Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction

verfasst von: Bharti Kataria, Jonas Nilsson Althén, Örjan Smedby, Anders Persson, Hannibal Sökjer, Michael Sandborg

Erschienen in: European Radiology | Ausgabe 6/2018

Abstract

Purpose

To estimate potential dose reduction in abdominal CT by visually comparing images reconstructed with filtered back projection (FBP) and strengths of 3 and 5 of a specific MBIR.

Material and methods

A dual-source scanner was used to obtain three data sets each for 50 recruited patients with 30, 70 and 100% tube loads (mean CTDIvol 1.9, 3.4 and 6.2 mGy). Six image criteria were assessed independently by five radiologists. Potential dose reduction was estimated with Visual Grading Regression (VGR).

Results

Comparing 30 and 70% tube load, improved image quality was observed as a significant strong effect of log tube load and reconstruction method with potential dose reduction relative to FBP of 22–47% for MBIR strength 3 (p < 0.001). For MBIR strength 5 no dose reduction was possible for image criteria 1 (liver parenchyma), but dose reduction between 34 and 74% was achieved for other criteria. Interobserver reliability showed agreement of 71–76% (κw 0.201–0.286) and intra-observer reliability of 82–96% (κw 0.525–0.783).

Conclusion

MBIR showed improved image quality compared to FBP with positive correlation between MBIR strength and increasing potential dose reduction for all but one image criterion.

Key Points

MBIR’s main advantage is its de-noising properties, which facilitates dose reduction.
MBIR allows for potential dose reduction in relation to FBP.
Visual Grading Regression (VGR) produces direct numerical estimates of potential dose reduction.
MBIR strengths 3 and 5 dose reductions were 22–34 and 34–74%.
MBIR strength 5 demonstrates inferior performance for liver parenchyma.
Abkürzungen
ADMIRE
Advanced Modeled Iterative Reconstruction
ALARA
As low as reasonably achievable
CT
Computed tomography
FBP
Filtered back projection
IR
Iterative reconstruction
kV
Kilovolt
MBIR
Model-based iterative reconstruction
mAs
Milliampere-second
mSv
Millisievert
Qref
Quality reference
VGR
Visual Grading Regression

Introduction

Technical developments and new applications have led to an increase in the use of computed tomography (CT) in medical imaging and the associated population doses that arise from it [1, 2]. CT contributes up to 70% of the collective effective dose, although it accounts for only 10–15% of the total medical imaging procedures that use ionising radiation [36]. Multiphase examinations are more common in abdominal CT, and approximately 30% of CT examinations are abdominal and pelvic, which deliver an effective dose of approximately 6–8 mSv [7]. In recognition of benefits of CT [8], the optimisation of the clinical protocols is motivated to keep the dose as low as reasonably achievable (ALARA principle) [9, 10].
Modern CT equipment presents a number of dose-reduction strategies such as automatic tube current modulation, iterative reconstruction algorithms (IR), dynamic collimation and dose efficient detectors, among others [1, 10, 11].
IR selectively reduces statistical noise in the images thus improving image quality of subtle details, and may facilitate dose reduction. There has been successful improvement in performance of the IR algorithms as they have evolved in the past decade from statistical to model-based algorithms (MBIR) [11, 12]. MBIR, which may be applied at different strengths, perform noise reduction in both raw data and image domains and incorporate physical models to accurately correct for a variety of image degrading effects [13, 14]. Several studies indicate that an increase in the strength of IR allows for larger dose reductions [13, 15, 16]. Evaluations of radiological imaging methods can be performed either by studying their ability to provide correct diagnoses [17] or by visual assessment of well-defined image quality features (visual grading) [18]. However, to our knowledge, there are no studies that directly estimate the dose-reduction potential of the Advanced Modeled Iterative Reconstruction (ADMIRE, Siemens, Erlangen, Germany) algorithm in the clinical setting using pair-wise comparison of images.
The aim of this study was to assess visual image quality between filtered back projection (FBP) and ADMIRE strengths 3 and 5 (out of 5) in abdominal CT, and to estimate the dose-reduction potential of the reconstruction algorithm.

Material and methods

This was a regional ethical board-approved prospective study conducted at the Centre for Medical Image Science and Visualization (CMIV), Linköping University, Sweden.
A Somatom Force 192-slice dual source CT (Siemens) was used to obtain three data sets of images per patient at 30, 70 and 100% dose levels from a single abdominal acquisition. These were achieved without additional patient exposure as the 30 and 70% tube loads were obtained simultaneously using the dual sources. The acquisition parameters from a standard clinical abdominal protocol are presented in Table 1. Due to a 35.5-cm diameter restriction of the small detector scan field of view (SFoV), ethical approval was obtained for 90 patients as anatomical fit to the smaller SFoV could only be determined after the scan was performed.
Table 1
Acquisition parameters for dual source Somatom Force (Siemens, Erlangen, Germany) for tube A and tube B (smaller detector) and Dual Energy Composition (DE comp) set at 0.5 i.e. equal kV weighting for each x-ray tube
Source
U (kV)
Qref (mAs)
Acquisition
Rotation (s)
Pitch
Care Dose 4D
Kernel
Dose level
Slice thickness/Increment (mm)
Tube A+B
120
140
192 × 0.6
0.5
0.6
Yes
Br36
100%
3/2
Tube A
120
98
192 × 0.6
0.5
0.6
Yes
Br36
70%
3/2
Tube B
120
42
192 × 0.6
0.5
0.6
Yes
Br36
30%
3/2
Qref quality reference
Patients were informed as to the intent of the study and written consent and approval were obtained. Inclusion criteria were patients over the age of 18 years undergoing a clinical abdominal CT with appropriate patient body habitus determined by visual estimation and use of a calliper to estimate patient size before the scan. Forty patients were excluded due to size and anatomical variations.
Of the 50 examinations, 25 were contrast-enhanced and 25 non-enhanced examinations.
Critical care was taken in patient positioning at isocentre in the gantry. Demographical data such as age, height and weight were also recorded.

Procedure

The images were anonymised so as to avoid identification of individual patients.
Images at each dose level (30, 70 and 100%) were reconstructed with FBP and ADMIRE strengths 3 and 5. Pairwise visual grading was carried out independently by five radiologists with varying experience (6–20 years), using four modified criteria (C1–C4) from the European guidelines for image quality in abdominal CT [19] together with image noise (C5) and overall image quality (C6) [20] to suit the purpose of this study. The criteria used were as follows:
  • C1. Visually sharp reproduction of the liver parenchyma
  • C2. Visually sharp reproduction of the pancreas contour
  • C3. Visually sharp reproduction of the contours of the kidneys and proximal ureters
  • C4. Visually sharp reproduction of the contours of lymph nodes smaller than 15 mm
  • C5. Image noise not affecting interpretation
  • C6. Overall image quality for diagnostic purposes.
All of the radiologists had 3–4 years’ of experience with SAFIRE strength 3 and one radiologist has been working with ADMIRE strength 3 for a year.
Prior to the study the participating radiologists were coached in grading the different aspects of subjective image quality so as to form a similar understanding of interpretation of the image criteria in order to minimise inter-observer variation. The data sets used in the coaching session were not included in the study population. Each reader rated the criteria in a randomised, blinded and pair-wise approach on DICOM-calibrated (EIZO RX 240) PACS version 17.3 (Sectra, Linköping, Sweden) workstations. The image pairs were graded on a 5-point Likert-type scale (Table 2).
Table 2
Ordinal grading scores used for each image criterion in the visual image quality assessment
Grading scores
−2 image on left monitor is better than image on right monitor
−1 image on left monitor is probably better than image on right monitor
0 images on left and right monitors are equivalent
+1 image on right monitor is probably better than image on left monitor
+2 image on right monitor is better than image on left monitor
Comparison of 12 pairs of image stacks (Fig. 1) per patient resulted in 600 (12 × 50) image pairs per radiologist and a total of 3,000 (600 × 5) image pair assessments. Five of the image pair assessments were replicated to calculate the intra-observer reliability.

Statistical analysis

Image quality scores were statistically analysed using visual grading regression (VGR) [18]. VGR is an ordinal logistic regression method applied to scores from observer ratings, controlling for dependencies between observers, patients, tube loads and reconstruction methods. Statistical analyses were performed with the software Stata 13.1 (Stata Corporation LP, College Station, TX, USA) using the multi-level mixed-effects ordered logistic regression (meologit) command. The regression coefficients describe how the image quality depends on the choice of tube load and reconstruction algorithm, respectively. By relating two of these coefficients to each other, it is possible to estimate the potential dose reduction (DR) when replacing one algorithm with another from the equation DR = 1 − e−(b/a), where a is the regression coefficient for log mAs and b that for the iterative reconstruction algorithm [21].
Inter-observer and intra-observer reliabilities were described with the weighted kappa (κw) [22] using the kappa2 command in Stata. The null hypothesis is that neither tube load nor reconstruction method influence perceived image quality. The significance limit was set at p=0.05.

Results

Of the 50 patients examined, 22 were women, age range 22–90 years (standard deviation (SD) 16.9, mean 64.6) with a body mass index (BMI) of 16.4–27.3 kg/m2 (SD 3.0, mean 21.8) and 28 men, age range 44–85 years (SD 10.9, mean 64.8) with a BMI of 16.6–26.2 kg/m2 (SD 2.1, mean 23.2). For the study population the CTDIvol ranged from 3.9 to 9.1 mGy (SD 1.3, mean 6.2 mGy), size-specific dose estimate (SSDE) ranged from 6.3 to 12.8 mGy (SD 1.5, mean 8.6 mGy), and dose-length product (DLP) ranged from 161 to 468 mGy.cm (SD 70, mean 292 mGy.cm).
The frequency histograms (Fig. 2) for each criterion show the percentage of favourable versus unfavourable scores (%) with respect to reconstruction algorithm and dose level. Highest scores are seen for quality reference (Qref) mAs 98 (70% dose level) for all image criteria. ADMIRE 3 also tended to yield higher scores when compared with FBP and ADMIRE 5. For ADMIRE 5, scores for overall image quality (criterion 6) were equivalent to FBP and inferior to FBP for criterion 1 (liver parenchyma). Surprisingly the scores for full dose images (Qref mAs 140) are lower than those at the 70% dose level (Qref mAs 98), suggesting that no image quality improvements are obtained with increase in dose. Visual demonstration of image quality in one of the study patients obtained with three tube loads and three reconstruction algorithms are presented in Fig. 3.
The effect on image quality of varying the tube load for each reconstruction algorithm separately is presented in Table 3. A positive value of the regression coefficient in the table indicates that image quality was improved by the change in Qref mAs. A negative value, on the other hand, indicated a reduction in image quality. For FBP, image quality increases with an increase in tube load. For ADMIRE 3 and ADMIRE 5, however, the image quality seems to increase between Qref mAs 42 and 98, but then to decrease for interval Qref mAs 98 and 140. VGR assumes that there is a linear relationship between the log mAs and the logit function score. As this assumption does not hold for the interval between Qref mAs 98 and 140, the subsequent VGR statistical analysis is based on Qref mAs interval 42 to 98.
Table 3
Dependency on tube load for different reconstruction algorithms. Visual grading regression (VGR) coefficients for log (mAs) in pairwise comparisons of two quality reference (Qref) mAs values
Criterion
Reconstruction algorithm
FBP
ADMIRE 3
ADMIRE 5
42 mAs vs. 98 mAs
98 mAs vs. 140 mAs
42 mAs vs. 98 mAs
98 mAs vs. 140 mAs
42 mAs vs. 98 mAs
98 mAs vs. 140 mAs
1. Liver parenchyma
3.36**
2.20***
–0.28°
1.49*
–0.37°
2. Pancreatic contours
1.89**
2.41***
–0.89*
1.37***
–2.62***
3. Kidneys and proximal ureters
7.25*
2.39*
2.29***
–0.92*
1.46***
–2.20***
4. Lymph nodes < 15 mm in diameter
6.50*
2.09**
1.83***
–1.93***
1.14***
–2.20*
5. Image noise
5.49**
4.70**
3.07**
4.79*
1.07°
6. Overall image quality
3.68*
0.03°
3.74**
–1.27**
* p < 0.05, ** p < 0.01, *** p < 0.001, ° not significant
– Convergence was not achieved with the statistical software available
The VGR result for log mAs and ADMIRE 3 and ADMIRE 5 comparing dose levels at Qref mAs 98 with 42 are presented in Table 4. A significant strong effect of log mAs (p < 0.001) for all six criteria was noted, which indicated an increase in dose for the same algorithm leads to a corresponding increase in image quality. ADMIRE strength 3 resulted in a strongly significant (p < 0.001) increase in image quality relative to FBP for all six criteria. This is reflected in the estimated potential dose reduction ranging from 22% (liver parenchyma) to 47% (image noise). Strong significant results are also seen for ADMIRE 5, criteria 2−6 with a slightly higher dose reduction ranging from 34% (overall image quality) to 74% (image noise), with the exception of criterion 1 (liver parenchyma), for which there was no significant difference between ADMIRE 5 and FBP when assessing the delineation of liver parenchyma.
Table 4
Visual grading regression (VGR) coefficients for all six criteria with estimated dose reduction values for comparison of images reconstructed with ADMIRE 3, ADMIRE 5 and filtered back projection (FBP) for dose interval between quality references (Qref) mAs 98 and 42
Criterion
Regression coefficients
Estimated dose reduction
(95% confidence limits)
log (mAs)
ADMIRE 3
ADMIRE 5
ADMIRE 3
ADMIRE 5
1. Liver parenchyma
2.28***
0.57***
–0.08°
22%
(11%; 33%)
2. Pancreatic contours
2.00***
0.92***
1.73***
37%
(26%; 48%)
58%
(53%; 63%)
3. Kidneys and proximal ureters
2.21***
1.11***
2.09***
40%
(31%; 49%)
61%
(57%; 66%)
4. Lymph nodes < 15 mm in diameter
1.72***
1.05***
1.93***
46%
(35%; 57%)
67%
(63%; 72%)
5. Image noise
2.38***
1.50***
3.16***
47%
(39%; 55%) 
74%
(71%; 76%)
6. Overall image quality
2.69***
1.06***
1.10***
33%
(24%; 41%)
34%
(27%; 40%)
***p < 0.001, ° not significant
Comparing results between ADMIRE 3 and ADMIRE 5 in relation to FBP (Table 4), criteria 2–6 suggest that the dose reduction potential is higher with increase in strength. However, for ADMIRE 5 criterion 1 (liver parenchyma) there was no dose reduction possible. Image noise decreases with increase in IR strength, which is reflected by the slightly higher estimated dose reduction for ADMIRE 5 compared to ADMIRE 3.
When comparing ADMIRE strength 3 with ADMIRE strength 5 (Table 5), the regression yielded significant (p<0.001) values for most criteria when dose levels for the same algorithm strength are compared, confirming that ADMIRE 5 is superior to ADMIRE 3. However, for criterion 1 (liver parenchyma) the negative coefficient value (−0.98) indicates that image quality in ADMIRE 5 is significantly lower compared to ADMIRE 3 and hence does not allow for any further dose reduction. Also, overall image quality with a regression coefficient value of −0.85 was not significant. For the criteria assessed, ADMIRE strengths 3 and 5 are superior when compared to FBP individually except for criterion 1 (ADMIRE 5). When it comes to delineation of the liver parenchyma and overall image quality, ADMIRE 5 was inferior to ADMIRE 3 with no dose reduction possible.
Table 5
Visual grading regression (VGR) coefficients for all six criteria with estimated dose reduction values for comparison between ADMIRE 3 and ADMIRE 5 in the dose interval quality references (Qref) mAs 98 and 42
Criterion
Regression coefficients
Estimated dose reduction
(95% confidence limits)
log (mAs)
ADMIRE 5 Reconstruction
1. Liver parenchyma
1.88***
–0.98***
–68%
(–102%; –35%)
2. Pancreatic contours
1.84***
0.61***
27%
(18%; 37%)
3. Kidneys and proximal ureters
2.06***
0.77***
31%
(23%; 39%)
4. Lymph nodes < 15 mm in diameter
1.49***
0.68***
37%
(26%; 47%)
5. Image noise
2.42***
1.66***
50%
(45%; 55%)
6. Overall image quality
3.18***
–0.85°
***p < 0.001, ° not significant
There were marginal differences in the results of the VGR analysis between the groups of patients who received intravenous contrast material and those who did not (data not shown).
The inter-observer agreement was fair, 71–76% with κw ranging from 0.201 (confidence interval (CI) 0.175–0.228) to 0.286 (CI 0.258–0.314). The intra-observer κw values ranged from 0.525 (CI 0.209–0.840) to 0.783 (CI 0.577–1.021), showing a moderate to substantial agreement between 82 and 96% for all the criteria.

Discussion

In radiology, several analysis methods can be used to describe image quality. The receiver operating characteristic (ROC) analysis method is used to evaluate and compare diagnostic performance [17, 23]. When determining potential dose reduction in the optimisation process, visual grading experiments may be useful [21, 24]. VGR seems to be the only analysis method that produces direct numerical estimates of potential dose reductions for new acquisition, reconstruction and post-processing techniques while image quality is maintained [25].
The present study aimed to estimate the dose reduction potential of ADMIRE strengths 3 and 5 compared with FBP in a standard-dose abdominal CT. It suggests that ADMIRE while preserving image quality allows for a dose reduction relative to FBP of 22−47% (ADMIRE 3) for all criteria assessed, and 34−74% (ADMIRE 5) for criteria 2–6 with the exception of liver parenchyma visualisation. Similar results are reported by Greffier et al, who studied the performance of SAFIRE strengths 1–5 compared to FBP in two data sets at 30 and 70% dose levels. They concluded that a 40–60% reduction in dose is possible [26]. Gordic et al. [27] evaluated both quantitative and qualitative image quality parameters in abdominal CT using ADMIRE. Results from their study showed an improved image quality with lower noise when comparing ADMIRE with FBP, where the amount of noise reduction (53% for ADMIRE 5) could be translated to a reduction in radiation dose (e.g. reduction in effective mAs). However, their study differs from the present study in that they evaluated visibility of only small structures such as small blood vessels, adrenal glands and lymph nodes.
In the present study, the VGR analysis in Table 3 revealed that the 70% dose level (Qref mAs 98) scores were, for the iterative algorithms, higher than full dose (Qref mAs 140) scores for all image criteria. This was an unexpected finding. For FBP, the results were in agreement with the general rule that image quality increases with increasing tube load. For the iterative reconstruction algorithms, one might speculate that a certain amount of noise is required for the algorithm to work optimally. It should be noted that due to technical advancements the Somatom Force scanner has been found to give better image quality than other Siemens equipment [28]. A possible conclusion is that the 70% dose level provides images of sufficiently high image quality. Thus, one can optimise the standard clinical abdominal protocol for the Somatom Force using ADMIRE 3 by reducing the tube load from Qref mAs 140 to 98 without changing the strength of the algorithm. However, as demonstrated in Table 5, by replacing ADMIRE 3 with ADMIRE 5, further dose reduction can be achieved for certain aspects of the image quality, but not all.
With ADMIRE strength 5 there is still a problem as non-linear effects of IR lead to smoothing of the anatomical features and a change in appearance of the anatomy in the images [29]. Mieville et al. [30] reported a change in the appearance of the MBIR images. Certain small objects that were not identified on the FBP images were visualised on the low-dose MBIR images. Suboptimal performance when evaluating small or subtle abdominal structures (i.e. common bile duct, adrenal glands and pancreatic duct) was also reported by Padole et al. [31] when comparing FBP with reduced dose MBIR and Adaptive Statistical Iterative Reconstruction (ASIR). This compromise in the visibility of structures could be attributed to the blotchy, pixelated and plastic-like appearance of the images. The liver parenchyma is a low-contrast object and although image quality is improved using MBIR, it is possible that no improvement is seen in detection of low-contrast details as was the case in the phantom study of Euler et al. [32]. Contrary results were presented by Solomon et al. [13] who studied low-contrast detectability using ADMIRE. There is a possibility that assessing low-contrast detectability in vivo is different to phantom images as the task of clinical assessment by a radiologist, who has access to full image data and patient information, is relatively complex compared to the simple task of assessing subtle lesions in a phantom. Solomon et al. [33] used a similar method to the present study to estimate the dose reduction potential of SAFIRE relative to FBP. They also investigated virtual liver lesion detectability in hybrid images. Although their study was simple compared to clinical reality, such experiments do provide valuable information about how different algorithms render the same lesion differently. In the present study, reader confidence in determining the visually sharp reproduction of liver parenchyma was lower for ADMIRE 5 than for FBP. Change in image texture due to the denoising properties of the reconstruction algorithm influences the potential dose reduction depending on the diagnostic task [30]. This may explain the slightly lower estimated dose reduction for ADMIRE 3 and the non-significant result for ADMIRE 5 when assessing the liver parenchyma.
On the other hand, anatomical contour assessments were not a problem even though the higher strength images were smoother in appearance. As our readers pointed out during the coaching session, when intra-abdominal fat is present, the delineation of contours is further enhanced. This was true for the assessment of proximal ureters, lymph nodes and pancreas. However, in patients with low BMI, when a smaller amount of intra-abdominal fat is present, the delineation of contours can be difficult [34].
Low kappa values were seen for inter-observer agreement between all five readers. It is not unusual that there is a variation in perception of image quality among radiologists as viewing strategies differ depending on the approach [35]. The intra-observer agreement in the present study was less than 100%, indicating variation in perception for the same reader at different points in time [36].
Future research is indicated as there is some scepticism among radiologists in using higher strengths of IR. This is mostly related to the change in image texture, due to substantial noise reduction, possibly affecting diagnostic confidence. When comparing delineation of structures in different planes, Mieville et al. [30] reported an improvement in detection of small structures in the coronal plane compared to the axial plane. It would be of interest to study if higher strengths of the algorithm could be made more clinically acceptable with a combination of IR and other post-processing methods. This might increase diagnostic confidence in IR images of higher strength and allow for further dose reductions.
The major limitation of our study was exclusion of overweight patients due to size limitations of the small detector, hence the estimated dose reduction is limited to the patients with a BMI of up to 27.3 kg/m2. Image quality might be inferior as noise increases in overweight patients with a BMI > 28, possibly leading to less or no dose reduction [24, 37]. Since there was a variation in patient body habitus, not all scans fitted the 35.5-cm diameter. Some of the patient anatomy, mostly in the pelvic region, slightly exceeded this limit, which may have affected the image quality. However, there were no anatomical image criteria present in this region, except for overall image noise and image quality. It is therefore unlikely that this would have affected the results. There are many image acquisition parameters that affect image quality. We have studied only change in tube load (mAs) and image reconstruction. Hence the study protocol differs from the clinical protocol as a fixed kV was used for both x-ray tubes without automatic adjustment of tube potential (Care kV). Visual grading is an easy and inexpensive method to assess image quality. However, it assumes that whenever normal anatomy is sharply reproduced, the same will apply to pathology. The extent to which this assumption is correct is generally not known. As reconstructed images with MBIR strengths 3 and 5 and FBP all have different appearances, it is difficult to perform a true blinded evaluation of subjective criteria [38]. This enhances the need for further research and analysis of objective image quality parameters to support the subjective findings of this study.

Conclusion

The model-based iterative reconstruction algorithm ADMIRE showed improved image quality compared to FBP. A positive correlation between ADMIRE strength and increasing potential dose reduction was found for the majority, but not all, of the image criteria.

Acknowledgements

We are grateful to Siemens for providing us with a research license for configuration of the Somatom Force to obtain three dose levels per patient. We acknowledge CMIV for providing the research facilities. Håkan Gustafsson is acknowledged for supporting our project and our participating radiologists, Anki Pozson, Johan Asplund, Jenny Öman, Senija Halilic and Peter Johansson, for grading the images. The study was presented both as an oral and an E-poster at the European Congress of Radiology 2017 in Vienna.

Compliance with ethical standards

Guarantor

The scientific guarantor of this publication is Prof. Michael Sandborg.

Conflict of interest

The authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.

Funding

This work was supported by ALF- and LFoU-grants from Region Östergötland and the Medical Faculty at Linköping University.

Statistics and biometry

One of the authors has adequate statistical expertise.
Written informed consent was obtained from all subjects (patients) in this study.

Ethical approval

Regional Ethical Review Board approval was obtained.

Methodology

• prospective
• cross-sectional study
• performed at one institution
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Mahesh M (2009) MDCT physics the basics technology, image quality and radiation dose. Lippincott Williams & Wilkins, Philadelphia Mahesh M (2009) MDCT physics the basics technology, image quality and radiation dose. Lippincott Williams & Wilkins, Philadelphia
2.
Zurück zum Zitat Moores M (2017) A review of the fundamental principles of radiation protection when applied to the patient in diagnostic radiology. Radiat Prot Dosim 175:1–9 Moores M (2017) A review of the fundamental principles of radiation protection when applied to the patient in diagnostic radiology. Radiat Prot Dosim 175:1–9
3.
Zurück zum Zitat Sun Z, Ng KA, Sarji SA (2010) Is utilisation of computed tomography justified in clinical practice? Part IV: applications of paediatric computed tomography. Singapore Med J 51:457–463PubMed Sun Z, Ng KA, Sarji SA (2010) Is utilisation of computed tomography justified in clinical practice? Part IV: applications of paediatric computed tomography. Singapore Med J 51:457–463PubMed
4.
Zurück zum Zitat Report UNSCEAR (2008) Sources and effects of ionizing radiation. United Nations Scientific Committee on the effects of ionizing radiation. Volume 1: Sources. Report to the general assembly Scientific annexes A & B Report UNSCEAR (2008) Sources and effects of ionizing radiation. United Nations Scientific Committee on the effects of ionizing radiation. Volume 1: Sources. Report to the general assembly Scientific annexes A & B
5.
Zurück zum Zitat Le Coultre R, Bize J, Champendal M et al (2016) Exposure of the Swiss population by radiodiagnostics: 2013 review. Radiat Prot Dosim 169:221–224CrossRef Le Coultre R, Bize J, Champendal M et al (2016) Exposure of the Swiss population by radiodiagnostics: 2013 review. Radiat Prot Dosim 169:221–224CrossRef
7.
Zurück zum Zitat Power SP, Maloney F, Twomey M, James K, O’Connor OJ, Maher MM (2016) Computed tomography and patient risks: facts, perceptions and uncertainties. World J Radiol 8:902–915CrossRefPubMedPubMedCentral Power SP, Maloney F, Twomey M, James K, O’Connor OJ, Maher MM (2016) Computed tomography and patient risks: facts, perceptions and uncertainties. World J Radiol 8:902–915CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Mayo-Smith WW, Hara AK, Mahesh M, Sahani DV, Pavlicek W (2014) How I do it: managing radiation dose in CT. Radiology 273:657–672CrossRefPubMed Mayo-Smith WW, Hara AK, Mahesh M, Sahani DV, Pavlicek W (2014) How I do it: managing radiation dose in CT. Radiology 273:657–672CrossRefPubMed
9.
Zurück zum Zitat Smith-Bindeman R, Lipson J, Markus R et al (2009) Radiation dose associated with common Computed Tomography examinations and the associated lifetime attributable risk for cancer. Arch Intern Med 169:2078–2086CrossRef Smith-Bindeman R, Lipson J, Markus R et al (2009) Radiation dose associated with common Computed Tomography examinations and the associated lifetime attributable risk for cancer. Arch Intern Med 169:2078–2086CrossRef
10.
Zurück zum Zitat Kalra MK, Sodickson AD, Mayo-Smith WW (2015) CT radiation: key concepts for gentle and wise use. Radiographics 35:1706–1721CrossRefPubMed Kalra MK, Sodickson AD, Mayo-Smith WW (2015) CT radiation: key concepts for gentle and wise use. Radiographics 35:1706–1721CrossRefPubMed
11.
Zurück zum Zitat Liu L (2014) Model based Iterative Reconstruction: a promising algorithm for today's Computed Tomography Imaging. J Med Radiat Sci 45:131–136CrossRef Liu L (2014) Model based Iterative Reconstruction: a promising algorithm for today's Computed Tomography Imaging. J Med Radiat Sci 45:131–136CrossRef
12.
Zurück zum Zitat Beister M, Kolditz D, Kalender WA (2012) Iterative reconstruction methods in X-ray CT. Phys Medica 28:94–108CrossRef Beister M, Kolditz D, Kalender WA (2012) Iterative reconstruction methods in X-ray CT. Phys Medica 28:94–108CrossRef
13.
Zurück zum Zitat Solomon J, Mileto A, Ramirez-Giraldo JC, Samei E (2015) Diagnostic Performance of an Advanced Modeled Iterative Reconstruction Algorithm for low-contrast detectability with a third generation Multidetector dual source CT Scanner. Radiology 275:735–745CrossRefPubMed Solomon J, Mileto A, Ramirez-Giraldo JC, Samei E (2015) Diagnostic Performance of an Advanced Modeled Iterative Reconstruction Algorithm for low-contrast detectability with a third generation Multidetector dual source CT Scanner. Radiology 275:735–745CrossRefPubMed
14.
Zurück zum Zitat Nuyts J, De Man B, Fessler JA, Zbijewski W, Beekman FJ (2013) Modelling the physics in iterative reconstruction for transmission computed tomography. Phys Med Biol 58:R63–R96CrossRefPubMedPubMedCentral Nuyts J, De Man B, Fessler JA, Zbijewski W, Beekman FJ (2013) Modelling the physics in iterative reconstruction for transmission computed tomography. Phys Med Biol 58:R63–R96CrossRefPubMedPubMedCentral
15.
Zurück zum Zitat Ott JG, Ba A, Racine D et al (2016) Patient exposure optimisation through task-based assessment of a new model-based iterative reconstruction technique. Radiat Prot Dosim 169:68–72CrossRef Ott JG, Ba A, Racine D et al (2016) Patient exposure optimisation through task-based assessment of a new model-based iterative reconstruction technique. Radiat Prot Dosim 169:68–72CrossRef
16.
Zurück zum Zitat Patino M, Fuentes JM, Singh S, Hahn PF, Sahani DV (2015) Iterative reconstruction techniques in abdominopelvic CT: technical concepts and clinical implementation. AJR Am J Roentgenol 205:W19–W31CrossRefPubMed Patino M, Fuentes JM, Singh S, Hahn PF, Sahani DV (2015) Iterative reconstruction techniques in abdominopelvic CT: technical concepts and clinical implementation. AJR Am J Roentgenol 205:W19–W31CrossRefPubMed
17.
18.
Zurück zum Zitat Smedby Ö, Fredrikson M (2010) Visual grading regression – analysing data from visual grading experiments with regression models. Br J Radiol 83:767–775CrossRefPubMedPubMedCentral Smedby Ö, Fredrikson M (2010) Visual grading regression – analysing data from visual grading experiments with regression models. Br J Radiol 83:767–775CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Borgen L, Kalra MK, Laerum F et al (2012) Application of adaptive non-linear 2 D and 3 D post processing filters for reduced dose abdominal CT. Acta Radiol 53:335–34219CrossRefPubMed Borgen L, Kalra MK, Laerum F et al (2012) Application of adaptive non-linear 2 D and 3 D post processing filters for reduced dose abdominal CT. Acta Radiol 53:335–34219CrossRefPubMed
21.
Zurück zum Zitat Smedby Ö, Fredrikson M, De Geer J, Borgen L, Sandborg M (2013) Quantifying the potential for dose reduction with visual grading regression. Br J Radiol 86:1–6CrossRef Smedby Ö, Fredrikson M, De Geer J, Borgen L, Sandborg M (2013) Quantifying the potential for dose reduction with visual grading regression. Br J Radiol 86:1–6CrossRef
22.
Zurück zum Zitat Abraira V, Pérez de Vargas A (1999) Generalization of the Kappa coefficient for ordinal categorical data, multiple observers and incomplete designs. Qüestiió 23:561–571 Abraira V, Pérez de Vargas A (1999) Generalization of the Kappa coefficient for ordinal categorical data, multiple observers and incomplete designs. Qüestiió 23:561–571
24.
Zurück zum Zitat Kataria B, Smedby Ö (2013) Patient dose and image quality in a low-dose abdominal CT: a comparison between iterative reconstruction and filtered back projection. Acta Radiol 54:540–548CrossRefPubMed Kataria B, Smedby Ö (2013) Patient dose and image quality in a low-dose abdominal CT: a comparison between iterative reconstruction and filtered back projection. Acta Radiol 54:540–548CrossRefPubMed
25.
Zurück zum Zitat Saffari SE, Löve Á, Fredrikson M, Smedby Ö (2015) Regression models for analyzing radiological visual grading studies – an empirical comparison. BMC Med Imaging 15:1–10CrossRef Saffari SE, Löve Á, Fredrikson M, Smedby Ö (2015) Regression models for analyzing radiological visual grading studies – an empirical comparison. BMC Med Imaging 15:1–10CrossRef
26.
Zurück zum Zitat Greffier J, Fernandez A, Macri F, Freitag C, Metge L, Beregi JP (2013) Which dose for what image? Iterative reconstruction for CT scan. Diagn Interv Imaging 94:1117–1121CrossRefPubMed Greffier J, Fernandez A, Macri F, Freitag C, Metge L, Beregi JP (2013) Which dose for what image? Iterative reconstruction for CT scan. Diagn Interv Imaging 94:1117–1121CrossRefPubMed
27.
Zurück zum Zitat Gordic S, Desbiolles L, Stolzmann P et al (2014) Advanced modelled iterative reconstruction for abdominal CT: qualitative and quantitative evaluation. Clin Radiol 69:497–504CrossRef Gordic S, Desbiolles L, Stolzmann P et al (2014) Advanced modelled iterative reconstruction for abdominal CT: qualitative and quantitative evaluation. Clin Radiol 69:497–504CrossRef
28.
Zurück zum Zitat Wichmann JL, Hardie AD, Schoepf JU et al (2016) Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT. Eur Radiol 27:642–650CrossRefPubMed Wichmann JL, Hardie AD, Schoepf JU et al (2016) Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT. Eur Radiol 27:642–650CrossRefPubMed
29.
Zurück zum Zitat Larsson J, Båth M, Ledenius K, Caisander H, Thilander-Klang A (2016) Assessment of clinical image quality in paediatric abdominal CT examinations: dependency on the level of adaptive statistical iterative reconstruction (ASiR) and the type of convolution kernel. Radiat Prot Dosimetry 169:123–129CrossRefPubMed Larsson J, Båth M, Ledenius K, Caisander H, Thilander-Klang A (2016) Assessment of clinical image quality in paediatric abdominal CT examinations: dependency on the level of adaptive statistical iterative reconstruction (ASiR) and the type of convolution kernel. Radiat Prot Dosimetry 169:123–129CrossRefPubMed
30.
Zurück zum Zitat Mieville FA, Berteloot L, Grandjean A et al (2013) Model-based iterative reconstruction in Pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis. Pediatr Radiol 43:558–567CrossRefPubMed Mieville FA, Berteloot L, Grandjean A et al (2013) Model-based iterative reconstruction in Pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis. Pediatr Radiol 43:558–567CrossRefPubMed
31.
Zurück zum Zitat Padole A, Singh S, Lira D et al (2015) Assessment of filtered back projection, adaptive statistical, and model-based iterative reconstruction for reduced dose abdominal computed tomography. J Comput Assist Tomogr 39:462–467CrossRefPubMed Padole A, Singh S, Lira D et al (2015) Assessment of filtered back projection, adaptive statistical, and model-based iterative reconstruction for reduced dose abdominal computed tomography. J Comput Assist Tomogr 39:462–467CrossRefPubMed
32.
Zurück zum Zitat Euler A, Stieltjes B, Szucs-Farkas Z et al. (2017) Impact of model-based iterative reconstruction on low-contrast lesion detection and image quality in abdominal CT: a 12-reader-based comparative phantom study with filtered back projection at different tube voltages. Eur Radiol. https://doi.org/10.1007/s00330-017-4825-9 Euler A, Stieltjes B, Szucs-Farkas Z et al. (2017) Impact of model-based iterative reconstruction on low-contrast lesion detection and image quality in abdominal CT: a 12-reader-based comparative phantom study with filtered back projection at different tube voltages. Eur Radiol. https://​doi.​org/​10.​1007/​s00330-017-4825-9
33.
Zurück zum Zitat Solomon J, Marin D, Roy Choudhury K, Patel B, Samei E (2017) Effect of radiation dose reduction and reconstruction algorithm on Image noise, contrast, resolution, and detectability of subtle hypoattenuating liver lesions at multidetector CT: filtered back projection versus a commercial model-based iterative reconstruction algorithm. Radiology. https://doi.org/10.1148/radiol.2017161736 Solomon J, Marin D, Roy Choudhury K, Patel B, Samei E (2017) Effect of radiation dose reduction and reconstruction algorithm on Image noise, contrast, resolution, and detectability of subtle hypoattenuating liver lesions at multidetector CT: filtered back projection versus a commercial model-based iterative reconstruction algorithm. Radiology. https://​doi.​org/​10.​1148/​radiol.​2017161736
34.
Zurück zum Zitat Sagara Y, Hara AK, Pavlicek W, Silva AC, Paden RG, Wu Q (2010) Abdominal CT: comparison of low-dose CT with adaptive statistic iterative reconstruction and routine dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol 195:713–719CrossRefPubMed Sagara Y, Hara AK, Pavlicek W, Silva AC, Paden RG, Wu Q (2010) Abdominal CT: comparison of low-dose CT with adaptive statistic iterative reconstruction and routine dose CT with filtered back projection in 53 patients. AJR Am J Roentgenol 195:713–719CrossRefPubMed
35.
Zurück zum Zitat Drew T, Vo MHL, Olwal A, Jacobson F, Seltzer SE, Wolfe J (2013) Scanners and drillers: characterizing expert visual search through volumetric images. J Vis 13:1–13CrossRef Drew T, Vo MHL, Olwal A, Jacobson F, Seltzer SE, Wolfe J (2013) Scanners and drillers: characterizing expert visual search through volumetric images. J Vis 13:1–13CrossRef
36.
Zurück zum Zitat Birkelo CC, Chamberlain WE, Phelps PS, Schools PE, Zacks D, Yerushalmy J (1947) Tuberculosis case finding. JAMA 133:359–366CrossRef Birkelo CC, Chamberlain WE, Phelps PS, Schools PE, Zacks D, Yerushalmy J (1947) Tuberculosis case finding. JAMA 133:359–366CrossRef
37.
Zurück zum Zitat Chang W, Lee JM, Lee K et al (2013) Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography. Invest Radiol 48:598–606CrossRefPubMed Chang W, Lee JM, Lee K et al (2013) Assessment of a model-based, iterative reconstruction algorithm (MBIR) regarding image quality and dose reduction in liver computed tomography. Invest Radiol 48:598–606CrossRefPubMed
38.
Zurück zum Zitat Hérin E, Gardavaud F, Chiaradia M et al (2015) Use of model-based iterative reconstruction (MBIR) in reduced dose CT for routine follow-up of patients with malignant lymphoma: dose savings, image quality & phantom study. Eur Radiol 25:2362–2370CrossRefPubMed Hérin E, Gardavaud F, Chiaradia M et al (2015) Use of model-based iterative reconstruction (MBIR) in reduced dose CT for routine follow-up of patients with malignant lymphoma: dose savings, image quality & phantom study. Eur Radiol 25:2362–2370CrossRefPubMed
Metadaten
Titel
Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction
verfasst von
Bharti Kataria
Jonas Nilsson Althén
Örjan Smedby
Anders Persson
Hannibal Sökjer
Michael Sandborg
Publikationsdatum
24.01.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 6/2018
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-5113-4

Weitere Artikel der Ausgabe 6/2018

European Radiology 6/2018 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.