Application of ASiR in combination with noise index in the chest CT examination of preschool-age children

verfasst von: Shilong Tang, Xianfan Liu, Ling He, Yu Zhou, Zhuo Cheng

Erschienen in: La radiologia medica | Ausgabe 6/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the application of adaptive statistical iterative reconstruction (ASIR) for chest CT scans of preschool-age children.

Method

Sixty children, ages 0 to < 1, 1 to < 3, and 3 to < 6 years, who underwent CT non-contrast-enhanced and enhanced scans were included. The non-contrast-enhanced scan sequences were performed with noise indexes (NIs) of 11, 14, and 16 for the 0 to < 1, 1 to < 3, and 3 to < 6 year age groups, respectively. Collected data were reconstructed using ASIR in increments of 10%, ranging from 0 to 100%, to generate 11 image groups. The signal-to-noise ratio, image noise, and other features of images obtained using ASIR with different weights were compared and analyzed. The best weight ranges for ASIR of chest CT scans of children at different ages within the range of 0–6 years were obtained. Enhanced scan sequence: The NI default was 9, and the data were subjected to the filtered back projection reconstruction algorithm. All other scanning parameters were the same as those used in the non-contrast-enhanced scan sequence.

Results

In the 0 to < 1 year group, the image qualities were scored as 3 or above with ASIR weights of 50% for the lung window and 40% for the mediastinal window; in the 1 to < 3 year group, the image qualities were scored as 3 or above with ASIR weights of 60% for the lung window and 50% for the mediastinal window; in the 3 to < 6 year group, the image qualities were scored as 3 or above with ASIR weights of 70% for the lung window and 60% for the mediastinal window.

Conclusion

For low-dose chest CT scans of preschool-age children, application of the ASIR technique significantly improved image quality and reduced image noise. The optimum weights of image ASIR were 50%, 60%, and 70% for the lung window and 40%, 50%, and 60% for the mediastinal window for the 0 to < 1, 1 to < 3, and 3 to < 6 year groups, respectively.
Literatur
1.
Zurück zum Zitat Ou P, Celermajer DS, Calcagni G, Brunelle F, Bonnet D, Sidi D (2007) Three-dimensional CT scanning: a new diagnostic modality in congenital heart disease. Heart 93(8):908–913CrossRefPubMed Ou P, Celermajer DS, Calcagni G, Brunelle F, Bonnet D, Sidi D (2007) Three-dimensional CT scanning: a new diagnostic modality in congenital heart disease. Heart 93(8):908–913CrossRefPubMed
2.
Zurück zum Zitat Smith EA, Dillman JR, Goodsitt MM, Christodoulou EG, Keshavarzi N, Strouse PJ (2014) Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology 270(2):526–534CrossRefPubMedPubMedCentral Smith EA, Dillman JR, Goodsitt MM, Christodoulou EG, Keshavarzi N, Strouse PJ (2014) Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology 270(2):526–534CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Brady SL, Moore BM, Yee BS, Kaufman RA (2014) Pediatric CT: implementation of ASIR for substantial radiation dose reduction while maintaining pre-ASIR image noise. Radiology 270:223–231CrossRefPubMed Brady SL, Moore BM, Yee BS, Kaufman RA (2014) Pediatric CT: implementation of ASIR for substantial radiation dose reduction while maintaining pre-ASIR image noise. Radiology 270:223–231CrossRefPubMed
4.
Zurück zum Zitat Gervaise A, Naulet P, Henry C, Pernin M, Portron Y, Lapierre-Combes M (2014) Low-dose CT with automatic tube current modulation, adaptive statistical iterative reconstruction, and low tube voltage for the diagnosis of renal colic: impact of body mass index. AJR Am J Roentgenol 202(3):553–560CrossRefPubMed Gervaise A, Naulet P, Henry C, Pernin M, Portron Y, Lapierre-Combes M (2014) Low-dose CT with automatic tube current modulation, adaptive statistical iterative reconstruction, and low tube voltage for the diagnosis of renal colic: impact of body mass index. AJR Am J Roentgenol 202(3):553–560CrossRefPubMed
5.
Zurück zum Zitat Mayer C, Meyer M, Sedlmair M, Schoenberg SO, Henzler T (2014) Potential for radiation dose savings in abdominal and chest CT using automatic tube voltage selection in combination with automatic tube current modulation. AJR Am J Roentgenol 203(2):292–299CrossRefPubMed Mayer C, Meyer M, Sedlmair M, Schoenberg SO, Henzler T (2014) Potential for radiation dose savings in abdominal and chest CT using automatic tube voltage selection in combination with automatic tube current modulation. AJR Am J Roentgenol 203(2):292–299CrossRefPubMed
6.
Zurück zum Zitat Kubo T, Ohno Y, Kauczor HU, Hatabu H (2014) Radiation dose reduction in chest CT—review of available options. Eur J Radiol 83(10):1953–1961CrossRefPubMed Kubo T, Ohno Y, Kauczor HU, Hatabu H (2014) Radiation dose reduction in chest CT—review of available options. Eur J Radiol 83(10):1953–1961CrossRefPubMed
7.
Zurück zum Zitat Chen JH, Jin EH, He W, Zhao LQ (2014) Combining automatic tube current modulation with adaptive statistical iterative reconstruction for low-dose chest CT screening. PLoS ONE 9(4):e92414CrossRefPubMedPubMedCentral Chen JH, Jin EH, He W, Zhao LQ (2014) Combining automatic tube current modulation with adaptive statistical iterative reconstruction for low-dose chest CT screening. PLoS ONE 9(4):e92414CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Singh S, Kalra MK, Hsieh J, Licato PE, Do S, Pien HH, Blake MA (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257(2):373–383CrossRefPubMed Singh S, Kalra MK, Hsieh J, Licato PE, Do S, Pien HH, Blake MA (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257(2):373–383CrossRefPubMed
10.
Zurück zum Zitat Hsieh SS, Chesler DA, Fleischmann D, Pelc NJ (2016) A limit on dose reduction possible with CT reconstruction algorithms without prior knowledge of the scan subject. Med Phys 43(3):1361–1368CrossRefPubMedPubMedCentral Hsieh SS, Chesler DA, Fleischmann D, Pelc NJ (2016) A limit on dose reduction possible with CT reconstruction algorithms without prior knowledge of the scan subject. Med Phys 43(3):1361–1368CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Ou P, Celermajer DS, Calcagni G, Brunelle F, Bonnet D, Sidi D (2007) Three-dimensional CT scanning: a new diagnostic modality in congenital heart disease. Heart 93(8):908–913CrossRefPubMed Ou P, Celermajer DS, Calcagni G, Brunelle F, Bonnet D, Sidi D (2007) Three-dimensional CT scanning: a new diagnostic modality in congenital heart disease. Heart 93(8):908–913CrossRefPubMed
12.
Zurück zum Zitat Young C, Taylor AM, Owens CM (2011) Paediatric cardiac computed tomography: a review of imaging techniques and radiation dose consideration. Eur Radiol 21(3):518–529CrossRefPubMed Young C, Taylor AM, Owens CM (2011) Paediatric cardiac computed tomography: a review of imaging techniques and radiation dose consideration. Eur Radiol 21(3):518–529CrossRefPubMed
13.
Zurück zum Zitat Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, Kim KP, Howe NL, Ronckers CM, Rajaraman P, Sir Craft AW, Parker L, Berrington deGonzalez A (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380(9840):499–505CrossRefPubMedPubMedCentral Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, Kim KP, Howe NL, Ronckers CM, Rajaraman P, Sir Craft AW, Parker L, Berrington deGonzalez A (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380(9840):499–505CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Widmann G, Bischel A, Stratis A, Kakar A, Bosmans H, Jacobs R, Gassner EM, Puelacher W, Pauwels R (2016) Ultralow dose dentomaxillofacial CT imaging and iterative reconstruction techniques: variability of Hounsfield units and contrast-to-noise ratio. Br J Radiol 89(1060):20151055CrossRefPubMedPubMedCentral Widmann G, Bischel A, Stratis A, Kakar A, Bosmans H, Jacobs R, Gassner EM, Puelacher W, Pauwels R (2016) Ultralow dose dentomaxillofacial CT imaging and iterative reconstruction techniques: variability of Hounsfield units and contrast-to-noise ratio. Br J Radiol 89(1060):20151055CrossRefPubMedPubMedCentral
15.
Zurück zum Zitat Cha MJ, Jeong WK, Choi D, Kim YK, Lim S, Choi SY, Lee WJ (2016) Iterative reconstruction: comparison of techniques for reduced-dose liver computed tomography following transarterial chemoembolization for hepatocellular carcinoma. Acta Radiol 57(12):1429–1437CrossRefPubMed Cha MJ, Jeong WK, Choi D, Kim YK, Lim S, Choi SY, Lee WJ (2016) Iterative reconstruction: comparison of techniques for reduced-dose liver computed tomography following transarterial chemoembolization for hepatocellular carcinoma. Acta Radiol 57(12):1429–1437CrossRefPubMed
16.
Zurück zum Zitat Yoon HJ, Chung MJ, Hwang HS, Moon JW, Lee KS (2015) Adaptive statistical iterative reconstruction-applied ultra-low-dose CT with radiography-comparable radiation dose: usefulness for lung nodule detection. Korean J Radiol 16(5):1132–1141CrossRefPubMedPubMedCentral Yoon HJ, Chung MJ, Hwang HS, Moon JW, Lee KS (2015) Adaptive statistical iterative reconstruction-applied ultra-low-dose CT with radiography-comparable radiation dose: usefulness for lung nodule detection. Korean J Radiol 16(5):1132–1141CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Wuest W, May M, Saake M et al (2016) Low-dose CT of the paranasal sinuses: minimizing X-ray exposure with spectral shaping. Eur Radiol 26:4155–4161CrossRefPubMed Wuest W, May M, Saake M et al (2016) Low-dose CT of the paranasal sinuses: minimizing X-ray exposure with spectral shaping. Eur Radiol 26:4155–4161CrossRefPubMed
18.
Zurück zum Zitat Lell MM, May MS, Brand M et al (2015) Imaging the parasinus region with a third-generation dual-source CT and the effect of tin filtration on image quality and radiation dose. AJNR Am J Neuroradiol 36:1225–1230CrossRefPubMed Lell MM, May MS, Brand M et al (2015) Imaging the parasinus region with a third-generation dual-source CT and the effect of tin filtration on image quality and radiation dose. AJNR Am J Neuroradiol 36:1225–1230CrossRefPubMed
19.
Zurück zum Zitat May MS, Brand M, Lell MM et al (2017) Radiation dose reduction in parasinus CT by spectral shaping. Neuroradiology 59:169–176CrossRefPubMed May MS, Brand M, Lell MM et al (2017) Radiation dose reduction in parasinus CT by spectral shaping. Neuroradiology 59:169–176CrossRefPubMed
20.
Zurück zum Zitat Smith EA, Dillman JR, Goodsitt MM et al (2014) Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology 270:526–534CrossRefPubMedPubMedCentral Smith EA, Dillman JR, Goodsitt MM et al (2014) Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology 270:526–534CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Zhao Y, Wu Y, Zuo Z, Cheng S (2017) CT angiography of the kidney using routine CT and the latest gemstone spectral imaging combination of different noise indexes: image quality and radiationdose. Radiol Med 122(5):327–336CrossRefPubMed Zhao Y, Wu Y, Zuo Z, Cheng S (2017) CT angiography of the kidney using routine CT and the latest gemstone spectral imaging combination of different noise indexes: image quality and radiationdose. Radiol Med 122(5):327–336CrossRefPubMed
22.
Zurück zum Zitat Lambert J, MacKenzie JD, Cody DD, Gould R (2014) Techniques and tactics for optimizing CT dose in adults and children: state of the art and future advances. J Am Coll Radiol 11:262–266CrossRefPubMed Lambert J, MacKenzie JD, Cody DD, Gould R (2014) Techniques and tactics for optimizing CT dose in adults and children: state of the art and future advances. J Am Coll Radiol 11:262–266CrossRefPubMed
23.
Zurück zum Zitat Phelps AS, Naeger DM, Courtier JL et al (2015) Pairwise comparison versus Likert scale for biomedical image assessment. Am J Roentgenol 204:8–14CrossRef Phelps AS, Naeger DM, Courtier JL et al (2015) Pairwise comparison versus Likert scale for biomedical image assessment. Am J Roentgenol 204:8–14CrossRef
24.
Zurück zum Zitat Giannoni CM, Guillerman RP (2015) Computed tomography for the evaluation of suspected airway foreign bodies. Clin Pediatr Emerg Med 16(4):230–234CrossRef Giannoni CM, Guillerman RP (2015) Computed tomography for the evaluation of suspected airway foreign bodies. Clin Pediatr Emerg Med 16(4):230–234CrossRef
25.
Zurück zum Zitat Kim HG, Chung YE, Lee YH et al (2015) Quantitative analysis of the effect of iterative reconstruction using a phantom: determining the appropriate blending percentage. Yonsei Med J 56:253–261CrossRefPubMed Kim HG, Chung YE, Lee YH et al (2015) Quantitative analysis of the effect of iterative reconstruction using a phantom: determining the appropriate blending percentage. Yonsei Med J 56:253–261CrossRefPubMed
26.
Zurück zum Zitat McKnight CD, Watcharotone K, Ibrahim M, Christodoulou E, Baer AH, Parmar HA (2014) Adaptive statistical iterative reconstruction: reducing dose while preserving image quality in the pediatric head CT examination. Pediatr Radiol 44:997–1003CrossRefPubMed McKnight CD, Watcharotone K, Ibrahim M, Christodoulou E, Baer AH, Parmar HA (2014) Adaptive statistical iterative reconstruction: reducing dose while preserving image quality in the pediatric head CT examination. Pediatr Radiol 44:997–1003CrossRefPubMed
27.
Zurück zum Zitat Mirro AE, Brady SL, Kaufman RA (2016) Full dose-reduction potential of statistical iterative reconstruction for head CT protocols in a predominantly pediatric population. AJNR Am J Neuroradiol 37:1199–1205CrossRefPubMedPubMedCentral Mirro AE, Brady SL, Kaufman RA (2016) Full dose-reduction potential of statistical iterative reconstruction for head CT protocols in a predominantly pediatric population. AJNR Am J Neuroradiol 37:1199–1205CrossRefPubMedPubMedCentral
28.
Zurück zum Zitat Vachha B, Brodoefel H, Wilcox C, Hackney DB, Moonis G (2013) Radiation dose reduction in soft tissue neck CT using adaptive statistical iterative reconstruction (ASIR). Eur J Radiol 82:2222–2226CrossRefPubMed Vachha B, Brodoefel H, Wilcox C, Hackney DB, Moonis G (2013) Radiation dose reduction in soft tissue neck CT using adaptive statistical iterative reconstruction (ASIR). Eur J Radiol 82:2222–2226CrossRefPubMed
Metadaten
Titel
Application of ASiR in combination with noise index in the chest CT examination of preschool-age children
verfasst von
Shilong Tang
Xianfan Liu
Ling He
Yu Zhou
Zhuo Cheng
Publikationsdatum
01.06.2019
Verlag
Springer Milan
Erschienen in
La radiologia medica / Ausgabe 6/2019
Print ISSN: 0033-8362
Elektronische ISSN: 1826-6983
DOI
https://doi.org/10.1007/s11547-018-00983-w

Neu im Fachgebiet Radiologie

KI-gestütztes Mammografiescreening überzeugt im Praxistest

Mit dem Einsatz künstlicher Intelligenz lässt sich die Detektionsrate im Mammografiescreening offenbar deutlich steigern. Mehr unnötige Zusatzuntersuchungen sind laut der Studie aus Deutschland nicht zu befürchten.

Stumme Schlaganfälle − ein häufiger Nebenbefund im Kopf-CT?

In 4% der in der Notfallambulanz initiierten zerebralen Bildgebung sind „alte“ Schlaganfälle zu erkennen. Gar nicht so selten handelt es sich laut einer aktuellen Studie dabei um unbemerkte Insulte. Bietet sich hier womöglich die Chance auf ein effektives opportunistisches Screening?

Die elektronische Patientenakte kommt: Das sollten Sie jetzt wissen

Am 15. Januar geht die „ePA für alle“ zunächst in den Modellregionen an den Start. Doch schon bald soll sie in allen Praxen zum Einsatz kommen. Was ist jetzt zu tun? Was müssen Sie wissen? Wir geben in einem FAQ Antworten auf 21 Fragen.

Stören weiße Wände und viel Licht die Bildqualitätskontrolle?

Wenn es darum geht, die technische Qualität eines Mammogramms zu beurteilen, könnten graue Wandfarbe und reduzierte Beleuchtung im Bildgebungsraum von Vorteil sein. Darauf deuten zumindest Ergebnisse einer kleinen Studie hin. 

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.