Open Access 01.12.2017 | Case report

An inappropriate pacing threshold increase after repeated electrical storm in a patient with implantable cardioverter defibrillator

verfasst von: Ye Zhu, Xiang Gu, Chao Xu

Erschienen in: BMC Cardiovascular Disorders | Ausgabe 1/2017

Abstract

Background

Implantable cardioverter defibrillators (ICD) are capable of effectively terminating malignant ventricular arrhythmia and are the most effective way to prevent sudden cardiac death. However, some evidences demonstrated that both anti-tachycardia pacing (ATP) and ICD shock can also bring adverse prognosis.

Case presentation

A 66-year-old Han Chinese man with prior ICD implantation was admitted to our hospital because of frequent ICD shocks. Although intravenous amiodarone and esmolol succinate were administered daily, the patient suffered 155 episodes of VT/VF during 8 weeks after implantation. After repeated discharge of the device, the pacing threshold of the patient increased gradually. Considering the inappropriate increase of the pacing threshold, we decided to reposition the right ventricular (RV) lead with good sensing and threshold parameters confirmed. Subsequent 22 months interrogation follow-up revealed a stable lead position and electrical specifications. Furthermore, antiarrhythmic drugs were maximally increased, while ATP burst was remarkably decreased and the inappropriate ICD shock never occurred until now.

Conclusion

An inappropriate pacing threshold was increased secondary to repeated ICD electrical storm. A timely active lead position adjustment reduced the pacing threshold and eliminated the risk of premature battery depletion.

Background

Implantable cardioverter defibrillator (ICD) can effectively terminate malignant ventricular tachycardia (VT)/ventricular fibrillation (VF) and prevent sudden death [13]. Electrical storm (ES), characterized by multiple attacks of VT or ventricular fibrillation (VF), is an unstable condition for which management and prevention are still a challenge. Whether ES is a causal factor or only an epiphenomenon remains unclear, although it is still a matter for debate that repeated shocks may provoke myocardial damage and result in further deterioration of the underlying disorders [4].
We report an ICD patient with severe ES caused by frequent VT/VF refractory to antiarrhythmic agents. The continuous discharge of the ICD might be related to an inappropriate increase of pacing threshold and it necessitated a repositioning of a lead.

Case presentation

A 66-year-old Han Chinese male was admitted after several VT attacks over the course of 1 month. The patient was diagnosed with rheumatic heart disease and impaired left ventricular ejection fraction (LVEF) of 42% in a functional New York Heart Association stage III, suffering from VT/VF and atrial fibrillation. He underwent surgery of the mitral valve and an aortic valve replacement 10 years ago. Electrolytic measurements showed sodium, calcium, and potassium values were within the normal range. He was treated with aldosterone antagonist, angiotensin converting enzyme inhibitor, lidocaine, β-blocker, and amiodarone, but VT recurred despite the optimized drug treatment. Additional doses of antiarrhythmic drugs were administered but unsuccessful. The patient was placed on amiodarone 200 mg orally once a day and metoprolol 25 mg orally four times a day. Attempts to reduce the frequency of VT using optimized antiarrhythmic drugs failed. VT could only be terminated by external cardioversion. The patient satisfied the requirement for ICD therapy according to the criteria of the Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT), [5] so a single chamber ICD (Model 1231–40, St. Jude Medical Cardiac Rhythm Management, USA) was implanted with lead in the right ventricular (RV) apex after symptom stabilization (Fig. 1a). During the implantation, electrical specifications were confirmed, including the R-wave of 9.0 mV, the slew rate of 1.7 V/s, the impedance of 574 Ohm, and pacing threshold of 0.6 V/0.5 ms.
Two days later, the patient was abruptly awakened by an episode of VT/VF. The initial administration of the ATP therapy through the ICD was unsuccessful. VF was finally terminated by a 36 J ICD shock. Frequent ICD shocks were associated with anxiety and intolerance of ICD therapy. The patient was thin and weak. ATP and Low energy cardioversion (15/25 J) were successfully used to relieve the patient’s pain, anxiety and possible myocardial injury (Fig. 2-a, b, c). Overall, the patient experienced 155 episodes of VT/VF and received a total of 27 shocks after the first implantation of ICD. The most severe attack was 34 VT/VF episodes within 24 h. Rarely, a 36.0 J shock at last reverted VT, which was not terminated by ATP burst and a 15.0 J/25 J shock (Fig. 2-d). Frequent ICD discharges depleted the battery life considerably. His treatment was aggressively escalated. Catheter ablation was contraindicated because of the patient’s metal valve and multiple original VTs. Renal artery denervation was recommended but refused. Attempts to reduce the frequency of VT only using antiarrhythmic drugs, such as lidocaine, were unsuccessful. The combination of ATP and intravenous esmolol seemed to reduce, but did not fully eliminate the episodes of VT.
After repeated discharge of the device for 8 weeks, the capture threshold was gradually increased. Initial threshold of device interrogation was respectively 2.37 V/0.5 ms, 2.75 V/0.5 ms, and 3.75 V/1.0 ms in 2, 4 and 9 weeks post- implantation (Additional file 1). Thorax X-ray at 9 weeks post-implant did not show apparent dislocation of the lead (Fig. 1b). Considering the patient’s reliance on the pacemaker, we decided to reposition the RV lead to better match the ventricular myocardium in the threshold field as a final alternative (Fig. 1c). Fortunately, this remedy did lower the pacing threshold. Lead measurements showed the ventricular sensing at 11.3 mV, a pacing threshold of 1.4 V/0.5 ms, and an impedance of 730 Ohms. The further clinical course of the patient was uncomplicated and he was discharged in good clinical status after the surgery. Recent device interrogation respectively showed a ventricular sensing of 8.9 mV, 7.9 mV and 9.1 mV, a pacing threshold of 1.75 V/0.5 ms, 1.5 V/1.0 ms and 1.5 V/1.0 ms at 13, 20, 24 months post-implantation follow-up (Additional file 1). Amiodarone 200 mg qd and metoprolol 25 mg q6h were given to enhance the efficacy of anti-arrhythmias. The patient’s pacing threshold and ventricular arrhythmias have carefully being monitored in the long term. Until now, the shock never occurred again and ATP burst significantly decreased.

Discussion and conclusions

This case highlights a rare clinical situation: an increased pacing threshold may be increased by repeated ES in a patient with an ICD.
In the era of ICD therapy, ES has become increasingly common. It is generally defined as the occurrence of three or more obvious episodes of VT /VF in 24 h, separated by bouts of normal rhythm after successful therapy, either ATP burst or shock [6]. Previous studies have shown that ES occurs in about 10 – 28% of ICD recipients and is associated with increased mortality [7]. The first therapy for ES is eliminating the risk factors as soon as possible, then following a positive comprehensive management protocol [8]. Structural heart disease, electrolyte imbalance, myocardial infarction, inherited arrhythmic syndrome and other factors could all lead to ES [9]. In this case, VT was definitively identified on admission. The patient had a history of heart failure, rheumatic heart disease, and valve replacement surgery. He did not have diabetes, hyperlipidemia, a history of smoking or any other risk factors for coronary heart disease. Risk factors for acidosis and electrolyte disorders were also excluded [10]. The patient had considerable VTs during the perioperative period of ICD implantation. ES before ICD implantation may cause myocardial damage, and the placement of the ICD electrode may further damaged the local myocardium. The inappropriate discharge during 8 weeks after ICD implantation was attributed to myocardial injury caused by ES before ICD implantation. Additionally, the occurrence of ES seriously affected the patient’s quality of life and induced psychological symptoms including anxiety and fear. This programming strategy should be minimal energy of the defibrillation and effectively terminate VT/VF. However, some patients at high risk for sudden cardiac death still require the maximum energy that ICDs can achieve. For such patients, a readjustment of the RV lead position is a successful strategy to lower the threshold [11].
Overdrive pacing, as implemented in this case, has been reported to stabilize an ES [12]. Comprehensive treatment of VT such as increasing the identifying frequency of VT, optimal medication, trigger or substrate ablation and denervation should be administered to reduce or eliminate false identification and inappropriate ICD discharges [13]. As illustrated in this case, the management of ES hinges on excluding every possible causative factor, no matter how rare. Catheter ablation appears to be effective in both short-term therapy and prevention of ES. Catheter ablation of the left ventricular was contraindicated because of the patient’s metal valve. Renal artery denervation was offered but refused. Amiodarone and beta­blockers might lower the incidence of shock therapy in patients with ICD. Optimal programming aimed at reducing the burden of ICD therapy without an increase in adverse outcomes [14]. For this case, post-implantation inspection showed that the pacing threshold gradually increased. Thorax X-ray at 9 weeks post-implant did not show apparent dislocation of the lead. However, a lead tiny dislodgement seldom be excluded or confirmed. The repeated ES may have facilitated the development of heart failure and myocardium injury, which in turn increased the pacing threshold and shortened the life of the device. We took into consideration the fact that RV lead repositioning is a surgical procedure that exposes the patient to several risks. RV-ICD lead position adjustment can also be successfully performed to avoid the pacing threshold increase. Low pacing threshold is not only critical for the safety and effectiveness of the ICD therapy, but also can extend the life of the device. We made several efforts, invasively and noninvasively, to decrease this patient’s pacing threshold, including lead repositioning and appropriate ICD adjustment [15]. Here, the positioning of the RV lead was used as a final remedy. The 22 months follow-up after the second ICD implantation showed a steady position of the RV lead and stable parameters for sensing and pacing. The inappropriate shock never occurred until now. However, lead repositioning may not lead to ES cessation. ES cessation was achieved by adjusting the administration and dosage of antiarrhythmic agents and myocardium rehabilitation.
In this case, the attack of VT/VF was safely terminated by maximal post-procedural administration of antiarrhythmic drugs, subsequent application of ATP and discharge by the ICD. The follow-up showed that an inappropriate pacing threshold by repeated ES could not be fully eliminated. A timely active lead position adjustment reduced the pacing threshold and eliminated the risk of premature battery depletion. A programmable strategy should be personally adjusted as the low, middle and high cardioversion energy for effectively terminating VT/VF and lessening myocardial injury secondary to ES.

Acknowledgements

The authors would like to appreciate the assistance of Loula Burton, Office of Research at Tulane University, in polishing the English manuscript. We are grateful to Dr.Lei Sun, Dr.Yi Zhang, Dr. Zhengyu Bao and Dr. Jianhua Shen for their valuable advice on the management of the patient. We thank the editor and the reviewers whose comments/suggestions helped improve and clarify this manuscript.

Funding

This work was supported by funds from science and technology department of Jiangsu province, China (No. BL2013022).

Availability of data and materials

All the data supporting our findings is contained within the manuscript.
The study was performed in accordance with the Declaration of Helsinki and approved by the institutional ethical board of Northern Jiangsu Province People’s Hospital, Jiangsu Province, China.
Written informed consent for the publication was obtained from the patient for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Chang PM, Powell BD, Jones PW, Carter N, Hayes DL, Saxon LA. Implantable Cardioverter defibrillator programming characteristics, shocked rhythms, and survival among patients under thirty years of age. J Cardiovasc Electrophysiol. 2016;27(10):1183–90.CrossRefPubMed Chang PM, Powell BD, Jones PW, Carter N, Hayes DL, Saxon LA. Implantable Cardioverter defibrillator programming characteristics, shocked rhythms, and survival among patients under thirty years of age. J Cardiovasc Electrophysiol. 2016;27(10):1183–90.CrossRefPubMed
2.
Zurück zum Zitat Jimenez-Candil J, Hernandez J, Martin A, Morinigo J, Perdiguero P, Bravo L, Ruiz S, Sanchez PL. Differences in ventricular tachyarrythmias and antitachycardia pacing effectiveness according to the ICD indication (primary versus secondary prevention): an analysis based on the stored electrograms. J Interv Card Electrophysiol. 2015;44(2):187–95.CrossRefPubMed Jimenez-Candil J, Hernandez J, Martin A, Morinigo J, Perdiguero P, Bravo L, Ruiz S, Sanchez PL. Differences in ventricular tachyarrythmias and antitachycardia pacing effectiveness according to the ICD indication (primary versus secondary prevention): an analysis based on the stored electrograms. J Interv Card Electrophysiol. 2015;44(2):187–95.CrossRefPubMed
3.
Zurück zum Zitat Malagu M, Ferri A, Mancuso O, Trevisan F, Nardozza M, Bertini M. Implantable cardioverter defibrillator management: an update. Futur Cardiol. 2016;12(6):673–88.CrossRef Malagu M, Ferri A, Mancuso O, Trevisan F, Nardozza M, Bertini M. Implantable cardioverter defibrillator management: an update. Futur Cardiol. 2016;12(6):673–88.CrossRef
4.
Zurück zum Zitat Nedios S, Darma A, Stevanello C, Richter S, Doering M, Rolf S, Sommer P, Hindricks G, Bollmann A, Arya A. Electrical storm in patients with implantable cardioverter-defibrillator in the era of catheter ablation: implications for better rhythm control. Heart Rhythm. 2015;12(12):2419–25.CrossRefPubMed Nedios S, Darma A, Stevanello C, Richter S, Doering M, Rolf S, Sommer P, Hindricks G, Bollmann A, Arya A. Electrical storm in patients with implantable cardioverter-defibrillator in the era of catheter ablation: implications for better rhythm control. Heart Rhythm. 2015;12(12):2419–25.CrossRefPubMed
5.
Zurück zum Zitat Au-Yeung WT, Reinhall PG, Poole JE, Anderson J, Johnson G, Fletcher RD, Moore HJ, Mark DB, Lee KL, Bardy GH. SCD-HeFT: use of R-R interval statistics for long-term risk stratification for arrhythmic sudden cardiac death. Heart Rhythm. 2015;12(10):2058–66.CrossRefPubMedPubMedCentral Au-Yeung WT, Reinhall PG, Poole JE, Anderson J, Johnson G, Fletcher RD, Moore HJ, Mark DB, Lee KL, Bardy GH. SCD-HeFT: use of R-R interval statistics for long-term risk stratification for arrhythmic sudden cardiac death. Heart Rhythm. 2015;12(10):2058–66.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Looi KL, Tang A, Agarwal S. Ventricular arrhythmia storm in the era of implantable cardioverter-defibrillator. Postgrad Med J. 2015;91(1079):519–26.CrossRefPubMed Looi KL, Tang A, Agarwal S. Ventricular arrhythmia storm in the era of implantable cardioverter-defibrillator. Postgrad Med J. 2015;91(1079):519–26.CrossRefPubMed
7.
Zurück zum Zitat Ginwalla M, Battula S, Dunn J, Lewis WR. Termination of electrocution-induced ventricular fibrillation by an implantable cardioverter defibrillator. Pacing Clin Electrophysiol. 2010;33(4):510–2.CrossRefPubMed Ginwalla M, Battula S, Dunn J, Lewis WR. Termination of electrocution-induced ventricular fibrillation by an implantable cardioverter defibrillator. Pacing Clin Electrophysiol. 2010;33(4):510–2.CrossRefPubMed
8.
Zurück zum Zitat Hohnloser SH, Al-Khalidi HR, Pratt CM, Brum JM, Tatla DS, Tchou P, Dorian P. Electrical storm in patients with an implantable defibrillator: incidence, features, and preventive therapy: insights from a randomized trial. Eur Heart J. 2006;27(24):3027–32.CrossRefPubMed Hohnloser SH, Al-Khalidi HR, Pratt CM, Brum JM, Tatla DS, Tchou P, Dorian P. Electrical storm in patients with an implantable defibrillator: incidence, features, and preventive therapy: insights from a randomized trial. Eur Heart J. 2006;27(24):3027–32.CrossRefPubMed
9.
Zurück zum Zitat Camsari A, Yilmaz DC, Ozcan T. Inappropriate shock delivery by implantable cardioverter defibrillator due to electrical leak from bathroom tap. Clin Res Cardiol. 2010;99(11):761–3.CrossRefPubMed Camsari A, Yilmaz DC, Ozcan T. Inappropriate shock delivery by implantable cardioverter defibrillator due to electrical leak from bathroom tap. Clin Res Cardiol. 2010;99(11):761–3.CrossRefPubMed
10.
Zurück zum Zitat Lee WC, Chen HC, Chen YL, Tsai TH, Pan KL, Lin YS, Chen MC. Left ventricle remodeling predicts the recurrence of ventricular tachyarrhythmias in implantable cardioverter defibrillator recipients for secondary prevention. BMC Cardiovasc Disord. 2016;16(1):231.CrossRefPubMedPubMedCentral Lee WC, Chen HC, Chen YL, Tsai TH, Pan KL, Lin YS, Chen MC. Left ventricle remodeling predicts the recurrence of ventricular tachyarrhythmias in implantable cardioverter defibrillator recipients for secondary prevention. BMC Cardiovasc Disord. 2016;16(1):231.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Cesario D, Bhargava M, Valderrabano M, Fonarow GC, Wilkoff B, Shivkumar K. Azygos vein lead implantation: a novel adjunctive technique for implantable cardioverter defibrillator placement. J Cardiovasc Electrophysiol. 2004;15(7):780–3.CrossRefPubMed Cesario D, Bhargava M, Valderrabano M, Fonarow GC, Wilkoff B, Shivkumar K. Azygos vein lead implantation: a novel adjunctive technique for implantable cardioverter defibrillator placement. J Cardiovasc Electrophysiol. 2004;15(7):780–3.CrossRefPubMed
12.
Zurück zum Zitat Kurisu S, Inoue I, Kawagoe T, Ishihara M, Shimatani Y, Mitsuba N, Hata T, Nakama Y, Kisaka T, Kijima Y. Temporary overdriving pacing as an adjunct to antiarrhythmic drug therapy for electrical storm in acute myocardial infarction. Circ J. 2005;69(5):613–6.CrossRefPubMed Kurisu S, Inoue I, Kawagoe T, Ishihara M, Shimatani Y, Mitsuba N, Hata T, Nakama Y, Kisaka T, Kijima Y. Temporary overdriving pacing as an adjunct to antiarrhythmic drug therapy for electrical storm in acute myocardial infarction. Circ J. 2005;69(5):613–6.CrossRefPubMed
13.
Zurück zum Zitat Gu X, Zhu Y, Sun L, Zhang Y, Cao K, Cheng J. Impact of rate control during atrial fibrillation on the proarrhythmic and antiarrhythmic effects of biventricular pacing. Int J Cardiol. 2012;159(2):162–3.CrossRefPubMed Gu X, Zhu Y, Sun L, Zhang Y, Cao K, Cheng J. Impact of rate control during atrial fibrillation on the proarrhythmic and antiarrhythmic effects of biventricular pacing. Int J Cardiol. 2012;159(2):162–3.CrossRefPubMed
14.
Zurück zum Zitat Tan VH, Wilton SB, Kuriachan V, Sumner GL, Exner DV. Impact of programming strategies aimed at reducing nonessential implantable cardioverter defibrillator therapies on mortality: a systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2014;7(1):164–70.CrossRefPubMed Tan VH, Wilton SB, Kuriachan V, Sumner GL, Exner DV. Impact of programming strategies aimed at reducing nonessential implantable cardioverter defibrillator therapies on mortality: a systematic review and meta-analysis. Circ Arrhythm Electrophysiol. 2014;7(1):164–70.CrossRefPubMed
15.
Zurück zum Zitat Grimard C, May MA, Mabo P, Babuty D. An original defibrillation lead implantation to avoid tricuspid prosthesis damage. Europace. 2010;12(4):589–90.CrossRefPubMed Grimard C, May MA, Mabo P, Babuty D. An original defibrillation lead implantation to avoid tricuspid prosthesis damage. Europace. 2010;12(4):589–90.CrossRefPubMed
Metadaten
Titel
An inappropriate pacing threshold increase after repeated electrical storm in a patient with implantable cardioverter defibrillator
verfasst von
Ye Zhu
Xiang Gu
Chao Xu
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
BMC Cardiovascular Disorders / Ausgabe 1/2017
Elektronische ISSN: 1471-2261
DOI
https://doi.org/10.1186/s12872-017-0695-y

Weitere Artikel der Ausgabe 1/2017

BMC Cardiovascular Disorders 1/2017 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.