Application of an advanced noise reduction algorithm for imaging of hands in rheumatic diseases: evaluation of image quality compared to standard-dose images

verfasst von: Katharina Ziegeler, Stefan Siepmann, Sandra Hermann, Alexander Beck, Alexander Lembcke, Bernd Hamm, Kay Geert A. Hermann

Erschienen in: Rheumatology International | Ausgabe 6/2020

Einloggen, um Zugang zu erhalten

Abstract

Objectives

X-ray is the fundamental imaging technique in both diagnosis and follow-up of rheumatic diseases. As patients often require sequential X-rays over many years, dose reduction is of great importance. New advanced noise reduction algorithms allow for a dose reduction of up to 50%. The aim of this study was to evaluate whether quality of low-dose images is non-inferior to standard-dose images and, therefore, application of this technique is possible in the context of imaging of rheumatic diseases.

Methods

A total of 298 patients with known or suspected rheumatic disease were enrolled prospectively into this study, separated into three consecutive groups: 80%, 64% and 50% tube charge reduction. All patients received imaging of one hand (laterality randomly assigned) with low-dose technique and imaging of the contralateral hand with standard-dose protocol. Images were evaluated by two independent readers who scored (on a scale of 1–5) the visualization of bony cortex, trabeculae and joint spaces of fingers and wrist separately. Additionally, soft tissue and overall contrast were evaluated on the same scale.

Results

Overall image quality (expressed by mean sum score out of 40) of the 50% low-dose images was 31.52 (SD 1.94) vs. 31.66 (SD 1.82) for standard images (p = 0.068). Bony contours as well as trabeculae were equally well visualized in both image sets. Median scores for soft tissue visualization was slightly lower for low dose compared to standard images [4 (IQR 3.5–4) vs. 4 (IQR 3.88–4); p = 0.001].

Conclusions

Overall image quality of low-dose images was not inferior to standard-dose images. Therefore, the application of low-dose technology based on advanced noise estimation algorithms in the context of imaging of rheumatic diseases is possible.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Smolen JS, Landewe R, Breedveld FC, Dougados M, Emery P, Gaujoux-Viala C, Gorter S, Knevel R, Nam J, Schoels M, Aletaha D, Buch M, Gossec L, Huizinga T, Bijlsma JW, Burmester G, Combe B, Cutolo M, Gabay C, Gomez-Reino J, Kouloumas M, Kvien TK, Martin-Mola E, McInnes I, Pavelka K, van Riel P, Scholte M, Scott DL, Sokka T, Valesini G, van Vollenhoven R, Winthrop KL, Wong J, Zink A, van der Heijde D (2010) EULAR recommendations for the management of rheumatoid arthritis with synthetic and biological disease-modifying antirheumatic drugs. Ann Rheum Dis 69(6):964–975. https://doi.org/10.1136/ard.2009.126532 CrossRefPubMedPubMedCentral Smolen JS, Landewe R, Breedveld FC, Dougados M, Emery P, Gaujoux-Viala C, Gorter S, Knevel R, Nam J, Schoels M, Aletaha D, Buch M, Gossec L, Huizinga T, Bijlsma JW, Burmester G, Combe B, Cutolo M, Gabay C, Gomez-Reino J, Kouloumas M, Kvien TK, Martin-Mola E, McInnes I, Pavelka K, van Riel P, Scholte M, Scott DL, Sokka T, Valesini G, van Vollenhoven R, Winthrop KL, Wong J, Zink A, van der Heijde D (2010) EULAR recommendations for the management of rheumatoid arthritis with synthetic and biological disease-modifying antirheumatic drugs. Ann Rheum Dis 69(6):964–975. https://​doi.​org/​10.​1136/​ard.​2009.​126532 CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Verhoeven MM, de Hair MJ, Tekstra J, Bijlsma JW, van Laar JM, Pethoe-Schramm A, Borm ME, Ter Borg EJ, Linn-Rasker SP, Teitsma XM, Lafeber FP, Jacobs JW, Welsing PM (2019) Initiating tocilizumab, with or without methotrexate, compared with starting methotrexate with prednisone within step-up treatment strategies in early rheumatoid arthritis: an indirect comparison of effectiveness and safety of the U-Act-Early and CAMERA-II treat-to-target trials. Ann Rheum Dis. https://doi.org/10.1136/annrheumdis-2019-215304 CrossRefPubMed Verhoeven MM, de Hair MJ, Tekstra J, Bijlsma JW, van Laar JM, Pethoe-Schramm A, Borm ME, Ter Borg EJ, Linn-Rasker SP, Teitsma XM, Lafeber FP, Jacobs JW, Welsing PM (2019) Initiating tocilizumab, with or without methotrexate, compared with starting methotrexate with prednisone within step-up treatment strategies in early rheumatoid arthritis: an indirect comparison of effectiveness and safety of the U-Act-Early and CAMERA-II treat-to-target trials. Ann Rheum Dis. https://​doi.​org/​10.​1136/​annrheumdis-2019-215304 CrossRefPubMed
6.
Zurück zum Zitat Wassenberg S (2015) Radiographic scoring methods in psoriatic arthritis. Clin Exp Rheumatol 33(5 Suppl 93):S55–59PubMed Wassenberg S (2015) Radiographic scoring methods in psoriatic arthritis. Clin Exp Rheumatol 33(5 Suppl 93):S55–59PubMed
7.
Zurück zum Zitat Colebatch AN, Edwards CJ, Ostergaard M, van der Heijde D, Balint PV, D'Agostino MA, Forslind K, Grassi W, Haavardsholm EA, Haugeberg G, Jurik AG, Landewe RB, Naredo E, O'Connor PJ, Ostendorf B, Potocki K, Schmidt WA, Smolen JS, Sokolovic S, Watt I, Conaghan PG (2013) EULAR recommendations for the use of imaging of the joints in the clinical management of rheumatoid arthritis. Ann Rheum Dis 72(6):804–814. https://doi.org/10.1136/annrheumdis-2012-203158 CrossRefPubMed Colebatch AN, Edwards CJ, Ostergaard M, van der Heijde D, Balint PV, D'Agostino MA, Forslind K, Grassi W, Haavardsholm EA, Haugeberg G, Jurik AG, Landewe RB, Naredo E, O'Connor PJ, Ostendorf B, Potocki K, Schmidt WA, Smolen JS, Sokolovic S, Watt I, Conaghan PG (2013) EULAR recommendations for the use of imaging of the joints in the clinical management of rheumatoid arthritis. Ann Rheum Dis 72(6):804–814. https://​doi.​org/​10.​1136/​annrheumdis-2012-203158 CrossRefPubMed
8.
Zurück zum Zitat Fiehn C, Holle J, Iking-Konert C, Leipe J, Weseloh C, Frerix M, Alten R, Behrens F, Baerwald C, Braun J, Burkhardt H, Burmester G, Detert J, Gaubitz M, Gause A, Gromnica-Ihle E, Kellner H, Krause A, Kuipers J, Lorenz HM, Muller-Ladner U, Nothacker M, Nusslein H, Rubbert-Roth A, Schneider M, Schulze-Koops H, Seitz S, Sitter H, Specker C, Tony HP, Wassenberg S, Wollenhaupt J, Kruger K (2018) S2e guideline: treatment of rheumatoid arthritis with disease-modifying drugs. Z Rheumatol 77(Suppl 2):35–53. https://doi.org/10.1007/s00393-018-0481-y CrossRefPubMed Fiehn C, Holle J, Iking-Konert C, Leipe J, Weseloh C, Frerix M, Alten R, Behrens F, Baerwald C, Braun J, Burkhardt H, Burmester G, Detert J, Gaubitz M, Gause A, Gromnica-Ihle E, Kellner H, Krause A, Kuipers J, Lorenz HM, Muller-Ladner U, Nothacker M, Nusslein H, Rubbert-Roth A, Schneider M, Schulze-Koops H, Seitz S, Sitter H, Specker C, Tony HP, Wassenberg S, Wollenhaupt J, Kruger K (2018) S2e guideline: treatment of rheumatoid arthritis with disease-modifying drugs. Z Rheumatol 77(Suppl 2):35–53. https://​doi.​org/​10.​1007/​s00393-018-0481-y CrossRefPubMed
11.
Zurück zum Zitat Symmons DP (2002) Epidemiology of rheumatoid arthritis: determinants of onset, persistence and outcome. Best Pract Res Clin Rheumatol 16(5):707–722CrossRef Symmons DP (2002) Epidemiology of rheumatoid arthritis: determinants of onset, persistence and outcome. Best Pract Res Clin Rheumatol 16(5):707–722CrossRef
14.
Zurück zum Zitat Dietrich TJ, Pfirrmann CW, Schwab A, Pankalla K, Buck FM (2013) Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography. Skeletal Radiol 42(7):959–967. https://doi.org/10.1007/s00256-013-1600-0 CrossRefPubMed Dietrich TJ, Pfirrmann CW, Schwab A, Pankalla K, Buck FM (2013) Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography. Skeletal Radiol 42(7):959–967. https://​doi.​org/​10.​1007/​s00256-013-1600-0 CrossRefPubMed
15.
Zurück zum Zitat Seeram E, Davidson R, Bushong S, Swan H (2016) Optimizing the exposure indicator as a dose management strategy in computed radiography. Radiol Technol 87(4):380–391PubMed Seeram E, Davidson R, Bushong S, Swan H (2016) Optimizing the exposure indicator as a dose management strategy in computed radiography. Radiol Technol 87(4):380–391PubMed
19.
Zurück zum Zitat Andre F, Fortner P, Vembar M, Mueller D, Stiller W, Buss SJ, Kauczor HU, Katus HA, Korosoglou G (2017) Improved image quality with simultaneously reduced radiation exposure: knowledge-based iterative model reconstruction algorithms for coronary CT angiography in a clinical setting. J Cardiovasc Comput Tomogr 11(3):213–220. https://doi.org/10.1016/j.jcct.2017.02.007 CrossRefPubMed Andre F, Fortner P, Vembar M, Mueller D, Stiller W, Buss SJ, Kauczor HU, Katus HA, Korosoglou G (2017) Improved image quality with simultaneously reduced radiation exposure: knowledge-based iterative model reconstruction algorithms for coronary CT angiography in a clinical setting. J Cardiovasc Comput Tomogr 11(3):213–220. https://​doi.​org/​10.​1016/​j.​jcct.​2017.​02.​007 CrossRefPubMed
21.
Zurück zum Zitat Hart D, Hillier M, Wall B (2002) Doses to patients from medical X-ray examinations in the UK-2000 review. United Kingdom Hart D, Hillier M, Wall B (2002) Doses to patients from medical X-ray examinations in the UK-2000 review. United Kingdom
23.
Zurück zum Zitat Chambers JM (1983) Graphical methods for data analysis. Wadsworth International Group Chambers JM (1983) Graphical methods for data analysis. Wadsworth International Group
27.
Zurück zum Zitat Hart D, Wall B, Hillier M, Shrimpton P (2010) HPA-CRCE-012: Frequency and collective dose for medical and dental X-ray examinations in the UK, 2008. United Kingdom Hart D, Wall B, Hillier M, Shrimpton P (2010) HPA-CRCE-012: Frequency and collective dose for medical and dental X-ray examinations in the UK, 2008. United Kingdom
29.
Zurück zum Zitat Ahmadzadeh A, Dehghan P, Rajaee A, Emam M, Enteshari K, Gachkar L (2013) Assessing rheumatologists and radiologists agreement rate regarding the diagnosis of focal bone erosions and osteopenic changes using hand X-rays radiography in patients with rheumatoid arthritis. Rheumatol Int 33(8):2019–2023. https://doi.org/10.1007/s00296-012-2645-4 CrossRefPubMed Ahmadzadeh A, Dehghan P, Rajaee A, Emam M, Enteshari K, Gachkar L (2013) Assessing rheumatologists and radiologists agreement rate regarding the diagnosis of focal bone erosions and osteopenic changes using hand X-rays radiography in patients with rheumatoid arthritis. Rheumatol Int 33(8):2019–2023. https://​doi.​org/​10.​1007/​s00296-012-2645-4 CrossRefPubMed
Metadaten
Titel
Application of an advanced noise reduction algorithm for imaging of hands in rheumatic diseases: evaluation of image quality compared to standard-dose images
verfasst von
Katharina Ziegeler
Stefan Siepmann
Sandra Hermann
Alexander Beck
Alexander Lembcke
Bernd Hamm
Kay Geert A. Hermann
Publikationsdatum
02.04.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Rheumatology International / Ausgabe 6/2020
Print ISSN: 0172-8172
Elektronische ISSN: 1437-160X
DOI
https://doi.org/10.1007/s00296-020-04560-1