A review of the atrial upper rate algorithms of St. Jude Medical (Abbott) cardiac implantable electronic devices

Incidence of repetitive nonreentrant ventriculoatrial synchrony (RNRVAS)

verfasst von: S. S. Barold, MD

Erschienen in: Herzschrittmachertherapie + Elektrophysiologie | Ausgabe 3/2017

Einloggen, um Zugang zu erhalten

Abstract

This review focuses on the manifestations of the three triggered atrial upper rate functions of St Jude Medical cardiac implantable electronic devices. The occurrence of repetitive nonreentrant ventriculoatrial synchrony (RNRVAS) is also evaluated as a basis for the development of automatic mode switching (AMS) and as a trigger for atrial tachycardia/atrial fibrillation (AT/AF) event recordings. RNRVAS is a common trigger for AMS because all the atrial events or intervals are used to calculate the filtered atrial rate interval (FARI). Once AMS is initiated, it will also effectively stop RNRVAS because entry into AMS also shortens the postventricular atrial refractory period (PVARP). Recent design developments to eliminate or minimize unusual upper rare responses include the following: (1) P waves in the PVARP are no longer counted towards the FARI if they are followed by an atrial paced event. (2) In new devices the AT/AF detection algorithm substitutes the Moving Average Interval (a relatively complex calculation) with the new FARI average. (3) Improved design of the rate-responsive PVARP with a far more aggressive response than in the past (enhanced atrial protection interval).
Literatur
1.
Zurück zum Zitat St Jude Medical, Cardiac Rhythm Management Division, Sylmar (2010) Bradycardia and tachycardia devices, merlin patient care system, help manual, 91342 St Jude Medical, Cardiac Rhythm Management Division, Sylmar (2010) Bradycardia and tachycardia devices, merlin patient care system, help manual, 91342
2.
Zurück zum Zitat St Jude Medical, Sylmar (2015) Bradycardia and tachycardia devices. Merlin patient care system, 91342 St Jude Medical, Sylmar (2015) Bradycardia and tachycardia devices. Merlin patient care system, 91342
3.
Zurück zum Zitat Barold SS, Levine PA (2001) Pacemaker repetitive nonreentrant ventriculoatrial synchronous rhythm. A review. J Interv Card Electrophysiol 5:45–58CrossRefPubMed Barold SS, Levine PA (2001) Pacemaker repetitive nonreentrant ventriculoatrial synchronous rhythm. A review. J Interv Card Electrophysiol 5:45–58CrossRefPubMed
4.
Zurück zum Zitat Barold SS (1991) Repetitive reentrant and non-reentrant ventriculoatrial synchrony in dual chamber pacing. Clin Cardiol 14:754–763CrossRefPubMed Barold SS (1991) Repetitive reentrant and non-reentrant ventriculoatrial synchrony in dual chamber pacing. Clin Cardiol 14:754–763CrossRefPubMed
5.
Zurück zum Zitat Sharma PS, Kaszala K, Tan AY, Koneru JN, Shepard R, Ellenbogen KA, Huizar JF (2016) Repetitive nonreentrant ventriculoatrial synchrony: an underrecognized cause of pacemaker-related arrhythmia. Heart Rhythm 13:1739–1747. doi:10.1016/j.hrthm.2016.04.002 Sharma PS, Kaszala K, Tan AY, Koneru JN, Shepard R, Ellenbogen KA, Huizar JF (2016) Repetitive nonreentrant ventriculoatrial synchrony: an underrecognized cause of pacemaker-related arrhythmia. Heart Rhythm 13:1739–1747. doi:10.​1016/​j.​hrthm.​2016.​04.​002
6.
Zurück zum Zitat Barold SS, Stroobandt RX, Sinnaeve A (2010) Cardiac pacemakers and cardiac resynchronization. Step-by-step, 2nd edn. Wiley Blackwell, Chichester, p 135CrossRef Barold SS, Stroobandt RX, Sinnaeve A (2010) Cardiac pacemakers and cardiac resynchronization. Step-by-step, 2nd edn. Wiley Blackwell, Chichester, p 135CrossRef
7.
Zurück zum Zitat Barold SS, Stroobandt RX (2012) Erroneous automatic pacemaker arrhythmia diagnosis. Is it malfunction or a design limitation? Heart Rhythm 9:998–1001CrossRefPubMed Barold SS, Stroobandt RX (2012) Erroneous automatic pacemaker arrhythmia diagnosis. Is it malfunction or a design limitation? Heart Rhythm 9:998–1001CrossRefPubMed
8.
Zurück zum Zitat Davis D (2016) Personal communication. St Jude Medical, Sylmar Davis D (2016) Personal communication. St Jude Medical, Sylmar
9.
Zurück zum Zitat Kohno R, Abe H, Oginosawa Y, Tamura M, Takeuchi M, Nagatomo T et al (2011) Reliability and characteristics of atrial tachyarrhythmias detection in dual chamber pacemakers. Circ J 75:1090–1097CrossRefPubMed Kohno R, Abe H, Oginosawa Y, Tamura M, Takeuchi M, Nagatomo T et al (2011) Reliability and characteristics of atrial tachyarrhythmias detection in dual chamber pacemakers. Circ J 75:1090–1097CrossRefPubMed
10.
Zurück zum Zitat ASSERT Investigators, Healey JS, Connolly SJ, Gold MR, Israel CW, Van Gelder IC, Capucci A et al (2012) Subclinical atrial fibrillation and the risk of stroke. N Engl J Med 366:120–129CrossRef ASSERT Investigators, Healey JS, Connolly SJ, Gold MR, Israel CW, Van Gelder IC, Capucci A et al (2012) Subclinical atrial fibrillation and the risk of stroke. N Engl J Med 366:120–129CrossRef
11.
Zurück zum Zitat Hohnloser SH, Healey JS, Gold MR, Israel CW, Yang S, van Gelder I, Capucci A, Lau CP, Fain E, Morillo CA, Ha A, Carlson M, Connolly SJ, ASSERT Investigators (2012) Atrial overdrive pacing to prevent atrial fibrillation: insights from ASSERT. Heart Rhythm 9:1667–1673CrossRefPubMed Hohnloser SH, Healey JS, Gold MR, Israel CW, Yang S, van Gelder I, Capucci A, Lau CP, Fain E, Morillo CA, Ha A, Carlson M, Connolly SJ, ASSERT Investigators (2012) Atrial overdrive pacing to prevent atrial fibrillation: insights from ASSERT. Heart Rhythm 9:1667–1673CrossRefPubMed
12.
Zurück zum Zitat ASSERT Steering Committee and Investigators, Kaufman ES, Israel CW, Nair GM, Armaganijan L, Divakaramenon S, Mairesse GH et al (2012) Positive predictive value of device-detected atrial high-rate episodes at different rates and durations: an analysis from ASSERT. Heart Rhythm 9:1241–1246CrossRef ASSERT Steering Committee and Investigators, Kaufman ES, Israel CW, Nair GM, Armaganijan L, Divakaramenon S, Mairesse GH et al (2012) Positive predictive value of device-detected atrial high-rate episodes at different rates and durations: an analysis from ASSERT. Heart Rhythm 9:1241–1246CrossRef
13.
Zurück zum Zitat Richter S, Muessigbrodt A, Salmas J, Doering M, Wetzel U, Arya A, Hindricks G, Brugada P, Israel CW (2013) Ventriculoatrial conduction and related pacemaker-mediated arrhythmias in patients implanted for atrioventricular block: an old problem revisited. Int J Cardiol 168:3300–3308CrossRefPubMed Richter S, Muessigbrodt A, Salmas J, Doering M, Wetzel U, Arya A, Hindricks G, Brugada P, Israel CW (2013) Ventriculoatrial conduction and related pacemaker-mediated arrhythmias in patients implanted for atrioventricular block: an old problem revisited. Int J Cardiol 168:3300–3308CrossRefPubMed
14.
Zurück zum Zitat Barold SS, Stroobandt RX, Van Heuverswyn F (2012) Pacemaker repetitive nonreentrant ventriculoatrial synchrony. Why did automatic mode switching occur? J Electrocardiol 45:420–425CrossRefPubMed Barold SS, Stroobandt RX, Van Heuverswyn F (2012) Pacemaker repetitive nonreentrant ventriculoatrial synchrony. Why did automatic mode switching occur? J Electrocardiol 45:420–425CrossRefPubMed
15.
Zurück zum Zitat Tzeis S, Pastromas S, Andrikopoulos G (2014) Repetitive non-reentrant ventriculoatrial synchrony: a rare cause of overestimating atrial fibrillation burden. Europace 16:1091CrossRefPubMed Tzeis S, Pastromas S, Andrikopoulos G (2014) Repetitive non-reentrant ventriculoatrial synchrony: a rare cause of overestimating atrial fibrillation burden. Europace 16:1091CrossRefPubMed
16.
Zurück zum Zitat Pakarinen S, Toivonen L (2012) Performance of atrial tachyarrhythmia-sensing algorithms in dual-chamber pacing using a fixed long AV delay in patients with sinus node dysfunction. J Interv Card Electrophysiol 35:207–213CrossRefPubMed Pakarinen S, Toivonen L (2012) Performance of atrial tachyarrhythmia-sensing algorithms in dual-chamber pacing using a fixed long AV delay in patients with sinus node dysfunction. J Interv Card Electrophysiol 35:207–213CrossRefPubMed
17.
Zurück zum Zitat St Jude Medical, Sylmar (2015) Filtered Atrial Rate Interval (FARI). Technical insight, November 2015 St Jude Medical, Sylmar (2015) Filtered Atrial Rate Interval (FARI). Technical insight, November 2015
18.
Zurück zum Zitat St Jude Medical, Sylmar (2014) Rate-Responsive PVARP = High. The new API!. Endurity, Assurity, and Allure. Technical insight, September 2014 St Jude Medical, Sylmar (2014) Rate-Responsive PVARP = High. The new API!. Endurity, Assurity, and Allure. Technical insight, September 2014
Metadaten
Titel
A review of the atrial upper rate algorithms of St. Jude Medical (Abbott) cardiac implantable electronic devices
Incidence of repetitive nonreentrant ventriculoatrial synchrony (RNRVAS)
verfasst von
S. S. Barold, MD
Publikationsdatum
28.06.2017
Verlag
Springer Medizin
Erschienen in
Herzschrittmachertherapie + Elektrophysiologie / Ausgabe 3/2017
Print ISSN: 0938-7412
Elektronische ISSN: 1435-1544
DOI
https://doi.org/10.1007/s00399-017-0517-z

Weitere Artikel der Ausgabe 3/2017

Herzschrittmachertherapie + Elektrophysiologie 3/2017 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.