Anatomical and biomechanical evaluation of the tension band technique in patellar fractures

verfasst von: Onder Baran, Metin Manisali, Berivan Cecen

Erschienen in: International Orthopaedics | Ausgabe 4/2009

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Tension band wiring for patellar fractures is common, but some recent reports refer to disadvantages of this approach. Our anatomical and biomechanical study focused on use of tension band techniques in patellar fractures. The anatomy of the patella and tendon insertion was examined with knee magnetic resonance imaging (MRI) and correlated with the technical requirements of the tension band. Tension band wiring over tendinous tissue was simulated and calculated with a cyclic biomechanical test on cow patellae. According to tension band templating on the MRI section, Kirschner wire insertion was needed for the tension band to turn over the tendinous tissue. The tension band became more stable while turning over less tendinous tissue and more adjacent bone surface. Nevertheless, cyclic loading tests indicate that all tension band applications in this study lose their initial stability. Excessive initial compression by the tension band resulted in bending of the Kirschner wire and thus reduction failure. For optimum stabilisation, tension force transfer should be done directly on bone or at least material that protects the tendon would be useful.

Introduction

The tension band technique is mostly used in patellar fractures. There are some recent reports indicating a high failure rate (50%) of the modified tension band technique despite previous good results. Several modifications have been used to increase the stabilisation of the tension band technique along with other techniques [4, 7, 9].
Although difficulties concerning the modified tension band technique are described and defined in textbooks and the literature in detail, technical inadequacies are often underestimated. This study aimed to explore the relationship between patellar anatomy, tension band and Kirschner wiring. Although the tension band is the main factor in achieving stability, Kirschner wiring designates tension band projection on tendon and bone. First of all, we explored this relationship with a magnetic resonance imaging (MRI) study. With the data obtained, technical requirements and anatomical and dynamic specifications were tested and assessed.

Material and methods

Anatomical study

Patellar variation was evaluated with knee MRI studies. Sagittal sections of patellae that pass the thickest point were obtained from 30 knee MRI studies; patellar diameter, tendon insertion, localisation and thickness were measured (Fig. 1).
The modified tension band technique was used according to the AO Manual of Internal Fixation. Templating the tension band technique was performed on MRI (Kirschner wire drawn 5–6 mm posteriorly from the anterior surface of the patella) (Fig. 2). Kirschner wiring position in relation to tendinous tissue was evaluated with three parameters: outside-anterior (A), inside-median (M) and outside-posterior (P).

Biomechanical study

Seven cow patellae with tendinous tissue were used as biomechanical test material. Cow patellae were cut transversely 2 cm from the pole to simulate human patellar fractures. Prepared cow patellar fragments were fixed on a traction apparatus by two 2-mm Kirschner wires, which passed 5 mm posteriorly from the anterior surface of the patella and through tendinous tissue. With a 1.2-mm tension band wire, four configurations were simulated: vertically oriented figure-of-eight passed anteriorly to the tendon (a), horizontally oriented figure-of-eight (b), vertically oriented figure-of-eight passed posteriorly to the tendon (c) and to eliminate the effect of the tendon a control group (d) was established as vertically oriented figure-of-eight without tendinous tissue. Figure 3 shows these orientations.
Distractions on the tension band wire were performed by a compression-distraction test device (AG-I 10 kN, Shimadzu, Kyoto, Japan) with 3 mm/min traction speed to a maximum load of 200 N and then ten cyclic loadings were performed between 100 and 200 N (Fig. 4).
The data on force-displacement variation were evaluated with the Kruskal-Wallis test using SPSS 11.0 for Windows. Statistical analyses were performed using a Mann-Whitney U test to estimate group differences.

Results

Knee MRI sagittal sections of 30 patients were obtained and evaluated. Patellar size measurements were taken with the following findings: mean patellar lengthening 34 ± 3 mm (30.6–42.9 mm), mean patellar thickness 19.6 ± 2.3 mm (15.5–26.1 mm), mean patellar tendon insertion thickness 4.8 ± 1 mm (3–6.8 mm) and mean quadriceps tendon insertion thickness 6.5 ± 1 mm (3.9–8.1 mm). The correlation between patellar length and thickness was 0.70, but a low correlation was observed between tendon thickness and patellar diameter (Table 1). Templating the modified tension band technique on the MRI slide indicated that all Kirschner wiring has to pass through tendinous soft tissue.
Table 1
MRI results for 30 patellae: patellar length (PL), patellar thickness (PT), quadriceps tendon thickness (QTt), patellar tendon thickness (PTt) and correlation between them
 
PL
PT
QTt
PTt
PL/PT
PTt/PT
QTt/PT
1
34.5
21.6
5.7
4.8
1.597222
0.222222
0.263889
2
35.2
19.4
6.8
5.9
1.814433
0.304124
0.350515
3
37.9
18.9
3.9
3.3
2.005291
0.174603
0.206349
4
36.6
22.9
6.8
5.7
1.598253
0.248908
0.296943
5
35.6
20.3
6.7
3.3
1.753695
0.162562
0.330049
6
35.2
16
6
4.5
2.2
0.28125
0.375
7
33.9
18
6.9
5.4
1.883333
0.3
0.383333
8
33.9
19.2
6.8
6.1
1.765625
0.317708
0.354167
9
42.9
26.1
8
6.8
1.643678
0.260536
0.306513
10
35.2
17.1
5.8
4.7
2.05848
0.274854
0.339181
11
31.8
17.6
6.9
3.9
1.806818
0.221591
0.392045
12
31.7
20
5.4
5.7
1.585
0.285
0.27
13
30.6
16.2
6.3
3.9
1.888889
0.240741
0.388889
14
31
15.5
8.1
4.8
2
0.309677
0.522581
15
38.8
22.8
5.9
3.6
1.701754
0.157895
0.258772
16
33.5
20.8
8.1
6
1.610577
0.288462
0.389423
17
33.9
19.3
4.3
3.6
1.756477
0.186528
0.222798
18
37.5
19.7
6.8
5.7
1.903553
0.28934
0.345178
19
40.3
20.3
5.7
4.8
1.985222
0.236453
0.280788
20
39.9
23.6
5.5
5.5
1.690678
0.233051
0.233051
21
32.8
19
6.8
4.3
1.726316
0.226316
0.357895
22
30.6
17.6
6.5
3
1.738636
0.170455
0.369318
23
31.4
19.5
7.6
5.4
1.610256
0.276923
0.389744
24
33.2
19.5
6.8
4.7
1.702564
0.241026
0.348718
25
34.6
19.5
7.2
6.5
1.774359
0.333333
0.369231
26
33.9
19.7
6.8
4.8
1.720812
0.243655
0.345178
27
33.6
20.5
7.8
6.3
1.639024
0.307317
0.380488
28
35.3
19.9
7.2
4.2
1.773869
0.211055
0.361809
29
33.9
17.9
6.5
4.5
1.893855
0.251397
0.363128
30
31.9
19.4
5.1
3.6
1.64433
0.185567
0.262887
Mean
34.70333
19.59333
6.49
4.843333
1.782433
0.248085
0.335262
Standard deviation
3.001549
2.267603
1.024308
1.0431
0.15495
0.049492
0.065327
Correlation
    
0.696617
0.397679
0.061432
Biomechanical study results are shown in Table 2. At the first 200 N loading, the mean displacement in method 1 was 0.23 mm (SD: 0.07), in method 2 it was 0.16 mm (SD: 0.14), in method 3 it was 0.23 mm (SD: 0.06) and in method 4 it was 0.14 mm (SD: 0.04). The mean displacement after the cyclic loading test in method 1 was 0.89 mm (SD: 0.31), in method 2 it was 0.57 mm (SD: 0.17), in method 3 it was 0.46 mm (SD: 0.11) and in method 4 it was 0.29 mm (SD: 0.09).
Table 2
Tension band cyclic loading test results
Tension band method
First displacement
Total displacement
A
0.230 ± 0.070
0.893 ± 0.313
B
0.228 ± 0.056
0.461 ± 0.114
C
0.164 ± 0.138
0.572 ± 0.170
D
0.141 ± 0.041
0.275 ± 0.086
Tension band configurations are vertically oriented figure-of-eight passing anteriorly to tendon (A), horizontally oriented figure-of-eight passing anteriorly to tendon (B), vertically oriented figure-of-eight passing posteriorly to tendon (C) and vertically oriented figure-of-eight without soft tissue (D). First displacement shows value is 200 N initial traction. Total displacement shows total displacement after ten cyclic loadings
The results indicate that there was a significant difference between all groups compared with the Kruskal-Wallis test (p = 0.000). Using the Mann-Whitney test, there was a significant difference between group 1 and group 2 (p = 0.012), group 1 and group 3 (p = 0.04), group 1 and group 4 (p = 0.02) and group 2 and group 4 (p = 0.02).

Discussion

The concept of more rigid fixation in patellar fractures is controversial. Yet there are high complication rates with the tension band technique and early mobilisation requires more rigid fixation methods.
Several biomechanical studies and several modifications of the tension band technique indicate that the relationship between the tension band and soft tissue around the patella affect stability in treatment of patellar fractures [110]. The tension band technique is important for patellar fractures; yet these techniques are more functional in the knee flexion position. Horizontally oriented figure-of-eight for tension band configuration is found to be stronger than the vertically oriented ones, which our test results support [6]. Improved wire twisting and flexible tension band material are helpful for good force distribution, thus decreasing failure rate [6, 8]. Other techniques are effective in the knee extension position, but their combination with the tension band technique is best and most effective in any position [2, 3].
According to the templating study of the tension band technique on the MRI section, the Kirschner wire must pass through tendinous tissue and tension band material must turn over this soft tissue. Our anatomical study shows that there is no statistical correlation between patellar length and thickness to confirm similar knowledge in the literature.
It is a fact that the expected result of the tension band techniques is affected by many factors such as variety of patellae, surgical practice and traumatic damage. But the Kirschner wire end point is placed deeply in some practices so that more interposed tendinous tissue can diminish stability. Many treatment methods are discussed in the literature with clinical results and biomechanical studies, and numerous results indicate the importance of the relationship between tension band and soft tissue.
For the biomechanical study, we used cow patellae because fresh tendinous attachment was needed as test material. The cow patella is too big to correlate with a human patella so a biomechanical test was designed to use only one part of the cow patella. The pole part of the cow patella with tendinous insertion was prepared similar in length to a human patellar fragment. The test was performed on one part of the patella only to eliminate practical error, as we were only interested in the tension band and soft tissue relationship.
In our study, we preferred cyclic loading test at 100–200 N because loading at over 200 N causes Kirschner wire bending. If this test were repeated in the flexion position, the Kirschner wire could not bend so we could reach a physiological loading force. Besides this, compression on the tension band is initially too high to prevent the progressive instability resulting in loss of reduction and angulation deformities. In our study, the tension band adjacent to bone surface had more stability than the tension band technique turned over tendinous tissue, but cyclic loading tests have shown that all tension band applications in this study lost their initial stability.

Conclusions

In conclusion, it is well known that the tension band is the main stability factor in patellar fractures in flexion. More stability and durability can be achieved in patellar fracture treatment, by transferring the tension force directly on to bone or at least choosing material that protects the tendon.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Benjamin J, Bried J, Dohm M, McMurtry M (1987) Biomechanical evaluation of various forms of fixation of transverse patellar fractures. J Orthop Trauma 1:219–222PubMedCrossRef Benjamin J, Bried J, Dohm M, McMurtry M (1987) Biomechanical evaluation of various forms of fixation of transverse patellar fractures. J Orthop Trauma 1:219–222PubMedCrossRef
2.
Zurück zum Zitat Berg EE (1997) Open reduction internal fixation of displaced transverse patella fractures with figure-eight wiring through parallel cannulated compression screws. J Orthop Trauma 11:573–576PubMedCrossRef Berg EE (1997) Open reduction internal fixation of displaced transverse patella fractures with figure-eight wiring through parallel cannulated compression screws. J Orthop Trauma 11:573–576PubMedCrossRef
3.
Zurück zum Zitat Burvant JG, Thomas KA, Alexander R, Harris MB (1994) Evaluation of methods of internal fixation of transverse patella fractures: a biomechanical study. J Orthop Trauma 8:147–153PubMedCrossRef Burvant JG, Thomas KA, Alexander R, Harris MB (1994) Evaluation of methods of internal fixation of transverse patella fractures: a biomechanical study. J Orthop Trauma 8:147–153PubMedCrossRef
4.
Zurück zum Zitat Carpenter JE, Kasman RA, Patel N, Lee ML, Goldstein SA (1997) Biomechanical evaluation of current patella fracture fixation techniques. J Orthop Trauma 11:351–356PubMedCrossRef Carpenter JE, Kasman RA, Patel N, Lee ML, Goldstein SA (1997) Biomechanical evaluation of current patella fracture fixation techniques. J Orthop Trauma 11:351–356PubMedCrossRef
5.
Zurück zum Zitat Curtis MJ (1990) Internal fixation for fractures of the patella. A comparison of two methods. J Bone Joint Surg Br 72:280–282PubMed Curtis MJ (1990) Internal fixation for fractures of the patella. A comparison of two methods. J Bone Joint Surg Br 72:280–282PubMed
6.
Zurück zum Zitat John J, Wagner WW, Kuiper JH (2007) Tension-band wiring of transverse fractures of patella. The effect of site of wire twists and orientation of stainless steel wire loop: a biomechanical investigation. Int Orthop 31:703–707PubMedCrossRef John J, Wagner WW, Kuiper JH (2007) Tension-band wiring of transverse fractures of patella. The effect of site of wire twists and orientation of stainless steel wire loop: a biomechanical investigation. Int Orthop 31:703–707PubMedCrossRef
7.
Zurück zum Zitat Labitzke R (1997) Proper and improper tension band fixation exemplified by patellar fracture (in German). Chirurg 68:638–642PubMedCrossRef Labitzke R (1997) Proper and improper tension band fixation exemplified by patellar fracture (in German). Chirurg 68:638–642PubMedCrossRef
8.
Zurück zum Zitat Scilaris TA, Grantham JL, Prayson MJ, Marshall MP, Hamilton JJ, Williams JL (1998) Biomechanical comparison of fixation methods in transverse patella fractures. J Orthop Trauma 12:356–359PubMedCrossRef Scilaris TA, Grantham JL, Prayson MJ, Marshall MP, Hamilton JJ, Williams JL (1998) Biomechanical comparison of fixation methods in transverse patella fractures. J Orthop Trauma 12:356–359PubMedCrossRef
9.
Zurück zum Zitat Smith ST, Cramer KE, Karges DE, Watson JT, Moed BR (1997) Early complications in the operative treatment of patella fractures. J Orthop Trauma 11:183–187PubMedCrossRef Smith ST, Cramer KE, Karges DE, Watson JT, Moed BR (1997) Early complications in the operative treatment of patella fractures. J Orthop Trauma 11:183–187PubMedCrossRef
10.
Zurück zum Zitat Weber MJ, Janecki CJ, McLeod P, Nelson CL, Thompson JA (1980) Efficacy of various forms of fixation of transverse fractures of the patella. J Bone Joint Surg Am 62:215–220PubMed Weber MJ, Janecki CJ, McLeod P, Nelson CL, Thompson JA (1980) Efficacy of various forms of fixation of transverse fractures of the patella. J Bone Joint Surg Am 62:215–220PubMed
Metadaten
Titel
Anatomical and biomechanical evaluation of the tension band technique in patellar fractures
verfasst von
Onder Baran
Metin Manisali
Berivan Cecen
Publikationsdatum
01.08.2009
Verlag
Springer-Verlag
Erschienen in
International Orthopaedics / Ausgabe 4/2009
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-008-0602-3