A multi-degree-of-freedom needle driver with a short tip and small shaft for pediatric laparoscopic surgery: in vivo assessment of multi-directional suturing on the vertical plane of the liver in rabbits
verfasst von:
Shinya Takazawa, Tetsuya Ishimaru, Masahiro Fujii, Kanako Harada, Kyoichi Deie, Jun Fujishiro, Naohiko Sugita, Mamoru Mitsuishi, Tadashi Iwanaka
Erschienen in:
Surgical Endoscopy
|
Ausgabe 8/2016
Einloggen, um Zugang zu erhalten
Abstract
Background
Laparoscopic Kasai portoenterostomy has been performed in infants with biliary atresia at several institutions, but laparoscopic anastomosis requiring multi-directional suturing on a vertical plane of the liver remains a challenge. To assist multi-directional suturing, we developed a multi-degree-of-freedom (DOF) needle driver whose tip length was 15 mm and shaft diameter was 3.5 mm. The tip of the multi-DOF needle driver has three DOFs for grasp, flection and rotation. The aim of this study was to evaluate the performance of the multi-DOF needle driver in two kinds of in vivo experiments.
Methods
Surgeons were asked to perform four-directional laparoscopic suturing on a vertical plane of the liver in six rabbits using the multi-DOF needle driver or a conventional needle driver. The needle grasping time, the needle handling time, the number of needle insertions, the number of liver lacerations, the suturing width and depth, and the area of necrotic tissues were analyzed and compared. Additionally, one surgeon was asked to perform laparoscopic hepato-jejunostomy in four rabbits to assess the feasibility of Kasai portoenterostomy using the multi-DOF needle driver.
Results
The suturing depth using the multi-DOF needle driver was significantly larger than that using the conventional needle driver in both the right and downward suturing directions. No statistically significant differences were found in other metrics. Liver lacerations were observed only when suturing was performed using the conventional needle driver. The experimental laparoscopic hepato-jejunostomy using the multi-DOF needle driver was successful.
Conclusions
Using the multi-DOF needle driver, uniform multi-directional suturing on a vertical plane of the liver could be performed. The short distal tip of the multi-DOF needle driver demonstrated its advantages in multi-directional suturing in a small body cavity. The multi-DOF needle driver may be able to be used to perform complex tasks in laparoscopic Kasai portoenterostomy.