08.05.2020 | Laboratory Investigation

Acute Thrombus Burden on Coated Flow Diverters Assessed by High Frequency Optical Coherence Tomography

Zeitschrift:
CardioVascular and Interventional Radiology
Autoren:
Robert M. King, Erin T. Langan, Giovanni J. Ughi, Christopher M. Raskett, Ajit S. Puri, Hans Henkes, Matthew J. Gounis
Wichtige Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s00270-020-02482-w) contains supplementary material, which is available to authorized users.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abstract

Purpose

The implantation of flow diverters requires administration of dual anti-platelet therapy, posing the potential for complications. The p48MW HPC (phenox, Bochüm, Germany) hydrophilic-coated flow diverting stent is designed to be anti-thrombotic, thus opening the potential for single anti-platelet therapy. We deploy a novel intravascular high-resolution imaging technique, high-frequency optical coherence tomography (HF-OCT), to study in an animal model the acute thrombus formation on coated p48MW devices versus uncoated control devices.

Methods

Three pigs were implanted with 4 flow diverters each, two test hydrophilic-coated devices, and two control uncoated devices (p48MW). Each pig was treated with a different anti-platelet regime: no anti-platelet therapy, aspirin only, aspirin and clopidogrel. Twenty minutes after the flow diverter was implanted, an HF-OCT data set was acquired. Acute clot formed on the flow diverter at each covered side branch was measured from the HF-OCT slices. Factors considered to be important were the device type (pHPC versus bare metal), aspirin, clopidogrel, and vessel location. A linear model was constructed from the significant factors.

Results

Both coating (p < 0.001) and aspirin (p = 0.003) were significantly related to reduction in clot burden, leading to an approximate 100-fold and 50-fold reduction in clot, respectively.

Conclusions

This study shows the power of HF-OCT not only in the detection of clot but also the quantification of clot burden. In an animal model, the pHPC-coated p48MW significantly reduced acute thrombus formation over jailed side branches as compared to the bare metal p48MW that was nearly eliminated when combined with aspirin administration.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Weitere Produktempfehlungen anzeigen
Zusatzmaterial
Nur für berechtigte Nutzer zugänglich
Literatur
Über diesen Artikel
  1. Sie können e.Med Innere Medizin 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

  2. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Radiologie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Radiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise