Open Access 01.12.2018 | Research

Assessment of the modulation degrees of intensity-modulated radiation therapy plans

verfasst von: So-Yeon Park, Jung-in Kim, Minsoo Chun, Hyunjun Ahn, Jong Min Park

Erschienen in: Radiation Oncology | Ausgabe 1/2018

Abstract

Background

To evaluate the modulation indices (MIs) for predicting the plan delivery accuracies of intensity-modulated radiation therapy (IMRT) plans.

Methods

A total of 100 dynamic IMRT plans that used TrueBeam STx and 102 dynamic IMRT plans that used Trilogy were selected. For each plan, various MIs were calculated, which included the modulation complexity score (MCS), plan-averaged beam area (PA), plan-averaged beam irregularity (PI), plan-averaged beam modulation (PM), MI quantifying multi-leaf collimator (MLC) speeds (MIs), MI quantifying MLC acceleration (MIa), and MI quantifying MLC acceleration and segment aperture irregularity (MIc,IMRT). To determine plan delivery accuracy, global gamma passing rates, MLC errors of log files, and dose-volumetric parameter differences between original and log file-reconstructed IMRT plans were obtained. To assess the ability of each MI for predicting plan delivery accuracy, Spearman’s rank correlation coefficients (rs) between MIs and plan delivery accuracy measures were calculated.

Results

PI showed moderately strong correlations with gamma passing rates in MapCHECK2 measurements of both TrueBeam STx and Trilogy (rs = − 0.591 with p <  0.001 and − 0.427 with p <  0.001 to with gamma criterion of 2%/2 mm, respectively). For ArcCHECK measurements, PI also showed moderately strong correlations with the gamma passing rates in the ArcCHECK measurements of TrueBeam STx and Trilogy (rs = − 0.545 with p <  0.001 and rs = − 0.581 with p <  0.001 with gamma criterion of 2%/2 mm, respectively). The PI showed the second strongest correlation with MLC errors in both TrueBeam STx and Trilogy (rs = 0.861 with p <  0.001 and rs = 0.767 with p <  0.001, respectively). In general, the PI showed moderately strong correlations with every plan delivery accuracy measure.

Conclusions

The PI showed moderately strong correlations with every plan delivery accuracy measure and therefore is a useful predictor of IMRT delivery accuracy.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1186/​s13014-018-1193-9) contains supplementary material, which is available to authorized users.
Abkürzungen
AAA
Anisotropic analytic algorithm
DVO
Dose volume optimizer
FFF
Flattening filter free
HN
Head and neck
IMRT
Intensity-modulated radiation therapy
IRB
Institutional review board
MCS
Modulation complexity score
MI
Modulation index
MLC
Multi-leaf collimator
MU
Monitor unit
OAR
Organ at risk
PA
Plan-averaged beam area
PI
Plan-averaged beam irregularity
PM
Plan-averaged beam modulation
QA
Quality assurance
r s
Spearman’s rho
SABR
Stereotactic ablative radiotherapy
SIB
Simultaneous integrated boost
TPS
Treatment planning system
VMAT
Volumetric modulated arc therapy

Background

Intensity-modulated radiation therapy (IMRT) delivers highly conformal radiation dose distributions to target volumes, while sparing normal tissues [18]. As the IMRT plan complexity increases to acquire the desired dose distribution, i.e., the modulation degree of IMRT increases, a large number of small and irregularly shaped beam apertures are utilized. It results in increased dose calculation and multi-leaf collimator (MLC) movement uncertainties [4, 911]. Therefore, the probability that the intended dose distribution is not accurately delivered to the patient during treatment increases. In this respect, verification of IMRT plan delivery accuracy is highly recommended before treatment in a clinical setting.
To verify IMRT plan delivery accuracy, gamma evaluations have been widely adopted in the clinic, as pre-treatment quality assurance (QA) methods [1214]. Although gamma evaluation is a practical method, it has some limitations. It has been reported that the gamma passing rates of individual dosimeters have been inconsistent with respect to each other [15]. Therefore, a thorough gamma evaluation validation of each device that is installed in the clinic should be performed. Moreover, recent studies have demonstrated the clinical irrelevance of the gamma passing rates [1618]. As an alternative to the gamma evaluation, an analysis of the machine log files that are recorded during delivery has been recommended as a pre-treatment QA by several studies [1922]. This method can be expanded to analyze the differences between dose-volume parameters of the original plan and the plans that are reconstructed with machine log files [2325]. However, this method has limitations, as it does not provide an independent verification of linac beam delivery.
On the other hand, to predict IMRT delivery accuracy and calculate the complexity of IMRT plans, several studies have introduced modulation indices (MIs) [4, 10, 26, 27]. Webb et al. measured the modulation degree of IMRT plans by quantifying the changes of adjacent beamlets that exceeded a pre-defined tolerance levels, derived with standard deviations of the fluence map [10]. McNiven et al. designed modulation complexity score (MCS) quantifications of beam aperture and MLC position variations. The MCSs were found to vary from 0 to 1 and decreased as the IMRT plan modulation degrees increased [27]. Du et al. quantified the complexities of both the IMRT and volumetric modulated arc therapy (VMAT) plans, using the beam aperture areas and perimeters, to quantify the usage of small and irregularly shaped beam apertures [4]. For VMAT, Park et al. expanded the concept of Webb’s MI to comprehensively assess MLC speeds and accelerations, the variation of gantry rotation speed, the variation of dose-rate, and the usage of small and irregularly shaped beam apertures [26].
Previous studies have investigated the performance of several MIs for predicting IMRT delivery accuracy, through the use of correlation analyses on conventional plan delivery accuracy measures, such as gamma passing rates. However, there are limitations to evaluating MIs with a single type of linac, single photon energy, limited treatment sites, or using a single type of measuring device. Therefore, further studies need to evaluate the performance of various MIs that are used as clinical predictors of IMRT plan deliveries. In this study, we assessed the performances of various MIs that have been suggested by previous studies to predict IMRT plan delivery accuracy with correlation analyses.

Methods

Patient selection and IMRT planning

Two machines, Trilogy and TrueBeam STx (Varian Medical Systems, Palo Alto, CA), which have different plan delivery systems and machine log file generation systems, were selected for this study. After approval by the institutional review board (IRB), a total of 181 patients with various treatment sites were retrospectively selected, and a total of 202 dynamic IMRT plans were generated for this study.
With the TrueBeam STx system, a total of 100 dynamic IMRT plans were generated for head and neck (HN) cancer (20 cases), brain tumor (20 cases), liver cancer (20 cases), spine tumor (20 cases), and lung cancer (20 cases). For HN cancer, a simultaneous integrated boost (SIB) technique was applied. For liver cancer, spine tumors, and lung cancer, stereotactic ablative radiotherapy (SABR) was performed using flattening filter free (FFF) photon beams of 6 MV (6 FFF) and 10 FFF. With Trilogy, a total of 102 dynamic IMRT plans were generated for HN cancer (40 cases), prostate cancer (21 primary plans and 21 boost plans), liver cancer (11 cases), and spine tumor (9 cases). The HN IMRT plans were generated with the SIB technique as with TrueBeam STx system. Information of IMRT plans generated in this study is summarized in Additional file 1: Table S1.
Every IMRT plan in this study was generated in Eclipse™ (Varian Medical Systems, Palo Alto, CA). All IMRT plans used in this study were optimized with the dose volume optimizer (DVO, ver. 10, Varian Medical Systems, Palo Alto, CA), and the dose distributions were calculated using the anisotropic analytic algorithm (AAA, ver. 10, Varian Medical Systems, Palo Alto, CA), with a calculation grid of 1 mm. All IMRT plans were normalized to cover 100% of the target volumes with 90% of the prescription doses.

Calculation of IMRT plan modulation indices

For each IMRT plan, the MCS from McNiven et al.; the plan-averaged beam area (PA), plan-averaged beam irregularity (PI), and plan-averaged beam modulation (PM) from Du et al.; the modulation index quantifying the speed of MLC (MIs); the modulation index for quantifying the MLC acceleration (MIa); and the modulation index for quantifying the MLC acceleration and aperture irregularity of the IMRT plan (MIc,IMRT) were calculated. Although MIc,IMRT values were originally designed for VMAT, we modified these indices to consider only MLC movements for application to IMRT plans. In other words, we excluded gantry rotation accuracy and dose-rate variability evaluations from the MIc,IMRT value. To eliminate the influence of various segment numbers on the MCS, MIs, MIa, and MIc,IMRT values, we normalized the values of these MIs with the total segment numbers. Equations of the MIs that were calculated in this study (including modifications) are summarized in Additional file 1: Table S2.

Measures of IMRT plan delivery accuracy

To evaluate plan delivery accuracy, three types of IMRT plan verification methods were used in this study, which included the global gamma evaluation with absolute doses, differences in MLC positions between planned and the actual deliveries utilizing machine log files, and the differences in the dose-volumetric parameters that were calculated from the original IMRT plans and from the IMRT plans that were reconstructed with machine log files.
For global gamma evaluation, reference dose distributions were calculated in the Eclipse system, with a calculation grid size of 1 mm. The calculated dose distributions were compared to the measured dose distributions using MapCHECK2™ and ArcCHECK™ detector arrays (Sun Nuclear Corporation, Melbourne, FL). Global gamma evaluations with gamma criteria of 3%/3 mm, 2%/2 mm, 2%/1 mm, and 1%/2 mm were performed with absolute doses. Percent dose differences in the global gamma evaluation were calculated relative to the maximum dose in the calculated dose distribution. Before the gamma evaluation measurements, MapCHECK2 and ArcCHECK dosimeters were calibrated, according to manufacturer guidelines.
When IMRT plans were delivered to the dosimeters for gamma evaluation, the actual MLC positions and delivered monitor units (MUs) at each segment were obtained from the DynaLog (Trilogy) and Trajectory (TrueBeam STx) files. These log files were reconstructed as DICOM-RT files, using an in-house program that was written in MATLAB (R2016a, Mathworks Inc., Natick, MA). The differences in the MLC positions between the original IMRT plan and the actual delivery were calculated and averaged for each IMRT plan. After the reconstructed DICOM-RT files from the log files were imported in the Eclipse system, dose distributions were calculated with the same dose calculation condition of the original IMRT plan. With the dose distributions calculated from reconstructed IMRT plans, dose-volumetric parameters for the target volumes and organs at risk (OARs) were calculated and compared to those of the original IMRT plans. The total number of analyzed dose volumetric parameters of TrueBeam STx and Trilogy were 156 and 152, respectively.

Correlation analysis

To evaluate the performance of each MI that was calculated in this study as an indicator of IMRT delivery accuracy, Spearman’s rho (rs) and corresponding p values were calculated between the MIs and the IMRT delivery accuracy measures. In the correlation analysis of different dose-volumetric parameters, the sample sizes were different from one another and the numbers of analyzed dose-volumetric parameters were large. For these reasons, we only counted the number of rs values with corresponding p values less than 0.05, which were considered as statistically significant in this study. The correlations with rs values from 0.2 to 0.39 are regarded as weak correlations. Those from 0.4 to 0.59 and 0.6 to 1 are regarded as moderately strong and very strong correlations, respectively [27, 28].

Results

The values of modulation indices

The values of MIs are shown in Table 1. The values of MIs showed various tendencies, depending on the type of linac and treatment sites.
Table 1
The average values of various modulation indices for intensity modulated radiation therapy plans
 
MIs (×  10− 3)
MIa (×  10− 2)
MIc, IMRT (×  10− 1)
MCS (×  10− 1)
PA (×  10− 1)
PI (×  10− 1)
PM
TrueBeam STx
 Lung SABR (n = 20)
7.97 ± 1.41
2.67 ± 0.69
0.31 ± 0.08
1.25 ± 0.18
1.01 ± 0.45
0.73 ± 0.18
0.67 ± 0.05
 Spine SABR (n = 20)
5.62 ± 1.48
1.46 ± 0.48
0.17 ± 0.05
0.64 ± 0.19
1.13 ± 0.53
1.53 ± 1.35
0.86 ± 0.05
 Liver SABR (n = 20)
7.83 ± 2.46
2.40 ± 0.60
0.27 ± 0.06
1.11 ± 0.42
1.39 ± 0.67
1.03 ± 0.65
0.73 ± 0.12
 Brain (n = 20)
9.20 ± 2.56
9.06 ± 2.62
1.01 ± 0.28
1.40 ± 0.46
3.72 ± 1.85
0.70 ± 0.27
0.67 ± 0.13
 Head and neck (n = 20)
6.12 ± 2.24
7.06 ± 1.52
0.82 ± 0.17
0.35 ± 0.08
3.72 ± 0.82
2.30 ± 0.43
0.93 ± 0.02
Trilogy
 Head and neck (n = 40)
4.14 ± 1.11
5.94 ± 0.83
0.70 ± 0.10
0.30 ± 0.05
2.27 ± 0.41
1.87 ± 0.22
0.94 ± 0.01
 Prostate primary (n = 21)
3.34 ± 0.71
4.95 ± 0.79
0.58 ± 0.09
0.81 ± 0.14
0.95 ± 0.22
0.64 ± 0.11
0.79 ± 0.04
 Prostate boost (n = 21)
3.18 ± 0.43
4.89 ± 0.65
0.57 ± 0.07
0.94 ± 0.13
0.90 ± 0.14
0.47 ± 0.10
0.72 ± 0.04
 Liver (n = 11)
5.14 ± 2.14
6.93 ± 1.40
0.80 ± 0.16
0.78 ± 0.24
1.72 ± 0.60
0.94 ± 0.54
0.79 ± 0.07
 Spine (n = 9)
5.46 ± 2.68
4.02 ± 1.55
0.47 ± 0.17
0.63 ± 0.25
2.01 ± 1.40
1.26 ± 0.82
0.86 ± 0.06
Note: MIs Modulation index considering multi-leaf collimator speed, MIa Modulation index considering multi-leaf collimator acceleration, MIc, IMRT Modulation index considering multi-leaf collimator acceleration and field aperture irregularity for intensity modulated radiation therapy, MCS Modulation complexity score, PA Plan-averaged beam area, PI Plan-averaged beam irregularity, PM Plan-averaged beam modulation, SABR Stereotactic ablative radiotherapy, Prostate primary Prostate primary plan, Prostate boost Prostate boost plan

Measurement of plan delivery accuracies

The global gamma passing rates with gamma criteria of 3%/3 mm, 2%/2 mm, 2%/1 mm, and 1%/2 mm are shown in Table 2. The mean gamma passing rate for each treatment site was remarkably different from one another at gamma criteria of 2%/1 mm and 1%/2 mm.
Table 2
Average values of global gamma passing rates of intensity modulated radiation therapy plans
 
3%/3 mm
2%/2 mm
2%/1 mm
1%/2 mm
MC
AC
p
MC
AC
p
MC
AC
p
MC
AC
p
TrueBeam
 Lung SABR (n = 2)
98.5 ± 2.3
99.4 ± 0.6
< 0.001
93.8 ± 4.0
97.7 ± 1.7
< 0.001
85.3 ± 7.1
91.1 ± 3.6
< 0.001
75.2 ± 11.7
95.9 ± 2.7
< 0.001
 Spine SABR (n = 20)
98.4 ± 2.7
99.3 ± 0.9
< 0.001
95.8 ± 4.1
97.9 ± 1.5
< 0.001
91.0 ± 7.1
87.6 ± 4.3
0.047
91.1 ± 4.3
95.6 ± 2.6
< 0.001
 Liver SABR (n = 20)
99.2 ± 0.9
99.4 ± 1.2
< 0.001
96.7 ± 2.4
97.9 ± 3.1
< 0.001
92.7 ± 4.0
88.8 ± 7.3
0.013
91.1 ± 5.3
95.8 ± 4.7
< 0.001
 Brain (n = 20)
98.8 ± 1.6
99.3 ± 2.4
0.002
96.4 ± 3.3
98.5 ± 4.2
0.002
91.6 ± 4.7
92.1 ± 7.3
0.761
89.8 ± 7.1
97.2 ± 4.9
< 0.001
 Head and neck (n = 20)
96.2 ± 3.4
97.5 ± 3.6
0.072
84.8 ± 8.3
89.1 ± 10.3
0.107
74.5 ± 10.8
80.5 ± 13.9
0.996
66.3 ± 9.9
71.6 ± 13.0
0.001
Trilogy
 Head and neck (n = 40)
99.4 ± 0.6
99.8 ± 0.7
0.088
96.3 ± 2.4
98.5 ± 2.1
< 0.001
90.1 ± 3.9
94.4 ± 3.7
0.003
87.8 ± 5.3
94.8 ± 3.9
< 0.001
 Prostate primary (n = 21)
98.2 ± 0.8
99.5 ± 0.4
0.162
93.9 ± 1.4
97.1 ± 1.0
0.042
84.8 ± 3.0
86.1 ± 2.7
0.077
85.4 ± 1.7
93.4 ± 1.8
< 0.001
 Prostate boost (n = 21)
97.3 ± 1.5
99.4 ± 0.5
0.624
89.4 ± 2.6
96.2 ± 1.9
0.187
78.9 ± 5.0
83.6 ± 4.9
0.043
73.1 ± 4.9
92.5 ± 2.9
0.005
 Liver (n = 11)
99.2 ± 0.8
99.8 ± 0.2
0.418
97.3 ± 2.6
98.2 ± 1.4
0.083
93.9 ± 3.9
92.3 ± 4.2
0.814
90.4 ± 4.1
94.5 ± 2.9
< 0.001
 Spine (n = 9)
99.2 ± 0.9
99.7 ± 0.4
0.246
96.4 ± 3.4
97.3 ± 1.6
0.152
91.5 ± 6.6
89.3 ± 6.8
0.135
91.0 ± 5.5
92.5 ± 2.5
0.153
Note: MC MapCHECK2™ detector array, AC ArcCHECK™ detector array, SABR Stereotactic ablative radiotherapy, Prostate primary Prostate primary plan, Prostate boost Prostate boost plan
The differences between the planned and actual MLC positions are shown in Fig. 1. The average MLC differences of TrueBeam STx and Trilogy were 0.06 mm and 0.25 mm, respectively. For TrueBeam STx, the average MLC difference of the brain IMRT plans was the highest, at 0.13 mm, while for Trilogy, the average MLC difference of the HN IMRT plans was the highest (0.37 mm). The MLC errors of TrueBeam STx were much smaller than those of the Trilogy system.

Correlations between gamma passing rates and modulation indices

The rs values of MIs for the global gamma passing rates, with gamma criteria of 3%/3 mm, 2%/2 mm, 2%/1 mm and 1%/2 mm, of the TrueBeam STx and Trilogy systems are shown in Tables 3 and 4, respectively. The results presented the rs values with statistically significant p values less than 0.05.
Table 3
Correlations between global gamma passing rates with various gamma criteria and the modulation indices for intensity modulated radiation therapy plans of TrueBeam STx (p ≤ 0.05)
 
3%/3 mm
2%/2 mm
2%/1 mm
1%/2 mm
r s
p
r s
p
r s
p
r s
p
MapCHECK2
 MIs
− 0.241
0.016
− 0.200
0.046
 MIa
 MIc, IMRT
− 0.201
0.045
 MCS
0.453
< 0.001
0.548
< 0.001
0.509
< 0.001
0.460
< 0.001
 PA
−0.223
0.026
− 0.236
0.018
− 0.196
0.050
 PI
−0.465
< 0.001
−0.591
< 0.001
− 0.550
< 0.001
− 0.474
< 0.001
 PM
− 0.476
< 0.001
− 0.566
< 0.001
− 0.516
< 0.001
− 0.438
< 0.001
ArcCHECK
 MIs
− 0.375
< 0.001
− 0.341
0.001
−0.518
< 0.001
− 0.292
0.003
 MIa
−0.234
0.019
− 0.234
0.019
 MIc, IMRT
−0.225
0.025
−0.228
0.022
 MCS
0.431
< 0.001
0.593
< 0.001
0.451
< 0.001
0.699
< 0.001
 PA
−0.235
0.019
− 0.225
0.024
 PI
−0.340
0.001
−0.545
< 0.001
−0.328
0.001
−0.677
< 0.001
 PM
−0.369
< 0.001
−0.545
< 0.001
− 0.389
< 0.001
− 0.667
< 0.001
Note: rs Spearman’s rho, MIs Modulation index considering multi-leaf collimator speed, MIa Modulation index considering multi-leaf collimator acceleration, MIc, IMRT Modulation index considering multi-leaf collimator acceleration and field aperture irregularity for intensity modulated radiation therapy, MCS Modulation complexity score, PA Plan-averaged beam area, PI Plan-averaged beam irregularity, PM Plan-averaged beam modulation
Table 4
Correlations between global gamma passing rates with various gamma criteria and the modulation indices for intensity modulated radiation therapy plans of Trilogy (p ≤ 0.05)
 
3%/3 mm
2%/2 mm
2%/1 mm
1%/2 mm
r s
p
r s
p
r s
p
r s
p
MI
MapCHECK2
 MIs
−0.398
< 0.001
−0.386
< 0.001
− 0.347
< 0.001
− 0.323
0.001
 MIa
−0.302
0.002
−0.305
0.002
−0.397
< 0.001
− 0.210
0.034
 MIc, IMRT
−0.315
< 0.001
−0.320
< 0.001
− 0.400
< 0.001
− 0.230
0.006
 MCS
0.391
<  0.001
0.347
<  0.001
0.377
<  0.001
0.272
0.004
 PA
−0.482
<  0.001
− 0.428
<  0.001
−0.655
<  0.001
− 0.286
<  0.001
 PI
−0.452
<  0.001
− 0.427
<  0.001
− 0.475
<  0.001
− 0.348
<  0.001
 PM
−0.431
0.001
−0.420
0.001
−0.455
<  0.001
− 0.357
0.020
ArcCHECK
 MIs
−0.271
0.006
−0.346
<  0.001
−0.355
<  0.001
 MIa
−0.364
<  0.001
−0.455
<  0.001
− 0.456
<  0.001
−0.273
0.006
 MIc, IMRT
−0.375
<  0.001
−0.471
<  0.001
− 0.472
<  0.001
−0.281
0.004
 MCS
0.426
<  0.001
0.589
<  0.001
0.660
<  0.001
 PA
−0.505
<  0.001
− 0.667
<  0.001
−0.739
<  0.001
− 0.370
<  0.001
 PI
−0.395
<  0.001
− 0.581
<  0.001
−0.663
<  0.001
 PM
−0.410
<  0.001
−0.607
<  0.001
− 0.671
<  0.001
−0.198
0.046
Note: rs Spearman’s rho, MIs Modulation index considering multi-leaf collimator speed, MIa Modulation index considering multi-leaf collimator acceleration, MIc, IMRT Modulation index considering multi-leaf collimator acceleration and field aperture irregularity for intensity modulated radiation therapy, MCS Modulation complexity score, PA Plan-averaged beam area, PI Plan-averaged beam irregularity, PM Plan-averaged beam modulation

Correlations between differences in machine parameters and modulation indices

The correlations between MLC errors and various MIs are shown in Table 5. For TrueBeam STx, every MI showed statistically significant correlations with MLC errors, except for MCS and PM. For Trilogy, every MI showed statistically significant correlations with MLC errors. The PI value had rs values of 0.861 and 0.767 in both the TrueBeam STx and Trilogy systems, respectively, showing the second strongest correlation (all with p <  0.001).
Table 5
Correlations of the averaged multi-leaf collimator positional errors to the modulation indices (p ≤ 0.05)
 
TrueBeam STx
Trilogy
r s
p
r s
p
MIs
0.411
< 0.001
0.213
0.032
MIa
0.783
< 0.001
0.489
< 0.001
MIc, IMRT
0.887
< 0.001
0.504
< 0.001
MCS
−0.727
< 0.001
PA
0.841
< 0.001
0.909
< 0.001
PI
0.861
< 0.001
0.767
< 0.001
PM
0.751
< 0.001
Note: rs Spearman’s rho, MIs Modulation index considering multi-leaf collimator speed, MIa Modulation index considering multi-leaf collimator acceleration, MIc, IMRT Modulation index considering multi-leaf collimator acceleration and field aperture irregularity for intensity modulated radiation therapy, MCS Modulation complexity score, PA Plan-averaged beam area, PI Plan-averaged beam irregularity, PM Plan-averaged beam modulation

Correlations between differences in dose-volume parameters and modulation indices

The numbers of statistically significant rs values (p <  0.05) of the various MIs to the differences in dose-volumetric parameters between original and reconstructed IMRT plans are shown in Fig. 2. The PI showed statistically significant rs values most frequently in the TrueBeam STx system (26 cases), whereas the PM showed statistically significant rs values most frequently in the Trilogy system (37 cases).

Discussion

In this study, we calculated the various MIs that have been previously suggested to predict IMRT delivery accuracy and then evaluated the performances of the MIs. When reviewing the mean values of MIs comprehensively, they showed inconsistent trends in IMRT plan complexities for the type of linac and treatment sites. This is because that the factors to be considered when calculating the values of MIs were different from one another. The MCS, PA, PI, and PM indices mainly focus on the beam aperture irregularity or the degree of segmentation, whereas the MIs, MIa, and MIc, IMRT indices mainly focus on the degree of mechanical movements. Nevertheless, for both the TrueBeam STx and Trilogy systems, all MIs except for the MIs and PA indices, indicated that the HN IMRT had the first or second highest complexity. The MIs values seem to have a limitation to calculate the plan complexities properly for various treatment sites since only MLC movements were considered for the MIs. The PA values increased as the modulation degree of IMRT increased unlike the existing definition and this was caused by the various target volume sizes for each treatment site in this study. Since the PA regards the use of small-sized apertures as high modulation of a plan, it is not appropriate to evaluate modulation degree of plans with various treatment sites. For example, the target volume sizes of the HN IMRT plans, which were highly modulated, were the largest among all IMRT plans, which resulted in the highest value of PA (the high value of PA means low modulation of a plan by its definition). On the contrary, the target volume sizes of the prostate boost IMRT plans, which were modulated less, were small, which resulted in the lowest value of PA. Therefore, PA increased owing to the target volume size variations not the modulation degree in this study. If we analyzed IMRT plans of a single treatment site with similar target volume sizes, this contradictory tendency to the definition of the PA would not be observed [26, 29]. PA values seem to not provide an appropriate modulation index for predicting IMRT delivery accuracy, through our analysis of IMRT plans with various treatment sites.
For TrueBeam STx, every gamma passing rate with both the MapCHECK2 and ArcCHECK indicated that HN IMRT plans had the lowest gamma passing rates for all gamma criteria. For Trilogy, every gamma passing rate with both the MapCHECK2 and ArcCHECK indicated that prostate boost plans had the lowest gamma passing rates for all gamma criteria, showing the inconsistency with the tendencies in the values of MIs. It was indicated that high-dose gradients of the field edge with low modulation and small field sizes increased the measurement uncertainty [30, 31] and it seems to decrease the gamma passing rates. This may be a factor in lowering correlations between MIs and gamma passing rates. Even for identical IMRT plan deliveries under identical gamma parameter settings, we acquired different gamma passing rates between the MapCHECK2 and ArcCHECK dosimeters, similar the findings of a previous study [15, 31]. Because the MapCHECK2 dosimeter has a high angular dependency, the gamma passing rates become inaccurate when the weight of delivered MUs increase at gantry angles near 90° or 270°. Jin et al. demonstrated that the MapCHECK2 measured doses near a gantry angle of 90° were inaccurate, and the inaccuracy increased as the photon beam energy decreased (30.6, 8.9, and 2.2% at gantry angles of 90° ± 10° for photon beam energies of 6, 10, and 15 MV, respectively) [30, 31]. Similarly, differences in the gamma passing rates between two dosimeters were large for lung SABR of the TrueBeam STx system, which used fields close to 90° or 270°. On the other hand, small differences in gamma passing rates between two dosimeters were observed for spine SABR and liver SABR of the TrueBeam STx and liver IMRT plans of Trilogy, which used fields close to 0° and 180°. The lower gamma passing rates of the MapCHECK2 dosimeter, compared to those of the ArcCHECK dosimeter, were due to dosimeter characteristics.
The DynaLog file is a record of actual motor values with an update rate of 50 ms. The motor values are converted to MLC position values using a conversion table (mlctable.txt). The Trajectory file is a record of the direct MLC position values, with an update rate of 20 ms [30]. The information of the Trajectory file is more accurate than that of the DynaLog file, owing to the short record time and direct-record method [21, 3234]. With respect to HN IMRT plans, the Trilogy MLC errors were much larger than those of the TrueBeam STx system, but the gamma passing rates of the TrueBeam STx system were smaller and as low as 11.5% of those of the Trilogy system, which were contradictory results to each other. Therefore, a direct absolute comparison between the DynaLog and Trajectory files seems inappropriate, since the recording systems differ. At the very least, we could identify the degree of IMRT plan modulation for various treatment sites within a single linac system.
For TrueBeam STx with the MapCHECK2, the PI showed the highest rs value for gamma passing rates with 2%/2 mm, 2%/1 mm and 1%/2 mm (rs = − 0.591 with p <  0.001, rs = − 0.550 with p <  0.001 and rs = − 0.474 with p <  0.001, respectively). For the ArcCHECK measurements, the rs values of the PI were − 0.545 and − 0.677 for gamma passing rates with 2%/2 mm and 1%/2 mm (all with p <  0.001), showing the second highest correlations. For the Trilogy system with the MapCHECK2, PI showed the second highest rs values for every passing rate gamma criterion (all with p <  0.001). For ArcCHECK measurements, the moderately strong correlations were observed between PI and the gamma passing rates with 2%/2 mm and 2%/1 mm (rs = − 0.581 with p <  0.001 and rs = − 0.663 with p <  0.001). By reviewing the results of the correlation analysis, PI mostly showed the first or second strongest correlations with every gamma passing rate, regardless of gamma criteria with the MapCHECK2 dosimeter, in both TrueBeam STx and Trilogy results (all with p <  0.001). Although the PI did not show the best performance with the ArcCHECK measurements, it showed moderately strong correlations with the gamma passing rates from the ArcCHECK measurements.
In the present study, the first attempt was made to evaluate performance of the previously developed MIs for IMRT by analyzing correlations of MIs with various plan delivery accuracy measures. By utilizing a large number of IMRT plans, various treatment sites, multiple photon energies and linac types, and multiple dosimeters, the correlations in this study were not as high as those found in earlier studies [3, 14, 16, 31]. Among various modulation indices, the PI showed moderately strong correlations with every plan delivery accuracy measure. Therefore, the PI seems to be used to verify IMRT plan delivery accuracy at the planning level in the clinical setting. Furthermore, it has a potential to be integrated in the treatment planning system (TPS) to generate IMRT plans with high delivery accuracy. It is expected to reduce resources in the clinical settings.
Although this study was expanded to comprehensively examine the performance of MIs as predictors of IMRT delivery accuracy by utilizing various treatment sites, photon beam energies, and dosimeters, the results were limited to only Varian linacs. This is a limitation of the present study. To acquire more comprehensive results, further study will be conducted by utilizing various types of linac from various manufacturers in the future.

Conclusions

The PI value showed best performance, among the modulation indices that were evaluated in this study, as a predictor of IMRT plan delivery accuracy showing strong correlations with various measures of IMRT delivery accuracy. The PI value could support the verification of IMRT plan delivery accuracies before patient treatment and reduce resource consumption in the clinic, as it can be calculated at the planning level.

Acknowledgements

Not applicable.

Funding

This work was supported by a National Research Foundation of Korea (NRF) grant from the Korea government (MSIP). (No.2017M2A2A7A02020639, No.2017M2A2A7A02020640, No.2017M2A2A7A02020641, No.2017M2A2A7A02020643).

Availability of data and materials

Data sharing is not applicable to this article, as no datasets were generated or analyzed during the current study.
Approval for this study was obtained from the institutional review board (IRB No. H-1802-069-922).
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Young R, Snyder BIMRT. (Intensity modulated radiation therapy): progress in technology and reimbursement. Radiol Manage. 2001;23:20–6.PubMed Young R, Snyder BIMRT. (Intensity modulated radiation therapy): progress in technology and reimbursement. Radiol Manage. 2001;23:20–6.PubMed
2.
Zurück zum Zitat Webb S. Optimization by simulated annealing of three-dimensional, conformal treatment planning for radiation fields defined by a multileaf collimator: II. Inclusion of two-dimensional modulation of the x-ray intensity. Phys Med Biol. 1992;37:1689–704.CrossRef Webb S. Optimization by simulated annealing of three-dimensional, conformal treatment planning for radiation fields defined by a multileaf collimator: II. Inclusion of two-dimensional modulation of the x-ray intensity. Phys Med Biol. 1992;37:1689–704.CrossRef
3.
Zurück zum Zitat Stein J, Bortfeld T, Dorschel B, Schlegel W. Dynamic X-ray compensation for conformal radiotherapy by means of multi-leaf collimation. Radiother Oncol. 1994;32:163–73.CrossRef Stein J, Bortfeld T, Dorschel B, Schlegel W. Dynamic X-ray compensation for conformal radiotherapy by means of multi-leaf collimation. Radiother Oncol. 1994;32:163–73.CrossRef
4.
Zurück zum Zitat Du W, Cho SH, Zhang X, Hoffman KE, Kudchadker RJ. Quantification of beam complexity in intensity-modulated radiation therapy treatment plans. Med Phys. 2014;41:021716.CrossRef Du W, Cho SH, Zhang X, Hoffman KE, Kudchadker RJ. Quantification of beam complexity in intensity-modulated radiation therapy treatment plans. Med Phys. 2014;41:021716.CrossRef
5.
Zurück zum Zitat Wang X, Spirou S, LoSasso T, Stein J, Chui CS, Mohan B. Dosimetric verification of intensity-modulated fields. Med Phys. 1996;23:317–27.CrossRef Wang X, Spirou S, LoSasso T, Stein J, Chui CS, Mohan B. Dosimetric verification of intensity-modulated fields. Med Phys. 1996;23:317–27.CrossRef
6.
Zurück zum Zitat Kim SY. Reirradiation of head and neck cancer in the era of intensity-modulated radiotherapy: patient selection, practical aspects, and current evidence. Radiat Oncol J. 2017;35:1–15.CrossRef Kim SY. Reirradiation of head and neck cancer in the era of intensity-modulated radiotherapy: patient selection, practical aspects, and current evidence. Radiat Oncol J. 2017;35:1–15.CrossRef
7.
Zurück zum Zitat Mani KR, Upadhayay S, Maria Das KJ. Influence of jaw tracking in intensity-modulated and volumetric-modulated arc radiotherapy for head and neck cancers: a dosimetric study. Radiat Oncol J. 2017;35:90–100.CrossRef Mani KR, Upadhayay S, Maria Das KJ. Influence of jaw tracking in intensity-modulated and volumetric-modulated arc radiotherapy for head and neck cancers: a dosimetric study. Radiat Oncol J. 2017;35:90–100.CrossRef
8.
Zurück zum Zitat Sakanaka K, Itasaka S, Ishida Y, Fujii K, Horimatsu T, Mizowaki T, et al. Dosimetric advantages and clinical outcomes of simultaneous integrated boost intensity-modulated radiotherapy for anal squamous cell carcinoma. Radiat Oncol J. 2017;35:368–79.CrossRef Sakanaka K, Itasaka S, Ishida Y, Fujii K, Horimatsu T, Mizowaki T, et al. Dosimetric advantages and clinical outcomes of simultaneous integrated boost intensity-modulated radiotherapy for anal squamous cell carcinoma. Radiat Oncol J. 2017;35:368–79.CrossRef
9.
Zurück zum Zitat Webb S. Optimization by simulated annealing of three-dimensional conformal treatment planning for radiation fields defined by a multileaf collimator. Phys Med Biol. 1991;36:1201–26.CrossRef Webb S. Optimization by simulated annealing of three-dimensional conformal treatment planning for radiation fields defined by a multileaf collimator. Phys Med Biol. 1991;36:1201–26.CrossRef
10.
Zurück zum Zitat Webb S. Use of a quantitative index of beam modulation to characterize dose conformality: illustration by a comparison of full beamlet IMRT, few-segment IMRT (fsIMRT) and conformal unmodulated radiotherapy. Phys Med Biol. 2003;48:2051–62.CrossRef Webb S. Use of a quantitative index of beam modulation to characterize dose conformality: illustration by a comparison of full beamlet IMRT, few-segment IMRT (fsIMRT) and conformal unmodulated radiotherapy. Phys Med Biol. 2003;48:2051–62.CrossRef
11.
Zurück zum Zitat Webb S, Poludniowski G. Intensity-modulated radiation therapy (IMRT) by a dynamic-jaws-only (DJO) technique in rotate-translate mode. Phys Med Biol. 2010;55:N495–506.CrossRef Webb S, Poludniowski G. Intensity-modulated radiation therapy (IMRT) by a dynamic-jaws-only (DJO) technique in rotate-translate mode. Phys Med Biol. 2010;55:N495–506.CrossRef
12.
Zurück zum Zitat Van Dyk J, Barnett RB, Cygler JE, Shragge PC. Commissioning and quality assurance of treatment planning computers. Int J Radiat Oncol Biol Phys. 1993;26:261–73.CrossRef Van Dyk J, Barnett RB, Cygler JE, Shragge PC. Commissioning and quality assurance of treatment planning computers. Int J Radiat Oncol Biol Phys. 1993;26:261–73.CrossRef
13.
Zurück zum Zitat Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656–61.CrossRef Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656–61.CrossRef
14.
Zurück zum Zitat Park SY, Park JM, Choi CH, Chun M, Han JH, Cho JD, et al. Optimal density assignment to 2D diode array detector for different dose calculation algorithms in patient specific VMAT QA. J Radiat Prot Res. 2017;42:177–88.CrossRef Park SY, Park JM, Choi CH, Chun M, Han JH, Cho JD, et al. Optimal density assignment to 2D diode array detector for different dose calculation algorithms in patient specific VMAT QA. J Radiat Prot Res. 2017;42:177–88.CrossRef
15.
Zurück zum Zitat Feygelman V, Zhang G, Stevens C, Nelms BE. Evaluation of a new VMAT QA device, or the “X” and “O” array geometries. J Appl Clin Med Phys. 2011;12:3346.CrossRef Feygelman V, Zhang G, Stevens C, Nelms BE. Evaluation of a new VMAT QA device, or the “X” and “O” array geometries. J Appl Clin Med Phys. 2011;12:3346.CrossRef
16.
Zurück zum Zitat Kim JI, Park SY, Kim HJ, Kim JH, Ye SJ, Park JM. The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT. Radiat Oncol. 2014;9:167.CrossRef Kim JI, Park SY, Kim HJ, Kim JH, Ye SJ, Park JM. The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT. Radiat Oncol. 2014;9:167.CrossRef
17.
Zurück zum Zitat Heilemann G, Poppe Bd, Laub W. On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance. Med Phys. 2013;40:031702.CrossRef Heilemann G, Poppe Bd, Laub W. On the sensitivity of common gamma-index evaluation methods to MLC misalignments in Rapidarc quality assurance. Med Phys. 2013;40:031702.CrossRef
18.
Zurück zum Zitat Bailey DW, Spaans JD, Kumaraswamy LK, Podgorsak MB. The MapCHECK measurement uncertainty function and its effect on planar dose pass rates. J Appl Clin Med Phys. 2016;17:165–73.CrossRef Bailey DW, Spaans JD, Kumaraswamy LK, Podgorsak MB. The MapCHECK measurement uncertainty function and its effect on planar dose pass rates. J Appl Clin Med Phys. 2016;17:165–73.CrossRef
19.
Zurück zum Zitat Vazquez-Quino LA, Huerta-Hernandez CI, Rangaraj D. Clinical experience with machine log file software for volumetric-modulated arc therapy techniques. Proc (Bayl Univ Med Cent). 2017;30:276–9.CrossRef Vazquez-Quino LA, Huerta-Hernandez CI, Rangaraj D. Clinical experience with machine log file software for volumetric-modulated arc therapy techniques. Proc (Bayl Univ Med Cent). 2017;30:276–9.CrossRef
20.
Zurück zum Zitat Tyagi N, Yang K, Yan D. Comparing measurement-derived (3DVH) and machine log file-derived dose reconstruction methods for VMAT QA in patient geometries. J Appl Clin Med Phys. 2014;15:4645.CrossRef Tyagi N, Yang K, Yan D. Comparing measurement-derived (3DVH) and machine log file-derived dose reconstruction methods for VMAT QA in patient geometries. J Appl Clin Med Phys. 2014;15:4645.CrossRef
21.
Zurück zum Zitat Agnew CE, Irvine DM, McGarry CK. Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT. J Appl Clin Med Phys. 2014;15:4994.CrossRef Agnew CE, Irvine DM, McGarry CK. Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT. J Appl Clin Med Phys. 2014;15:4994.CrossRef
22.
Zurück zum Zitat Chung K. A pilot study of the scanning beam quality assurance using machine log files in proton beam therapy. Prog Med Phys. 2017;28:129–33.CrossRef Chung K. A pilot study of the scanning beam quality assurance using machine log files in proton beam therapy. Prog Med Phys. 2017;28:129–33.CrossRef
23.
Zurück zum Zitat Park SY, Kim IH, Ye SJ, Carlson J, Park JM. Texture analysis on the fluence map to evaluate the degree of modulation for volumetric modulated arc therapy. Med Phys. 2014;41:111718.CrossRef Park SY, Kim IH, Ye SJ, Carlson J, Park JM. Texture analysis on the fluence map to evaluate the degree of modulation for volumetric modulated arc therapy. Med Phys. 2014;41:111718.CrossRef
24.
Zurück zum Zitat Park SY, Park JM, Sung W, Kim IH, Ye SJ. Texture analysis on the edge-enhanced fluence of VMAT. Radiat Oncol. 2015;10:74.CrossRef Park SY, Park JM, Sung W, Kim IH, Ye SJ. Texture analysis on the edge-enhanced fluence of VMAT. Radiat Oncol. 2015;10:74.CrossRef
25.
Zurück zum Zitat Park JM, Park SY, Kim H, Kim JH, Carlson J, Ye SJ. Modulation indices for volumetric modulated arc therapy. Phys Med Biol. 2014;59:7315–40.CrossRef Park JM, Park SY, Kim H, Kim JH, Carlson J, Ye SJ. Modulation indices for volumetric modulated arc therapy. Phys Med Biol. 2014;59:7315–40.CrossRef
26.
Zurück zum Zitat Park JM, Park SY, Kim H. Modulation index for VMAT considering both mechanical and dose calculation uncertainties. Phys Med Biol. 2015;60:7101–25.CrossRef Park JM, Park SY, Kim H. Modulation index for VMAT considering both mechanical and dose calculation uncertainties. Phys Med Biol. 2015;60:7101–25.CrossRef
27.
Zurück zum Zitat Chowdhuty AK, Debsarkar A, Chakrabarty S. Novel methods for assessing urban air quality: combined air and noise pollution approach. J Atmos Poll. 2015;3(1):1–8. Chowdhuty AK, Debsarkar A, Chakrabarty S. Novel methods for assessing urban air quality: combined air and noise pollution approach. J Atmos Poll. 2015;3(1):1–8.
28.
Zurück zum Zitat Hinkle DE, Wiersma W, Jurs SG. Applied statistics for the behavioral sciences. 5th ed. Wadsworth: Houghton Mifflin Harcourt; 2003. Hinkle DE, Wiersma W, Jurs SG. Applied statistics for the behavioral sciences. 5th ed. Wadsworth: Houghton Mifflin Harcourt; 2003.
29.
Zurück zum Zitat McNiven AL, Sharpe MB, Purdie TG. A new metric for assessing IMRT modulation complexity and plan deliverability. Med Phys. 2010;37:505–15.CrossRef McNiven AL, Sharpe MB, Purdie TG. A new metric for assessing IMRT modulation complexity and plan deliverability. Med Phys. 2010;37:505–15.CrossRef
30.
Zurück zum Zitat Jin H, Keeling VP, Johnson DA, Ahmad S. Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance. J Appl Clin Med Phys. 2014;15:4638.PubMed Jin H, Keeling VP, Johnson DA, Ahmad S. Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance. J Appl Clin Med Phys. 2014;15:4638.PubMed
31.
Zurück zum Zitat Keeling VP, Ahmad S. Jin H. a comprehensive comparison study of three different planar IMRT QA techniques using MapCHECK 2. J Appl Clin Med Phys. 2013;14:4398.CrossRef Keeling VP, Ahmad S. Jin H. a comprehensive comparison study of three different planar IMRT QA techniques using MapCHECK 2. J Appl Clin Med Phys. 2013;14:4398.CrossRef
32.
Zurück zum Zitat Sun B, Rangaraj D, Palaniswaamy G, Yaddanapudi S, Wooten O, Yang D, et al. Initial experience with TrueBeam trajectory log files for radiation therapy delivery verification. Pract Radiat Oncol. 2013;3:e199–208.CrossRef Sun B, Rangaraj D, Palaniswaamy G, Yaddanapudi S, Wooten O, Yang D, et al. Initial experience with TrueBeam trajectory log files for radiation therapy delivery verification. Pract Radiat Oncol. 2013;3:e199–208.CrossRef
33.
Zurück zum Zitat Teke T, Bergman AM, Kwa W, Gill B, Duzenli C, Popescu IA. Monte Carlo based, patient-specific RapidArc QA using Linac log files. Med Phys. 2010;37:116–23.CrossRef Teke T, Bergman AM, Kwa W, Gill B, Duzenli C, Popescu IA. Monte Carlo based, patient-specific RapidArc QA using Linac log files. Med Phys. 2010;37:116–23.CrossRef
34.
Zurück zum Zitat Ramsey CR, Spencer KM, Alhakeem R, Oliver AL. Leaf position error during conformal dynamic arc and intensity modulated arc treatments. Med Phys. 2001;28:67–72.CrossRef Ramsey CR, Spencer KM, Alhakeem R, Oliver AL. Leaf position error during conformal dynamic arc and intensity modulated arc treatments. Med Phys. 2001;28:67–72.CrossRef
Metadaten
Titel
Assessment of the modulation degrees of intensity-modulated radiation therapy plans
verfasst von
So-Yeon Park
Jung-in Kim
Minsoo Chun
Hyunjun Ahn
Jong Min Park
Publikationsdatum
01.12.2018
Verlag
BioMed Central
Erschienen in
Radiation Oncology / Ausgabe 1/2018
Elektronische ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-018-1193-9

Weitere Artikel der Ausgabe 1/2018

Radiation Oncology 1/2018 Zur Ausgabe

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.