Skip to main content
Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology 6/2016

15.09.2015 | Retinal Disorders

Assessment of potential vessel segmentation pitfalls in the analysis of blood flow velocity using the Retinal Function Imager

verfasst von: Gábor Márk Somfai, Jing Tian, Delia Cabrera DeBuc

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 6/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purpose of our study was to investigate the potential pitfalls associated with different vessel segmentation methods using the built-in software of the Retinal Function Imager (RFI) for the analysis of retinal blood flow velocity (BFV).

Methods

Ten eyes of nine healthy subjects were enrolled in the study. Retinal blood flow measurements were obtained with the RFI device with a 20° field of view imaging. The same grader segmented the retinal vasculature using the RFI software in both sessions, with segments ranging in length from 50 to 100 pixels (“short segments”) and 100–200 pixels (“long segments”). The blood flow velocities for the arteriolar and venular system were calculated, and the percentage of excluded vessel segments with high coefficients of variation (>45 %) was recorded and compared by paired t test. Spearman's correlation was used to analyze the relationship between measurements by the two vessel segmentation methods.

Results

The number of analyzed vessel segments did not differ significantly between the two groups (28.6 ± 2.6 short and 26.7 ± 4.6 long segments), while the percentage of acceptable segments was significantly higher in the long segment group (65.2 ± 11.4 % vs 85.2 ± 5.87 %, p = 0.001). In the short segment group, more than 15 % of vessel segments were rejected in all subjects, while in the long segment group only three subjects had a rejection rate of greater than 15 % (16.7 %, 18.7 % and 28 %). Both arteriolar and venular velocities were lower in the short segment group, although it reached significance only for arteriolar velocities (3.93 ± 0.55 vs. 4.45 ± 0.76 mm/s, p = 0.036 and 2.95 ± 0.56 vs. 3.17 ± 0.84 mm/s, p = 0.201 for arterioles and venules, respectively). Only venular velocities showed significant correlation (p = 0.003, R 2 = 0.67) between the two groups.

Conclusions

Our results suggest that BFV measurements by the RFI may be affected by segment length, and care should therefore be taken in choosing vessel segment lengths used during the analysis of RFI data. Long segments of 100–200 pixels (400–800 μm) seem to provide more robust measurements, which can be explained by the analysis methodology of the RFI device.
Literatur
1.
Zurück zum Zitat Arend O, Remky A, Harris A, Bertram B, Reim M, Wolf S (1995) Macular microcirculation in cystoid maculopathy of diabetic patients. B J Ophthalmol 79:628–632CrossRef Arend O, Remky A, Harris A, Bertram B, Reim M, Wolf S (1995) Macular microcirculation in cystoid maculopathy of diabetic patients. B J Ophthalmol 79:628–632CrossRef
2.
Zurück zum Zitat Bohdanecka Z, Orgül S, Meyer AB, Prünte C, Flammer J (1999) Relationship between blood flow velocities in retrobulbar vessels and laser Doppler flowmetry at the optic disk in glaucoma patients. Ophthalmologica 213:145–149CrossRefPubMed Bohdanecka Z, Orgül S, Meyer AB, Prünte C, Flammer J (1999) Relationship between blood flow velocities in retrobulbar vessels and laser Doppler flowmetry at the optic disk in glaucoma patients. Ophthalmologica 213:145–149CrossRefPubMed
3.
Zurück zum Zitat Loukovaara S, Kaaja R, Immonen I (2002) Macular capillary blood flow velocity by blue-field entoptoscopy in diabetic and healthy women during pregnancy and the postpartum period. Graefe’s Arch Clin Exp Ophthalmol 240:977–982CrossRef Loukovaara S, Kaaja R, Immonen I (2002) Macular capillary blood flow velocity by blue-field entoptoscopy in diabetic and healthy women during pregnancy and the postpartum period. Graefe’s Arch Clin Exp Ophthalmol 240:977–982CrossRef
4.
Zurück zum Zitat Kimura I, Shinoda K, Tanino T, Ohtake Y, Mashima Y, Oguchi Y (2003) Scanning laser Doppler flowmeter study of retinal blood flow in macular area of healthy volunteers. B J Ophthalmol 87:1469–1473CrossRef Kimura I, Shinoda K, Tanino T, Ohtake Y, Mashima Y, Oguchi Y (2003) Scanning laser Doppler flowmeter study of retinal blood flow in macular area of healthy volunteers. B J Ophthalmol 87:1469–1473CrossRef
5.
Zurück zum Zitat Polska E, Ehrlich P, Luksch A, Fuchsjäger-Mayrl G, Schmetterer L (2003) Effects of adenosine on intraocular pressure, optic nerve head blood flow, and choroidal blood flow in healthy humans. Invest Ophthalmol Vis Sci 44:3110–3114CrossRefPubMed Polska E, Ehrlich P, Luksch A, Fuchsjäger-Mayrl G, Schmetterer L (2003) Effects of adenosine on intraocular pressure, optic nerve head blood flow, and choroidal blood flow in healthy humans. Invest Ophthalmol Vis Sci 44:3110–3114CrossRefPubMed
6.
Zurück zum Zitat Nagaoka T, Kitaya N, Sugawara R et al (2004) Alteration of choroidal circulation in the foveal region in patients with type 2 diabetes. B J Ophthalmol 88:1060–1063CrossRef Nagaoka T, Kitaya N, Sugawara R et al (2004) Alteration of choroidal circulation in the foveal region in patients with type 2 diabetes. B J Ophthalmol 88:1060–1063CrossRef
7.
Zurück zum Zitat Nelson DA, Krupsky S, Pollack A et al (2004) Special report: Noninvasive multi-parameter functional optical imaging of the eye. Ophthalmic Surg Las Imaging 36:57–66 Nelson DA, Krupsky S, Pollack A et al (2004) Special report: Noninvasive multi-parameter functional optical imaging of the eye. Ophthalmic Surg Las Imaging 36:57–66
8.
Zurück zum Zitat Landa G, Garcia PMT, Rosen RB (2010) Correlation between retina blood flow velocity assessed by retinal function imager and retina thickness estimated by scanning laser ophthalmoscopy/optical coherence tomography. Ophthalmologica 223:155–161CrossRef Landa G, Garcia PMT, Rosen RB (2010) Correlation between retina blood flow velocity assessed by retinal function imager and retina thickness estimated by scanning laser ophthalmoscopy/optical coherence tomography. Ophthalmologica 223:155–161CrossRef
9.
Zurück zum Zitat Burgansky-Eliash Z, Nelson DA, Bar-Tal OP, Lowenstein A, Grinvald A, Barak A (2010) Reduced retinal blood flow velocity in diabetic retinopathy. Retina 30:765–773CrossRefPubMed Burgansky-Eliash Z, Nelson DA, Bar-Tal OP, Lowenstein A, Grinvald A, Barak A (2010) Reduced retinal blood flow velocity in diabetic retinopathy. Retina 30:765–773CrossRefPubMed
10.
Zurück zum Zitat Burgansky-Eliash Z, Lowenstein A, Neuderfer M et al (2013) The correlation between retinal blood flow velocity measured by the retinal function imager and various physiological parameters. Ophthalmic Surg Las Imaging Retina 44:51–58CrossRef Burgansky-Eliash Z, Lowenstein A, Neuderfer M et al (2013) The correlation between retinal blood flow velocity measured by the retinal function imager and various physiological parameters. Ophthalmic Surg Las Imaging Retina 44:51–58CrossRef
11.
Zurück zum Zitat Barak A, Burgansky-Eliash Z, Barash H, Nelson DA, Grinvald A, Loewenstein A (2012) The effect of intravitreal bevacizumab (Avastin) injection on retinal blood flow velocity in patients with choroidal neovascularization. Eur J Ophthalmol 22:423CrossRefPubMed Barak A, Burgansky-Eliash Z, Barash H, Nelson DA, Grinvald A, Loewenstein A (2012) The effect of intravitreal bevacizumab (Avastin) injection on retinal blood flow velocity in patients with choroidal neovascularization. Eur J Ophthalmol 22:423CrossRefPubMed
12.
Zurück zum Zitat Burgansky-Eliash Z, Barash H, Nelson D et al (2014) Retinal Blood Flow Velocity in Patients with Age-Related Macular Degeneration. Curr Eye Res 39:304–311CrossRefPubMed Burgansky-Eliash Z, Barash H, Nelson D et al (2014) Retinal Blood Flow Velocity in Patients with Age-Related Macular Degeneration. Curr Eye Res 39:304–311CrossRefPubMed
13.
Zurück zum Zitat Beutelspacher SC, Serbecic N, Barash H, Burgansky-Eliash Z, Grinvald A, Jonas JB (2011) Central serous chorioretinopathy shows reduced retinal flow circulation in retinal function imaging (RFI). Acta Ophthalmol 89:e479–e482CrossRefPubMed Beutelspacher SC, Serbecic N, Barash H, Burgansky-Eliash Z, Grinvald A, Jonas JB (2011) Central serous chorioretinopathy shows reduced retinal flow circulation in retinal function imaging (RFI). Acta Ophthalmol 89:e479–e482CrossRefPubMed
14.
Zurück zum Zitat Landa G, Rosen RB (2010) A new vascular pattern for idiopathic juxtafoveal telangiectasia revealed by the retinal function imager. Ophthalmic Surg Las Imaging 41:413–417CrossRef Landa G, Rosen RB (2010) A new vascular pattern for idiopathic juxtafoveal telangiectasia revealed by the retinal function imager. Ophthalmic Surg Las Imaging 41:413–417CrossRef
15.
Zurück zum Zitat Birger Y, Blumenfeld O, Bartov E, Burgansky-Eliash Z (2011) Reduced retinal blood flow-velocity in severe hyperlipidemia measured by the retinal function imager. Graefe’s Arch Clin Exp Ophthalmol 249:1587–1590CrossRef Birger Y, Blumenfeld O, Bartov E, Burgansky-Eliash Z (2011) Reduced retinal blood flow-velocity in severe hyperlipidemia measured by the retinal function imager. Graefe’s Arch Clin Exp Ophthalmol 249:1587–1590CrossRef
16.
Zurück zum Zitat Jensen PS, Glucksberg MR (1998) Regional variation in capillary hemodynamics in the cat retina. Invest Ophthalmol Vis Sci 39:407–415PubMed Jensen PS, Glucksberg MR (1998) Regional variation in capillary hemodynamics in the cat retina. Invest Ophthalmol Vis Sci 39:407–415PubMed
17.
Zurück zum Zitat Optical Imaging RFI 3K Series: RFI 3000/3006 System Operation Manual 2011 Optical Imaging RFI 3K Series: RFI 3000/3006 System Operation Manual 2011
18.
Zurück zum Zitat Gutfreund S, Izkhakov E, Pokroy R et al (2013) Retinal blood flow velocity in metabolic syndrome. Graefe’s Arch Clin Exp Ophthalmol 251:1507–1513CrossRef Gutfreund S, Izkhakov E, Pokroy R et al (2013) Retinal blood flow velocity in metabolic syndrome. Graefe’s Arch Clin Exp Ophthalmol 251:1507–1513CrossRef
19.
Zurück zum Zitat Izhaky D, Nelson DA, Burgansky-Eliash Z, Grinvald A (2009) Functional imaging using the retinal function imager: direct imaging of blood velocity, achieving fluorescein angiography-like images without any contrast agent, qualitative oximetry, and functional metabolic signals. Jpn J Ophthalmol 53:345–351CrossRefPubMed Izhaky D, Nelson DA, Burgansky-Eliash Z, Grinvald A (2009) Functional imaging using the retinal function imager: direct imaging of blood velocity, achieving fluorescein angiography-like images without any contrast agent, qualitative oximetry, and functional metabolic signals. Jpn J Ophthalmol 53:345–351CrossRefPubMed
20.
Zurück zum Zitat Landa G, Amde W, Haileselassie Y, Rosen RB (2011) Cilioretinal arteries in diabetic eyes are associated with increased retinal blood flow velocity and occurrence of diabetic macular edema. Retina 31:304–311CrossRefPubMed Landa G, Amde W, Haileselassie Y, Rosen RB (2011) Cilioretinal arteries in diabetic eyes are associated with increased retinal blood flow velocity and occurrence of diabetic macular edema. Retina 31:304–311CrossRefPubMed
21.
Zurück zum Zitat Landa G, Rosen RB (2010) New patterns of retinal collateral circulation are exposed by a retinal functional imager (RFI). B J Ophthalmol 94:54–58CrossRef Landa G, Rosen RB (2010) New patterns of retinal collateral circulation are exposed by a retinal functional imager (RFI). B J Ophthalmol 94:54–58CrossRef
22.
Zurück zum Zitat Nelson DA, Burgansky-Eliash Z, Barash H et al (2011) High-resolution wide-field imaging of perfused capillaries without the use of contrast agent. Clin Ophthalmol 5:1095PubMedPubMedCentral Nelson DA, Burgansky-Eliash Z, Barash H et al (2011) High-resolution wide-field imaging of perfused capillaries without the use of contrast agent. Clin Ophthalmol 5:1095PubMedPubMedCentral
23.
Zurück zum Zitat Arbel Y, Sternfeld A, Barak A et al (2014) Inverse correlation between coronary and retinal blood flows in patients with normal coronary arteries and slow coronary blood flow. Atherosclerosis 232(1):149–54CrossRefPubMed Arbel Y, Sternfeld A, Barak A et al (2014) Inverse correlation between coronary and retinal blood flows in patients with normal coronary arteries and slow coronary blood flow. Atherosclerosis 232(1):149–54CrossRefPubMed
24.
Zurück zum Zitat Landa G, Jangi AA, Garcia PM, Rosen RB (2012) Initial report of quantification of retinal blood flow velocity in normal human subjects using the Retinal Functional Imager (RFI). Int Ophthalmol 32(3):211–5CrossRefPubMed Landa G, Jangi AA, Garcia PM, Rosen RB (2012) Initial report of quantification of retinal blood flow velocity in normal human subjects using the Retinal Functional Imager (RFI). Int Ophthalmol 32(3):211–5CrossRefPubMed
25.
Zurück zum Zitat Witkin AJ, Alshareef RA, Rezeq SS et al (2012) Comparative analysis of the retinal microvasculature visualized with fluorescein angiography and the retinal function imager. Am J Ophthalmol 154:901–907CrossRefPubMed Witkin AJ, Alshareef RA, Rezeq SS et al (2012) Comparative analysis of the retinal microvasculature visualized with fluorescein angiography and the retinal function imager. Am J Ophthalmol 154:901–907CrossRefPubMed
26.
Zurück zum Zitat Faria JML, Duarte DA, Chavez RFL, Arthur AM, Arthur R, Iano Y (2014) Reliability and validity of digital assessment of perifoveal capillary network measurement using high-resolution imaging. B J Ophthalmol 98:725–729 Faria JML, Duarte DA, Chavez RFL, Arthur AM, Arthur R, Iano Y (2014) Reliability and validity of digital assessment of perifoveal capillary network measurement using high-resolution imaging. B J Ophthalmol 98:725–729
27.
Zurück zum Zitat Chhablani J, Bartsch DU, Cheng L et al (2013) Segmental reproducibility of retinal blood flow velocity measurements using retinal function imager. Graefe’s Arch Clin Exp Ophthalmol 251:2665–2670CrossRef Chhablani J, Bartsch DU, Cheng L et al (2013) Segmental reproducibility of retinal blood flow velocity measurements using retinal function imager. Graefe’s Arch Clin Exp Ophthalmol 251:2665–2670CrossRef
Metadaten
Titel
Assessment of potential vessel segmentation pitfalls in the analysis of blood flow velocity using the Retinal Function Imager
verfasst von
Gábor Márk Somfai
Jing Tian
Delia Cabrera DeBuc
Publikationsdatum
15.09.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 6/2016
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-015-3166-0

Weitere Artikel der Ausgabe 6/2016

Graefe's Archive for Clinical and Experimental Ophthalmology 6/2016 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.