An improvised pressure gauge for regional nerve blockade/anesthesia injections: an initial study

verfasst von: Jayaprakash Patil, Hari Ankireddy, Antony Wilkes, David Williams, Michael Lim

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 6/2015

Einloggen, um Zugang zu erhalten

Abstract

High injection pressure is one of the warning signs of intraneural injection, with animal models suggesting pressures higher than 69 or 176 kPa as high risk, and is normally detected subjectively and inaccurately. We describe a system improvised from common clinical components that uses Boyle’s law to objectively measure injection pressure. The objectives of the study were to (1) Validate our improvised pressure gauge (IPG) by comparing the injection pressure as calculated by Boyle’s law against the measured pressure and (2) Use the IPG to measure the range of injection pressures by two groups of anesthetic professionals using the “syringe feel” technique. Our IPG system consists of an extended 1 ml syringe attached to a 3-way stopcock, inserted between the syringe containing the local anesthetic injectate and the needle. The IPG was validated against a pressure calibration reference. 20 anesthesiologists and 20 anesthetic assistants were recruited to apply pressure to the 20 ml syringe in vitro while blinded to the attached IPG. The pressures were measured on three separate occasions for each participant. There was good agreement (<8 percent difference) between the measured and theoretical pressure values. Anesthesiologists exceeded the threshold of 69 kPa in 18 of a total of 60 attempts whereas anesthetic assistants exceeded the threshold in 30 attempts out of 60 attempts. Anesthetic assistants exerted a higher overall pressure of 80 kPa compared to 51 kPa for anesthesiologists—this was statistically significant (p = 0.027). Our improvised system is easily and rapidly assembled from common clinical equipment and shows promise as a monitor for inadvertent intraneural injection.
Literatur
1.
Zurück zum Zitat Contractor S, Hardman JG. Injury during anaesthesia. Contin Educ Anaesth Crit Care Pain. 2006;6:67–70.CrossRef Contractor S, Hardman JG. Injury during anaesthesia. Contin Educ Anaesth Crit Care Pain. 2006;6:67–70.CrossRef
2.
Zurück zum Zitat Selander D, Brattsand R, Lundborg G, Nordborg C, Olsson Y. Local anesthetics: importance of mode of application, concentration and adrenaline for the appearance of nerve lesions. An experimental study of axonal degeneration and barrier damage after intrafascicular injection or topical application of bupivacaine (Marcain). Acta Anaesth Scand. 1979;23:127–36.CrossRefPubMed Selander D, Brattsand R, Lundborg G, Nordborg C, Olsson Y. Local anesthetics: importance of mode of application, concentration and adrenaline for the appearance of nerve lesions. An experimental study of axonal degeneration and barrier damage after intrafascicular injection or topical application of bupivacaine (Marcain). Acta Anaesth Scand. 1979;23:127–36.CrossRefPubMed
3.
Zurück zum Zitat Bigeleisen PE. Nerve puncture and apparent intraneural injection during ultrasound-guided axillary block does not invariably result in neurologic injury. Anesthesiology. 2006;105:779–83.CrossRefPubMed Bigeleisen PE. Nerve puncture and apparent intraneural injection during ultrasound-guided axillary block does not invariably result in neurologic injury. Anesthesiology. 2006;105:779–83.CrossRefPubMed
4.
5.
Zurück zum Zitat Vuckovic I, Hadzic A, Dilberovic F, et al. Detection of neurovascular structures using injection pressure in blockade of brachial plexus in rat. Bosn J Basic Med. 2005;5:79–85. Vuckovic I, Hadzic A, Dilberovic F, et al. Detection of neurovascular structures using injection pressure in blockade of brachial plexus in rat. Bosn J Basic Med. 2005;5:79–85.
6.
Zurück zum Zitat Hadzic A, Dilberovic F, Shah S, et al. Combination of intraneural injection and high injection pressure leads to fascicular injury and neurologic deficits in dogs. Reg Anesth Pain Med. 2004;29:417–23.CrossRefPubMed Hadzic A, Dilberovic F, Shah S, et al. Combination of intraneural injection and high injection pressure leads to fascicular injury and neurologic deficits in dogs. Reg Anesth Pain Med. 2004;29:417–23.CrossRefPubMed
7.
Zurück zum Zitat Theron PS, Mackay Z, Gonzalez JG, Donaldson N, Blanco R. An animal model of “syringe feel” during peripheral nerve block. Reg Anesth Pain Med. 2009;34:330–2.CrossRefPubMed Theron PS, Mackay Z, Gonzalez JG, Donaldson N, Blanco R. An animal model of “syringe feel” during peripheral nerve block. Reg Anesth Pain Med. 2009;34:330–2.CrossRefPubMed
8.
Zurück zum Zitat Claudio R, Hadzic A, Shih H, et al. Injection pressures by anesthesiologists during simulated peripheral nerve block. Reg Anesth Pain Med. 2004;29:201–5.CrossRefPubMed Claudio R, Hadzic A, Shih H, et al. Injection pressures by anesthesiologists during simulated peripheral nerve block. Reg Anesth Pain Med. 2004;29:201–5.CrossRefPubMed
11.
Zurück zum Zitat Choyce A, Chan VW, Middleton WJ, Knight PR, Peng P, McCartney CJ. What is the relationship between paresthesia and nerve stimulation for axillary brachial plexus block? Reg Anesth Pain Med. 2001;26:100–4.CrossRefPubMed Choyce A, Chan VW, Middleton WJ, Knight PR, Peng P, McCartney CJ. What is the relationship between paresthesia and nerve stimulation for axillary brachial plexus block? Reg Anesth Pain Med. 2001;26:100–4.CrossRefPubMed
12.
Zurück zum Zitat Urmey WF, Stanton J. Inability to consistently elicit a motor response following sensory paresthesia during interscalene block administration. Anesthesiology. 2002;96:552–4.CrossRefPubMed Urmey WF, Stanton J. Inability to consistently elicit a motor response following sensory paresthesia during interscalene block administration. Anesthesiology. 2002;96:552–4.CrossRefPubMed
15.
Zurück zum Zitat Chan VW, Brull R, McCartney CJ, Xu D, Abbas S, Shannon P. An ultrasonographic and histological study of intraneural injection and electrical stimulation in pigs. Anesth Analg. 2007;104:1281–4.CrossRefPubMed Chan VW, Brull R, McCartney CJ, Xu D, Abbas S, Shannon P. An ultrasonographic and histological study of intraneural injection and electrical stimulation in pigs. Anesth Analg. 2007;104:1281–4.CrossRefPubMed
16.
Zurück zum Zitat Cohen JM, Gray AT. Functional deficits after intraneural injection during interscalene block. Reg Anesth Pain Med. 2010;35:397–9.CrossRefPubMed Cohen JM, Gray AT. Functional deficits after intraneural injection during interscalene block. Reg Anesth Pain Med. 2010;35:397–9.CrossRefPubMed
17.
Zurück zum Zitat Russon K, Blanco R. Accidental intraneural injection into the musculocutaneous nerve visualized with ultrasound. Anesth Analg. 2007;105:1504–5.CrossRefPubMed Russon K, Blanco R. Accidental intraneural injection into the musculocutaneous nerve visualized with ultrasound. Anesth Analg. 2007;105:1504–5.CrossRefPubMed
19.
Zurück zum Zitat Abrahams MS, Aziz MF, Fu RF, Horn JL. Ultrasound guidance compared with electrical neurostimulation for peripheral nerve block: a systematic review and meta-analysis of randomised controlled trials. Br J Anaesth. 2009;102:408–17. doi:10.1093/bja/aen384 Epub 2009 Jan 26.CrossRefPubMed Abrahams MS, Aziz MF, Fu RF, Horn JL. Ultrasound guidance compared with electrical neurostimulation for peripheral nerve block: a systematic review and meta-analysis of randomised controlled trials. Br J Anaesth. 2009;102:408–17. doi:10.​1093/​bja/​aen384 Epub 2009 Jan 26.CrossRefPubMed
21.
Zurück zum Zitat Liu SS, Zayas VM, Gordon MA, et al. A prospective, randomized, controlled trial comparing ultrasound versus nerve stimulator guidance for interscalene block for ambulatory shoulder surgery for postoperative neurological symptoms. Anesth Analg. 2009;109:265–71. doi:10.1213/ane.0b013e3181a3272c.CrossRefPubMed Liu SS, Zayas VM, Gordon MA, et al. A prospective, randomized, controlled trial comparing ultrasound versus nerve stimulator guidance for interscalene block for ambulatory shoulder surgery for postoperative neurological symptoms. Anesth Analg. 2009;109:265–71. doi:10.​1213/​ane.​0b013e3181a3272c​.CrossRefPubMed
24.
Zurück zum Zitat Schafhalter-Zoppoth I, Zeitz ID, Gray AT. Inadvertent femoral nerve impalement and intraneural injection visualized by ultrasound. Anesth Analg. 2004;99:627–8.CrossRefPubMed Schafhalter-Zoppoth I, Zeitz ID, Gray AT. Inadvertent femoral nerve impalement and intraneural injection visualized by ultrasound. Anesth Analg. 2004;99:627–8.CrossRefPubMed
Metadaten
Titel
An improvised pressure gauge for regional nerve blockade/anesthesia injections: an initial study
verfasst von
Jayaprakash Patil
Hari Ankireddy
Antony Wilkes
David Williams
Michael Lim
Publikationsdatum
01.12.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 6/2015
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9701-z

Weitere Artikel der Ausgabe 6/2015

Journal of Clinical Monitoring and Computing 6/2015 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.