Accuracy of tibial cuts with patient-specific instrumentation is not influenced by the surgeon’s level of experience

verfasst von: Alexander Antoniadis, Roland S. Camenzind, Michael O. Schär, Dario Bergadano, Näder Helmy

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 5/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

It was hypothesized that surgeon’s experience as well as bone density play a significant role in achieving accurate cuts with patient-specific instrumentation (PSI). The aim of this study was to compare the accuracy of the tibial cuts in different bone densities made by a highly experienced orthopedic surgeon on one hand and a less experienced orthopedic surgeon on the other.

Methods

Tibial models from three different sawbone densities were developed for this study. Each surgeon performed 21 cuts. A coordinate measuring machine was used to analyse the cuts. The K-Cohen test was performed to evaluate the results. The analyzed parameters were guide positioning and deviation from the guide cut to the tibial cut, including varus/valgus angle, the tibial slope, cut height, planarity (mm2), and rugosity (mm).

Results

There was a significant difference in the positioning of the tibial cut guide between the two surgeons for the tibial slope (p < 0.05), while no difference was observed for the varus/valgus angle (n.s.) and the cut height (n.s.). No significant difference in the tibial cut was observed between the surgeons for the tibial slope angle (n.s.), varus/valgus angle (n.s.), planarity (n.s.), and rugosity (n.s.). In the different bone types, no significant difference was observed for the tibial slope (n.s.) and varus/valgus angle (n.s.), while planarity and rugosity showed significant differences (p < 0.05). Our study showed no significant difference in the tibial cuts for the tibial slope, varus/valgus angle, planarity, and rugosity between the two surgeons.

Conclusions

In the present study, it could be demonstrated that accuracy of the cuts is ensured by PSI not depending on the surgeon’s experience and the bone mineral density. This speaks to its clinical significance: PSI might be suited for less experienced surgeons to reduce outliers in total knee arthroplasty (TKA).
Literatur
1.
Zurück zum Zitat Baker PN, Deehan DJ, Lees D, Jameson S, Avery PJ, Gregg PJ et al (2012) The effect of surgical factors on early patient-reported outcome measures (PROMS) following total knee replacement. J Bone Jt Surg Br 94:1058–1066CrossRef Baker PN, Deehan DJ, Lees D, Jameson S, Avery PJ, Gregg PJ et al (2012) The effect of surgical factors on early patient-reported outcome measures (PROMS) following total knee replacement. J Bone Jt Surg Br 94:1058–1066CrossRef
2.
Zurück zum Zitat Baker PN, van der Meulen JH, Lewsey J, Gregg PJ (2007) The role of pain and function in determining patient satisfaction after total knee replacement. Data from the National Joint Registry for England and Wales. J Bone Jt Surg Br 89:893–900CrossRef Baker PN, van der Meulen JH, Lewsey J, Gregg PJ (2007) The role of pain and function in determining patient satisfaction after total knee replacement. Data from the National Joint Registry for England and Wales. J Bone Jt Surg Br 89:893–900CrossRef
3.
Zurück zum Zitat Bali K, Walker P, Bruce W (2012) Custom-fit total knee arthroplasty: our initial experience in 32 knees. J Arthroplast 27:1149–1154CrossRef Bali K, Walker P, Bruce W (2012) Custom-fit total knee arthroplasty: our initial experience in 32 knees. J Arthroplast 27:1149–1154CrossRef
4.
Zurück zum Zitat Barrack RL, Ruh EL, Williams BM, Ford AD, Foreman K, Nunley RM (2012) Patient specific cutting blocks are currently of no proven value. J Bone Jt Surg Br 94:95–99CrossRef Barrack RL, Ruh EL, Williams BM, Ford AD, Foreman K, Nunley RM (2012) Patient specific cutting blocks are currently of no proven value. J Bone Jt Surg Br 94:95–99CrossRef
5.
Zurück zum Zitat Berend ME, Ritter MA, Meding JB, Faris PM, Keating EM, Redelman R et al (2004) Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res 428:26–34CrossRef Berend ME, Ritter MA, Meding JB, Faris PM, Keating EM, Redelman R et al (2004) Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res 428:26–34CrossRef
6.
Zurück zum Zitat Boonen B, Schotanus MG, Kerens B, van der Weegen W, van Drumpt RA, Kort NP (2013) Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee Surg Sports Traumatol Arthrosc 21:2206–2212CrossRefPubMed Boonen B, Schotanus MG, Kerens B, van der Weegen W, van Drumpt RA, Kort NP (2013) Intra-operative results and radiological outcome of conventional and patient-specific surgery in total knee arthroplasty: a multicentre, randomised controlled trial. Knee Surg Sports Traumatol Arthrosc 21:2206–2212CrossRefPubMed
7.
Zurück zum Zitat Bottomley N, Jones LD, Rout R, Alvand A, Rombach I, Evans T et al (2016) A survival analysis of 1084 knees of the Oxford unicompartmental knee arthroplasty: a comparison between consultant and trainee surgeons. Bone Jt J 98-B:22–27CrossRef Bottomley N, Jones LD, Rout R, Alvand A, Rombach I, Evans T et al (2016) A survival analysis of 1084 knees of the Oxford unicompartmental knee arthroplasty: a comparison between consultant and trainee surgeons. Bone Jt J 98-B:22–27CrossRef
8.
Zurück zum Zitat Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KD (2010) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res 468:57–63CrossRefPubMed Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KD (2010) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res 468:57–63CrossRefPubMed
9.
Zurück zum Zitat Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C (2013) A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Jt J 95-B:354–359CrossRef Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C (2013) A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Jt J 95-B:354–359CrossRef
10.
Zurück zum Zitat Chen JY, Yeo SJ, Yew AK, Tay DK, Chia SL, Lo NN et al (2014) The radiological outcomes of patient-specific instrumentation versus conventional total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22:630–635CrossRefPubMed Chen JY, Yeo SJ, Yew AK, Tay DK, Chia SL, Lo NN et al (2014) The radiological outcomes of patient-specific instrumentation versus conventional total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 22:630–635CrossRefPubMed
11.
Zurück zum Zitat Helmy N, Dao Trong ML, Kuhnel SP (2014) Accuracy of patient specific cutting blocks in total knee arthroplasty. Biomed Res Int 2014:562919PubMedPubMedCentral Helmy N, Dao Trong ML, Kuhnel SP (2014) Accuracy of patient specific cutting blocks in total knee arthroplasty. Biomed Res Int 2014:562919PubMedPubMedCentral
12.
Zurück zum Zitat Heyse TJ, Tibesku CO (2015) Improved tibial component rotation in TKA using patient-specific instrumentation. Arch Orthop Trauma Surg 135:697–701CrossRefPubMed Heyse TJ, Tibesku CO (2015) Improved tibial component rotation in TKA using patient-specific instrumentation. Arch Orthop Trauma Surg 135:697–701CrossRefPubMed
13.
Zurück zum Zitat Higuera CA, Deirmengian C (2012) The new technologies in knee arthroplasty. J Clin Rheumatol 18:345–348CrossRefPubMed Higuera CA, Deirmengian C (2012) The new technologies in knee arthroplasty. J Clin Rheumatol 18:345–348CrossRefPubMed
14.
Zurück zum Zitat Insall JN, Binazzi R, Soudry M, Mestriner LA (1985) Total knee arthroplasty. Clin Orthop Relat Res 192):13–22 Insall JN, Binazzi R, Soudry M, Mestriner LA (1985) Total knee arthroplasty. Clin Orthop Relat Res 192):13–22
15.
Zurück zum Zitat Jenny JY, Clemens U, Kohler S, Kiefer H, Konermann W, Miehlke RK (2005) Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a case-control study of 235 cases compared with 235 conventionally implanted prostheses. J Arthroplast 20:832–839CrossRef Jenny JY, Clemens U, Kohler S, Kiefer H, Konermann W, Miehlke RK (2005) Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a case-control study of 235 cases compared with 235 conventionally implanted prostheses. J Arthroplast 20:832–839CrossRef
16.
Zurück zum Zitat Katz JN, Barrett J, Mahomed NN, Baron JA, Wright RJ, Losina E (2004) Association between hospital and surgeon procedure volume and the outcomes of total knee replacement. J Bone Jt Surg Am 86-A:1909–1916CrossRef Katz JN, Barrett J, Mahomed NN, Baron JA, Wright RJ, Losina E (2004) Association between hospital and surgeon procedure volume and the outcomes of total knee replacement. J Bone Jt Surg Am 86-A:1909–1916CrossRef
17.
Zurück zum Zitat Katz JN, Mahomed NN, Baron JA, Barrett JA, Fossel AH, Creel AH et al (2007) Association of hospital and surgeon procedure volume with patient-centered outcomes of total knee replacement in a population-based cohort of patients age 65 years and older. Arthritis Rheum 56:568–574CrossRefPubMed Katz JN, Mahomed NN, Baron JA, Barrett JA, Fossel AH, Creel AH et al (2007) Association of hospital and surgeon procedure volume with patient-centered outcomes of total knee replacement in a population-based cohort of patients age 65 years and older. Arthritis Rheum 56:568–574CrossRefPubMed
18.
Zurück zum Zitat Kotela A, Kotela I (2014) Patient-specific computed tomography based instrumentation in total knee arthroplasty: a prospective randomized controlled study. Int Orthop 38:2099–2107CrossRefPubMed Kotela A, Kotela I (2014) Patient-specific computed tomography based instrumentation in total knee arthroplasty: a prospective randomized controlled study. Int Orthop 38:2099–2107CrossRefPubMed
19.
Zurück zum Zitat Krishnan SP, Dawood A, Richards R, Henckel J, Hart AJ (2012) A review of rapid prototyped surgical guides for patient-specific total knee replacement. J Bone Jt Surg Br 94:1457–1461CrossRef Krishnan SP, Dawood A, Richards R, Henckel J, Hart AJ (2012) A review of rapid prototyped surgical guides for patient-specific total knee replacement. J Bone Jt Surg Br 94:1457–1461CrossRef
20.
Zurück zum Zitat Labek G, Thaler M, Janda W, Agreiter M, Stockl B (2011) Revision rates after total joint replacement: cumulative results from worldwide joint register datasets. J Bone Jt Surg Br 93:293–297CrossRef Labek G, Thaler M, Janda W, Agreiter M, Stockl B (2011) Revision rates after total joint replacement: cumulative results from worldwide joint register datasets. J Bone Jt Surg Br 93:293–297CrossRef
21.
Zurück zum Zitat Lee DH, Padhy D, Lee SH, Nha KW, Park JH, Han SB (2012) Osteoporosis affects component positioning in computer navigation-assisted total knee arthroplasty. Knee 19:203–207CrossRefPubMed Lee DH, Padhy D, Lee SH, Nha KW, Park JH, Han SB (2012) Osteoporosis affects component positioning in computer navigation-assisted total knee arthroplasty. Knee 19:203–207CrossRefPubMed
22.
Zurück zum Zitat Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R (2009) Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplast 24:570–578CrossRef Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R (2009) Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplast 24:570–578CrossRef
23.
Zurück zum Zitat Nam D, Maher PA, Rebolledo BJ, Nawabi DH, McLawhorn AS, Pearle AD (2013) Patient specific cutting guides versus an imageless, computer-assisted surgery system in total knee arthroplasty. Knee 20:263–267CrossRefPubMed Nam D, Maher PA, Rebolledo BJ, Nawabi DH, McLawhorn AS, Pearle AD (2013) Patient specific cutting guides versus an imageless, computer-assisted surgery system in total knee arthroplasty. Knee 20:263–267CrossRefPubMed
24.
Zurück zum Zitat Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107CrossRefPubMed Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107CrossRefPubMed
25.
Zurück zum Zitat Noble JW Jr, Moore CA, Liu N (2012) The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplast 27:153–155CrossRef Noble JW Jr, Moore CA, Liu N (2012) The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplast 27:153–155CrossRef
26.
Zurück zum Zitat Nunley RM, Ellison BS, Ruh EL, Williams BM, Foreman K, Ford AD et al (2012) Are patient-specific cutting blocks cost-effective for total knee arthroplasty? Clin Orthop Relat Res 470:889–894CrossRefPubMed Nunley RM, Ellison BS, Ruh EL, Williams BM, Foreman K, Ford AD et al (2012) Are patient-specific cutting blocks cost-effective for total knee arthroplasty? Clin Orthop Relat Res 470:889–894CrossRefPubMed
27.
Zurück zum Zitat Ravi B, Croxford R, Austin PC, Hollands S, Paterson JM, Bogoch E et al (2014) Increased surgeon experience with rheumatoid arthritis reduces the risk of complications following total joint arthroplasty. Arthritis Rheumatol 66:488–496CrossRefPubMed Ravi B, Croxford R, Austin PC, Hollands S, Paterson JM, Bogoch E et al (2014) Increased surgeon experience with rheumatoid arthritis reduces the risk of complications following total joint arthroplasty. Arthritis Rheumatol 66:488–496CrossRefPubMed
28.
Zurück zum Zitat Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA (2011) The effect of alignment and BMI on failure of total knee replacement. J Bone Jt Surg Am 93:1588–1596CrossRef Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA (2011) The effect of alignment and BMI on failure of total knee replacement. J Bone Jt Surg Am 93:1588–1596CrossRef
29.
Zurück zum Zitat Russell R, Brown T, Huo M, Jones R (2014) Patient-specific instrumentation does not improve alignment in total knee arthroplasty. J Knee Surg 27:501–504CrossRefPubMed Russell R, Brown T, Huo M, Jones R (2014) Patient-specific instrumentation does not improve alignment in total knee arthroplasty. J Knee Surg 27:501–504CrossRefPubMed
30.
Zurück zum Zitat Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME et al (2013) Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplast 28:116–119CrossRef Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME et al (2013) Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplast 28:116–119CrossRef
31.
Zurück zum Zitat Scott CE, Howie CR, MacDonald D, Biant LC (2010) Predicting dissatisfaction following total knee replacement: a prospective study of 1217 patients. J Bone Jt Surg Br 92:1253–1258CrossRef Scott CE, Howie CR, MacDonald D, Biant LC (2010) Predicting dissatisfaction following total knee replacement: a prospective study of 1217 patients. J Bone Jt Surg Br 92:1253–1258CrossRef
32.
Zurück zum Zitat Tibesku CO, Hofer P, Portegies W, Ruys CJ, Fennema P (2013) Benefits of using customized instrumentation in total knee arthroplasty: results from an activity-based costing model. Arch Orthop Trauma Surg 133:405–411CrossRefPubMed Tibesku CO, Hofer P, Portegies W, Ruys CJ, Fennema P (2013) Benefits of using customized instrumentation in total knee arthroplasty: results from an activity-based costing model. Arch Orthop Trauma Surg 133:405–411CrossRefPubMed
33.
Zurück zum Zitat Tibesku CO, Innocenti B, Wong P, Salehi A, Labey L (2012) Can CT-based patient-matched instrumentation achieve consistent rotational alignment in knee arthroplasty? Arch Orthop Trauma Surg 132:171–177CrossRefPubMed Tibesku CO, Innocenti B, Wong P, Salehi A, Labey L (2012) Can CT-based patient-matched instrumentation achieve consistent rotational alignment in knee arthroplasty? Arch Orthop Trauma Surg 132:171–177CrossRefPubMed
34.
Zurück zum Zitat Victor J, Dujardin J, Vandenneucker H, Arnout N, Bellemans J (2014) Patient-specific guides do not improve accuracy in total knee arthroplasty: a prospective randomized controlled trial. Clin Orthop Relat Res 472:263–271CrossRefPubMed Victor J, Dujardin J, Vandenneucker H, Arnout N, Bellemans J (2014) Patient-specific guides do not improve accuracy in total knee arthroplasty: a prospective randomized controlled trial. Clin Orthop Relat Res 472:263–271CrossRefPubMed
35.
Zurück zum Zitat Voleti PB, Hamula MJ, Baldwin KD, Lee GC (2014) Current data do not support routine use of patient-specific instrumentation in total knee arthroplasty. J Arthroplast 29:1709–1712CrossRef Voleti PB, Hamula MJ, Baldwin KD, Lee GC (2014) Current data do not support routine use of patient-specific instrumentation in total knee arthroplasty. J Arthroplast 29:1709–1712CrossRef
Metadaten
Titel
Accuracy of tibial cuts with patient-specific instrumentation is not influenced by the surgeon’s level of experience
verfasst von
Alexander Antoniadis
Roland S. Camenzind
Michael O. Schär
Dario Bergadano
Näder Helmy
Publikationsdatum
05.06.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 5/2019
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-018-4992-5

Weitere Artikel der Ausgabe 5/2019

Knee Surgery, Sports Traumatology, Arthroscopy 5/2019 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.