18.03.2016

A Practice Quality Improvement Project: Reducing Dose of Routine Chest CT Imaging in a Busy Clinical Practice

verfasst von: Edwin A. Takahashi, Marc D. Kohli, Shawn D. Teague

Erschienen in: Journal of Imaging Informatics in Medicine | Ausgabe 5/2016

Einloggen, um Zugang zu erhalten

Abstract

The purpose of this report is to describe our experience with the implementation of a practice quality improvement (PQI) project in thoracic imaging as part of the American Board of Radiology Maintenance of Certification process. The goal of this PQI project was to reduce the effective radiation dose of routine chest CT imaging in a busy clinical practice by employing the iDose4 (Philips Healthcare) iterative reconstruction technique. The dose reduction strategy was implemented in a stepwise process on a single 64-slice CT scanner with a volume of 1141 chest CT scans during the year. In the first annual quarter, a baseline effective dose was established using the standard filtered back projection (FBP) algorithm protocol and standard parameters such as kVp and mAs. The iDose4 technique was then applied in the second and third annual quarters while keeping all other parameters unchanged. In the fourth quarter, a reduction in kVp was also implemented. Throughout the process, the images were continually evaluated to assure that the image quality was comparable to the standard protocol from multiple other scanners. Utilizing a stepwise approach, the effective radiation dose was reduced by 23.62 and 43.63 % in quarters two and four, respectively, compared to our initial standard protocol with no perceived difference in diagnostic quality. This practice quality improvement project demonstrated a significant reduction in the effective radiation dose of thoracic CT scans in a busy clinical practice.
Literatur
1.
Zurück zum Zitat Hawkins CM, Alsip CN, Pryor RM, et al: Quality improvement and confirmation projects: facilitating rapid, measurable performance improvement. Radiographics 33(7):e225–35, 2013CrossRefPubMed Hawkins CM, Alsip CN, Pryor RM, et al: Quality improvement and confirmation projects: facilitating rapid, measurable performance improvement. Radiographics 33(7):e225–35, 2013CrossRefPubMed
2.
Zurück zum Zitat Strife JL, Kun LE, Becker GJ, et al: The American board of radiology perspective on maintenance of certification: part IV—practice quality improvement for diagnostic radiology. Radiology 243(2):309–13, 2007CrossRefPubMed Strife JL, Kun LE, Becker GJ, et al: The American board of radiology perspective on maintenance of certification: part IV—practice quality improvement for diagnostic radiology. Radiology 243(2):309–13, 2007CrossRefPubMed
4.
Zurück zum Zitat Griffith B, Brown ML, Jain R: Improving imaging utilization through practice quality improvement (maintenance of certification part IV): a review of requirements and approach to implementation. AJR Am J Roentgenol 202(4):797–802, 2014CrossRefPubMed Griffith B, Brown ML, Jain R: Improving imaging utilization through practice quality improvement (maintenance of certification part IV): a review of requirements and approach to implementation. AJR Am J Roentgenol 202(4):797–802, 2014CrossRefPubMed
5.
Zurück zum Zitat Kouo T: Experience with a practice quality improvement system in a university radiology department. J Am Coll Radiol 9(11):814–9, 2012CrossRefPubMed Kouo T: Experience with a practice quality improvement system in a university radiology department. J Am Coll Radiol 9(11):814–9, 2012CrossRefPubMed
6.
Zurück zum Zitat Keegan J, Miglioretti DL, Gould R, et al: Radiation dose metrics in CT: assessing dose using the National Quality Forum CT patient safety measure. J Am Coll Radiol 11(3):309–15, 2014CrossRefPubMed Keegan J, Miglioretti DL, Gould R, et al: Radiation dose metrics in CT: assessing dose using the National Quality Forum CT patient safety measure. J Am Coll Radiol 11(3):309–15, 2014CrossRefPubMed
7.
Zurück zum Zitat Choudhery S, Richter M, Anene A, et al: Practice quality improvement during residency: where do we stand and where can we improve? Acad Radiol 21(7):851–58, 2014CrossRefPubMed Choudhery S, Richter M, Anene A, et al: Practice quality improvement during residency: where do we stand and where can we improve? Acad Radiol 21(7):851–58, 2014CrossRefPubMed
8.
9.
Zurück zum Zitat Tamm EP, Szklaruk J, Puthooran L, et al: Quality initiatives: planning, setting up, and carrying out radiology process improvement projects. Radiographics 32(5):1529–42, 2012CrossRefPubMed Tamm EP, Szklaruk J, Puthooran L, et al: Quality initiatives: planning, setting up, and carrying out radiology process improvement projects. Radiographics 32(5):1529–42, 2012CrossRefPubMed
10.
Zurück zum Zitat Lee CS, Larson DB: Beginner’s guide to practice quality improvement using the model for improvement. J Am Coll Radiol 11(12ptA):1131–6, 2014CrossRefPubMed Lee CS, Larson DB: Beginner’s guide to practice quality improvement using the model for improvement. J Am Coll Radiol 11(12ptA):1131–6, 2014CrossRefPubMed
11.
Zurück zum Zitat Hirsch JA, Becker GJ, Derdeyn CP, et al: Maintenance of certification: part 2-continuous certification. J Neurointerv Surg 6(2):156–60, 2014CrossRefPubMed Hirsch JA, Becker GJ, Derdeyn CP, et al: Maintenance of certification: part 2-continuous certification. J Neurointerv Surg 6(2):156–60, 2014CrossRefPubMed
13.
Zurück zum Zitat Brenner DJ, Doll R, Goodhead DT, et al: Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. PNAS 100(24):13761–6, 2003CrossRefPubMedPubMedCentral Brenner DJ, Doll R, Goodhead DT, et al: Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. PNAS 100(24):13761–6, 2003CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Takahashi EA, Yoon HC: Four-year cumulative radiation exposure in patients undergoing computed tomography angiography for suspected pulmonary embolism. Radiol Res Pract. 2013:482403, 2013PubMedPubMedCentral Takahashi EA, Yoon HC: Four-year cumulative radiation exposure in patients undergoing computed tomography angiography for suspected pulmonary embolism. Radiol Res Pract. 2013:482403, 2013PubMedPubMedCentral
15.
Zurück zum Zitat Beister M, Kolditz D, Kalender WA: Iterative reconstruction methods in X-ray CT. Phys Med 28(2):94–108, 2012CrossRefPubMed Beister M, Kolditz D, Kalender WA: Iterative reconstruction methods in X-ray CT. Phys Med 28(2):94–108, 2012CrossRefPubMed
16.
Zurück zum Zitat Gonzalez-Guindalini FD, Ferreira Botelho MP, Tore HG, et al: MDCT of chest, abdomen, and pelvis using attenuation-based automated tube voltage selection in combination with iterative reconstruction: an intrapatient study of radiation dose and image quality. AJR Am J Roentgenol 201(5):1075–82, 2013CrossRefPubMed Gonzalez-Guindalini FD, Ferreira Botelho MP, Tore HG, et al: MDCT of chest, abdomen, and pelvis using attenuation-based automated tube voltage selection in combination with iterative reconstruction: an intrapatient study of radiation dose and image quality. AJR Am J Roentgenol 201(5):1075–82, 2013CrossRefPubMed
17.
18.
Zurück zum Zitat Lee SW, Kim Y, Shim SS, et al: Image quality assessment of ultra low-dose chest CT using sonogram-affirmed iterative reconstruction. Eur Radiol 24(4):817–26, 2014CrossRefPubMed Lee SW, Kim Y, Shim SS, et al: Image quality assessment of ultra low-dose chest CT using sonogram-affirmed iterative reconstruction. Eur Radiol 24(4):817–26, 2014CrossRefPubMed
19.
Zurück zum Zitat Ning P, Zhu S, Shi D, et al: X-ray dose reduction in abdominal computed tomography using advanced iterative reconstruction algorithms. PLoS One 9(3):e92568, 2014CrossRefPubMedPubMedCentral Ning P, Zhu S, Shi D, et al: X-ray dose reduction in abdominal computed tomography using advanced iterative reconstruction algorithms. PLoS One 9(3):e92568, 2014CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Matsuki M, Murakami T, Juri H, et al: Impact of adaptive iterative dose reduction (AIDR) 3D on low-dose abdominal CT: comparison with routine-dose CT using filtered back projection. Acta Radiol 54(8):869–75, 2013CrossRefPubMed Matsuki M, Murakami T, Juri H, et al: Impact of adaptive iterative dose reduction (AIDR) 3D on low-dose abdominal CT: comparison with routine-dose CT using filtered back projection. Acta Radiol 54(8):869–75, 2013CrossRefPubMed
21.
Zurück zum Zitat Kim M, Lee JM, Yoon JH, et al: Adaptive iterative dose reduction algorithm in CT: effect on image quality compared to filtered back projection in body phantoms of different sizes. Korean J Radiol 15(2):195–204, 2014CrossRefPubMedPubMedCentral Kim M, Lee JM, Yoon JH, et al: Adaptive iterative dose reduction algorithm in CT: effect on image quality compared to filtered back projection in body phantoms of different sizes. Korean J Radiol 15(2):195–204, 2014CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Prakash P, Kalra MK, Digumarthy SR, et al: Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience. J Comput Assist Tomogr 34(1):40–5, 2010CrossRefPubMed Prakash P, Kalra MK, Digumarthy SR, et al: Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience. J Comput Assist Tomogr 34(1):40–5, 2010CrossRefPubMed
23.
Zurück zum Zitat Noel PB, Renger B, Fiebich M, et al: Does iterative reconstruction lower CT radiation dose: evaluation of 15,000 examinations. PLoS One 8(11):e81141, 2013CrossRefPubMedPubMedCentral Noel PB, Renger B, Fiebich M, et al: Does iterative reconstruction lower CT radiation dose: evaluation of 15,000 examinations. PLoS One 8(11):e81141, 2013CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Arapakis I, Efstathopoulos E, Tsitsia V, et al: Using “iDose4” iterative reconstruction algorithm in adults’ chest-abdomen-pelvis CT examinations: effect on image quality in relation to patient radiation exposure. Br J Radiol. 87:20130613, 2014CrossRefPubMedPubMedCentral Arapakis I, Efstathopoulos E, Tsitsia V, et al: Using “iDose4” iterative reconstruction algorithm in adults’ chest-abdomen-pelvis CT examinations: effect on image quality in relation to patient radiation exposure. Br J Radiol. 87:20130613, 2014CrossRefPubMedPubMedCentral
25.
Zurück zum Zitat Lagmani A, Buhk JH, Henes FO, et al: Impact of a 4th generation iterative reconstruction technique on image quality in low-dose computed tomography of the chest in immunocompromised patients. Rofo 185(8):749–57, 2013CrossRef Lagmani A, Buhk JH, Henes FO, et al: Impact of a 4th generation iterative reconstruction technique on image quality in low-dose computed tomography of the chest in immunocompromised patients. Rofo 185(8):749–57, 2013CrossRef
26.
Zurück zum Zitat Li Q, Yu H, Zhang L, et al: Combining low tube voltage and iterative reconstruction for contrast-enhanced CT imaging of the chest—initial clinical experience. Clin Radiol 68(5):e249–53, 2013CrossRefPubMed Li Q, Yu H, Zhang L, et al: Combining low tube voltage and iterative reconstruction for contrast-enhanced CT imaging of the chest—initial clinical experience. Clin Radiol 68(5):e249–53, 2013CrossRefPubMed
27.
Zurück zum Zitat Cook TS, Zimmerman S, Maidment AD, et al: Automated extraction of radiation dose information for CT examinations. J Am Coll Radiol 7(11):871–7, 2010CrossRefPubMed Cook TS, Zimmerman S, Maidment AD, et al: Automated extraction of radiation dose information for CT examinations. J Am Coll Radiol 7(11):871–7, 2010CrossRefPubMed
28.
Zurück zum Zitat Yousem DM, Nidecker A: Maintenance of certification: update on attitudes of members of the American Society of Neuroradiology. AJNR Am J Neuroradiol 31(8):1369–72, 2010CrossRefPubMed Yousem DM, Nidecker A: Maintenance of certification: update on attitudes of members of the American Society of Neuroradiology. AJNR Am J Neuroradiol 31(8):1369–72, 2010CrossRefPubMed
29.
Zurück zum Zitat Khawaja RD, Singh S, Gilman M, et al: Computed tomography (CT) of the chest at less than 1 mSv: an ongoing prospective clinical trial of chest CT at submillisievert radiation doses with iterative model image reconstruction and iDose4 technique. J Comput Assist Tomogr 38(4):613–9, 2014CrossRefPubMed Khawaja RD, Singh S, Gilman M, et al: Computed tomography (CT) of the chest at less than 1 mSv: an ongoing prospective clinical trial of chest CT at submillisievert radiation doses with iterative model image reconstruction and iDose4 technique. J Comput Assist Tomogr 38(4):613–9, 2014CrossRefPubMed
Metadaten
Titel
A Practice Quality Improvement Project: Reducing Dose of Routine Chest CT Imaging in a Busy Clinical Practice
verfasst von
Edwin A. Takahashi
Marc D. Kohli
Shawn D. Teague
Publikationsdatum
18.03.2016
Verlag
Springer International Publishing
Erschienen in
Journal of Imaging Informatics in Medicine / Ausgabe 5/2016
Print ISSN: 2948-2925
Elektronische ISSN: 2948-2933
DOI
https://doi.org/10.1007/s10278-016-9877-x

Weitere Artikel der Ausgabe 5/2016

Journal of Digital Imaging 5/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.