varus knees and greater lateral in valgus knees) [18, 25]. Malrotation might cause abnormal stresses and premature wear of the polyethylene components, followed by peri-prosthetic and implant loosening [7, 8, 10, 11, 18, 19]. However, the relationship between implant orientation and wear patterns in retrieved knee prostheses is still poorly understood.
The aim of this study was, for the first time, to correlate highly accurate 3D-CT measurements of pre-revision TKA position, provided by an innovative 3D imaging technique [8], with retrieval analysis findings post-revision to better understand implant orientation effects on TKA.

Materials and methods

This was a retrieval study involving 53 contemporary TKAs that had been consecutively revised at a single institution. The implants were revised from 30 female and 23 male patients with a median (range) age of 62 (42–78) years and a median time to revision of 38 (5–162) months. The reasons for revision were instability (n = 26), malposition (n = 11), patella maltracking (n = 7), aseptic loosening (n = 3), pain (n = 2), stiffness (n = 2), infection (n = 1) and arthrofibrosis (n = 1). The retrieved TKAs consisted of 40 cruciate retaining (CR) and 13 posterior stabilized (PS) TKA (Fig. 1). Table 1 summarizes the TKA specifications and patient demographics for each case.
Table 1
Implant and patient demographic
Case number
Design
Manufacturer
Patient age
Gender
Time to revision [months]
Reason for revision
1
NexGen LCCK, PS
Zimmer
69
M
29
Infection
2
Innex, CR
Zimmer
65
F
71
Instability
3
PFC Sigma, CR
DePuy
69
M
15
Aseptic loosening
4
PFC Sigma, CR
DePuy
66
M
53
Malposition
5
Triathlon, PS
Stryker
53
F
79
Instability
6
Natural Knee II, CR
Zimmer
69
M
36
Instability
7
Triathlon, PS
Stryker
75
M
99
Aseptic loosening
8
Triathlon, CR
Stryker
66
M
72
Instability
9
BalanSys, CR
Mathys
61
F
40
Patella maltracking
10
Natural Knee II, CR
Zimmer
71
F
90
Patella maltracking
11
PFC Sigma, CR
DePuy
46
F
20
Pain
12
TC PLUS, CR
Smith&Nephew
62
M
10
Instability
13
Type LCS, CR
DePuy
46
M
120
Instability
14
ATTUNE, PS
DePuy
68
F
15
Instability
15
Signature, PS
Zimmer
64
M
10
Malposition
16
NexGen, PS
Zimmer
51
F
6
Pain
17
Type LCS, CR
DePuy
51
F
22
Instability
18
Synthes-LCS, CR
DePuy
71
F
24
Patella maltracking
19
NexGen, PS
Zimmer
48
M
36
Instability
20
Synthes-LCS, CR
DePuy
64
F
162
Instability
21
Persona, CR
Zimmer
57
M
60
Patella maltracking
22
Type LCS, CR
DePuy
70
F
12
Instability
23
Journey, CR
Smith&Nephew
48
M
21
Instability
24
PFC Sigma, PS
DePuy
53
F
45
Malposition
25
BalanSys, CR
Mathys
62
F
18
Instability
26
ATTUNE, PS
DePuy
78
F
5
Instability
27
Triathlon, CR
Stryker
60
M
91
Instability
28
BalanSys, CR
Mathys
56
M
25
Instability
29
BalanSys, CR
Mathys
53
F
22
Instability
30
PFC Sigma, CR
DePuy
62
F
60
Instability
31
Innex, CR
Zimmer
66
F
13
Malposition
32
TC PLUS, CR
Smith&Nephew
59
M
101
Stiffness
33
BalanSys, CR
Mathys
42
F
51
Malposition
34
Gemini, PS
Link
72
F
58
Malposition
35
PFC Sigma, CR
DePuy
73
F
37
Malposition
36
PFC Sigma, PS
DePuy
72
F
19
Stiffness
37
PFC Sigma, PS
DePuy
63
F
61
Aseptic loosening
38
Triathlon, CR
Stryker
49
F
17
Instability
39
BalanSys, CR
Mathys
66
M
77
Instability
40
Colombus, CR
Aesculap
59
F
60
Patella maltracking
41
BalanSys, CR
Mathys
46
M
46
Malposition
42
PFC Sigma, CR
DePuy
70
M
64
Instability
43
ATTUNE, CR
DePuy
67
F
13
Patella maltracking
44
ATTUNE, CR
DePuy
64
F
22
Instability
45
Synthes-LCS, CR
DePuy
62
F
24
Instability
46
BalanSys, CR
Mathys
68
M
156
Malposition
47
Natural Knee II, CR
Zimmer
59
M
15
Malposition
48
BalanSys, PS
Mathys
55
F
100
Instability
49
Natural Knee II, CR
Zimmer
74
M
38
Malposition
50
Natural Knee II, CR
Zimmer
56
M
79
Malposition
51
PFC Sigma, CR
DePuy
60
M
101
Instability
52
PFC Sigma, CR
DePuy
61
F
10
Instability
53
Triathlon, CR
Stryker
62
F
66
Patella maltracking
All investigations were conducted in conformity with ethical principles of research, that informed consent for participation in the study was obtained and that institutional approval of the human protocol for this investigation was obtained.

3D computed tomography (CT) position

Pre-revision CT scans of both the femoral and tibial components from each patient were taken using the imperial CT protocol, which obtains the relevant bony landmarks (hip–knee–ankle) and includes specific metal artefact reduction sequences [8, 26]. From the CT images, the reference axes (anatomical and mechanical) were defined, the images standardized and the angles computed, to provide the following positioning measurements: (1) axial rotation in the transverse plane of both the femoral and tibial components, with reference to the transepicondylar and anteroposterior axes, respectively, (2) the degree of varus/valgus of single components and the tibiofemoral angle in the coronal plane and (3) the tibial component slope in the sagittal plane, Fig. 2.
Differences between femoral and tibial axial rotation angles to describe the relative rotational mismatch [27] between these two components in the transverse plane were also computed; the measured angles were used to estimate the magnitude of this mismatch, while the sign explained the relative position between the femoral and the tibial components (negative values meant that the femoral component was internally rotated compared to the tibial one; positive values meant that the femoral component was externally rotated compared to the tibial one).
The measured values were classified into three ranges (Fig. 3), in agreement with surgical standard aims [28, 29].

Explant preparation

All implants were decontaminated using 10% formalin solution. The polyethylene components were then stored in a freezer at − 18 °C, to minimise the oxidation process.

Analysis of polyethylene surfaces

The tibial polyethylene inserts (n = 53) were forensically analyzed using a Keyence VHX-700F series (Keyence Co., Japan) digital microscope, with magnification from 50× up to 200×. On the articular surface, both the medial and lateral sides were divided into four sections, while the central part of the insert into two sections (Fig. 4). Each of the 10 total sections were analyzed using the Hood score, according to the presence and severity of seven modes of surface degradation (surface deformation, pitting, embedded debris, scratching, burnishing, abrasion and delamination) [30], Table 2. The maximum damage grade was 21 for a single section (grade 3 for each of the seven damage modes) and 210 for the entire surface (grade 3 for each of the seven damage modes for each of the 10 sections). Grading was performed by two different examiners. In case of disagreement, the examiners discussed the results together, to agree a final grade.
Table 2
Damage-grading criteria used to assess polyethylene component surface, according to the Hood score [30]
 
Polyethylene surfaces
Modes of surface degradation
Deformation, pitting, embedded debris, scratching, burnishing, abrasion, delamination
Score
 0
Not visible
 1
< 10% of surface
 2
10–50% of surface
 3
> 50% of surface
Maximum section score
21
Maximum total score
210

Statistical analysis

Statistical analysis was performed to determine if there were any significant correlations between orientation of components and expected location of the surface damage, described in the literature [18, 20, 24, 25]. Potential associations between (1) internal relative rotational mismatch and the amount of damage on the posterior compartment in the medial side and on the anterior compartment in the lateral side, Fig. 5a; (2) external relative rotational mismatch and the amount of damage on the posterior compartment in the lateral side and on the anterior compartment in the medial side, Fig. 5b; (3) varus tibiofemoral angle and the amount of damage on the medial compartment, Fig. 5c; (4) valgus tibiofemoral angle and the amount of damage on the lateral compartment, Fig. 5d; (5) posterior slope and amount of damage on the posterior compartments, Fig. 5e were analyzed (two-tailed non-parametric Spearman correlation, p value < 0.05 was considered as significant).
Potential significant differences in the total Hood score between optimal and suboptimal orientation were also investigated (Mann–Whitney test, p value < 0.05 was considered as significant).
Separate statistical analyses on implants revised for malposition were performed and significant differences between implants with short (≤ 2 years) and long (> 2 years) implantation time were also investigated (Mann–Whitney test, p value < 0.05 was considered as significant). All statistical analyses were performed using Prism 7 (GraphPad, USA).

Results

3D-CT Position, axial plane

The 3D-CT imaging revealed that 68% of the femoral components had a rotation within ± 3°, whilst the rest showed rotations smaller than − 3°. 43% of the tibial components showed orientations ± 3°, whilst 45% and 12% had axial rotations greater than 3° and smaller than − 3°, respectively.
It was found that in 26% of cases, implants showed relative rotational mismatch within ± 3°. 66% showed internal rotations smaller than − 3° and 8% showed external rotations greater than 3°. Table 3 summarizes all the results, showing median and range values.
Table 3
Number of cases, median and range values of orientations of the femoral and tibial components and relative rotational mismatch in the axial plane
 
Axial orientation
Number of cases
Median (range) [°]
Femoral component
± 3°
36
− 2 (− 3 to 3)
> 3°
< − 3°
17
− 6 (− 11 to − 4)
Tibial component
± 3°
23
1 (− 3 to 3)
> 3°
24
8 (4 to 19)
< − 3°
6
− 9.5 (− 12 to − 6)
Relative rotational mismatch
± 3°
14
− 1 (− 3 to 3)
> 3°
4
10 (7 to 12)
< − 3°
35
− 9 (− 22 to − 4)

3D-CT Position, coronal plane

87% of the femoral components had orientations within ± 3°, whilst 11% had angulations smaller than − 3° and the 2% showed angulation greater than 3°.
The majority of tibial components showed angulations within ± 3°, whilst 11% and 2% had angulations smaller than − 3° and greater than 3°, respectively.
Considering the tibiofemoral angle, 70% of cases showed orientations within ± 3°, whilst 21% and 9% had angulations smaller than − 3° and greater than 3°, respectively.
Table 4 summarizes all the results, showing median and range values.
Table 4
Number of cases, median and range values of orientations of the femoral and tibial components and femorotibial angle in the coronal plane
 
Coronal orientation
Number of cases
Median (range) [°]
Femoral component
± 3°
46
0 (− 3 to 3)
> 3°
1
4 (4)
<− 3°
6
− 4 (− 7 to − 4)
Tibial component
± 3°
46
− 1 (− 3 to 3)
> 3°
1
5 (5)
< − 3°
6
− 4 (− 5 to − 4)
Femorotibial angle
± 3°
37
− 1 (− 3 to 3)
> 3°
5
5 (4 to 6)
< − 3°
11
− 4 (− 14 to − 4)

3D-CT Position, Sagittal plane

64% of the tibial components showed posterior tibial slopes that were not within 0° to 3° or 5° to 7. 36% had positions of tibial trays within 0° to 3° or 5° to 7°.
Table 5 summarizes all the results, showing median and range values.
Table 5
Number of cases, median and range values of orientations of the tibial components in the sagittal plane: negative values mean posterior tibial slope, whilst positive values mean anterior tibial slope
 
Sagittal orientation
Number of cases
Median (range) [°]
Posterior Staibilized design
− 3° to 0°
7
− 1 (− 3 to 0)
Out of − 3° to 0°
6
− 4.5 (− 9 to − 2, 4)
Cruciate Retaining design
− 7° to − 5°
12
− 5 (− 7 to − 5)
Out of − 7° to − 5°
28
− 3 (− 12 to − 8, − 4 to 2)

Analysis of polyethylene surfaces

The most common modes of surface degradation were scratching, pitting and burnishing, whilst deformation and delamination were rare (Fig. 6). The median total Hood score (range) was 39 (17–72). There was no significant difference (p = 0.5459) between lateral and medial compartments (defined by the sum of 0–3 and 4–7 sections, respectively, in a right tibial insert; viceversa in a left tibial insert), that showed mean (range) damage scores of 19 (7–39) and 20 (8–41), respectively.

Correlations

There was a significant correlation between tibial slope and damage score on polyethylene tibial inserts: posterior tibial slope was associated with higher Hood scores on the anterior areas of the polyethylene, Fig. 7. Table 6 shows further data.
Table 6
Correlation between implant orientation and Hood score in all the anatomical planes and differences in Hood scores between normal and suboptimal orientation
 
Correlation between orientation and Hood score in expected areas?
Significant difference in the Hood score between normal and suboptimal orientation?
Axial plane
Axial mismatch
n.s.
r = 0.02978; p = 0.8826
n.s.
p = 0.3042
Coronal plane
Tibiofemoral angle
n.s.
r = − 0.1168; p = 0.4051
n.s.
p = 0.0964
Sagittal plane
Tibial slope
Significant
r = − 0.2856; p = 0.0382
PS → n.s. p = 0.1678
CR → n.s., p = 0.0851
n.s. not significant
Separate analyses of the group revised for malposition revealed a significant correlation between implants’ position in the axial plane and damage score on polyethylene tibial inserts, Fig. 8: internal rotation mismatch was associated with higher Hood scores on the posterior compartment in the medial side and on the anterior compartment in the lateral side, while external ones were associated with higher Hood scores on the posterior compartment in the lateral side and on the anterior compartment in the medial side, Table 7.
Table 7
Correlation between implant orientation and Hood score in all the anatomical planes in implants revised for malposition
 
Correlation between orientation and Hood score in expected areas?
Relative rotational mismatch
Significant
r = 0.6537, p = 0.0240
Femorotibial angle
n.s.
r = − 0.02154, p = 0.8714
Posterior tibial slope
n.s.
r = − 0.1485, p = 0.5814
n.s. not significant
There was a significant difference in total Hood score between implants with short (≤ 2 years) and long (> 2 years) implantation time.
There was a significant correlation between time to revision and total Hood score, Fig. 9.

Discussion

The most important finding of the present study was a significant correlation between relative rotational mismatch and the severity and location of polyethylene damage in implants revised for malposition. There was also a significant correlation between the tibial slope in the sagittal plane, and amount and location of surface damage, but no significant correlation was found between implant orientation and surface damage in polyethylene tibial inserts in both the axial and coronal planes. Moreover, a significant correlation was found between time to revision and hood score. These findings support a multifactorial aetiology of polyethylene wear: implant position alone cannot explain all complex mechanisms involved in the generation of surface damage in polyethylene, especially in early revised implants. This is the first retrieval study to use 3D-CT pre-revision images to measure implant position and correlate findings with results from polyethylene surface damage inspection.
3D-CT-measured implant orientation revealed that the majority of both femoral and tibial components showed optimal orientations in the axial plane, with few cases of excessive internal and external rotations of femoral and tibial components, respectively. Calculations of the relative rotational mismatch revealed that the majority were internally rotated. Regarding the coronal plane, both the femoral and tibial showed mainly optimal orientation, as well as the tibiofemoral angle. In the sagittal plane, almost the totality of the tibial trays showed suboptimal posterior slope.
The current study was the first to investigate the relationship between wear patterns in retrieved knee replacements and 3D-CT-based computed measurements in all the three anatomical planes. In particular, the orientation in the axial plane is difficult to measure with traditional imaging methods [8]; other research groups have identified position in this orientation using roentgenographic score systems [23], X-ray [18, 24] or planar CT images [12, 27, 31, 32]. Axial rotation was measured with accuracy and repeatability using an innovative 3D-CT imaging technique with a standardized, published method and correlates these measurements with results from retrievals. Interestingly, investigating implants revised only for malposition, 9 out of 12 samples showed severe cases of malposition in the axial plane: internal rotational mismatches had a median value (range) of 10 (1–22) degrees, while external rotational mismatches had a median value (range) of 9 (1–12) degrees. In these cases, a significant correlation with the Hood score was found; this could suggest that only severe malposition can lead to asymmetry in the surface damage of polyethylene tibial inserts.
The main characteristic degradation features found on polyethylene inserts were scratching, pitting and burnishing, with no significant difference between medial and lateral sides. This symmetric wear pattern was in contrast to the predominantly asymmetric patterns found in previous studies [18, 20, 2224]. This discrepancy might be caused by the subjective nature of the widely used Hood score. Indeed, surface damage was found to be only a moderate predictor of wear in polyethylene [33]. Moreover, our components were revised relatively early with median time to revision of 37 months (5–162). The main reasons for revision were instability, malposition and patella maltracking. It is most likely that all patients had poor clinical function and, therefore, low levels of use. This would imply a lower gait cycle number (steps or joint cycles of use per year) and well below the average of 0.9–1.4 million gait cycles per annum. This relative lack of use may also explain why wear patterns did not correlate with components’ positions in all the anatomical planes. Furthermore, the multifactorial nature of the wear pattern mechanisms has to be taken into consideration: this includes a combination of surgical, implant and patient variables. For example, the precision in determining the rotational alignment during a total knee replacement procedure is crucial; correct component positioning can prevent post-cam impingement and reduce the polyethylene wear [32]. Knee replacement design is an important factor that can influence wear [10, 11, 18, 2023 27, 18]. It has been shown that a new posterior stabilized design can lead to a reduction in polyethylene surface damage and peg deformation [34]. It has also previously been reported that more active patients show greater wear in polyethylene inserts [35].
A considerable number of limitations have to be considered. First, the evaluation of wear pattern was performed using visual scoring for plastic components. Although previous papers [6, 8, 10, 23, 24] used these methods and demonstrated that they can give an acceptable estimation about the quality of the damage, they may be not accurate enough to measure the amount of wear [33]. Recently, the utility of alternative techniques, such as a coordinate measure machine (CMM), laser and micro-CT scanning [24, 25], was demonstrated. Further analyses involving these types of tool may lead to new findings and results, which in combination with 3D CT might allow us to fully understand the role of components’ orientation in total knee replacements and to define a complete “safe zone” for TKA. Future studies should recruit greater number of patients and include a greater number of implants of a single design.
These results showed the importance of optimizing component position to minimize polyethylene damage.

Conclusion

This is the first study to use accurate measurements from pre-revision 3D-CT to compare tibial and femoral orientation in all three planes with retrieval findings in 53 total knee replacements. Our results revealed a significant correlation between implant axial alignment, and the severity and location of polyethylene damage in implants revised for malposition as well as between tibial slope and damage score on polyethylene tibial inserts.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

All investigations were conducted in conformity with ethical principles of research and that institutional approval of the human protocol for this investigation was obtained.
Informed consent for participation in the study was obtained.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
2.
Zurück zum Zitat Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89(4):780–785 Kurtz S, Ong K, Lau E, Mowat F, Halpern M (2007) Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am 89(4):780–785
4.
Zurück zum Zitat Robertsson O, Bizjajeva S, Fenstad AM et al (2010) Knee arthroplasty in Denmark, Norway and Sweden: a pilot study from the Nordic Arthroplasty Register Association. Acta Orthop 81(1):82–89CrossRef Robertsson O, Bizjajeva S, Fenstad AM et al (2010) Knee arthroplasty in Denmark, Norway and Sweden: a pilot study from the Nordic Arthroplasty Register Association. Acta Orthop 81(1):82–89CrossRef
5.
Zurück zum Zitat Callaghan JJ, O’roukre MR, Saleh K (2004) Why knees fail: lessons learned. J Arthroplasty 19(4 Suppl 1):31–34CrossRef Callaghan JJ, O’roukre MR, Saleh K (2004) Why knees fail: lessons learned. J Arthroplasty 19(4 Suppl 1):31–34CrossRef
6.
Zurück zum Zitat Dalury DF, Pomeroy DL, Gorab RS, Adams MJ (2013) Why are total knee arthroplasties being revised? J Arthroplasty 28(8 Suppl):120–121CrossRef Dalury DF, Pomeroy DL, Gorab RS, Adams MJ (2013) Why are total knee arthroplasties being revised? J Arthroplasty 28(8 Suppl):120–121CrossRef
7.
Zurück zum Zitat Hirschmann MT, Becker R (2015) The unhappy knee replacement. A comprehensive review and management guide. Roland Becker Editors. Springer, New York Hirschmann MT, Becker R (2015) The unhappy knee replacement. A comprehensive review and management guide. Roland Becker Editors. Springer, New York
8.
Zurück zum Zitat Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP (2011) The position and orientation of total knee replacement components. J Bone Joint Surg Br 93(5):629–633CrossRef Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP (2011) The position and orientation of total knee replacement components. J Bone Joint Surg Br 93(5):629–633CrossRef
9.
Zurück zum Zitat Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME, Ritter MA, Nunley RM (2013) Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplasty 28(8 Suppl):116–119CrossRef Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME, Ritter MA, Nunley RM (2013) Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplasty 28(8 Suppl):116–119CrossRef
10.
Zurück zum Zitat Armstrong AD, Brien HJ, Dunning CE, King GJ, Johnson JA, Chess DG (2003) Patellar position after total knee arthroplasty: influence of femoral component malposition. J Arthroplasty 18(4):458–465CrossRef Armstrong AD, Brien HJ, Dunning CE, King GJ, Johnson JA, Chess DG (2003) Patellar position after total knee arthroplasty: influence of femoral component malposition. J Arthroplasty 18(4):458–465CrossRef
11.
Zurück zum Zitat Berger RA, Crossett LS, Jacobs JJ, Rubash HE (1998) Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. (356):144–153CrossRef Berger RA, Crossett LS, Jacobs JJ, Rubash HE (1998) Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. (356):144–153CrossRef
12.
Zurück zum Zitat Nicoll D, Rowley DI (2010) Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg Br 92(9):1238–4124CrossRef Nicoll D, Rowley DI (2010) Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg Br 92(9):1238–4124CrossRef
13.
Zurück zum Zitat Slevin O, Amsler F, Hirschmann MT (2016) No correlation between coronal alignment of total knee arthroplasty and clinical outcomes: a prospective clinical study using 3D-CT. Knee Surg Sports Traumatol Arthrosc 25(12):3892–3900CrossRef Slevin O, Amsler F, Hirschmann MT (2016) No correlation between coronal alignment of total knee arthroplasty and clinical outcomes: a prospective clinical study using 3D-CT. Knee Surg Sports Traumatol Arthrosc 25(12):3892–3900CrossRef
14.
15.
Zurück zum Zitat Slevin O, Schmid FA, Schiapparelli FF, Rasch H, Amsler F, Hirschmann MT (2017) Coronal femoral TKA position significantly influences in vivo patellar loading in unresurfaced patellae after primary total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 25(11):3605–3610CrossRef Slevin O, Schmid FA, Schiapparelli FF, Rasch H, Amsler F, Hirschmann MT (2017) Coronal femoral TKA position significantly influences in vivo patellar loading in unresurfaced patellae after primary total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 25(11):3605–3610CrossRef
16.
Zurück zum Zitat Chambers AW, Wood AR, Kosmopoulos V, Sanchez HB, Wagner RA (2016) Effect of posterior tibial slope on flexion and anterior-posterior tibial translation in posterior cruciate-retaining total knee arthroplasty. J Arthroplasty 31(1):103–106CrossRef Chambers AW, Wood AR, Kosmopoulos V, Sanchez HB, Wagner RA (2016) Effect of posterior tibial slope on flexion and anterior-posterior tibial translation in posterior cruciate-retaining total knee arthroplasty. J Arthroplasty 31(1):103–106CrossRef
17.
Zurück zum Zitat Di Laura A, Whittaker R, Hothi H, Kwon YM, Skinner J, Hart A (2016) Dual mobility vs standard bearings: what is the effect on taper junction? Bone Joint J 98(sup 11):3 Di Laura A, Whittaker R, Hothi H, Kwon YM, Skinner J, Hart A (2016) Dual mobility vs standard bearings: what is the effect on taper junction? Bone Joint J 98(sup 11):3
18.
Zurück zum Zitat Wasielewski RC, Galante JO, Leighty RM, Roseneberg AG (1994) Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical consideration during total knee arthroplasty. Clin Orthop Relat Res 299:31–43 Wasielewski RC, Galante JO, Leighty RM, Roseneberg AG (1994) Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical consideration during total knee arthroplasty. Clin Orthop Relat Res 299:31–43
19.
Zurück zum Zitat Metthies AK, Henckel J, Skinner JA, Hart AJ (2011) A retrieval analysis of explanted Durom metal-on-metal hip arthroplasties. Hip Int 21(06):724–731CrossRef Metthies AK, Henckel J, Skinner JA, Hart AJ (2011) A retrieval analysis of explanted Durom metal-on-metal hip arthroplasties. Hip Int 21(06):724–731CrossRef
20.
Zurück zum Zitat Currier JH, Bill MA, Mayor MB (2005) Analysis of wear asymmetry in a series of 94 retrieved polyethylene tibial bearings. J Biomech 38(2):367–375CrossRef Currier JH, Bill MA, Mayor MB (2005) Analysis of wear asymmetry in a series of 94 retrieved polyethylene tibial bearings. J Biomech 38(2):367–375CrossRef
21.
Zurück zum Zitat Flannery M, Jones E, Birkinshaw C (2010) Compliant layer knee bearings. Part II: preliminary wear assessment. Wear 269(5–6):331–338CrossRef Flannery M, Jones E, Birkinshaw C (2010) Compliant layer knee bearings. Part II: preliminary wear assessment. Wear 269(5–6):331–338CrossRef
22.
Zurück zum Zitat Liau JJ, Cheng CK, Huang CH, Lo WH (2002) The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol Avon) 17(2):140–146CrossRef Liau JJ, Cheng CK, Huang CH, Lo WH (2002) The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol Avon) 17(2):140–146CrossRef
23.
Zurück zum Zitat Lokte PA, Ecker ML (1977) Influence of positioning of Prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79CrossRef Lokte PA, Ecker ML (1977) Influence of positioning of Prosthesis in total knee replacement. J Bone Joint Surg Am 59(1):77–79CrossRef
24.
Zurück zum Zitat Pang HN, Bin Abd Razak HR, Jamiecson P, Teeter MG, Naudie DD, MacDonald SJ (2016) Factors affecting wear of constrained polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty 31(6):1340–1345CrossRef Pang HN, Bin Abd Razak HR, Jamiecson P, Teeter MG, Naudie DD, MacDonald SJ (2016) Factors affecting wear of constrained polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty 31(6):1340–1345CrossRef
25.
Zurück zum Zitat Srivastava A, Lee GY, Steklov N, Colwell CW Jr, Ezzet KA, D’Lima DD (2012) Effect of tibial component varus on wear in total knee arthroplasty. Knee 19(5):560–563CrossRef Srivastava A, Lee GY, Steklov N, Colwell CW Jr, Ezzet KA, D’Lima DD (2012) Effect of tibial component varus on wear in total knee arthroplasty. Knee 19(5):560–563CrossRef
26.
Zurück zum Zitat Henckel J, Richards R, Lozhkin K, Harris S, Rodriguez y Baena FM, Barrett ARW, Cobb JP (2006) Very low-dose computed tomography for planning and outcome measurement in knee replacement. The imperial knee protocol. J Bone Joint Surg 88-B:1513–1518CrossRef Henckel J, Richards R, Lozhkin K, Harris S, Rodriguez y Baena FM, Barrett ARW, Cobb JP (2006) Very low-dose computed tomography for planning and outcome measurement in knee replacement. The imperial knee protocol. J Bone Joint Surg 88-B:1513–1518CrossRef
27.
Zurück zum Zitat Harman MK, Banks SA, Kirschner S, Lützner J (2012) Prosthesis alignment affects axial rotation motion after total knee replacement: a prospective in vivo study combining computed tomography and fluoroscopic evaluations. BMC Musculoskelet Disord 13:206CrossRef Harman MK, Banks SA, Kirschner S, Lützner J (2012) Prosthesis alignment affects axial rotation motion after total knee replacement: a prospective in vivo study combining computed tomography and fluoroscopic evaluations. BMC Musculoskelet Disord 13:206CrossRef
28.
Zurück zum Zitat Gromov K, Korchi M, Thomsen MG, Husted H, Troelsen A (2014) What is the optimal alignment of the tibial and femoral components in knee arthroplasty? An overview of the literature. Acta Orthop 85(5):480–487CrossRef Gromov K, Korchi M, Thomsen MG, Husted H, Troelsen A (2014) What is the optimal alignment of the tibial and femoral components in knee arthroplasty? An overview of the literature. Acta Orthop 85(5):480–487CrossRef
29.
Zurück zum Zitat Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73(5):709–714CrossRef Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73(5):709–714CrossRef
30.
Zurück zum Zitat Hood RW, Wright TM, Burstein AH (1983) Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res 17:829–842CrossRef Hood RW, Wright TM, Burstein AH (1983) Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res 17:829–842CrossRef
31.
Zurück zum Zitat Ferrara F, Cipriani A, Rapisarda S, Iacobucci M, Magarelli N, Leone A, Bonomo L (2016) Assessment of implant position after total knee arthroplasty by dual-energy computed tomography. Acta Radiol 57(5):612–619CrossRef Ferrara F, Cipriani A, Rapisarda S, Iacobucci M, Magarelli N, Leone A, Bonomo L (2016) Assessment of implant position after total knee arthroplasty by dual-energy computed tomography. Acta Radiol 57(5):612–619CrossRef
32.
Zurück zum Zitat Yau WP, Chiu KY, Tang WM (2007) How precise is the determination of rotational alignment of the femoral prosthesis in total knee arthroplasty: an in vivo study. J Arthroplasty 22(7):1042–1048CrossRef Yau WP, Chiu KY, Tang WM (2007) How precise is the determination of rotational alignment of the femoral prosthesis in total knee arthroplasty: an in vivo study. J Arthroplasty 22(7):1042–1048CrossRef
33.
Zurück zum Zitat Knowlton CB, Bhutani P, Wimmer MA (2016) Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners. J Biomed Mater Res Part B 105(7):2053–2059CrossRef Knowlton CB, Bhutani P, Wimmer MA (2016) Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners. J Biomed Mater Res Part B 105(7):2053–2059CrossRef
34.
Zurück zum Zitat Gilbert SL, Rana AJ, Lipman JD, Wright GH, Westrich GH (2014) Design changes improve contact patterns and articular surface damage in total knee arthroplasty. Knee 21(6):1129–1134CrossRef Gilbert SL, Rana AJ, Lipman JD, Wright GH, Westrich GH (2014) Design changes improve contact patterns and articular surface damage in total knee arthroplasty. Knee 21(6):1129–1134CrossRef
35.
Zurück zum Zitat McClung CD, Zahiri CA, Higa JK, Amstutz HC, Schmalzried TP (2000) Relationship between body mass index and activity in hip or knee arthroplasty patients. J Orthop Res 18(1):35–39CrossRef McClung CD, Zahiri CA, Higa JK, Amstutz HC, Schmalzried TP (2000) Relationship between body mass index and activity in hip or knee arthroplasty patients. J Orthop Res 18(1):35–39CrossRef
Metadaten
Titel
3D patient imaging and retrieval analysis help understand the clinical importance of rotation in knee replacements
verfasst von
Arianna Cerquiglini
Johann Henckel
Harry Hothi
Niccoló Rotigliano
Michael T. Hirschmann
Alister J. Hart
Publikationsdatum
08.03.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 11/2018
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-018-4891-9

Weitere Artikel der Ausgabe 11/2018

Knee Surgery, Sports Traumatology, Arthroscopy 11/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.