08.06.2016 | Original Paper

Accuracy of measuring acetabular cup position after total hip arthroplasty: comparison between a radiographic planning software and three-dimensional computed tomography

verfasst von: Vahdettin Bayraktar, Markus Weber, Frederik von Kunow, Florian Zeman, Benjamin Craiovan, Tobias Renkawitz, Joachim Grifka, Michael Woerner

Erschienen in: International Orthopaedics | Ausgabe 4/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Various methods are available for measuring acetabular cup position after total hip arthroplasty (THA) on standard anterior-posterior (AP) radiographs. We compared the accuracy of a commercial radiographic planning software program with that of three-dimensional computed tomography (3D-CT) scans.

Methods

We obtained plain AP radiographs and 3D-CTs from 65 patients after THA. In addition to calculating cup anteversion and inclination with 3D-CT, we determined the cup position using the radiographic planning software program mediCAD® 2.5 (Hectec, Niederviehbach, Germany). Furthermore, we compared the measurements using the inter-teardrop and bi-ischial lines as pelvic landmarks.

Results

The mean difference in anteversion between 3D-CT and mediCAD® software was 0.1° using the inter-teardrop line (standard deviation [SD], 8.8°; range, −21° to 23°; p = 0.97) and 0.4° using the bi-ischial line (SD, 8.8°; range, −23° to 21°; p = 0.72). Inclination showed a mean difference of 0.6° using the inter-teardrop line (SD, 4.4°; range, −9° to 21°; p = 0.24) and 0.5° using bi-ischial line (SD, 4.6°; range, −9° to 22°; p = 0.35). The means for absolute differences were 7.2° for anteversion and 3.1° for inclination. With regard to using the bi-ischial or inter-teardrop line, no significant difference was found between the two pelvic landmarks. The intra-class correlation coefficient (ICC) was analysed for anteversion and inclination using either the inter-teardrop line or the bi-ischial line as radiographic baseline.

Conclusions

A radiographic planning software program (mediCAD®) is a helpful tool for measuring cup inclination on AP radiographs. With respect to anteversion, measurements are rather susceptible to mistakes with mean inaccuracies of over 7°. Thus, 3D-CT remains the “gold standard” if a lower tolerance limit (±3°) is required for more complex biomechanical evaluations. As a pelvic landmark, the interteardrop line is preferential to the bi-ischial line because of its lower impact on the position of the pelvis.
Literatur
1.
Zurück zum Zitat Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220CrossRefPubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220CrossRefPubMed
2.
Zurück zum Zitat Hube R, Dienst M, von Roth P (2014) Complications after minimally invasive total hip arthroplasty. Orthopade 43(1):47–53CrossRefPubMed Hube R, Dienst M, von Roth P (2014) Complications after minimally invasive total hip arthroplasty. Orthopade 43(1):47–53CrossRefPubMed
3.
Zurück zum Zitat Widmer KH (2007) Containment versus impingement: finding a compromise for cup placement in total hip arthroplasty. Int Orthop 31(Suppl 1):S29–S33CrossRefPubMed Widmer KH (2007) Containment versus impingement: finding a compromise for cup placement in total hip arthroplasty. Int Orthop 31(Suppl 1):S29–S33CrossRefPubMed
4.
Zurück zum Zitat Preininger B, Haschke F, Perka C (2014) Diagnostics and therapy of luxation after total hip arthroplasty. Orthopade 43(1):54–63CrossRefPubMed Preininger B, Haschke F, Perka C (2014) Diagnostics and therapy of luxation after total hip arthroplasty. Orthopade 43(1):54–63CrossRefPubMed
5.
Zurück zum Zitat Wan Z, Boutary M, Dorr LD (2008) The influence of acetabular component position on wear in total hip arthroplasty. J Arthroplasty 23(1):51–56CrossRefPubMed Wan Z, Boutary M, Dorr LD (2008) The influence of acetabular component position on wear in total hip arthroplasty. J Arthroplasty 23(1):51–56CrossRefPubMed
6.
Zurück zum Zitat Harrison CL, Thomson AI, Cutts S, Rowe PJ, Riches PE (2014) Research synthesis of recommended acetabular cup orientations for total hip arthroplasty. J Arthroplast 29(2):377–382CrossRef Harrison CL, Thomson AI, Cutts S, Rowe PJ, Riches PE (2014) Research synthesis of recommended acetabular cup orientations for total hip arthroplasty. J Arthroplast 29(2):377–382CrossRef
7.
Zurück zum Zitat Ghelman B, Kepler CK, Lyman S, Della Valle AG (2009) CT outperforms radiography for determination of acetabular cup version after THA. Clin Orthop Relat Res 467(9):2362–2370CrossRefPubMedPubMedCentral Ghelman B, Kepler CK, Lyman S, Della Valle AG (2009) CT outperforms radiography for determination of acetabular cup version after THA. Clin Orthop Relat Res 467(9):2362–2370CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M (2008) Validation of a computer navigation system and a CT method for determination of the orientation of implanted acetabular cup in total hip arthroplasty: a cadaver study. Clin Biomech 23(8):1004–1011CrossRef Lin F, Lim D, Wixson RL, Milos S, Hendrix RW, Makhsous M (2008) Validation of a computer navigation system and a CT method for determination of the orientation of implanted acetabular cup in total hip arthroplasty: a cadaver study. Clin Biomech 23(8):1004–1011CrossRef
9.
Zurück zum Zitat Dargel J, Oppermann J, Brüggemann GP, Eysel P (2014) Dislocation following total hip replacement. Dtsch Arztebl Int 111(51–52):884–890PubMedPubMedCentral Dargel J, Oppermann J, Brüggemann GP, Eysel P (2014) Dislocation following total hip replacement. Dtsch Arztebl Int 111(51–52):884–890PubMedPubMedCentral
10.
Zurück zum Zitat Craiovan B, Renkawitz T, Weber M, Grifka J, Nolte L, Zheng G (2014) Is the acetabular cup orientation after total hip arthroplasty on a two dimension or three dimension model accurate? Int Orthop 38:2009–2015CrossRefPubMed Craiovan B, Renkawitz T, Weber M, Grifka J, Nolte L, Zheng G (2014) Is the acetabular cup orientation after total hip arthroplasty on a two dimension or three dimension model accurate? Int Orthop 38:2009–2015CrossRefPubMed
11.
Zurück zum Zitat Kim YH, Cho KH, Park YG (2015) Is the acetabular cup orientation after total hip arthroplasty on a two-dimensional or three-dimensional model accurate? Int Orthop 39(4):819–820CrossRefPubMed Kim YH, Cho KH, Park YG (2015) Is the acetabular cup orientation after total hip arthroplasty on a two-dimensional or three-dimensional model accurate? Int Orthop 39(4):819–820CrossRefPubMed
12.
Zurück zum Zitat Tsai TY, Dimitriou D, Li G, Kwon YM (2014) Does total hip arthroplasty restore native hip anatomy? Three-dimensional reconstruction analysis. Int Orthop 38(8):1577–1583CrossRefPubMedPubMedCentral Tsai TY, Dimitriou D, Li G, Kwon YM (2014) Does total hip arthroplasty restore native hip anatomy? Three-dimensional reconstruction analysis. Int Orthop 38(8):1577–1583CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Widmer KH (2004) A simplified method to determine acetabular cup anteversion from plain radiographs. J Arthroplasty 19(3):387–390CrossRefPubMed Widmer KH (2004) A simplified method to determine acetabular cup anteversion from plain radiographs. J Arthroplasty 19(3):387–390CrossRefPubMed
14.
Zurück zum Zitat Murray DW (1993) The definition and measurement of acetabular orientation. J Bone Joint Surg (Br) 75(2):228–232CrossRef Murray DW (1993) The definition and measurement of acetabular orientation. J Bone Joint Surg (Br) 75(2):228–232CrossRef
15.
Zurück zum Zitat Lembeck B, Mueller O, Reize P, Wuelker N (2005) Pelvic tilt makes acetabular cup navigation inaccurate. Acta Orthop 76(4):517–523CrossRefPubMed Lembeck B, Mueller O, Reize P, Wuelker N (2005) Pelvic tilt makes acetabular cup navigation inaccurate. Acta Orthop 76(4):517–523CrossRefPubMed
16.
Zurück zum Zitat DiGioia AM, Hafez MA, Jaramaz B, Levison TJ, Moody JE (2006) Functional pelvic orientation measured from lateral standing and sitting radiographs. Clin Orthop Relat Res 453:272–276CrossRefPubMed DiGioia AM, Hafez MA, Jaramaz B, Levison TJ, Moody JE (2006) Functional pelvic orientation measured from lateral standing and sitting radiographs. Clin Orthop Relat Res 453:272–276CrossRefPubMed
17.
Zurück zum Zitat Renkawitz T, Haimerl M, Dohmen L, Gneiting S, Wegner M, Ehret N, Buchele C, Schubert M, Lechler P, Woerner M et al (2011) Minimally invasive computer-navigated total hip arthroplasty, following the concept of femur first and combined anteversion: design of a blinded randomized controlled trial. BMC Musculoskelet Disord 12:192CrossRefPubMedPubMedCentral Renkawitz T, Haimerl M, Dohmen L, Gneiting S, Wegner M, Ehret N, Buchele C, Schubert M, Lechler P, Woerner M et al (2011) Minimally invasive computer-navigated total hip arthroplasty, following the concept of femur first and combined anteversion: design of a blinded randomized controlled trial. BMC Musculoskelet Disord 12:192CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33(1):159–174CrossRefPubMed Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33(1):159–174CrossRefPubMed
19.
Zurück zum Zitat Echeverri S, Leyvraz PF, Zambelli PY, Jolles BM (2006) Reliable acetabular cup orientation with a new gravity-assisted guidance system. J Arthroplasty 21(3):413–419CrossRefPubMed Echeverri S, Leyvraz PF, Zambelli PY, Jolles BM (2006) Reliable acetabular cup orientation with a new gravity-assisted guidance system. J Arthroplasty 21(3):413–419CrossRefPubMed
20.
Zurück zum Zitat Bosker BH, Verheyen CC, Horstmann WG, Tulp NJ (2007) Poor accuracy of freehand cup positioning during total hip arthroplasty. Arch Orthop Trauma Surg 127(5):375–379CrossRefPubMedPubMedCentral Bosker BH, Verheyen CC, Horstmann WG, Tulp NJ (2007) Poor accuracy of freehand cup positioning during total hip arthroplasty. Arch Orthop Trauma Surg 127(5):375–379CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Varghese B, Muthukumar N, Balasubramaniam M, Scally A (2011) Reliability of measurements with digital radiographs—a myth. Acta Orthop Belg 77:622–625PubMed Varghese B, Muthukumar N, Balasubramaniam M, Scally A (2011) Reliability of measurements with digital radiographs—a myth. Acta Orthop Belg 77:622–625PubMed
22.
Zurück zum Zitat Babisch JW, Layher F, Amiot LP (2008) The rationale for tilt-adjusted acetabular cup navigation. J Bone Joint Surg Am 90:357–365CrossRefPubMed Babisch JW, Layher F, Amiot LP (2008) The rationale for tilt-adjusted acetabular cup navigation. J Bone Joint Surg Am 90:357–365CrossRefPubMed
23.
Zurück zum Zitat Goergen TG, Resnick D (1975) Evaluation of acetabular anteversion following total hip arthroplasty: necessity of proper centering. Br J Radiol 48:259–260CrossRefPubMed Goergen TG, Resnick D (1975) Evaluation of acetabular anteversion following total hip arthroplasty: necessity of proper centering. Br J Radiol 48:259–260CrossRefPubMed
24.
Zurück zum Zitat Woolson ST, Hartford JM, Sawyer A (1999) Results of a method of leg-length equalization for patients undergoing primary total hip replacement. J Arthroplasty 14(2):159–164CrossRefPubMed Woolson ST, Hartford JM, Sawyer A (1999) Results of a method of leg-length equalization for patients undergoing primary total hip replacement. J Arthroplasty 14(2):159–164CrossRefPubMed
25.
Zurück zum Zitat Lu M, Zhou YX, Du H, Zhang J, Liu J (2013) Reliability and validity of measuring of acetabular component orientation by plain anteroposterior radiographs. Clin Orthop Relat Res 471(9):2987–2994CrossRefPubMedPubMedCentral Lu M, Zhou YX, Du H, Zhang J, Liu J (2013) Reliability and validity of measuring of acetabular component orientation by plain anteroposterior radiographs. Clin Orthop Relat Res 471(9):2987–2994CrossRefPubMedPubMedCentral
26.
Zurück zum Zitat Conn KS, Clarke MT, Hallett JP (2002) A simple guide to determine the magnification of radiographs and to improve the accuracy of preoperative templating. J Bone Joint Surg (Br) 84-B:269–272CrossRef Conn KS, Clarke MT, Hallett JP (2002) A simple guide to determine the magnification of radiographs and to improve the accuracy of preoperative templating. J Bone Joint Surg (Br) 84-B:269–272CrossRef
27.
Zurück zum Zitat Westacott DJ, McArthur J, King RJ, Foguet P (2013) Assessment of cup orientation in hip resurfacing: a comparison of TraumaCad and computed tomography. J Orthop Surg Res 8:8CrossRefPubMedPubMedCentral Westacott DJ, McArthur J, King RJ, Foguet P (2013) Assessment of cup orientation in hip resurfacing: a comparison of TraumaCad and computed tomography. J Orthop Surg Res 8:8CrossRefPubMedPubMedCentral
28.
Zurück zum Zitat Malik A, Wan Z, Jaramaz B, Bowman G, Dorr LD (2010) A validation model for measurement of acetabular component position. J Arthroplasty 25:812–819CrossRefPubMed Malik A, Wan Z, Jaramaz B, Bowman G, Dorr LD (2010) A validation model for measurement of acetabular component position. J Arthroplasty 25:812–819CrossRefPubMed
29.
Zurück zum Zitat Lazennec JY, Rousseau MA, Rangel A, Gorin M, Belicourt C, Brusson A, Catonne Y (2011) Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: measurements reproducibility with EOS imaging system versus conventional radiographies. Orthop Traumatol Surg Res 97(4):373–380CrossRefPubMed Lazennec JY, Rousseau MA, Rangel A, Gorin M, Belicourt C, Brusson A, Catonne Y (2011) Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: measurements reproducibility with EOS imaging system versus conventional radiographies. Orthop Traumatol Surg Res 97(4):373–380CrossRefPubMed
Metadaten
Titel
Accuracy of measuring acetabular cup position after total hip arthroplasty: comparison between a radiographic planning software and three-dimensional computed tomography
verfasst von
Vahdettin Bayraktar
Markus Weber
Frederik von Kunow
Florian Zeman
Benjamin Craiovan
Tobias Renkawitz
Joachim Grifka
Michael Woerner
Publikationsdatum
08.06.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
International Orthopaedics / Ausgabe 4/2017
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-016-3240-1

Weitere Artikel der Ausgabe 4/2017

International Orthopaedics 4/2017 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.