A novel method to measure femoral component migration by computed tomography: a cadaver study

verfasst von: Friedrich Boettner, Peter Sculco, Joseph Lipman, Lisa Renner, Martin Faschingbauer

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 6/2016

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Radiostereometric analysis (RSA) is the most accurate technique to measure implant migration. However, it requires special equipment, technical expertise and analysis software and has not gained wide acceptance. The current paper analyzes a novel method to measure implant migration utilizing widely available computer tomography (CT).

Methods

Three uncemented total hip replacements were performed in three human cadavers and six tantalum beads were inserted into the femoral bone similar to RSA. Six different 28 mm heads (−3, 0, 2.5, 5.0, 7.5 and 10 mm) were added to simulate five reproducible translations (maximum total point migration) of the center of the head. Implant migration was measured in a 3-D analysis software (Geomagic Studio 7). Repeat manual reconstructions of the center of the head were performed by two investigators to determine repeatability and accuracy.

Results

The accuracy of measurements between the centers of two head sizes was 0.11 mm with a CI 95 % of 0.22 mm. The intra-observer repeatability was 0.13 mm (CI 95 % 0.25 mm). The interrater-reliability was 0.943.

Conclusion

CT based measurement of head displacement in a cadaver model were highly accurate and reproducible.
Literatur
1.
Zurück zum Zitat Freeman MAR P-B (1994) Early migration and late aseptic failure of proximal femoral prosteses. J Bone Jt Surg Br 76(3):432–438 Freeman MAR P-B (1994) Early migration and late aseptic failure of proximal femoral prosteses. J Bone Jt Surg Br 76(3):432–438
2.
Zurück zum Zitat Kiran A, Bottomley N, Biant LC, Javaid MK, Carr AJ, Cooper C, Field RE, Murray DW, Price A, Beard DJ, Arden NK (2015) Variations in good patient reported outcomes after total knee arthroplasty. J Arthroplast. doi:10.1016/j.arth.2015.02.039 Kiran A, Bottomley N, Biant LC, Javaid MK, Carr AJ, Cooper C, Field RE, Murray DW, Price A, Beard DJ, Arden NK (2015) Variations in good patient reported outcomes after total knee arthroplasty. J Arthroplast. doi:10.​1016/​j.​arth.​2015.​02.​039
3.
Zurück zum Zitat Karrholm J, Herberts P, Hultmark P, Malchau H, Nivbrant B, Thanner J (1997) Radiostereometry of hip prostheses. Review of methodology and clinical results. Clin Orthop Relat Res 344:94–110PubMed Karrholm J, Herberts P, Hultmark P, Malchau H, Nivbrant B, Thanner J (1997) Radiostereometry of hip prostheses. Review of methodology and clinical results. Clin Orthop Relat Res 344:94–110PubMed
4.
Zurück zum Zitat Walker PSMS, Cobb AG, Bentley G, Hua J (1995) Prediction of clinical outcome of THR from migration measurements on standard radiographs. J Bone Jt Surg Br 77:705–714 Walker PSMS, Cobb AG, Bentley G, Hua J (1995) Prediction of clinical outcome of THR from migration measurements on standard radiographs. J Bone Jt Surg Br 77:705–714
5.
Zurück zum Zitat Pijls BG, Nieuwenhuijse MJ, Fiocco M, Plevier JW, Middeldorp S, Nelissen RG, Valstar ER (2012) Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survival studies. Acta Orthop 83(6):583–591. doi:10.3109/17453674.2012.745353 CrossRefPubMedPubMedCentral Pijls BG, Nieuwenhuijse MJ, Fiocco M, Plevier JW, Middeldorp S, Nelissen RG, Valstar ER (2012) Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survival studies. Acta Orthop 83(6):583–591. doi:10.​3109/​17453674.​2012.​745353 CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Hanley JA, McNeil BJ (1982) The meaning and use of the area under a receiver operating characteristic (ROC) curve. Radiology 143:29–36CrossRefPubMed Hanley JA, McNeil BJ (1982) The meaning and use of the area under a receiver operating characteristic (ROC) curve. Radiology 143:29–36CrossRefPubMed
7.
Zurück zum Zitat Bojan AJ, Bragdon C, Jonsson A, Ekholm C, Karrholm J (2015) Three-dimensional bone-implant movements in trochanteric hip fractures: precision and accuracy of radiostereometric analysis in a phantom model. J Orthop Res. doi:10.1002/jor.22822 PubMed Bojan AJ, Bragdon C, Jonsson A, Ekholm C, Karrholm J (2015) Three-dimensional bone-implant movements in trochanteric hip fractures: precision and accuracy of radiostereometric analysis in a phantom model. J Orthop Res. doi:10.​1002/​jor.​22822 PubMed
9.
Zurück zum Zitat Olivecrona H, Olivecrona L, Weidenhielm L, Noz ME, Kardos J, Maguire GQ Jr, Zeleznik MP, Aspelin P (2008) A new technique for diagnosis of acetabular cup loosening using computed tomography: preliminary experience in 10 patients. Acta Orthop 79(3):346–353. doi:10.1080/17453670710015247 CrossRefPubMed Olivecrona H, Olivecrona L, Weidenhielm L, Noz ME, Kardos J, Maguire GQ Jr, Zeleznik MP, Aspelin P (2008) A new technique for diagnosis of acetabular cup loosening using computed tomography: preliminary experience in 10 patients. Acta Orthop 79(3):346–353. doi:10.​1080/​1745367071001524​7 CrossRefPubMed
10.
Zurück zum Zitat Olivecrona H, Weidenhielm L, Olivecrone L, Noz ME, Maguire JR, Zeleznik MP, Svensson L, Jonson T (2003) Spatial component position in total hip arthroplasty. Acta Radiol 44:84–91CrossRefPubMed Olivecrona H, Weidenhielm L, Olivecrone L, Noz ME, Maguire JR, Zeleznik MP, Svensson L, Jonson T (2003) Spatial component position in total hip arthroplasty. Acta Radiol 44:84–91CrossRefPubMed
11.
Zurück zum Zitat Olivecrona L, Crafoord J, Olivecrona H, Noz ME, Maguire GQ, Zeleznik MP, Svensson L, Weidenhielm L (2002) Acetabular component migration in total hip arthroplasty using ct and a semiautomated program for volume merging. Acta Radiol 43:517–527CrossRefPubMed Olivecrona L, Crafoord J, Olivecrona H, Noz ME, Maguire GQ, Zeleznik MP, Svensson L, Weidenhielm L (2002) Acetabular component migration in total hip arthroplasty using ct and a semiautomated program for volume merging. Acta Radiol 43:517–527CrossRefPubMed
12.
Zurück zum Zitat Ranstam J, Ryd L, Önsten I (2000) Accurate accuracy assessment. Acta Orthop Scand 71(1):106–108CrossRefPubMed Ranstam J, Ryd L, Önsten I (2000) Accurate accuracy assessment. Acta Orthop Scand 71(1):106–108CrossRefPubMed
13.
Zurück zum Zitat McGraw KO, Wong SP (1996) Forming inferences about some intraclass correlation coefficients. Psychol Methods 1:30–46CrossRef McGraw KO, Wong SP (1996) Forming inferences about some intraclass correlation coefficients. Psychol Methods 1:30–46CrossRef
14.
Zurück zum Zitat Iannotti JP, Balderston RA, Booth RE, Rothman RH, Cohn JC, Pickens G (1986) Aseptic loosening after total hip arthroplasty. Incidence, clinical significance, and etiology. J Arthroplast 1(2):99–107CrossRef Iannotti JP, Balderston RA, Booth RE, Rothman RH, Cohn JC, Pickens G (1986) Aseptic loosening after total hip arthroplasty. Incidence, clinical significance, and etiology. J Arthroplast 1(2):99–107CrossRef
15.
Zurück zum Zitat Nunn DFM, Hill PF, Evans SJW (1989) The measurement of migration of the acetabular component of hip prostheses. J Bone Jt Surg Br 71:629–631 Nunn DFM, Hill PF, Evans SJW (1989) The measurement of migration of the acetabular component of hip prostheses. J Bone Jt Surg Br 71:629–631
16.
Zurück zum Zitat Sutherland CJ, Bresina SJ (1992) Measurement of acetabular component migration using two-dimensional radiography. J Arthroplast 7(Suppl):377–379CrossRef Sutherland CJ, Bresina SJ (1992) Measurement of acetabular component migration using two-dimensional radiography. J Arthroplast 7(Suppl):377–379CrossRef
17.
Zurück zum Zitat Wetherell RGAA, Heatley FW (1989) Measurement of Acetabular erosion. J Bone Jt Surg Br 71:447–451 Wetherell RGAA, Heatley FW (1989) Measurement of Acetabular erosion. J Bone Jt Surg Br 71:447–451
18.
Zurück zum Zitat Phillips NJ, Stockley I, Wilkinson JM (2002) Direct plain radiographic methods versus EBRA-Digital for measuring implant migration after total hip arthroplasty. J Arthroplast 17(7):917–925. doi:10.1054/arth.2002.34529 CrossRef Phillips NJ, Stockley I, Wilkinson JM (2002) Direct plain radiographic methods versus EBRA-Digital for measuring implant migration after total hip arthroplasty. J Arthroplast 17(7):917–925. doi:10.​1054/​arth.​2002.​34529 CrossRef
19.
Zurück zum Zitat Dominkus M, Wanivenhaus AH, Morscher M, Powell G, Krismer M, Wolfl G (1998) Different cup migration in rheumatoid arthritis and arthrosis: a radiographic analysis of 127 uncemented acetabular cups. Acta Orthop Scand 69(5):455–462CrossRefPubMed Dominkus M, Wanivenhaus AH, Morscher M, Powell G, Krismer M, Wolfl G (1998) Different cup migration in rheumatoid arthritis and arthrosis: a radiographic analysis of 127 uncemented acetabular cups. Acta Orthop Scand 69(5):455–462CrossRefPubMed
20.
Zurück zum Zitat Hendrich CBJ, Eulert J (1997) Migration of the uncemented Harris-Galante acetabular cup. J Arthroplast 12(8):889–895CrossRef Hendrich CBJ, Eulert J (1997) Migration of the uncemented Harris-Galante acetabular cup. J Arthroplast 12(8):889–895CrossRef
21.
Zurück zum Zitat Ilchmann TML, Joshi A, Hardinge K, Murphy J, Wingstrand H (1998) Migration and wear of long-term successful Charnley total hip replacements. J Bone Jt Surg Br 80(3):377–381CrossRef Ilchmann TML, Joshi A, Hardinge K, Murphy J, Wingstrand H (1998) Migration and wear of long-term successful Charnley total hip replacements. J Bone Jt Surg Br 80(3):377–381CrossRef
22.
Zurück zum Zitat Krismer MBR, Stöckl B, Fischer M, Bauer R, Haid C (1999) The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. J Bone Jt Surg Br 81:273–280CrossRef Krismer MBR, Stöckl B, Fischer M, Bauer R, Haid C (1999) The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. J Bone Jt Surg Br 81:273–280CrossRef
23.
Zurück zum Zitat Krismer M, Fischer M, Mayrhofer P, Stockl F, Bittner C, Trojer C, Stockl B (1994) A prospective study of the migration of two acetabular components PCA versus RM cups. Int Orthop 18(1):23–28CrossRefPubMed Krismer M, Fischer M, Mayrhofer P, Stockl F, Bittner C, Trojer C, Stockl B (1994) A prospective study of the migration of two acetabular components PCA versus RM cups. Int Orthop 18(1):23–28CrossRefPubMed
24.
Zurück zum Zitat Roffman M, Kligman M (1999) Cementless coated and noncoated Mathys acetabular cups: radiographic and histologic evaluation. Orthopedics 22(1):39–41PubMed Roffman M, Kligman M (1999) Cementless coated and noncoated Mathys acetabular cups: radiographic and histologic evaluation. Orthopedics 22(1):39–41PubMed
25.
Zurück zum Zitat Biedermann R, Krismer M, Stöckl B, Mayrhofer P, Ornstein E, Franzen H (1999) Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement. J Bone Jt Surg Br 81:266–272CrossRef Biedermann R, Krismer M, Stöckl B, Mayrhofer P, Ornstein E, Franzen H (1999) Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement. J Bone Jt Surg Br 81:266–272CrossRef
26.
Zurück zum Zitat Krismer MBR, Tschupik J, Mayrhofer P (1995) EBRA: a Method to measure migration of acetabular components. J Biomech 28(10):1225–1236CrossRefPubMed Krismer MBR, Tschupik J, Mayrhofer P (1995) EBRA: a Method to measure migration of acetabular components. J Biomech 28(10):1225–1236CrossRefPubMed
27.
Zurück zum Zitat Selvik G (1989) Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop Scand 60(Sup232):1–51CrossRef Selvik G (1989) Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop Scand 60(Sup232):1–51CrossRef
30.
Zurück zum Zitat Malchau HPH (2008) How are wear-related problems diagnosed and what forms of surveillance are necessary? J Am Acad Orthop Surg 16:14–19CrossRef Malchau HPH (2008) How are wear-related problems diagnosed and what forms of surveillance are necessary? J Am Acad Orthop Surg 16:14–19CrossRef
31.
Zurück zum Zitat Jedenmalm A, Noz ME, Olivecrona H, Olivecrona L, Stark A (2008) A new approach for assessment of wear in metal-backed acetabular cups using computed tomography: a phantom study with retrievals. Acta orthopaedica 79(2):218–224. doi:10.1080/17453670710015003 CrossRefPubMed Jedenmalm A, Noz ME, Olivecrona H, Olivecrona L, Stark A (2008) A new approach for assessment of wear in metal-backed acetabular cups using computed tomography: a phantom study with retrievals. Acta orthopaedica 79(2):218–224. doi:10.​1080/​1745367071001500​3 CrossRefPubMed
32.
Zurück zum Zitat Goldvasser D, Noz ME, Maguire GQ Jr, Olivecrona H, Bragdon CR, Malchau H (2012) A new technique for measuring wear in total hip arthroplasty using computed tomography. J Arthroplast 27(9):1636–1640. doi:10.1016/j.arth.2012.03.053 (e1631)CrossRef Goldvasser D, Noz ME, Maguire GQ Jr, Olivecrona H, Bragdon CR, Malchau H (2012) A new technique for measuring wear in total hip arthroplasty using computed tomography. J Arthroplast 27(9):1636–1640. doi:10.​1016/​j.​arth.​2012.​03.​053 (e1631)CrossRef
33.
Zurück zum Zitat Karrholm J (1994) Does early micromotion of femoral protheses matter? J Bone Jt Surg Br 76:912–991 Karrholm J (1994) Does early micromotion of femoral protheses matter? J Bone Jt Surg Br 76:912–991
Metadaten
Titel
A novel method to measure femoral component migration by computed tomography: a cadaver study
verfasst von
Friedrich Boettner
Peter Sculco
Joseph Lipman
Lisa Renner
Martin Faschingbauer
Publikationsdatum
16.03.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 6/2016
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-016-2442-8

Weitere Artikel der Ausgabe 6/2016

Archives of Orthopaedic and Trauma Surgery 6/2016 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.