Open Access 01.12.2020 | Research article

A 3D-printed, personalized, biomechanics-specific beta-tricalcium phosphate bioceramic rod system: personalized treatment strategy for patients with femoral shaft non-union based on finite element analysis

verfasst von: Jian Lu, Qi-Yang Wang, Jia-Gen Sheng, Shang-Chun Guo, Shi-Cong Tao

Erschienen in: BMC Musculoskeletal Disorders | Ausgabe 1/2020

Abstract

Background

Although double-plate fixation (DP), i.e., fixation with a combination of a main lateral plate (LP) and a support medial plate (MP), is a relatively mature method for treating femoral shaft non-union with bone defect causes complications. The purpose of this study was to evaluate LP fixation with a 3D-printed, personalized, biomechanics-specific β-TCP bioceramic rod system (LP + 3DpbsBRS) as an alternative with less collateral damage.

Methods

Structure-specific finite element modelling was used to simulate femoral shaft non-union with bone defects and treatment with an LP only as the blank control. Then, the peak von Mises stress (VMS), the VMS distribution, and the plate displacement were determined to compare the effectiveness of LP + CBG (cancellous bone grafting), DP + CBG, and LP + 3DpbsBRS under 850 N of axial force.

Results

Our results indicated that the peak VMS was 260.2 MPa (LP + 3DpbsBRS), 249.6 MPa (MP in DP + CBG), 249.3 MPa (LP in DP + CBG), and 502.4 MPa (LP + CBG). The bending angle of the plate was 1.2° versus 1.0° versus 1.1° versus 2.3° (LP + 3DpbsBRS versus MP in DP + CBG versus LP in DP + CBG versus LP + CBG).

Conclusion

The 3DpbsBRS in the LP + 3DpbsBRS group could replace the MP in the DP + CBG group by providing similar medial mechanical support. Furthermore, avoiding the use of an MP provides better protection of the soft tissue and vasculature.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
DP
Double-plate fixation
LP
Lateral plate
MP
Support medial plate
β-TCP
Beta-tricalcium phosphate
3DpbsBRS
3D-printed, personalized, biomechanics-specific β-TCP bioceramic rod system
VMS
Von Mises stress
CBG
Cancellous bone grafting
CT
Computed tomography

Background

Fractures are orthopaedic conditions that can occur at any age and are mostly caused by high-energy trauma; approximately 1.1 to 2.9 million fractures occur per year worldwide [1]. The probability of non-union after fracture is as high as 5–10% [2, 3], and non-union occurs in 1–20% of femoral shaft fractures [4]. Treating femoral non-union causes an economic [5] and psychological burden [6] on patients and is a major challenge for orthopaedic surgeons worldwide. According to imaging features, non-union can be divided into hypertrophic non-union and atrophic non-union [7, 8]. Hypertrophic non-union, also known as mechanical non-union, is characterized by excessive bone formation and poor mechanical fixation [9]. To promote mechanical stability, the most common clinical treatments include supplemental fixation, e.g., nail dynamization [10], exchanging nailing with augmentation plating [11], and augmentation plating [12]; however, the healing rate is variable (range, 53–96%) for these procedures. Atrophic non-union is characterized by the absence of a callus and cartilage due to a lack of cells and blood supply. Therefore, the fracture site may be sclerotic or osteopenic, which may lead to mechanical instability. Herein, we focus on two issues associated with the treatment of long bone non-union: mechanical stability and biological activity [13].
Autologous bone grafts with mechanical stability are the “gold standard” for the treatment of femoral shaft non-union with bone defects due to their complete histocompatibility and strong osteoconductive, osteoinductive, and osteogenic activities [13]. The rate of complications of autologous bone grafting is as high as 23% [14]; complications include pain at the donor site, haematoma, infection, loss of sensation, scar formation, and limited source of bone [15, 16]. Donor site injury problems and complications have spurred research to identify other treatment methods. Due to its excellent rigidity and stability, fixation with double-locking compression plates is one method for addressing the instability in long bone non-union [17]. However, at the same time, the medial aspect of the femur often loses a large amount of soft tissue, which leads to a reduction in the blood supply and secondary damage to bone continuity [17, 18].
The purpose of the surgical treatment of patients with femoral shaft non-union with bone defects is to provide a rigid, stable structure and create a healthy, biological environment conducive to fracture healing [9], which often is challenging for orthopaedists. In recent years, many clinical studies [1922] have demonstrated the efficacy of beta-tricalcium phosphate (β-TCP) bioceramics as a bone graft substitute for repairing bone defects. β-TCP bioceramics have good biocompatibility and biodegradability. Furthermore, they have excellent microporosity, which is beneficial for inducing vascularization, improving osteoconductivity, and promoting cell proliferation and differentiation. However, porous bioceramics have very weak tension, which limits their application in the treatment of bone defects in weight-bearing areas. Dense bioceramics have improved mechanical properties and the ability to degrade in vitro, which could be complementary to the low mechanical properties and high bioactivity of porous bioceramics. Thus, we designed a model in which the distribution of dense and porous bioceramics would be determined according to the stress distribution of implants used to treat bone defects. In the early stages after implantation, the dense bioceramic could provide excellent mechanical support at the site of non-union, while the porous bioceramic could induce vascularization, allowing the transport of nutrients and bone ingrowth. In the late stage after implantation, the material would gradually degrade with new bone formation, and the desired biomechanical support, at the site of the bone defect, would be provided by the reconstructed bone.
In this study, we established a standardized model of femoral shaft non-union, and then, according to the predicted stress distribution of the implants that were to be implanted into the defective bone, developed a 3D-printed, personalized, biomechanics-specific β-TCP bioceramic rod system (3DpbsBRS). In a series of follow-up finite element analyses, we evaluated the biomechanical properties of the 3DpbsBRS and determined whether fixation with the 3DpbsBRS and only a lateral plate (LP) could offer similar medial support as fixation with a double plate (DP), which involves a medial plate (MP) to provide medial support and is associated with more soft tissue injury and loss of blood supply to the periosteum. This study will provide a new perspective for follow-up studies; the 3DpbsBRS is expected to provide a personalized clinical solution for individual patients in various situations, based on the idea that the combination of predictive biomechanical computation and 3D printing technology could provide personalized mechanical support, reduced plate use, and better protection of tissue and blood vessels.

Methods

Ethical review

This study was approved by the Ethics Committee of the Shanghai Jiao Tong University Affiliated Sixth People’s Hospital and involved the examination of an adult volunteer with a written informed consent before the study began (sex: male, age: 20, height: 178 cm, weight: 75 kg) by enhanced computed tomography (CT) to obtain raw imaging data of a normal femur.

Establishment of a finite element model for femoral fixation

The raw data of slices at a 0.6-mm interval in Digital Imaging and Communications in Medicine (DICOM) format were imported into Mimics 20.0 (The Materialise Group, Leuven, Belgium) to establish 3D geometric models. Herein, sampling and surface building for geometry were performed using Geomagic software. Next, the fundamental 3D models obtained were compiled and meshed in HyperMesh 14.0 (USA). Figure 1a-c depicts the sequence of software used in this study. Then, in 3-Matic 11.0 (Materialise, Leuven, Belgium), a 15-mm transverse osteotomy plane was made at the mid-end of the femur (168 mm from the lateral femoral condyle) to simulate femoral shaft non-union with bone defects (Fig. 1d).
According to the blueprint provided by the manufacturer, we reconstructed the geometric 3D model of the LP, MP and screws (Synthes, 3.5-mm LCP) using Solid Works 14.0 (Solid Works Corp, Dassault Systèmes, Concord, MA, USA). The plates and the femur were assembled into four case models in 3-Matic 11.0 (Fig. 2). The threaded surface of the screws was replaced by a smooth surface; the size of the surface corresponded to the average diameter of the screw provided by the manufacturer [4].
Case 1 (LP only group: lateral main plate only): a complete femoral defect of 15 mm, with fixation of the lateral femur with a 9-hole, 3.5-mm LCP (Fig. 2a).
Case 2 (LP + CBG group: lateral main plate with cancellous bone grafting): same as case 1 with the addition of filling the defect with cancellous bone (Fig. 2b).
Case 3 (DP + CBG group: double plates with cancellous bone grafting): same as case 2 with the addition of a 6-hole, 3.5-mm LCP to the medial femoral aspect (Fig. 2c).
Case 4 (LP + 3DpbsBRS group: lateral main plate with 3D-printed, personalized, biomechanics-specific β-TCP bioceramic rod system): fixation with a 3.5-mm LCP on the lateral femur and filling of the defect with the 3DpbsBRS, unlike in case 2 (Fig. 2d). The steps for 3DpbsBRS acquisition were as follows: select the case 2 model of LP + CBG at the non-union bone ends for finite element analysis (Fig. 3b) and obtain the stress distribution of the grafted cancellous bone (Fig. 3c). According to the stress distribution, design the BRS with porous bioceramic when the stress is less than 2 MPa and dense bioceramic when the stress is greater than 2 MPa, yielding the 3DpbsBRS (Fig. 3d).
All of these case models underwent follow-up biomechanical simulation. Subsequently, the combined fixation and femoral model were meshed at 1 mm using HyperMesh 14.0 (USA). Finally, the combined models were imported into Abaqus 6.14 (Dassault Systèmes, USA) to generate a finite element model for mechanical analysis.

Material properties and value assignment for finite element analysis

Based on earlier findings, the femur and strut were assumed to be linear, isotropic and elastic [23]. The LCP and screws were made of a titanium alloy (Ti-6AL-4 V). Values for the β-TCP BRS were based on data provided by Shanghai Bio-Lu Biomaterials Co., Ltd. (Shanghai, China). The porous bioceramic had a porosity of approximately 70%, a pore size of approximately 500 μm, and a pore interconnect diameter of approximately 150 μm. Table 1 shows the elastic modulus and Poisson ratio of the material, 3DpbsBRS model, and bone.
Table 1
Material attributes for value assignment in the finite element models (Ti-6AL-4 V, cortical, trabecular, porous ceramic granules and dense ceramic granules)
Components
Ti-6AL-4 V
Bone
β-TCP Bioceramic
Cortical
Trabecular
Porous ceramic granules
Dense ceramic granules
E (GPa)
105
16.7
0.155
0.2
7.49
Poisson ratio
0.35
0.26
0.3
0.3
0.3
Porosity
   
70%
5–10%
Compressive strength (MPa)
   
2.15
62
β-TCP beta-tricalcium phosphate; E, Young’s modulus

Finite element analysis

According to previously published studies [24], it was assumed that there was frictional interaction between different parts of the model. The internal fixation was considered to be in the locked state, so the interface of the screw and the LP was set to be bonded, screws were tied to the bone, therefore not allowing any movement between those parts. The coefficient of friction between the cortical bone and the cancellous graft, bone and bone graft were both 0.46, and the coefficient of friction between bone and the steel plate was 0.3 [25]. To prevent rigid body motions during analysis, and all nodes on the distal surface of the femur were placed under a 0 degree-of-freedom constraint [26], under boundary conditions. Then, 850 N was applied to the centre of the femoral head of the finite element model, which was equivalent to 100% of the body weight (Fig. 3a) [27, 28]. According to Eberle et al. [14], the force vector pointed laterally 13° on the coronal plane and 8° on the sagittal plane.

Main outcome measures

Three parameters were used to capture mechanical factors involved in fixation stability and fracture healing: the peak von Mises stress (VMS) of the implant, the VMS distribution of the implants and the displacement and deformation of the model.

Results

The counts of element and node of four models were shownin the Table 2. The peak VMS of the plate was concentrated on the surface of the plate near the bone defect. The four fixation models showed great differences in the stress distribution.
Table 2
The counts of element and node of four models
Model
LP-only
LP + CBG
DP + CBG
LP + 3DpbsBRS
Element
366,127
371,747
393,283
371,747
Node
84,237
85,605
92,020
85,605
3DpbsBRS 3D-printed, personalized, biomechanics-specific β-TCP bioceramic rod system; LP lateral plate; CBG cancellous bone grafting; DP double-plate
Specifically, during computational simulation, the LP-only group simulated a bone defect after fracture that was prone to fixation failure (Fig. 4a). In this model of a bone defect treated without grafting that showed failure under 850 N of axial force, specific values could not be calculated. The stress in the LP + CBG group was approximately 2 times higher than that in the DP + CBG and LP + 3DpbsBRS groups. The peak VMS of the LP in the LP + CBG group was 502.4 MPa, compared with 249.3 MPa and 260.2 MPa in the DP + CBG and LP + 3DpbsBRS groups, respectively (Figs. 5 and 7a). In the DP + CBG group, some of the stress was dissipated by the MP, which showed a peak VMS of 249.6 MPa.
We calculated the bending angle of the plate based on the yield strength to evaluate the strength of fixation under axial loading. In the DP + CBG and LP + 3DpbsBRS groups, the bending angles were 1.1° and 1.2°, respectively, which were significantly smaller than the bending angle of 2.3° in the LP + CBG group (Figs. 6 and 7b).

Discussion

Based on a previous study [15] and our results of finite element analysis (Fig. 7), for the treatment of long bone non-union, it is important to solve the issue of bone reconstruction at the site of non-union and compressive stability of the medial femur. The current study is the first known finite element analysis of LP + 3DpbsBRS to explore the possibility of finding an alternative treatment method that can provide similar medial mechanical support as the MP without causing additional soft tissue and vascular damage.
Multiple studies [16, 29] have reported that double-locking compression plates, an advanced management strategy for femoral shaft non-union, have recognized therapeutic effects. Double-locking compression plates provide constant non-union end compression and the opportunity to remove fibrous scar tissue; thus, they are considered effective for treating femoral shaft non-union with bone defects [30]. Furthermore, regarding mechanical stability, double-locking compression plates are considered to be the best existing method for providing medial mechanical support to the femur because they provide 3D fixation [31, 32]. However, the addition of an MP to the medial femur requires reduced blood supply of the fractured bone ends. Also, the MP itself will cause compression of the periosteum, which will continuously affect the blood supply of the periosteum [31]. Thus, the addition of an MP carries the risk of damaging the blood supply of the bone and inhibiting bone regeneration.
Finite element analysis was used to verify our conjecture that the 3DpbsBRS could provide the same medial mechanical support as the MP and that the combined utilization of LP + 3DpbsBRS could provide the same mechanical support as DP fixation, with less soft tissue damage and blood supply disruption. As shown in Fig. 7, the stress on the LP in the LP + CBG group was 2 times higher than that on the LP in the DP + CBG group, and the bending angle of the LP in the LP + CBG group was also twice that of the LP in the DP + CBG group. In the DP model, we found that some of the stress was dispersed by the MP, resulting in a decrease in the bending angle and stress of the LP. Under axial loading, the LP + 3DpbsBRS and DP groups showed similar results in terms of the bending angle and stress distribution of the steel plate. This series of results indicates that the combined application of LP + 3DpbsBRS provides stability, creating an excellent mechanical environment with limited micromotion for non-union repair and, thus, promoting indirect healing of the non-union [33]. Moreover, the 3DpbsBRS in the LP + 3DpbsBRS group dispersed the medial stress during the treatment of long bone non-union, resulting in less stress on the LP and providing greater shear resistance. The entire plate fixation system showed more stability than LP + CBG and stability equivalent to that of the DP + CBG model. Furthermore, in the LP + 3DpbsBRS group, exposing the non-union end allows removal of fibrous scar tissue and filling with the 3DpbsBRS, when entering from the original incision.
An ideal bone graft substitute should provide a 3D structure to support bone cells, stem cells, and bone ingrowth during degradation and treatment. To avoid these problems, β-TCP bioceramics have been widely used in bone regeneration grafts, which have been proposed for the treatment of bone defects and tested in clinical and animal models [2931]. The structure of porous bioceramics promotes the growth of fibrovascular tissue, followed by bone apposition on the porous inner surface. Meanwhile, porous bioceramics have also exhibited superior biocompatibility, osteoconductivity and resorption characteristics and are associated with a low infection risk [32]. Degrading β-TCP could also release large amounts of sulphate (SO42−) and calcium (Ca2+) ions, which are key inorganic salts for forming new bone. β-TCP bioceramics are more biodegradable than hydroxyapatite and can be completely replaced by new bone tissues [32]. In terms of mechanical properties, porous bioceramics tend to perform poorly. β-TCP bioceramics activate cells and signals for the development of new bone and degradation of the implanted material and support the pressure side of the bone. Dense bioceramics have an elastic modulus ranging from 180 MPa to 1.0 GPa and exhibit excellent mechanical properties, with a compressive strength of 10–80 MPa [34]. Recent studies [3436] have revealed that the compressive strength of β-TCP bioceramics, after 4 weeks of biodegradation, was 24–43 MPa, which is more than 11 times that of porous bioceramics (2 MPa). Moreover, the hardened bone and scar bone tissue at the non-union ends could be cleared and the β-TCP bioceramic could guide the necessary nutrients and stem cells to both ends of the non-union for repair. Dense bioceramics provide immediate structural continuity at the non-union site and early postoperative mechanical support at the site of the non-union while protecting the structure of the porous bioceramic so that the porous bioceramic can continue to induce bone formation. Osteogenesis and biodegradation occur simultaneously, and new bone formation is associated with increased mechanical properties until permanent biomechanical support is achieved. Importantly, considering that personalized and precision medicine should always be the most effective treatment for individual patients [37], personalized treatment strategies, such as the 3DpbsBRS, may be a clinical solution for patients with femoral shaft non-union with bone defects. In addition, the 3DpbsBRS can be combined with bioactive molecules, stem cells and exosomesin future research to potentially yield better regenerative functional and therapeutic results [3840].
Furthermore, this study offers a novel solution; for other types of bone defects at various fracture sites, finite element analysis based on a mechanical model can help produce a personalized and precise 3DpbsBRS. In these multitudinous scenarios, the 3DpbsBRS can not only serve as a biological substitute for bone but also provide 3D support to reduce the use of additional plates, enhance the therapeutic effect and relieve the financial burden of patients.
Of course, our research has its limitations because it is a study based on finite element simulation with some reasonable simplifications. We are now pushing forward with relevant animal experiments as our follow-up research, and we hope to present more evidence to prove that our new method has a good prospect in the future.

Conclusion

The 3DpbsBRS in the LP + 3DpbsBRS group could replace the MP in the DP + CBG group by providing similar medial mechanical support. Furthermore, avoiding the use of an MP provides better protection of the soft tissue and vasculature. The 3DpbsBRS is expected to provide a personalized clinical solution for individual patients in various situations, based on the idea that the combination of predictive biomechanical computation and 3D printing technology could provide personalized mechanical support, reduced plate use, and better protection of tissue and blood vessels.

Acknowledgements

We are very grateful to the engineer, Shi Zhan, in our institute, and the biomechanics specialist, Hai Hu, for their technical support and consultation.
This study has been approved by the Independent Ethics Committee of the Evaluation Committee of Shanghai Sixth People’s Hospital, and the written informed consent of volunteers has been obtained.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Agarwal-Harding KJ, Meara JG, Greenberg SL, Hagander LE, Zurakowski D, Dyer GS. Estimating the global incidence of femoral fracture from road traffic collisions: a literature review. J Bone Joint Surg Am. 2015;97(6):e31.PubMedCrossRef Agarwal-Harding KJ, Meara JG, Greenberg SL, Hagander LE, Zurakowski D, Dyer GS. Estimating the global incidence of femoral fracture from road traffic collisions: a literature review. J Bone Joint Surg Am. 2015;97(6):e31.PubMedCrossRef
2.
Zurück zum Zitat Tzioupis C, Giannoudis PV. Prevalence of long-bone non-unions. Injury. 2007;38(Suppl 2):S3–9.PubMedCrossRef Tzioupis C, Giannoudis PV. Prevalence of long-bone non-unions. Injury. 2007;38(Suppl 2):S3–9.PubMedCrossRef
3.
Zurück zum Zitat van Griensven M. Preclinical testing of drug delivery systems to bone. Adv Drug Deliv Rev. 2015;94:151–64.PubMedCrossRef van Griensven M. Preclinical testing of drug delivery systems to bone. Adv Drug Deliv Rev. 2015;94:151–64.PubMedCrossRef
4.
Zurück zum Zitat Zhang H, Li J, Zhou J, Li L, Hao M, Wang K, Xu G, Li C, Zhang W, Tang P. Finite element analysis of different double-plate angles in the treatment of the femoral shaft nonunion with no cortical support opposite the primary lateral plate. Biomed Res Int. 2018;2018:3267107.PubMedPubMedCentral Zhang H, Li J, Zhou J, Li L, Hao M, Wang K, Xu G, Li C, Zhang W, Tang P. Finite element analysis of different double-plate angles in the treatment of the femoral shaft nonunion with no cortical support opposite the primary lateral plate. Biomed Res Int. 2018;2018:3267107.PubMedPubMedCentral
5.
Zurück zum Zitat Bozic KJ, Rosenberg AG, Huckman RS, Herndon JH. Economic evaluation in orthopaedics. J Bone Joint Surg-Am Vol. 2003;85A(1):129–42.CrossRef Bozic KJ, Rosenberg AG, Huckman RS, Herndon JH. Economic evaluation in orthopaedics. J Bone Joint Surg-Am Vol. 2003;85A(1):129–42.CrossRef
6.
Zurück zum Zitat Hak DJ, Fitzpatrick D, Bishop JA, Marsh JL, Tilp S, Schnettler R, Simpson H, Alt V. Delayed union and nonunions: epidemiology, clinical issues, and financial aspects. Injury. 2014;45(Suppl 2):S3–7.PubMedCrossRef Hak DJ, Fitzpatrick D, Bishop JA, Marsh JL, Tilp S, Schnettler R, Simpson H, Alt V. Delayed union and nonunions: epidemiology, clinical issues, and financial aspects. Injury. 2014;45(Suppl 2):S3–7.PubMedCrossRef
7.
Zurück zum Zitat Rupp M, Biehl C, Budak M, Thormann U, Heiss C, Alt V. Diaphyseal long bone nonunions - types, aetiology, economics, and treatment recommendations. Int Orthop. 2018;42(2):247–58.PubMedCrossRef Rupp M, Biehl C, Budak M, Thormann U, Heiss C, Alt V. Diaphyseal long bone nonunions - types, aetiology, economics, and treatment recommendations. Int Orthop. 2018;42(2):247–58.PubMedCrossRef
8.
Zurück zum Zitat Weber BG, Cech O. Pseudoarthrosis: pathology biomechanics therapy results berne; 1976. Weber BG, Cech O. Pseudoarthrosis: pathology biomechanics therapy results berne; 1976.
9.
Zurück zum Zitat Gelalis ID, Politis AN, Arnaoutoglou CM, Korompilias AV, Pakos EE, Vekris MD, Karageorgos A, Xenakis TA. Diagnostic and treatment modalities in nonunions of the femoral shaft: a review. Injury. 2012;43(7):980–8.PubMedCrossRef Gelalis ID, Politis AN, Arnaoutoglou CM, Korompilias AV, Pakos EE, Vekris MD, Karageorgos A, Xenakis TA. Diagnostic and treatment modalities in nonunions of the femoral shaft: a review. Injury. 2012;43(7):980–8.PubMedCrossRef
10.
11.
Zurück zum Zitat Utvag SE, Grundnes O, Reikeras O. Graded exchange reaming and nailing of non-unions. Strength and mineralisation in rat femoral bone. Arch Orthop Trauma Surg. 1998;118(1–2):1–6.PubMed Utvag SE, Grundnes O, Reikeras O. Graded exchange reaming and nailing of non-unions. Strength and mineralisation in rat femoral bone. Arch Orthop Trauma Surg. 1998;118(1–2):1–6.PubMed
12.
Zurück zum Zitat Lin CJ, Chiang CC, Wu PK, Chen CF, Huang CK, Su AW, Chen WM, Liu CL, Chen TH. Effectiveness of plate augmentation for femoral shaft nonunion after nailing. J Chin Med Assoc. 2012;75(8):396–401.PubMedCrossRef Lin CJ, Chiang CC, Wu PK, Chen CF, Huang CK, Su AW, Chen WM, Liu CL, Chen TH. Effectiveness of plate augmentation for femoral shaft nonunion after nailing. J Chin Med Assoc. 2012;75(8):396–401.PubMedCrossRef
13.
Zurück zum Zitat Jones CB, Mayo KA. Nonunion treatment: iliac crest bone graft techniques. J Orthop Trauma. 2005;19(10 Suppl):S11–3.PubMedCrossRef Jones CB, Mayo KA. Nonunion treatment: iliac crest bone graft techniques. J Orthop Trauma. 2005;19(10 Suppl):S11–3.PubMedCrossRef
14.
Zurück zum Zitat Eberle S, Gerber C, von Oldenburg G, Hungerer S, Augat P. Type of hip fracture determines load share in intramedullary osteosynthesis. Clin Orthop Relat Res. 2009;467(8):1972–80.PubMedPubMedCentralCrossRef Eberle S, Gerber C, von Oldenburg G, Hungerer S, Augat P. Type of hip fracture determines load share in intramedullary osteosynthesis. Clin Orthop Relat Res. 2009;467(8):1972–80.PubMedPubMedCentralCrossRef
15.
Zurück zum Zitat Rizzo E, Ghisellini F, Cordey J, Wahl D, Perren S, Cannas M. Biomechanical behaviour at the distal third of the femur: possible use of a medial metaphyseal plate. Injury. 1998;29(6):451–6.PubMedCrossRef Rizzo E, Ghisellini F, Cordey J, Wahl D, Perren S, Cannas M. Biomechanical behaviour at the distal third of the femur: possible use of a medial metaphyseal plate. Injury. 1998;29(6):451–6.PubMedCrossRef
16.
Zurück zum Zitat Steinberg EL, Elis J, Steinberg Y, Salai M, Ben-Tov T. A double-plating approach to distal femur fracture: a clinical study. Injury. 2017;48(10):2260–5.PubMedCrossRef Steinberg EL, Elis J, Steinberg Y, Salai M, Ben-Tov T. A double-plating approach to distal femur fracture: a clinical study. Injury. 2017;48(10):2260–5.PubMedCrossRef
17.
Zurück zum Zitat Schütz M, Müller M, Regazzoni P, Höntzsch D, Krettek C, Van der Werken C, Haas N. Use of the less invasive stabilization system (LISS) in patients with distal femoral (AO33) fractures: a prospective multicenter study. Arch Orthop Trauma Surg. 2005;125(2):102–8.PubMedCrossRef Schütz M, Müller M, Regazzoni P, Höntzsch D, Krettek C, Van der Werken C, Haas N. Use of the less invasive stabilization system (LISS) in patients with distal femoral (AO33) fractures: a prospective multicenter study. Arch Orthop Trauma Surg. 2005;125(2):102–8.PubMedCrossRef
18.
Zurück zum Zitat Kregor PJ, Hughes JL, Cole PA. Fixation of distal femoral fractures above total knee arthroplasty utilizing the Less Invasive Stabilization System (L.I.S.S.). Injury. 2001;32(Suppl 3):Sc64–75.PubMedCrossRef Kregor PJ, Hughes JL, Cole PA. Fixation of distal femoral fractures above total knee arthroplasty utilizing the Less Invasive Stabilization System (L.I.S.S.). Injury. 2001;32(Suppl 3):Sc64–75.PubMedCrossRef
19.
Zurück zum Zitat Sponer P, Filip S, Kucera T, Brtkova J, Urban K, Palicka V, Koci Z, Syka M, Bezrouk A, Sykova E. Utilizing autologous multipotent Mesenchymal stromal cells and beta-Tricalcium phosphate scaffold in human bone defects: a prospective, Controlled Feasibility Trial. BioMed Res Int. 2016;2016:2076061.PubMedPubMedCentralCrossRef Sponer P, Filip S, Kucera T, Brtkova J, Urban K, Palicka V, Koci Z, Syka M, Bezrouk A, Sykova E. Utilizing autologous multipotent Mesenchymal stromal cells and beta-Tricalcium phosphate scaffold in human bone defects: a prospective, Controlled Feasibility Trial. BioMed Res Int. 2016;2016:2076061.PubMedPubMedCentralCrossRef
20.
Zurück zum Zitat Sotome S, Ae K, Okawa A, Ishizuki M, Morioka H, Matsumoto S, Nakamura T, Abe S, Beppu Y, Shinomiya K. Efficacy and safety of porous hydroxyapatite/type 1 collagen composite implantation for bone regeneration: a randomized controlled study. Journal of orthopaedic science : official journal of the Japanese Orthopaedic Association. 2016;21(3):373–80.CrossRef Sotome S, Ae K, Okawa A, Ishizuki M, Morioka H, Matsumoto S, Nakamura T, Abe S, Beppu Y, Shinomiya K. Efficacy and safety of porous hydroxyapatite/type 1 collagen composite implantation for bone regeneration: a randomized controlled study. Journal of orthopaedic science : official journal of the Japanese Orthopaedic Association. 2016;21(3):373–80.CrossRef
21.
Zurück zum Zitat Blom AW, Wylde V, Livesey C, Whitehouse MR, Eastaugh-Waring S, Bannister GC, Learmonth ID. Impaction bone grafting of the acetabulum at hip revision using a mix of bone chips and a biphasic porous ceramic bone graft substitute. Acta Orthop. 2009;80(2):150–4.PubMedPubMedCentralCrossRef Blom AW, Wylde V, Livesey C, Whitehouse MR, Eastaugh-Waring S, Bannister GC, Learmonth ID. Impaction bone grafting of the acetabulum at hip revision using a mix of bone chips and a biphasic porous ceramic bone graft substitute. Acta Orthop. 2009;80(2):150–4.PubMedPubMedCentralCrossRef
22.
Zurück zum Zitat Whitehouse MR, Dacombe PJ, Webb JC, Blom AW. Impaction grafting of the acetabulum with ceramic bone graft substitute mixed with femoral head allograft: high survivorship in 43 patients with a median follow-up of 7 years: a follow-up report. Acta Orthop. 2013;84(4):365–70.PubMedPubMedCentralCrossRef Whitehouse MR, Dacombe PJ, Webb JC, Blom AW. Impaction grafting of the acetabulum with ceramic bone graft substitute mixed with femoral head allograft: high survivorship in 43 patients with a median follow-up of 7 years: a follow-up report. Acta Orthop. 2013;84(4):365–70.PubMedPubMedCentralCrossRef
23.
Zurück zum Zitat Grassi L, Väänänen SP, Amin Yavari S, Weinans H, Jurvelin JS, Zadpoor AA, Isaksson H. Experimental validation of finite element model for proximal composite femur using optical measurements. J Mech Behav Biomed Mater. 2013;21:86–94.PubMedCrossRef Grassi L, Väänänen SP, Amin Yavari S, Weinans H, Jurvelin JS, Zadpoor AA, Isaksson H. Experimental validation of finite element model for proximal composite femur using optical measurements. J Mech Behav Biomed Mater. 2013;21:86–94.PubMedCrossRef
24.
Zurück zum Zitat Eberle S, Gerber C, von Oldenburg G, Hogel F, Augat P. A biomechanical evaluation of orthopaedic implants for hip fractures by finite element analysis and in-vitro tests. Proc Inst Mech Eng H J Eng Med. 2010;224(10):1141–52.CrossRef Eberle S, Gerber C, von Oldenburg G, Hogel F, Augat P. A biomechanical evaluation of orthopaedic implants for hip fractures by finite element analysis and in-vitro tests. Proc Inst Mech Eng H J Eng Med. 2010;224(10):1141–52.CrossRef
25.
Zurück zum Zitat Nuño N, Amabili M, Groppetti R, Rossi A. Static coefficient of friction between Ti-6Al-4V and PMMA for cemented hip and knee implants. J Biomed Mater Res. 2002;59(1):191–200.PubMedCrossRef Nuño N, Amabili M, Groppetti R, Rossi A. Static coefficient of friction between Ti-6Al-4V and PMMA for cemented hip and knee implants. J Biomed Mater Res. 2002;59(1):191–200.PubMedCrossRef
26.
Zurück zum Zitat Goffin JM, Pankaj P, Simpson AH. The importance of lag screw position for the stabilization of trochanteric fractures with a sliding hip screw: a subject-specific finite element study. J Orthop Res. 2013;31(4):596–600.PubMedCrossRef Goffin JM, Pankaj P, Simpson AH. The importance of lag screw position for the stabilization of trochanteric fractures with a sliding hip screw: a subject-specific finite element study. J Orthop Res. 2013;31(4):596–600.PubMedCrossRef
27.
Zurück zum Zitat Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34(7):859–71.PubMedCrossRef Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34(7):859–71.PubMedCrossRef
28.
Zurück zum Zitat Ramlee MH, Kadir MRA, Murali MR, Kamarul T. Biomechanical evaluation of two commonly used external fixators in the treatment of open subtalar dislocation--a finite element analysis. Med Eng Phys. 2014;36(10):1358–66.PubMedCrossRef Ramlee MH, Kadir MRA, Murali MR, Kamarul T. Biomechanical evaluation of two commonly used external fixators in the treatment of open subtalar dislocation--a finite element analysis. Med Eng Phys. 2014;36(10):1358–66.PubMedCrossRef
29.
Zurück zum Zitat Peng Y, Ji X, Zhang L, Tang P. Double locking plate fixation for femoral shaft nonunion. Eur J Orthop Surg Traumatol. 2016;26(5):501–7.PubMedCrossRef Peng Y, Ji X, Zhang L, Tang P. Double locking plate fixation for femoral shaft nonunion. Eur J Orthop Surg Traumatol. 2016;26(5):501–7.PubMedCrossRef
30.
Zurück zum Zitat Jiang Y, Guo YF, Meng YK, Zhu L, Chen AM. A report of a novel technique: the comprehensive fibular autograft with double metal locking plate fixation (cFALP) for refractory post-operative diaphyseal femur fracture non-union treatment. Injury. 2016;47(10):2307–11.PubMedCrossRef Jiang Y, Guo YF, Meng YK, Zhu L, Chen AM. A report of a novel technique: the comprehensive fibular autograft with double metal locking plate fixation (cFALP) for refractory post-operative diaphyseal femur fracture non-union treatment. Injury. 2016;47(10):2307–11.PubMedCrossRef
31.
Zurück zum Zitat Maimaitiyiming A, Amat A, Rehei A, Tusongjiang M, Li C. Treatment of the femoral shaft nonunion with double plate fixation and bone grafting: a case series of 14 patients. Injury. 2015;46(6):1102–7.PubMedCrossRef Maimaitiyiming A, Amat A, Rehei A, Tusongjiang M, Li C. Treatment of the femoral shaft nonunion with double plate fixation and bone grafting: a case series of 14 patients. Injury. 2015;46(6):1102–7.PubMedCrossRef
32.
Zurück zum Zitat Zhang W, Zhang Z, Li J, Zhang L, Chen H, Tang P. Clinical outcomes of femoral shaft non-union: dual plating versus exchange nailing with augmentation plating. J Orthop Surg Res. 2018;13(1):295.PubMedPubMedCentralCrossRef Zhang W, Zhang Z, Li J, Zhang L, Chen H, Tang P. Clinical outcomes of femoral shaft non-union: dual plating versus exchange nailing with augmentation plating. J Orthop Surg Res. 2018;13(1):295.PubMedPubMedCentralCrossRef
33.
Zurück zum Zitat Green E, Lubahn JD, Evans J. Risk factors, treatment, and outcomes associated with nonunion of the midshaft humerus fracture. J Surg Orthop Adv. 2005;14(2):64–72.PubMed Green E, Lubahn JD, Evans J. Risk factors, treatment, and outcomes associated with nonunion of the midshaft humerus fracture. J Surg Orthop Adv. 2005;14(2):64–72.PubMed
34.
Zurück zum Zitat Lu Y, Lu X, Li M, Chen X, Liu Y, Feng X, Yu J, Zhang C, Niu D, Wang S, et al. Minimally invasive treatment for osteonecrosis of the femoral head with angioconductive bioceramic rod. Int Orthop. 2018;42(7):1567–73.PubMedCrossRef Lu Y, Lu X, Li M, Chen X, Liu Y, Feng X, Yu J, Zhang C, Niu D, Wang S, et al. Minimally invasive treatment for osteonecrosis of the femoral head with angioconductive bioceramic rod. Int Orthop. 2018;42(7):1567–73.PubMedCrossRef
35.
Zurück zum Zitat Walsh WR, Vizesi F, Michael D, Auld J, Langdown A, Oliver R, Yu Y, Irie H, Bruce W. Beta-TCP bone graft substitutes in a bilateral rabbit tibial defect model. Biomaterials. 2008;29(3):266–71.PubMedCrossRef Walsh WR, Vizesi F, Michael D, Auld J, Langdown A, Oliver R, Yu Y, Irie H, Bruce W. Beta-TCP bone graft substitutes in a bilateral rabbit tibial defect model. Biomaterials. 2008;29(3):266–71.PubMedCrossRef
36.
Zurück zum Zitat Zhang F, Chang J, Lu J, Lin K, Ning C. Bioinspired structure of bioceramics for bone regeneration in load-bearing sites. Acta Biomater. 2007;3(6):896–904.PubMedCrossRef Zhang F, Chang J, Lu J, Lin K, Ning C. Bioinspired structure of bioceramics for bone regeneration in load-bearing sites. Acta Biomater. 2007;3(6):896–904.PubMedCrossRef
37.
Zurück zum Zitat König IR, Fuchs O, Hansen G, von Mutius E, Kopp MV. What is precision medicine? Eur Respir J. 2017;50:4.CrossRef König IR, Fuchs O, Hansen G, von Mutius E, Kopp MV. What is precision medicine? Eur Respir J. 2017;50:4.CrossRef
38.
Zurück zum Zitat Tao SC, Guo SC, Zhang CQ. Modularized Extracellular Vesicles: The Dawn of Prospective Personalized and Precision Medicine. Adv Sci (Weinheim, Baden-Wurttemberg, Germany). 2018;5(2):1700449. Tao SC, Guo SC, Zhang CQ. Modularized Extracellular Vesicles: The Dawn of Prospective Personalized and Precision Medicine. Adv Sci (Weinheim, Baden-Wurttemberg, Germany). 2018;5(2):1700449.
39.
Zurück zum Zitat Tao SC, Guo SC. Extracellular vesicles in bone: "dogrobbers" in the "eternal battle field". Cell Communication Signaling. 2019;17(1):6.PubMedCrossRef Tao SC, Guo SC. Extracellular vesicles in bone: "dogrobbers" in the "eternal battle field". Cell Communication Signaling. 2019;17(1):6.PubMedCrossRef
40.
Zurück zum Zitat Lu J, Wang QY, Sheng JG. Exosomes in the repair of bone defects: next-generation therapeutic tools for the treatment of nonunion. Biomed Res Int. 2019;2019:1983131.PubMedPubMedCentral Lu J, Wang QY, Sheng JG. Exosomes in the repair of bone defects: next-generation therapeutic tools for the treatment of nonunion. Biomed Res Int. 2019;2019:1983131.PubMedPubMedCentral
Metadaten
Titel
A 3D-printed, personalized, biomechanics-specific beta-tricalcium phosphate bioceramic rod system: personalized treatment strategy for patients with femoral shaft non-union based on finite element analysis
verfasst von
Jian Lu
Qi-Yang Wang
Jia-Gen Sheng
Shang-Chun Guo
Shi-Cong Tao
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Musculoskeletal Disorders / Ausgabe 1/2020
Elektronische ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-020-03465-1