A duplex real-time RT-PCR assay for detecting H5N1 avian influenza virus and pandemic H1N1 influenza virus

verfasst von: Xiao-ping Kang, Tao Jiang, Yong-qiang Li, Fang Lin, Hong Liu, Guo-hui Chang, Qing-yu Zhu, E-de Qin, Cheng-feng Qin, Yin-hui Yang

Erschienen in: Virology Journal | Ausgabe 1/2010

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

A duplex real-time reverse transcriptase polymerase chain reaction (RT-PCR) assay was improved for simultaneous detection of highly pathogenic H5N1 avian influenza virus and pandemic H1N1 (2009) influenza virus, which is suitable for early diagnosis of influenza-like patients and for epidemiological surveillance. The sensitivity of this duplex real-time RT-PCR assay was 0.02 TCID50 (50% tissue culture infective dose) for H5N1 and 0.2 TCID50 for the pandemic H1N1, which was the same as that of each single-target RT-PCR for pandemic H1N1 and even more sensitive for H5N1 with the same primers and probes. No cross reactivity of detecting other subtype influenza viruses or respiratory tract viruses was observed. Two hundred and thirty-six clinical specimens were tested by comparing with single real-time RT-PCR and result from the duplex assay was 100% consistent with the results of single real-time RT-PCR and sequence analysis.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1743-422X-7-113) contains supplementary material, which is available to authorized users.
Xiao-ping Kang, Tao Jiang contributed equally to this work.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

XK and TJ: designed the study, did laboratory testing, analysed the test results, co-wrote and edited the manuscript. YL, FL, LH and GC took samples and did laboratory testing. QZ, CQ and YY organized the overall project and helped edit the manuscript. All authors have read and approved the final manuscript.

Findings

In March and April 2009, a novel swine-origin pandemic H1N1 influenza virus appeared and spread worldwide. The pandemic of H1N1 2009 influenza virus poses a public health threat. According to the World Health Organization (WHO), as of 30 April 2010, worldwide at least 17919 death cases have been reported [1]. Compared with influenza A H1N1 virus, the highly pathogenic avian influenza virus H5N1 has a much higher mortality rate: H5N1 has a mortality rate of more than 50% while novel pandemic H1N1 has only about 1% [2]. Nevertheless, pandemic H1N1 and highly pathogenic H5N1 manifest similar clinical symptoms at the early stage of infection [3]. Early detection of these two pathogens is an essential prerequisite for effective control and prevention of the pandemic.
Real-time RT-PCR/PCR is a powerful method for the sensitive and specific detection of virus-derived nucleic acids in clinical samples [4]. It is time-saving and more specific compared with endpoint PCR. Therefore, real-time RT-PCR assay has been widely used and recommended by WHO for many rapid viral pathogen detection. Moreover, multiplex assays allow to measure several fluorophores in one well, which can simultaneously detect different target sequences [5]. Since the outbreak of pandemic H1N1 influenza, many real-time RT-PCR assays have been developed for detecting of the novel H1N1 [613]. Some multiplex real-time RT-PCR assays for simultaneous typing (A/B) and subtyping of H1, H2, H3, H5, H7 and H9 of influenza A viruses have also been reported [1417]. However, no studies were reported for subtyping of the novel pandemic H1N1 and H5N1 simultaneously.
In this study, a duplex TaqMan real-time RT-PCR assay was improved by adjusting the concentrations of primers and probes in the WHO protocols. The assay could simultaneously detect H5N1 avian influenza virus and pandemic H1N1 influenza virus, which could be used for early diagnosis of influenza-like patients and for epidemiological surveillance.
The primers and probes for the novel pandemic H1N1 and the influenza H5N1 were derived from the protocols recommended by WHO [18, 19]. The sequence of primers for H5N1 were H5-sense: 5'-GGA ACT TAC CAA ATA CTG TCA ATT TAT TCA-3', H5-antisense: 5'-CCA TAA AGA TAG ACC AGC TAC CAT GA-3' and H5-probe: 5'-HEX-TTG CCA GTG CTA GGG AAC TCG CCA C-BHQ1-3' which was labeled with the reporter hexachloro-6-carboxyfluorescein (HEX) at the 5' end and a black hole quencher (BHQ1) at the 3' end [19]. The sequence of primers and probe for pandemic H1N1 were H1-sense: 5'-GTG CTA TAA ACA CCA GCC TYC CA-3', H1-antisense: 5'-CGG GAT ATT CCT TAA TCC TGT RGC-3', H1-probe: 5'-FAM-CAG AAT ATA CAT* CCR GTC ACA ATT GGA RAA-3' which was labeled with the reporter 6-carboxyfluorescein (6-FAM) at the 5' end and a black hole quencher (BHQ1) at the a modified T residue [18]. All primers and probes were synthesized by Dalian Takara Company.
RNA was extracted from the supernatants of cultured viruses or from clinical specimens by RNeasy Mini kit (Qiagen, Hilden, Germany) according to the manufacture's instruction. Virus RNA extractions were conducted in Biosafety Level 3 (BSL-3) facilities.
The duplex real-time RT-PCR amplification was carried out in a 20 μL volume reaction with the Quantitect Probe RT-PCR kit (Qiagen, Hilden, Germany) with the LightCycler 2.0 system (Roche, Mannheim, Germany). Different concentrations of the primers and probes for both H5N1 and H1N1 were combined and adjusted to improve the sensitivity of the assay [20]. The concentrations of primers and probe for H5 gene in the duplex system was optimized by using a serial of 10-fold diluted H5N1 RNAs as templates. The sensitivity of the duplex assays was evaluated at the concentration of H5 primers as 0.4 μM, 0.8 μM, 1.6 μM and 3.2 μM. The results showed that the sensitivity of the duplex assays dramatically increased with the concentration of H5 primers. The duplex assay had the best detecting result at the H5 primer concentration of 1.6 μM. The concentration of probes for H5 gene was also optimized by examining the concentration at 0.1 μM, 0.2 μM, 0.4 μM and 0.8 μM, and any significant different results at different concentrations was not observed. Therefore, 0.1 μM was selected as the final concentration of H5 probe in the duplex real-time RT-PCR system. Similar to the duplex assay for H5 gene, the concentrations of primers and probe for H1 also were optimized by using RNA from 2, 0.2, 0.02 and 0.002 TCID50 H1N1 as templates. The concentration of H1 primers was optimized from 0.1 μM, 0.2 μM, 0.4 μM, 0.8 μM and 1.6 μM, while the concentration of H1 probe was optimized at 0.1 μM, 0.2 μM, 0.4 μM and 0.8 μM. Thus, 0.4 μM of primers and 0.2 μM of probe were selected as the optimal concentration for H1 gene. The reactions were incubated at 50°C for 30 min, followed by 95°C for 10 min, 45 cycles of 95°C for 15 s, and 52°C for 1 min. Fluorescence was recorded at 52°C.
The final optimized reaction mixture consisted of 10 μL of 2× reaction buffer, 0.2 μL reverse transcription enzyme, 1.6 μM of each H5N1 primer, 400 nM of each novel H1N1 primer, 100 nM of H5N1 probe, 200 nM of pandemic H1N1 probe and 2 μL RNA templates. The protocol of real-time RT-PCR for influenza A (H1N1) recommended by WHO used a concentration of 1000 nM each of novel SW H1 primers and 250 nM of SW H1 probe [18]. The WHO real-time RT-PCR protocol for H5N1 used a concentration of 800 nM each of H5 primers and 200 nM of H5 probe [19].
The sensitivity of the single and duplex real-time RT-PCR for novel H1N1 and H5N1 virus was evaluated by 10-fold diluted virus RNA, respectively. The pandemic H1N1 influenza virus strain A/Beijing/501/2009(H1N1) (GenBank: GQ223408-GQ223415) and avian influenza virus H5N1 strain A/Beijing/01/2003 (GenBank: EF587274-EF587281) were cultured on MDCK cells and tested by the assay. The analysis of threshold cycles (Ct) signals as a function of log10 TCID50 titers of tested viruses showed a nearly linear decrease of Ct value with increased virus titer. The sensitivity of the single real-time RT-PCR assay was 0.2 TCID50 for both novel pandemic H1N1 and H5N1 virus. The detection threshold of the duplex real-time RT-PCR assay was 0.2 TCID50 for the pandemic H1N1 and 0.02 TCID50 for H5N1 virus. Compared with single real-time RT-PCR assay, the duplex real-time RT-PCR assay had the same sensitivity for pandemic H1N1 virus, and about 10-fold more sensitive for H5N1 virus (0.02 TCID50) (Figure 1). This was likely because the procedure of the duplex assay for H5N1 was optimized for Roche LightCycler 2.0 system, while the single real-time RT-PCR assay recommended by WHO was optimized for ABI instruments [18].
The specificity of the duplex real time RT-PCR assay was validated by using human genomes and a panel of respiratory tract viruses including human seasonal H3N2 influenza viruses, seasonal H1N1 virus, human respiratory syncytial virus A and B, human coronavirus 229E, human coronavirus OC43, influenza B virus, human parainfluenza virus and human adenovirus. All samples were tested negative (data not shown).
A total of 236 clinical throat swab specimens from suspected cases of novel pandemic H1N1 patients were collected. The samples were first detected by WHO real-time RT-PCR protocol for influenza A (H1N1). As the WHO protocol also included a set of primers for universal detection of type A influenza viruses, thus, only the 182 Influenza A positive samples confirmed by WHO protocol were further tested by the duplex system for comparison with the results of WHO protocol. Among the 182 influenza type A positive specimens, 124 specimens were positive for the pandemic H1N1 and none for the H5N1 tested by the duplex system. Duplex real-time RT-PCR assays showed 100% coincident results with the single real-time assays. Sequence analysis was also conducted with 20 samples selected randomly from the novel H1N1 positive specimens.
Since few clinical specimens was available currently in China, a mock clinical H5N1 specimen was extracted from the spleen of BALB/C mice infected with high pathogenic H5N1 influenza virus to validate the specificity of high pathogenic H5N1 virus. The duplex real-time RT-PCR assay showed strong positive result, which was confirmed by single real-time PCR and subsequent sequencing analyses. This demonstrated that the duplex real-time RT-PCR assay was a substitute method for the current single real-time RT-PCR method.
So far, the number of confirmed severe novel H1N1 cases has reached about seven hundred thousand in China and increased quickly. The lessons from 1918 and 1957 influenza suggested that the pandemic might return in a much more lethal form in the second wave [21]. Recently, the increasing geographic distribution of high pathogenic H5N1 as well as its ability to transfer to humans and to cause severe infection has raised serious concerns regarding the control measures of this virus. The duplex real-time RT-PCR assay for pandemic H1N1 and high pathogenic H5N1 would play an important role for control and prevention of pandemic caused by these viruses.

Acknowledgements

This work was supported in part by the National 973 project of China (2010CB534002), and the Major Special Program of National Science and Technology of China (2008ZX10004-401 and 2009ZX10004-204).
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://​creativecommons.​org/​licenses/​by/​2.​0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

XK and TJ: designed the study, did laboratory testing, analysed the test results, co-wrote and edited the manuscript. YL, FL, LH and GC took samples and did laboratory testing. QZ, CQ and YY organized the overall project and helped edit the manuscript. All authors have read and approved the final manuscript.
Anhänge

Authors’ original submitted files for images

Literatur
1.
Zurück zum Zitat WHO: Pandemic (H1N1) 2009-Update 98. 2010 WHO: Pandemic (H1N1) 2009-Update 98. 2010
2.
Zurück zum Zitat WHO: Cumulative Number of Confirmed Human Cases of Avian Influenza A/(H5N1) Reported to WHO. 2010 WHO: Cumulative Number of Confirmed Human Cases of Avian Influenza A/(H5N1) Reported to WHO. 2010
3.
Zurück zum Zitat Rumschlag-Booms E, Guo Y, Wang J, Caffrey M, Rong L: Comparative analysis between a low pathogenic and a high pathogenic influenza H5 hemagglutinin in cell entry. Virol J. 2009, 6: 76-81. 10.1186/1743-422X-6-76.PubMedPubMedCentralCrossRef Rumschlag-Booms E, Guo Y, Wang J, Caffrey M, Rong L: Comparative analysis between a low pathogenic and a high pathogenic influenza H5 hemagglutinin in cell entry. Virol J. 2009, 6: 76-81. 10.1186/1743-422X-6-76.PubMedPubMedCentralCrossRef
5.
Zurück zum Zitat Wittwer CT, Herrmann MG, Gundry CN, Elenitoba-Johnson JKS: Real-time multiplex PCR assays. Methods. 2001, 25: 430-442. 10.1006/meth.2001.1265.PubMedCrossRef Wittwer CT, Herrmann MG, Gundry CN, Elenitoba-Johnson JKS: Real-time multiplex PCR assays. Methods. 2001, 25: 430-442. 10.1006/meth.2001.1265.PubMedCrossRef
6.
Zurück zum Zitat Dong H, Zhang Y, Xiong H, Yan A, Ding G, Chen Y, Xie L, Chen J, Zhang G, Hao P, Cong L, Lu Y, Che X, Wang X, Li Y, Yuen KY, Zhao G, Jin W: Detection of human novel influenza A (H1N1) viruses using multi-fluorescent real-time RT-PCR. Virus Res. 2010, 147: 85-90. 10.1016/j.virusres.2009.10.011.PubMedCrossRef Dong H, Zhang Y, Xiong H, Yan A, Ding G, Chen Y, Xie L, Chen J, Zhang G, Hao P, Cong L, Lu Y, Che X, Wang X, Li Y, Yuen KY, Zhao G, Jin W: Detection of human novel influenza A (H1N1) viruses using multi-fluorescent real-time RT-PCR. Virus Res. 2010, 147: 85-90. 10.1016/j.virusres.2009.10.011.PubMedCrossRef
7.
Zurück zum Zitat LeBlanc JJ, Li Y, Bastien N, Forward KR, Davidson RJ, Hatchette TF: Switching gears for an influenza pandemic: validation of a duplex reverse transcriptase PCR assay for simultaneous detection and confirmatory identification of pandemic (H1N1) 2009 influenza virus. JClin Microbiol. 2009, 47: 3805-3813. 10.1128/JCM.01344-09.CrossRef LeBlanc JJ, Li Y, Bastien N, Forward KR, Davidson RJ, Hatchette TF: Switching gears for an influenza pandemic: validation of a duplex reverse transcriptase PCR assay for simultaneous detection and confirmatory identification of pandemic (H1N1) 2009 influenza virus. JClin Microbiol. 2009, 47: 3805-3813. 10.1128/JCM.01344-09.CrossRef
8.
Zurück zum Zitat Jiang T, Kang X, Deng Y, Zhao H, Li X, Yu X, Yu M, Qin E, Zhu Q, Yang Y, Qin C: Development of a real-time RT-PCR assay for a novel influenza A (H1N1) virus. J Virol Method. 2010, 163: 470-473. 10.1016/j.jviromet.2009.09.021.CrossRef Jiang T, Kang X, Deng Y, Zhao H, Li X, Yu X, Yu M, Qin E, Zhu Q, Yang Y, Qin C: Development of a real-time RT-PCR assay for a novel influenza A (H1N1) virus. J Virol Method. 2010, 163: 470-473. 10.1016/j.jviromet.2009.09.021.CrossRef
9.
Zurück zum Zitat Panning M, Eickmann M, Landt O, Monazahian M, Olschläger S, Baumgarte S, Reischl U, Wenzel JJ, Niller HH, Günther S, Hollmann B, Huzly D, Drexler JF, Helmer A, Becker S, Matz B, Eis-Hübinger A, Drosten C: Detection of influenza A(H1N1) virus by real-time RT-PCR. Euro Surveill. 2009, 14: 19329-PubMed Panning M, Eickmann M, Landt O, Monazahian M, Olschläger S, Baumgarte S, Reischl U, Wenzel JJ, Niller HH, Günther S, Hollmann B, Huzly D, Drexler JF, Helmer A, Becker S, Matz B, Eis-Hübinger A, Drosten C: Detection of influenza A(H1N1) virus by real-time RT-PCR. Euro Surveill. 2009, 14: 19329-PubMed
10.
Zurück zum Zitat Yang JR, Lo J, Liu JL, Lin CH, Ho YL, Chen CJ, Wu HS, Liu MT: Rapid SYBR green I and modified probe real-time reverse transcription-PCR assays identify influenza H1N1 viruses and distinguish between pandemic and seasonal strains. J Clin Microbiol. 2009, 47: 3714-1716. 10.1128/JCM.01646-09.PubMedPubMedCentralCrossRef Yang JR, Lo J, Liu JL, Lin CH, Ho YL, Chen CJ, Wu HS, Liu MT: Rapid SYBR green I and modified probe real-time reverse transcription-PCR assays identify influenza H1N1 viruses and distinguish between pandemic and seasonal strains. J Clin Microbiol. 2009, 47: 3714-1716. 10.1128/JCM.01646-09.PubMedPubMedCentralCrossRef
11.
Zurück zum Zitat Wang R, Sheng ZM, Taubenberger JK: Detection of novel (swine origin) H1N1 influenza A virus by quantitative real-time reverse transcription-PCR. J Clin Microbiol. 2009, 47: 2675-2677. 10.1128/JCM.01087-09.PubMedPubMedCentralCrossRef Wang R, Sheng ZM, Taubenberger JK: Detection of novel (swine origin) H1N1 influenza A virus by quantitative real-time reverse transcription-PCR. J Clin Microbiol. 2009, 47: 2675-2677. 10.1128/JCM.01087-09.PubMedPubMedCentralCrossRef
12.
Zurück zum Zitat Lau SK, Chan KH, Yip CC, Ng TK, Tsang OT, Woo PC, Yuen KY: Confirmation of the first Hong Kong case of human infection by novel swine origin influenza A (H1N1) virus diagnosed using ultrarapid, real-time reverse transcriptase PCR. J Clin Microbiol. 2009, 47: 2344-2346. 10.1128/JCM.00924-09.PubMedPubMedCentralCrossRef Lau SK, Chan KH, Yip CC, Ng TK, Tsang OT, Woo PC, Yuen KY: Confirmation of the first Hong Kong case of human infection by novel swine origin influenza A (H1N1) virus diagnosed using ultrarapid, real-time reverse transcriptase PCR. J Clin Microbiol. 2009, 47: 2344-2346. 10.1128/JCM.00924-09.PubMedPubMedCentralCrossRef
13.
Zurück zum Zitat Lorusso A, Faaberg KS, Killian ML, Koster L, Vincent AL: One-step real-time RT-PCR for pandemic influenza A virus (H1N1) 2009 matrix gene detection in swine samples. J Virol Methods. 2010, 164: 83-87. 10.1016/j.jviromet.2009.12.002.PubMedCrossRef Lorusso A, Faaberg KS, Killian ML, Koster L, Vincent AL: One-step real-time RT-PCR for pandemic influenza A virus (H1N1) 2009 matrix gene detection in swine samples. J Virol Methods. 2010, 164: 83-87. 10.1016/j.jviromet.2009.12.002.PubMedCrossRef
14.
Zurück zum Zitat Wang W, Ren P, Mardi S, Hou L, Tsai C, Chan KH, Cheng P, Sheng J, Buchy P, Sun B, Toyoda T, Lim W, Peiris JS, Zhou P, Deubel V: Design of multiplexed detection assays for identification of avian influenza a virus subtypes pathogenic to humans by SmartCycler real-time reverse transcription-PCR. J Clin Microbiol. 2009, 47: 86-92. 10.1128/JCM.01090-08.PubMedPubMedCentralCrossRef Wang W, Ren P, Mardi S, Hou L, Tsai C, Chan KH, Cheng P, Sheng J, Buchy P, Sun B, Toyoda T, Lim W, Peiris JS, Zhou P, Deubel V: Design of multiplexed detection assays for identification of avian influenza a virus subtypes pathogenic to humans by SmartCycler real-time reverse transcription-PCR. J Clin Microbiol. 2009, 47: 86-92. 10.1128/JCM.01090-08.PubMedPubMedCentralCrossRef
15.
Zurück zum Zitat Suwannakarn K, Payungporn S, Chieochansin T, Samransamruajkit R, Amonsin A, Songserm T, Chaisingh A, Chamnanpood P, Chutinimitkul S, Theamboonlers A, Poovorawan Y: Typing (A/B) and subtyping (H1/H3/H5) of influenza A viruses by multiplex real-time RT-PCR assays. J Virol Methods. 2008, 152: 25-31. 10.1016/j.jviromet.2008.06.002.PubMedCrossRef Suwannakarn K, Payungporn S, Chieochansin T, Samransamruajkit R, Amonsin A, Songserm T, Chaisingh A, Chamnanpood P, Chutinimitkul S, Theamboonlers A, Poovorawan Y: Typing (A/B) and subtyping (H1/H3/H5) of influenza A viruses by multiplex real-time RT-PCR assays. J Virol Methods. 2008, 152: 25-31. 10.1016/j.jviromet.2008.06.002.PubMedCrossRef
16.
Zurück zum Zitat Wu C, Cheng X, He J, Lv X, Wang J, Deng R, Long Q, Wang X: A multiplex real-time RT-PCR for detection and identification of influenza virus types A and B and subtypes H5 and N1. J Virol Methods. 2008, 148: 81-88. 10.1016/j.jviromet.2007.10.023.PubMedCrossRef Wu C, Cheng X, He J, Lv X, Wang J, Deng R, Long Q, Wang X: A multiplex real-time RT-PCR for detection and identification of influenza virus types A and B and subtypes H5 and N1. J Virol Methods. 2008, 148: 81-88. 10.1016/j.jviromet.2007.10.023.PubMedCrossRef
17.
Zurück zum Zitat Li PQ, Zhang J, Muller CP, Chen JX, Yang ZF, Zhang R, Li J, He YS: Development of a multiplex real-time polymerase chain reaction for the detection of influenza virus type A including H5 and H9 subtypes. Diagn Microbiol Infect Dis. 2008, 61: 192-197. 10.1016/j.diagmicrobio.2008.01.007.PubMedCrossRef Li PQ, Zhang J, Muller CP, Chen JX, Yang ZF, Zhang R, Li J, He YS: Development of a multiplex real-time polymerase chain reaction for the detection of influenza virus type A including H5 and H9 subtypes. Diagn Microbiol Infect Dis. 2008, 61: 192-197. 10.1016/j.diagmicrobio.2008.01.007.PubMedCrossRef
18.
Zurück zum Zitat WHO: CDC protocol of realtime RTPCR for swine influenza A(H1N1). 2009 WHO: CDC protocol of realtime RTPCR for swine influenza A(H1N1). 2009
19.
Zurück zum Zitat WHO: Recommendations and laboratory procedures for detection of avian influenza A(H5N1) virus in specimens from suspected human cases. 2007 WHO: Recommendations and laboratory procedures for detection of avian influenza A(H5N1) virus in specimens from suspected human cases. 2007
20.
Zurück zum Zitat Hull R, Nattanmai S, Kramer LD, Bernard KA, Tavakoli NP: A duplex real-time reverse transcriptase polymerase chain reaction assay for the detection of St. Louis encephalitis and eastern equine encephalitis viruses. Diagn Microbiol Infect Dis. 2008, 62: 272-279. 10.1016/j.diagmicrobio.2008.07.004.PubMedPubMedCentralCrossRef Hull R, Nattanmai S, Kramer LD, Bernard KA, Tavakoli NP: A duplex real-time reverse transcriptase polymerase chain reaction assay for the detection of St. Louis encephalitis and eastern equine encephalitis viruses. Diagn Microbiol Infect Dis. 2008, 62: 272-279. 10.1016/j.diagmicrobio.2008.07.004.PubMedPubMedCentralCrossRef
Metadaten
Titel
A duplex real-time RT-PCR assay for detecting H5N1 avian influenza virus and pandemic H1N1 influenza virus
verfasst von
Xiao-ping Kang
Tao Jiang
Yong-qiang Li
Fang Lin
Hong Liu
Guo-hui Chang
Qing-yu Zhu
E-de Qin
Cheng-feng Qin
Yin-hui Yang
Publikationsdatum
01.12.2010
Verlag
BioMed Central
Erschienen in
Virology Journal / Ausgabe 1/2010
Elektronische ISSN: 1743-422X
DOI
https://doi.org/10.1186/1743-422X-7-113

Weitere Artikel der Ausgabe 1/2010

Virology Journal 1/2010 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.