A more accurate and safer method for the measurement of scattered radiation in X-ray examination rooms

verfasst von: Tokiko Nakamura, Suzuki Shoichi, Yasutaka Takei, Masanao Kobayashi, Vergil Cruz, Ikuo Kobayashi, Satoshi Asegawa, Kyoichi Kato

Erschienen in: Radiological Physics and Technology | Ausgabe 1/2020

Einloggen, um Zugang zu erhalten

Abstract

Advancements in and increasing usage of complex diagnostic examinations with interventional procedures and surgeries has led to an increase in the occupational radiation doses received by physicians and other medical staff. Measuring the scattered radiation doses received by these staff is vital for the development-effective radiation protection programs. In this study, we measured scattered doses during angiography and digital breast tomosynthesis examinations with small-type dosimeters using our jungle gym (JG) method with measurement points at 50-cm intervals. The results were compared with measurements taken using the conventional ion chamber method. The JG method uses paper pipe tubes and a plastic joint structure and allows measurements at different points inside an examination room. The difference between measurements can be attributed to the radiation absorption characteristics of the components used in the JG method. A maximum radiation dose reduction of 20% was observed due to absorption by the JG components. This effect was smaller than the measurement error produced because of reproducibility issues and other limitations of the conventional method. The conventional measurement has disadvantages that are associated with the reproducibility of measurement points, equipment load, and the radiation exposure experienced by the measurer. The proposed JG method exhibits significant improvements in all these aspects. Furthermore, the measurer does not have to be present in the measurement room; therefore, the JG method is extremely safe and useful for radiation protection.
Literatur
1.
Zurück zum Zitat ICRP 85. Avoidance of radiation injuries from medical interventional procedures. International Commission on Radiological Protection (ICRP) Publication 85. Ann ICRP. 2000;30(2):64. ICRP 85. Avoidance of radiation injuries from medical interventional procedures. International Commission on Radiological Protection (ICRP) Publication 85. Ann ICRP. 2000;30(2):64.
2.
Zurück zum Zitat ICRP 139. Occupational radiological protection in interventional procedures. International Commission on Radiological Protection Publication 139. Ann ICRP. 2018;47(2):72–83. ICRP 139. Occupational radiological protection in interventional procedures. International Commission on Radiological Protection Publication 139. Ann ICRP. 2018;47(2):72–83.
3.
Zurück zum Zitat Willams JR. The interdependence of staff and patient doses in interventional radiology. BJR. 1997;70:498–503.CrossRef Willams JR. The interdependence of staff and patient doses in interventional radiology. BJR. 1997;70:498–503.CrossRef
4.
Zurück zum Zitat Balter S. Stray radiation in the cardiac catheterisation laboratory. Radiat Prot Dosimetry. 2001;94:183–8.CrossRef Balter S. Stray radiation in the cardiac catheterisation laboratory. Radiat Prot Dosimetry. 2001;94:183–8.CrossRef
5.
Zurück zum Zitat Iida H, et al. The measurement of scatter radiation distributions in X-ray rooms. J Radiat Prot Sec Jpn Soc Radiat Technol. 2004;18:48. Iida H, et al. The measurement of scatter radiation distributions in X-ray rooms. J Radiat Prot Sec Jpn Soc Radiat Technol. 2004;18:48.
6.
Zurück zum Zitat Nakamura T, Suzuki S, Takei Y, Kobayashi I, Pongnapang N, Kato K. Simultaneous measurement of patient dose and distribution of indoor scatter radiation during digital breast tomosynthesis. Radiography. 2019;25(1):72–6.CrossRef Nakamura T, Suzuki S, Takei Y, Kobayashi I, Pongnapang N, Kato K. Simultaneous measurement of patient dose and distribution of indoor scatter radiation during digital breast tomosynthesis. Radiography. 2019;25(1):72–6.CrossRef
9.
Zurück zum Zitat Fetterly KA, Schueler BA, Grams MP, Sturchio GM. Estimating head and neck tissue dose from x-ray scatter to physicians performing X-ray guided cardiovascular procedures: a phantom study. J Radiol Prot. 2017;37(16):43–58.CrossRef Fetterly KA, Schueler BA, Grams MP, Sturchio GM. Estimating head and neck tissue dose from x-ray scatter to physicians performing X-ray guided cardiovascular procedures: a phantom study. J Radiol Prot. 2017;37(16):43–58.CrossRef
10.
Zurück zum Zitat Sato H, Mori K, Fujisaku K, Yoneda H. Development of EGS5 Monte Carlo toolkit for estimation of dose distribution in X-ray diagnostic room. Jpn J Med Phys. 2010;30:25–38. Sato H, Mori K, Fujisaku K, Yoneda H. Development of EGS5 Monte Carlo toolkit for estimation of dose distribution in X-ray diagnostic room. Jpn J Med Phys. 2010;30:25–38.
11.
Zurück zum Zitat Shimada M. Application of Monte Carlo simulations (4) area of general radiography. Nihon Houshasen Gijyutsu Gakkai Zasshi. 2015;71(3):271–6.CrossRef Shimada M. Application of Monte Carlo simulations (4) area of general radiography. Nihon Houshasen Gijyutsu Gakkai Zasshi. 2015;71(3):271–6.CrossRef
12.
Zurück zum Zitat Tomita H, Morozumi K. Measurement of the distribution of scatter radiation in CT rooms with MDCT equipment using OSL dosimeters. Nihon Houshasen Gijyutsu Gakkai Zasshi. 2004;60(11):1550–4.CrossRef Tomita H, Morozumi K. Measurement of the distribution of scatter radiation in CT rooms with MDCT equipment using OSL dosimeters. Nihon Houshasen Gijyutsu Gakkai Zasshi. 2004;60(11):1550–4.CrossRef
13.
Zurück zum Zitat Kuribayashi S, Onobori K, Hayashida A, et al. Report on airborne dosimetry during IVR (PTCA). Nihon Houshasen Gijyutsu Gakkai Zasshi. 2001;57(1):33–48.CrossRef Kuribayashi S, Onobori K, Hayashida A, et al. Report on airborne dosimetry during IVR (PTCA). Nihon Houshasen Gijyutsu Gakkai Zasshi. 2001;57(1):33–48.CrossRef
14.
Zurück zum Zitat Hironobu T, Kunihiro M. Measurement of scatter radiation on MDCT equipment using an OSL dosimeter. Nihon Houshasen Gijyutsu Gakkai Zasshi. 2004;60(11):1550–4.CrossRef Hironobu T, Kunihiro M. Measurement of scatter radiation on MDCT equipment using an OSL dosimeter. Nihon Houshasen Gijyutsu Gakkai Zasshi. 2004;60(11):1550–4.CrossRef
15.
Zurück zum Zitat Kusama T, Ono K. Radiology protection manual. 3rd ed. Tokyo: Japan Medical Association; 2013. p. 75–77. Kusama T, Ono K. Radiology protection manual. 3rd ed. Tokyo: Japan Medical Association; 2013. p. 75–77.
16.
Zurück zum Zitat Reynolds TA, Higgins P. Surface dose measurements with commonly used detectors: a consistent thickness correction method. J Appl Clin Med Phys. 2015;16:5572.CrossRef Reynolds TA, Higgins P. Surface dose measurements with commonly used detectors: a consistent thickness correction method. J Appl Clin Med Phys. 2015;16:5572.CrossRef
17.
Zurück zum Zitat Asahara T, Hayashi H, Goto S, Tomita E, Kimoto N, Mihara Y, Asakawa T, Kanazawa Y, Katsumata A, Higashino K, Yamashita K, Okazaki T, Hashizume T. Exposure dose measurement during diagnostic pediatric X-ray examination using an optically stimulated luminescence (OSL) dosimeter based on precise dose calibration taking into consideration variation of X-ray spectra. Radiat Meas. 2018;119:209–19.CrossRef Asahara T, Hayashi H, Goto S, Tomita E, Kimoto N, Mihara Y, Asakawa T, Kanazawa Y, Katsumata A, Higashino K, Yamashita K, Okazaki T, Hashizume T. Exposure dose measurement during diagnostic pediatric X-ray examination using an optically stimulated luminescence (OSL) dosimeter based on precise dose calibration taking into consideration variation of X-ray spectra. Radiat Meas. 2018;119:209–19.CrossRef
18.
Zurück zum Zitat Jursinic PA. Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements. Med Phys. 2007;34:4594–604.CrossRef Jursinic PA. Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements. Med Phys. 2007;34:4594–604.CrossRef
19.
Zurück zum Zitat Viamonte A, da Rosa LA, Buckley LA, et al. Radiotherapy dosimetry using a commercial OSL system. Med Phys. 2008;35:1261–6.CrossRef Viamonte A, da Rosa LA, Buckley LA, et al. Radiotherapy dosimetry using a commercial OSL system. Med Phys. 2008;35:1261–6.CrossRef
20.
Zurück zum Zitat Al-Senan RM, Hatab MR. Characteristics of an OSLD in the diagnostic energy range. Med Phys. 2011;38:4396–405.CrossRef Al-Senan RM, Hatab MR. Characteristics of an OSLD in the diagnostic energy range. Med Phys. 2011;38:4396–405.CrossRef
21.
Zurück zum Zitat ICRP 118. ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs-threshold doses for tissue reactions in a radiation protection context. International Commission on Radiological Protection (ICRP) Publication 118. Ann ICRP. 2012;41:1–322. ICRP 118. ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs-threshold doses for tissue reactions in a radiation protection context. International Commission on Radiological Protection (ICRP) Publication 118. Ann ICRP. 2012;41:1–322.
Metadaten
Titel
A more accurate and safer method for the measurement of scattered radiation in X-ray examination rooms
verfasst von
Tokiko Nakamura
Suzuki Shoichi
Yasutaka Takei
Masanao Kobayashi
Vergil Cruz
Ikuo Kobayashi
Satoshi Asegawa
Kyoichi Kato
Publikationsdatum
01.03.2020
Verlag
Springer Singapore
Erschienen in
Radiological Physics and Technology / Ausgabe 1/2020
Print ISSN: 1865-0333
Elektronische ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-019-00550-6

Weitere Artikel der Ausgabe 1/2020

Radiological Physics and Technology 1/2020 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.