A novel device for air removal from vascular access line: a bench study

verfasst von: Zeping Xu, Lingyan Jin, Byron Smith, Yiping Bai, Haoxiang Luo, Lars Axel Strombergsson, Min Fei, Yandong Jiang

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 6/2018

Einloggen, um Zugang zu erhalten

Abstract

Efficient air removal from a vascular access line is a key step to prevent air embolism. Existing devices, especially for rapid infusers, are far from optimum. In this study, we developed a novel device, vascular access line air removal device (VALARD), and compared its efficiency of air removal and pause time of forward bulk flow with a commonly used device, the Belmont pump. Part I experiment, saline was infused at a forward bulk flow rate of 250, 500, and 750 mL/min. Meanwhile, air was introduced into the infusion line at a rate of 5, 10, and 15 mL/min for each bulk flow rate. Air bubbles > 10 µL downstream from either the VALARD or the Belmont pump and the fraction of pause time of the forward bulk flow were determined. Part II experiment, 120 mL of air was rapidly introduced into the VALARD at a bulk flow rate of about 500 mL/min. Air bubbles > 10 µL downstream from the VALARD, fraction of pause time of the forward bulk flow, and the transit time of the 120 mL of air at the working chamber were recorded. The VALARD: no air bubbles > 10 µL were detected during any tested combination of air injection and bulk flow rates without pause of forward flow. The Belmont pump: air bubbles > 10 µL were detected in 60% of the tests with pause of the forward flow. The VALARD eliminates air efficiently without pause of the forward bulk flow. Further clinical trials are needed to compare the VALARD with other devices and to assess its efficiency, safety, and user friendliness.
Literatur
3.
Zurück zum Zitat Mirski MA, Lele AV, Fitzsimmons L, Toung TJ. Diagnosis and treatment of vascular air embolism. Anesthesiology. 2007;106(1):164–77.CrossRef Mirski MA, Lele AV, Fitzsimmons L, Toung TJ. Diagnosis and treatment of vascular air embolism. Anesthesiology. 2007;106(1):164–77.CrossRef
6.
Zurück zum Zitat Philip BK, Philip JH. Avoiding air infusion with pressurized infusion systems: a new hazard. Anesth Analg. 1985;64(3):381–2.CrossRef Philip BK, Philip JH. Avoiding air infusion with pressurized infusion systems: a new hazard. Anesth Analg. 1985;64(3):381–2.CrossRef
9.
Zurück zum Zitat Vesely TM. Air embolism during insertion of central venous catheters. J Vasc Interv Radiol. 2001;12(11):1291–5.CrossRef Vesely TM. Air embolism during insertion of central venous catheters. J Vasc Interv Radiol. 2001;12(11):1291–5.CrossRef
11.
Zurück zum Zitat Comunale ME. A laboratory evaluation of the level 1 rapid infuser (H1025) and the Belmont instrument fluid management system (FMS 2000) for rapid transfusion. Anesth Analg. 2003;97(4):1064–9.CrossRef Comunale ME. A laboratory evaluation of the level 1 rapid infuser (H1025) and the Belmont instrument fluid management system (FMS 2000) for rapid transfusion. Anesth Analg. 2003;97(4):1064–9.CrossRef
14.
Zurück zum Zitat Adhikary GS, Massey SR. Massive air embolism: a case report. J Clin Anesth. 1998;10(1):70–72.CrossRef Adhikary GS, Massey SR. Massive air embolism: a case report. J Clin Anesth. 1998;10(1):70–72.CrossRef
15.
Zurück zum Zitat Eaton MP, Dhillon AK. Relative performance of the level 1 and ranger pressure infusion devices. Anesth Analg. 2003;97(4):1074–7.CrossRef Eaton MP, Dhillon AK. Relative performance of the level 1 and ranger pressure infusion devices. Anesth Analg. 2003;97(4):1074–7.CrossRef
Metadaten
Titel
A novel device for air removal from vascular access line: a bench study
verfasst von
Zeping Xu
Lingyan Jin
Byron Smith
Yiping Bai
Haoxiang Luo
Lars Axel Strombergsson
Min Fei
Yandong Jiang
Publikationsdatum
17.02.2018
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 6/2018
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-018-0114-7

Weitere Artikel der Ausgabe 6/2018

Journal of Clinical Monitoring and Computing 6/2018 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.