03.07.2019 | KNEE

A synthetic bone insert may protect the lateral cortex and fixation plate following a high tibial osteotomy by reducing the tensile strains

Zeitschrift:
Knee Surgery, Sports Traumatology, Arthroscopy
Autoren:
Rachel MacDonald, Kyla Huebner, Jack Farr, Cynthia E. Dunning, Alan Getgood, Timothy A. Burkhart
Wichtige Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abstract

Purpose

To determine the effectiveness of a synthetic bone insert on improving medial opening wedge high tibial osteotomy integrity in response to post-surgical cyclical loading.

Materials and methods

A medial opening wedge high tibial osteotomy, secured with a compression fixation plate, was performed on 12 cadaveric knee specimens that were randomised to either: (1) a synthetic insert condition (n = 6), in which a 9 mm bio-absorbable wedge was inserted into the gap space; or (2) a plate-only condition (n = 6). Uniaxial strain gauges, placed on the lateral cortex and fixation plate, measured the strain response as the specimens were subjected to a staircase cyclical loading protocol; a sinusoidal waveform between 100 and 800 N was applied and increased by increments of 200 N every 5000 cycles until failure. Peak strains at failure were compared between conditions using a one-tailed independent samples t test.

Results

The strains from the fixation plate were significantly different between the insert and plate only conditions (p = 0.02), transitioning from a compressive strain with the wedge (mean [SD] = − 8.6 [− 3.6] µε) to a tensile strain without the wedge (mean [SD] = 12.9 [23] µε). The strains measured at the lateral cortex were also significantly affected by the inclusion of a synthetic bone insert (p = 0.016), increasing from − 55.6 (− 54.3) µε when the insert was utilised to 23.7 (55.7) µε when only the plate was used.

Conclusions

The addition of a synthetic insert limited the tensile strains at the plate and lateral cortex, suggesting that this may protect these regions from fracture during prolonged loading.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel
  1. Sie können e.Med Orthopädie & Unfallchirurgie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Orthopädie und Unfallchirurgie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Orthopädie und Unfallchirurgie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise