Open Access 01.12.2023 | Review

A systematic review of the accuracy of digital surgical guides for dental implantation

verfasst von: Yiting Shi, JunKai Wang, Chao Ma, Jiayi Shen, Xian Dong, Dan Lin

Erschienen in: International Journal of Implant Dentistry | Ausgabe 1/2023

Abstract

Purpose

This review aimed to reveal the influence of implant guides on surgical accuracy with regard to supporting types, manufacturing methods and design (including fixation screws and sleeves).

Methods

A literature search related to accuracy of surgical guides for dental implantation was performed in Web of Science and PubMed. Studies with in vivo or in vitro deviation data published in recent 5 years (2018–2022) were included and assessed by Newcastle–Ottawa Scale with regard to risk of bias and reliability degree of clinical studies. Accuracy-related deviation data were summarized as forest plots and normal distributions.

Results

Forty-one articles were included with high degree of credibility. Data showed that implant surgery accuracy can be achieved with mean distance deviation < 2 mm (most < 1 mm) and angular deviation < 8° (most < 5°).

Conclusions

Bilateral tooth-supported guides exhibited highest in vitro accuracy and similar in vivo accuracy to unilateral tooth-supported guides; mucosa-supported guides exhibit lowest in vivo accuracy, while its in vitro data showed low credibility due to mechanical complexity of living mucosa tissue. Milling exhibited higher in vivo accuracy of guides than 3d-printing, though further data support was needed. Design of fixation screws and sleeves of implant guides affected the surgical accuracy and might remain a research focus in near future. However, lack of universal evaluation standards for implantation accuracy remained a major problem in this field. The influence of implant guides on surgical accuracy revealed in this review might shed light on future development of dental implantology.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s40729-023-00507-w.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AR
Augmented reality
CBCT
Cone beam CT
ISO
International Organization for Standardization
IOSs
Intraoral scanners
PICO
Participants, Intervention, Comparison or control and Outcome measures
3D
Three-dimensional
ID
Identity document
NOS
Newcastle–Ottawa Scale
WOS
Web of Science
SD
Standard deviation
n
Number of records
CAD
Computer Aided Design
CAM
Computer Aided Manufacturing
SLA
Stereo lithography appearance
FFF
Fused filament fabrication
DLP
Digital light processing
Ref.
Reference
No.
Numero sign/numero symbol
SUMHS
Shanghai University of Medicine & Health Sciences

Introduction

Global population aging results in increasing demand for dental implant surgery, emphasizing the necessity of improving the surgical accuracy, which directly increases the success rate and reduces surgical trauma [1]. In recent years, digital technology that realizes the visualization of planting schemes significantly raises the surgical accuracy of dental implantation [2]. Digital surgical guides, as the information carrier of implant direction, position, angle, can effectively enhance the surgical accuracy, reduce surgical time and complications [3]. The surgical accuracy of dental implantation is influenced by data acquisition method, manufacturing procedure, supporting types, fixation screws and sleeve design of the surgical guide.
Accuracy of implanting guide comprises trueness and precision (ISO 5725–1:1994). Trueness refers to the deviation between postoperative placement and preoperative plan of the implant; precision refers to the deviation of repetitive test results. Generally, accuracy discussed in clinical studies refers to trueness, while in vitro studies (e.g., implant on plaster models) may involve both trueness and precision. The accuracy compared and discussed in this review mainly refers to trueness. Despite the lack of a universal evaluation standard for implantation surgical accuracy, common indicators including coronal deviation (mm), apical deviation (mm), depth deviation/vertical deviation (mm), angular deviation (°) are applied in existing literatures and discussed in this review (Fig. 1).
Preoperative data acquisition, including intraoral (collected via intraoral scanning or extraoral scanning of impressions) and CBCT data, is the prerequisite for implant guide design [4]. Existing studies indicated a higher accuracy of intraoral scanners (IOSs) than extraoral scanning of impressions [5], and IOSs are the developing trend in the future and are highly accepted among the patients due to its flexibility and simplicity [6]. Accuracy of commercial IOSs ranges from 20 to100 μm for dentition, and 50–250 μm for edentulous jaws [7, 8]. By contrast, CBCT exhibits lower accuracy: within an applicable range of radiation dose (20–100 μSv, generally considered as a balance of safety and accuracy), the voxel size of CBCT-constructed 3D models ranges from 0.1 to 0.5 mm3, and the accuracy ranges from 200 to 1000 μm [9]. Considering the accuracy range of CBCT, it is generally recommended to maintain a safety margin of 2 mm from adjacent anatomical structures in clinical practice [10]. As the accuracy of dentition model obtained by CBCT is relatively low to meet the requirements of surgical guide design, integration of CBCT-constructed model and scanned dentition model is commonly applied [4]. As data acquisition accuracy will be projected onto design of surgical guides, and current evaluation of implant surgery accuracy is mainly based on CBCT data acquisition, comparability of the accuracy indicator values in different literatures should be reviewed dialectically.
Two common narrative aspects of existing reviews on the accuracy of digital surgical guides are comparing static surgical guides to dynamic or augmented reality (AR) navigation systems [11] and analyzing the accuracy of surgical guides for specific conditions. Examples of the later include comparison of the applicability, accuracy and clinical effects of digital and traditional surgical guides in flapless implant surgery reviewed by Emitis Natali Naeini in 2020 [12] demonstrating higher accuracy of digital surgical guides than traditional ones; comparison of the accuracy of digital surgical guides between flap and flapless implant surgery by Karthikeyan Subramani in 2022 [13] indicating that flapless surgery resulted in higher accuracy; analysis of the accuracy of edentulous mucosa-supported surgical guides prepared by Stereolithography by Cheongbeom Seo in 2018 [14] and surgical guides for partially edentulous patients analyzed by Ramadhan Hardani Putra in 2022 [3], both exhibiting no significant differences in any specific influencing factors. In addition, R. Eftekhar Ashtiani in 2021 [15] compares the accuracy among different guiding systems and concludes that both the implant and its design software influenced the accuracy, though no final statement could be made on an optimized system. Fernando Bover-Ramos in 2018 [16] compares the accuracy of surgical guides studied in cadaveric, clinical, and in vitro models and finds that in vitro studies showed higher accuracy than clinical and cadaveric studies, and the accuracy of full-guided surgical guides are higher than half-guided ones. A review published by Firas Al Yafi in 2019 [1] summarizes the operational procedures of digital surgical guides in detail, and though lack of in-depth data analyses, provides a comprehensive list of accuracy-affecting factors. A review by Ali Tahmaseb in 2018 [17], based on published research data from 2012–2015, concludes that the accuracy of surgical guide was within the clinically acceptable range in most cases and was higher in partially edentulous patients than edentulous patients.
Considering the lack of a systematic and comprehensive review of the factors influencing the accuracy of surgical guides, this review analyzed and discussed the accuracies of all static digital surgical guides in aspects including guide supporting types, manufacturing methods and design of implant guides (including fixation screws and sleeves), covering different influencing factors to provide a comprehensive guidance for implant design in future. This review aimed to reveal the influence of implant guides on surgical accuracy, and to provide reference for future development of digital dental implantology. It was hypothesized that these aspects influence the accuracies of static digital surgical guides, and the hypothesis was verified by collecting and categorizing the numerical data of surgical accuracy indicator reported in literature of the last five years.

Materials and methods

This review has been registered in PROSPERO, with the registration ID of 416029.
The search was performed using keywords based on the PICO approach. The PICO was formulated as follows: Participants (P) = patients of dental implantation; Intervention (I) = implants placed using digital surgical guides; Comparison or control (C) = different supporting types, design of fixation screws, design of sleeve, manufacturing methods of the surgical guides; Outcome measures (O) = coronal deviation (mm), apical deviation (mm), depth deviation (mm), angular deviation (°).
Based on the above PICO analysis, this review applied the keywords: computer-aided implant surgery (CAIS); static surgical guide; accuracy; deviation; dental implants) and MeSH terms (Surgery, Computer-Assisted) AND (Dental Implants). Advanced searching strategies were established based on the above keywords to perform an extensive search of the literature for papers related to accuracy of digital surgical guides for dental implantation on the databases of Web of Science (WOS) and PubMed as follows:
WOS: TS = (oral OR dental) AND TS = (surgical guide) AND TS = (accuracy OR precision OR rightness) AND TS = (implant OR implantation).
PubMed: (oral OR dental) AND (surgical guide) AND (accuracy OR precision OR rightness) AND (implant OR implantation).
Literatures were screened using predetermined inclusion and exclusion criteria as follows.
Inclusion criteria:
1.
Types of literature were limited to research articles, case reports and clinical trials that were peer-reviewed and published in WOS or PubMed cited scientific journals.
 
2.
Titles and abstracts of the articles were related to the accuracy of digital surgical guides for dental implantation.
 
3.
At least one of the following in vivo or in vitro deviation data must be involved: coronal deviation (mm), apical deviation (mm), depth deviation (mm), angular deviation (°).
 
4.
Written in English.
 
5.
The year of publication was restricted in recent five years (2018–2022).
 
Exclusion criteria:
1.
Reviews, meeting abstracts, grey literature or non-peer-reviewed literature were excluded.
 
2.
Written in languages other than English.
 
3.
Published before 2018.
 
To minimize the potential for reviewer bias, two reviewers (CM and JS) independently conducted literature searches and performed the study selection. Both reviewers strictly followed the inclusion and exclusion criteria, and any disagreement was resolved by discussion.
Data were extracted by one reviewer (JS) and examined by another reviewer (CM). The following data were directly collected from the included articles: literature information (authors, year, and title), research type (clinical/cadaver/in vitro), number of patients/cadavers/models, number of implants, surgical information (full-/half-guided, planning software, implant site, jaw position, bone quality and implant length and diameter), types of surgical guide (bilateral tooth-supported, unilateral tooth-supported, and mucosa-supported guides), guide fabrication method (3D printing/milling), number of fixation pins, deviation data including global/horizontal coronal deviation (mm), global/horizontal apical deviation (mm), angular deviation (°) and vertical deviation (mm). The form of deviation data included mean ± SD and/or median (min, max).
To assess the risk of bias and degree of reliability, clinical studies were scored based on the Newcastle–Ottawa Scale (NOS) adapted by Chambrone et al. [18] including evaluation of four subcategories: sample selection of study groups, comparability, outcome and statistical analysis. Specific scoring items are listed in Additional file 1: Methods. A maximum of 13 points could be obtained for each study, with a score of 10–13 indicating high study quality, a score of 7–9 indicating moderate study quality, and a score of less than 7 indicating low study quality.

Results

Following the PRISMA guideline (Fig. 2), the search strategy reported 954 records, among which 249 duplicate records were firstly removed. After overviewing the titles, abstracts and keywords, the investigators excluded 54 reviews, 30 articles written in languages other than English, and 580 records with no considerable information about accuracy of surgical guides for dental implantation. The remaining 41 records were sought for full-text retrieval and assessment of data availability, and all 41 articles involved available deviation data. The 41 articles were ultimately included in this review, among which 21 were in vitro studies, 19 were in vivo studies, and 1 was a comparison of in vitro and in vivo accuracy. Among the 19 in vivo studies, 2 were performed on cadaver, and 17 were clinical researches. Among 17 clinical research types, 3 were case–control studies, 11 were clinical trials, and 3 were cohort studies. The comparison of in vitro and in vivo accuracy was a case–control study (Table 1).
Table 1
Classification of research types
Type of research
Number
Type of clinical study
Number
In vitro
21
In vivo
   
 Cadaver
2
Clinical
17
Case–control study
3
Clinical trial
11
Cohort study
3
Comparison of in vitro and in vivo
1
Case–control study
1
NOS analysis was performed in the 20 in vivo studies in this paper, of which 3 were of medium quality and 17 were of high quality (Additional file 1: Fig. S1). Specifically, all studies exhibited high scores in sample selection, comparability and statistics, though the outcomes of patient follow-up adequacy were presented in only 3 studies. The NOS results indicated a high degree of credibility of this review.
A descriptive table of the 41 studies included in this review with deviation data including global/horizontal coronal deviation (mm), global/horizontal apical deviation (mm), angular deviation (°) and vertical deviation (mm) in the form of mean ± SD and/or median (min, max) was provided as Table 2, and the data listed and discussed in following part of this review are based on the corresponding information. The comparison criteria (supporting types, design of fixation screws, design of sleeve, manufacturing methods of the surgical guides) and accuracy-related deviation data of the included literatures were summarized as forest plots (Tables 3, 4, 5, Additional file 1: Tables S1–S4) and normal distributions (Figs. 4, 5).
Table 2
Descriptive table of the 41 studies included in this review with deviation data including global/horizontal coronal deviation (mm), global/horizontal apical deviation (mm), angular deviation (°) and vertical deviation (mm) in the form of mean ± SD and/or median (min, max)
Ref. No.
Author (year)
Research type
Clinical research type
Full- or half-guided
No. of patients/models
No. of implants
Implant site
Supported type
Fabrication
1
Xiaoqian Liu (2022) [20]
In vitro
/
/
/
54
/
/
Milling
2
Roberto Pessoa (2022) [21]
In vitro
/
Full
16
48
26, 27, 28
Unilateral tooth-supported
3D printer
In vitro
/
Full
16
48
26, 27, 28
Unilateral tooth-supported
3D printer
3
Rani D'haese (2022) [22]
In vitro
/
Full
15
15
26
Bilateral tooth-supported
3D printer
4
Yao Sun (2022) [23]
In vitro
/
/
10
20
36, 46
Bilateral tooth-supported
3D printer
In vitro
/
/
10
20
36, 46
Bilateral tooth-supported
3D printer
5
Jeanette K Li-Rodríguez (2022) [24]
In vitro
/
Full
8
24
15, 36, 46
Bilateral tooth-supported
Milling
In vitro
/
Full
8
8
21
Bilateral tooth-supported
Milling
6
Nicole Báez-Marrero (2022) [25]
In vitro
/
/
8
15
2, 7, 8
Unilateral tooth-supported
3D printer
7
Kristof Orban (2022) [26]
In vivo
Cohort study
Half
20
20
4, 5 or 6, 7
Unilateral tooth-supported
3D printer
8
Jordi Gargallo-Albiol (2022) [27]
In vivo
Cohort study
Full
30
60
4, 5 or 6, 7
Unilateral tooth-supported
3D printer
9
Wanwanat Singthong (2022) [28]
In vivo
Clinical trial
Full
12
12
4, 5 or 6, 7
Bilateral tooth-supported
3D printer
In vivo
Clinical trial
Full
12
12
4, 5 or 6, 7
Bilateral tooth-supported
3D printer
10
Yuzhang Feng (2022) [29]
In vivo
Clinical trial
Full
20
20
1, 2
Bilateral tooth-supported
3D printer
11
Fangzhi Lou (2021) [30]
In vivo
Clinical trial
Half
20
36
11, 12, 13, 21, 22, 23
Bilateral tooth-supported
3D printer
In vivo
Clinical trial
Full
20
33
11, 12, 13, 21, 22
Bilateral tooth-supported
3D printer
12
David Schneider (2021) [31]
In vitro
/
Half
24
72
34, 35, 36
Bilateral tooth-supported
/
  
In vitro
/
Full
24
72
34, 35, 36
Bilateral tooth-supported
3D printer
  
In vitro
/
Half
24
24
46
Bilateral tooth-supported
/
  
In vitro
/
Full
24
24
46
Bilateral tooth-supported
3D printer
13
Young Woo Song (2021) [32]
In vitro
/
Full
20
20
36
Bilateral tooth-supported
Milling
  
In vitro
/
Full
20
20
36
Bilateral tooth-supported
3D printer
14
Jaafar Abduo (2021) [33]
In vitro
/
Full
14
14
13
Bilateral tooth-supported
Milling
  
In vitro
/
Full
14
14
16
Bilateral tooth-supported
Milling
  
In vitro
/
Pilot
14
14
13
Bilateral tooth-supported
3D printer
  
In vitro
/
Pilot
14
14
16
Bilateral tooth-supported
3D printer
15
Laura Herschdorfer (2021) [34]
In vitro
/
Full
10
10
46
Bilateral tooth-supported
3D printer
  
In vitro
/
Full
10
10
46
Bilateral tooth-supported
3D printer
  
In vitro
/
Full
10
10
46
Bilateral tooth-supported
3D printer
16
Chalermchai Ngamprasertkit (2021) [35]
In vivo
Clinical trial
Half
15
15
1, 2, 3, 4, 5
Bilateral tooth-supported
3D printer
  
In vivo
Clinical trial
Full
15
15
1, 2, 3, 4, 5
Bilateral tooth-supported
3D printer
17
Johannes Spille (2021) [36]
In vitro
/
Half
6
48
1, 2, 4, 5, 6, 7
Mucosa-supported
/
18
Arndt Guentsch (2021)[37]
In vitro
/
Half (6 mm-sleeve)
20
20
46
Bilateral tooth-supported
3D printer
  
In vitro
/
Full (2 mm-sleeve)
20
20
46
Bilateral tooth-supported
3D printer
  
In vitro
/
Full (4 mm-sleeve)
20
20
46
Bilateral tooth-supported
3D printer
  
In vitro
/
Full (6 mm-sleeve)
20
20
46
Bilateral tooth-supported
3D printer
19
Yen-Ting Han (2021)[38]
In vivo
Clinical trial
Half
30 (Total)
18
/
/
3D printer
  
In vivo
Clinical trial
Full
30 (Total)
56
/
/
3D printer
  
In vivo
Clinical trial
Full
30 (Total)
23
/
/
3D printer
  
In vivo
Clinical trial
Full
30 (Total)
33
/
/
3D printer
  
In vivo
Clinical trial
Full
30 (Total)
28
/
Mucosa-supported
3D printer
  
In vivo
Clinical trial
Full
30 (Total)
20
/
Bilateral tooth-supported
3D printer
20
Paknisa Sittikornpaiboon (2021) [39]
In vitro
/
/(5 mm-sleeve)
5
10
14, 24
Bilateral tooth-supported
3D printer
  
In vitro
/
/(5 mm-sleeve)
5
10
14, 24
Bilateral tooth-supported
3D printer
  
In vitro
/
/(4 mm-sleeve)
5
10
14, 24
Bilateral tooth-supported
3D printer
  
In vitro
/
/(4 mm-sleeve)
5
10
14, 24
Bilateral tooth-supported
3D printer
  
In vitro
/
/(4 mm-sleeve)
5
10
14, 24
Bilateral tooth-supported
3D printer
21
Lirong Huang (2021)[40]
In vivo
Case–control study
Full
20
21
1, 2, 3
Bilateral tooth-supported
/
  
In vivo
Case–control study
Full
20
31
1, 2, 3
Bilateral tooth-supported
/
22
Rani D’haese (2021)[41]
In vitro
/
Full
15
90
46, 44, 42, 32, 34, 36
Mucosa-supported
3D printer
  
In vitro
/
Full
15
90
46, 44, 42, 32, 34, 36
Mucosa-supported
3D printer
23
Chia-Cheng Lin (2020) [42]
In vitro
/
Full
30
30
36
Unilateral tooth-supported
3D printer
  
In vitro
/
Full
30
30
46
Unilateral tooth-supported
3D printer
  
In vitro
/
Full
30
30
47
Unilateral tooth-supported
3D printer
24
Yuan Chen (2020) [43]
In vivo
Case–control study
Full
30
37
1, 2, 3
Bilateral tooth-supported
3D printer
  
In vivo
Case–control study
Half
33
39
1, 2, 3
Bilateral tooth-supported
3D printer
25
Dong Wu (2020) [44]
In vivo
Case–control study
/
29
57
1, 2, 4, 5, 6, 7
Bilateral tooth-supported
3D printer
26
Kang-jie Cheng (2020) [45]
In vitro
/
Full
1
5
35, 36, 45, 46, 47
Unilateral tooth-supported
3D printer
27
Pantip Henprasert (2020) [46]
In vitro
/
Full
15
15
46
Bilateral tooth-supported
milling
  
In vitro
/
Full
15
15
46
Bilateral tooth-supported
3D printer
28
Kristian Kniha (2020) [47]
In vivo
Clinical trial
Full
5
10
1, 2, 3
Bilateral tooth-supported
3D printer
  
In vivo
Clinical trial
Full
5
10
4, 5, 6, 7
Bilateral tooth-supported
3D printer
  
In vivo
Clinical trial
Full
5
10
1, 2, 3
Mucosa-supported
3D printer
  
In vivo
Clinical trial
Full
5
10
4, 5, 6, 7
Mucosa-supported
3D printer
29
R. Vinci (2020) [48]
In vivo
Cohort study
Full
12
51
1, 2, 3, 4, 5, 6, 7
Mucosa-supported
Milling
  
In vivo
Cohort study
Full
11
49
1, 2, 3, 4, 5, 6, 7
Mucosa-supported
Milling
  
In vivo
Cohort study
Full
14
46
1, 2, 3
Mucosa-supported
Milling
  
In vivo
Cohort study
Full
14
54
4, 5, 6, 7
Mucosa-supported
Milling
30
Nopparat Suksod (2020) [49]
In vivo
Clinical trial
Half
20
20
1, 2
Bilateral tooth-supported
3D printer
31
Márton Kivovics (2020) [50]
In vivo
Clinical trial
Half
6
18
/
Mucosa-supported
3D printer
  
In vivo
Clinical trial
Half
7
22
/
Mucosa-supported
3D printer
32
Palita Smitkarn (2019) [51]
In vivo
Clinical trial
Full
52
60
1, 2, 4, 5, 6, 7
Unilateral tooth-supported
3D printer
33
Karim El Kholy (2019) [52]
In vitro
/
Full
40
240
15, 12, 21, 23, 25, 26
Unilateral tooth-supported
3D printer
  
In vitro
/
Full
15
45
15, 12, 23
Bilateral tooth-supported
3D printer
  
In vitro
/
Full
15
45
15, 12, 23
Bilateral tooth-supported
3D printer
  
In vitro
/
Full
15
45
15, 12, 23
Bilateral tooth-supported
3D printer
34
Henrik Skjerven (2019) [53]
In vivo
Clinical trial
Full
20
27
1, 2, 3, 4, 5, 6, 7
Bilateral tooth-supported
3D printer
35
Rai-Jei Chang (2018) [54]
In vitro
/
Half
17
20
36, 37
Unilateral tooth-supported
/
  
In vivo
Case–control study
Half
17
20
36, 37
Unilateral tooth-supported
/
36
Zhaozhao Chen (2018) [55]
In vivo
Clinical trial
/
4
12
11, 12, 21, 22
Bilateral tooth-supported
/
37
Jan Brandt (2018) [56]
In vitro
/
Full
/
30
46
Bilateral tooth-supported
/
38
Philipp Kauffmann (2018) [57]
In vitro
/
/
/
20
44, 45, 46, 47
Mucosa-supported
/
  
In vitro
/
/
/
20
44, 45, 46, 47
Mucosa-supported
/
39
Yiqin Fang (2018) [58]
In vivo
Clinical trial
/
32
40
11, 22
Bilateral tooth-supported
3D printer
40
Boyoung Ma (2018) [59]
In vivo
Clinical trial
/
17
28
36, 15
Bilateral tooth-supported
3D printer
  
In vivo
Clinical trial
/
17
28
36, 15
Bilateral tooth-supported
3D printer
41
Björn Gjelvold (2018) [60]
In vitro
/
Full
10
10
44
Bilateral tooth-supported
3D printer
  
In vitro
/
Full
10
10
44
Bilateral tooth-supported
3D printer
Ref. No.
Author (year)
Data extraction (mean ± SD, median, (min, max))
No. of fixation screws
Global coronal deviation (mm)
Horizontal coronal deviation (mm)
Global apical deviation (mm)
Horizontal apical deviation (mm)
Angular deviation (°)
Vertical deviation (mm)
1
Xiaoqian Liu (2022) [20]
0.16 ± 0.06, /, (0.03, 0.29)
/
/
/
0.61 ± 0.40, /,
(0.01, 1.86)
0.11 ± 0.07, /, (0.00, 0.25)
/
2
Roberto Pessoa (2022) [21]
0.88 ± 0.36, /, /
0.67 ± 0.22, /, /
1.60 ± 0.69, /, /
1.72 ± 0.70, /, /
4.53 ± 2.04, /, /
–0.16 ± 0.62, /, /
4
0.88 ± 0.36, /, /
0.55 ± 0.32, /, /
1.44 ± 0.75, /, /
1.63 ± 0.69, /, /
4.28 ± 2.01, /, /
–0.5 ± 0.5, /, /
0
3
Rani D'haese (2022) [22]
0.52 ± 0.25, /, (0.09, 1.07)
/
0.90 ± 0.47, /, (0.14, 1.74)
/
2.63 ± 1.69, /, (0.38, 5.99)
0.32 ± 0.27, /, (0.02, 1.00)
/
4
Yao Sun (2022) [23]
0.35 ± 0.11, /, (0.20, 0.64)
0.21 ± 0.13, /, (0.03, 0.44)
0.75 ± 0.28, /, (0.21, 1.17)
0.48 ± 0.30, /, (0.01, 0.91)
2.74 ± 1.24, /, (0.50, 4.80)
0.11 ± 0.09, /, (0.00, 0.36)
/
0.41 ± 0.13, /, (0.16, 0.66)
0.28 ± 0.14, /, (0.04, 0.52)
0.91 ± 0.34, /, (0.34, 1.38)
0.60 ± 0.33, /, (0.10, 1.13)
3.22 ± 1.55, /, (1.00, 6.90)
0.11 ± 0.08, /, (0.00. 0.29)
/
5
Jeanette K Li-Rodríguez (2022) [24]
/
0.2 ± 0.126, /, /
/
/
1.1 ± 0.834, /, /
/
/
/
0.2 ± 0.126, /, /
/
/
1.1 ± 0.834, /, /
/
/
6
Nicole Báez-Marrero (2022) [25]
1.43 ± 0.60, /, /
/
2.19 ± 0.63, /, /
/
6.81 ± 3.10, /, /
/
2
7
Kristof Orban (2022) [26]
1.20 ± 0.46, /, /
1.06 ± 0.52, /, /
1.45 ± 0.79, /, /
1.28 ± 0.83, /, /
4.82 ± 2.07, /, /
0.55 ± 0.28, /, /
/
8
Jordi Gargallo-Albiol (2022) [27]
0.21 ± 0.69, /, /
/
0.67 ± 1.06, /, /
/
5.62 ± 4.09, /, /
/
/
9
Wanwanat Singthong (2022) [28]
/
1.07 ± 0.36, /, (0.47, 1.60)
/
/
3.52 ± 1.64, /, (0.60, 6.10)
− 0.71 ± 0.29, /, (− 0.04, − 1.15)
/
/
0.97 ± 0.33, /, (0.33. 1.38)
/
/
3.77 ± 2.16, /, (1.05, 7.20)
− 0.84 ± 0.30, /, (− 0.26, − 1.28)
/
10
Yuzhang Feng (2022) [29]
0.99 ± 0.63, /, /
/
1.50 ± 0.75, /, /
/
3.07 ± 2.18, /, /
/
1
11
Fangzhi Lou (2021) [30]
0.69 ± 0.10, 0.675, (0.51, 0.87)
/
0.80 ± 0.08, 0.80, (0.68, 0.95)
/
3.16 ± 0.70, 3.15, (1.43, 4.73)
0.52 ± 0.11, 0.505, (0.33, 0.73)
/
0.39 ± 0.12, 0.39, (0.15, 0.61)
/
0.28 ± 0.09, 0.29, (0.11, 0.44)
/
2.05 ± 0.45, 2.09, (0.89, 3.00)
0.24 ± 0.06, 0.25, (0.12, 0.34)
/
12
David Schneider (2021) [31]
/
0.70 ± 0.48, 0.56, (0.14, 1.80)
/
0.77 ± 0.53, 0.64, (0.16, 2.04)
1.70 ± 0.67, 1.65, (0.80, 3.20)
0.46 ± 0.33, 0.36, (0.00, 1.26)
 
  
/
0.18 ± 0.11, 0.15, (0.02, 0.49)
/
0.31 ± 0.17, 0.27, (0.09, 0.83)
1.57 ± 0.84, 1.40, (0.20, 3.30)
0.19 ± 0.13, 0.17, (0.01, 0.44)
 
  
/
0.49 ± 0.33, 0.43, (0.06, 1.22)
/
0.51 ± 0.33, 0.53, (0.06, 1.24)
1.36 ± 0.78, 1.20, (0.00, 2.80)
0.45 ± 0.46, 0.24, (0.04, 1.70)
 
  
/
0.24 ± 0.13, 0.21, (0.03, 0.52)
/
0.34 ± 0.20, 0.30, (0.07, 0.93)
1.32 ± 0.88, 1.05, (0.30, 3.40)
0.28 ± 0.19, 0.29, (0.02, 0.78)
 
13
Young Woo Song (2021) [32]
/
1.37, 1.01, /, /
/
1.68, 1.41, /, /
3.49, 3.62, /, /
0.95, 0.71, /, /
 
  
/
0.95, 0.78, /, /
/
1.34, 1.25, /, /
3.04, 2.69, /, /
0.64, 0.44, /, /
 
14
Jaafar Abduo (2021) [33]
/
0.46 ± 0.23, /, (0.11, 0.89)
/
0.62 ± 0.42, /, (0.12, 1.46)
1.25 ± 0.84, /, (0.30, 2.66)
0.31 ± 0.26, /, (0.03, 0.84)
 
  
/
0.39 ± 0.24, /, (0.01, 0.75)
/
0.71 ± 0.41, /, (0.21, 1.53)
1.59 ± 1.13, /, (0.17, 4.20)
0.37 ± 0.30, /, (0.03, 1.08)
 
  
/
0.53 ± 0.26, /, (0.21, 0.98)
/
1.49 ± 0.54, /, (0.16, 2.27)
6.76 ± 2.49, /, (1.60, 9.65)
0.61 ± 0.35, /, (0.11, 1.18)
 
  
/
0.34 ± 0.24, /, (0.02, 0.83)
/
0.76 ± 0.52, /, (0.22, 2.16)
4.00 ± 2.62, /, (0.76, 9.07)
0.51 ± 0.47, /, (0.05, 1.77)
 
15
Laura Herschdorfer (2021) [34]
0.24 ± 0.19, 0.19, (0.07, 0.71)
/
0.40 ± 0.23, 0.36, (0.08, 0.92)
/
1.44 ± 0.61, 1.30, (0.70, 2.60)
/
 
  
0.23 ± 0.13, 0.20, (0.08, 0.50)
/
0.37 ± 0.22, 0.34, (0.37, 0.80)
/
1.37 ± 0.71, 1.15, (0.60, 2.40)
/
 
  
0.22 ± 0.06, 0.23, (0.13, 0.31)
/
0.30 ± 0.08, 0.32, (0.16, 0.39)
/
0.94 ± 0.48, 0.56, (0.00, 1.50)
/
 
16
Chalermchai Ngamprasertkit (2021) [35]
0.74 ± 0.36, /, (0.09, 1.38)
0.57 ± 0.39, /, (0.08, 1.38)
1.29 ± 0.61, /, (0.36, 2.32)
1.17 ± 0.68, /, (0.04, 2.31)
3.44 ± 1.61, /, (0.95, 6.68)
(coronal)0.36 ± 0.27, /, (0.01, 0.95) (apical)0.37 ± 0.27, /, (0.03, 1.01)
 
  
0.48 ± 0.22, /, (0.20, 0.87)
0.39 ± 0.26, /, (0.08, 0.87)
0.71 ± 0.31, /, (0.18, 1.34)
0.64 ± 0.37, /, (0.03, 1.33)
2.03 ± 1.00, /, (0.88, 4.03)
(coronal)0.19 ± 0.14, /, (0.01, 0.51) (apical)0.20 ± 0.13, /, (0.03, 0.51)
 
17
Johannes Spille (2021) [36]
/
1.009 ± 0.415, /, /
/
1.068 ± 0.384, /, /
2.67 ± 1.58, /, /
/
 
18
Arndt Guentsch (2021)[37]
0.20 ± 0.14, /, /
/
0.19 ± 0.13, /, /
/
2.85 ± 1.47, /, /
/
/
  
0.10 ± 0.13, /, /
/
0.12 ± 0.11, /, /
/
1.35 ± 0.52, /, /
/
/
  
0.37 ± 0.17, /, /
/
0.37 ± 0.17, /, /
/
1.47 ± 0.62, /, /
/
/
  
0.23 ± 0.17, /, /
/
0.23 ± 0.17, /, /
/
1.79 ± 0.57, /, /
/
/
19
Yen-Ting Han (2021)[38]
1.84 ± 0.64, /, (0.92, 3.13)
1.12 ± 0.40, /, (0.41, 1.68)
2.24 ± 0.97, /, (0.65, 3.54)
1.57 ± 0.96, /, (0.34, 3.32)
6.44 ± 3.02, /, (2.09, 14.00)
1.26 ± 0.90, /, (0.03, 2.90)
/
  
0.97 ± 0.45, /, (0.00, 2.19)
0.69 ± 0.41, /, (0.00, 1.82)
1.27 ± 0.58, /, (0.00, 3.18)
1.04 ± 0.58, /, (0.00, 2.90)
3.21 ± 1.72, /, (0.01, 7.25)
0.57 ± 0.43, /, (0.00, 1.77)
/
  
1.07 ± 0.54, /, /
0.90 ± 0.47, /, /
1.34 ± 0.77, /, /
1.18 ± 0.75, /, /
3.67 ± 2.14, /, /
0.49 ± 0.41, /, /
/
  
0.90 ± 0.37, /, /
0.54 ± 0.29, /, /
1.22 ± 0.40, /, /
0.95 ± 0.41, /, /
2.89 ± 1.30, /, /
0.62 ± 0.44, /, /
/
  
0.98 ± 0.37, /, /
0.62 ± 0.33, /, /
1.18 ± 0.47, /, /
0.88 ± 0.49, /, /
3.12 ± 1.53, /, /
0.64 ± 0.44, /, /
/
  
0.89 ± 0.45, /, /
0.65 ± 0.36, /, /
1.25 ± 0.62, /, /
1.08 ± 0.57, /, /
3.04 ± 1.89, /, /
0.53 ± 0.40, /, /
/
20
Paknisa Sittikornpaiboon (2021) [39]
0.56 ± 0.19, 0.51, (0.32, 0.96)
/
0.83 ± 0.32, 0.75, (0.49, 1.49)
/
2.70 ± 1.37, 2.95, (0.90, 5.10)
/
/
  
0.42 ± 0.12, 0.41, (0.25, 0.63)
/
0.76 ± 0.22, 0.73, (0.45, 1.14)
/
2.50 ± 0.89, 2.70, (0.70, 3.60)
/
/
  
1.18 ± 0.19, 1.13, (0.86, 1.48)
/
1.70 ± 0.41, 1.63, (1.08, 2.38)
/
4.37 ± 1.34, 4.00, (2.70, 6.50)
/
/
  
1.09 ± 0.12, 1.09, (0.90, 1.25)
/
1.95 ± 0.48, 1.98, (0.94, 2.53)
/
5.13 ± 1.86, 5.45, (0.70, 6.90)
/
/
  
0.81 ± 0.15, 0.83, (0.47, 1.01)
/
1.73 ± 0.23, 1.72, (1.41, 2.07)
/
5.30 ± 1.04, 5.45, (3.60, 6.50)
/
/
21
Lirong Huang (2021)[40]
0.82 ± 0.32, /, /
/
1.18 ± 0.41, /, /
/
3.24 ± 1.33, /, /
0.64 ± 0.36, /, /
1
  
0.89 ± 0.34, /, /
/
1.10 ± 0.38, /, /
/
3.38 ± 1.86, /, /
0.25 ± 0.77, /, /
1
22
Rani D’haese (2021)[41]
0.82 ± 0.43, /, (0.17, 2.08)
/
0.99 ± 0.45, /, (0.12, 2.06)
/
3.25 ± 1.69, /, (0.16, 8.70)
/
4
  
0.45 ± 0.31, /, (0.05, 1.62)
/
0.71 ± 0.43, /, (0.15, 2.14)
/
2.39 ± 1.42, /, (0.37, 8.16)
/
4
23
Chia-Cheng Lin (2020) [42]
0.78 ± 0.57, /, (0.06, 2.97)
0.47 ± 0.33, /, (0.04, 1.39)
1.29 ± 0.88, /, (0.07, 3.53)
1.10 ± 0.80, /, (0.05, 3.35)
3.67 ± 2.73, /, (0.16, 11.32)
0.57 ± 0.53, /, (0.05, 2.76)
/
  
0.75 ± 0.47, /, (0.22, 2.37)
0.49 ± 0.39, /, (0.02, 1.92)
1.30 ± 1.00, /, (0.42, 4.38)
1.12 ± 1.01, /, (0.17, 4.19)
3.68 ± 3.66, /, (0.23, 16.21)
0.49 ± 0.39, /, (0.03, 1.39)
/
  
0.75 ± 0.33, /, (0.23, 1.67)
0.63 ± 0.35, /, (0.21, 1.66)
1.24 ± 0.85, /, (0.21, 3.83)
1.15 ± 0.88, /, (0.20, 3.81)
3.55 ± 2.97, /, (0.12, 12.65)
0.31 ± 0.24, /, (0.01, 0.94)
/
24
Yuan Chen (2020) [43]
0.59 ± 0.28, /, (0.10, 1.30)
/
0.99 ± 0.41, /, (0.20, 1.80)
/
1.91 ± 1.02, /, (0.20, 4.20)
0.38 ± 0.26, /, (0.00, 1.10)
1
  
1.04 ± 0.64, /, (0.10, 3.10)
/
1.46 ± 0.64, /, (0.50, 3.30)
/
2.77 ± 1.72, /, (0.40, 6.30)
0.84 ± 0.68, /, (0.10, 3.10)
0
25
Dong Wu (2020) [44]
1.22 ± 0.70, /, /
/
1.33 ± 0.73, /, /
/
4.34 ± 2.22, /, /
/
/
26
Kang-jie Cheng (2020) [45]
0.79 ± 0.17, /, /
0.61 ± 0.19, /, /
1.26 ± 0.27, /, /
0.91 ± 0.55, /, /
3.77 ± 1.57, /, /
(coronal)0.38 ± 0.17, /, / (apical)0.37 ± 0.20, /, /
/
27
Pantip Henprasert (2020) [46]
0.27 ± 0.12, /, /
/
0.81 ± 0.28, /, /
/
(mesio-distal)
0.77 ± 0.72, /, / (bucco-lingual)
1.77 ± 0.76, /, /
(buccal)0.21 ± 0.24, /, /
(lingual)0.23 ± 0.12, /, /
(distal)0.40 ± 0.32, /, /
(mesial)0.33 ± 0.37, /, /
/
  
0.32 ± 0.15, /, /
/
0.84 ± 0.47, /, /
/
(mesio-distal) 0.78 ± 0.80, /, / (bucco-lingual)
1.60 ± 1.22, /, /
(buccal)0.24 ± 0.23, /, /
(lingual)0.25 ± 0.17, /, /
(distal)0.33 ± 0.23, /, /
(mesial)0.37 ± 0.28, /, /
/
28
Kristian Kniha (2020) [47]
1.47 ± 0.86, /, (0.0, 3.40)
/
1.77 ± 0.85, /, (0.50, 3.20)
/
2.81 ± 2.69, /, (0.00, 8.40)
0.10 ± 0.46, /, (− 0.70, 0.90)
/
  
1.47 ± 0.86, /, (0.0, 3.40)
/
1.77 ± 0.85, /, (0.50, 3.20)
/
2.81 ± 2.69, /, (0.00, 8.40)
− 0.07 ± 0.54, /, (− 0.90, 1.00)
/
  
1.31 ± 0.61, /, (0.10, 2.60)
/
1.91 ± 0.79, /, (0.50, 3.10)
/
6.22 ± 4.26, /, (0.00, 15.30)
0.22 ± 0.58, /, (− 1.00, 1.10)
/
  
1.31 ± 0.61, /, (0.10, 2.60)
/
1.91 ± 0.79, /, (0.50, 3.10)
/
6.22 ± 4.26, /, (0.00, 15.30)
− 0.31 ± 0.66, /, (− 1.40, 0.90)
/
29
R. Vinci (2020) [48]
/
0.67 ± 0.37, /, (0.30, 1.77)
/
0.89 ± 0.30, /, (0.10, 1.57)
/
/
3
  
/
0.12 ± 0.28, /, (0.08, 1.18)
/
0.31 ± 0.43, /, (0.30, 1.77)
/
/
3
  
/
0.41 ± 0.31, /, (0.08, 1.30)
/
0.88 ± 0.44, /, (0.17, 2.66)
/
/
3
  
/
0.31 ± 0.38, /, (0.27, 1.77)
/
0.79 ± 0.40, /, (0.10, 3.54)
/
/
3
30
Nopparat Suksod (2020) [49]
0.98 ± 0.48, /, /
/
1.57 ± 0.46, /, /
/
4.23 ± 1.84, /, /
/
/
31
Márton Kivovics (2020) [50]
1.987 ± 0.7049, /, /
/
1.954 ± 0.6853, /, /
/
6.544 ± 5.393, /, /
/
3
  
1.879 ± 0.7893, /, /
/
2.124 ± 0.8373, /, /
/
7.177 ± 4.214, /, /
/
3
32
Palita Smitkarn (2019) [51]
1.0 ± 0.6, 0.9, (0.20, 2.67)
/
1.3 ± 0.6, 1.2, (0.24, 2.57)
/
3.1 ± 2.3, 2.8, (0.00, 8.60)
/
/
33
Karim El Kholy (2019) [52]
0.284 ± 0.133, /, (0.051, 0.583)
/
0.675 ± 0.429, /, (0.140, 1.980)
/
4.363 ± 1.682, /, (1.180, 8.800)
/
/
  
1.015 ± 0.124, /, (0.840, 1.230)
/
1.657 ± 0.209, /, (1.310, 1.940)
/
7.713 ± 1.236, /, (5.500, 10.500)
/
/
  
0.562 ± 0.086, /, (0.410, 0.710)
/
1.195 ± 0.397, /, (0.850, 1.920)
/
5.688 ± 1.521, /, (3.200, 8.300)
/
/
  
0.289 ± 0.159, /, (0.060, 0.591)
/
0.616 ± 0.255, /, (0.220, 1.000)
/
4.731 ± 1.601, /, (2.600, 8.500)
/
/
34
Henrik Skjerven (2019) [53]
1.05 ± 0.59, /, (0.36, 2.74)
/
1.63 ± 1.05, /,
(0.56, 5.16)
/
3.85 ± 1.83, /, (1.25, 8.60)
0.48 ± 0.5, /, (0.52, 1.34)
/
35
Rai-Jei Chang (2018) [54]
/, 0.40, (0.00, 1.00)
/
/, 0.65, (0.10, 1.90)
/
/, 2.16, (0.17, 6.91)
/
1
  
/, 0.95, (0.30, 1.30)
/
/, 1.35, (0.10, 3.60)
/
/, 3.92, (0.44, 11.66)
/
1
36
Zhaozhao Chen (2018) [55]
0.85 ± 0.38, /, (0.42, 1.51)
/
0.93 ± 0.34, /, (0.64, 1.72)
/
3.11 ± 1.55, /, (0.66, 4.95)
(Considering direction) -0.32 ± 0.48, /, (− 1.00, 0.64)
(Absolute value) 0.50 ± 0.26, /, (0.18, 1.00)
/
37
Jan Brandt (2018) [56]
/
0.725 ± 0.142, /, (0.518, 1.112)
/
0.990 ± 0.244, /, (0.633, 1.526)
2.011 ± 0.855, /, (0.366, 4.036)
0.541 ± 0.129, /, (0.242, 0.848)
/
38
Philipp Kauffmann (2018) [57]
/, 0.47, (0.05, 1.31)
/
/, 0.86, (0.21, 1.68)
/
/, 3.41, (0.48, 5.79)
/, 0.44, (0.03, 1.54)
3
  
/, 0.49, (0.10, 1.11)
/
/, 0.77, (0.16, 1.86)
/
/, 2.76, (0.32, 7.54)
/, 0.52, (0.06, 1.69)
0
39
Yiqin Fang (2018) [58]
0.46, /, (0.00, 1.15)
/
0.67, /, (0.14, 1.19)
/
1.40, /, (0.30, 2.57)
0.15, /, (0.10, 0.82)
/
40
Boyoung Ma (2018) [59]
0.82 ± 0.44, /, (0.13, 1.85)
/
1.19 ± 0.46, /, (0.37, 2.51)
/
2.43 ± 1.13, /,
(1.15, 5.70)
-0.03 ± 0.65, /,
(-2.12, 1.58)
/
  
1.37 ± 0.80, /, (0.18, 3.76)
/
1.77 ± 0.86, /, (0.45, 3.76)
/
4.74 ± 2.06, /,
(0.00, 8.86)
/
/
41
Björn Gjelvold (2018) [60]
0.27 ± 0.08, /, /
/
0.34 ± 0.14, /, /
/
0.99 ± 0.57, /, /
0.16 ± 0.11, /, /
/
  
0.39 ± 0.01, /, /
/
0.49 ± 0.17, /, /
/
1.25 ± 0.49, /, /
0.34 ± 0.18, /, /
/
Table 3
Forest plot showing the global coronal deviations of the reviewed studies concerning different guide supporting types in different research types
https://static-content.springer.com/image/art%3A10.1186%2Fs40729-023-00507-w/MediaObjects/40729_2023_507_Tab3_HTML.png
Table 4
Forest plot showing the global coronal deviations, global apical deviations and angular deviations of the reviewed studies concerning fixation screws application
https://static-content.springer.com/image/art%3A10.1186%2Fs40729-023-00507-w/MediaObjects/40729_2023_507_Tab4_HTML.png
Table 5
Forest plot showing the angular deviations of the reviewed studies concerning different guide fabrication strategies in different research types
https://static-content.springer.com/image/art%3A10.1186%2Fs40729-023-00507-w/MediaObjects/40729_2023_507_Tab5_HTML.png
As presented in Table 3 and Additional file 1: Tables S1, S2 forest plots of global coronal deviations, global apical deviations and angular deviations from the reviewed studies are summarized according to different guide supporting types. Tooth-supported guides are the most frequently applied in studies published in recent five years: among the 41 involved literatures, 27 involved bilateral tooth-supported guides, 9 involved mixed tooth-/bone- or tooth-/mucosa-supported guides (unilateral tooth-supported), and only 7 applied mucosa-supported guides. The accuracy-related numerical data (mean value, or median when mean value is not available) of bilateral tooth-supported implant guide ranges within 0.1–1.18 mm (in vitro)/0.46–1.47 mm (in vivo) in global coronal deviation, 0.18 ~ 1.37 mm (in vitro)/0.39–1.07 mm (in vivo) in horizontal coronal deviation, 0.12–1.95 mm (in vitro)/0.28–1.77 mm (in vivo) in global apical deviation, 0.31–1.68 mm (in vitro)/0.64–1.17 mm (in vivo) in horizontal apical deviation, 0.77–7.713° (in vitro)/1.4–4.74° (in vivo) in angular deviation, and 0.11–0.95 mm (in vitro)/0.03–0.84 mm (in vivo) in vertical deviation. Global coronal deviations of unilateral tooth-supported guides (including mixed tooth-/bone- or tooth-/mucosa-supported guides) are between 0.284–1.43 mm (in vitro)/0.21–1.2 mm (in vivo), horizontal coronal deviations between 0.47–0.67 mm (in vitro), global apical deviations between 0.65–2.19 mm (in vitro)/0.67–1.45 mm (in vivo), horizontal apical deviations between 0.91–1.72 mm (in vitro), angular deviations between 2.16–6.81 mm (in vitro)/3.1–5.62 mm (in vivo)°, and vertical deviations between 0.16–0.57 mm (in vitro). Global coronal deviations of mucosa-supported guides are between 0.45–0.82 mm (in vitro)/0.98–1.987 mm (in vivo), horizontal coronal deviations between 0.12–0.67 mm (in vivo), global apical deviations between 0.71–0.99 mm (in vitro)/1.18–2.124 mm (in vivo), horizontal apical deviations between 0.31–0.89 mm (in vivo), angular deviations between 2.39–3.41° (in vitro)/3.12–7.177° (in vivo), and vertical deviations between 0.44–0.52 mm (in vitro)/0.22–0.64 mm (in vivo) (Fig. 4, more data see Table 3).
Only 3 studies involved the comparison of guides with/without fixation screws, among which 2 were in vitro studies, and 1 was in vivo study (Table 4). Only 2 studies involved the comparison of sleeve length of guides.
Among the 41 involved literatures in this review, 29 involved 3D printing-fabricated guides, 3 involved milling-fabricated guides, 3 involved both, and 6 involved no information of fabricating methods. The in vitro accuracy-related numerical data (mean value, or median when mean value is not available) of milling guides ranges within 0.2–1.37 mm in horizontal coronal deviation, 0.62–1.68 mm in horizontal apical deviation, 0.61–3.49° in angular deviation; and the in vitro data of 3d printing guides ranges within 0.1–1.43 mm in global coronal deviation, 0.18–0.95 mm in horizontal coronal deviation, 0.12–2.19 mm in global apical deviation, 0.31–1.72 mm in horizontal apical deviation, 0.78–7.713° in angular deviation., and 0.11–0.64 mm in vertical deviation. The in vivo data of milling guides ranges within 0.12–0.67 mm in horizontal coronal deviation, 0.31–0.89 mm in horizontal apical deviation; and the data of 3d printing guides ranges within 0.21–1.987 mm in global coronal deviation, 0.39–1.12 mm in horizontal coronal deviation, 0.28–2.24 mm in global apical deviation, 0.64–1.57 mm in horizontal apical deviation, 1.4–7.177° in angular deviation, and 0.03–1.26 mm in vertical deviation (Fig. 5, more data see Table 2). Studies of milled guides mainly apply the horizontal deviation indicators following its coordinate system, less than 3 literatures apply the global deviation indicators, therefore the forest plots (Table 5 and Supplementary Table S3-S4) and normal distributions (Fig. 5) concerning different fabrication approaches are based on horizontal coronal deviations and horizontal apical deviations.
As visually shown in Tables 3, 4, 5 and Additional file 1: Tables S1–S4, with current technology of digital implant guides, implant surgery accuracy can be achieved with the mean distance deviation < 2 mm (most < 1 mm) and angular deviation < 8° (most < 5°).

Discussion

Guide supporting type

To verify the hypothesis that supporting types influence the accuracies of surgical guides, we collectively categorized and analyzed guide type and deviation data in existing literature. Implant guides are divided into categories according to its support types, including bone-supported, mucosa-supported, tooth-supported, and any combination (Fig. 3). Theoretically, the anatomical differences among teeth, bone and mucosa may lead to different accuracy of guides with different support types. Bilateral tooth-supported guides provide best retention and biomechanical stability with anchorage on hard tissue, therefore theoretically endow highest accuracy.
Although with advantages in accuracy and operability, bilateral tooth-supported guides are indicated for patients with intact teeth both mesial and distal to the edentulous area. As for distal extension edentulism, one of the most common clinical manifestation, unilateral tooth-supported guides including mixed tooth-/bone- or tooth-/mucosa-supported guides are often used to provide efficient retention.
Bone-supported guides are overlaid on the alveolar crest exposed via full‐thickness mucoperiosteal flap operation and fixed with fixation screws. Its larger surgical wound, upturned tissue flap affects its repositioning, resulting in relatively low theoretical accuracy. Simple bone-supported guides are seldom reported in recent five years [62], and among the 41 researches included in this review, only two studies applied mixed tooth-/bone-supported guides [63, 64].
Mucosa-supported guides are indicated for completely edentulous patients or patients who barely have residual teeth. Without flap operation, it is anchored to the bone through the mucosa with fixation screws. To be noted, a recent research reported that calculation of implant angular deviation of mucosa-supported guides by tissue or implant alignment resulted in different values [41], emphasizing the lack of standard for accuracy measurement, and indicating that comparability of the accuracy indicator values in different literatures should be reviewed dialectically.
Since different guide support types are used in various situations, the existing literature rarely directly compares the accuracy among different guide types. A systematic review in 2012 compares the accuracy of different guide types, and reveals a significant higher accuracy of tooth- and mucosa-supported guides than bone-supported ones, with no significant difference between tooth- and mucosa-supported guides [63]. The review mentions only two researches that directly compare different support types [65, 66]. A systematic review in 2018 showed that the implantation placement in cases using tooth-supported guides is more accurate than bone- and mucosa-supported [16].
In this review, as accuracy-related parameters differ in different studies, three parameters with relatively high frequency of application (global coronal deviations, global apical deviations and angular deviations) are selected for normal distribution analyses (Fig. 4). It is interesting to notice that the normal distribution of deviations varied with research types, especially for mucosa-supported guides, which exhibited the highest deviation peaks in all three parameters in in vivo researches (Fig. 4A–C), and the lowest in deviation peaks and narrowest distribution in vitro studies (Fig. 4D–F). This opposite trend was probably attributed to the different elasticity between in vitro experimental model and in vivo mucosa. In vitro experimental mucosa-imitating models exhibited simpler mechanical properties than living tissues, resulting in higher accuracies.
In addition, the use of anesthetics may affect the accuracy of the procedure due to elasticity of mucosa and its deformation upon penetration [67]. Compared to bilateral tooth-supported guides, unilateral tooth guides exhibited higher deviations in most categories (Fig. 4C–F), except in in vivo global coronal deviation where unilateral tooth-supported guides exhibited slightly lower distribution than bilateral guides (Fig. 4A), and in in vivo global apical deviation where bilateral and unilateral tooth-supported guides showed a similar data range (Fig. 4B).
In summary, bilateral tooth-supported guides exhibited the highest in vitro accuracy and similar in vivo accuracy to unilateral tooth-supported guides; mucosa-supported guides exhibit the lowest in vivo accuracy, while its in vitro data showed low credibility due to the mechanical complexity of living mucosa tissue.

Design of fixation screws

For bone- and mucosa-supported guides, fixation screws can be further introduced to fix the surgical guide and avoid displacement. The accuracy of implantation is reported to be improved by the application of fixation screws and influenced by its distribution [63, 68]. Since fixation screws are regular designs for implant guides, guides without fixation screws are only involved in three studies [21, 43, 57] as included in this review, all compared to groups with fixation screws in the same research. The difference between the two in vitro studies [21, 57] was not significant, and though the in vivo study [43] reported significantly higher accuracy in experimental group using fixation screws, whether the accuracy was influenced by the guide type (full-/half-) or the usage of fixation screws still requires further study due to the small sample size (Table 4).
For mucosa-supported guides used in edentulous patients, the use of fixation screws provide larger surface support and reduce the intraoperative displacement [69], efficiently reduce the angular deviation [68], depth deviation and horizontal deviation [57]. Therefore, in cases demanding a high depth precision and avoiding injury to the mandibular nerve, application of fixation screws contribute to better implant results [57].
For mixed tooth-/mucosa-supported guide in free-end dental implantation, application of fixation screws also results in a significant improvement in the accuracy regarding horizontal apical and depth deviation (direction considered) [21]. Apart from mucosa-supported and mixed tooth-/mucosa-supported guides, fixation screws can also be introduced into tooth-supported guides to achieve improved stability in both maxillary and mandibular anterior implantation [40, 43].
Concerning the number of fixation screws, most design applied three-point fixation [48, 50, 57], though the number can be adjusted [21, 25, 40, 43, 54]. A systematic review in 2012 indicates that when fixation screws are used, the mean deviation for all indicators are reduced, and the deviations decrease as the number of fixation screws increases, but the indicators exhibit no significant difference [63]. Design of two equally distributed fixation screws can also efficiently stabilize the surgical guide, located approximately at the lateral incisor [70]. Ideally, for fully edentulous maxillae, four-point fixation (two in anterior area and distal to the implant site) that covers the entire maxillary arch can efficiently avoid bending of the guide in the distal areas of the surgical field [71].

Design of sleeve

The guidance of drill hole, implant direction, depth, and angle are realized via design of sleeves, which can also reduce surgical time [7274]. Sleeves can be classified as open or closed. Open sleeves with C-shaped buccally opening are applied in posterior areas where mouth opening and interarch space are limited or insufficient [74]. To ensure implant accuracy, the drill should be in the center and parallel to the inner wall of sleeves during hole preparation [75]. As summarized in Table 2, among the 41 involved literatures in this review, only 2 involved the comparison of sleeve design of guides, indicating that smaller distance from sleeve to bone leads to more accurate results [37], and that sleeve design might affect the accuracy [39].
Implant accuracy is affected by the design of height, drilling distance, and sleeve–bone distance, but the sleeve–implant distance and the sleeve axis angle do not affect the accuracy of digital implant guides [20]. By using shorter sleeve heights or shorter implants, decreasing the drilling distance below the guided sleeve can significantly increase the implant accuracy and reduce lateral movement of the drill [76]. However, sleeve heights ≤ 5 mm lead to implant placement deviation and decrease of the accuracy [77]. The increased drilling distance beyond the guiding sleeve results in a significant global and angular deviation at both the implant crest and apex [76]. Decreased sleeve–bone distance results in higher accuracy of the implant surgical guide. With the sleeve–bone distance of 2 or 4 mm, the implant accuracy of closed and open sleeve is similar; whereas with the sleeve–bone distance of 6 mm, lower accuracy is shown in both open and closed sleeves, and open sleeves exhibited a more significant trend [78].
In addition, material of the sleeve also affects implant accuracy. Metal sleeves are common in early surgery guides, and with the development of material science and technology, it is reported that plastic sleeves endow lower angle deviation, depth deviation, placement deviation than metal ones, as well as ensure a faster and easier guided surgery work-flow [79].

Manufacturing accuracy

Currently, digital implant guides can be manufactured using additive manufacturing (3D printing) or subtractive manufacturing (CAD/CAM milling). Typical 3D printing includes stereo lithography appearance (SLA), PolyJet, MultiJet, fused filament fabrication (FFF), digital light processing (DLP), etc. The difference in manufacturing accuracy between these two strategies remains inconclusive. It has been reported that the processing accuracy of milling (0.02–0.25 mm) is higher than printing (0.03–0.44 mm) in the aspect of inner surface, vertical fit discrepancy, guide seating distortion, and error range of anterior and posterior implants [80]. However, some literatures indicated no significant difference in manufacturing accuracy between milling and 3D printing [81]. Difference of manufacturing accuracy among different 3D printing manufacturing technologies has also been reported. For example, DLP printer exhibited a lower manufacturing error compared to FFF printer [80].
Despite of the differences in manufacturing accuracy, clinical researches indicated no significant differences in all surgical accuracy indicators between additive and subtractive manufactured guides [46]. Implant guides manufactured by different additive technologies (SLA, PolyJet and MultiJet) exhibited similar surgical accuracy regarding the angle, coronal and apical deviation [34]. DLP printer showed higher accuracy than SLA printer in coronal and vertical deviation, and no significant difference in apical, horizontal and angle deviation [60].
As shown in Table 5 and Additional file 1: Tables S3–S4, in terms of surgical guide fabrication, 3d printing is of wider application than milling. As indicated in Fig. 5, among the included literatures in this review, 3D printed guides exhibited higher peaks of horizontal coronal and horizontal apical deviations in in vivo researches than milled guides (Fig. 5A, D, in vivo angular deviation data were not sufficient to perform normal distribution analysis herein). While in in vitro researches, different trends were observed in the three deviation parameters (Fig. 5B, C, E).
In summary, though milled guides exhibited higher in vivo accuracy than 3d printed guides, the small sample size of milled guides reviewed in this article resulted in relatively low reliability of this conclusion, and further data support might be needed.

Conclusion and prospect

This review has verified the hypothesis that guide supporting types, manufacturing methods and design of implant guides (including fixation screws and sleeves) could influence the accuracies of static digital surgical guides by collecting and categorizing the numerical data of surgical accuracy indicator reported in literature of the last five years. Bilateral tooth-supported guides exhibited the highest in vitro accuracy and similar in vivo accuracy to unilateral tooth-supported guides; mucosa-supported guides exhibit the lowest in vivo accuracy, while its in vitro data showed low credibility due to the mechanical complexity of living mucosa tissue. Milled guides exhibited higher in vivo accuracy than 3d printed guides, though further data support might be needed. Apart from operator's skill and standardization that may affect the accuracy of implantation, this review has revealed that with current medical technology and the aid of digital implant guides, implant surgery accuracy can be achieved with the distance deviation < 2 mm (most < 1 mm) and angular deviation < 8° (most < 5°). The bottleneck of surgical accuracy improvement resulted from the difficulty of guide fixation in edentulous patients. In addition, the lack of a universal evaluation standard for implantation surgical accuracy remained a major problem in this research field.
As the design of supporting types, fixation screws and sleeves of implant guides can affect the accuracy of implant surgeries, existing studies focus on improving the accuracy via selecting appropriate supporting types, optimizing and customizing the guide design (including fixation screws and sleeves) according to individual demands. Future developing trend of this field may continuously focus on standardization of the evaluation of surgical accuracy and improving minimally invasive surgical methods (such as gradually phasing out bone-support guides that involve flap surgeries). The improvement of implant accuracy for edentulous patients has been a field of intense research in recent years and may remain a research focus in the near future.
The influence of implant guide design on surgical accuracy revealed in this review may shed light on future improvement of digital implant guides. However, this review only analyzed and discussed four influencing factors that affected the implantation accuracy, other factors including the guiding protocol (full/half guide), implant position (maxillary/mandibular, anterior/posterior, etc.), implant size, bone quality, etc., remained undiscussed and required further analysis in future reviews.

Declarations

Not applicable.
All of the authors are in agreement with the content of the manuscript.

Competing interests

The authors report no conflicts of interest related to the subject of this review.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Al Yafi F, Camenisch B, Al-Sabbagh M. Is digital guided implant surgery accurate and reliable? Dent Clin North Am. 2019;63(3):381–97.PubMedCrossRef Al Yafi F, Camenisch B, Al-Sabbagh M. Is digital guided implant surgery accurate and reliable? Dent Clin North Am. 2019;63(3):381–97.PubMedCrossRef
2.
Zurück zum Zitat Chen P, Nikoyan L. Guided implant surgery: a technique whose time has come. Dent Clin North Am. 2021;65(1):67–80.PubMedCrossRef Chen P, Nikoyan L. Guided implant surgery: a technique whose time has come. Dent Clin North Am. 2021;65(1):67–80.PubMedCrossRef
3.
Zurück zum Zitat Putra RH, Yoda N, Astuti ER, Sasaki K. The accuracy of implant placement with computer-guided surgery in partially edentulous patients and possible influencing factors: a systematic review and meta-analysis. J Prosthodont Res. 2022;66(1):29–39.PubMedCrossRef Putra RH, Yoda N, Astuti ER, Sasaki K. The accuracy of implant placement with computer-guided surgery in partially edentulous patients and possible influencing factors: a systematic review and meta-analysis. J Prosthodont Res. 2022;66(1):29–39.PubMedCrossRef
4.
Zurück zum Zitat Flügge T, Derksen W, Te Poel J, Hassan B, Nelson K, Wismeijer D. Registration of cone beam computed tomography data and intraoral surface scans—a prerequisite for guided implant surgery with CAD/CAM drilling guides. Clin Oral Implants Res. 2017;28(9):1113–8.PubMedCrossRef Flügge T, Derksen W, Te Poel J, Hassan B, Nelson K, Wismeijer D. Registration of cone beam computed tomography data and intraoral surface scans—a prerequisite for guided implant surgery with CAD/CAM drilling guides. Clin Oral Implants Res. 2017;28(9):1113–8.PubMedCrossRef
5.
Zurück zum Zitat Chandran SK, Jaini JL, Babu AS, Mathew A, Keepanasseril A. Digital versus conventional impressions in dentistry: a systematic review. J Clin Diagn Res. 2019;13(4):1–6. Chandran SK, Jaini JL, Babu AS, Mathew A, Keepanasseril A. Digital versus conventional impressions in dentistry: a systematic review. J Clin Diagn Res. 2019;13(4):1–6.
6.
Zurück zum Zitat Cicciu M, Fiorillo L, D’Amico C, Gambino D, Amantia EM, Laino L, et al. 3D digital impression systems compared with traditional techniques in dentistry: a recent data systematic review. Materials. 2020;13(8):1982.PubMedPubMedCentralCrossRef Cicciu M, Fiorillo L, D’Amico C, Gambino D, Amantia EM, Laino L, et al. 3D digital impression systems compared with traditional techniques in dentistry: a recent data systematic review. Materials. 2020;13(8):1982.PubMedPubMedCentralCrossRef
7.
Zurück zum Zitat Kihara H, Hatakeyama W, Komine F, Takafuji K, Takahashi T, Yokota J, et al. Accuracy and practicality of intraoral scanner in dentistry: a literature review. J Prosthodont Res. 2020;64(2):109–13.PubMedCrossRef Kihara H, Hatakeyama W, Komine F, Takafuji K, Takahashi T, Yokota J, et al. Accuracy and practicality of intraoral scanner in dentistry: a literature review. J Prosthodont Res. 2020;64(2):109–13.PubMedCrossRef
8.
Zurück zum Zitat Giachetti L, Sarti C, Cinelli F, Russo DS. Accuracy of digital impressions in fixed prosthodontics: a systematic review of clinical studies. Int J Prosthodont. 2020;33(2):192–201.PubMedCrossRef Giachetti L, Sarti C, Cinelli F, Russo DS. Accuracy of digital impressions in fixed prosthodontics: a systematic review of clinical studies. Int J Prosthodont. 2020;33(2):192–201.PubMedCrossRef
9.
Zurück zum Zitat Jacobs R, Salmon B, Codari M, Hassan B, Bornstein MM. Cone beam computed tomography in implant dentistry: recommendations for clinical use. BMC Oral Health. 2018;18(1):88.PubMedPubMedCentralCrossRef Jacobs R, Salmon B, Codari M, Hassan B, Bornstein MM. Cone beam computed tomography in implant dentistry: recommendations for clinical use. BMC Oral Health. 2018;18(1):88.PubMedPubMedCentralCrossRef
10.
Zurück zum Zitat Fokas G, Vaughn VM, Scarfe WC, Bornstein MM. Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: a systematic review. Clin Oral Implants Res. 2018;29:393–415.PubMedCrossRef Fokas G, Vaughn VM, Scarfe WC, Bornstein MM. Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: a systematic review. Clin Oral Implants Res. 2018;29:393–415.PubMedCrossRef
11.
Zurück zum Zitat Mai HN, Dam VV, Lee DH. Accuracy of augmented reality-assisted navigation in dental implant surgery: systematic review and meta-analysis. J Med Internet Res. 2023;25: e42040.PubMedPubMedCentralCrossRef Mai HN, Dam VV, Lee DH. Accuracy of augmented reality-assisted navigation in dental implant surgery: systematic review and meta-analysis. J Med Internet Res. 2023;25: e42040.PubMedPubMedCentralCrossRef
12.
Zurück zum Zitat Naeini EN, Atashkadeh M, De Bruyn H, D’Haese J. Narrative review regarding the applicability, accuracy, and clinical outcome of flapless implant surgery with or without computer guidance. Clin Implant Dent Relat Res. 2020;22(4):454–67.PubMedPubMedCentralCrossRef Naeini EN, Atashkadeh M, De Bruyn H, D’Haese J. Narrative review regarding the applicability, accuracy, and clinical outcome of flapless implant surgery with or without computer guidance. Clin Implant Dent Relat Res. 2020;22(4):454–67.PubMedPubMedCentralCrossRef
13.
Zurück zum Zitat Subramani K. Is computer-guided implant placement with a flapless approach more accurate than with a flapped surgical approach? Evid Based Dent. 2022;23(3):110–1.PubMedCrossRef Subramani K. Is computer-guided implant placement with a flapless approach more accurate than with a flapped surgical approach? Evid Based Dent. 2022;23(3):110–1.PubMedCrossRef
14.
Zurück zum Zitat Seo C, Juodzbalys G. Accuracy of guided surgery via stereolithographic mucosa-supported surgical guide in implant surgery for edentulous patient: a systematic review. J Oral Maxillofac Res. 2018;9(1): e1.PubMedPubMedCentralCrossRef Seo C, Juodzbalys G. Accuracy of guided surgery via stereolithographic mucosa-supported surgical guide in implant surgery for edentulous patient: a systematic review. J Oral Maxillofac Res. 2018;9(1): e1.PubMedPubMedCentralCrossRef
15.
Zurück zum Zitat EftekharAshtiani R, Ghasemi Z, Nami M, Mighani F, Namdari M. Accuracy of static digital surgical guides for dental implants based on the guide system: a systematic review. J Stomatol Oral Maxillofac Surg. 2021;122(6):600–7.CrossRef EftekharAshtiani R, Ghasemi Z, Nami M, Mighani F, Namdari M. Accuracy of static digital surgical guides for dental implants based on the guide system: a systematic review. J Stomatol Oral Maxillofac Surg. 2021;122(6):600–7.CrossRef
16.
Zurück zum Zitat Bover-Ramos F, Vina-Almunia J, Cervera-Ballester J, Penarrocha-Diago M, Garcia-Mira B. Accuracy of implant placement with computer-guided surgery: a systematic review and meta-analysis comparing cadaver, clinical, and in vitro studies. Int J Oral Maxillofac Implants. 2018;33(1):101–15.PubMedCrossRef Bover-Ramos F, Vina-Almunia J, Cervera-Ballester J, Penarrocha-Diago M, Garcia-Mira B. Accuracy of implant placement with computer-guided surgery: a systematic review and meta-analysis comparing cadaver, clinical, and in vitro studies. Int J Oral Maxillofac Implants. 2018;33(1):101–15.PubMedCrossRef
17.
Zurück zum Zitat Tahmaseb A, Wu V, Wismeijer D, Coucke W, Evans C. The accuracy of static computer-aided implant surgery: a systematic review and meta-analysis. Clin Oral Implants Res. 2018;29(Suppl 16):416–35.PubMedCrossRef Tahmaseb A, Wu V, Wismeijer D, Coucke W, Evans C. The accuracy of static computer-aided implant surgery: a systematic review and meta-analysis. Clin Oral Implants Res. 2018;29(Suppl 16):416–35.PubMedCrossRef
18.
Zurück zum Zitat Raico Gallardo YN, da Silva-Olivio IRT, Mukai E, Morimoto S, Sesma N, Cordaro L. Accuracy comparison of guided surgery for dental implants according to the tissue of support: a systematic review and meta-analysis. Clin Oral Implants Res. 2017;28(5):602–12.PubMedCrossRef Raico Gallardo YN, da Silva-Olivio IRT, Mukai E, Morimoto S, Sesma N, Cordaro L. Accuracy comparison of guided surgery for dental implants according to the tissue of support: a systematic review and meta-analysis. Clin Oral Implants Res. 2017;28(5):602–12.PubMedCrossRef
19.
Zurück zum Zitat Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021;372: n71.PubMedPubMedCentralCrossRef Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021;372: n71.PubMedPubMedCentralCrossRef
20.
Zurück zum Zitat Liu X, Liu J, Feng H, Pan S. Accuracy of a milled digital implant surgical guide: an in vitro study. J Prosthet Dent. 2022;127(3):453–61.PubMedCrossRef Liu X, Liu J, Feng H, Pan S. Accuracy of a milled digital implant surgical guide: an in vitro study. J Prosthet Dent. 2022;127(3):453–61.PubMedCrossRef
21.
Zurück zum Zitat Pessoa R, Siqueira R, Li J, Saleh I, Meneghetti P, Bezerra F, et al. The impact of surgical guide fixation and implant location on accuracy of static computer-assisted implant surgery. J Prosthodont. 2022;31(2):155–64.PubMedCrossRef Pessoa R, Siqueira R, Li J, Saleh I, Meneghetti P, Bezerra F, et al. The impact of surgical guide fixation and implant location on accuracy of static computer-assisted implant surgery. J Prosthodont. 2022;31(2):155–64.PubMedCrossRef
22.
Zurück zum Zitat D’Haese R, Vrombaut T, Hommez G, De Bruyn H, Vandeweghe S. Accuracy of guided implant surgery using an intraoral scanner and desktop 3D-printed tooth-supported guides. Int J Oral Maxillofac Implants. 2022;37(3):479–84.PubMedCrossRef D’Haese R, Vrombaut T, Hommez G, De Bruyn H, Vandeweghe S. Accuracy of guided implant surgery using an intraoral scanner and desktop 3D-printed tooth-supported guides. Int J Oral Maxillofac Implants. 2022;37(3):479–84.PubMedCrossRef
23.
Zurück zum Zitat Sun Y, Ding Q, Tang L, Zhang L, Sun Y, Xie Q. Accuracy of a chairside fused deposition modeling 3D-printed single-tooth surgical template for implant placement: an in vitro comparison with a light cured template. J Craniomaxillofac Surg. 2019;47(8):1216–21.PubMedCrossRef Sun Y, Ding Q, Tang L, Zhang L, Sun Y, Xie Q. Accuracy of a chairside fused deposition modeling 3D-printed single-tooth surgical template for implant placement: an in vitro comparison with a light cured template. J Craniomaxillofac Surg. 2019;47(8):1216–21.PubMedCrossRef
24.
Zurück zum Zitat Li-Rodriguez JK, Diaz-Durany M, Romeo-Rubio M, Paz Salido M, Pradies G. Accuracy of a guided implant system with milled surgical templates. J Oral Sci. 2022;64(2):145–50.PubMedCrossRef Li-Rodriguez JK, Diaz-Durany M, Romeo-Rubio M, Paz Salido M, Pradies G. Accuracy of a guided implant system with milled surgical templates. J Oral Sci. 2022;64(2):145–50.PubMedCrossRef
25.
Zurück zum Zitat Baez-Marrero N, Rafel JL, Rodriguez-Cardenas YA, Aliaga-Del Castillo A, Dias-Da Silveira HL, Arriola-Guillen LE. Accuracy of computer-assisted surgery in immediate implant placement: an experimental study. J Indian Soc Periodontol. 2022;26(3):219–23.PubMedCrossRef Baez-Marrero N, Rafel JL, Rodriguez-Cardenas YA, Aliaga-Del Castillo A, Dias-Da Silveira HL, Arriola-Guillen LE. Accuracy of computer-assisted surgery in immediate implant placement: an experimental study. J Indian Soc Periodontol. 2022;26(3):219–23.PubMedCrossRef
26.
Zurück zum Zitat Orban K, Varga E Jr, Windisch P, Braunitzer G, Molnar B. Accuracy of half-guided implant placement with machine-driven or manual insertion: a prospective, randomized clinical study. Clin Oral Investig. 2022;26(1):1035–43.PubMedCrossRef Orban K, Varga E Jr, Windisch P, Braunitzer G, Molnar B. Accuracy of half-guided implant placement with machine-driven or manual insertion: a prospective, randomized clinical study. Clin Oral Investig. 2022;26(1):1035–43.PubMedCrossRef
27.
Zurück zum Zitat Gargallo-Albiol J, Zilleruelo-Pozo MJ, Lucas-Taule E, Munoz-Penalver J, Paternostro-Betancourt D, Hernandez-Alfaro F. Accuracy of static fully guided implant placement in the posterior area of partially edentulous jaws: a cohort prospective study. Clin Oral Investig. 2022;26(3):2783–91.PubMedCrossRef Gargallo-Albiol J, Zilleruelo-Pozo MJ, Lucas-Taule E, Munoz-Penalver J, Paternostro-Betancourt D, Hernandez-Alfaro F. Accuracy of static fully guided implant placement in the posterior area of partially edentulous jaws: a cohort prospective study. Clin Oral Investig. 2022;26(3):2783–91.PubMedCrossRef
28.
Zurück zum Zitat Singthong W, Serichetaphongse P, Chengprapakorn W. A randomized clinical trial on the accuracy of guided implant surgery between two implant-planning programs used by inexperienced operators. J Prosthet Dent. 2022. Singthong W, Serichetaphongse P, Chengprapakorn W. A randomized clinical trial on the accuracy of guided implant surgery between two implant-planning programs used by inexperienced operators. J Prosthet Dent. 2022.
29.
Zurück zum Zitat Feng Y, Su Z, Mo A, Yang X. Comparison of the accuracy of immediate implant placement using static and dynamic computer-assisted implant system in the esthetic zone of the maxilla: a prospective study. Int J Implant Dent. 2022;8(1):65.PubMedPubMedCentralCrossRef Feng Y, Su Z, Mo A, Yang X. Comparison of the accuracy of immediate implant placement using static and dynamic computer-assisted implant system in the esthetic zone of the maxilla: a prospective study. Int J Implant Dent. 2022;8(1):65.PubMedPubMedCentralCrossRef
30.
Zurück zum Zitat Lou F, Rao P, Zhang M, Luo S, Lu S, Xiao J. Accuracy evaluation of partially guided and fully guided templates applied to implant surgery of anterior teeth: a randomized controlled trial. Clin Implant Dent Relat Res. 2021;23(1):117–30.PubMedCrossRef Lou F, Rao P, Zhang M, Luo S, Lu S, Xiao J. Accuracy evaluation of partially guided and fully guided templates applied to implant surgery of anterior teeth: a randomized controlled trial. Clin Implant Dent Relat Res. 2021;23(1):117–30.PubMedCrossRef
31.
Zurück zum Zitat Schneider D, Sax C, Sancho-Puchades M, Hammerle CHF, Jung RE. Accuracy of computer-assisted, template-guided implant placement compared with conventional implant placement by hand—an in vitro study. Clin Oral Implants Res. 2021;32(9):1052–60.PubMedPubMedCentralCrossRef Schneider D, Sax C, Sancho-Puchades M, Hammerle CHF, Jung RE. Accuracy of computer-assisted, template-guided implant placement compared with conventional implant placement by hand—an in vitro study. Clin Oral Implants Res. 2021;32(9):1052–60.PubMedPubMedCentralCrossRef
32.
Zurück zum Zitat Song YW, Kim J, Kim JH, Park JM, Jung UW, Cha JK. Accuracy of dental implant placement by a novel in-house model-free and zero-setup fully guided surgical template made of a light-cured composite resin (VARO Guide((R))): a comparative in vitro study. Materials (Basel). 2021;14(14). Song YW, Kim J, Kim JH, Park JM, Jung UW, Cha JK. Accuracy of dental implant placement by a novel in-house model-free and zero-setup fully guided surgical template made of a light-cured composite resin (VARO Guide((R))): a comparative in vitro study. Materials (Basel). 2021;14(14).
33.
Zurück zum Zitat Abduo J, Lau D. Accuracy of static computer-assisted implant placement in long span edentulous area by novice implant clinicians: a cross-sectional in vitro study comparing fully-guided, pilot-guided, and freehand implant placement protocols. Clin Implant Dent Relat Res. 2021;23(3):361–72.PubMedCrossRef Abduo J, Lau D. Accuracy of static computer-assisted implant placement in long span edentulous area by novice implant clinicians: a cross-sectional in vitro study comparing fully-guided, pilot-guided, and freehand implant placement protocols. Clin Implant Dent Relat Res. 2021;23(3):361–72.PubMedCrossRef
34.
Zurück zum Zitat Herschdorfer L, Negreiros WM, Gallucci GO, Hamilton A. Comparison of the accuracy of implants placed with CAD-CAM surgical templates manufactured with various 3D printers: an in vitro study. J Prosthet Dent. 2021;125(6):905–10.PubMedCrossRef Herschdorfer L, Negreiros WM, Gallucci GO, Hamilton A. Comparison of the accuracy of implants placed with CAD-CAM surgical templates manufactured with various 3D printers: an in vitro study. J Prosthet Dent. 2021;125(6):905–10.PubMedCrossRef
35.
Zurück zum Zitat Ngamprasertkit C, Aunmeungthong W, Khongkhunthian P. The implant position accuracy between using only surgical drill guide and surgical drill guide with implant guide in fully digital workflow: a randomized clinical trial. Oral Maxillofac Surg. 2022;26(2):229–37.PubMedCrossRef Ngamprasertkit C, Aunmeungthong W, Khongkhunthian P. The implant position accuracy between using only surgical drill guide and surgical drill guide with implant guide in fully digital workflow: a randomized clinical trial. Oral Maxillofac Surg. 2022;26(2):229–37.PubMedCrossRef
36.
Zurück zum Zitat Spille J, Jin F, Behrens E, Acil Y, Lichtenstein J, Naujokat H, et al. Comparison of implant placement accuracy in two different preoperative digital workflows: navigated vs. pilot-drill-guided surgery. Int J Implant Dent. 2021;7(1):45.PubMedPubMedCentralCrossRef Spille J, Jin F, Behrens E, Acil Y, Lichtenstein J, Naujokat H, et al. Comparison of implant placement accuracy in two different preoperative digital workflows: navigated vs. pilot-drill-guided surgery. Int J Implant Dent. 2021;7(1):45.PubMedPubMedCentralCrossRef
37.
Zurück zum Zitat Guentsch A, Sukhtankar L, An H, Luepke PG. Precision and trueness of implant placement with and without static surgical guides: an in vitro study. J Prosthet Dent. 2021;126(3):398–404.PubMedCrossRef Guentsch A, Sukhtankar L, An H, Luepke PG. Precision and trueness of implant placement with and without static surgical guides: an in vitro study. J Prosthet Dent. 2021;126(3):398–404.PubMedCrossRef
38.
Zurück zum Zitat Han YT, Lin WC, Fan FY, Chen CL, Lin CC, Cheng HC. Comparison of dental surface image registration and fiducial marker registration: an in vivo accuracy study of static computer-assisted implant surgery. J Clin Med. 2021;10(18):4183.PubMedPubMedCentralCrossRef Han YT, Lin WC, Fan FY, Chen CL, Lin CC, Cheng HC. Comparison of dental surface image registration and fiducial marker registration: an in vivo accuracy study of static computer-assisted implant surgery. J Clin Med. 2021;10(18):4183.PubMedPubMedCentralCrossRef
39.
Zurück zum Zitat Sittikornpaiboon P, Arunjaroensuk S, Kaboosaya B, Subbalekha K, Mattheos N, Pimkhaokham A. Comparison of the accuracy of implant placement using different drilling systems for static computer-assisted implant surgery: a simulation-based experimental study. Clin Implant Dent Relat Res. 2021;23(4):635–43.PubMedCrossRef Sittikornpaiboon P, Arunjaroensuk S, Kaboosaya B, Subbalekha K, Mattheos N, Pimkhaokham A. Comparison of the accuracy of implant placement using different drilling systems for static computer-assisted implant surgery: a simulation-based experimental study. Clin Implant Dent Relat Res. 2021;23(4):635–43.PubMedCrossRef
40.
Zurück zum Zitat Huang L, Zhang X, Mo A. A retrospective study on the transferring accuracy of a fully guided digital template in the anterior zone. Materials (Basel). 2021;14(16):4631.PubMedPubMedCentralCrossRef Huang L, Zhang X, Mo A. A retrospective study on the transferring accuracy of a fully guided digital template in the anterior zone. Materials (Basel). 2021;14(16):4631.PubMedPubMedCentralCrossRef
41.
Zurück zum Zitat D’Haese R, Vrombaut T, Hommez G, De Bruyn H, Vandeweghe S. Accuracy of guided implant surgery in the edentulous jaw using desktop 3D-printed mucosal supported guides. J Clin Med. 2021;10(3):391.PubMedPubMedCentralCrossRef D’Haese R, Vrombaut T, Hommez G, De Bruyn H, Vandeweghe S. Accuracy of guided implant surgery in the edentulous jaw using desktop 3D-printed mucosal supported guides. J Clin Med. 2021;10(3):391.PubMedPubMedCentralCrossRef
42.
Zurück zum Zitat Lin CC, Ishikawa M, Maida T, Cheng HC, Ou KL, Nezu T, et al. Stereolithographic surgical guide with a combination of tooth and bone support: accuracy of guided implant surgery in distal extension situation. J Clin Med. 2020;9(3):709.PubMedPubMedCentralCrossRef Lin CC, Ishikawa M, Maida T, Cheng HC, Ou KL, Nezu T, et al. Stereolithographic surgical guide with a combination of tooth and bone support: accuracy of guided implant surgery in distal extension situation. J Clin Med. 2020;9(3):709.PubMedPubMedCentralCrossRef
43.
Zurück zum Zitat Chen Y, Zhang X, Wang M, Jiang Q, Mo A. Accuracy of full-guided and half-guided surgical templates in anterior immediate and delayed implantation: a retrospective study. Materials (Basel). 2020;14(1):26.PubMedPubMedCentralCrossRef Chen Y, Zhang X, Wang M, Jiang Q, Mo A. Accuracy of full-guided and half-guided surgical templates in anterior immediate and delayed implantation: a retrospective study. Materials (Basel). 2020;14(1):26.PubMedPubMedCentralCrossRef
44.
Zurück zum Zitat Wu D, Zhou L, Yang J, Zhang B, Lin Y, Chen J, et al. Accuracy of dynamic navigation compared to static surgical guide for dental implant placement. Int J Implant Dent. 2020;6(1):78.PubMedPubMedCentralCrossRef Wu D, Zhou L, Yang J, Zhang B, Lin Y, Chen J, et al. Accuracy of dynamic navigation compared to static surgical guide for dental implant placement. Int J Implant Dent. 2020;6(1):78.PubMedPubMedCentralCrossRef
45.
Zurück zum Zitat Cheng KJ, Kan TS, Liu YF, Zhu WD, Zhu FD, Wang WB, et al. Accuracy of dental implant surgery with robotic position feedback and registration algorithm: an in-vitro study. Comput Biol Med. 2021;129: 104153.PubMedCrossRef Cheng KJ, Kan TS, Liu YF, Zhu WD, Zhu FD, Wang WB, et al. Accuracy of dental implant surgery with robotic position feedback and registration algorithm: an in-vitro study. Comput Biol Med. 2021;129: 104153.PubMedCrossRef
46.
Zurück zum Zitat Henprasert P, Dawson DV, El-Kerdani T, Song X, Couso-Queiruga E, Holloway JA. Comparison of the accuracy of implant position using surgical guides fabricated by additive and subtractive techniques. J Prosthodont. 2020;29(6):534–41.PubMedCrossRef Henprasert P, Dawson DV, El-Kerdani T, Song X, Couso-Queiruga E, Holloway JA. Comparison of the accuracy of implant position using surgical guides fabricated by additive and subtractive techniques. J Prosthodont. 2020;29(6):534–41.PubMedCrossRef
47.
Zurück zum Zitat Kniha K, Brandt M, Bock A, Modabber A, Prescher A, Holzle F, et al. Accuracy of fully guided orthodontic mini-implant placement evaluated by cone-beam computed tomography: a study involving human cadaver heads. Clin Oral Investig. 2021;25(3):1299–306.PubMedCrossRef Kniha K, Brandt M, Bock A, Modabber A, Prescher A, Holzle F, et al. Accuracy of fully guided orthodontic mini-implant placement evaluated by cone-beam computed tomography: a study involving human cadaver heads. Clin Oral Investig. 2021;25(3):1299–306.PubMedCrossRef
48.
Zurück zum Zitat Vinci R, Manacorda M, Abundo R, Lucchina AG, Scarano A, Crocetta C, et al. Accuracy of edentulous computer-aided implant surgery as compared to virtual planning: a retrospective multicenter study. J Clin Med. 2020;9(3):774.PubMedPubMedCentralCrossRef Vinci R, Manacorda M, Abundo R, Lucchina AG, Scarano A, Crocetta C, et al. Accuracy of edentulous computer-aided implant surgery as compared to virtual planning: a retrospective multicenter study. J Clin Med. 2020;9(3):774.PubMedPubMedCentralCrossRef
49.
Zurück zum Zitat Suksod N, Kunavisarut C, Kitisubkanchana J. Accuracy of computer-guided implantation in the placement of one-piece ceramic dental implants in the anterior region: a prospective clinical study. PLoS ONE. 2020;15(9): e0237229.PubMedPubMedCentralCrossRef Suksod N, Kunavisarut C, Kitisubkanchana J. Accuracy of computer-guided implantation in the placement of one-piece ceramic dental implants in the anterior region: a prospective clinical study. PLoS ONE. 2020;15(9): e0237229.PubMedPubMedCentralCrossRef
50.
Zurück zum Zitat Kivovics M, Penzes D, Nemeth O, Mijiritsky E. The influence of surgical experience and bone density on the accuracy of static computer-assisted implant surgery in edentulous jaws using a mucosa-supported surgical template with a half-guided implant placement protocol—a randomized clinical study. Materials (Basel). 2020;13(24):5759.PubMedPubMedCentralCrossRef Kivovics M, Penzes D, Nemeth O, Mijiritsky E. The influence of surgical experience and bone density on the accuracy of static computer-assisted implant surgery in edentulous jaws using a mucosa-supported surgical template with a half-guided implant placement protocol—a randomized clinical study. Materials (Basel). 2020;13(24):5759.PubMedPubMedCentralCrossRef
51.
Zurück zum Zitat Smitkarn P, Subbalekha K, Mattheos N, Pimkhaokham A. The accuracy of single-tooth implants placed using fully digital-guided surgery and freehand implant surgery. J Clin Periodontol. 2019;46(9):949–57.PubMedCrossRef Smitkarn P, Subbalekha K, Mattheos N, Pimkhaokham A. The accuracy of single-tooth implants placed using fully digital-guided surgery and freehand implant surgery. J Clin Periodontol. 2019;46(9):949–57.PubMedCrossRef
52.
Zurück zum Zitat El Kholy K, Lazarin R, Janner SFM, Faerber K, Buser R, Buser D. Influence of surgical guide support and implant site location on accuracy of static computer-assisted implant surgery. Clin Oral Implants Res. 2019;30(11):1067–75.PubMedCrossRef El Kholy K, Lazarin R, Janner SFM, Faerber K, Buser R, Buser D. Influence of surgical guide support and implant site location on accuracy of static computer-assisted implant surgery. Clin Oral Implants Res. 2019;30(11):1067–75.PubMedCrossRef
53.
Zurück zum Zitat Skjerven H, Riis UH, Herlofsson BB, Ellingsen JE. In vivo accuracy of implant placement using a full digital planning modality and stereolithographic guides. Int J Oral Maxillofac Implants. 2019;34(1):124–32.PubMedCrossRef Skjerven H, Riis UH, Herlofsson BB, Ellingsen JE. In vivo accuracy of implant placement using a full digital planning modality and stereolithographic guides. Int J Oral Maxillofac Implants. 2019;34(1):124–32.PubMedCrossRef
54.
Zurück zum Zitat Chang RJ, Chen HL, Huang LG, Wong YK. Accuracy of implant placement with a computer-aided fabricated surgical template with guided parallel pins: a pilot study. J Chin Med Assoc. 2018;81(11):970–6.PubMedCrossRef Chang RJ, Chen HL, Huang LG, Wong YK. Accuracy of implant placement with a computer-aided fabricated surgical template with guided parallel pins: a pilot study. J Chin Med Assoc. 2018;81(11):970–6.PubMedCrossRef
55.
Zurück zum Zitat Chen Z, Li J, Sinjab K, Mendonca G, Yu H, Wang HL. Accuracy of flapless immediate implant placement in anterior maxilla using computer-assisted versus freehand surgery: a cadaver study. Clin Oral Implants Res. 2018;29(12):1186–94.PubMedCrossRef Chen Z, Li J, Sinjab K, Mendonca G, Yu H, Wang HL. Accuracy of flapless immediate implant placement in anterior maxilla using computer-assisted versus freehand surgery: a cadaver study. Clin Oral Implants Res. 2018;29(12):1186–94.PubMedCrossRef
56.
Zurück zum Zitat Brandt J, Brenner M, Lauer HC, Brandt S. Accuracy of a template-guided implant surgery system with a CAD/CAM-based measurement method: an in vitro study. Int J Oral Maxillofac Implants. 2018;33(2):328–34.PubMedCrossRef Brandt J, Brenner M, Lauer HC, Brandt S. Accuracy of a template-guided implant surgery system with a CAD/CAM-based measurement method: an in vitro study. Int J Oral Maxillofac Implants. 2018;33(2):328–34.PubMedCrossRef
57.
Zurück zum Zitat Kauffmann P, Rau A, Engelke W, Troeltzsch M, Brockmeyer P, Dagmar LS, et al. Accuracy of navigation-guided dental implant placement with screw versus hand template fixation in the edentulous mandible. Int J Oral Maxillofac Implants. 2018;33(2):383–8.PubMedCrossRef Kauffmann P, Rau A, Engelke W, Troeltzsch M, Brockmeyer P, Dagmar LS, et al. Accuracy of navigation-guided dental implant placement with screw versus hand template fixation in the edentulous mandible. Int J Oral Maxillofac Implants. 2018;33(2):383–8.PubMedCrossRef
58.
Zurück zum Zitat Fang Y, An X, Jeong SM, Choi BH. Accuracy of computer-guided implant placement in anterior regions. J Prosthet Dent. 2019;121(5):836–42.PubMedCrossRef Fang Y, An X, Jeong SM, Choi BH. Accuracy of computer-guided implant placement in anterior regions. J Prosthet Dent. 2019;121(5):836–42.PubMedCrossRef
59.
Zurück zum Zitat Ma B, Park T, Chun I, Yun K. The accuracy of a 3D printing surgical guide determined by CBCT and model analysis. J Adv Prosthodont. 2018;10(4):279–85.PubMedPubMedCentralCrossRef Ma B, Park T, Chun I, Yun K. The accuracy of a 3D printing surgical guide determined by CBCT and model analysis. J Adv Prosthodont. 2018;10(4):279–85.PubMedPubMedCentralCrossRef
60.
Zurück zum Zitat Gjelvold B, Mahmood DJH, Wennerberg A. Accuracy of surgical guides from 2 different desktop 3D printers for computed tomography-guided surgery. J Prosthet Dent. 2019;121(3):498–503.PubMedCrossRef Gjelvold B, Mahmood DJH, Wennerberg A. Accuracy of surgical guides from 2 different desktop 3D printers for computed tomography-guided surgery. J Prosthet Dent. 2019;121(3):498–503.PubMedCrossRef
61.
Zurück zum Zitat Tang W, Liu Q, Zeng X, Yu J, Shu D, Shen G, et al. Accuracy of half-way mucosa-supported implant guides for edentulous jaws: a retrospective study with a median follow-up of 2 years. J Int Med Res. 2021;49(3):300060521999739.PubMedCrossRef Tang W, Liu Q, Zeng X, Yu J, Shu D, Shen G, et al. Accuracy of half-way mucosa-supported implant guides for edentulous jaws: a retrospective study with a median follow-up of 2 years. J Int Med Res. 2021;49(3):300060521999739.PubMedCrossRef
62.
Zurück zum Zitat Siqueira R, Chen Z, Galli M, Saleh I, Wang HL, Chan HL. Does a fully digital workflow improve the accuracy of computer-assisted implant surgery in partially edentulous patients? A systematic review of clinical trials. Clin Implant Dent Relat Res. 2020;22(6):660–71.PubMedCrossRef Siqueira R, Chen Z, Galli M, Saleh I, Wang HL, Chan HL. Does a fully digital workflow improve the accuracy of computer-assisted implant surgery in partially edentulous patients? A systematic review of clinical trials. Clin Implant Dent Relat Res. 2020;22(6):660–71.PubMedCrossRef
63.
Zurück zum Zitat Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M. Accuracy of computer-aided implant placement. Clin Oral Implants Res. 2012;23(Suppl 6):112–23.PubMedCrossRef Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M. Accuracy of computer-aided implant placement. Clin Oral Implants Res. 2012;23(Suppl 6):112–23.PubMedCrossRef
64.
Zurück zum Zitat Zhou W, Liu Z, Song L, Kuo CL, Shafer DM. Clinical factors affecting the accuracy of guided implant surgery—a systematic review and meta-analysis. J Evid Based Dent Pract. 2018;18(1):28–40.PubMedCrossRef Zhou W, Liu Z, Song L, Kuo CL, Shafer DM. Clinical factors affecting the accuracy of guided implant surgery—a systematic review and meta-analysis. J Evid Based Dent Pract. 2018;18(1):28–40.PubMedCrossRef
65.
Zurück zum Zitat Ozan O, Turkyilmaz I, Yilmaz B. A preliminary report of patients treated with early loaded implants using computerized tomography-guided surgical stents: flapless versus conventional flapped surgery. J Oral Rehabil. 2007;34(11):835–40.PubMedCrossRef Ozan O, Turkyilmaz I, Yilmaz B. A preliminary report of patients treated with early loaded implants using computerized tomography-guided surgical stents: flapless versus conventional flapped surgery. J Oral Rehabil. 2007;34(11):835–40.PubMedCrossRef
66.
Zurück zum Zitat Ersoy AE, Turkyilmaz I, Ozan O, McGlumphy EA. Reliability of implant placement with stereolithographic surgical guides generated from computed tomography: clinical data from 94 implants. J Periodontol. 2008;79(8):1339–45.PubMedCrossRef Ersoy AE, Turkyilmaz I, Ozan O, McGlumphy EA. Reliability of implant placement with stereolithographic surgical guides generated from computed tomography: clinical data from 94 implants. J Periodontol. 2008;79(8):1339–45.PubMedCrossRef
67.
Zurück zum Zitat Cunha RM, Souza FA, Hadad H, Poli PP, Maiorana C, Carvalho PSP. Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides. J Prosthet Dent. 2021;125(2):266–72.PubMedCrossRef Cunha RM, Souza FA, Hadad H, Poli PP, Maiorana C, Carvalho PSP. Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides. J Prosthet Dent. 2021;125(2):266–72.PubMedCrossRef
68.
Zurück zum Zitat Cassetta M, Di Mambro A, Giansanti M, Stefanelli LV, Cavallini C. The intrinsic error of a stereolithographic surgical template in implant guided surgery. Int J Oral Maxillofac Surg. 2013;42(2):264–75.PubMedCrossRef Cassetta M, Di Mambro A, Giansanti M, Stefanelli LV, Cavallini C. The intrinsic error of a stereolithographic surgical template in implant guided surgery. Int J Oral Maxillofac Surg. 2013;42(2):264–75.PubMedCrossRef
69.
Zurück zum Zitat Cassetta M, Giansanti M, Di Mambro A, Stefanelli LV. Accuracy of positioning of implants inserted using a mucosa-supported stereolithographic surgical guide in the edentulous maxilla and mandible. Int J Oral Maxillofac Implants. 2014;29(5):1071–8.PubMedCrossRef Cassetta M, Giansanti M, Di Mambro A, Stefanelli LV. Accuracy of positioning of implants inserted using a mucosa-supported stereolithographic surgical guide in the edentulous maxilla and mandible. Int J Oral Maxillofac Implants. 2014;29(5):1071–8.PubMedCrossRef
70.
Zurück zum Zitat Di Giacomo GA, da Silva JV, da Silva AM, Paschoal GH, Cury PR, Szarf G. Accuracy and complications of computer-designed selective laser sintering surgical guides for flapless dental implant placement and immediate definitive prosthesis installation. J Periodontol. 2012;83(4):410–9.PubMedCrossRef Di Giacomo GA, da Silva JV, da Silva AM, Paschoal GH, Cury PR, Szarf G. Accuracy and complications of computer-designed selective laser sintering surgical guides for flapless dental implant placement and immediate definitive prosthesis installation. J Periodontol. 2012;83(4):410–9.PubMedCrossRef
71.
Zurück zum Zitat D’Haese J, Van De Velde T, Elaut L, De Bruyn H. A prospective study on the accuracy of mucosally supported stereolithographic surgical guides in fully edentulous maxillae. Clin Implant Dent Relat Res. 2012;14(2):293–303.PubMedCrossRef D’Haese J, Van De Velde T, Elaut L, De Bruyn H. A prospective study on the accuracy of mucosally supported stereolithographic surgical guides in fully edentulous maxillae. Clin Implant Dent Relat Res. 2012;14(2):293–303.PubMedCrossRef
72.
Zurück zum Zitat Kuhl S, Payer M, Zitzmann NU, Lambrecht JT, Filippi A. Technical accuracy of printed surgical templates for guided implant surgery with the coDiagnostiX software. Clin Implant Dent Relat Res. 2015;17(Suppl 1):e177–82.PubMed Kuhl S, Payer M, Zitzmann NU, Lambrecht JT, Filippi A. Technical accuracy of printed surgical templates for guided implant surgery with the coDiagnostiX software. Clin Implant Dent Relat Res. 2015;17(Suppl 1):e177–82.PubMed
73.
Zurück zum Zitat Suriyan N, Sarinnaphakorn L, Deeb GR, Bencharit S. Trephination-based, guided surgical implant placement: a clinical study. J Prosthet Dent. 2019;121(3):411–6.PubMedCrossRef Suriyan N, Sarinnaphakorn L, Deeb GR, Bencharit S. Trephination-based, guided surgical implant placement: a clinical study. J Prosthet Dent. 2019;121(3):411–6.PubMedCrossRef
74.
Zurück zum Zitat Oh KC, Shim JS, Park JM. In vitro comparison between metal sleeve-free and metal sleeve-incorporated 3D-printed computer-assisted implant surgical guides. Materials (Basel). 2021;14(3):615.PubMedPubMedCentralCrossRef Oh KC, Shim JS, Park JM. In vitro comparison between metal sleeve-free and metal sleeve-incorporated 3D-printed computer-assisted implant surgical guides. Materials (Basel). 2021;14(3):615.PubMedPubMedCentralCrossRef
75.
Zurück zum Zitat Lee DH, An SY, Hong MH, Jeon KB, Lee KB. Accuracy of a direct drill-guiding system with minimal tolerance of surgical instruments used for implant surgery: a prospective clinical study. J Adv Prosthodont. 2016;8(3):207–13.PubMedPubMedCentralCrossRef Lee DH, An SY, Hong MH, Jeon KB, Lee KB. Accuracy of a direct drill-guiding system with minimal tolerance of surgical instruments used for implant surgery: a prospective clinical study. J Adv Prosthodont. 2016;8(3):207–13.PubMedPubMedCentralCrossRef
76.
Zurück zum Zitat El Kholy K, Janner SFM, Schimmel M, Buser D. The influence of guided sleeve height, drilling distance, and drilling key length on the accuracy of static computer-assisted implant surgery. Clin Implant Dent Relat Res. 2019;21(1):101–7.PubMedCrossRef El Kholy K, Janner SFM, Schimmel M, Buser D. The influence of guided sleeve height, drilling distance, and drilling key length on the accuracy of static computer-assisted implant surgery. Clin Implant Dent Relat Res. 2019;21(1):101–7.PubMedCrossRef
77.
Zurück zum Zitat Schnutenhaus S, Edelmann C, Rudolph H. Does the macro design of an implant affect the accuracy of template-guided implantation? A prospective clinical study. Int J Implant Dent. 2021;7(1):42.PubMedPubMedCentralCrossRef Schnutenhaus S, Edelmann C, Rudolph H. Does the macro design of an implant affect the accuracy of template-guided implantation? A prospective clinical study. Int J Implant Dent. 2021;7(1):42.PubMedPubMedCentralCrossRef
78.
Zurück zum Zitat Guentsch A, An H, Dentino AR. Precision and trueness of computer-assisted implant placement using static surgical guides with open and closed sleeves: an in vitro analysis. Clin Oral Implants Res. 2022;33(4):441–50.PubMedPubMedCentralCrossRef Guentsch A, An H, Dentino AR. Precision and trueness of computer-assisted implant placement using static surgical guides with open and closed sleeves: an in vitro analysis. Clin Oral Implants Res. 2022;33(4):441–50.PubMedPubMedCentralCrossRef
79.
Zurück zum Zitat Tallarico M, Czajkowska M, Cicciu M, Giardina F, Minciarelli A, Zadrozny L, et al. Accuracy of surgical templates with and without metallic sleeves in case of partial arch restorations: a systematic review. J Dent. 2021;115: 103852.PubMedCrossRef Tallarico M, Czajkowska M, Cicciu M, Giardina F, Minciarelli A, Zadrozny L, et al. Accuracy of surgical templates with and without metallic sleeves in case of partial arch restorations: a systematic review. J Dent. 2021;115: 103852.PubMedCrossRef
80.
Zurück zum Zitat Abduo J, Lau D. Effect of manufacturing technique on the accuracy of surgical guides for static computer-aided implant surgery. Int J Oral Maxillofac Implants. 2020;35(5):931–8.PubMedCrossRef Abduo J, Lau D. Effect of manufacturing technique on the accuracy of surgical guides for static computer-aided implant surgery. Int J Oral Maxillofac Implants. 2020;35(5):931–8.PubMedCrossRef
81.
Zurück zum Zitat Mukai S, Mukai E, Santos-Junior JA, Shibli JA, Faveri M, Giro G. Assessment of the reproducibility and precision of milling and 3D printing surgical guides. BMC Oral Health. 2021;21(1):1.PubMedPubMedCentralCrossRef Mukai S, Mukai E, Santos-Junior JA, Shibli JA, Faveri M, Giro G. Assessment of the reproducibility and precision of milling and 3D printing surgical guides. BMC Oral Health. 2021;21(1):1.PubMedPubMedCentralCrossRef
Metadaten
Titel
A systematic review of the accuracy of digital surgical guides for dental implantation
verfasst von
Yiting Shi
JunKai Wang
Chao Ma
Jiayi Shen
Xian Dong
Dan Lin
Publikationsdatum
01.12.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Implant Dentistry / Ausgabe 1/2023
Elektronische ISSN: 2198-4034
DOI
https://doi.org/10.1186/s40729-023-00507-w

Weitere Artikel der Ausgabe 1/2023

International Journal of Implant Dentistry 1/2023 Zur Ausgabe

Neu im Fachgebiet Zahnmedizin

Sechs Monate E-Rezept – Erfolgsgeschichte mit Schattenseiten

IT für Ärzte Nachrichten

244 Millionen E-Rezepte sind in sechs Monaten eingelöst worden: Nach einem halben Jahr zieht die Betriebsgesellschaft gematik eine positive Zwischenbilanz. Doch noch sind nicht alle Akteure begeistert.

So können sich Praxen vor Cyberkriminalität schützen

IT für Ärzte Nachrichten

Immer wieder werden Arztpraxen Opfer von Netzverbrechern. Der typische Hacker war einmal, Cyberkriminalität ist sehr vielfältig, berichtet Lars Korunalp. Der IT-Berater gibt Tipps und nennt konkrete Maßnahmen, wie sich Praxen vor Zugriffen aus dem Internet schützen können.

Thüringer Landtag beschließt Vorab-Quote im Medizinstudium

Bewerber, die sich zu einer langen Tätigkeit als Haus- oder Zahnarzt in Thüringen verpflichten, bekommen in Jena künftig leichter einen Studienplatz. Sechs Prozent der Plätze sollen so vergeben werden.

Notfallreform: Lauterbach nimmt KVen und ausgewählte Kliniken in die Pflicht

06.06.2024 Klinik aktuell Nachrichten

Die Ampelkoalition nimmt einen neuen Anlauf für die Reform der Notfallversorgung. Der Gesetzentwurf zeigt: Die Vertragsärzte müssen sich auf erhebliche Veränderungen in der Organisation der Notdienste einstellen.

Update Zahnmedizin

Bestellen Sie unseren kostenlosen Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.