Introduction
The two most widely used surgical procedures for the treatment of osteoarthritis (OA) of the trapeziometacarpal (TMC) joint are the trapeziectomy and TMC joint arthroplasty. In recently published research it is hypothesized that TMC joint arthroplasty is superior compared to trapeziectomy in terms of pain, strength, range of motion (ROM), satisfaction and recovery [
1‐
5]. However, most studies present short-term follow-up and thus long-term data are of interest.
Previously, our group presented the 5-year results of a radiostereometric analysis (RSA) of the Surface Replacement (SR) TMC joint prosthesis (SR™TMC, Avanta®, San Diego, CA) [
6]. This study showed a survival of eight out of ten prostheses with satisfying clinical outcomes.
The experience with RSA in the TMC joint is limited: two experimental and two small clinical studies have been published up to now [
6‐
9]. These studies have learned us that RSA of the TMC joint is feasible with high precision for translations, but precision for rotation measurement is poor. Long-term RSA studies of the TMC joint have not been published before and thus long-term migration data of TMC joint prostheses are unknown.
The aim of the present study is to determine long-term survival and clinical outcomes of the SR™TMC joint prosthesis 10 years after placement and to evaluate the migration of the prosthesis during follow-up.
Discussion
In this first long-term RSA-study of the TMC joint we present 10-year survival rate, clinical outcomes and migration of the SR™TMC-joint prosthesis. After 10 years of follow-up, two out of ten prostheses were revised and two additional loose implants were found at the 10-year follow-up moment. Long-term survival is worse compared to recently published long-term results of the Roseland®, ARPE®, Ivory® and Rubis II prostheses, varying from 85 to 95% [
13‐
18]. This is mainly explained by two early revisions. Migration analysis of these implants showed a stable fixation of the trapezium component in the first 12 months postoperatively and both implants turned out to be well fixated during revision surgery. Therefore, both revisions were not considered as loose implants. No additional implants were revised during further follow-up. However, at 10-year follow-up, two patients (patient 5 and 9) had clinical signs of loosening including pain and loss of function. Clinical suspicion was supported in patient 9 by RSA as we found the implant migrating substantially, while conventional radiographs did not show any sign of loosening. Loosening could not be confirmed by RSA in patient 5 since analysis of the RSA radiographs was inaccurate as a result of unstable markers.
Six patients were highly satisfied with high DASH and Nelson scores and low VAS pain scores. Especially key pinch grip was high in these patients and comparable with pinch grip in the normal population corrected for sex and age. These satisfying results are comparable with previously published long-term results [
13‐
18].
Several studies have been published comparing total joint arthroplasty with trapeziectomy, but only reporting short-term results [
1‐
5]. Jager et al. described higher satisfaction, mobility, strength, pain reduction and functional scores in favor of the MAIA® total joint prosthesis [
1]. Robles-Molina et al. reported similar pain relief and functional improvement, but superior pinch strength and range of motion in the ARPE group [
2]. Besides a significantly better ROM, pinch strength, DASH, pain relief and satisfaction, Cebrian-Gomez et al. described a faster return to work in the Ivory prosthesis group [
3]. Unlike the results of Cebrian-Gomez et al., Thorkildsen and Røkkum did not find any significant difference in DASH scores between TMC joint arthroplasty and trapeziectomy but did find better motion and strength in the prosthesis group [
4].
On the other hand, most studies show higher complication and revision rates in total joint arthroplasty. Taking this into consideration, together with the assumed higher costs of TMC joint arthroplasty in comparison with trapeziectomy, the optimal surgical treatment for TMC joint OA remains a topic of debate. In our opinion further research should be done to investigate which individual patients do actually have benefit from the described advantages of total joint arthroplasty and which do not.
Worth noticing is that all controlled trials comparing trapeziectomy and total joint arthroplasty have investigated ball-and-socket design implants and not saddle-shaped SR implants as used in this study. The SR TMC joint prosthesis has been developed to preserve normal anatomy and kinematics of the thumb, striving for better survival [
19]. Although controlled trials comparing the two implants have not been performed, short-term survival of the SR implant appeared to be inferior to ball-and-socket implants whereafter the prosthesis was withdrawn from the market [
20]. Apart from our study, no long-term results of the SR TMC joint prosthesis are available to compare with long-term results of ball-and-socket designs. Given the lack of long-term survival data and the absence of controlled studies, no firm conclusions can be drawn about superiority of one of both implants in the long term.
Concerning the role of RSA in TMC joint arthroplasty scientific support is limited. RSA studies analyzing TMC joint prostheses are sparse and patient cohorts are small. Furthermore, the technique faces some significant challenges that have to do with the small size of the joint. As in previously published studies, it was not possible to calculate rotations of the implant in our study. The main reason for this is the lack of stable (ME < 0.35), sufficient (
N > 2) and well three-dimensional spread (CN < 150) markers. Using marker rigid bodies containing unstable and poor spread markers results in large and incorrect rotations. Simply ignoring these rotations is not the right strategy since rotations do affect translation calculation. An incorrect rotational alignment of the reference model can lead to incorrect rotations and hence inaccurate translation measurement. This effect of rotation on translation is explained by Beardsley et al. and Van Hamersveld et al. [
21,
22] Therefore, striving for a well three-dimensional spread and stable rigid body as a reference for migration calculation is still important, even if rotations are left out of consideration. However, the small size of the surrounding bone makes it difficult to ensure this in the TMC joint. In cases were RSA data cannot be analyzed due to said reasons the alternative strategy of
reversed migration calculation as applied in this study may be a better option for two reasons.
First, when the implant is used as the reference model, RSA scenes with only two bone markers can be used for translation calculation. The reference model can be used to correct for the different positions of the TMC joint with respect to the calibration cage in different RSA acquisitions. With only two markers in the reference model, this cannot be accurately done since two markers create a line around which the implant model can be rotated 360 degrees, resulting in potentially inaccurate migrations.
Secondly, for marker rigid bodies with a CN > 150 similar issue arises. The higher the CN, the more the markers in the rigid body are aligned in a column like fashion, resulting in similar rotational inaccuracies as described above (Fig.
1). The strategy of reversed migration calculation will not solve all marker-related problems but does allow for more data to be analyzed. Furthermore, an improvement can be expected in translation measurements when marker rigid bodies have a high CN.
To avoid the problem of invisible or unstable markers, a CT-based method has been developed with comparable accuracy and precision to that of RSA of the hip [
23]. More recently, Broden et al. described the use of a CT-based method to calculate migration of shoulder implants [
24]. In an experimental setting, accuracy and precision were comparable to that of RSA with similar effective doses. Given the mentioned marker-related problems, accuracy and precision of CT-based migration calculation is worth to be investigated in the TMC joint.
Considering the proved predictive value of early migration in hip- and knee arthroplasty for future loosening, RSA plays an important role in the introduction and surveillance of orthopedic implants [
25‐
27]. However, the relation between early migration and future loosening has not yet been proved in other joints than the hip and knee. If existent, this relation will be difficult to demonstrate in TMC joint arthroplasty because of small patient numbers. Thus, the question arises ‘What to do with RSA in the TMC joint?’
An important principle in total joint replacement is to strive for the best possible fixation of implants into the surrounding bone. Although the predictive value of early migration of TMC joint implants is unclear, RSA remains the most accurate method available to assess implant migration and fixation. Comparing different implant designs, RSA may play an important role in the early distinction of good and bad performing implants preventing implants with suboptimal fixation coming into the market.
A not so often discussed feature of RSA is the use of the technique as a diagnostic tool in individual cases. In daily practice, confirming clinical suspicion of loosening may be challenging, expensive and time consuming. Generally, the first step in the diagnostic algorithm of implant failure is taking conventional radiographs. However, the value of conventional radiographs in diagnosing loosening is limited [
28]. Magnetic Resonance Imaging (MRI) and Computed Tomography (CT) scans to assess implant loosening are more expensive and generally difficult to read because of metal artefacts [
29]. Additional bone scintigraphy and SPECT may be helpful but are expensive and time consuming.
In our study we found two patients with clinical symptoms of loosening. Conventional radiographs did not show signs of loosening, while RSA enabled us to easily confirm implant migration in one of both patients. Although the numbers in this study are too small to come to conclusions, future research comparing diagnostic accuracy and cost-effectiveness of RSA versus other diagnostic modalities in the detection of implant loosening could be interesting. The diagnostic value of RSA could be investigated in patients who are already involved in RSA studies and could undergo both RSA examinations and other diagnostic tests in case of loosening.. Alternatively, tantalum beads could be implanted in new patient cohorts undergoing total joint arthroplasty. After obtaining two postoperative RSA radiographs as reference and to demonstrate stabilization of implants, RSA radiographs could be repeated in case of clinically suspicion of loosening during follow-up.