Assessment of the corneal biomechanical features of granular corneal dystrophy type 2 using dynamic ultra-high-speed Scheimpflug imaging

verfasst von: Akira Tanikawa, Takeshi Soma, Atsuya Miki, Shizuka Koh, Yoshiyuki Kitaguchi, Naoyuki Maeda, Yoshinori Oie, Satoshi Kawasaki, Kohji Nishida

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 3/2023

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the corneal biomechanical features of eyes with granular corneal dystrophy type 2 (GCD2) by analyzing corneal biomechanical indices obtained using a Corvis ST (CST) dynamic ultra-high-speed Scheimpflug imaging device.

Methods

In this retrospective case–control study, 35 CST parameters were compared in normal eyes (control) and eyes of patients with GCD2 treated at Osaka University Hospital, Osaka, Japan. The parameters included the Corvis Biomechanical Index (CBI), which is important in differentiating eyes with keratoconus from normal eyes. We measured the deposition rates of lesions in the central 7-mm region of the eye and assessed the correlation between the deposition rate and the CBI.

Results

Twenty-one eyes with GCD2 and 23 control eyes were analyzed. Eyes with GCD2 showed significantly less corneal stiffness in 15 CST parameters than did control eyes. In particular, the CBI was remarkably higher in eyes with GCD2 than in control eyes (P = 0.000006). Additionally, the deposition rate and the CBI were positively correlated.

Conclusions

GCD2 eyes had softer corneas than did control eyes in most biomechanical CST parameters, and one of the parameters (the CBI) was linked to the rate of deposited lesions. Since IOP may be underestimated in GCD2 eyes, management should be especially careful in GCD2 cases complicated by glaucoma.
Literatur
16.
18.
Zurück zum Zitat Hanan F, Ali Z, Naeem M (2020) Accuracy of the Corvis biomechanical index in keratoconus screening. Pak J Ophthalmol 36:3CrossRef Hanan F, Ali Z, Naeem M (2020) Accuracy of the Corvis biomechanical index in keratoconus screening. Pak J Ophthalmol 36:3CrossRef
Metadaten
Titel
Assessment of the corneal biomechanical features of granular corneal dystrophy type 2 using dynamic ultra-high-speed Scheimpflug imaging
verfasst von
Akira Tanikawa
Takeshi Soma
Atsuya Miki
Shizuka Koh
Yoshiyuki Kitaguchi
Naoyuki Maeda
Yoshinori Oie
Satoshi Kawasaki
Kohji Nishida
Publikationsdatum
30.09.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 3/2023
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-022-05847-8

Weitere Artikel der Ausgabe 3/2023

Graefe's Archive for Clinical and Experimental Ophthalmology 3/2023 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.