19.12.2015 | Original Research

Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients

verfasst von: Werner Baulig, Burkhardt Seifert, Donat R. Spahn, Oliver M. Theusinger

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2017

Einloggen, um Zugang zu erhalten

Abstract

The Masimo Radical-7 Pulse CO-Oximeter (Masimo Corp., USA) non-invasively computes hemoglobin concentration (SpHb). SpHb was compared to Co-Oximeter readings (CoOxHb) of arterial samples in surgery patients of the emergency department. Forty-six patients were enrolled. The Masimo R1 25L (revision F and G) adult adhesive sensor was attached to the ring finger of the arterially cannulated hand. Before start, every 30 min during surgery and in the case of severe bleeding SpHb and CoOxHb values were documented. SpHb and post hoc adjusted SpHb (AdSpHb) values were analyzed. Linear regression analysis and Bland–Altman plot for agreement were performed. The detection failure rate of SpHb was 24.5 %. CoOxHb and SpHb showed a strong correlation (r = +0.81), but agreement was moderate [bias (LOA) of −0.6 (−3.0; +1.9)] g/dl. Positive and negative predicted value was 0.49 and 0.69. Exclusion of changes of CoOxHb values ≤1 g/dl resulted in a positive and negative predictive value of 0.66 and 1.00. Post hoc adjustment of the SpHb (AdSpHb) improved linear correlation of CoOxHb and AdSpHb [r = +0.90 (p < 0.001)] but less the agreement [bias (LOA) of CoOxHb and AdSpHb = −0.1 (−2.1/+1.9) g/dl]. SpHb agreed only moderately with CoOxHb values and predicted decreases of CoOxHb only if changes of SpHb ≤ 1.0 g/dl were excluded. The detection failure rate of SpHb was high. At present, additional refinements of the current technology are necessary to further improve performance of non-invasive hemoglobin measurement in the clinical setting.
Literatur
1.
Zurück zum Zitat Cothren CC, Moore EE, Hedegaard HB, Meng K. Epidemiology of urban trauma deaths: a comprehensive reassessment 10 years later. World J Surg. 2007;31:1507–11.CrossRefPubMed Cothren CC, Moore EE, Hedegaard HB, Meng K. Epidemiology of urban trauma deaths: a comprehensive reassessment 10 years later. World J Surg. 2007;31:1507–11.CrossRefPubMed
2.
Zurück zum Zitat Wen Y, Yang H, Wei W, Shan-shou L. The outcomes of 1120 severe multiple trauma patients with hemorrhagic shock in an emergency department: a retrospective study. BMC Emerg Med. 2013;13(Suppl 1):S6.CrossRefPubMedPubMedCentral Wen Y, Yang H, Wei W, Shan-shou L. The outcomes of 1120 severe multiple trauma patients with hemorrhagic shock in an emergency department: a retrospective study. BMC Emerg Med. 2013;13(Suppl 1):S6.CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Stadlbauer V, Wallner S, Stojakovic T, Smolle KH. Comparison of 3 different multianalyte point-of-care devices during clinical routine on a medical intensive care unit. J Crit Care. 2011;26(433):e1–11. Stadlbauer V, Wallner S, Stojakovic T, Smolle KH. Comparison of 3 different multianalyte point-of-care devices during clinical routine on a medical intensive care unit. J Crit Care. 2011;26(433):e1–11.
4.
Zurück zum Zitat Shamir MY, Avramovich A, Smaka T. The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration. Anesth Analg. 2012;114:972–8.CrossRefPubMed Shamir MY, Avramovich A, Smaka T. The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration. Anesth Analg. 2012;114:972–8.CrossRefPubMed
5.
Zurück zum Zitat O’Brien D. Audible alarms in medical equipment. Med Device Diagn Ind. 2006;28:98–103. O’Brien D. Audible alarms in medical equipment. Med Device Diagn Ind. 2006;28:98–103.
6.
Zurück zum Zitat Bergek C, Zdolsek JH, Hahn RG. Accuracy of noninvasive haemoglobin measurement by pulse oximetry depends on the type of infusion fluid. Eur J Anaesthesiol. 2013;30:73–9.CrossRefPubMed Bergek C, Zdolsek JH, Hahn RG. Accuracy of noninvasive haemoglobin measurement by pulse oximetry depends on the type of infusion fluid. Eur J Anaesthesiol. 2013;30:73–9.CrossRefPubMed
7.
Zurück zum Zitat Macknet MR, Allard M, Applegate RL 2nd, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111:1424–6.CrossRefPubMed Macknet MR, Allard M, Applegate RL 2nd, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111:1424–6.CrossRefPubMed
8.
Zurück zum Zitat Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7 pulse co-oximetry, HemoCue((R)) and laboratory haemoglobin measurements in obstetric patients. Anaesthesia. 2013;68:40–5.CrossRefPubMed Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7 pulse co-oximetry, HemoCue((R)) and laboratory haemoglobin measurements in obstetric patients. Anaesthesia. 2013;68:40–5.CrossRefPubMed
9.
Zurück zum Zitat Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011;113:1396–402.CrossRefPubMed Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011;113:1396–402.CrossRefPubMed
10.
Zurück zum Zitat Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111:946–54.CrossRefPubMed Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111:946–54.CrossRefPubMed
11.
Zurück zum Zitat Sjostrand F, Rodhe P, Berglund E, Lundstrom N, Svensen C. The use of a noninvasive hemoglobin monitor for volume kinetic analysis in an emergency room setting. Anesth Analg. 2013;116:337–42.CrossRefPubMed Sjostrand F, Rodhe P, Berglund E, Lundstrom N, Svensen C. The use of a noninvasive hemoglobin monitor for volume kinetic analysis in an emergency room setting. Anesth Analg. 2013;116:337–42.CrossRefPubMed
12.
Zurück zum Zitat Knutson T, Della-Giustina D, Tomich E, Wills B, Luerssen E, Reynolds P. Evaluation of a new nonnvasive device in determining hemoglobin levels in emergency department patients. West J Emerg Med. 2013;14:283–6.CrossRefPubMedPubMedCentral Knutson T, Della-Giustina D, Tomich E, Wills B, Luerssen E, Reynolds P. Evaluation of a new nonnvasive device in determining hemoglobin levels in emergency department patients. West J Emerg Med. 2013;14:283–6.CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Coquin J, Bertarrex A, Dewitte A, Lefevre L, Joannes-Boyau O, Fleureau C, Winnock S, Leuillet S, Janvier G, Ouattara A. Accuracy of determining hemoglobin level using occlusion spectroscopy in patients with severe gastrointestinal bleeding. Anesthesiology. 2013;118:640–8.CrossRefPubMed Coquin J, Bertarrex A, Dewitte A, Lefevre L, Joannes-Boyau O, Fleureau C, Winnock S, Leuillet S, Janvier G, Ouattara A. Accuracy of determining hemoglobin level using occlusion spectroscopy in patients with severe gastrointestinal bleeding. Anesthesiology. 2013;118:640–8.CrossRefPubMed
14.
Zurück zum Zitat Naftalovich R, Naftalovich D. Error in noninvasive spectrophotometric measurement of blood hemoglobin concentration under conditions of blood loss. Med Hypotheses. 2011;77:665–7.CrossRefPubMed Naftalovich R, Naftalovich D. Error in noninvasive spectrophotometric measurement of blood hemoglobin concentration under conditions of blood loss. Med Hypotheses. 2011;77:665–7.CrossRefPubMed
15.
Zurück zum Zitat Butcher N, Balogh ZJ. AIS>2 in at least two body regions: a potential new anatomical definition of polytrauma. Injury. 2012;43:196–9.CrossRefPubMed Butcher N, Balogh ZJ. AIS>2 in at least two body regions: a potential new anatomical definition of polytrauma. Injury. 2012;43:196–9.CrossRefPubMed
16.
Zurück zum Zitat Miyashita R, Hirata N, Sugino S, Mimura M, Yamakage M. Improved non-invasive total haemoglobin measurements after in vivo adjustment. Anaesthesia. 2014;69:752–6.CrossRefPubMed Miyashita R, Hirata N, Sugino S, Mimura M, Yamakage M. Improved non-invasive total haemoglobin measurements after in vivo adjustment. Anaesthesia. 2014;69:752–6.CrossRefPubMed
17.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed
18.
Zurück zum Zitat Hahn RG, Li Y, Zdolsek J. Non-invasive monitoring of blood haemoglobin for analysis of fluid volume kinetics. Acta Anaesthesiol Scand. 2010;54:1233–40.CrossRefPubMed Hahn RG, Li Y, Zdolsek J. Non-invasive monitoring of blood haemoglobin for analysis of fluid volume kinetics. Acta Anaesthesiol Scand. 2010;54:1233–40.CrossRefPubMed
19.
Zurück zum Zitat Vos JJ, Kalmar AF, Struys MM, Porte RJ, Wietasch JK, Scheeren TW, Hendriks HG. Accuracy of non-invasive measurement of haemoglobin concentration by pulse co-oximetry during steady-state and dynamic conditions in liver surgery. Br J Anaesth. 2012;109:522–8.CrossRefPubMed Vos JJ, Kalmar AF, Struys MM, Porte RJ, Wietasch JK, Scheeren TW, Hendriks HG. Accuracy of non-invasive measurement of haemoglobin concentration by pulse co-oximetry during steady-state and dynamic conditions in liver surgery. Br J Anaesth. 2012;109:522–8.CrossRefPubMed
20.
Zurück zum Zitat Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115:548–54.CrossRefPubMed Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115:548–54.CrossRefPubMed
21.
Zurück zum Zitat Causey MW, Miller S, Foster A, Beekley A, Zenger D, Martin M. Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg. 2011;201:592–8.CrossRefPubMed Causey MW, Miller S, Foster A, Beekley A, Zenger D, Martin M. Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station. Am J Surg. 2011;201:592–8.CrossRefPubMed
22.
Zurück zum Zitat Tsuei BJ, Hanseman DJ, Blakeman MJ, Blakeman TC, Yang SH, Branson RD, Gerlach TW. Accuracy of noninvasive hemoglobin monitoring in patients at risk for hemorrhage. J Trauma Acute Care Surg. 2014;77:S134–9.CrossRefPubMed Tsuei BJ, Hanseman DJ, Blakeman MJ, Blakeman TC, Yang SH, Branson RD, Gerlach TW. Accuracy of noninvasive hemoglobin monitoring in patients at risk for hemorrhage. J Trauma Acute Care Surg. 2014;77:S134–9.CrossRefPubMed
23.
Zurück zum Zitat Applegate RL 2nd, Barr SJ, Collier CE, Rook JL, Mangus DB, Allard MW. Evaluation of pulse cooximetry in patients undergoing abdominal or pelvic surgery. Anesthesiology. 2012;116:65–72.CrossRefPubMed Applegate RL 2nd, Barr SJ, Collier CE, Rook JL, Mangus DB, Allard MW. Evaluation of pulse cooximetry in patients undergoing abdominal or pelvic surgery. Anesthesiology. 2012;116:65–72.CrossRefPubMed
24.
Zurück zum Zitat Colquhoun DA, Forkin KT, Durieux ME, Thiele RH. Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin. J Clin Monit Comput. 2012;26:69–73.CrossRefPubMed Colquhoun DA, Forkin KT, Durieux ME, Thiele RH. Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin. J Clin Monit Comput. 2012;26:69–73.CrossRefPubMed
25.
Zurück zum Zitat Yang SH, Peter F, Anazodo A, Gao C, Chen H, Wade Ch., Hartsky L, Miller C, Imle C, Fang R, Mackenzie CF. Trends of hemoglobin oximetry: do they help predict blood transfusion during trauma patient resuscitation? Anesth Analg. 2015. Yang SH, Peter F, Anazodo A, Gao C, Chen H, Wade Ch., Hartsky L, Miller C, Imle C, Fang R, Mackenzie CF. Trends of hemoglobin oximetry: do they help predict blood transfusion during trauma patient resuscitation? Anesth Analg. 2015.
26.
Zurück zum Zitat Galvagno SM Jr, Hu P, Yang S, Gao C, Hanna D, Shackelford S, Mackenzie C. Accuracy of continuous noninvasive hemoglobin monitoring for the prediction of blood transfusions in trauma patients. J Clin Monit Comput. 2015;29:815–21.CrossRefPubMed Galvagno SM Jr, Hu P, Yang S, Gao C, Hanna D, Shackelford S, Mackenzie C. Accuracy of continuous noninvasive hemoglobin monitoring for the prediction of blood transfusions in trauma patients. J Clin Monit Comput. 2015;29:815–21.CrossRefPubMed
27.
Zurück zum Zitat Gehring H, Hornberger C, Dibbelt L, Rothsigkeit A, Gerlach K, Schumacher J, Schmucker P. Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand. 2002;46:980–6.CrossRefPubMed Gehring H, Hornberger C, Dibbelt L, Rothsigkeit A, Gerlach K, Schumacher J, Schmucker P. Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand. 2002;46:980–6.CrossRefPubMed
28.
Zurück zum Zitat Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39:2277–82.CrossRefPubMed Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39:2277–82.CrossRefPubMed
29.
Zurück zum Zitat Isosu T, Obara S, Hosono A, Ohashi S, Nakano Y, Imaizumi T, Mogami M, Murakawa M. Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients. J Clin Monit Comput. 2013;27:55–60.CrossRefPubMed Isosu T, Obara S, Hosono A, Ohashi S, Nakano Y, Imaizumi T, Mogami M, Murakawa M. Validation of continuous and noninvasive hemoglobin monitoring by pulse CO-oximetry in Japanese surgical patients. J Clin Monit Comput. 2013;27:55–60.CrossRefPubMed
30.
Zurück zum Zitat Gehring H, Duembgen L, Peterlein M, Hagelberg S, Dibbelt L. Hemoximetry as the “gold standard”? Error assessment based on differences among identical blood gas analyzer devices of five manufacturers. Anesth Analg. 2007;105:S24–30.CrossRefPubMed Gehring H, Duembgen L, Peterlein M, Hagelberg S, Dibbelt L. Hemoximetry as the “gold standard”? Error assessment based on differences among identical blood gas analyzer devices of five manufacturers. Anesth Analg. 2007;105:S24–30.CrossRefPubMed
31.
Zurück zum Zitat Mokken FC, van der Waart FJ, Henny CP, Goedhart PT, Gelb AW. Differences in peripheral arterial and venous hemorheologic parameters. Ann Hematol. 1996;73:135–7.CrossRefPubMed Mokken FC, van der Waart FJ, Henny CP, Goedhart PT, Gelb AW. Differences in peripheral arterial and venous hemorheologic parameters. Ann Hematol. 1996;73:135–7.CrossRefPubMed
32.
Zurück zum Zitat Yang ZW, Yang SH, Chen L, Qu J, Zhu J, Tang Z. Comparison of blood counts in venous, fingertip and arterial blood and their measurement variation. Clin Lab Haematol. 2001;23:155–9.CrossRefPubMed Yang ZW, Yang SH, Chen L, Qu J, Zhu J, Tang Z. Comparison of blood counts in venous, fingertip and arterial blood and their measurement variation. Clin Lab Haematol. 2001;23:155–9.CrossRefPubMed
33.
Zurück zum Zitat Prist R, Rocha-e-Silva M, Scalabrini A, Coelho IJ, Franca ES, Meneghetti C, Braga GA. A quantitative analysis of transcapillary refill in severe hemorrhagic hypotension in dogs. Shock. 1994;1:188–95.CrossRefPubMed Prist R, Rocha-e-Silva M, Scalabrini A, Coelho IJ, Franca ES, Meneghetti C, Braga GA. A quantitative analysis of transcapillary refill in severe hemorrhagic hypotension in dogs. Shock. 1994;1:188–95.CrossRefPubMed
Metadaten
Titel
Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients
verfasst von
Werner Baulig
Burkhardt Seifert
Donat R. Spahn
Oliver M. Theusinger
Publikationsdatum
19.12.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2017
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9816-2

Weitere Artikel der Ausgabe 1/2017

Journal of Clinical Monitoring and Computing 1/2017 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.