01.05.2014 | Breast Oncology | Ausgabe 5/2014 Open Access

Annals of Surgical Oncology 5/2014

A Randomized Prospective Study of Lumpectomy Margin Assessment with Use of MarginProbe in Patients with Nonpalpable Breast Malignancies

Zeitschrift:
Annals of Surgical Oncology > Ausgabe 5/2014
Autoren:
MD Freya Schnabel, MD Susan K. Boolbol, MD Mark Gittleman, MD Tami Karni, MD Lorraine Tafra, MD Sheldon Feldman, MD Alice Police, MD Neil B. Friedman, MD Scott Karlan, MD Dennis Holmes, MD Shawna C. Willey, MD Moshe Carmon, MD Kristen Fernandez, MD Stephanie Akbari, MD Jay Harness, MD Lisa Guerra, MD Thomas Frazier, MD Karen Lane, MD Rache M. Simmons, MD Alison Estabrook, MD Tanir Allweis

Abstract

Background

The presence of tumor cells at the margins of breast lumpectomy specimens is associated with an increased risk of ipsilateral tumor recurrence. Twenty to 30 % of patients undergoing breast-conserving surgery require second procedures to achieve negative margins. This study evaluated the adjunctive use of the MarginProbe device (Dune Medical Devices Ltd, Caesarea, Israel) in providing real-time intraoperative assessment of lumpectomy margins.

Methods

This multicenter randomized trial enrolled patients with nonpalpable breast malignancies. The study evaluated MarginProbe use in addition to standard intraoperative methods for margin assessment. After specimen removal and inspection, patients were randomized to device or control arms. In the device arm, MarginProbe was used to examine the main lumpectomy specimens and direct additional excision of positive margins. Intraoperative imaging was used in both arms; no intraoperative pathology assessment was permitted.

Results

In total, 596 patients were enrolled. False-negative rates were 24.8 and 66.1 % and false-positive rates were 53.6 and 16.6 % in the device and control arms, respectively. All positive margins on positive main specimens were resected in 62 % (101 of 163) of cases in the device arm, versus 22 % (33 of 147) in the control arm (p < 0.001). A total of 19.8 % (59 of 298) of patients in the device arm underwent a reexcision procedure compared with 25.8 % (77 of 298) in the control arm (6 % absolute, 23 % relative reduction). The difference in tissue volume removed was not significant.

Conclusions

Adjunctive use of the MarginProbe device during breast-conserving surgery improved surgeons’ ability to identify and resect positive lumpectomy margins in the absence of intraoperative pathology assessment, reducing the number of patients requiring reexcision. MarginProbe may aid performance of breast-conserving surgery by reducing the burden of reexcision procedures for patients and the health care system.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Bis zum 22.10. bestellen und 100 € sparen!

e.Med Chirurgie

Kombi-Abonnement

Mit e.Med Chirurgie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Chirurgie, den Premium-Inhalten der chirurgischen Fachzeitschriften, inklusive einer gedruckten chirurgischen Zeitschrift Ihrer Wahl.

Nicht verpassen: e.Med bis 22. Oktober 100 € günstiger.

e.Med Onkologie

Kombi-Abonnement

Mit e.Med Onkologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Onkologie, den Premium-Inhalten der onkologischen Fachzeitschriften, inklusive einer gedruckten onkologischen Zeitschrift Ihrer Wahl.

Nicht verpassen: e.Med bis 22. Oktober 100 € günstiger.

Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 5/2014

Annals of Surgical Oncology 5/2014Zur Ausgabe

Neu im Fachgebiet Chirurgie

 

 

 
 

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Chirurgie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise