Open Access 01.12.2014 | Research article

Association between low-frequency ultrasound and hip fractures - comparison with DXA-based BMD

verfasst von: Mikko Määttä, Petro Moilanen, Jussi Timonen, Pasi Pulkkinen, Raija Korpelainen, Timo Jämsä

Erschienen in: BMC Musculoskeletal Disorders | Ausgabe 1/2014

Abstract

Background

New methods for diagnosing osteoporosis and evaluating fracture risk are being developed. We aim to study the association between low-frequency (LF) axial transmission ultrasound and hip fracture risk in a population-based sample of older women.

Methods

The study population consisted of 490 community-dwelling women (78–82 years). Ultrasound velocity (VLF) at mid-tibia was measured in 2006 using a low-frequency scanning axial transmission device. Bone mineral density (BMD) at proximal femur measured using dual-energy x-ray absorptiometry (DXA) was used as the reference method. The fracture history of the participants was collected from December 1997 until the end of 2010. Lifestyle-related risk factors and mobility were assessed at 1997.

Results

During the total follow-up period (1997–2010), 130 women had one or more fractures, and 20 of them had a hip fracture. Low VLF (the lowest quartile) was associated with increased hip fracture risk when compared with VLF in the normal range (Odds ratio, OR = 3.3, 95% confidence interval (CI) 1.3-8.4). However, VLF was not related to fracture risk when all bone sites were considered. Osteoporotic femoral neck BMD was associated with higher risk of a hip fracture (OR = 4.1, 95% CI 1.6-10.5) and higher risk of any fracture (OR = 2.4, 95% CI 1.6-3.8) compared to the non-osteoporotic femoral neck BMD. Decreased VLF remained a significant risk factor for hip fracture when combined with lifestyle-related risk factors (OR = 3.3, 95% CI 1.2-9.0).

Conclusion

Low VLF was associated with hip fracture risk in older women even when combined with lifestyle-related risk factors. Further development of the method is needed to improve the measurement precision and to confirm the results.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1471-2474-15-208) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

MM participated in the fracture data collection, carried out the main analyses, and drafted the manuscript. PM participated in the analyses and interpretation of ultrasound data and manuscript writing. JT and PP made a contribution to the design of the study and revising the manuscript. RK and TJ participated in study conception and design, data interpretation and made substantial contributions to the manuscript. RK also participated significantly to collection of lifestyle-related risk factors data. All authors have read, revised, and given their final approval of the version to be published.

Background

Osteoporotic fractures possess a significant public health problem that is increasing due to aging population. At the moment, the golden standard used in fracture risk assessment is bone mineral density (BMD) measurements using dual energy x-ray absorptiometry (DXA). Recently, World Health Organization (WHO) introduced FRAX®, a fracture risk calculator that combines easily obtained clinical information and DXA-based femoral neck BMD, if available, to estimate the 10-year osteoporotic fracture probability [1].
Quantitative ultrasound (QUS) has raised interest as an alternative method to x-ray-based imaging for measuring bone status. There is a number of different QUS methods that have been used to assess bone status and to evaluate fracture risk [2]. The advantages of ultrasound include a relatively low cost and a portable technique with no ionizing radiation. The results of QUS measurements have been shown to be associated with fracture risk [3]. A question has arisen whether QUS parameters can serve as a surrogate for DXA-based BMD to improve FRAX® estimate [4].
A novel low-frequency (LF) axial transmission QUS method for assessing the properties of cortical bone has been shown to reflect bone density and cortical thickness [57]. A relationship between the LF tibial velocity (VLF) and geometry of proximal femur in older women has also been suggested [8]. However, the suitability of the method for individual fracture risk assessment has not yet been established.
In this study we evaluated the association between the LF axial transmission ultrasound method and fractures in a population-based sample of older women, using the standard DXA as the reference method. We hypothesized that decreased ultrasound velocity measured on tibia is associated with increased hip fracture risk in elderly females. In our previous population based prospective study we assessed life-style related determinants of hip fracture in elderly females [9]. Here we analyzed how the combination of low-frequency ultrasound and previously determined lifestyle-related factors are related to the risk of hip fracture.

Methods

Subjects and clinical assessment

The study population consisted of 490 women born between 1924 and 1927, originally recruited in 1997 as a population-based cohort to study the risk factors for osteoporosis and fractures [10]. All those still alive belonging to the original cohort were invited to clinical measurements in 2006. A total of 618 women attended these measurements, and 490 were measured using quantitative ultrasound. The fracture history between December 1, 1997 and December 31, 2010 was collected from hospital discharge registers. The fractures were confirmed manually from medical records to avoid the bias of recording multiple hospitalizations due to a single fracture. Health and lifestyle information, including medical history, age at menopause, smoking habits, alcohol and coffee consumption, physical activity and mobility, calcium and vitamin D intakes, and fracture history was collected at study baseline in 1997 using self-administrative questionnaires and interviews. The assessment of functional mobility at baseline was done using the “Timed Up & Go” (TUG) test [11]. All subjects gave a written informed consent and the study protocol was approved by the Ethics Committee of the Northern Ostrobothnia Hospital District. The study was done in accordance with the Declaration of Helsinki.

Quantitative ultrasound measurements

The speed of sound was recorded in the medial mid-shaft of the left tibia using a scanning low-frequency (LF) axial transmission device. The principle of the device and the measurement setup has been published previously [7, 8]. In brief, two separate transducers (f c  = 200 kHz) mounted on a rail scanned a 30 mm distance. The time of flight of the first arriving signal (FAS) was determined using the first maximum. The apparent velocity of the FAS (VLF) was determined by measuring the time of flight for a number of source-receiver distances. The precision error was characterized by CVrms which was 3.2%. The standardized coefficient of variation (SCV) [12] of the method was 6.6%. Due to the lack of established reference population data, we used the present study population to define the range of SCV.

DXA measurements

A Hologic DXA device (Delphi QDR series, Hologic, Bedford, MA, USA) was used to get the reference data. Standard anteroposterior positioning was used to measure the femoral neck bone mineral density (BMD) of the left proximal femur.

Statistical analysis

The subjects of the study were divided into three groups based on their fracture history (Figure 1): a) women without fractures (NF, controls), b) women with any fracture (Fx), and c) women with a hip fracture (Hip Fx). All the hip fracture patients were also included in group (b). The fracture data of three different time periods were analyzed: a) from the beginning of the follow-up period until the time of bone measurements (1997–2006), b) from the time of the bone measurements until the end of the follow-up period (2006–2010), and c) the whole follow-up period (1997–2010). The women with fractures were compared to those without fractures. Since the data were normally distributed the independent samples t-test was used to analyze the statistical significance of the differences between the NF, Fx, and Hip Fx groups. The study subjects were also classified in to normal, osteopenic, and osteoporotic according to their femoral neck BMD T-score in accordance with the WHO definition [13]. Due to the lack of an established reference population, T-scores could not be calculated for low-frequency ultrasound velocity (VLF). Thus, VLF results were divided into three groups based on quartiles within the study population: a) low VLF (0-25%), b) moderate VLF (25-50% and 50-75% combined), and c) high VLF (75-100%). Crosstabulation and the χ2 test were used to compare the distribution of the subjects with fractures and those without fractures within these groups. To further analyze the statistical significance of the association of bone measurement results (QUS/DXA) with the fracture risk, a multivariate logistic regression analysis was used. The results of those analyses are reported as the odds ratios (OR) and 95% confidence intervals (CI). The imaging modality (QUS/DXA) used was included in the model and the forward stepwise (likelihood ratio) method was used to form the final models. All models were adjusted with age and BMI. For the fractures that occurred after the measurements in 2006, the predictive ability of each bone measurement value was assessed using the Cox proportional hazards model and the corresponding hazard ratios (HR), and the 95% CIs were calculated. A similar protocol to logistic regression was used. In addition, previous fractures (1997–2006) (yes, no) was added to models as covariates. The follow-up time between the measurement time and the time of the first fracture, death, or end of the follow-up period was recorded. Logistic regression models were also calculated to analyze the effect of combining VLF and previously determined [9] lifestyle-related risk factors for hip fracture. In our previous population based cohort study BMI, functional mobility, physical activity, hypertension, coffee consumption, and daily smoking predicted hip fractures in 1222 old women. These factors were used as a covariates to determine the best fit model for the current subpopulation with ultrasound and DXA measurements.
Statistical analyses were done using SPSS for Windows (Release 18.0, SPSS Inc., Chicago, IL, USA). In all tests, p-values less than 0.05 were considered statistically significant.

Results

During the first follow-up period (1997–2006), 98 fractures occurred in 81 women (Figure 1, Table 1). Between 2006 and 2010, 61 subjects sustained fractures; in total 76 different fractures occurred. Altogether, between the years 1997 and 2010, 174 different fractures occurred in 130 subjects. Wrist fracture was the most common fracture (n = 51). There were 9 hip fractures between 1997 and 2006, and 11 hip fractures after the 2006 measurements, totaling in 20 hip fractures during the whole follow-up period. Thirty-five subjects died during the follow-up period.
Table 1
Distribution of fractures (n (%)) according to the ICD-10 code with selected subdivisions in a population-based sample of older women (n = 490)
  
1997–2006
2006–2010
1997–2010
M80
Osteoporosis with pathological fracture
 
2 (2.6)
2 (1.1)
S12
Fracture of neck
 
1 (1.3)
1 (0.6)
S22
Fracture of rib(s), sternum and thoracic spine
3 (3.1)
3 (3.9)
6 (3.4)
S32
Fracture of lumbar spine and pelvis
   
S32.0
Fracture of lumbar vertebra
 
9 (11.8)
9 (5.2)
 
Other
3 (3.1)
4 (5.3)
7 (4.0)
S42
Fracture of shoulder and upper arm
11 (11.2)
6 (7.9)
17 (9.8)
S52
Fracture of forearm
   
S52.5
Fracture of lower end of radius
29 (29.6)
22 (28.9)
51 (29.3)
 
Other
5 (5.1)
1 (1.3)
6 (3.4)
S62
Fracture at wrist and hand level
7 (7.1)
5 (6.6)
12 (6.9)
S72
Fracture of femur
   
S72.0
Fracture of neck of femur
4 (4.1)
7 (9.2)
11 (6.3)
S72.1
Pertrochanteric fracture
1 (1.0)
4 (5.3)
5 (2.9)
S72.2
Subtrochanteric fracture
4 (4.1)
 
4 (2.3)
 
Other
3 (3.1)
6 (7.9)
9 (5.2)
S82
Fracture of lower leg, including ankle
26 (26.5)
5 (6.6)
31 (17.8)
S92
Fracture of foot, except ankle
2 (2.0)
1 (1.3)
3 (1.7)
Total number of fractures
98 (100)
76 (100)
174 (100)
ICD-10 = International Classification of Diseases, 10th revision. Time periods represent previous fractures (1997–2006) and future fractures (2006–2010) with respect to the bone measurements, and all fractures during the follow-up period (1997–2010).
The characteristics of different fracture groups are shown in Table 2. The women with a hip fracture were taller than the women without fractures (p < 0.05). Those with a previous fracture (1997–2006) had a lower VLF than the ones without fractures (p < 0.05). Subjects who sustained a fracture during the follow-up period had a lower femoral neck BMD than those in the NF group (p < 0.05), except when comparing the Hip Fx 1997–2006 and NF groups.
Table 2
Characteristics of the women with different fracture status (NF, Fx, and Hip Fx) in different time periods (1997–2006, 2006–2010, and 1997–2010) at the time of the measurements (2006) (n = 490)
 
NF
Fx 1997-2010
Hip Fx 1997-2010
n =360
n = 130
n = 20
  
Fx 1997-2006
Fx 2006-2010
Hip Fx 1997-2006
Hip Fx 2006-2010
   
n = 81
n = 61
n = 9
n = 11
Age [years]
79.9 (1.2)
80.0 (1.2)
80.0 (1.3)
79.7 (1.2)
80.3 (1.1)*
79.3 (1.4)
80.5 (0.9)
Weight [kg]
67.7 (11.3)
67.8 (12.2)
69.8 (15.7)
69.5 (12.3)
65.8 (12.2)
75.7 (10.2)*
65.0 (18.2)
Height [cm]
155.2 (5.4)
155.6 (5.8)
158.9 (7.2)*
156.6 (5.4)*
154.5 (6.4)
160.7 (4.4)*
157.4 (8.8)
BMI [kg/m2]
28.1 (4.5)
28.0 (4.7)
27.4 (4.6)
28.4 (4.8)
27.5 (4.6)
29.3 (3.6)
25.9 (4.9)
VLF [m/s]
3583 (193)
3547 (200)
3515 (245)
3534 (204)*
3564 (203)
3500 (291)
3527 (215)
Femoral neck BMD [g/cm2]
0.654 (0.102)
0.606 (0.083)*
0.589 (0.102)*
  
0.609 (0.078)*
0.605 (0.096)*
0.611 (0.089)
0.570 (0.112)*
Values are Mean (SD).
*Independent samples t-test p-value < 0.05 when compared to NF group.
Differences between distributions in the NF and Hip Fx groups were observed in both VLF and femoral neck BMD (p < 0.05) (Table 3). Also, there was a difference in the distribution of subjects based on femoral neck BMD between the Fx and NF groups.
Table 3
The distribution of women with and without a fracture and hip fracture according to the V LF and BMD values
 
NF
Fx 1997–2010
Hip Fx 1997–2010
 
(n = 360)
(n = 130)
(n = 20)
VLF
Highest 25%
101 (28)
24 (18)
5 (25)*
25% to 75%
170 (47)
69 (53)
5 (25)
Lowest 25%
89 (25)
37 (28)
10 (50)
Femoral neck BMD
Normal
73 (20)
11 (8)*
2 (10)*
Osteopenic
206 (57)
65 (50)
8 (40)
 
Osteoporotic
81 (23)
54 (42)
10 (50)
Values are n (%). *Distribution of the group is different than that of the NF group (p-value of χ2 test < 0.05).
Based on the regression analysis, low VLF was not associated with previous fractures (1997–2006) or with the fractures that occurred after the measurements (2006–2010) (Table 4). Decreased femoral neck BMD, however, was associated with increased fracture risk in all follow-up periods; estimated risks (odds and hazard ratios) varied between 1.8 and 2.5 compared to the women with femoral neck T-score > -2.5.
Table 4
Association of low V LF and osteoporotic femoral neck BMD with fractures that occurred during different follow-up periods
 
Fx 1997-2006
Fx 2006-2010
Fx 1997-2010
NF n = 409, Fx n = 81
NF n = 429, Fx n = 61
NF n = 334, Fx n = 121*
 
n
OR (95% CI)
p
HR (95% CI)
p
OR (95% CI)
p
VLF
Moderate or high (25-100%)
364
      
Low (0-25%)
126
1.6 (0.9 - 2.6)
0.088
0.8 (0.4 - 1.4)
0.359
1.3 (0.8 - 2.0)
0.326
Femoral neck BMD
T-score > -2.5
355
      
 
T-score ≤ -2.5
135
1.8 (1.1 - 3.0)
0.019
2.4 (1.4 - 3.9)
0.001
2.5 (1.6 - 3.9)
<0.001
All regression models were adjusted by age and BMI. In Cox regression previous fracture (during 1997–2006) was also used as a covariate. NF = non-fractured group, Fx = fractured group, n = number of subjects, OR = odds ratio of logistic regression, HR = hazard ratio of Cox regression, CI = confidence interval, p = p-value of the covariate in the model. *Subjects who died during the follow-up were excluded (NF n = 26, Fx n = 9).
Decreased VLF was associated with an increased risk of hip fracture occurred before the measurements (1997–2006) (OR = 6.3; Table 5). Additionally, across the whole follow-up period (1997–2010), low VLF was associated with higher risk of hip fracture compared to moderate or high VLF (OR = 3.3). An osteoporotic femoral neck BMD predicted hip fractures after the measurements (2006–2010) with a hazard ratio (HR) of 4.8 (95% CI 1.4-16.6) for subjects with a T-score ≤ -2.5 compared to the women with normal or osteopenic T-score. Low femoral neck BMD was also associated with an increased risk of hip fracture (OR = 4.1, 95% CI 1.6-10.5) during the whole follow-up period (1997–2010) compared to women with normal or osteopenic femoral neck BMD. When VLF was combined with previously assessed [9] hip fracture risk factors regression model for the same population (Table 6), the final model included VLF, low physical activity, and impaired functional mobility (p = 0.001, Table 6), VLF being the strongest independent factor for hip fracture (OR = 3.3, 95% CI 1.2–9.0). Femoral neck BMD did not reach statistical significance when included in the regression analyses along with lifestyle-related risk factors.
Table 5
Association of low V LF and osteoporotic femoral neck BMD with hip fractures that occurred during different follow-up periods
 
Hip Fx 1997-2006
Hip Fx 2006-2010
Hip Fx 1997-2010
NF n = 409, Hip Fx n = 9
NF n = 429, Hip Fx n = 11
NF n = 334, Hip Fx n = 19*
 
n
OR (95% CI)
p
n
HR (95% CI)
p
n
OR (95% CI)
p
VLF
Moderate or high (25-100%)
313
  
324
  
260
  
Low (0-25%)
105
6.3 (1.5 - 25.5)
0.010
116
1.7 (0.5 - 5.7)
0.415
93
3.3 (1.3 - 8.4)
0.012
Femoral neck BMD
T-score > -2.5
311
  
325
  
267
  
 
T-score ≤ -2.5
107
1.5 (0.4 - 6.0)
0.593
115
4.8 (1.4 - 16.6)
0.013
86
4.1 (1.6 – 10.5)
0.003
All regression models were adjusted by age and BMI. In Cox regression previous fracture (during 1997–2006) was also used as a covariate. NF = non-fractured group, Hip Fx = hip fracture group, n = number of subjects, OR = odds ratio of logistic regression, HR = hazard ratio of Cox regression, CI = confidence interval, p = p-value of the covariate in the model. *Subjects who died during the follow-up were excluded (NF n = 26, Hip Fx n = 1).
Table 6
Logistic regression models without and with V LF for having a hip fracture in a population-based sample of older women
 
OR
(95% CI)
p-value
Lifestyle-related risk factors1
  
0.002*
TUG ≥ 11 s vs. less (referent)
3.4
(1.2 - 9.9)
0.026
Low PA vs. moderate to high (referent)
2.8
(1.0 - 7.5)
0.046
Coffee consumption > 3 cups/day vs. less (referent)
0.3
(0.1 - 1.0)
0.051
Lifestyle-related risk factors and VLF 2
  
0.001*
Low VLF (0-25%) vs. Moderate or high (25-100%) (referent)
3.3
(1.2 - 9.0)
0.018
Low PA vs. moderate to high (referent)
3.1
(1.1 - 8.5)
0.028
TUG ≥ 11 s vs. less (referent)
3.1
(1.0 - 8.9)
0.042
Odds ratios (OR) are calculated compared to the NF group. TUG “Timed Up & Go” test, PA physical activity, CI confidence interval. The number of subjects in the analyses was Hip Fx n = 18, NF n = 296. 1Age, BMI, TUG, PA, hypertension, coffee consumption, and smoking were included in the analysis to form the best model, 2The best model after including VLF in the analysis, *p-value for the full model.

Discussion

In this population-based study, we used a low-frequency ultrasound scanner to assess the association between decreased ultrasound velocity and fracture risk. To our knowledge, this is the first time that low-frequency axial transmission ultrasound is being used in a population-based cohort of older women. The reported results show that decreased ultrasound velocity (VLF) was associated with higher risk of hip fracture compared to moderate or high VLF. DXA and LF ultrasound yielded similar results when comparing the hip fracture and non-fracture groups.
Our findings are in line with earlier studies on the fracture-discrimination ability of similar axial transmission ultrasound devices. Using the Omnisense (Sunlight, BeamMed Ltd, Petah Tikva, Israel) device, operating at a center frequency of 1.25 MHz, Nguyen et al. [14] reported that decreased tibial SOS is associated with increased fracture risk independently of BMD and age with an odds ratio (OR) of 1.75. The corresponding OR as determined for the femoral neck BMD here was 2.11. Damilakis et al. [15] found no difference using the Omnisense device in the tibial SOS of healthy subjects and subjects with osteoporotic fracture. Nevertheless, they reported increased osteoporotic fracture ORs for ultrasound measurements on the radius and phalanx (ORs between 1.7 and 2.7). In other studies with an Omnisense device, SOS measured on the radius has been shown to discriminate subjects with hip fracture from controls with no fractures (ORs varying between 1.9 and 2.7) [1618]. Another axial transmission QUS device, Myriad Soundscan (Myriad Ultrasound systems, Israel) uses a center frequency of 250 kHz which is closer to that used in this study. Using this device, Stegman et al. [19] reported a low-energy appendicular fracture OR of 1.4 for the tibial SOS. This observation is in good agreement with the present study, where we found a similar trend in the retrospective part of the study. Augat et al. [20] found an increased OR of 1.7 for the tibial SOS of subjects who sustained hip fracture compared to non-fracture controls. The corresponding OR for the femoral neck BMD was 3.5 in their study. This SOS OR is somewhat lower than those observed in the present study for hip fractures. Also, Talmant et al. [21] studied the fracture discrimination ability of a 1 MHz bidirectional ultrasonometer. They reported a fracture OR of 1.8 for the radius SOS and of 2.1 for the femoral neck BMD. Direct comparison between the ultrasound methods used is difficult due to differences in the ultrasound methodology and measurement setup. Also, the bone site and type of fracture, as well as the study populations, affect the results. However, the majority of the studies imply that axial transmission ultrasound is capable of assessing fracture risk.
Some studies have assessed the combining clinical risk factors and QUS parameters [2224]. In the current study low VLF remained a significant risk factor for hip fracture when included in the lifestyle-related risk factor model observed in our previous study with the same population [9]. However, the lifestyle-related risk factors were collected at baseline whereas QUS measurements were made at eight years from baseline. Thus, the results need to be confirmed in future prospective studies.
In the present study ultrasound velocity was measured at mid-tibia. Even though osteoporosis is a systemic disease, the degree of changes in bone structure varies between bone sites. In weight-bearing sites (e.g. the tibia) bone loss is not as big as in non-weight-bearing sites (e.g. the radius and phalanx) [25]. Thus, even if the tibia is an easily accessible site for axial ultrasound measurements, it may not be the most responsive site to osteoporotic changes. Recently, we have reported that LF axial-transmission ultrasound in the radius is able to retrospectively discriminate postmenopausal women with fractures from age-matched controls with no fractures using an improved version of the LF ultrasound device used here [26].
The strength of the present study was the use of a population-based cohort of older women. The other strength was the relatively long follow-up time before and after the bone measurements. However, this study also had some limitations. One major limitation was the limited measurement precision of the current prototype device. The bulky mechanical scanning setup limits the positioning accuracy and does not enable proper correction for the impacts of overlying soft tissue [8]. As a result, the in vivo precision (CVrms) was limited. During the follow-up period of the study, significant mechanical improvements were implemented in the device, and a CV of 0.5% could be achieved with the latest device version [5, 26]. Obviously, this valuable information available after the device improvement cannot retrospectively improve the quality of the results presented here. The other limitation is the retrospective nature of the study. It is possible that the fracture events before the measurements caused the changes in bone properties (e.g. via altered loading conditions) and affected on outcome of the measurement. Also, the number of hip fractures was relatively low, which significantly limited the statistical reliability of the results. Additionally, a minor limitation was the lack of reference data for young and healthy population, which disabled the determination of the T-score. Instead, we used quartiles to define the subjects with low VLF values.

Conclusions

In conclusion, decreased low-frequency ultrasound velocity was associated to increased hip fracture risk despite the limited measurement precision. The results reported here can be used to further improve the measurement precision of the method so as to reliably predict future fractures. To this end carefully planned follow-up studies are needed.

Acknowledgements

The work done is a part of the Osteoporotic Fracture Risk study supported by the Finnish Funding Agency for Technology and Innovation (Tekes, projects no. 40463/05, 40403/06). The work was also supported by the Academy of Finland. MM was supported by Tauno Tönning Foundation and Finnish Cultural Foundation.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

MM participated in the fracture data collection, carried out the main analyses, and drafted the manuscript. PM participated in the analyses and interpretation of ultrasound data and manuscript writing. JT and PP made a contribution to the design of the study and revising the manuscript. RK and TJ participated in study conception and design, data interpretation and made substantial contributions to the manuscript. RK also participated significantly to collection of lifestyle-related risk factors data. All authors have read, revised, and given their final approval of the version to be published.
Anhänge

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.
Literatur
1.
Zurück zum Zitat Kanis JA, Oden A, Johansson H, Borgström F, Ström O, McCloskey E: FRAX and its applications to clinical practice. Bone. 2009, 44 (5): 734-743. 10.1016/j.bone.2009.01.373.CrossRefPubMed Kanis JA, Oden A, Johansson H, Borgström F, Ström O, McCloskey E: FRAX and its applications to clinical practice. Bone. 2009, 44 (5): 734-743. 10.1016/j.bone.2009.01.373.CrossRefPubMed
2.
Zurück zum Zitat Glüer CC: A new quality of bone ultrasound research. IEEE Trans Ultrason Ferroelectr Freq Control. 2008, 55 (7): 1524-1528.CrossRefPubMed Glüer CC: A new quality of bone ultrasound research. IEEE Trans Ultrason Ferroelectr Freq Control. 2008, 55 (7): 1524-1528.CrossRefPubMed
3.
Zurück zum Zitat Marín F, González-Macías J, Díez-Pérez A, Palma S, Delgado-Rodríguez M: Relationship between bone quantitative ultrasound and fractures: a meta-analysis. J Bone Miner Res. 2006, 21 (7): 1126-1135. 10.1359/jbmr.060417.CrossRefPubMed Marín F, González-Macías J, Díez-Pérez A, Palma S, Delgado-Rodríguez M: Relationship between bone quantitative ultrasound and fractures: a meta-analysis. J Bone Miner Res. 2006, 21 (7): 1126-1135. 10.1359/jbmr.060417.CrossRefPubMed
4.
Zurück zum Zitat Lewiecki EM, Compston JE, Miller PD, Adachi JD, Adams JE, Leslie WD, Kanis JA, Moayyeri A, Adler RA, Hans DB, Kendler DL, Díez-Pérez A, Krieg MA, Masri BK, Lorenc RR, Bauer DC, Blake GM, Josse RG, Clark P, Khan AA, FRAX® Position Development Conference Members: Official Positions for FRAX® Bone Mineral Density and FRAX® simplification from Joint Official Positions Development Conference of the International Society for Clinical Densitometry and International Osteoporosis Foundation on FRAX®. J Clin Densitom. 2011, 14 (3): 226-236. 10.1016/j.jocd.2011.05.017.CrossRefPubMed Lewiecki EM, Compston JE, Miller PD, Adachi JD, Adams JE, Leslie WD, Kanis JA, Moayyeri A, Adler RA, Hans DB, Kendler DL, Díez-Pérez A, Krieg MA, Masri BK, Lorenc RR, Bauer DC, Blake GM, Josse RG, Clark P, Khan AA, FRAX® Position Development Conference Members: Official Positions for FRAX® Bone Mineral Density and FRAX® simplification from Joint Official Positions Development Conference of the International Society for Clinical Densitometry and International Osteoporosis Foundation on FRAX®. J Clin Densitom. 2011, 14 (3): 226-236. 10.1016/j.jocd.2011.05.017.CrossRefPubMed
5.
Zurück zum Zitat Kilappa V, Moilanen P, Xu L, Nicholson PH, Timonen J, Cheng S: Low-frequency axial ultrasound velocity correlates with bone mineral density and cortical thickness in the radius and tibia in pre- and postmenopausal women. Osteoporos Int. 2011, 22 (4): 1103-1113. 10.1007/s00198-010-1273-7.CrossRefPubMed Kilappa V, Moilanen P, Xu L, Nicholson PH, Timonen J, Cheng S: Low-frequency axial ultrasound velocity correlates with bone mineral density and cortical thickness in the radius and tibia in pre- and postmenopausal women. Osteoporos Int. 2011, 22 (4): 1103-1113. 10.1007/s00198-010-1273-7.CrossRefPubMed
6.
Zurück zum Zitat Moilanen P, Nicholson PH, Kärkkäinen T, Wang Q, Timonen J, Cheng S: Assessment of the tibia using ultrasonic guided waves in pubertal girls. Osteoporos Int. 2003, 14 (12): 1020-1027. 10.1007/s00198-003-1528-7.CrossRefPubMed Moilanen P, Nicholson PH, Kärkkäinen T, Wang Q, Timonen J, Cheng S: Assessment of the tibia using ultrasonic guided waves in pubertal girls. Osteoporos Int. 2003, 14 (12): 1020-1027. 10.1007/s00198-003-1528-7.CrossRefPubMed
7.
Zurück zum Zitat Nicholson PH, Moilanen P, Kärkkäinen T, Timonen J, Cheng S: Guided ultrasonic waves in long bones: modelling, experiment and in vivo application. Physiol Meas. 2002, 23 (4): 755-768. 10.1088/0967-3334/23/4/313.CrossRefPubMed Nicholson PH, Moilanen P, Kärkkäinen T, Timonen J, Cheng S: Guided ultrasonic waves in long bones: modelling, experiment and in vivo application. Physiol Meas. 2002, 23 (4): 755-768. 10.1088/0967-3334/23/4/313.CrossRefPubMed
8.
Zurück zum Zitat Määttä M, Moilanen P, Nicholson P, Cheng S, Timonen J, Jämsä T: Correlation of Tibial Low-Frequency Ultrasound Velocity with Femoral Radiographic Measurements and BMD in Elderly Women. Ultrasound Med Biol. 2009, 35 (6): 903-911. 10.1016/j.ultrasmedbio.2008.12.003.CrossRefPubMed Määttä M, Moilanen P, Nicholson P, Cheng S, Timonen J, Jämsä T: Correlation of Tibial Low-Frequency Ultrasound Velocity with Femoral Radiographic Measurements and BMD in Elderly Women. Ultrasound Med Biol. 2009, 35 (6): 903-911. 10.1016/j.ultrasmedbio.2008.12.003.CrossRefPubMed
9.
Zurück zum Zitat Määttä M, Terho E, Jokinen H, Pulkkinen P, Korpelainen J, Heikkinen J, Keinänen-Kiukaanniemi S, Jämsä T, Korpelainen R: Lifestyle factors and site-specific risk of hip fracture in community dwelling older women – a 13-year prospective population-based cohort study. BMC Musculoskelet Disord. 2012, 13 (1): 173-10.1186/1471-2474-13-173.CrossRefPubMedPubMedCentral Määttä M, Terho E, Jokinen H, Pulkkinen P, Korpelainen J, Heikkinen J, Keinänen-Kiukaanniemi S, Jämsä T, Korpelainen R: Lifestyle factors and site-specific risk of hip fracture in community dwelling older women – a 13-year prospective population-based cohort study. BMC Musculoskelet Disord. 2012, 13 (1): 173-10.1186/1471-2474-13-173.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Korpelainen R, Korpelainen J, Heikkinen J, Väänänen K, Keinänen-Kiukaanniemi S: Lifestyle factors are associated with osteoporosis in lean women but not in normal and overweight women: a population-based cohort study of 1222 women. Osteoporos Int. 2003, 14 (1): 34-43. 10.1007/s00198-002-1319-6.CrossRefPubMed Korpelainen R, Korpelainen J, Heikkinen J, Väänänen K, Keinänen-Kiukaanniemi S: Lifestyle factors are associated with osteoporosis in lean women but not in normal and overweight women: a population-based cohort study of 1222 women. Osteoporos Int. 2003, 14 (1): 34-43. 10.1007/s00198-002-1319-6.CrossRefPubMed
11.
Zurück zum Zitat Podsiadlo D, Richardson S: The timed “Up & Go”: a test of basic functional mobility for frail elderly persons. J Am Geriatr Soc. 1991, 39 (2): 142-148.CrossRefPubMed Podsiadlo D, Richardson S: The timed “Up & Go”: a test of basic functional mobility for frail elderly persons. J Am Geriatr Soc. 1991, 39 (2): 142-148.CrossRefPubMed
12.
Zurück zum Zitat Orgee JM, Foster H, McCloskey EV, Khan S, Coombes G, Kanis JA: A precise method for the assessment of tibial ultrasound velocity. Osteoporos Int. 1996, 6 (1): 1-7. 10.1007/BF01626530.CrossRefPubMed Orgee JM, Foster H, McCloskey EV, Khan S, Coombes G, Kanis JA: A precise method for the assessment of tibial ultrasound velocity. Osteoporos Int. 1996, 6 (1): 1-7. 10.1007/BF01626530.CrossRefPubMed
13.
Zurück zum Zitat Report of a WHO Study Group: Assessment of fracture risk and its application to screening for postmenopausal osteoporosis. World Health Organ Tech Rep Ser. 1994, 843: 1-129. Report of a WHO Study Group: Assessment of fracture risk and its application to screening for postmenopausal osteoporosis. World Health Organ Tech Rep Ser. 1994, 843: 1-129.
14.
Zurück zum Zitat Nguyen TV, Center JR, Eisman JA: Bone mineral density-independent association of quantitative ultrasound measurements and fracture risk in women. Osteoporos Int. 2004, 15 (12): 942-947. 10.1007/s00198-004-1717-z.CrossRefPubMed Nguyen TV, Center JR, Eisman JA: Bone mineral density-independent association of quantitative ultrasound measurements and fracture risk in women. Osteoporos Int. 2004, 15 (12): 942-947. 10.1007/s00198-004-1717-z.CrossRefPubMed
15.
Zurück zum Zitat Damilakis J, Papadokostakis G, Vrahoriti H, Tsagaraki I, Perisinakis K, Hadjipavlou A, Gourtsoyiannis N: Ultrasound velocity through the cortex of phalanges, radius, and tibia in normal and osteoporotic postmenopausal women using a new multisite quantitative ultrasound device. Invest Radiol. 2003, 38 (4): 207-211.PubMed Damilakis J, Papadokostakis G, Vrahoriti H, Tsagaraki I, Perisinakis K, Hadjipavlou A, Gourtsoyiannis N: Ultrasound velocity through the cortex of phalanges, radius, and tibia in normal and osteoporotic postmenopausal women using a new multisite quantitative ultrasound device. Invest Radiol. 2003, 38 (4): 207-211.PubMed
16.
Zurück zum Zitat Hans D, Srivastav SK, Singal C, Barkmann R, Njeh CF, Kantorovich E, Glüer CC, Genant HK: Does combining the results from multiple bone sites measured by a new quantitative ultrasound device improve discrimination of hip fracture?. J Bone Miner Res. 1999, 14 (4): 644-651. 10.1359/jbmr.1999.14.4.644.CrossRefPubMed Hans D, Srivastav SK, Singal C, Barkmann R, Njeh CF, Kantorovich E, Glüer CC, Genant HK: Does combining the results from multiple bone sites measured by a new quantitative ultrasound device improve discrimination of hip fracture?. J Bone Miner Res. 1999, 14 (4): 644-651. 10.1359/jbmr.1999.14.4.644.CrossRefPubMed
17.
Zurück zum Zitat Hans D, Genton L, Allaoua S, Pichard C, Slosman DO: Hip fracture discrimination study: QUS of the radius and the calcaneum. J Clin Densitom. 2003, 6 (2): 163-172. 10.1385/JCD:6:2:163.CrossRefPubMed Hans D, Genton L, Allaoua S, Pichard C, Slosman DO: Hip fracture discrimination study: QUS of the radius and the calcaneum. J Clin Densitom. 2003, 6 (2): 163-172. 10.1385/JCD:6:2:163.CrossRefPubMed
18.
Zurück zum Zitat Weiss M, Ben-Shlomo A, Hagag P, Ish-Shalom S: Discrimination of proximal hip fracture by quantitative ultrasound measurement at the radius. Osteoporos Int. 2000, 11 (5): 411-416. 10.1007/s001980070108.CrossRefPubMed Weiss M, Ben-Shlomo A, Hagag P, Ish-Shalom S: Discrimination of proximal hip fracture by quantitative ultrasound measurement at the radius. Osteoporos Int. 2000, 11 (5): 411-416. 10.1007/s001980070108.CrossRefPubMed
19.
Zurück zum Zitat Stegman MR, Heaney RP, Travers-Gustafson D, Leist J: Cortical ultrasound velocity as an indicator of bone status. Osteoporos Int. 1995, 5 (5): 349-353. 10.1007/BF01622257.CrossRefPubMed Stegman MR, Heaney RP, Travers-Gustafson D, Leist J: Cortical ultrasound velocity as an indicator of bone status. Osteoporos Int. 1995, 5 (5): 349-353. 10.1007/BF01622257.CrossRefPubMed
20.
Zurück zum Zitat Augat P, Fan B, Lane NE, Lang TF, LeHir P, Lu Y, Uffmann M, Genant HK: Assessment of bone mineral at appendicular sites in females with fractures of the proximal femur. Bone. 1998, 22 (4): 395-402. 10.1016/S8756-3282(97)00302-5.CrossRefPubMed Augat P, Fan B, Lane NE, Lang TF, LeHir P, Lu Y, Uffmann M, Genant HK: Assessment of bone mineral at appendicular sites in females with fractures of the proximal femur. Bone. 1998, 22 (4): 395-402. 10.1016/S8756-3282(97)00302-5.CrossRefPubMed
21.
Zurück zum Zitat Talmant M, Kolta S, Roux C, Haguenauer D, Vedel I, Cassou B, Bossy E, Laugier P: In vivo performance evaluation of bi-directional ultrasonic axial transmission for cortical bone assessment. Ultrasound Med Biol. 2009, 35 (6): 912-919. 10.1016/j.ultrasmedbio.2008.12.008.CrossRefPubMed Talmant M, Kolta S, Roux C, Haguenauer D, Vedel I, Cassou B, Bossy E, Laugier P: In vivo performance evaluation of bi-directional ultrasonic axial transmission for cortical bone assessment. Ultrasound Med Biol. 2009, 35 (6): 912-919. 10.1016/j.ultrasmedbio.2008.12.008.CrossRefPubMed
22.
Zurück zum Zitat Durosier C, Hans D, Krieg MA, Ruffieux C, Cornuz J, Meunier PJ, Schott AM: Combining clinical factors and quantitative ultrasound improves the detection of women both at low and high risk for hip fracture. Osteoporos Int. 2007, 18 (12): 1651-1659. 10.1007/s00198-007-0414-0.CrossRefPubMed Durosier C, Hans D, Krieg MA, Ruffieux C, Cornuz J, Meunier PJ, Schott AM: Combining clinical factors and quantitative ultrasound improves the detection of women both at low and high risk for hip fracture. Osteoporos Int. 2007, 18 (12): 1651-1659. 10.1007/s00198-007-0414-0.CrossRefPubMed
23.
Zurück zum Zitat Karjalainen JP, Riekkinen O, Töyräs J, Hakulinen M, Kröger H, Rikkonen T, Salovaara K, Jurvelin JS: Multi-site bone ultrasound measurements in elderly women with and without previous hip fractures. Osteoporos Int. 2012, 23 (4): 1287-1295. 10.1007/s00198-011-1682-2.CrossRefPubMed Karjalainen JP, Riekkinen O, Töyräs J, Hakulinen M, Kröger H, Rikkonen T, Salovaara K, Jurvelin JS: Multi-site bone ultrasound measurements in elderly women with and without previous hip fractures. Osteoporos Int. 2012, 23 (4): 1287-1295. 10.1007/s00198-011-1682-2.CrossRefPubMed
24.
Zurück zum Zitat Hans D, Durosier C, Kanis JA, Johansson H, Schott-Pethelaz AM, Krieg MA: Assessment of the 10-year probability of osteoporotic hip fracture combining clinical risk factors and heel bone ultrasound: the EPISEM prospective cohort of 12,958 elderly women. J Bone Miner Res. 2008, 23 (7): 1045-1051. 10.1359/jbmr.080229.CrossRefPubMed Hans D, Durosier C, Kanis JA, Johansson H, Schott-Pethelaz AM, Krieg MA: Assessment of the 10-year probability of osteoporotic hip fracture combining clinical risk factors and heel bone ultrasound: the EPISEM prospective cohort of 12,958 elderly women. J Bone Miner Res. 2008, 23 (7): 1045-1051. 10.1359/jbmr.080229.CrossRefPubMed
25.
Zurück zum Zitat Knapp KM, Blake GM, Spector TD, Fogelman I: Multisite quantitative ultrasound: precision, age- and menopause-related changes, fracture discrimination, and T-score equivalence with dual-energy X-ray absorptiometry. Osteoporos Int. 2001, 12 (6): 456-464. 10.1007/s001980170090.CrossRefPubMed Knapp KM, Blake GM, Spector TD, Fogelman I: Multisite quantitative ultrasound: precision, age- and menopause-related changes, fracture discrimination, and T-score equivalence with dual-energy X-ray absorptiometry. Osteoporos Int. 2001, 12 (6): 456-464. 10.1007/s001980170090.CrossRefPubMed
26.
Zurück zum Zitat Moilanen P, Määttä M, Kilappa V, Xu L, Nicholson PH, Alén M, Timonen J, Jämsä T, Cheng S: Discrimination of fractures by low-frequency axial transmission ultrasound in postmenopausal females. Osteoporos Int. 2012, 24 (2): 723-730.CrossRefPubMed Moilanen P, Määttä M, Kilappa V, Xu L, Nicholson PH, Alén M, Timonen J, Jämsä T, Cheng S: Discrimination of fractures by low-frequency axial transmission ultrasound in postmenopausal females. Osteoporos Int. 2012, 24 (2): 723-730.CrossRefPubMed
Metadaten
Titel
Association between low-frequency ultrasound and hip fractures - comparison with DXA-based BMD
verfasst von
Mikko Määttä
Petro Moilanen
Jussi Timonen
Pasi Pulkkinen
Raija Korpelainen
Timo Jämsä
Publikationsdatum
01.12.2014
Verlag
BioMed Central
Erschienen in
BMC Musculoskeletal Disorders / Ausgabe 1/2014
Elektronische ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-15-208

Weitere Artikel der Ausgabe 1/2014

BMC Musculoskeletal Disorders 1/2014 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.