68Ga-PSMA-11 dose reduction for dedicated pelvic imaging with simultaneous PET/MR using TOF BSREM reconstructions

verfasst von: Hanna Svirydenka, Urs J. Muehlematter, Hannes W. Nagel, Gaspar Delso, Daniela A. Ferraro, Ken Kudura, Irene A. Burger, Edwin E. G. W. ter Voert

Erschienen in: European Radiology | Ausgabe 6/2020

Einloggen, um Zugang zu erhalten

Abstract

Objectives

When increasing the PET acquisition time to match the longer MRI protocol in simultaneous PET/MR, the injected PET tracer dose can possibly be lowered to reduce radiation exposure. Moreover, applying new commercially available time-of-flight (TOF) block sequential regularized expectation maximization (BSREM)–based reconstruction algorithms could allow for further dose reductions. The purpose of this study was to find the minimal dose of the tracer targeting the prostate specific membrane antigen (68Ga-PSMA-11) for a dedicated 15-min pelvic PET/MR scan that still matches the image quality of a reference 3-min scan at 100% (150 MBq) dose.

Methods

In this retrospective analysis, 25 patients were included. PET emission datasets were edited to simulate stepwise reductions of injected tracer dose. Reference TOF ordered subset expectation maximum (OSEM) and new TOF BSREM reconstructions were performed and differences in the resulting PET images were visually and quantitatively assessed.

Results

Visually, TOF BSREM reconstructions with relatively high regularization parameter (β) values are preferred. Quantitatively, however, high β-values result in lower lesion maximum standardized uptake values (SUVmax) compared to the reference. A β-value of 550 was considered the optimal compromise for the lowest possible 10% dose reconstructions, resulting in comparable visual assessment and lesion SUVmax.

Conclusions

This study indicates that the injected 68Ga-PSMA-11 tracer dose for a standard 3-min PET scan can be reduced to approximately 10% (15 MBq) when the PET acquisition time is matched to the 15-min pelvic MRI protocol, and when reconstructed with TOF BSREM using β = 550. This decreases the effective dose from 3.54 to 0.35 mSv.

Key Points

• Low-dose dedicated pelvic 68 Ga-PSMA-11 PET/MR reduces radiation exposure for patients.
• Retrospective study investigating the minimal dose needed for adequate image quality for 15-min PET frames over the pelvis showed using quantitative and qualitative analysis that a substantial dose reduction is possible without significant loss of image quality when using the TOF BSREM reconstruction algorithm.
• With the introduction of low-dose pelvic 68 Ga-PSMA-11 PET/MR, new potential applications of 68 Ga-PSMA-11 PET for local staging or investigation of equivocal MRI findings could become applicable, even for patients without confirmed prostate cancer.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Eiber M, Weirich G, Holzapfel K et al (2016) Simultaneous (68)Ga-PSMA HBED-CC PET/MRI improves the localization of primary prostate cancer. Eur Urol 70:829–836CrossRef Eiber M, Weirich G, Holzapfel K et al (2016) Simultaneous (68)Ga-PSMA HBED-CC PET/MRI improves the localization of primary prostate cancer. Eur Urol 70:829–836CrossRef
2.
Zurück zum Zitat Park SY, Zacharias C, Harrison C et al (2018) Gallium 68 PSMA-11 PET/MR imaging in patients with intermediate- or high-risk prostate cancer. Radiology 288:495–505CrossRef Park SY, Zacharias C, Harrison C et al (2018) Gallium 68 PSMA-11 PET/MR imaging in patients with intermediate- or high-risk prostate cancer. Radiology 288:495–505CrossRef
3.
Zurück zum Zitat Silver DA, Pellicer I, Fair WR, Heston WD, Cordon-Cardo C (1997) Prostate-specific membrane antigen expression in normal and malignant human tissues. Clin Cancer Res 3:81–85PubMed Silver DA, Pellicer I, Fair WR, Heston WD, Cordon-Cardo C (1997) Prostate-specific membrane antigen expression in normal and malignant human tissues. Clin Cancer Res 3:81–85PubMed
5.
Zurück zum Zitat Boellaard R, Quick HH (2015) Current image acquisition options in PET/MR. Semin Nucl Med 45:192–200CrossRef Boellaard R, Quick HH (2015) Current image acquisition options in PET/MR. Semin Nucl Med 45:192–200CrossRef
6.
Zurück zum Zitat Qi J, Leahy RM (2006) Iterative reconstruction techniques in emission computed tomography. Phys Med Biol 51:R541–R578CrossRef Qi J, Leahy RM (2006) Iterative reconstruction techniques in emission computed tomography. Phys Med Biol 51:R541–R578CrossRef
7.
Zurück zum Zitat Boellaard R, van Lingen A, Lammertsma AA (2001) Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling. J Nucl Med 42:808–817PubMed Boellaard R, van Lingen A, Lammertsma AA (2001) Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling. J Nucl Med 42:808–817PubMed
8.
Zurück zum Zitat Liow JS, Strother SC (1993) The convergence of object dependent resolution in maximum likelihood based tomographic image reconstruction. Phys Med Biol 38:55–70CrossRef Liow JS, Strother SC (1993) The convergence of object dependent resolution in maximum likelihood based tomographic image reconstruction. Phys Med Biol 38:55–70CrossRef
9.
Zurück zum Zitat Asma E, Manjeshwar R (2007) Analysis of organ uniformity in low count density penalized likelihood PET images. 2007 IEEE Nuclear Science Symposium Conference Record, pp 4426–4432 Asma E, Manjeshwar R (2007) Analysis of organ uniformity in low count density penalized likelihood PET images. 2007 IEEE Nuclear Science Symposium Conference Record, pp 4426–4432
10.
Zurück zum Zitat Asma E, Ahn S, Ross SG, Chen A, Manjeshwar RM (2012) Accurate and consistent lesion quantitation with clinically acceptable penalized likelihood images. 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC), pp 4062–4066 Asma E, Ahn S, Ross SG, Chen A, Manjeshwar RM (2012) Accurate and consistent lesion quantitation with clinically acceptable penalized likelihood images. 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC), pp 4062–4066
11.
Zurück zum Zitat Geman S, Geman D (1984) Stochastic relaxation, Gibbs distributions, and the Bayesian restoration of images. IEEE Trans Pattern Anal Mach Intell 6:721–741CrossRef Geman S, Geman D (1984) Stochastic relaxation, Gibbs distributions, and the Bayesian restoration of images. IEEE Trans Pattern Anal Mach Intell 6:721–741CrossRef
12.
Zurück zum Zitat Geman S, McClure D (1985) Bayesian image analysis methods: an application to single photon emission computed tomography. Proc Statistical Computation Section, pp 12–18 Geman S, McClure D (1985) Bayesian image analysis methods: an application to single photon emission computed tomography. Proc Statistical Computation Section, pp 12–18
13.
Zurück zum Zitat Mumcuoglu EU, Leahy RM, Cherry SR (1996) Bayesian reconstruction of PET images: methodology and performance analysis. Phys Med Biol 41:1777–1807CrossRef Mumcuoglu EU, Leahy RM, Cherry SR (1996) Bayesian reconstruction of PET images: methodology and performance analysis. Phys Med Biol 41:1777–1807CrossRef
14.
Zurück zum Zitat Ahn S, Ross SG, Asma E et al (2015) Quantitative comparison of OSEM and penalized likelihood image reconstruction using relative difference penalties for clinical PET. Phys Med Biol 60:5733CrossRef Ahn S, Ross SG, Asma E et al (2015) Quantitative comparison of OSEM and penalized likelihood image reconstruction using relative difference penalties for clinical PET. Phys Med Biol 60:5733CrossRef
15.
Zurück zum Zitat Sah BR, Stolzmann P, Delso G et al (2017) Clinical evaluation of a block sequential regularized expectation maximization reconstruction algorithm in 18F-FDG PET/CT studies. Nucl Med Commun 38:57–66CrossRef Sah BR, Stolzmann P, Delso G et al (2017) Clinical evaluation of a block sequential regularized expectation maximization reconstruction algorithm in 18F-FDG PET/CT studies. Nucl Med Commun 38:57–66CrossRef
16.
Zurück zum Zitat Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, Gleeson FV (2015) Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT system. J Nucl Med 56:1447–1452CrossRef Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, Gleeson FV (2015) Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT system. J Nucl Med 56:1447–1452CrossRef
17.
Zurück zum Zitat Ter Voert EEGV, Muehlematter UJ, Delso G et al (2018) Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical (68)Ga-PSMA PET/MR. EJNMMI Res 8:70 Ter Voert EEGV, Muehlematter UJ, Delso G et al (2018) Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical (68)Ga-PSMA PET/MR. EJNMMI Res 8:70
18.
Zurück zum Zitat Fendler WP, Eiber M, Beheshti M et al (2017) (68)Ga-PSMA PET/CT: joint EANM and SNMMI procedure guideline for prostate cancer imaging: version 1.0. Eur J Nucl Med Mol Imaging 44:1014–1024CrossRef Fendler WP, Eiber M, Beheshti M et al (2017) (68)Ga-PSMA PET/CT: joint EANM and SNMMI procedure guideline for prostate cancer imaging: version 1.0. Eur J Nucl Med Mol Imaging 44:1014–1024CrossRef
19.
Zurück zum Zitat Huang SY, Savic D, Yang J, Shrestha U, Seo Y (2014) The effect of magnetic field on positron range and spatial resolution in an integrated whole-body time-of-flight PET/MRI system. IEEE Nucl Sci Symp Conf Rec (1997) 2014 Huang SY, Savic D, Yang J, Shrestha U, Seo Y (2014) The effect of magnetic field on positron range and spatial resolution in an integrated whole-body time-of-flight PET/MRI system. IEEE Nucl Sci Symp Conf Rec (1997) 2014
20.
Zurück zum Zitat Levin C, Glover G, Deller T, McDaniel D, Peterson W, Maramraju SH (2013) Prototype time-of-flight PET ring integrated with a 3T MRI system for simultaneous whole-body PET/MR imaging. J Nucl Med Meeting Abstracts 54:148 Levin C, Glover G, Deller T, McDaniel D, Peterson W, Maramraju SH (2013) Prototype time-of-flight PET ring integrated with a 3T MRI system for simultaneous whole-body PET/MR imaging. J Nucl Med Meeting Abstracts 54:148
21.
Zurück zum Zitat Wollenweber SD, Ambwani S, Delso G et al (2013) Evaluation of an atlas-based PET head attenuation correction using PET/CT & MR patient data. IEEE Trans Nucl Sci 60:3383–3390CrossRef Wollenweber SD, Ambwani S, Delso G et al (2013) Evaluation of an atlas-based PET head attenuation correction using PET/CT & MR patient data. IEEE Trans Nucl Sci 60:3383–3390CrossRef
22.
Zurück zum Zitat Wollenweber SD, Ambwani S, Lonn AHR et al (2013) Comparison of 4-class and continuous fat/water methods for whole-body, MR-based PET attenuation correction. IEEE Trans Nucl Sci 60:3391–3398CrossRef Wollenweber SD, Ambwani S, Lonn AHR et al (2013) Comparison of 4-class and continuous fat/water methods for whole-body, MR-based PET attenuation correction. IEEE Trans Nucl Sci 60:3391–3398CrossRef
23.
Zurück zum Zitat Caribe PPRV, Koole M, D'Asseler Y, Deller TW, Van Laere K, Vandenberghe S (2019) NEMA NU 2-2007 performance characteristics of GE Signa integrated PET/MR for different PET isotopes. EJNMMI Phys 6:11 Caribe PPRV, Koole M, D'Asseler Y, Deller TW, Van Laere K, Vandenberghe S (2019) NEMA NU 2-2007 performance characteristics of GE Signa integrated PET/MR for different PET isotopes. EJNMMI Phys 6:11
24.
Zurück zum Zitat Sah BR, Ghafoor S, Burger IA et al (2018) Feasibility of (18)F-FDG dose reductions in breast cancer PET/MRI. J Nucl Med 59:1817–1822CrossRef Sah BR, Ghafoor S, Burger IA et al (2018) Feasibility of (18)F-FDG dose reductions in breast cancer PET/MRI. J Nucl Med 59:1817–1822CrossRef
25.
Zurück zum Zitat Sekine T, Delso G, Zeimpekis KG et al (2018) Reduction of (18)F-FDG dose in clinical PET/MR imaging by using silicon photomultiplier detectors. Radiology 286:249–259CrossRef Sekine T, Delso G, Zeimpekis KG et al (2018) Reduction of (18)F-FDG dose in clinical PET/MR imaging by using silicon photomultiplier detectors. Radiology 286:249–259CrossRef
26.
Zurück zum Zitat Gandhi H, Holley D, Gulaka P, Iagaru A (2017) 68Ga-PSMA 11 PET/MRI influence of acquisition time on image quality. J Nucl Med 58:798 Gandhi H, Holley D, Gulaka P, Iagaru A (2017) 68Ga-PSMA 11 PET/MRI influence of acquisition time on image quality. J Nucl Med 58:798
27.
Zurück zum Zitat Conti M, Eriksson L (2016) Physics of pure and non-pure positron emitters for PET: a review and a discussion. EJNMMI Phys 3:8–8CrossRef Conti M, Eriksson L (2016) Physics of pure and non-pure positron emitters for PET: a review and a discussion. EJNMMI Phys 3:8–8CrossRef
28.
Zurück zum Zitat Teoh EJ, McGowan DR, Schuster DM, Tsakok MT, Gleeson FV, Bradley KM (2018) Bayesian penalised likelihood reconstruction (Q.Clear) of (18)F-fluciclovine PET for imaging of recurrent prostate cancer: semi-quantitative and clinical evaluation. Br J Radiol 91:20170727CrossRef Teoh EJ, McGowan DR, Schuster DM, Tsakok MT, Gleeson FV, Bradley KM (2018) Bayesian penalised likelihood reconstruction (Q.Clear) of (18)F-fluciclovine PET for imaging of recurrent prostate cancer: semi-quantitative and clinical evaluation. Br J Radiol 91:20170727CrossRef
29.
Zurück zum Zitat Lawhn Heath C, Flavell R, Deller T, Lake S, Carroll P, Hope T (2017) Scatter artifact with 68Ga PSMA-PET: severity reduced with furosemide diuresis and improved time-of-flight scatter correction. J Nucl Med 58:738 Lawhn Heath C, Flavell R, Deller T, Lake S, Carroll P, Hope T (2017) Scatter artifact with 68Ga PSMA-PET: severity reduced with furosemide diuresis and improved time-of-flight scatter correction. J Nucl Med 58:738
30.
Zurück zum Zitat Wangerin KA, Baratto L, Khalighi MM et al (2018) Clinical evaluation of (68)Ga-PSMA-II and (68)Ga-RM2 PET images reconstructed with an improved scatter correction algorithm. AJR Am J Roentgenol 211:655–660CrossRef Wangerin KA, Baratto L, Khalighi MM et al (2018) Clinical evaluation of (68)Ga-PSMA-II and (68)Ga-RM2 PET images reconstructed with an improved scatter correction algorithm. AJR Am J Roentgenol 211:655–660CrossRef
Metadaten
Titel
68Ga-PSMA-11 dose reduction for dedicated pelvic imaging with simultaneous PET/MR using TOF BSREM reconstructions
verfasst von
Hanna Svirydenka
Urs J. Muehlematter
Hannes W. Nagel
Gaspar Delso
Daniela A. Ferraro
Ken Kudura
Irene A. Burger
Edwin E. G. W. ter Voert
Publikationsdatum
14.02.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 6/2020
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-06667-2

Weitere Artikel der Ausgabe 6/2020

European Radiology 6/2020 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.