26.08.2016 | Original Article

Additive value of 320-section low-dose dynamic volume CT in relation to 3-T MRI for the preoperative evaluation of brain tumors

verfasst von: Eri Hayashida, Toshinori Hirai, Hideo Nakamura, Masafumi Kidoh, Minako Azuma, Yasuhiko Iryo, Mika Kitajima, Seitaro Oda, Daisuke Utsunomiya, Takeshi Nakaura, Yasuyuki Yamashita

Erschienen in: Japanese Journal of Radiology | Ausgabe 10/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To assess whether 320-section low-dose dynamic volume computed tomography (320-LDVCT) with adaptive iterative dose reduction (AIDR) adds value to 3-T MRI for the preoperative evaluation of brain tumors.

Methods

The study population was comprised of 16 consecutive patients with brain tumors who, in addition to preoperative 3-T MRI, underwent 320-LDVCT with AIDR. Two radiologists independently evaluated the CT and MRI studies; one measured the relative cerebral blood volume (rCBV) in the tumor and contralateral brain on CT and MR perfusion maps. Interobserver agreement was assessed by κ statistics.

Results

In 3 of 16 patients (19 %), 320-LDVCT added diagnostic value to 3-T MRI studies with respect to the visualization of feeders (κ = 0.77), and in 12 (75 %) it helped the delineation of venous structures (κ = 0.71) and the relationship between the tumor and adjacent vessels (κ = 0.85). The average standardized rCBV value was 12.2 ± 2.40 (range 0.7–36.6) on MR and 8.80 ± 2.77 (range 0.8–38.0) on CT perfusion studies; the correlation between these values was very strong (r = 0.92, p < 0.0001). According to the neurosurgeons, 320-LDVCT added helpful information for surgery in 4 patients (25 %).

Conclusion

The 320-LDVCT can add value to 3-T MRI for the tumor feeders and relationship between the tumor and adjacent vessels.
Literatur
1.
Zurück zum Zitat Essig M, Anzalone N, Combs SE, Dorfler A, Lee SK, Picozzi P, et al. MR imaging of neoplastic central nervous system lesions: review and recommendations for current practice. AJNR Am J Neuroradiol. 2012;33:803–17.CrossRefPubMed Essig M, Anzalone N, Combs SE, Dorfler A, Lee SK, Picozzi P, et al. MR imaging of neoplastic central nervous system lesions: review and recommendations for current practice. AJNR Am J Neuroradiol. 2012;33:803–17.CrossRefPubMed
2.
Zurück zum Zitat Uetani H, Akter M, Hirai T, Shigematsu Y, Kitajima M, Kai Y, et al. Can 3 T MR angiography replace DSA for the identification of arteries feeding intracranial meningiomas? AJNR Am J Neuroradiol. 2013;34:765–72.CrossRefPubMed Uetani H, Akter M, Hirai T, Shigematsu Y, Kitajima M, Kai Y, et al. Can 3 T MR angiography replace DSA for the identification of arteries feeding intracranial meningiomas? AJNR Am J Neuroradiol. 2013;34:765–72.CrossRefPubMed
3.
Zurück zum Zitat Murayama K, Katada K, Nakane M, Toyama H, Anno H, Hayakawa M, et al. Whole-brain perfusion CT performed with a prototype 256-detector row CT system: initial experience. Radiology. 2009;250:202–11.CrossRefPubMed Murayama K, Katada K, Nakane M, Toyama H, Anno H, Hayakawa M, et al. Whole-brain perfusion CT performed with a prototype 256-detector row CT system: initial experience. Radiology. 2009;250:202–11.CrossRefPubMed
4.
Zurück zum Zitat Hayakawa M, Maeda S, Sadato A, Tanaka T, Kaito T, Hattori N, et al. Detection of pulsation in ruptured and unruptured cerebral aneurysms by electrocardiographically gated 3-dimensional computed tomographic angiography with a 320-row area detector computed tomography and evaluation of its clinical usefulness. Neurosurgery. 2011;69:843–51 (discussion 51).CrossRefPubMed Hayakawa M, Maeda S, Sadato A, Tanaka T, Kaito T, Hattori N, et al. Detection of pulsation in ruptured and unruptured cerebral aneurysms by electrocardiographically gated 3-dimensional computed tomographic angiography with a 320-row area detector computed tomography and evaluation of its clinical usefulness. Neurosurgery. 2011;69:843–51 (discussion 51).CrossRefPubMed
5.
Zurück zum Zitat Willems PW, Brouwer PA, Barfett JJ, terBrugge KG, Krings T. Detection and classification of cranial dural arteriovenous fistulas using 4D-CT angiography: initial experience. AJNR Am J Neuroradiol. 2011;32:49–53.PubMed Willems PW, Brouwer PA, Barfett JJ, terBrugge KG, Krings T. Detection and classification of cranial dural arteriovenous fistulas using 4D-CT angiography: initial experience. AJNR Am J Neuroradiol. 2011;32:49–53.PubMed
6.
Zurück zum Zitat Kim DJ, Krings T. Whole-brain perfusion CT patterns of brain arteriovenous malformations: a pilot study in 18 patients. AJNR Am J Neuroradiol. 2011;32:2061–6.CrossRefPubMed Kim DJ, Krings T. Whole-brain perfusion CT patterns of brain arteriovenous malformations: a pilot study in 18 patients. AJNR Am J Neuroradiol. 2011;32:2061–6.CrossRefPubMed
7.
Zurück zum Zitat Willems PW, Taeshineetanakul P, Schenk B, Brouwer PA, Terbrugge KG, Krings T. The use of 4D-CTA in the diagnostic work-up of brain arteriovenous malformations. Neuroradiology. 2012;54:123–31.CrossRefPubMed Willems PW, Taeshineetanakul P, Schenk B, Brouwer PA, Terbrugge KG, Krings T. The use of 4D-CTA in the diagnostic work-up of brain arteriovenous malformations. Neuroradiology. 2012;54:123–31.CrossRefPubMed
8.
Zurück zum Zitat Chen W, Xing W, Peng Y, He Z, Wang C, Wang Q. Cerebral aneurysms: accuracy of 320-detector row nonsubtracted and subtracted volumetric CT angiography for diagnosis. Radiology. 2013;269:841–9.CrossRefPubMed Chen W, Xing W, Peng Y, He Z, Wang C, Wang Q. Cerebral aneurysms: accuracy of 320-detector row nonsubtracted and subtracted volumetric CT angiography for diagnosis. Radiology. 2013;269:841–9.CrossRefPubMed
9.
Zurück zum Zitat Kidoh M, Hirai T, Oda S, Utsunomiya D, Kawano T, Yano S, et al. Can CT angiography reconstructed from CT perfusion source data on a 320-section volume CT scanner replace conventional CT angiography for the evaluation of intracranial arteries? Jpn J Radiol. 2015;33:353–9.CrossRefPubMed Kidoh M, Hirai T, Oda S, Utsunomiya D, Kawano T, Yano S, et al. Can CT angiography reconstructed from CT perfusion source data on a 320-section volume CT scanner replace conventional CT angiography for the evaluation of intracranial arteries? Jpn J Radiol. 2015;33:353–9.CrossRefPubMed
10.
Zurück zum Zitat Funama Y, Utsunomiya D, Taguchi K, Oda S, Shimonobo T, Yamashita Y. Automatic exposure control at single- and dual-heartbeat CTCA on a 320-MDCT volume scanner: effect of heart rate, exposure phase window setting, and reconstruction algorithm. Phys Med. 2014;30:385–90.CrossRefPubMed Funama Y, Utsunomiya D, Taguchi K, Oda S, Shimonobo T, Yamashita Y. Automatic exposure control at single- and dual-heartbeat CTCA on a 320-MDCT volume scanner: effect of heart rate, exposure phase window setting, and reconstruction algorithm. Phys Med. 2014;30:385–90.CrossRefPubMed
11.
Zurück zum Zitat Nitta N, Ikeda M, Sonoda A, Nagatani Y, Ohta S, Takahashi M, et al. Images acquired using 320-MDCT with adaptive iterative dose reduction with wide-volume acquisition: visual evaluation of image quality by 10 radiologists using an abdominal phantom. AJR Am J Roentgenol. 2014;202:2–12.CrossRefPubMed Nitta N, Ikeda M, Sonoda A, Nagatani Y, Ohta S, Takahashi M, et al. Images acquired using 320-MDCT with adaptive iterative dose reduction with wide-volume acquisition: visual evaluation of image quality by 10 radiologists using an abdominal phantom. AJR Am J Roentgenol. 2014;202:2–12.CrossRefPubMed
12.
Zurück zum Zitat Tabuchi S, Nakajima S. Usefulness of 320-row area detector computed tomography for the diagnosis of cystic falx meningioma. Case Rep Oncol. 2013;6:362–6.CrossRefPubMedPubMedCentral Tabuchi S, Nakajima S. Usefulness of 320-row area detector computed tomography for the diagnosis of cystic falx meningioma. Case Rep Oncol. 2013;6:362–6.CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Chen T, Guo D, Fang Z, Zhong W, Zhao J, Jiang Y. Preliminary study of whole-brain CT perfusion imaging in patients with intracranial tumours adjacent to large blood vessels. Clin Radiol. 2014;69:e25–32.CrossRefPubMed Chen T, Guo D, Fang Z, Zhong W, Zhao J, Jiang Y. Preliminary study of whole-brain CT perfusion imaging in patients with intracranial tumours adjacent to large blood vessels. Clin Radiol. 2014;69:e25–32.CrossRefPubMed
14.
Zurück zum Zitat Rosen BR, Belliveau JW, Buchbinder BR, McKinstry RC, Porkka LM, Kennedy DN, et al. Contrast agents and cerebral hemodynamics. Magn Reson Med. 1991;19:285–92.CrossRefPubMed Rosen BR, Belliveau JW, Buchbinder BR, McKinstry RC, Porkka LM, Kennedy DN, et al. Contrast agents and cerebral hemodynamics. Magn Reson Med. 1991;19:285–92.CrossRefPubMed
15.
Zurück zum Zitat Mert A, Buehler K, Sutherland GR, Tomanek B, Widhalm G, Kasprian G, et al. Brain tumor surgery with 3-dimensional surface navigation. Neurosurgery. 2012;71:ons286–94 (discussion ons94-5).PubMed Mert A, Buehler K, Sutherland GR, Tomanek B, Widhalm G, Kasprian G, et al. Brain tumor surgery with 3-dimensional surface navigation. Neurosurgery. 2012;71:ons286–94 (discussion ons94-5).PubMed
16.
Zurück zum Zitat Nabavi DG, Cenic A, Craen RA, Gelb AW, Bennett JD, Kozak R, et al. CT assessment of cerebral perfusion: experimental validation and initial clinical experience. Radiology. 1999;213:141–9.CrossRefPubMed Nabavi DG, Cenic A, Craen RA, Gelb AW, Bennett JD, Kozak R, et al. CT assessment of cerebral perfusion: experimental validation and initial clinical experience. Radiology. 1999;213:141–9.CrossRefPubMed
17.
Zurück zum Zitat Eastwood JD, Provenzale JM. Cerebral blood flow, blood volume, and vascular permeability of cerebral glioma assessed with dynamic CT perfusion imaging. Neuroradiology. 2003;45:373–6.CrossRefPubMed Eastwood JD, Provenzale JM. Cerebral blood flow, blood volume, and vascular permeability of cerebral glioma assessed with dynamic CT perfusion imaging. Neuroradiology. 2003;45:373–6.CrossRefPubMed
18.
Zurück zum Zitat Ellika SK, Jain R, Patel SC, Scarpace L, Schultz LR, Rock JP, et al. Role of perfusion CT in glioma grading and comparison with conventional MR imaging features. AJNR Am J Neuroradiol. 2007;28:1981–7.CrossRefPubMed Ellika SK, Jain R, Patel SC, Scarpace L, Schultz LR, Rock JP, et al. Role of perfusion CT in glioma grading and comparison with conventional MR imaging features. AJNR Am J Neuroradiol. 2007;28:1981–7.CrossRefPubMed
19.
Zurück zum Zitat Nishimura S, Hirai T, Shigematsu Y, Kitajima M, Morioka M, Kai Y, et al. Evaluation of brain and head and neck tumors with 4D contrast-enhanced MR angiography at 3T. AJNR Am J Neuroradiol. 2012;33:445–8.CrossRefPubMed Nishimura S, Hirai T, Shigematsu Y, Kitajima M, Morioka M, Kai Y, et al. Evaluation of brain and head and neck tumors with 4D contrast-enhanced MR angiography at 3T. AJNR Am J Neuroradiol. 2012;33:445–8.CrossRefPubMed
Metadaten
Titel
Additive value of 320-section low-dose dynamic volume CT in relation to 3-T MRI for the preoperative evaluation of brain tumors
verfasst von
Eri Hayashida
Toshinori Hirai
Hideo Nakamura
Masafumi Kidoh
Minako Azuma
Yasuhiko Iryo
Mika Kitajima
Seitaro Oda
Daisuke Utsunomiya
Takeshi Nakaura
Yasuyuki Yamashita
Publikationsdatum
26.08.2016
Verlag
Springer Japan
Erschienen in
Japanese Journal of Radiology / Ausgabe 10/2016
Print ISSN: 1867-1071
Elektronische ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-016-0576-9

Weitere Artikel der Ausgabe 10/2016

Japanese Journal of Radiology 10/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.