A controlled comparison of the BacT/ALERT® 3D and VIRTUO™ microbial detection systems

verfasst von: H. Totty, M. Ullery, J. Spontak, J. Viray, M. Adamik, B. Katzin, W. M. Dunne Jr, P. Deol

Erschienen in: European Journal of Clinical Microbiology & Infectious Diseases | Ausgabe 10/2017

Abstract

The performance of the next-generation BacT/ALERT® VIRTUO™ Microbial Detection System (VIRTUO™, bioMérieux Inc., Hazelwood, MO) was compared to the BacT/ALERT® 3D Microbial Detection System (3D, bioMérieux Inc., Durham, NC) using BacT/ALERT® FA Plus (FA Plus), BacT/ALERT® PF Plus (PF Plus), BacT/ALERT® FN Plus (FN Plus), BacT/ALERT® Standard Aerobic (SA), and BacT/ALERT® Standard Anaerobic (SN) blood culture bottles (bioMérieux Inc., Durham, NC). A seeded limit of detection (LoD) study was performed for each bottle type in both systems. The LoD studies demonstrated that both systems were capable of detecting organisms at nearly identical levels [<10 colony-forming units (CFU) per bottle], with no significant difference. Following LoD determination, a seeded study was performed to compare the time to detection (TTD) between the systems using a panel of clinically relevant microorganisms inoculated at or near the LoD with 0, 4, or 10 mL of healthy human blood. VIRTUO™ exhibited a faster TTD by an average of 3.5 h, as well as demonstrated a significantly improved detection rate of 99.9% compared to 98.8% with 3D (p-value <0.05).
Begleitmaterial
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1007/​s10096-017-2994-8) contains supplementary material, which is available to authorized users.

Introduction

Each year in the USA, there are approximately 200,000 documented bloodstream infections [1]. Delayed laboratory diagnosis of bacteremia or fungemia can lead to increased mortality [2]. Therefore, rapid detection of microorganisms in blood and sterile body fluids is essential for optimized patient care. The BacT/ALERT® VIRTUO™ Microbial Detection System is the next generation of BacT/ALERT® instrumentation with an improved user interface capable of providing automated processes that were previously performed manually. The VIRTUO™ system consists of an incubator, agitation mechanism, robotic apparatus for automated loading and unloading of bottles, and a tactile graphical user interface. The versatile and scalable design of VIRTUO™ allows for enhanced bottle capacity with a smaller footprint in the laboratory. Employing the same colorimetric technology used in previous generations of BacT/ALERT® instruments, VIRTUO™ measures the change in pH of the media associated with microbial growth and CO2 production via a colorimetric sensor in each bottle and optically monitors the reflectance of the sensor over time [3]. The system stores and interprets these readings using algorithms embedded in the firmware and/or software. More importantly, VIRTUO™ uses a new algorithm designed to optimize test sensitivity and specificity, while significantly reducing the time to detection (TTD) of microbial growth when paired with BacT/ALERT® FA Plus, PF Plus, FN Plus, Standard Aerobic (SA), or Standard Anaerobic (SN) bottles.
The BacT/ALERT® bottles are each formulated to recover and detect a subset of clinically relevant microorganisms. SA culture bottles are intended for aerobic microorganisms, while the SN bottles contain media and atmospheric conditions designed for anaerobic microorganisms [4, 5]. FA Plus and PF Plus culture bottles contain antimicrobial neutralizing resins as well as media and atmospheric conditions suitable for aerobic and facultative microorganisms [6, 7]. PF Plus bottles recover and detect microorganisms when only a small volume of blood is available [16]. FN Plus culture bottles also employ antimicrobial neutralizing resins, while using media and atmosphere essential for anaerobes [8].
A series of seeded studies directly compared the instrumentation and growth-based detection algorithms of the 3D and VIRTUO™ systems. The limit of detection (LoD) was evaluated using a test set of clinically relevant organisms seeded into bottles to determine the lowest inoculum [colony-forming units (CFU) per bottle] in which a 95% detection rate is obtained. Once the LoD was established, a controlled seeded study was performed at/near the LoD for each test organism with different blood volumes to compare the detection rates and TTD between systems.

Materials and methods

Limit of detection

For each organism–inocula evaluated, 60 bottles were seeded. A single inoculum was prepared for each microorganism and tested in FA Plus, PF Plus, FN Plus, SA, and SN bottles (Table 1). BioBalls™ or colonies from solid media were serially diluted to achieve a target suspension of 10 CFU per mL. Bottles were inoculated with 0.5 mL of the target suspension. Plate counts were conducted for each inoculum before and after bottle inoculation. Thirty bottles were loaded into 3D and 30 were loaded into VIRTUO™. Once bottles flagged positive, they were promptly subcultured to confirm purity. Negative bottles remained in the instrument for 5 days until declared negative, at which time they were subcultured to verify the absence of microorganism.
Table 1
Limit of detection (LoD) test panel by bottle type
Bottle type
Microorganism
Strain
FA Plus and PF Plus
Candida albicans
ATCC 14053
Enterococcus faecalis
NCTC 12697
Escherichia coli
NCTC 12923
Haemophilus influenzae a
ATCC 10211
Pseudomonas aeruginosa
ATCC 9027
Staphylococcus aureus
NCTC 10788
Streptococcus pneumoniae
ATCC 6305
FN Plus
Bacteroides fragilis c
ATCC 25285
Clostridium perfringens
NCTC 8789
Enterococcus faecalis
NCTC 12697
Escherichia coli
NCTC 12923
Staphylococcus aureus
NCTC 10788
Streptococcus pneumoniae
ATCC 6305
SA
Aspergillus brasiliensis
NCPF 2275
Candida albicans
ATCC 14053
Enterococcus faecalis
ATCC 12697
Escherichia coli
NCTC 12923
Haemophilus influenzae a
ATCC 10211
Pseudomonas aeruginosa
ATCC 9027
Staphylococcus aureus
NCTC 10788
Streptococcus pneumoniae b
ATCC 6305
SN
Bacteroides fragilis
ATCC 25285
Clostridium perfringens
NCTC 8789
Enterococcus faecalis
NCTC 12697
Escherichia coli
NCTC 12923
Staphylococcus aureus
NCTC 10788
Streptococcus pneumoniae b
ATCC 6305
aSupplemented with 1 mL human blood
bSupplemented with 1 mL human blood for SA and SN only
cSupplemented with 4 mL human blood for FN Plus only
The LoD was defined for each organism/bottle/system as the lowest inocula that resulted in at least 95% of the bottles being declared positive. Based on a Poisson distribution, in a target inoculum of 3 CFU, approximately 5% of the bottles would not receive a single organism, 42% would receive less than 3 CFU, and 53% would receive 3 CFU or greater.

Comparison of detection systems

Human blood was collected from healthy donors in tubes containing 0.35% sodium polyanethol sulfonate (SPS) in 0.85% sodium chloride and pooled prior to testing [9]. Prior to inoculation, bottles were supplemented at three blood volumes: 0 mL, representing sterile body fluid; 4 mL, representing a pediatric blood sample volume; and 10 mL, representing an adult blood sample volume (Table 2). Uninoculated bottles containing 0, 4, or 10 mL blood served as negative controls. Both systems received nine seeded bottles at three blood volumes, for a total of 27 replicates per microorganism and bottle type.
Table 2
Instrument and algorithm comparison panel by bottle type
Bottle type
Organism
Bottles per organism in 3D
Bottles per organism in VIRTUO™
Blood volume (mL)
FA Plus, PF Plus, and SA
Abiotrophia defectiva
27
27
0a, 4b, 10c
Aggregatibacter actinomycetemcomitans d
Aspergillus fumigatus
Aspergillus brasiliensis
Campylobacter jejuni e
Candida albicans
Candida glabrata
Candida krusei
Cardiobacterium hominis
Corynebacterium jeikeium d
Cryptococcus neoformans
Eikenella corrodens d
Enterobacter aerogenes
Enterococcus faecalis
Escherichia coli
Haemophilus influenzae d
Klebsiella pneumoniae
Listeria monocytogenes
Micrococcus luteus
Neisseria meningitidis d
Proteus vulgaris
Pseudomonas aeruginosa
Salmonella enterica
Serratia marcescens
Staphylococcus aureus
Staphylococcus epidermidis
Stenotrophomonas maltophilia
Streptococcus agalactiae
Streptococcus mitis
Streptococcus pneumoniae
Streptococcus pyogenes
FN Plus and SN
Bacteroides fragilis
Bacteroides thetaiotaomicron
Bacteroides vulgatus
Clostridium perfringens
Clostridium septicum
Eggerthella lenta
Enterococcus faecalis
Escherichia coli
Fusobacterium nucleatum
Parvimonas micra
Peptoniphilus asaccharolyticus
Staphylococcus aureus
Streptococcus pneumoniae
Streptococcus pyogenes
aReflects sterile body fluid sample
bReflects pediatric patient blood volume
cReflects adult patient blood volume
dFastidious organisms not expected to grow in the absence of blood
eMicroaerophilic organism not expected to grow in SA bottle type
Thirty-two aerobic microorganism and 14 anaerobic microorganisms were used to evaluate the performance of the detection systems. Each microorganism was serially diluted to a target inoculum of ≤ 30 CFU/bottle. Plate counts were performed on each inocula before and after each testing event to verify purity and appropriate microorganism density. Bottles were incubated until flagged positive or declared negative after 5 days. All inoculated negative bottles were examined for the presence of viable organisms by subculture. Two positive bottles from each test combination were randomly selected for subculture to verify purity of the inoculated bottles.
All possible differences between 3D and VIRTUO™ TTD were calculated for bottle type, organism, and fill volume using pairwise analysis. Only matched test bottles (bottle type, organism, and blood volume) for which data were available from both culture systems were included in the final comparison analysis. TTD detections were compared between each system at the level of bottle type, organism, and blood volume. Ratios were calculated from TTD results for each microorganism by dividing the TTD from VIRTUO™ by the TTD from 3D. A ratio less than 1 indicates an improvement in TTD on VIRTUO™. A ratio approximately equal to 1 indicates that the TTD on 3D and VIRTUO™ were equivalent. Detection rates between instrument types were compared using Fisher’s exact test.

Results

Limit of detection

A total of 479 seeded BacT/ALERT® bottles were tested in the 3D system. Of these, 473 (98.8%) were declared positive and confirmed by subculture. A total of 1616 seeded culture bottles were tested in the VIRTUO™ system, with 1605 (99.3%) declared positive. All negative bottles were determined to be true-negative results by terminal subculture. Both systems were capable of detecting less than CFU/bottle for all bottle types. There was no difference in the LoD of the VIRTUO™ system for different bottle types when compared to the established LoD of the 3D system (Supplemental Tables 1a–d).

Comparison of time to detection between systems

The VIRTUO™ algorithm provided an overall mean reduction in TTD of approximately 3.48 h (Table 3). FN Plus bottles exhibited the greatest TTD improvement, with an average decrease of 3.90 h. The SA bottle type had the least TTD improvement, with an average decrease of 3.14 h.
Table 3
Time to detection (TTD) and detection rates from BacT/ALERT® 3D and VIRTUO™ by bottle type and blood volume
Bottle type
Blood volume (mL)
3D
VIRTUO™
TTD difference (3D - VIRTUO™)
# (+)/#
Detection rate (%)
# (+)/#
Detection rate (%)
Mean TTD difference (h)
95% CI for mean TTD difference (h)
FA Plus
0
227/239
95.0
233/233
100.0*
3.62
(3.37, 3.86)
4**
271/272
99.6
275/275
100.0
3.94
(3.47, 4.41)
10
273/278
98.2
279/279
100.0*
3.24
(2.96, 3.51)
Combined
771/789
97.7
787/787
100.0*
3.59
(3.39, 3.80)
FN Plus
0
93/93
100.0
91/91
100.0
3.62
(3.14, 4.10)
4
117/121
96.7
122/122
100.0*
3.78
(3.27, 4.29)
10
110/115
95.7
114/114
100.0*
4.28
(3.93, 4.64)
Combined
320/329
97.3
327/327
100.0*
3.90
(3.64, 4.17)
SA
0
215/215
100.0
213/213
100.0
2.35
(2.13, 2.56)
4
276/276
100.0
279/279
100.0
3.47
(3.33, 3.61)
10
270/270
100.0
276/276
100.0
3.38
(3.14, 3.62)
Combined
761/761
100.0
768/768
100.0
3.14
(3.02, 3.25)
SN
0
125/125
100.0
123/123
100.0
4.65
(4.46, 4.84)
4
124/124
100.0
116/117
99.1
4.02
(3.76, 4.29)
10
119/120
99.2
119/120
99.2
2.20
(1.58, 2.82)
Combined
368/369
99.7
358/360
99.4
3.65
(3.42, 3.88)
Overall
2220/2248
98.8
2240/2242
99.9*
3.48
(3.38, 3.58)
*Significant difference between detection rates; Fisher’s exact test p-value <0.05; **represents PF Plus
Based on the comparison ratios (Table 4), all organisms had faster or equivalent TTD with VIRTUO™, except for Stenotrophomonas maltophilia, which generated slower TTD in FA Plus bottles (data not shown). Over half of the Gram-negative organisms tested had an average TTD of less than 24 h in the VIRTUO™ system, with an overall TTD ratio of 0.85. Overall, Gram-positive organisms demonstrated the greatest TTD improvement. Abiotrophia defectiva and Corynebacterium jeikeium exhibited the largest TTD reductions in VIRTUO™, with comparison ratios of 0.66 and 0.65, respectively (Table 4). Large decreases in TTD across both anaerobic bottle types were also observed in VIRTUO™ with Clostridium perfringens and Clostridium septicum (Table 4). Yeasts demonstrated smaller TTD improvements with VIRTUO™. Although the test panel for moulds was limited, the TTD of both Aspergillus species benefited from the VIRTUO™ detection algorithm (Table 4).
Table 4
Overall TTD comparisons between BacT/ALERT® 3D and VIRTUO™ by organism
Group
Organism
3D
VIRTUO™
Comparison ratio (VIRTUO™ TTD/3D TTD)
Mean TTD (h)
Recovery rate (%)
Mean TTD (h)
Recovery rate (%)
Gram-positive
Abiotrophia defectiva
25.7
100.0
17.0
100.0
0.66
Clostridium perfringens
12.1
100.0
9.1
100.0
0.76
Clostridium septicum
16.6
100.0
12.3
100.0
0.74
Corynebacterium jeikeium
50.5
100.0
32.8
100.0
0.65
Eggerthella lenta
35.3
94.4
34.7
96.2
0.98
Enterococcus faecalis
14.2
100.0
11.3
100.0
0.80
Listeria monocytogenes
21.7
100.0
20.2
100.0
0.93
Micrococcus luteus
35.1
100.0
32.5
100.0
0.93
Parvimonas micra
47.0
100.0
41.0
100.0
0.87
Peptoniphilus asaccharolyticus
51.3
100.0
46.9
100.0
0.91
Staphylococcus aureus
16.1
100.0
13.1
100.0
0.81
Staphylococcus epidermidis
19.8
100.0
17.1
100.0
0.86
Streptococcus agalactiae
15.1
100.0
11.6
100.0
0.77
Streptococcus mitis
13.2
100.0
10.2
100.0
0.77
Streptococcus pneumoniae
17.1
100.0
14.5
100.0
0.85
Streptococcus pyogenes
15.3
100.0
12.6
100.0
0.82
Gram-negative
Aggregatibacter actinomycetemcomitans
46.5
100.0
43.4
100.0
0.93
Bacteroides fragilis
34.7
100.0
32.4
100.0
0.94
Bacteroides thetaiotaomicron
42.9
100.0
36.7
100.0
0.86
Bacteroides vulgatus
43.3
100.0
40.7
100.0
0.94
Campylobacter jejuni
46.1
100.0
41.6
100.0
0.90
Cardiobacterium hominis
54.9
80.0
47.9
100.0
0.87
Eikenella corrodens
26.2
100.0
22.5
100.0
0.86
Enterobacter aerogenes
12.9
100.0
10.7
100.0
0.83
Escherichia coli
11.4
100.0
9.0
100.0
0.79
Fusobacterium nucleatum
53.9
84.4
43.7
100.0
0.81
Haemophilus influenzae
20.0
100.0
16.9
100.0
0.85
Klebsiella pneumoniae
12.2
100.0
9.7
100.0
0.80
Neisseria meningitidis
22.3
100.0
19.3
100.0
0.86
Proteus vulgaris
13.8
100.0
11.4
100.0
0.82
Pseudomonas aeruginosa
17.6
100.0
14.3
100.0
0.81
Salmonella enterica
13.6
100.0
11.1
100.0
0.82
Serratia marcescens
15.1
100.0
11.3
100.0
0.75
Stenotrophomonas maltophilia
29.4
100.0
36.6
100.0
1.25
Mould
Aspergillus fumigatus
34.9
100.0
30.2
100.0
0.87
Aspergillus brasiliensis
58.4
83.0
46.3
100.0
0.79
Yeast
Candida albicans
27.3
100.0
26.1
100.0
0.95
Candida glabrata
51.9
100.0
44.3
100.0
0.85
Candida krusei
21.9
100.0
19.8
100.0
0.90
Cryptococcus neoformans
63.2
100.0
63.2
100.0
1.00

Comparison of detection rates between systems

In some instances, bottles inoculated at or near the LoD remained negative, likely due to the non-uniform distribution (Poisson distribution) of the inoculum. Only matched test bottles (bottle type, organism, and blood volume) for which data were available from both detection systems were included in the final comparison analysis.
A total of 2242 inoculated culture bottles were tested in the VIRTUO™ system, of which 2240 (99.9%) were declared positive. A total of 2248 inoculated bottles tested in the 3D system were available for comparison to VIRTUO™, of which 2220 (98.8%) were declared positive (Table 3).
For individual bottle types, only the rates of detection observed for SN bottles containing blood were not significantly different between systems (p > 0.05; Table 3). All other bottle types tested demonstrated a detection rate in the VIRTUO™ greater than or equal to the detection rate of 3D. All seeded SA, FA Plus, and FN Plus bottles were declared positive in the VIRTUO™ system (Table 4). Furthermore, FN Plus bottles in the 3D system exhibited decreased detection rates as blood volume increased, resulting in seven false-negative results for Fusobacterium nucleatum (Table 3). False-negative results were also observed in FA Plus bottles seeded with Cardiobacterium hominis and Aspergillus brasiliensis in the 3D system.

Discussion

While testing at lower inoculum levels may delay TTD, comparing the detection systems at or near the LoD provides a more stringent evaluation compared to clinical samples. The data presented showed no difference in the LoD between 3D and VIRTUO™ for the panel of microorganisms tested. Even at low inoculum levels, 55% of the organisms evaluated in this study had an average TTD of less than 24 h in VIRTUO™. Further, our data demonstrates that the VIRTUO™ system provided a significantly improved rate of detection and faster TTD compared to the 3D (p < 0.05).
Overall, VIRTUO™ exhibited an improved detection rate of 99.9% compared to the 98.8% detection rate of 3D, with an average decrease in TTD of 3.48 h. FN Plus bottles had the most noticeable TTD improvement, with an average reduction of 3.90 h. Another seeded study reported significantly faster TTD with VIRTUO™ compared to 3D for several bacterial species in FA Plus and FN Plus bottles, with a median reduction of 2.8 h [10]. Additionally, the TTD for Candida glabrata in FA Plus bottles was reduced from a median of 65 h on 3D to 54 h on VIRTUO™ [10]. Previous studies demonstrate faster TTD for antibiotic-free samples and coagulase-negative staphylococci with BacT/ALERT® FA Plus and FN Plus bottles compared to BD BACTEC™ culture bottles [1113]. The improved detection algorithm of the VIRTUO™ system could further enhance this advantage in cases where blood cultures are collected from patients receiving concurrent antimicrobial therapy [6, 8, 10].
Gram-positive organisms had the greatest reduction in TTD with VIRTUO™. In addition, VIRTUO™ improved the TTD for Fusobacterium nucleatum by an average of 1.8 days (31 to 57 h, depending on bottle type). Earlier case studies report that the mean number of days required to recover this organism from blood culture in the BACTEC™ system (BACTEC 9120, BD Diagnostic Systems) was 2.6 days (range 35–87 h) [12]. Additionally, the VIRTUO™ algorithm detected 100% of F. nucleatum in FN Plus bottles, as well as 100% of Aspergillus brasiliensis and Cardiobacterium hominis in FA Plus bottles, whereas for these same combinations, the 3D system generated a few false-negative results. Although our panel evaluated a limited subset of yeast and moulds, a high number of Aspergillus false-negatives have been observed using various blood culture systems, including the BACTEC™ and 3D systems [14]. In some instances, organisms may be encountered that grow in the culture bottle but do not produce sufficient CO2 to signal positive [48]. A previous report indicated that C. hominis generally produces small incremental changes to the growth indices in automated blood culture systems after 3–5 days and recommended incubation for at least 14 days before cultures are presumed negative [15, 16]. The improved algorithm of VIRTUO™ is capable of detecting these minute changes, reducing false-negative results, and allowing for an overall reduction in C. hominis TTD from 54.9 h in 3D to 47.9 h.
Automated blood culture systems provide improved detection of bacteremia and fungemia and, as a result, have become a staple in the diagnostic clinical microbiology laboratory. Our studies compared the instrumentation and growth-based algorithms of the 3D and VIRTUO™ systems using a panel of compatible blood culture bottles seeded with tightly controlled inoculum levels at/near the LoD and at multiple blood volumes. The VIRTUO™ algorithm improved positive blood culture TTD while reducing false-negative results when directly compared to the current 3D system. Further comparison studies performed with patient samples in a clinical setting are needed to validate the improved performance of VIRTUO™.

Compliance with ethical standards

Conflict of interest

All authors are employees of bioMérieux, Inc., manufacturer of the BacT/ALERT® 3D, BacT/ALERT® VIRTUO™, and BacT/ALERT® blood culture bottles used in this manuscript.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Anhänge

Electronic supplementary material

Literatur
1.
Zurück zum Zitat Magadia RR, Weinstein MP (2001) Laboratory diagnosis of bacteremia and fungemia. Infect Dis Clin North Am 15:1009–1024CrossRefPubMed Magadia RR, Weinstein MP (2001) Laboratory diagnosis of bacteremia and fungemia. Infect Dis Clin North Am 15:1009–1024CrossRefPubMed
2.
Zurück zum Zitat Kumar A, Roberts D, Wood KE, Light B, Parrillo JE, Sharma S, Suppes R, Feinstein D, Zanotti S, Taiberg L, Gurka D, Kumar A, Cheang M (2006) Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med 34:1589–1596CrossRefPubMed Kumar A, Roberts D, Wood KE, Light B, Parrillo JE, Sharma S, Suppes R, Feinstein D, Zanotti S, Taiberg L, Gurka D, Kumar A, Cheang M (2006) Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med 34:1589–1596CrossRefPubMed
3.
Zurück zum Zitat BacT/ALERT® VIRTUO™ Microbial Detection System User Manual (2014-06) 414998-EN BacT/ALERT® VIRTUO™ Microbial Detection System User Manual (2014-06) 414998-EN
4.
Zurück zum Zitat BacT/ALERT® SA [package insert] (2010) bioMérieux, Inc., Durham, NC BacT/ALERT® SA [package insert] (2010) bioMérieux, Inc., Durham, NC
5.
Zurück zum Zitat BacT/ALERT® SN [package insert] (2011) bioMérieux, Inc., Durham, NC BacT/ALERT® SN [package insert] (2011) bioMérieux, Inc., Durham, NC
6.
Zurück zum Zitat BacT/ALERT® FA Plus [package insert] (2011) bioMérieux, Inc., Durham, NC BacT/ALERT® FA Plus [package insert] (2011) bioMérieux, Inc., Durham, NC
7.
Zurück zum Zitat BacT/ALERT® PF Plus [package insert] (2016) bioMérieux, Inc., Durham, NC BacT/ALERT® PF Plus [package insert] (2016) bioMérieux, Inc., Durham, NC
8.
Zurück zum Zitat BacT/ALERT® FN Plus [package insert] (2011) bioMérieux, Inc., Durham, NC BacT/ALERT® FN Plus [package insert] (2011) bioMérieux, Inc., Durham, NC
9.
Zurück zum Zitat Clinical and Laboratory Standards Institute (CLSI) (2007) Principles and procedures for blood cultures; approved guideline. CLSI document M47-A. CLSI, Wayne Clinical and Laboratory Standards Institute (CLSI) (2007) Principles and procedures for blood cultures; approved guideline. CLSI document M47-A. CLSI, Wayne
10.
Zurück zum Zitat Altun O, Almuhayawi M, Lüthje P, Taha R, Ullberg M, Özenci V (2016) Controlled evaluation of the new BacT/Alert Virtuo blood culture system for detection and time to detection of bacteria and yeasts. J Clin Microbiol 54:1148–1151CrossRefPubMedPubMedCentral Altun O, Almuhayawi M, Lüthje P, Taha R, Ullberg M, Özenci V (2016) Controlled evaluation of the new BacT/Alert Virtuo blood culture system for detection and time to detection of bacteria and yeasts. J Clin Microbiol 54:1148–1151CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Kirn TJ, Mirrett S, Reller LB, Weinstein MP (2014) Controlled clinical comparison of BacT/Alert FA Plus and FN Plus blood culture media with BacT/alert FA and FN blood culture media. J Clin Microbiol 52:839–843CrossRefPubMedPubMedCentral Kirn TJ, Mirrett S, Reller LB, Weinstein MP (2014) Controlled clinical comparison of BacT/Alert FA Plus and FN Plus blood culture media with BacT/alert FA and FN blood culture media. J Clin Microbiol 52:839–843CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Terhes G, Piukovics K, Urbán E, Nagy E (2011) Four cases of bacteraemia caused by Fusobacterium nucleatum in febrile, neutropenic patients. J Med Microbiol 60:1046–1049CrossRefPubMed Terhes G, Piukovics K, Urbán E, Nagy E (2011) Four cases of bacteraemia caused by Fusobacterium nucleatum in febrile, neutropenic patients. J Med Microbiol 60:1046–1049CrossRefPubMed
13.
Zurück zum Zitat Fiori B, D’Inzeo T, Di Florio V, De Maio F, De Angelis G, Giaquinto A, Campana L, Tanzarella E, Tumbarello M, Antonelli M, Sanguinetti M, Spanu T (2014) Performance of two resin-containing blood culture media in detection of bloodstream infections and in direct matrix-assisted laser desorption ionization–time of flight mass spectrometry (MALDI-TOF MS) broth assays for isolate identification: clinical comparison of the BacT/Alert Plus and BACTEC Plus systems. J Clin Microbiol 52:3558–3567 Fiori B, D’Inzeo T, Di Florio V, De Maio F, De Angelis G, Giaquinto A, Campana L, Tanzarella E, Tumbarello M, Antonelli M, Sanguinetti M, Spanu T (2014) Performance of two resin-containing blood culture media in detection of bloodstream infections and in direct matrix-assisted laser desorption ionization–time of flight mass spectrometry (MALDI-TOF MS) broth assays for isolate identification: clinical comparison of the BacT/Alert Plus and BACTEC Plus systems. J Clin Microbiol 52:3558–3567
14.
Zurück zum Zitat Rosa C, Araujo R, Rodrigues AG, Pinto-de-Sousa MI, Pina-Vaz C (2011) Detection of Aspergillus species in BACTEC blood cultures. J Med Microbiol 60:1467–1471CrossRefPubMed Rosa C, Araujo R, Rodrigues AG, Pinto-de-Sousa MI, Pina-Vaz C (2011) Detection of Aspergillus species in BACTEC blood cultures. J Med Microbiol 60:1467–1471CrossRefPubMed
15.
Zurück zum Zitat Lu PL, Hsueh PR, Hung CC, Teng LJ, Jang TN, Luh KT (2000) Infective endocarditis complicated with progressive heart failure due to beta-lactamase-producing Cardiobacterium hominis. J Clin Microbiol 38:2015–2017PubMedPubMedCentral Lu PL, Hsueh PR, Hung CC, Teng LJ, Jang TN, Luh KT (2000) Infective endocarditis complicated with progressive heart failure due to beta-lactamase-producing Cardiobacterium hominis. J Clin Microbiol 38:2015–2017PubMedPubMedCentral
16.
Zurück zum Zitat Manso E, Strusi P, Massacci C, Ferrini L (1987) Endocarditis due to Cardiobacterium hominis. Boll Ist Sieroter Milan 66:489–490PubMed Manso E, Strusi P, Massacci C, Ferrini L (1987) Endocarditis due to Cardiobacterium hominis. Boll Ist Sieroter Milan 66:489–490PubMed
Metadaten
Titel
A controlled comparison of the BacT/ALERT® 3D and VIRTUO™ microbial detection systems
verfasst von
H. Totty
M. Ullery
J. Spontak
J. Viray
M. Adamik
B. Katzin
W. M. Dunne Jr
P. Deol
Publikationsdatum
12.05.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Clinical Microbiology & Infectious Diseases / Ausgabe 10/2017
Print ISSN: 0934-9723
Elektronische ISSN: 1435-4373
DOI
https://doi.org/10.1007/s10096-017-2994-8

Weitere Artikel der Ausgabe 10/2017

European Journal of Clinical Microbiology & Infectious Diseases 10/2017 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.