A comparison of three different tonometric methods in the measurement of intraocular pressure in the pediatric age group

verfasst von: Dilek Uzlu, Nurettin Akyol, Adem Türk, Yavuz Oruç

Erschienen in: International Ophthalmology | Ausgabe 8/2020

Einloggen, um Zugang zu erhalten

Abstract

Introduction and aim

To compare intraocular pressure (IOP) measurements obtained with non-contact tonometry (NCT), Icare rebound tonometry (IRT) and Goldmann applanation tonometry (GAT) in pediatric cases and to examine the effect of topical anesthesia on measurements obtained using IRT.

Materials and methods

Pediatric cases in a healthy general condition and with ophthalmic examination findings within normal limits were included in the study. IOP measurements were taken with NCT and IRT devices in all cases. Topical anesthesia was subsequently applied, and IOP measurements were then repeated using IRT and GAT tonometer devices. Differences between measurements were subjected to statistical analysis.

Results

One hundred ten eyes of 55 patients, 27 male, with a mean age of 11.44 ± 2.31 years (7–17) were included. Mean pre-anesthesia IOP values were 16.47 ± 2.89 mmHg with NCT and 17.49 ± 2.57 mmHg with IRT. Mean IOP values after topical anesthesia were 16.91 ± 2.17 mmHg with IRT and 15.51 ± 2.41 mmHg with GAT. IOP measurement values obtained with all three devices exhibited positive correlation with central corneal thickness values. Statistically significant correlation was present in terms of IOP measurement values between all three devices. However, IOP values obtained with the three devices exhibited statistically significant differences. The application of topical anesthesia caused a statistically significant decrease in IRT measurements; however, statistically significant this small change (0.58 mmHg) in recorded IOP would not be considered clinically significant.

Conclusion

IOP measurements obtained with NCT, IRT and GAT devices in the pediatric age group correlate with one another, but differ from one another in a statistically significant manner. Application of topical anesthesia affects IRT measurements; however, these small changes would not be considered clinically significant.
Literatur
1.
Zurück zum Zitat Heijl A, Leske MC, Hyman L et al (2002) Reduction of intraocular pressure and glaucoma progression: results from the early manifest glaucoma trial. Arch Ophthalmol 120(10):1268–1279PubMedCrossRef Heijl A, Leske MC, Hyman L et al (2002) Reduction of intraocular pressure and glaucoma progression: results from the early manifest glaucoma trial. Arch Ophthalmol 120(10):1268–1279PubMedCrossRef
2.
Zurück zum Zitat Leske MC, Heijl A, Hussein M et al (2003) Factors for glaucoma progression and the effect of treatment: the early manifest glaucoma trial. Arch Ophthalmol 121(1):48–56PubMedCrossRef Leske MC, Heijl A, Hussein M et al (2003) Factors for glaucoma progression and the effect of treatment: the early manifest glaucoma trial. Arch Ophthalmol 121(1):48–56PubMedCrossRef
3.
Zurück zum Zitat Group CN-TGS (1998) Comparison of glaucomatous progression between untreated patients with normal-tension glaucoma and patients with therapeutically reduced intraocular pressures. Am J Ophthalmol 126(4):487–497CrossRef Group CN-TGS (1998) Comparison of glaucomatous progression between untreated patients with normal-tension glaucoma and patients with therapeutically reduced intraocular pressures. Am J Ophthalmol 126(4):487–497CrossRef
4.
Zurück zum Zitat Kao SF, Lichter PR, Bergstrom TJ et al (1987) Clinical comparison of the OculabTono-Pen to the Goldmann applanation tonometer. Ophthalmology 94:15411544CrossRef Kao SF, Lichter PR, Bergstrom TJ et al (1987) Clinical comparison of the OculabTono-Pen to the Goldmann applanation tonometer. Ophthalmology 94:15411544CrossRef
5.
Zurück zum Zitat Eraslan M, Çerman E, Sümmen S (2017) Comparison of intraocular pressure measurements in healthy pediatric patients using three types of tonometers. Turk J Ophthalmol 47:1PubMedPubMedCentralCrossRef Eraslan M, Çerman E, Sümmen S (2017) Comparison of intraocular pressure measurements in healthy pediatric patients using three types of tonometers. Turk J Ophthalmol 47:1PubMedPubMedCentralCrossRef
6.
Zurück zum Zitat Cook JA, Botello AP, Elders A et al (2012) Systematic review of the agreement of tonometers with Goldmann applanation tonometry. Ophthalmology 119:1552–1557PubMedCrossRef Cook JA, Botello AP, Elders A et al (2012) Systematic review of the agreement of tonometers with Goldmann applanation tonometry. Ophthalmology 119:1552–1557PubMedCrossRef
7.
Zurück zum Zitat Tamçelik N, Atalay E, Cicik E et al (2015) Comparability of Icare pro rebound tonometer with Goldmann applanation and noncontact tonometer in a wide range of intraocular pressure and central corneal thickness. Ophthalmic Res 54(1):18–25PubMedCrossRef Tamçelik N, Atalay E, Cicik E et al (2015) Comparability of Icare pro rebound tonometer with Goldmann applanation and noncontact tonometer in a wide range of intraocular pressure and central corneal thickness. Ophthalmic Res 54(1):18–25PubMedCrossRef
8.
Zurück zum Zitat Recep OF, Hasiripi H, Çağil N et al (2001) Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry. J Cataract Refract Surg 27(11):1787–1791PubMedCrossRef Recep OF, Hasiripi H, Çağil N et al (2001) Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry. J Cataract Refract Surg 27(11):1787–1791PubMedCrossRef
9.
Zurück zum Zitat Farhood QK (2013) Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer. ClinmOphthalmol 7:23–27 Farhood QK (2013) Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer. ClinmOphthalmol 7:23–27
10.
Zurück zum Zitat García-Resúa C, Pena-Verdeal H, Miñones M et al (2013) Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in health yeyes. Clin Exp Optom 96:286–294PubMedCrossRef García-Resúa C, Pena-Verdeal H, Miñones M et al (2013) Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in health yeyes. Clin Exp Optom 96:286–294PubMedCrossRef
11.
Zurück zum Zitat Ogbuehi KC (2006) Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer. Clin Exp Optom 89:310–314PubMedCrossRef Ogbuehi KC (2006) Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer. Clin Exp Optom 89:310–314PubMedCrossRef
12.
Zurück zum Zitat Feng CS, Jin KW, Yi K et al (2015) Comparison of intraocular pressure measurements obtained by rebound, noncontact, and Goldmann applanation tonometry in children. Am J Ophthalmol 160:937–943PubMedCrossRef Feng CS, Jin KW, Yi K et al (2015) Comparison of intraocular pressure measurements obtained by rebound, noncontact, and Goldmann applanation tonometry in children. Am J Ophthalmol 160:937–943PubMedCrossRef
13.
Zurück zum Zitat Flemmons MS, Hsiao YC, Dzau J et al (2011) Icare rebound tonometry in children with known and suspected glaucoma. J AAPOS 15(2):153–157PubMedCrossRef Flemmons MS, Hsiao YC, Dzau J et al (2011) Icare rebound tonometry in children with known and suspected glaucoma. J AAPOS 15(2):153–157PubMedCrossRef
14.
Zurück zum Zitat Sahin A, Basmak H, Niyaz L et al (2007) Reproducibility and tolerability of the ICare rebound tonometer in school children. J Glaucoma 16(2):185–188PubMedCrossRef Sahin A, Basmak H, Niyaz L et al (2007) Reproducibility and tolerability of the ICare rebound tonometer in school children. J Glaucoma 16(2):185–188PubMedCrossRef
15.
Zurück zum Zitat Pakrou N, Gray T (2008) Clinical comparison of the Icare tonometer and Goldmann applanation tonometry. J Glaucoma 17:43–47PubMedCrossRef Pakrou N, Gray T (2008) Clinical comparison of the Icare tonometer and Goldmann applanation tonometry. J Glaucoma 17:43–47PubMedCrossRef
16.
Zurück zum Zitat Brusini P, Salvetat ML, Zeppieri M (2006) Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 15:213–217PubMedCrossRef Brusini P, Salvetat ML, Zeppieri M (2006) Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 15:213–217PubMedCrossRef
17.
Zurück zum Zitat Iliev ME, Goldblum D, Katsoulis K et al (2006) Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 90(7):833–835PubMedPubMedCentralCrossRef Iliev ME, Goldblum D, Katsoulis K et al (2006) Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 90(7):833–835PubMedPubMedCentralCrossRef
18.
Zurück zum Zitat Almubrad TM, Ogbuehi KC (2007) Clinical investigation of the effect of topical anesthesia on intraocular pressure. Clin Ophthalmol 1:305–309PubMedPubMedCentral Almubrad TM, Ogbuehi KC (2007) Clinical investigation of the effect of topical anesthesia on intraocular pressure. Clin Ophthalmol 1:305–309PubMedPubMedCentral
19.
Zurück zum Zitat Montero JA, Ruiz-Moreno JM, Fernandez-Munoz M et al (2008) Effect of topical anesthetics on intraocular pressure and pachymetry. Eur J Ophthalmol 18(5):748–750PubMedCrossRef Montero JA, Ruiz-Moreno JM, Fernandez-Munoz M et al (2008) Effect of topical anesthetics on intraocular pressure and pachymetry. Eur J Ophthalmol 18(5):748–750PubMedCrossRef
20.
Zurück zum Zitat Dosunmu EO, Marcus I, Tung I et al (2014) The effect of repeated measurements and the use of topical anesthetic on rebound tonometry values in children. J AAPOS 18(6):619–621PubMedCrossRef Dosunmu EO, Marcus I, Tung I et al (2014) The effect of repeated measurements and the use of topical anesthetic on rebound tonometry values in children. J AAPOS 18(6):619–621PubMedCrossRef
21.
Zurück zum Zitat Yılmaz I, Altan C, Aygıt DE et al (2014) Comparison of three methods of tonometry in normal subjects: goldmann applanationmtonometer, non-contact airpuff tonometer and Tono-PenXl. Clin Ophthalmol 8:1069–1074PubMedPubMedCentralCrossRef Yılmaz I, Altan C, Aygıt DE et al (2014) Comparison of three methods of tonometry in normal subjects: goldmann applanationmtonometer, non-contact airpuff tonometer and Tono-PenXl. Clin Ophthalmol 8:1069–1074PubMedPubMedCentralCrossRef
22.
Zurück zum Zitat Salim S, Linn DJ, Echols JR II et al (2009) Comparison of intraocular pressure measurements with the portable PT100 noncontact tonometer and goldmann applanation tonometry. Clin Ophthalmol 3:341–344PubMedPubMedCentral Salim S, Linn DJ, Echols JR II et al (2009) Comparison of intraocular pressure measurements with the portable PT100 noncontact tonometer and goldmann applanation tonometry. Clin Ophthalmol 3:341–344PubMedPubMedCentral
23.
Zurück zum Zitat Gao F, Lıu X, Zhao Q et al (2017) Comparison of the icare rebound tonometer and the Goldmann applanation tonometer. Exp Therap Med 13:1912–1916CrossRef Gao F, Lıu X, Zhao Q et al (2017) Comparison of the icare rebound tonometer and the Goldmann applanation tonometer. Exp Therap Med 13:1912–1916CrossRef
24.
Zurück zum Zitat Kato Y, Nakakura S, Matsuo N et al (2018) Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects. Int Ophthalmol 38(2):687–696PubMedCrossRef Kato Y, Nakakura S, Matsuo N et al (2018) Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects. Int Ophthalmol 38(2):687–696PubMedCrossRef
25.
Zurück zum Zitat Güler M, Bilak Ş, Bilgin B et al (2015) Comparison of intraocular pressure measurements obtained byIcare PRO rebound tonometer, Tomey FT-1000 noncontact tonometer, and Goldmann applanationtonometer in healthy subjects. J Glaucoma 24(8):613–618PubMedCrossRef Güler M, Bilak Ş, Bilgin B et al (2015) Comparison of intraocular pressure measurements obtained byIcare PRO rebound tonometer, Tomey FT-1000 noncontact tonometer, and Goldmann applanationtonometer in healthy subjects. J Glaucoma 24(8):613–618PubMedCrossRef
26.
Zurück zum Zitat Grigorian F, Grigorian AP, Li A et al (2015) Comparison of the Icare rebound tonometry with the Goldmann applanation tonometry in a pediatric population. J AAPOS 19(6):572–574PubMedCrossRef Grigorian F, Grigorian AP, Li A et al (2015) Comparison of the Icare rebound tonometry with the Goldmann applanation tonometry in a pediatric population. J AAPOS 19(6):572–574PubMedCrossRef
27.
Zurück zum Zitat Martinez-de-la-Casa JM, Jimenez-Santos M, Saenz-Frances F et al (2011) Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice. Acta Ophthalmol 89(7):676–680PubMedCrossRef Martinez-de-la-Casa JM, Jimenez-Santos M, Saenz-Frances F et al (2011) Performance of the rebound, noncontact and Goldmann applanation tonometers in routine clinical practice. Acta Ophthalmol 89(7):676–680PubMedCrossRef
28.
Zurück zum Zitat Rao A, Kumar M, Prakash B et al (2014) Relationship of central corneal thickness and intraocular pressure by icare rebound tonometer. J Glaucoma 23(6):380–384PubMedCrossRef Rao A, Kumar M, Prakash B et al (2014) Relationship of central corneal thickness and intraocular pressure by icare rebound tonometer. J Glaucoma 23(6):380–384PubMedCrossRef
29.
Zurück zum Zitat Moreno-Montañés J, Gosende I, Caire J et al (2011) Comparation of the new rebound tonometer IOPen and the Goldmanntonometer, and their relationship to corneal properties. Eye (Lond) 25(1):50–56CrossRef Moreno-Montañés J, Gosende I, Caire J et al (2011) Comparation of the new rebound tonometer IOPen and the Goldmanntonometer, and their relationship to corneal properties. Eye (Lond) 25(1):50–56CrossRef
30.
Zurück zum Zitat Burdová MC, Ferrová K, Filous A et al (2011) Correlation of intraocular pressure measured by applanation tonometry, noncontact tonometry and Tonopen with central corneal thickness. CeskSlov Oftalmol 67(5–6):154–157 Burdová MC, Ferrová K, Filous A et al (2011) Correlation of intraocular pressure measured by applanation tonometry, noncontact tonometry and Tonopen with central corneal thickness. CeskSlov Oftalmol 67(5–6):154–157
Metadaten
Titel
A comparison of three different tonometric methods in the measurement of intraocular pressure in the pediatric age group
verfasst von
Dilek Uzlu
Nurettin Akyol
Adem Türk
Yavuz Oruç
Publikationsdatum
23.04.2020
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 8/2020
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-020-01375-8

Weitere Artikel der Ausgabe 8/2020

International Ophthalmology 8/2020 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.