Open Access 13.10.2022 | Original Article

Assessment of Risk Factors for Rupture in Breast Reconstruction Patients with Macrotextured Breast Implants

verfasst von: Guido Paolini, Guido Firmani, Francesca Briganti, Mattia Macino, Simone Nigrelli, Michail Sorotos, Fabio Santanelli di Pompeo

Erschienen in: Aesthetic Plastic Surgery | Ausgabe 2/2023

Abstract

Background

Breast implants (BI) are widely used in plastic surgery, though they are not lifetime devices. Average life before rupture is reported to be around 10–15 years. No consensus exists regarding which factors are involved.

Objectives

Following FDA recommendations, this study aims at identifying potential risk factors by evaluating their effect on BI rupture cases.

Methods

In this observational study, 763 BI patients were operated between 2003 and 2019, with a mean implant indwelling of 12.2 years. Patients that returned for follow-up were administered a questionnaire regarding postoperative lifestyle and habits. Implant rupture rate was 15.1%, while BI lifespan was 10.1 years. We obtained complete data from 191 breast implant patients (288 implants). Twenty-three potential risk factors were evaluated and divided in four categories: patient-related, surgery-related, postoperative complications/symptoms, and postoperative care/lifestyle habits. Odds Ratio (OR) for each factor was calculated. Linear regression analysis was calculated for those with a significant OR.

Results

We report 120 patients (195 implants) with intact and 71 (93 implants) with ruptured devices. BIs were macrotextured in 95.1% of cases (86.8% Allergan BIOCELL). OR was significant for underwire bra use (OR: 2.708), car seat belts (OR: 3.066), mammographic imaging (OR: 2.196), weightlifting (OR: 0.407) and carry-on heavy purses and backpacks (OR: 0.347).

Conclusion

Wearing underwire bras, seat belts and undergoing mammography increases the risk of rupture. Weightlifting and carry heavy bags do not increase that risk. Implant rupture is directly linked with time of indwelling. Postoperative recommendations in BI patients should consider findings from our study, though larger multicenter studies should be encouraged.

Level of Evidence IV

This journal requires that authors assign a level of evidence to each article. For a full description of these Evidence-Based Medicine ratings, please refer to the Table of Contents or the online Instructions to Authors www.​springer.​com/​00266.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s00266-022-03118-9.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Breast implants (BIs) are class III medical devices according to the Food and Drug Administration (FDA) [1], a category reserved for high-risk devices which may cause illness or injury [2]. Despite this fact, recent estimates suggest that 3% of female adults have BIs [3]. Their popularity is owed to breast augmentations being among the top five most frequent cosmetic surgeries, while implant-based techniques represent the most commonly performed reconstructive procedures [4].
Industrial and clinical research have been conducted to improve upon BI characteristics, with the goal of reducing complications and enhancing aesthetic outcomes [5, 6]. Despite these efforts, the FDA cautions that BIs are not lifetime devices [7]. Some complications leading to removal or replacement are related to surgical technique (i.e., bleeding/hematoma, early seroma, infection), thus they are potentially predictable and possibly avoidable [8]. Implant rupture, which represents one cause for revisional surgery [9], is unpredictable, as it is unclear whether it depends on device failure or whether certain risk factors increase its likelihood. In fact, little is known regarding which factors can negatively affect BI survival, other than longevity of device indwelling.
Long-term observations from our prospectively-maintained breast surgery database highlighted a high number of BI rupture cases, which compelled us to identify which factors were associated with a higher likelihood of rupture, in accordance with the FDA’s “Guidance for Industry and Staff” on BIs [10]. Potential risk factors regarding patient characteristics, surgery, complications, postoperative care and lifestyle habits were assessed to verify their impact on rupture.

Materials and methods

This retrospective observational study was conducted in accordance with the World Medical Association’s Declaration of Helsinki for ethical principles in medical research involving human subjects [11]. A prospectively-maintained breast surgery database was used to determine BI rupture rate, and calculate mean indwelling and lifespan. BI lifespan was calculated for patients with ruptured devices and was defined as the period of time between placement and rupture. BI indwelling was calculated for the overall population, by averaging the time between implantation and latest follow-up in patients with intact BIs, and BI lifespan in patients with ruptured devices.
The primary outcome was to determine which variables were associated with BI rupture [10]. These variables were subdivided into four categories: patient-related factors, surgery-related factors, postoperative complications/symptoms, and postoperative care/lifestyle habits. A total of 23 potential risk factors were identified and are further subdivided in Table 1.
Table 1
Identified risk factors for breast implant rupture, subdivided into four categories
Patient-related risk factors
 Age
  Body mass index (BMI)
  Handedness (right-handedness, left-handedness or ambidexterity)
  Time between implantation and removal or exchange for rupture
Surgery-related factors
 Implant manufacturer
 Implant degree of texturization
 Implant size (in cc)
 Indication for placement (reconstructive vs cosmetic procedure)
 Type of breast reconstruction procedure (only for reconstruction)
 Submuscular vs subglandular pocket (only for cosmetic procedure)
 Adjunct surgical procedures (i.e., autologous fat transfer, nipple-areola complex reconstruction, scar revisions)
 Pre- or postoperative radiotherapy
Postoperative complications
 Capsular contracture (according to Baker’s classification, range: soft [I], minimal [II], moderate [III] and severe [IV])
 History of trauma of the chest region following surgery
Postoperative care and lifestyle habits
 Using an underwire bra
  Wearing a car seat belt
  Sleeping in a prone position
  Implant manipulations or massages following surgery
  Physical activity and sports
  Weightlifting (i.e., lifting weights heavier than 5 kg using arms on a daily basis)
  Use of heavy purses or backpacks
  Type of housework (subdivided, according to the frequency and duration of household chores, into: none, light, moderate and heavy)
  History of mammographic imaging following surgery, with Eklund-modified compression technique

Patient population

In our database, we identified 763 patients who received BI placement for reconstructive or aesthetic purposes at our institution between 2003 and 2019, with a follow-up of at least 3 years. The database was used to retrieve surgery-related factors and postoperative complications. All patients were sent via email a questionnaire regarding postoperative care and lifestyle habits [Supplemental Material 1]. This study was conducted between June 2021 and January 2022. Patients who responded to the questionnaire were included, while those without recent diagnostic breast imaging according to FDA recommendations [12], and/or who did not respond were excluded.

Statistical analysis

All variables were subdivided into two categories: dichotomous and continuous (Table 2). Time of implant indwelling was considered a continuous variable expressed in 1-month increments. We initially performed descriptive statistic calculations for continuous variables, assessing the average value, mode, median value and standard deviations. Nominal and ordinal variables were reported adding proportions and percentages. Pearson correlation coefficient was used for the following continuous variables: patient BMI, breast implant weight, time of implant indwelling and the dependent dichotomous variable implant integrity status (ruptured vs intact). In regard to the continuous variable “time of implant indwelling,” we compared averages between cases with BI rupture and cases with intact BIs. Odds Ratio (OR) for each factor was calculated for all dichotomous variables. Association between rupture and potential risk factors was evaluated by multivariate analysis. A contingency table was prepared, linking breast implant integrity status with type of implant-based surgery, and degree of capsular contracture, then with all dichotomous variables in our database. Later, we separately analyzed whether breast pain was correlated with BI rupture using Pearson correlation coefficient.
Table 2
Dichotomous vs continuous variable subdivision
Dichotomous variables
Continuous variables
  Adjunct surgical procedures (i.e., autologous fat grafting)
  Capsular contracture
  Handedness
  History of trauma of the breast region
  History of mammographic imaging
  History of radiotherapy
  Breast implant integrity status (ruptured vs. intact)
  Implant manipulations or massages
  Physical activity and sports
  Sleeping in prone position
  Use of underwire bra
  Use of car seatbelt
  Weightlifting and use of heavy bags
  Body mass index (BMI)
  Breast implant weight
  Time of implant indwelling
  Type of housework
We used OR values to determine the association between BI rupture rate and all variables from the four selected categories. An OR equal to or above 1 was considered significant. Finally, logistic regression was used to identify independent predictors of implant rupture. All variables were considered statistically significant for p values less than 0.05, with a confidence interval of 95%. Hosmer–Lemeshow test was performed to assess goodness of fit for logistic regression model. All statistical analyses were performed using SPSS® software vers. 28.0 (IBM® Corporation; Armonk, New York).

Results

The starting population of 763 patients had a mean age of 48.9 years and consisted of 580 primary (76%) and 183 revisional cases (24%). Mean BI indwelling was of 12.2 years (range: 3.2–30.5). One-hundred-fifteen BI rupture patients were identified (15.1%). Mean BI lifespan was of 10.1 years (range: 0.6–29.4). One-hundred-ninety-one (25.0%) patients responded to the questionnaire between June 2021 and January 2022, for a total of 288 breasts. The population included 254 (88.1%) breast reconstructions following mastectomy and 34 (11.9%) non-oncologic cases (including cosmetic augmentations and corrections of congenital breast deformities). Placement was submuscular in 255 (88.5%) and subglandular in 33 cases (11.5%). We report that 250 (86.8%) BIs were manufactured by Allergan (Allergan Inc., Irvine, CA, USA), and that overall 95.1% of all analyzed devices had a macrotextured outer surface. Additionally, 253 (87.8%) implants were shaped, while 35 (12.2%) were round. Patient demographics, surgery and BI characteristics are detailed in Table 3. The study included 93 (32.3%) BI ruptures: 84 had been placed for breast reconstruction following mastectomy, four during cosmetic breast augmentation and five for correction of congenital breast deformities. In 12 patients, rupture of the implant was bilateral, while it was unilateral in 69. Capsular contracture was grade I–II in 235 (81.6%) and grade III-IV in 53 patients (18.4%) (Table 4). Rupture was defined according to preoperative Magnetic Resonance Imaging findings as intracapsular when the periprosthetic capsule was intact despite a breach of the implant shell, which occurred in 69 instances (74.2%). Extracapsular rupture was defined as the event in which silicone extravasation occurred, which represents 24 cases (25.8%). Location of BI rupture was detected during revisional surgery and accurately described in 73 surgical reports, but was unknown in 20 (21.5%). Rupture was located in one of four quadrants. If two or more quadrants were affected, the implant was divided in an anterior and posterior aspect. If three or more quadrants were affected, the implant was considered completely disrupted. Frequency of BI rupture location is depicted in Figure 1, but was not deemed statistically significant (Figure 1).
Table 3
Demographic, surgery and breast implant characteristics of the patient population that responded to the questionnaire
Patient characteristics
Value (percentage) [ranges]
Number of patients
191
Number of BIs
288 (100%)
Mean age
49.2 years [19–81]
Mean weight
61.4 kg [41–93]
Mean height
1.63 m [1.50–1.87]
Mean BMI
23.2 kg/m2 [14.9–36.3]
Median year of implantation
2007 [2003–2019]
Median year of explantation/replacement
2014 [2011–2022]
Implantation at our institution:
 
  Primary surgery
  244 (84.7%)
  Revisional surgery
  44 (15.3%)
Laterality of BI placement:
 
  Bilateral placement
  198 implants (68.8%)
  Unilateral placement
  90 implants (31.2%)
    Right side
    46 (51.1%)
    Left side
    44 (48.9%)
Indications for BI placement:
 
  Unilateral breast cancer
  183 (63.5%)
  Bilateral breast cancer
  58 (20.1%)
  Prophylactic mastectomy
  13 (4.5%)
  Cosmetic breast augmentation
  19 (6.6%)
  Correction of congenital breast deformity
  15 (5.2%)
Device placement:
 
  Submuscular
  255 (88.5%)
  Subglandular
  33 (11.5%)
Breast reconstruction timing:
 
  Immediate
  135 (53.1%)
  Delayed
  119 (46.9%)
Type of breast reconstruction:
 
  Implant-enhanced latissimus dorsi
  137 (53.9%)
  Tissue expander/implant exchange
  19 (7.5%)
  Wise pattern direct implant
  66 (26.0%)
  ADM assisted subpectoral
  32 (12.6%)
Breast implant manufacturers:
 
  Third generation BIs or older
  4 (1.4%)
  Allergan Inc.
  250 (86.8%)
  GC aesthetics/eurosilicone
  10 (3.5%)
  Mentor worldwide LLC
  7 (2.4%)
  Establishment labs
  8 (2.8%)
  Sientra/silimed
  7 (2.4%)
  Unknown
  2 (0.7%)
Breast implant shape:
 
  Anatomical
  253 (87.8%)
  Round
  35 (12.2%)
Breast implant outer texture:
 
  Macrotextured
  274 (95.1%)
  Microtextured
  0 (0%)
  Smooth
  12 (4.2%)
  Unknown
  2 (0.7%)
Mean BI size
351.8 cc [125–600]
BI integrity status:
 
  Intact
  195 (67.7%)
  Ruptured
  93 (32.3%)
    Intracapsular rupture
    69 (74.2%)
    Extracapsular rupture
    24 (25.8%)
Indication of ruptured implants
 
  Breast reconstruction following mastectomy (unilateral, bilateral and prophylactic)
  84 out of 254 (33.1%)
  Cosmetic breast augmentation
  4 out of 19 (21.5%)
  Correction of congenital breast deformity
  5 out of 15 (33.3%)
Laterality of patients with intact breast implants
110 (57.6%)
  Bilateral
  71 (64.5%)
  Unilateral
  39 (35.5%)
    Right
    22 (56.4%)
    Left
    17 (43.6%)
Laterality of patients with ruptured breast implants
81 (42.4 %)
  Bilateral rupture
  12 (14.8%)
  Unilateral rupture
  69 (85.2%)
    Right
    28 (40.5%)
    Left
    41 (59.4%)
Table 4
Postoperative care and lifestyle habits of the patient population
Patient characteristics
Value (percentage) [ranges]
Capsular contracture
 
  Baker I–II
  235 (81.6)
  Baker III–IV
  53 (18.4)
History of trauma
  Yes: 51 (17.7)
  No: 237 (82.3)
Use of underwire bra
  Yes: 45 (15.6)
  No: 243 (84.4)
Wear car seat belts
  Yes: 260 (90.3)
  No: 28 (9.7)
Prone sleep
  Yes: 91 (31.6)
  No: 197 (68.4)
Implant manipulations or massages
  Yes: 20 (6.9)
  No: 268 (93.1)
Physical activity and sports
  Yes: 139 (48.3)
  No: 149 (51.7)
Weightlifting
  Yes: 181 (62.8)
  No: 107 (37.2)
Type of housework
  None: 17 (5.9)
  Mild: 196 (68.1)
  Moderate: 36 (12.5)
  Heavy: 39 (13.5)
Use heavy purses and backpacks
  Yes: 182 (63.2)
  No: 106 (36.8)
Mammography imaging
  Yes: 136 (47.2)
  No: 152 (52.8)
In terms of postoperative care and lifestyle habits, the use of underwire bras [OR: 2.708 (1.242–5.903), 95% CI, p = 0.012], wear car seat belts [OR: 3.066 (1.959–9.805) 95% CI, p = 0.049] and prior mammographic imaging [OR: 2.196 (1.252–3.853), 95% CI, p = 0.006] increased risk for BI rupture; while weightlifting [OR: 0.407 (0.319–0.754), 95% CI, p = 0.004] and use of heavy purses or backpacks [OR: 0.347 (0.193–0.625), 95% CI, p < 0.001] did not increase BI rupture rate. All other factors, including prone sleep (p = 0.403), history of fall or trauma of the chest region (p = 110), physical activity (p = 0.114), implant manipulations (p) and radiotherapy (p = 0.806), were not deemed statistically significant (Table 5). Regarding patient-related risk factors for implant rupture, longer implant indwelling was statistically significant [OR: 1.040 (0.999–1,009), 95% confidence interval (CI), p = 0.015], while age, BMI and handedness were not. No surgery-related risk factor was statistically significant, including indication for surgery and adjunct surgical procedures (i.e., fat grafting, nipple reconstruction, balancing procedures). Among postoperative complications, capsular contracture and history of trauma were not significant (Figures 2 and 3).
Table 5
Multivariate analysis: significant factors associated with breast implant rupture
Parameter
Odds ratio
95% confidence limits
p value
Lower
Upper
Time
1.040
0.999
1.009
0.015
Underwire bras
2.708
1.242
5.903
0.012
Car seat belts
3.066
1.959
9.805
0.049
Mammographic imaging
2.196
1.252
3.853
0.006
Weightlifting
0.407
0.319
0.754
0.004
Heavy purses and backpacks
0.347
0.193
0.625
< 0.001

Discussion

Device rupture is considered one of the primary safety concerns presented by gel-filled BIs [10]. Despite great technological advancements achieved in the last decades, these devices still cannot be expected to last a lifetime and there is no agreement on their lifespan [7], though most surgeons will estimate it to be around 10–15 years [13, 14]. Nowadays, BI manufacturers recognize rupture as a potential complication [15].It is also fairly accepted that the longer the indwelling, the likelier it is for patients to experience adverse outcomes requiring surgery [16]. To strengthen BI safety, the FDA proposed black box warning that includes requesting manufacturers to improve labelling and patient information, outlining implant rupture as a relevant concern [17], as well as updating rupture screening recommendations [18]. Other means include pushing for improved reporting strategy of confirmed rupture cases, and conducting Premarket Approval (PMA) Core studies. Some Plastic Surgery societies have been proactive regarding the reporting of rupture cases. In fact, the European Association of Plastic Surgeons (EURAPS) was requested in 2018 by the European Commission to provide their expert opinion concerning the Device Specific Vigilance Guidance (DSVG) on BIs. The EURAPS stated that events such as rupture are well-known and highly clinically significant incidents, thus proposed them as appropriate to be reported as individual cases, rather than using Manufacturers’ Periodic Summary or Trend Reports [19]. However, individual reporting is a precious source of knowledge that will collectively take time to produce viable information. On the other hand, core studies are characterized by relatively limited sample sizes, thus we believe that they may not fully address all our concerns regarding device rupture. For this reason, the FDA advocates using other sources of information, such as retrospective or prospective studies to garner long-term rupture rate data [10].
On a related subject, there are staggering inconsistencies in literature regarding terminology and the definition of “BI lifespan.” Some authors consider it as the time between implantation and removal or replacement [14], while others define it as the time between implantation and rupture [14]. We chose the latter definition in our study. Terminology is key since implant rupture accounts for several cases of removal or exchange, but remains highly unpredictable [20]. Core studies report rupture rates of BIs produced by various manufacturers at 10 years follow-up, which range from 4.9 to 24.2% in cosmetic augmentations, and 9.8–35.4% in reconstruction cases [2127]. This data is summarized in Tables 5 and 6.
Table 6
Comparison of breast implant rupture rates across primary augmentation, revision augmentation, primary reconstruction and revision reconstruction patients
      
Primary augmentation
Revision augmentation
Primary reconstruction
Revision reconstruction
 
Tested implant
Manufacturer
Implant shape
Total population
Follow-up
Breast implant rupture rates
Caplin et al. [24]
MemoryGel
Mentor
Round
1008
10 years
24.2%
23.7%
32.7%
38.7%
Hammond et al. [25]
MemoryShape (contour profile gel)
Mentor
Shaped
955
10 years
6.6%
9.6%
18.9%
0%
El-Haddad et al. [27]
Sebbin silicone textured and smooth
Sebbin
Round
205
10 years
7.4%
21.2%
Duteille et al. [28]
Cristalline paragel
Eurosilicone S.A.S. (GC aesthetics)
Round and anatomical
995
10 years
4.9%
2.3%
0%
0%
Spear et al. [22]
Natrelle round
Allergan
Round
715
10 years
9.3%
5.4%
35.4%
Maxwell et al. [23]
Natrelle style 410
Allergan
Anatomical
941
10 years
17.7%
14.7%
12.4%
19.6%
Stevens et al. [26]
High-strength cohesive (HSC), high-strength cohesive plus (HSC+)
Sientra
Round and shaped
1788
10 years
7.8%
5.2%
9.8%
Data were retrieved from industry-funded core studies. “–“ is used when data in not available. The terms “Anatomical” and “Shaped” are used interchangeably throughout the table
BI rupture rate is defined as the ratio between ruptured implants (numerator) over the total number of implanted individuals (denominator). Obtaining a realistic account of rupture rate that is not limited by industry-sponsored studies is particularly challenging for three reasons. Firstly, data provided by manufacturers might overestimate implant longevity due to the nature of laboratory stress tests performed on BIs [28]. These stress tests are limited in their ability to reproduce in-vivo conditions of implant indwelling, meaning that they are not necessarily predictive of device failure [10]. Secondly, the number of women with BI in each country (i.e., the denominator) is not known, since National Implant Registries are either inexistent, not operational or not currently efficient [29, 30]. Thirdly, most BI ruptures are silent and clinically undetectable [31], which makes the numerator difficult to assess. In fact, some ruptures may come across as incidental findings from revisional surgeries performed for other reasons or diagnostic images requested for other health concerns [32]. It is worth nothing that 25.8% of all ruptures were extracapsular. Extracapsular rupture can sometimes represent the progression of an intracapsular one, with additional silicone seepage occurring over time [16, 33]. We speculate that BI ruptures likely occurred earlier than when radiologically detected.
Some authors proposed recommendations for optimal pre- and postoperative care of textured implants [34], or breast augmentations in general [35]. Nevertheless, standards of care are still elusive and many of the still used recommendations are not evidence-based, as they have not been proven to reduce incidence of rupture. Questions remain whether this event is a device-related complication or whether it may be influenced by certain factors. The most notable study addressing this topic was conducted by Feng et al. [36] in 1999. They analyzed potential risk factors in a population of 842 patients who received implant removal between 1990 and 1996. The authors found that the following factors were associated with a significantly higher risk of rupture rate: increasing age of implant, retroglandular placement, severe capsular contracture (Baker’s grades III and IV), and presence of local symptoms. In our study, 23 variables were selected, in accordance with the FDA’s recommendations regarding demographics, baseline characteristics and complications for patients in study cohorts [10]. Similarly to Feng et al.’s study, we found longer implant indwelling to increase rupture rate. However, his population differs significantly from our own, as 69.5% of their patients had undergone a cosmetic procedure, 28.2% received a reconstruction following breast cancer and 2.3% corrected a congenital deformity. Additionally, they reported that 67.4% received a subglandular placement while 32.6% had their BIs placed submuscularly. In our study, 88.1% of patients received a reconstruction following mastectomy, while 6.6% a cosmetic breast augmentation and 5.2% a congenital deformity correction. Implant pocket was submuscular in 88.5% of cases, and subglandular in 11.5%. Subglandular placement was not statistically significant in our study. The role of submuscular pockets in BI rupture is a controversial topic. Though Feng et al. found it to be protective compared to subglandular placements, Hadad et al. [37] found it to be a significant risk factor in their retrospective cohort study on 362 breast augmentation patients. Conversely, Hölmich et al. [38] concluded in their observational study on 533 cosmetic patients that no significant association was found with the position (subglandular or submuscular) of the implant. Additionally, despite featuring several implant-based reconstruction techniques in our study (submuscular direct-to-implant, implant-enhanced latissimus dorsi, tissue expander/implant exchange) we did not find submuscular placement to affect rupture rate in a statistically significant manner. Regarding the location of rupture, we found that most occurred in the upper quadrants. This is peculiar since the most exposed area and thus theoretically more at risk of rupture in cosmetic augmentations is the lower outer quadrant. This might be a result of the fact that our cohort mostly features reconstructive cases.
Our study, though limited by a small patient population, is focused on reconstructive surgery and features a long follow-up period. Furthermore, it was able to clarify the role of patient-related factors and lifestyle habits, which are poorly investigated in the literature. We found the use of underwire bras, car seat belts and prior mammographic screening to increase implant rupture rate. Avoidance of bras with an underwire in the postoperative period are a common recommendation for implant-based breast reconstructions [39], as they could impair healing of an inferior incision [40]. We inquired our patients on long-term use of underwire bras after surgery in order to clarify whether proscribing their use postoperatively was a myth or had a rationale. The reason why underwire bras might be associated with an increased rupture rate is unclear. We could speculate that it might stem from repeated and consistent mechanical wear-and-tear. However, if that were the case, we would have expected more ruptures in the inferior quadrants while the opposite occurred in our study. The mechanism of rupture might indeed be caused by the reduced expansibility of the lower pole which leads to its compression and migration of pressure cranially. Nevertheless, we currently advise patients against their use in the postoperative period.
Regarding the effect of car seat belts, we would like to acknowledge that their use is a legal obligation in Italy. Individuals may be exempt from their use when they present a health condition which poses a contraindication. Those include but are not limited to at risk pregnancies, severe obesity, severe incisional hernias, severe respiratory failure, use of specific shoulder/hip braces or orthopedic corsets, certain amputations or mutilations, and the insertion of medical devices, including BIs. The legal exemption is only valid when the health condition has been certified by a physician who agrees that the condition poses a contraindication. Song et al. [41] stated in their systematic review that rupture or damage correlated to car seat belts is caused by a moderate-to-severe compression of the chest which crushes BIs between the belt and the ribcage. While it is true that seat belts might cause implant rupture, in our practice we strongly recommend patients to abide by traffic safety rules and to use them anyway. The health risks stemming from not using the seat belt are likely more concerning than the risk of BI rupture. Therefore, despite the possibility of a legal exemption, we recommend patients to use the car seat belt. In our study group, 174 patients wore the seat belts and experience 90 ruptures (51.7%), while 17 patients (8.9%) did not use them anyway, 3 of which experienced implant rupture (17.6%). Additionally, the use of seat belts was considered a dichotomic variable: patients either used them consistently or did not. It was not correlated to their profession (i.e., cab drivers) or to the number of hours spent driving, which could have been used to distinguish continuous from incidental car seat belt usage. This potentially limits the weight of our findings and deserves further investigation. In regards to diagnostic imaging, BIs are radiopaque objects which might obscure breast tissue [42]. Patients with BIs can still undergo mammography imaging using the Eklund maneuver which consists in displacing the BI pushing it back against the chest wall while pulling the breast tissue forward so it can be seen in the mammogram [43]. Implant rupture is an inherent risk of this procedure, which radiographers may inadvertently cause due to lack of guidance about when to stop compression [44]. Nevertheless, despite our findings, the benefits provided by mammographic imaging for cancer detection largely outweigh the risk of BI rupture, which is why we do not recommend discontinuation of their use.
Regarding other statistically significant factors, we found the use of heavy purses/backpacks and daily weightlifting not to increase BI rupture rate. This should be clarified since those activities affect the chest causing repeated contraction of pectoralis major muscle, which has been shown to cause elevation of the breasts and development of wrinkles or ripples in the caudal and cranial quadrants [45]. The constant muscle contraction and stress exerted by physical activity could cause an intrinsic and regular pressure on the implant, with a wear-and-tear which should result in a higher likelihood of rupture. However, no reputable evidence has currently linked repeated contraction of pectoralis major muscle with implant rupture, and neither our study does. Besides the high prevalence of submuscular implants in our population, the low number of subglandularly placed devices hinders us from being able to obtain a significant comparison. Another possible confounding factor could be that we routinely perform primary division of the thoracodorsal nerve in our implant-enhanced latissimus dorsi patients, which represent more than half of our patient population (53.9%). This approach has been shown to effectively reduce muscle contraction surrounding the BI [46]. Of note, our results are in contrast with some of the beliefs supported by other authors regarding the role of capsular contracture in the likelihood of implant rupture. In fact, Feng et al. found that grade III/IV capsular contracture is associated with a higher risk of BI rupture [38].
When a cause for BI rupture can be identified, iatrogenic damage appears to be the most frequent culprit [9, 47]. It can take many forms which include surgical instrument damage. Autologous fat transfers (AFT) represent some of these instances, where infiltration cannulas may result in the accidental BI puncture [48]. AFT is commonly performed in reconstructive patients, either as adjunct procedure to correct implant-related deformities such as waving and rippling, or to mitigate the outcomes of radiotherapy[49, 50]. Of note, out of 77 cases where AFT procedures were performed, 20 cases had implant rupture, and three were specifically found from their surgical report to have puncture sites breaching the implants’ shell with fat tissue collections inside the devices. Nevertheless, AFT and adjunct surgical procedures in general were not found to statistically increase the risk of BI rupture in our study. We believe that this might be due to certain intraoperative strategies which have been implemented to reduce the risk of accidental BI damage, such as ensuring subcutaneous visibility of the injection cannula’s tip while it is being inserted, to avoid plunging the cannula too deep.

Conclusion

We believe that providing accurate patient information before undergoing any BI procedure is instrumental, and we believe that disregarding recommendations with no scientific backing is one major step toward that objective. In our mostly reconstructive population, patient and operative variables did not significantly influence implant survival. The same could not be said regarding the use of underwire bras, car seat belts and mammography imaging which have shown to increase the likelihood of BI rupture. Despite our findings, while we can recommend patients not to use underwire bras postoperatively, the life-saving benefits from using car seat belts and mammographic imaging cannot be dismissed, regardless of the increased rupture risk. On the other hand, several historical concerns regarding postoperative management are long outmoded, including the fact that patients should be discouraged from lifting heavy weights to reduce the risk of rupture, and we speculate that increased muscular activity does not constitute a higher risk factor. Special considerations should be mentioned for AFT procedures in implant-based reconstructions, which are safe and are not linked with a higher likelihood of rupture. Finally, our evidence confirms that implant rupture rate is directly correlated with duration of implant indwelling. Our findings aim at improving everyday clinical practice with BIs. Nevertheless a larger scale study with greater cohorts could further highlight the influence of operative variables and patient lifestyle habits on implant survival and define standard guidelines.

Declarations

Conflict of interest

The NESMOS Department has received research funds from Motiva, Establishment Labs in 2017, and from GC Aesthetics in 2018 and 2020. The Department also received miniature breast implants from Establishment Labs, GC Aesthetics and Sebbin for research purposes in 2017. The Senior author would like to disclose that they serve as a paid consultant for BellaSeno GmbH, and for Notified Body 0373 carrying out CE Mark certification activities for the Italian Ministry of Health for the year 2021 and 2022, and that they received reimbursements from ICEAG in 2015 and SCHEER-WG in 2019, 2020 and 2021. They have no ownership or investment to disclose. Neither the corresponding author nor any of the coauthors have received any type of funding or other support from a sponsor/commercial entity/third party. All other authors hereby certify that, to the best of their knowledge, no financial support or benefit has been received, neither by themselves directly, nor by any member of their immediate family or any individual or entity with whom or with which they may have a significant relationship from any commercial source which is related directly or indirectly to the scientific work which is reported on in the article.
The datasets obtained from the questionnaire used in this study did not contain personal, identifiable data, thus ethical committee approval and is not required for this type of study, but informed consent was administered to all patients.

Human or Animal Rights

This retrospective observational study was conducted in accordance with the World Medical Association’s Declaration of Helsinki for ethical principles in medical research involving human subjects.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Dent – Das Online-Abo der Zahnmedizin

Online-Abonnement

Mit e.Dent erhalten Sie Zugang zu allen zahnmedizinischen Fortbildungen und unseren zahnmedizinischen und ausgesuchten medizinischen Zeitschriften.

Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
12.
Zurück zum Zitat U.S. Department of Health and Human Services, Food and Drug Administration, Center for Devices and Radiological Health (2020) Breast implants - certain labeling recommendations to improve patient communication. [Online] available at: https://www.fda.gov/media/131885/download. Accessed 28 Nov 2021 U.S. Department of Health and Human Services, Food and Drug Administration, Center for Devices and Radiological Health (2020) Breast implants - certain labeling recommendations to improve patient communication. [Online] available at: https://​www.​fda.​gov/​media/​131885/​download. Accessed 28 Nov 2021
13.
Zurück zum Zitat Benadiba L, Pétoin DS, Berlie J, Rouëssé J, Girard M (2000) Durée de vie des implants mammaires en reconstruction mammaire. A propos de 949 implants [Survivorship of breast implants used in breast reconstruction. 949 implants]. Ann Chir Plast Esthet 45(1):31–40PubMed Benadiba L, Pétoin DS, Berlie J, Rouëssé J, Girard M (2000) Durée de vie des implants mammaires en reconstruction mammaire. A propos de 949 implants [Survivorship of breast implants used in breast reconstruction. 949 implants]. Ann Chir Plast Esthet 45(1):31–40PubMed
22.
Zurück zum Zitat Maxwell GP, Van Natta BW, Bengtson BP, Murphy DK (2015) Ten-year results from the Natrelle 410 anatomical form-stable silicone breast implant core study [published correction appears in Aesthet Surg J. 2015;35(8):1044]. Aesthet Surg J 35(2): 145–155. https://doi.org/10.1093/asj/sju084 Maxwell GP, Van Natta BW, Bengtson BP, Murphy DK (2015) Ten-year results from the Natrelle 410 anatomical form-stable silicone breast implant core study [published correction appears in Aesthet Surg J. 2015;35(8):1044]. Aesthet Surg J 35(2): 145–155. https://​doi.​org/​10.​1093/​asj/​sju084
25.
26.
Zurück zum Zitat El-Haddad R, Lafarge-Claoue B, Garabedian C, Staub S (2018) A 10-year prospective study of implant-based breast augmentation and reconstruction. Eplasty 18:e7PubMedPubMedCentral El-Haddad R, Lafarge-Claoue B, Garabedian C, Staub S (2018) A 10-year prospective study of implant-based breast augmentation and reconstruction. Eplasty 18:e7PubMedPubMedCentral
29.
Zurück zum Zitat Santanelli di Pompeo F, Sorotos M, Clemens MW, Firmani G (2021) European association of plastic surgeons (EURAPS) committee on device safety and development. Breast implant-associated anaplastic large cell lymphoma (BIA-ALCL): review of epidemiology and prevalence assessment in Europe. Aesthet Surg J 41(9):1014–1025. https://doi.org/10.1093/asj/sjaa285CrossRefPubMed Santanelli di Pompeo F, Sorotos M, Clemens MW, Firmani G (2021) European association of plastic surgeons (EURAPS) committee on device safety and development. Breast implant-associated anaplastic large cell lymphoma (BIA-ALCL): review of epidemiology and prevalence assessment in Europe. Aesthet Surg J 41(9):1014–1025. https://​doi.​org/​10.​1093/​asj/​sjaa285CrossRefPubMed
30.
Zurück zum Zitat Santanelli di Pompeo F, Clemens MW, Atlan M, et al. (2022) Practice recommendation updates from the world consensus conference on BIA-ALCL [published online ahead of print, 2022 May 26]. Aesthet Surg J 2022;sjac133. https://doi.org/10.1093/asj/sjac133 Santanelli di Pompeo F, Clemens MW, Atlan M, et al. (2022) Practice recommendation updates from the world consensus conference on BIA-ALCL [published online ahead of print, 2022 May 26]. Aesthet Surg J 2022;sjac133. https://​doi.​org/​10.​1093/​asj/​sjac133
33.
Zurück zum Zitat O’Brien W (2010) Top 3 differentials in radiology: a case review. Section XII “breast imaging.” Thieme Medical Pub, New York O’Brien W (2010) Top 3 differentials in radiology: a case review. Section XII “breast imaging.” Thieme Medical Pub, New York
35.
Zurück zum Zitat Tebbetts JB (2002) Achieving a predictable 24-hour return to normal activities after breast augmentation: part I. Refining practices by using motion and time study principles. Plast Reconstr Surg 109(1):273–292CrossRefPubMed Tebbetts JB (2002) Achieving a predictable 24-hour return to normal activities after breast augmentation: part I. Refining practices by using motion and time study principles. Plast Reconstr Surg 109(1):273–292CrossRefPubMed
42.
Zurück zum Zitat Institute of Medicine (US) (1999) Committee on the safety of silicone breast implants. In: Bondurant S, Ernster V, Herdman R, eds. Safety of silicone breast implants. Washington (DC): National Academies Press (US). 12, Silicone Implants and Breast Imaging. Available from: https://www.ncbi.nlm.nih.gov/books/NBK44781/ Institute of Medicine (US) (1999) Committee on the safety of silicone breast implants. In: Bondurant S, Ernster V, Herdman R, eds. Safety of silicone breast implants. Washington (DC): National Academies Press (US). 12, Silicone Implants and Breast Imaging. Available from: https://​www.​ncbi.​nlm.​nih.​gov/​books/​NBK44781/​
Metadaten
Titel
Assessment of Risk Factors for Rupture in Breast Reconstruction Patients with Macrotextured Breast Implants
verfasst von
Guido Paolini
Guido Firmani
Francesca Briganti
Mattia Macino
Simone Nigrelli
Michail Sorotos
Fabio Santanelli di Pompeo
Publikationsdatum
13.10.2022
Verlag
Springer US
Erschienen in
Aesthetic Plastic Surgery / Ausgabe 2/2023
Print ISSN: 0364-216X
Elektronische ISSN: 1432-5241
DOI
https://doi.org/10.1007/s00266-022-03118-9

Neu im Fachgebiet Chirurgie

Verbände und Cremes gegen Dekubitus: „Wir wissen nicht, was sie bringen!“

Die Datenlage zur Wirksamkeit von Verbänden oder topischen Mitteln zur Prävention von Druckgeschwüren sei schlecht, so die Verfasser einer aktuellen Cochrane-Studie. Letztlich bleibe es unsicher, ob solche Maßnahmen den Betroffenen nutzen oder schaden.

Nackenschmerzen nach Bandscheibenvorfall: Muskeltraining hilft!

Bei hartnäckigen Schmerzen aufgrund einer zervikalen Radikulopathie schlägt ein Team der Universität Istanbul vor, lokale Steroidinjektionen mit einem speziellen Trainingsprogramm zur Stabilisierung der Nackenmuskulatur zu kombinieren.

US-Team empfiehlt Gastropexie nach Hiatushernien-Op.

Zur Vermeidung von Rezidiven nach Reparatur einer paraösophagealen Hiatushernie sollte einem US-Team zufolge der Magen bei der Op. routinemäßig an der Bauchwand fixiert werden. Das Ergebnis einer randomisierten Studie scheint dafür zu sprechen.

Mit Lidocain kommt der Darm nicht schneller in Schwung

Verzögertes Wiederanspringen der Darmfunktion ist ein Hauptfaktor dafür, wenn Patientinnen und Patienten nach einer Kolonresektion länger als geplant im Krankenhaus bleiben müssen. Ob man diesem Problem mit Lidocain vorbeugen kann, war Thema einer Studie.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.