Assessing transpulmonary pressure via direct pleural manometry

verfasst von: Christopher Rugg, Stefan Schmid, Janett Kreutziger, Mathias Ströhle

Erschienen in: Annals of Intensive Care | Ausgabe 1/2020

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN
Hinweise
This comment refers to the article available at https://​doi.​org/​10.​1186/​s13613-019-0554-3.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
P L
Transpulmonary pressure
P es
Esophageal pressure
P pl
Pleural pressure
To the Editor,
With great interest we read the multicentre, prospective study on PEEP titration conducted in ARDS patients by Bergez et al. [1]. PEEP was set on the basis of two cleverly devised protocols—one referring to plateau the other to transpulmonary pressure (PL). Both were able to improve oxygenation compared to baseline settings with the latter slightly increasing possible over-distension. In accordance with an also recently published expert review by Gattinoni et al., they recognized diverse limitations and assumptions while using esophageal pressure (Pes) as a surrogate for pleural pressure (Ppl) [2]. Nonetheless, another animal and human cadaver study revealed that Pes accurately reflects Ppl in the mid to dependent lung [3]. Regarding studies on mechanical ventilation, direct measurement of Ppl is usually restricted to animal or cadaver studies and is typically described as not feasible in clinical practice [2]. In 2018, a detailed review by Zielinska-Krawczyk et al. revealed that regarding therapeutic thoracenteses literature on direct pleural manometry actually is quite copious [4]. Direct pleural manometry has been used for studying various pathophysiological aspects of thoracentesis including the safety of high-volume pleural fluid removal. Furthermore, a portable electronic manometer, easy to build and only using widely accessible elements has been described [5]. Therefore, a small-bore catheter, inserted into the pleural cavity was connected with a vascular pressure transducer via a 3-way stopcock. After referencing the pressure to the catheter insertion point into the chest, Ppl was precisely measured.
We would like to share our experience in performing such measurements in an ICU setting. From a total of over 600 patients per year in our department (23 beds, mixed ICU), over 100 receive pleural pigtail-catheters, mostly indicated for gas exchange-limiting pleural effusions. These patients are usually decompensated of heart or pulmonary failure, intubated and to some extent mechanically ventilated. Our anatomical landmark for catheter insertion is the inferior angle of the scapula, with the patient in lateral position. The catheter tip then usually lies dorsally in the middle to caudal section of the pleural cavity, with position controlled by chest radiography and sonography. Awareness of catheter tip position is fundamental for adequate interpretation of measured pressures, as is the fact that pleural surface pressure may differ from pleural fluid pressure [4]. Measurements can be intermittently displayed on our standard ICU monitor as a pressure curve for example, monitoring also negative values (Fig. 1). Noteworthy is that the units displayed are usually mmHg and one must convert to cmH2O. In so doing we are able to conduct reliable, real-time measurements of Ppl during the complete respiratory cycle, regardless of ventilation mode. Complications are scarce, include pneumo- or haematothorax and occur mainly during insertion or removal of the catheter.
Of course, in critically ill patients pleural effusions are not always present, but if they are, drainage with a small-bore catheter not only helps improve oxygenation, but also gives a possibility to easily and directly measure Ppl and thus titrate ventilator settings without further deployment of additional devices. This can be beneficial in settings of limited resources. Direct pleural manometry via thoracic pigtail-catheters can be seen as an alternative to rather than a replacement of esophageal measurements in a special subgroup of patients.

Acknowledgements

Not applicable.
Not applicable.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Bergez M, Fritsch N, Tran-Van D, Saghi T, Bounkim T, Gentile A, Labadie P, Fontaine B, Ouattara A, Rozé H. PEEP titration in moderate to severe ARDS: plateau versus transpulmonary pressure. Ann Intensive Care. 2019;9:81.CrossRef Bergez M, Fritsch N, Tran-Van D, Saghi T, Bounkim T, Gentile A, Labadie P, Fontaine B, Ouattara A, Rozé H. PEEP titration in moderate to severe ARDS: plateau versus transpulmonary pressure. Ann Intensive Care. 2019;9:81.CrossRef
2.
Zurück zum Zitat Gattinoni L, Giosa L, Bonifazi M, Pasticci I, Busana M, Macri M, Romitti F, Vassalli F, Quintel M. Targeting transpulmonary pressure to prevent ventilator-induced lung injury. Expert Rev Respir Med. 2019;13(8):737–46.CrossRef Gattinoni L, Giosa L, Bonifazi M, Pasticci I, Busana M, Macri M, Romitti F, Vassalli F, Quintel M. Targeting transpulmonary pressure to prevent ventilator-induced lung injury. Expert Rev Respir Med. 2019;13(8):737–46.CrossRef
3.
Zurück zum Zitat Yoshida T, Amato MBP, Grieco D, Chen L, Lima CAS, Roldan R, Morais CCA, Gomes S, Costa ELV, Cardoso PFG, Charbonney E, Richard JCM, Brochard L, Kavanagh BP. Esophageal manometry and regional transpulmonary pressure in lung injury. Am J Respir Crit Care Med. 2018;197:1018–26.CrossRef Yoshida T, Amato MBP, Grieco D, Chen L, Lima CAS, Roldan R, Morais CCA, Gomes S, Costa ELV, Cardoso PFG, Charbonney E, Richard JCM, Brochard L, Kavanagh BP. Esophageal manometry and regional transpulmonary pressure in lung injury. Am J Respir Crit Care Med. 2018;197:1018–26.CrossRef
4.
Zurück zum Zitat Zielinska-Krawczyk M, Krenke R, Grabczak EM, Light RW. Pleural manometry-historical background, rationale for use and methods of measurement. Respir Med. 2018;136:21–8.CrossRef Zielinska-Krawczyk M, Krenke R, Grabczak EM, Light RW. Pleural manometry-historical background, rationale for use and methods of measurement. Respir Med. 2018;136:21–8.CrossRef
5.
Zurück zum Zitat Krenke R, Guć M, Grabczak EM, Michnikowski M, Pałko KJ, Chazan R, Gólczewski T. Development of an electronic manometer for intrapleural pressure monitoring. Respiration. 2011;136:21–8. Krenke R, Guć M, Grabczak EM, Michnikowski M, Pałko KJ, Chazan R, Gólczewski T. Development of an electronic manometer for intrapleural pressure monitoring. Respiration. 2011;136:21–8.
Metadaten
Titel
Assessing transpulmonary pressure via direct pleural manometry
verfasst von
Christopher Rugg
Stefan Schmid
Janett Kreutziger
Mathias Ströhle
Publikationsdatum
01.12.2020
Verlag
Springer International Publishing
Erschienen in
Annals of Intensive Care / Ausgabe 1/2020
Elektronische ISSN: 2110-5820
DOI
https://doi.org/10.1186/s13613-020-00718-z

Weitere Artikel der Ausgabe 1/2020

Annals of Intensive Care 1/2020 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.