Erschienen in: Surgical Endoscopy 8/2016

28.10.2015 | New Technology

A multi-degree-of-freedom needle driver with a short tip and small shaft for pediatric laparoscopic surgery: in vivo assessment of multi-directional suturing on the vertical plane of the liver in rabbits

verfasst von: Shinya Takazawa, Tetsuya Ishimaru, Masahiro Fujii, Kanako Harada, Kyoichi Deie, Jun Fujishiro, Naohiko Sugita, Mamoru Mitsuishi, Tadashi Iwanaka

Erschienen in: Surgical Endoscopy | Ausgabe 8/2016

Einloggen, um Zugang zu erhalten

Abstract

Background

Laparoscopic Kasai portoenterostomy has been performed in infants with biliary atresia at several institutions, but laparoscopic anastomosis requiring multi-directional suturing on a vertical plane of the liver remains a challenge. To assist multi-directional suturing, we developed a multi-degree-of-freedom (DOF) needle driver whose tip length was 15 mm and shaft diameter was 3.5 mm. The tip of the multi-DOF needle driver has three DOFs for grasp, flection and rotation. The aim of this study was to evaluate the performance of the multi-DOF needle driver in two kinds of in vivo experiments.

Methods

Surgeons were asked to perform four-directional laparoscopic suturing on a vertical plane of the liver in six rabbits using the multi-DOF needle driver or a conventional needle driver. The needle grasping time, the needle handling time, the number of needle insertions, the number of liver lacerations, the suturing width and depth, and the area of necrotic tissues were analyzed and compared. Additionally, one surgeon was asked to perform laparoscopic hepato-jejunostomy in four rabbits to assess the feasibility of Kasai portoenterostomy using the multi-DOF needle driver.

Results

The suturing depth using the multi-DOF needle driver was significantly larger than that using the conventional needle driver in both the right and downward suturing directions. No statistically significant differences were found in other metrics. Liver lacerations were observed only when suturing was performed using the conventional needle driver. The experimental laparoscopic hepato-jejunostomy using the multi-DOF needle driver was successful.

Conclusions

Using the multi-DOF needle driver, uniform multi-directional suturing on a vertical plane of the liver could be performed. The short distal tip of the multi-DOF needle driver demonstrated its advantages in multi-directional suturing in a small body cavity. The multi-DOF needle driver may be able to be used to perform complex tasks in laparoscopic Kasai portoenterostomy.
Literatur
1.
Zurück zum Zitat Wong KK, Chung PH, Chan KL, Fan ST, Tam PK (2008) Should open Kasai portoenterostomy be performed for biliary atresia in the era of laparoscopy? Pediatr Surg Int 24:931–933CrossRefPubMed Wong KK, Chung PH, Chan KL, Fan ST, Tam PK (2008) Should open Kasai portoenterostomy be performed for biliary atresia in the era of laparoscopy? Pediatr Surg Int 24:931–933CrossRefPubMed
2.
Zurück zum Zitat Ure BM, Kuebler JF, Schukfeh N, Engelmann C, Dingemann J, Petersen C (2011) Survival with the native liver after laparoscopic versus conventional Kasai portoenterostomy in infants with biliary atresia: a prospective trial. Ann Surg 253:826–830CrossRefPubMed Ure BM, Kuebler JF, Schukfeh N, Engelmann C, Dingemann J, Petersen C (2011) Survival with the native liver after laparoscopic versus conventional Kasai portoenterostomy in infants with biliary atresia: a prospective trial. Ann Surg 253:826–830CrossRefPubMed
3.
Zurück zum Zitat Lishuang M, Zhen C, Guoliang Q, Zhen Z, Chen W, Long L, Shuli L (2015) Laparoscopic portoenterostomy versus open portoenterostomy for the treatment of biliary atresia: a systematic review and meta-analysis of comparative studies. Pediatr Surg Int 31:261–269CrossRefPubMed Lishuang M, Zhen C, Guoliang Q, Zhen Z, Chen W, Long L, Shuli L (2015) Laparoscopic portoenterostomy versus open portoenterostomy for the treatment of biliary atresia: a systematic review and meta-analysis of comparative studies. Pediatr Surg Int 31:261–269CrossRefPubMed
4.
Zurück zum Zitat Wada M, Nakamura H, Koga H, Miyano G, Lane GJ, Okazaki T, Urao M, Murakami H, Kasahara M, Sakamoto S, Ishizaki Y, Kawasaki S, Yamataka A (2014) Experience of treating biliary atresia with three types of portoenterostomy at a single institution: extended, modified Kasai, and laparoscopic modified Kasai. Pediatr Surg Int 30:863–870CrossRefPubMed Wada M, Nakamura H, Koga H, Miyano G, Lane GJ, Okazaki T, Urao M, Murakami H, Kasahara M, Sakamoto S, Ishizaki Y, Kawasaki S, Yamataka A (2014) Experience of treating biliary atresia with three types of portoenterostomy at a single institution: extended, modified Kasai, and laparoscopic modified Kasai. Pediatr Surg Int 30:863–870CrossRefPubMed
5.
Zurück zum Zitat Dutta S, Woo R, Albanese CT (2007) Minimal access portoenterostomy: advantages and disadvantages of standard laparoscopic and robotic techniques. J Laparoendosc Adv Surg Tech A 17:258–264CrossRefPubMed Dutta S, Woo R, Albanese CT (2007) Minimal access portoenterostomy: advantages and disadvantages of standard laparoscopic and robotic techniques. J Laparoendosc Adv Surg Tech A 17:258–264CrossRefPubMed
6.
Zurück zum Zitat Meehan JJ, Elliott S, Sandler A (2007) The robotic approach to complex hepatobiliary anomalies in children: preliminary report. J Pediatr Surg 42:2110–2114CrossRefPubMed Meehan JJ, Elliott S, Sandler A (2007) The robotic approach to complex hepatobiliary anomalies in children: preliminary report. J Pediatr Surg 42:2110–2114CrossRefPubMed
7.
Zurück zum Zitat Melich G, Hong YK, Kim J, Hur H, Baik SH, Kim NK, Sender Liberman A, Min BS (2015) Simultaneous development of laparoscopy and robotics provides acceptable perioperative outcomes and shows robotics to have a faster learning curve and to be overall faster in rectal cancer surgery: analysis of novice MIS surgeon learning curves. Surg Endosc 29:558–568CrossRefPubMed Melich G, Hong YK, Kim J, Hur H, Baik SH, Kim NK, Sender Liberman A, Min BS (2015) Simultaneous development of laparoscopy and robotics provides acceptable perioperative outcomes and shows robotics to have a faster learning curve and to be overall faster in rectal cancer surgery: analysis of novice MIS surgeon learning curves. Surg Endosc 29:558–568CrossRefPubMed
8.
Zurück zum Zitat Lee WJ, Chan CP, Wang BY (2013) Recent advances in laparoscopic surgery. Asian J Endosc Surg 6:1–8CrossRefPubMed Lee WJ, Chan CP, Wang BY (2013) Recent advances in laparoscopic surgery. Asian J Endosc Surg 6:1–8CrossRefPubMed
9.
10.
Zurück zum Zitat Di Lorenzo N, Camperchioli I, Gaspari AL (2007) Radius surgical system and conventional laparoscopic instruments in abdominal surgery: application, learning curve and ergonomy. Surg Oncol 16:S69–S72CrossRefPubMed Di Lorenzo N, Camperchioli I, Gaspari AL (2007) Radius surgical system and conventional laparoscopic instruments in abdominal surgery: application, learning curve and ergonomy. Surg Oncol 16:S69–S72CrossRefPubMed
11.
Zurück zum Zitat Waseda M, Inaki N, Torres Bermudez JR, Manukyan G, Gacek IA, Schurr MO, Braun M, Buess GF (2007) Precision in stitches: radius surgical system. Surg Endosc 21:2056–2062CrossRefPubMed Waseda M, Inaki N, Torres Bermudez JR, Manukyan G, Gacek IA, Schurr MO, Braun M, Buess GF (2007) Precision in stitches: radius surgical system. Surg Endosc 21:2056–2062CrossRefPubMed
12.
Zurück zum Zitat Frede T, Hammady A, Klein J, Teber D, Inaki N, Waseda M, Buess G, Rassweiler J (2007) The radius surgical system—a new device for complex minimally invasive procedures in urology? Eur Urol 51:1015–1022CrossRefPubMed Frede T, Hammady A, Klein J, Teber D, Inaki N, Waseda M, Buess G, Rassweiler J (2007) The radius surgical system—a new device for complex minimally invasive procedures in urology? Eur Urol 51:1015–1022CrossRefPubMed
13.
Zurück zum Zitat Torres Bermudez JR, Buess G, Waseda M, Gacek I, Becerra Garcia F, Manukyan GA, Inaki N (2009) Laparoscopic intracorporal colorectal sutured anastomosis using the Radius Surgical System in a phantom model. Surg Endosc 23:1624–1632CrossRefPubMed Torres Bermudez JR, Buess G, Waseda M, Gacek I, Becerra Garcia F, Manukyan GA, Inaki N (2009) Laparoscopic intracorporal colorectal sutured anastomosis using the Radius Surgical System in a phantom model. Surg Endosc 23:1624–1632CrossRefPubMed
14.
Zurück zum Zitat Hirano Y, Inaki N, Ishikawa N, Watanabe G (2013) Laparoscopic treatment for esophageal achalasia and gastro-esophago-reflex disease using radius surgical system. Indian J Surg 75:s160–s162CrossRef Hirano Y, Inaki N, Ishikawa N, Watanabe G (2013) Laparoscopic treatment for esophageal achalasia and gastro-esophago-reflex disease using radius surgical system. Indian J Surg 75:s160–s162CrossRef
15.
Zurück zum Zitat Hackethal A, Koppan M, Eskef K, Tinneberg HR (2012) Handheld articulating laparoscopic instruments driven by robotic technology. First clinical experience in gynecological surgery. Gynecol Surg 9:203–206CrossRef Hackethal A, Koppan M, Eskef K, Tinneberg HR (2012) Handheld articulating laparoscopic instruments driven by robotic technology. First clinical experience in gynecological surgery. Gynecol Surg 9:203–206CrossRef
16.
Zurück zum Zitat Bittner JG, Hathaway CA, Brown JA (2008) Three-dimensional visualisation and articulating instrumentation: impact on simulated laparoscopic tasks. J Minim Access Surg 4:31–38CrossRefPubMed Bittner JG, Hathaway CA, Brown JA (2008) Three-dimensional visualisation and articulating instrumentation: impact on simulated laparoscopic tasks. J Minim Access Surg 4:31–38CrossRefPubMed
17.
18.
Zurück zum Zitat Ohdaira T, Tsutsumi N, Xu H, Mori M, Uemura M, Ieiri S, Hashizume M (2011) Ultra-minimally invasive local immune cell therapy and regenerative therapy by multi-piercing surgery for abdominal solid tumor: therapeutic simulation by natural orifice translumenal endoscopic surgery-assisted needlescopic surgery using 3-mm diameter robots. J Hepatobiliary Pancreat Sci 18:499–505CrossRefPubMed Ohdaira T, Tsutsumi N, Xu H, Mori M, Uemura M, Ieiri S, Hashizume M (2011) Ultra-minimally invasive local immune cell therapy and regenerative therapy by multi-piercing surgery for abdominal solid tumor: therapeutic simulation by natural orifice translumenal endoscopic surgery-assisted needlescopic surgery using 3-mm diameter robots. J Hepatobiliary Pancreat Sci 18:499–505CrossRefPubMed
19.
Zurück zum Zitat Fujii M, Fukushima K, Sugita N, Ishimaru T, Iwanaka T, Mitsuishi M (2011) Design of intuitive user interface for multi-DOF forceps for laparoscopic surgery. In: International conference of robotics and automation, pp 5743–5748 Fujii M, Fukushima K, Sugita N, Ishimaru T, Iwanaka T, Mitsuishi M (2011) Design of intuitive user interface for multi-DOF forceps for laparoscopic surgery. In: International conference of robotics and automation, pp 5743–5748
20.
Zurück zum Zitat Fujii M, Sugita N, Ishimaru T, Iwanaka T, Mitsuishi M (2013) A novel approach to the design of a needle driver with multiple DOFs for pediatric laparoscopic surgery. Minim Invasive Ther Allied Technol 22:9–16CrossRefPubMed Fujii M, Sugita N, Ishimaru T, Iwanaka T, Mitsuishi M (2013) A novel approach to the design of a needle driver with multiple DOFs for pediatric laparoscopic surgery. Minim Invasive Ther Allied Technol 22:9–16CrossRefPubMed
21.
Zurück zum Zitat Ishimaru T, Takazawa S, Uchida H, Kawashima H, Fujii M, Harada K, Sugita N, Mitsuishi M, Iwanaka T (2013) Development of a needle driver with multiple degrees of freedom for neonatal laparoscopic surgery. J Laparoendosc Adv Surg Tech A 23:644–648CrossRefPubMed Ishimaru T, Takazawa S, Uchida H, Kawashima H, Fujii M, Harada K, Sugita N, Mitsuishi M, Iwanaka T (2013) Development of a needle driver with multiple degrees of freedom for neonatal laparoscopic surgery. J Laparoendosc Adv Surg Tech A 23:644–648CrossRefPubMed
22.
Zurück zum Zitat Takazawa S, Ishimaru T, Fujii M, Harada K, Sugita N, Mitsuishi M, Iwanaka T (2013) Assessment of suturing in the vertical plane shows the efficacy of the multi-degree-of-freedom needle driver for neonatal laparoscopy. Pediatr Surg Int 29:1177–1182CrossRefPubMed Takazawa S, Ishimaru T, Fujii M, Harada K, Sugita N, Mitsuishi M, Iwanaka T (2013) Assessment of suturing in the vertical plane shows the efficacy of the multi-degree-of-freedom needle driver for neonatal laparoscopy. Pediatr Surg Int 29:1177–1182CrossRefPubMed
23.
Zurück zum Zitat Takazawa S, Ishimaru T, Fujii M, Harada K, Fujishiro J, Konishi K, Miyakawa K, Kakihara T, Sugita N, Mitsuishi M, Iwanaka T (2014) Multi-DOF needle driver with a short, flexible distal tip is useful for multi-directional suturing and anastomosis in a small cavity: in vivo experiment. Int J Comput Assist Radiol Surg 9:S336–S337 Takazawa S, Ishimaru T, Fujii M, Harada K, Fujishiro J, Konishi K, Miyakawa K, Kakihara T, Sugita N, Mitsuishi M, Iwanaka T (2014) Multi-DOF needle driver with a short, flexible distal tip is useful for multi-directional suturing and anastomosis in a small cavity: in vivo experiment. Int J Comput Assist Radiol Surg 9:S336–S337
24.
Zurück zum Zitat De S, Rosen J, Dagan A, Hannaford B, Swanson P, Sinanan M (2007) Assessment of tissue damage due to mechanical stresses. Int J Rob Res 26:1159–1171CrossRef De S, Rosen J, Dagan A, Hannaford B, Swanson P, Sinanan M (2007) Assessment of tissue damage due to mechanical stresses. Int J Rob Res 26:1159–1171CrossRef
25.
Zurück zum Zitat MacDonald JD, Bowers CA, Chin SS, Burns G (2014) Comparison of the effects of surgical dissection devices on the rabbit liver. Surg Today 44:1116–1122CrossRefPubMed MacDonald JD, Bowers CA, Chin SS, Burns G (2014) Comparison of the effects of surgical dissection devices on the rabbit liver. Surg Today 44:1116–1122CrossRefPubMed
26.
Zurück zum Zitat Lorincz A, Knight CG, Langenburg SE, Rabah R, Gidell K, Dawe E, Grant S, Klein MD (2004) Robot-assisted minimally invasive Kasai portoenterostomy: a survival porcine study. Surg Endosc 18:1136–1139CrossRefPubMed Lorincz A, Knight CG, Langenburg SE, Rabah R, Gidell K, Dawe E, Grant S, Klein MD (2004) Robot-assisted minimally invasive Kasai portoenterostomy: a survival porcine study. Surg Endosc 18:1136–1139CrossRefPubMed
27.
Zurück zum Zitat Emam TA, Hanna GB, Kimber C, Dunkley P, Cuschieri A (2000) Effect of intracorporeal-extracorporeal instrument length ratio on endoscopic task performance and surgeon movements. Arch Surg 135:62–66CrossRefPubMed Emam TA, Hanna GB, Kimber C, Dunkley P, Cuschieri A (2000) Effect of intracorporeal-extracorporeal instrument length ratio on endoscopic task performance and surgeon movements. Arch Surg 135:62–66CrossRefPubMed
28.
Zurück zum Zitat Lee AC, Haddad MJ, Hanna GB (2007) Influence of instrument size on endoscopic task performance in pediatric intracorporeal knot tying: smaller instruments are better in infants. Surg Endosc 21:2086–2090CrossRefPubMed Lee AC, Haddad MJ, Hanna GB (2007) Influence of instrument size on endoscopic task performance in pediatric intracorporeal knot tying: smaller instruments are better in infants. Surg Endosc 21:2086–2090CrossRefPubMed
29.
Zurück zum Zitat Takemoto H, Tanaka K, Inomata Y, Matsukawa Y, Satomura K, Ozawa K (1989) Granulation at the porta hepatis following hepatic portoenterostomy for biliary atresia: the healing of experimental hepatoenterostomy. J Pediatr Surg 24:271–275CrossRefPubMed Takemoto H, Tanaka K, Inomata Y, Matsukawa Y, Satomura K, Ozawa K (1989) Granulation at the porta hepatis following hepatic portoenterostomy for biliary atresia: the healing of experimental hepatoenterostomy. J Pediatr Surg 24:271–275CrossRefPubMed
Metadaten
Titel
A multi-degree-of-freedom needle driver with a short tip and small shaft for pediatric laparoscopic surgery: in vivo assessment of multi-directional suturing on the vertical plane of the liver in rabbits
verfasst von
Shinya Takazawa
Tetsuya Ishimaru
Masahiro Fujii
Kanako Harada
Kyoichi Deie
Jun Fujishiro
Naohiko Sugita
Mamoru Mitsuishi
Tadashi Iwanaka
Publikationsdatum
28.10.2015
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 8/2016
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-015-4616-9

Weitere Artikel der Ausgabe 8/2016

Surgical Endoscopy 8/2016 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

CME: 2 Punkte

Prof. Dr. med. Gregor Antoniadis Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

CME: 2 Punkte

Dr. med. Benjamin Meyknecht, PD Dr. med. Oliver Pieske Das Webinar S2e-Leitlinie „Distale Radiusfraktur“ beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

CME: 2 Punkte

Dr. med. Mihailo Andric
Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.