Automated semi-quantitative amyloid PET analysis technique without MR images for Alzheimer’s disease

verfasst von: Etsuko Imabayashi, Naoyuki Tamamura, Yuzuho Yamaguchi, Yuto Kamitaka, Muneyuki Sakata, Kenji Ishii

Erschienen in: Annals of Nuclear Medicine | Ausgabe 10/2022

Abstract

Objective

Although beta-amyloid (Aβ) positron emission tomography (PET) images are interpreted visually as positive or negative, approximately 10% are judged as equivocal in Alzheimer’s disease. Therefore, we aimed to develop an automated semi-quantitative analysis technique using 18F-flutemetamol PET images without anatomical images.

Methods

Overall, 136 cases of patients administered 18F-flutemetamol were enrolled. Of 136 cases, five PET images each with the highest and lowest values of standardized uptake value ratio (SUVr) of cerebral cortex-to-pons were used to create positive and negative templates. Using these templates, PET images of the remaining 126 cases were standardized, and SUVr images were produced with the pons as a reference region. The mean of SUVr values in the volume of interest delineated on the cerebral cortex was compared to those in the CortexID Suite (GE Healthcare). Furthermore, centiloid (CL) values were calculated for the 126 cases using data from the Centiloid Project (http://​www.​gaain.​org/​centiloid-project) and both templates. 18F-flutemetamol-PET was interpreted visually as positive/negative based on Aβ deposition in the cortex. However, the criterion "equivocal" was added for cases with focal or mild Aβ accumulation that were difficult to categorize. Optimal cutoff values of SUVr and CL maximizing sensitivity and specificity for Aβ detection were determined by receiver operating characteristic (ROC) analysis using the visual evaluation as a standard of truth.

Results

SUVr calculated by our method and CortexID were highly correlated (R2 = 0.9657). The 126 PET images comprised 84 negative and 42 positive cases of Aβ deposition by visual evaluation, of which 11 and 10 were classified as equivocal, respectively. ROC analyses determined the optimal cutoff values, sensitivity, and specificity for SUVr as 0.544, 89.3%, and 92.9%, respectively, and for CL as 12.400, 94.0%, and 92.9%, respectively. Both semi-quantitative analyses showed that 12 and 9 of the 21 equivocal cases were negative and positive, respectively, under the optimal cutoff values.

Conclusions

This semi-quantitative analysis technique using 18F-flutemetamol-PET calculated SUVr and CL automatically without anatomical images. Moreover, it objectively and homogeneously interpreted positive or negative Aβ burden in the brain as a supplemental tool for the visual reading of equivocal cases in routine clinical practice.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s12149-022-01769-x.
Etsuko Imabayashi and Naoyuki Tamamura contributed equally to this work.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

As of 2021, three amyloid positron-emission tomography (PET) tracers, 18F-florbetapir, 18F-florbetaben, and 18F-flutemetamol, are approved as pharmaceutical diagnostic agents by the Ministry of Health, Labour, and Welfare in Japan. These tracers are indicated for PET imaging of the brain to estimate β-amyloid (Aβ) neuritic plaque density in adult patients with cognitive impairment being evaluated for Alzheimer's disease (AD) and other causes of cognitive decline. Readers visually interpret the PET images distinctly as defined for each tracer, but not quantitatively, and classify them as positive or negative depending on the Aβ deposition in the brain. Regarding the positivity/negativity of Aβ deposition in the brain, the inter-rater agreement of visual assessment of 11C-Pittuberg compound B (PiB) [13] PET images, the most commonly used amyloid PET tracer worldwide, was higher (κ = 0.9) than those of 18F-florbetapir [4, 5], 18F-florbetaben [6], and 18F-flutemetamol [7 − 9] (κ = 0.8 each). Approximately 10% of Aβ-PET images are interpreted as equivocal as a border case between positivity and negativity, causing false-positive/negative results [10 − 12]. There have been reports that a semi-quantitative analysis using cortical standardized uptake value ratio (SUVr) to the cerebellar cortex or pons was useful for the evaluation of equivocal cases [3, 10, 11]. However, the SUVr is influenced by and varies according to differences in tracers, PET scanners, and/or imaging protocol such as radioactivity, time before images are acquired, and scan time, and hence, it is not suitable for quantification analysis of amyloid imaging [13]. This causes issues in a multicenter study.
One of the methods to solve this problem is the centiloid (CL) scale developed by Klunk et al. [13]. Briefly, CL values are calculated by converting the SUVr of each 18F-amyloid PET image to the SUVr obtained from images at 50–70 min postinjection of 11C-PiB and standardizing the semi-quantitative amyloid imaging measures to a scale from 0 to 100. The 0-anchor is intended to represent a definitively non-amyloid brain, while the 100-anchor is intended to represent the amount of global amyloid deposition found in a typical mild-to-moderate AD patient [13]. The CL methodology allows the calculation of a calibrated measure of Aβ binding for different tracers at different institutes if data are acquired with a consistent acquisition/reconstruction methodology [13]; therefore, this method has been adopted in recent clinical trials [1316]. Furthermore, it has been reported that the CL method is useful for the positive/negative evaluation of Aβ burden in equivocal cases by visual interpretation [17].
Generally, MR images are required for anatomical standardization of PET images, followed by the calculation of SUVr using the standardized PET images; then, the resulting SUVr is converted to the CL values by referring to the recommended protocol of the Global Alzheimer’s Association Information Network (http://​www.​gaain.​org). Therefore, it is difficult to utilize the CL method for PET images without the corresponding MR images, in clinical practice. To overcome this, several groups have reported quantitative analysis methods to calculate CL values using only PET images without anatomical imaging data such as MRI or CT images [1820]. We also developed a new technique that enabled to standardize PET images anatomically without MR images, which can set the composite volume of interest (VOI) and calculate SUVr and CL values in the VOIs automatically with simple operations. In addition, we found an optimal cutoff value for each SUVr and CL value in the composite VOI for separating positive and negative Aβ deposition in the brain by the receiver operating characteristic (ROC) curve, based on the visual evaluation of 18F-flutemetamol-PET as a standard of truth (SOT). Our semi-quantitative analysis technique is versatile and may improve the objectivity and reproducibility of the evaluation with 18F-flutemetamol-PET for AD as a supplemental tool in routine clinical practice.

Materials and methods

Participants

Data of 136 participants (40 men and 96 women) with a mean age of 79.2 years (range 68–86 years) were obtained from Tokyo Metropolitan Geriatric Hospital and Institute of Gerontology (TMIG) between July 2020 and January 2021. The selection criteria included patients diagnosed with AD, mild cognitive impairment (MCI), or cognitive unimpaired at our center; those who underwent both MRI (T1-weighted imaging) and PET scan using 18F-flutemetamol within the past 6 months; and those who were analyzed with VSRAD™ (Eisai Co., Ltd., Tokyo, Japan) (but not a requirement). The exclusion criteria were patients with poor quality MR and PET images (e.g., low resolution, distortion, artifacts, and lack of whole-brain coverage); patients with brain tumors or intracerebral hemorrhage that would interfere with the analysis; and patients deemed inappropriate by the principal investigator (K.I.). Among PET images from the 136 participants, images from 10 participants were used for creating positive and negative templates. The remaining 126 participants were classified as negative in 84 cases and positive in 42 cases for Aβ deposition by visual reading, and 11 of the negative and 10 of positive cases were classified as equivocal. The final clinical diagnosis was normal in 92 cases for normal, while in 34 cases, the diagnosis was MCI due to AD or no dementia due to AD. The other epidemiological characteristics of the participants were summarized in Table 1. The study was approved by the institutional ethics committee of TMIG, and written informed consent was obtained from all participants.

MR and PET imaging

Amyloid PET images were obtained using an integrated PET/CT scanner, Discovery 710, and Discovery MI (GE Healthcare, Milwaukee, United States). At approximately 90 min postinjection of 180.9 ± 7.9 (range, 140.1–200.4) MBq 18F-flutemetamol synthesized by FASTlab™ (GE Healthcare), 30-min list-mode PET scan was started, and the PET images were reconstructed using the 3D ordered subset expectation maximization with a time-of-flight procedure (Iteration 4, subset 16, and filter cut off 4 mm).
MRI was performed using Achieva 1.5 T (Philips Healthcare, Andover, MA, the United States). T1-weighted 3D turbo field echo SENSE1 sequence was used for volumetric MRI. The clinical report including physical examination, MRI, and other clinical examinations were all obtained not earlier than 5 months before the PET/CT scan.

Visual reading

PET/CT images were visually assessed by two experienced nuclear medicine specialists (E.I. and K.I.) certified by the Japanese Society of Nuclear Medicine and evaluated with the MRI results. The brightness of the pons on the PET image was adjusted to 90% of the maximum intensity of the color scale, and the accumulation of Aβ was evaluated as negative or positive in five regions of the brain (frontal lobe, temporal lobe, parietal lobe, posterior cingulate gyrus and precuneus, and striatum). If any one of the five regions was positive, it was judged as Aβ positive; if all five regions were negative, it was considered as Aβ negative. Participants ruled out as negative or positive because of mild accumulation and/or uncertain extent of Aβ in the brain were classified as equivocal by consensus between the two readers. The SUVr of the cerebral cortex-to-pons was measured using CortexID Suite (GE Healthcare) in all cases.

Image processing

Figure 1 shows the workflow of the procedure to create the optimal template for each participant and the data analyses. For the deformation of PET images, we used a program that implements a technique for representing deformation fields using basic functions as well as Statistical Parametric Mapping (SPM). This is an in-house program based on a previous report [21].

PET template-based spatial normalization

Five deformed PET images each with the highest and lowest SUVr in cerebral cortex measured using the CortexID Suite were selected; arithmetic mean-based positive/negative templates were created. The accumulation of Aβ varied from patient to patient, and the standardization error increased with a single template. Therefore, the positive and negative templates were weighted from 0.1% to 99.9% and from 99.9% to 0.1% in 0.1% increments, and the weighted average was used as a candidate template for evaluating the similarity with the participant's PET images by Zero-mean normalized cross-correlation (ZNCC) [22], The candidate template with the highest similarity was adopted as the optimal template. The similarity was calculated as follows:
Step 1: Subject PET images were affine-transformed based on the averaged images of the positive and negative templates.
Step 2: ZNCC was calculated for the weighted-average template candidates in 0.1% increments and the affine-transformed images.
Step 3: The template candidate with the highest ZNCC value was adopted as the optimal template.
Finally, the PET images of 126 cases excluding the 10 cases used for the template creation were standardized directly based on the optimal template without MR images.

Centiloid process

The SUVr values were converted to the CL values by referring to the method by Klunk et al. [13]. Firstly, the standard VOI of 1.0-mm pixel pitch published in the Centiloid Project (http://​www.​gaain.​org/​centiloid-project) was converted to 1.5-mm pixel pitch and used to calculate the CL values (Fig. 2a). The mean pixel values were measured in each VOI put on the cerebral cortex, total cerebellum, and pons; the SUVr in the cerebral cortex was calculated as the mean pixel value of the cerebral cortex divided by that of the total cerebellum; and then, the SUVr was transferred to CL values using the formula mentioned above (the mean pixel value of pons was used for the calculation of SUVr in each region of the brain). Secondly, the anatomical standardized images for the 18F-flutemetamol-PET images published in the Centiloid Project were deformed using arithmetic mean-based positive/negative templates created in this study, followed by calculation of SUVrFlute for those images. Then, the SUVr values corresponding to the SUVrPiB in the Centiloid Project were calculated from the SUVrFlute.

Quantitative analyses

The anatomically standardized 126 PET images (Montreal Neurological Institute [MNI] coordinates) were transformed into Talairach standard brain PET images (Talairach) using the pre-determined deformation parameters from MNI standard brain to Talairach standard brain [23]. The MNI to Talairach deformation parameters were obtained by using the Talairach-shaped FDG-PET images included in NEUROSTAT (https://​neurostat.​neuro.​utah.​edu/​) as templates and linearly and nonlinearly deforming the MNI-shaped PET images included in SPM. An in-house program based on previous reports [21] was used for this deformation.
The VOIs for each region (frontal lobes, parietal lobes, temporal lobes, posterior cingulate gyrus, precuneus, and basal ganglia) were drawn to measure the pixel values in only the gray matter to avoid counting signals from white matter, based on the Talairach Daemon (Fig. 2b, c), and the VOIs of these six regions were integrated to calculate the SUVr (defined as "Composite") with the pons as the reference region. To verify the validity of the automatically calculated Composite SUVr, the values were compared with those calculated by CortexID Suite, which is used worldwide for quantitative analysis of 18F-flutemetamol PET images. Furthermore, the CL values for each participant were calculated for the 126 anatomically standardized PET images (MNI) using the SUVr to CL value conversion formula mentioned above. Finally, the ROC curve was drawn using EZR [24] and the optimal CL cutoff using the evaluation of visual reading of 18F-flutemetamol-PET images as SOT was determined by ROC analysis, maximizing the Youden’s index (sensitivity + specificity—1). The area under the curve (AUC), sensitivity and specificity at the optimal CL value were then calculated.

Statistics

The correlation between SUVr and CL values calculated by our method and SUVr calculated by CortexID Suite were examined with linear regression and coefficient of determination (R2) using Microsoft Excel 2016 (Microsoft Japan, Tokyo, Japan). The differences of SUVr and CL values calculated by the current method between the groups (negative, equivocal, positive) evaluated by visual reading were compared using the Mann–Whitney U test, and p < 0.05 was set as the level of significance. The reproducibility of evaluation between visual reading and semi-quantitative analysis for the participants was assessed using Cohen’s kappa.

Results

Validity of anatomical standardization

Five PET images each with the lowest and highest value of SUVr in 18F-flutemetamol accumulation between the cerebral cortex and pons were standardized anatomically and then normalized based on the mean pixel in the pons. The negative and positive standard brain templates obtained by weighted averaging are shown in Fig. 3a, b, respectively. Then, the weighted average image of these templates was used as the optimal template to obtain the highest similarity to each participant PET image. For example, the optimal template (d − f) for each case (a − c) with CL values of -8.98 (a and d), 22.41 (b and e), and 91.65 (c and f) are shown in Fig. 4, respectively.

Validity of SUVr

To verify the validity of the composite SUVr calculated automatically using our semi-quantitative analysis method, we compared them with that calculated by CortexID Suite. As shown in Fig. 5A, the composite SUVr in the current method correlated well with that in Cortex ID Suite, with a coefficient of determination (R2) of 0.9657 (slope = 0.9516, intercept = 0.0679).

Comparison with GAAIN data

The formula to convert the SUVr to the CL values directly was defined as follows: CL = 122.83 × SUVrFlute − 126.13. The coefficient of determination (R2) between the CL value calculated by this conversion formula and the CL value published in the Centiloid Project was 0.983 (slope: 1.000, intercept: 0.538).

Validity of the results of quantitative analyses

The Composite SUVr and CL values of flutemetamol-PET imaging calculated by the current method showed a high correlation (R2 = 0.9188). The values of SUVr and CL assessed by visual reading are shown in Table 1 and Fig. 6. The mean ± standard deviation (median, range) of SUVr values in each negative, equivocal, and positive group classified by visual reading were 0.45 ± 0.04 (0.48, 0.39 − 0.55), 0.48 ± 0.05 (0.51, 0.41 − 0.61), and 0.65 ± 0.07 (0 0.70, 0.48 − 0.80), respectively. Similarly, those of CL values were − 1.91 ± 8.08 (− 2.50, − 18.5 − 22.4), 10.2 ± 16.1 (6.83, − 14.2 − 48.1), and 61.1 ± 18.9 (64.30, − 4.16 − 94.4), respectively. Significant differences in SUVr and CL values between all groups were observed (p < 0.01).
We investigated the optimal cutoff value of CL units for separating positive and negative Aβ deposition in the brain by ROC curve analysis (Fig. 7), based on the evaluation of visual reading of 18F-flutemetamol-PET images as a SOT. As a result, the maximum value of AUC was 0.973 (95% CI: 0.943–1) and the optimal CL cutoff value was 12.400 for CL values, and the sensitivity and specificity at the optimal cutoff value were 94.0% and 92.9%, respectively (Fig. 7a). On the other hand, the maximum value of AUC, 95% CI, optimal cutoff, sensitivity, and specificity for SUVr were 0.977, 0.957 − 0.997, 0.544, 89.3%, and 92.9%, respectively (Fig. 7b). The reproducibility of the results of evaluation between visual reading and quantitative analysis at the optimal cutoff was examined using Cohen's kappa for all 126 cases (visual reading: 84 negative and 42 positive) and 21 equivocal cases (visual reading: 11 negative and 10 positive) (Table 2). In 126 participants, the optimal SUVr cutoff value was reassessed as 78 negative and 48 positive, and kappa coefficient was 0.7932, while those values for the optimal CL cutoff value were 82 negative and 44 positive, kappa coefficient was 0.8588. Both the quantitative analyses showed that 12 and 9 cases of the 21 equivocal cases were negative and positive, respectively, under the optimal cutoff values, and the kappa coefficient was 0.7123.
Table 1
Characteristics of patients
  
Visual evaluation
All
Negative
Equivocal
Positive
Number
126
73
21
32
Age (range)
79.3 ± 3.9 (68 − 86)
78.6 ± 4.1 (68 − 86)
80.7 ± 3.6 (75 − 86)
80.1 ± 3.3 (72 − 86)
MMSE
27.7 ± 2.4
(21 − 30)
27.9 ± 2.3 (21 − 30)
28.1 ± 1.8 (24 − 30)
27.0 ± 2.7 (21 − 30)
Clinical diagnosis
(normal, MCI)
90, 36
57, 16
16, 5
17, 15
SUVr
0.51 ± 01 (0.39 − 0.80)
0.45 ± 0.04
(0.39 − 0.55)
0.48 ± 0.05 (0.41 − 0.61)
0.65 ± 0.07 (0.48 − 0.80)
CL
16.1 ± 29.7
(-18.5 − 94.4)
-1.91 ± 8.08
(-18.5 − 22.4)
10.2 ± 16.1
(-14.2 − 48.1)
61.1 ± 18.9
(-4.16 − 94.4)
Mean ± SD (range). CL centiloid, MCI mild cognitive impairment, MMSE minimental state examination, SUVr standard uptake value ratio
Table 2
Reproducibility of the visual and semi-quantitative assessments
All
VR
 
Equivocal
VR
 
 
P
N
  
P
N
 
SUVr
P
39
9
48
SUVr
P
8
1
9
N
3
75
78
N
2
10
12
κ = 0.7931
 
42
84
126
κ = 0.7123
 
10
11
21
All
VR
 
Equivocal
VR
 
 
P
N
  
P
N
 
CL
P
39
5
44
CL
P
8
1
9
N
3
79
82
N
2
10
12
κ = 0.8588
 
42
84
126
κ = 0.7123
 
10
11
21
CL centiloid, N negative, P positive, SUVr standard uptake value ratio, VR visual reading

Discussion

In this study, a new technique was developed to quantify the accumulation of Aβ in the brain automatically using PET images alone, without anatomical imaging data such as MRI or CT. As the accumulation of Aβ in the brains of patients with Alzheimer's disease varies broadly, using a single template for anatomical standardization of PET images increases the error [25]. Edison et al. reported that the SUVr was overestimated if the quantification was performed with standardized PET images using an average template when compared to those performed with standardized MR images [26]. However, Bourgeat et al. reported that an optimal template, which is the weighted average image of the positive and negative templates to obtain the highest similarity to the participant PET image, can reduce the error of standardization for each patient [25]. We also found that the use of a single average template in the quantitative analysis of 18F-flutemetamol-PET images underestimated the values of SUVr in the high SUVr group and overestimated those in the low SUVr group and that the weighted average of the positive and negative PET images could improve these under/overestimations (Supplemental Fig. 1). However, we developed a new semi-quantitative analysis method to automatically calculate SUVr and CL values without any anatomical images and validated the accuracy of the spatial normalization of Aβ images. In the spatial normalization of PET images, we utilized a program package that includes a method to express (or represent) a deformation field using basis functions; this program was developed based on a previous report [26]. As a result, the composite SUVr calculated automatically with the current semi-quantitative analysis method showed a high correlation with that in CortexID Suite (R2 = 0.9657) and CL values (R2 = 0.9188) (Fig. 5a, b). Therefore, our method will enable the following:
  • Semi-quantitative analysis of Aβ-PET images without anatomical data such as MRI or CT
  • Easy calculation of SUVr and CL and evaluation of Aβ deposition on a uniform scale
  • Reduction of inter-operator variability of data and improvement of reproducibility through automation of data processing
As shown in Figs. 5b and 6b, although there was a significant difference in the CL values between cases evaluated visually to be negative and equivocal, their ranges overlapped and it was difficult to discriminate the two groups by even quantitative analysis using CL values. In contrast, cases that were positive and both negative and/or equivocal in visual reading could be distinguished by CL values, and the sensitivity and specificity at the optimal cutoff CL value of 12.400 were 94.0% and 92.9%, respectively, if the evaluation of visual reading was used as SOT. In the ROC analysis for SUVr and CL values, both the sensitivity and specificity at the optimal cutoff values were higher in the CL values (Fig. 7), which may be due to the difference in the shape of the VOIs and standard brain used for calculating each semi-quantitative parameter. While CL VOIs were created with 18F-FDG-PET [13] and values were calculated directly from 18F-flutemetamol-PET images, which were deformed referring to the MNI standard brain (asymmetrically shaped structure), the SUVr values were calculated in two steps: deformation of the PET images with reference to the MNI standard brain, followed by transformation of the deformed PET images to fit the Talairach standard (symmetry) and VOIs were created. Since the shape of the human brain is asymmetrical, the process of deformation from the MNI to the Talairach standard brain probably affected the sensitivity and specificity calculated based on the SUVr. Therefore, we believe that if SUVr could be calculated using the same methodology as that used to calculate CL, the sensitivity and specificity of SUVr would be equivalent to those of CL. Furthermore, the VOIs used to measure CL were created by identifying regions specific for the positive group by comparing two groups, namely positive and negative Aβ deposition, whereas the VOIs for each brain region used to calculate SUVr were generated based on anatomical locations, referring to the Talairach Daemon. In other words, although the VOIs for each region excluded white matter, they did not exclusively consist of pixels specific to the Aβ-PET-positive group but also included some pixels in the VOIs that involved high or low accumulation in both the positive and negative groups. These differences in creating VOIs used in our study probably also affected the calculation of sensitivity and specificity at the optimal cutoff values for SUVr and CL.
Of the 21 cases assessed to be equivocal by visual reading, 12 and 9 cases were objectively classified as negative and positive, respectively, by the optimal cutoff values of both SUVr and CL. Furthermore, the reproducibility (Cohen's kappa = 0.7123) of the evaluation between visual reading and our semi-quantitative analyses for the equivocal cases was acceptable. These results suggest that the current semi-quantitative evaluation is useful as an assistant tool for the visual reading of 18F-flutemetamol-PET images (the global standard for the evaluation of 18F-flutemetamol-PET imaging is visual reading.).
There were three cases in which the CL values were below the optimal cutoff value (CL = 6.6, 3.8, and − 4.2) although the visual readings were positive. In the two cases with CL = 6.6 and 3.8, Aβ accumulation was visually observed only in the temporoparietal lobe and posterior cingulate gyrus in addition to a part of the composite VOI, respectively, which could not be assessed as positive for Aβ accumulation in the quantitative analysis. Moreover, because the visually positive cases with negative CL values showed a significant accumulation of 18F-flutemetamol visually in the occipital lobe, which is not included in the evaluation area of our composite VOI, there was a discrepancy between the visual reading and current semi-quantitative analysis method (Table 2).
The optimal cutoff value for CL in the current study was 12.400, which was lower than that reported by Amadoru et al. for participants with AD [27]. This may be because participants in our study were healthy participants and patients with MCI, in whom Aβ burden is at an early stage, supported by the report that the optimal cutoff for CL for healthy participants aged 70 and above was 11.9 [28]. Considering the importance of assessing Aβ deposition in the brain at an earlier stage in the treatment of AD, our semi-quantitative analysis technique of measuring Aβ in the brain using the CL scale is considered to be useful in the diagnosis of AD, and it is expected that our cutoff value will be useful in clinical use. However, it is important to conduct further studies to evaluate Aβ accumulation in various clinical stages of AD before the cutoff can be used for actual discrimination.
There are some limitations to our semi-quantitative evaluation of Aβ burden. As mentioned, participants with Aβ deposition in some cortical areas only (e.g., occipital lobe) that cannot be detected by our method are evaluated as false negatives. The software might be unable to exert anatomical standardization precisely in case of severe brain infarction or ventriculomegaly. Cases in which a part of the brain is out of the visual field of PET images would lead to a mistake in the standardization during data processing according to our methodology. Further, the protocol of Aβ-PET imaging must be fixed to achieve more accurate analytical results because SUVr and CL values may fluctuate if the time between the administration of 18F-flutemetamol and the initiation of imaging is not kept constant. In addition, data processing for anatomical transformation and semi-quantitative parameter estimation is complex and relatively time-consuming. Therefore, improving processing speed is a future challenge to satisfy the requirements for clinical practice.
In conclusion, our semi-quantitative analysis technique using 18F-flutemetamol-PET can be used to calculate SUVr and CL automatically without anatomical images. Moreover, it can also be used to interpret positive or negative Aβ burden in the brain objectively and homogeneously and serve as a supplemental tool for the visual reading of equivocal cases in routine clinical practice.

Declarations

Ethical approval

All procedures performed in studies involving human participants were approved by the Ethics Committee of the Tokyo Metropolitan Institute of Gerontology (Approved number: R21-041).
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

© Springer Medizin

Bis 11. April 2024 bestellen und im ersten Jahr 50 % sparen!

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

© Springer Medizin

Bis 11. April 2024 bestellen und im ersten Jahr 50 % sparen!

Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Ng S, Villemagne VL, Berlangieri S, Lee ST, Cherk M, Gong SJ, et al. Visual assessment versus quantitative assessment of 11C-PIB PET and 18F-FDG PET for detection of Alzheimer’s disease. J Nucl Med. 2007;48:547–52.CrossRef Ng S, Villemagne VL, Berlangieri S, Lee ST, Cherk M, Gong SJ, et al. Visual assessment versus quantitative assessment of 11C-PIB PET and 18F-FDG PET for detection of Alzheimer’s disease. J Nucl Med. 2007;48:547–52.CrossRef
2.
Zurück zum Zitat Suotunen T, Hirvonen J, Immonen-Räihä P, Aalto S, Lisinen I, Arponen E, et al. Visual assessment of 11C-PIB PET in patients with cognitive impairment. Eur J Nucl Med Mol Imaging. 2010;37:1141–7.CrossRef Suotunen T, Hirvonen J, Immonen-Räihä P, Aalto S, Lisinen I, Arponen E, et al. Visual assessment of 11C-PIB PET in patients with cognitive impairment. Eur J Nucl Med Mol Imaging. 2010;37:1141–7.CrossRef
3.
Zurück zum Zitat Yamane T, Ishii K, Sakata M, Ikari Y, Nishio T, Ishii K, et al. Inter-rater variability of visual interpretation and comparison with quantitative evaluation of 11C-PiB PET amyloid images of the Japanese Alzheimer’s Disease Neuroimaging Initiative (J-ADNI) multicenter study. Eur J Nucl Med Mol Imaging. 2017;44:850–7.CrossRef Yamane T, Ishii K, Sakata M, Ikari Y, Nishio T, Ishii K, et al. Inter-rater variability of visual interpretation and comparison with quantitative evaluation of 11C-PiB PET amyloid images of the Japanese Alzheimer’s Disease Neuroimaging Initiative (J-ADNI) multicenter study. Eur J Nucl Med Mol Imaging. 2017;44:850–7.CrossRef
4.
Zurück zum Zitat Namiki C, Takita Y, Iwata A, Momose T, Senda M, Okubo Y, et al. Imaging characteristics and safety of florbetapir (18F) in Japanese healthy volunteers, patients with mild cognitive impairment and patients with Alzheimer’s disease. Ann Nucl Med. 2015;29:570–81.CrossRef Namiki C, Takita Y, Iwata A, Momose T, Senda M, Okubo Y, et al. Imaging characteristics and safety of florbetapir (18F) in Japanese healthy volunteers, patients with mild cognitive impairment and patients with Alzheimer’s disease. Ann Nucl Med. 2015;29:570–81.CrossRef
5.
Zurück zum Zitat Fakhry-Darian D, Patel NH, Khan S, Barwick T, Svensson W, Khan S, et al. Optimisation and usefulness of quantitative analysis of 18F-florbetapir PET. Br J Radiol. 2019;92:20181020.CrossRef Fakhry-Darian D, Patel NH, Khan S, Barwick T, Svensson W, Khan S, et al. Optimisation and usefulness of quantitative analysis of 18F-florbetapir PET. Br J Radiol. 2019;92:20181020.CrossRef
6.
Zurück zum Zitat Kim JY, Oh D, Sung K, Choi H, Paeng JC, Cheon GJ, et al. Visual interpretation of 18F-Florbetaben PET supported by deep learning-based estimation of amyloid burden. Eur J Nucl Med Mol Imaging. 2021;48:1116–23.CrossRef Kim JY, Oh D, Sung K, Choi H, Paeng JC, Cheon GJ, et al. Visual interpretation of 18F-Florbetaben PET supported by deep learning-based estimation of amyloid burden. Eur J Nucl Med Mol Imaging. 2021;48:1116–23.CrossRef
7.
Zurück zum Zitat Hatashita S, Yamasaki H, Suzuki Y, Tanaka K, Wakebe D, Hayakawa H. 18F-Flutemetamol amyloid-beta PET imaging compared with 11C-PIB across the spectrum of Alzheimer’s disease. Eur J Nucl Med Mol Imaging. 2014;41:290–300.CrossRef Hatashita S, Yamasaki H, Suzuki Y, Tanaka K, Wakebe D, Hayakawa H. 18F-Flutemetamol amyloid-beta PET imaging compared with 11C-PIB across the spectrum of Alzheimer’s disease. Eur J Nucl Med Mol Imaging. 2014;41:290–300.CrossRef
8.
Zurück zum Zitat Mountz JM, Laymon CM, Cohen AD, Zhang Z, Price JC, Boudhar S, et al. Comparison of qualitative and quantitative imaging characteristics of 11C-PiB and 18F-flutemetamol in normal control and Alzheimer’s subjects. Neuroimage Clin. 2015;9:592–8.CrossRef Mountz JM, Laymon CM, Cohen AD, Zhang Z, Price JC, Boudhar S, et al. Comparison of qualitative and quantitative imaging characteristics of 11C-PiB and 18F-flutemetamol in normal control and Alzheimer’s subjects. Neuroimage Clin. 2015;9:592–8.CrossRef
9.
Zurück zum Zitat Collij LE, Konijnenberg E, Reimand J, Kate MT, Braber AD, Lopes Alves I, et al. Assessing amyloid pathology in cognitively normal subjects using 18F-flutemetamol PET: comparing visual reads and quantitative methods. J Nucl Med. 2019;60:541–7.CrossRef Collij LE, Konijnenberg E, Reimand J, Kate MT, Braber AD, Lopes Alves I, et al. Assessing amyloid pathology in cognitively normal subjects using 18F-flutemetamol PET: comparing visual reads and quantitative methods. J Nucl Med. 2019;60:541–7.CrossRef
10.
Zurück zum Zitat Son HJ, Oh JS, Oh M, Kim SJ, Lee JH, Roh JH, et al. The clinical feasibility of deep learning-based classification of amyloid PET images in visually equivocal cases. Eur J Nucl Med Mol Imaging. 2020;47:332–41.CrossRef Son HJ, Oh JS, Oh M, Kim SJ, Lee JH, Roh JH, et al. The clinical feasibility of deep learning-based classification of amyloid PET images in visually equivocal cases. Eur J Nucl Med Mol Imaging. 2020;47:332–41.CrossRef
11.
Zurück zum Zitat Okada Y, Kato T, Iwata K, Kimura Y, Nakamura A, Hattori H, et al. Evaluation of PiB visual interpretation with CSF Aβ and longitudinal SUVr in J-ADNI study. Ann Nucl Med. 2020;34:108–18.CrossRef Okada Y, Kato T, Iwata K, Kimura Y, Nakamura A, Hattori H, et al. Evaluation of PiB visual interpretation with CSF Aβ and longitudinal SUVr in J-ADNI study. Ann Nucl Med. 2020;34:108–18.CrossRef
12.
Zurück zum Zitat Payoux P, Delrieu J, Gallini A, Adel D, Salabert AS, Hitzel A, et al. Cognitive and functional patterns of nondemented subjects with equivocal visual amyloid PET findings. Eur J Nucl Med Mol Imaging. 2015;42:1459–68.CrossRef Payoux P, Delrieu J, Gallini A, Adel D, Salabert AS, Hitzel A, et al. Cognitive and functional patterns of nondemented subjects with equivocal visual amyloid PET findings. Eur J Nucl Med Mol Imaging. 2015;42:1459–68.CrossRef
13.
Zurück zum Zitat Klunk WE, Koeppe RA, Price JC, Benzinger TL, Devous MD Sr, Jagust WJ, et al. The Centiloid Project: standardizing quantitative amyloid plaque estimation by PET. Alzheimers Dement. 2015;11:1–15.CrossRef Klunk WE, Koeppe RA, Price JC, Benzinger TL, Devous MD Sr, Jagust WJ, et al. The Centiloid Project: standardizing quantitative amyloid plaque estimation by PET. Alzheimers Dement. 2015;11:1–15.CrossRef
14.
Zurück zum Zitat Mintun MA, Lo AC, Duggan Evans C, Wessels AM, Ardayfio PA, Andersen SW, et al. Donanemab in early Alzheimer’s disease. N Engl J Med. 2021;384:1691–704.CrossRef Mintun MA, Lo AC, Duggan Evans C, Wessels AM, Ardayfio PA, Andersen SW, et al. Donanemab in early Alzheimer’s disease. N Engl J Med. 2021;384:1691–704.CrossRef
15.
Zurück zum Zitat Klein G, Delmar P, Kerchner GA, Hofmann C, Abi-Saab D, Davis A, et al. Thirty-Six-month amyloid positron emission tomography results show continued reduction in amyloid burden with subcutaneous gantenerumab. J Prev Alzheimers Dis. 2021;8:3–6.PubMed Klein G, Delmar P, Kerchner GA, Hofmann C, Abi-Saab D, Davis A, et al. Thirty-Six-month amyloid positron emission tomography results show continued reduction in amyloid burden with subcutaneous gantenerumab. J Prev Alzheimers Dis. 2021;8:3–6.PubMed
16.
Zurück zum Zitat Lopes Alves I, Collij LE, Altomare D, Frisoni GB, Saint-Aubert L, Payoux P, et al. Quantitative amyloid PET in Alzheimer’s disease: the AMYPAD prognostic and natural history study. Alzheimers Dement. 2020;16:750–8.CrossRef Lopes Alves I, Collij LE, Altomare D, Frisoni GB, Saint-Aubert L, Payoux P, et al. Quantitative amyloid PET in Alzheimer’s disease: the AMYPAD prognostic and natural history study. Alzheimers Dement. 2020;16:750–8.CrossRef
17.
Zurück zum Zitat Matsuda H, Ito K, Ishii K, Shimosegawa E, Okazawa H, Mishina M, et al. Quantitative evaluation of 18F-flutemetamol PET in patients with cognitive impairment and suspected alzheimer’s disease: a multicenter study. Front Neurol. 2021;11: 578753.CrossRef Matsuda H, Ito K, Ishii K, Shimosegawa E, Okazawa H, Mishina M, et al. Quantitative evaluation of 18F-flutemetamol PET in patients with cognitive impairment and suspected alzheimer’s disease: a multicenter study. Front Neurol. 2021;11: 578753.CrossRef
18.
Zurück zum Zitat Lilja J, Leuzy A, Chiotis K, Savitcheva I, Sörensen J, Nordberg A. Spatial normalization of 18F-flutemetamol PET images using an adaptive principal-component template. J Nucl Med. 2019;60:285–91.CrossRef Lilja J, Leuzy A, Chiotis K, Savitcheva I, Sörensen J, Nordberg A. Spatial normalization of 18F-flutemetamol PET images using an adaptive principal-component template. J Nucl Med. 2019;60:285–91.CrossRef
19.
Zurück zum Zitat Iaccarino L, La Joie R, Koeppe R, Siegel BA, Hillner BE, Gatsonis C, et al. rPOP: Robust PET-only processing of community acquired heterogeneous amyloid-PET data. Neuroimage. 2022;246: 118775.CrossRef Iaccarino L, La Joie R, Koeppe R, Siegel BA, Hillner BE, Gatsonis C, et al. rPOP: Robust PET-only processing of community acquired heterogeneous amyloid-PET data. Neuroimage. 2022;246: 118775.CrossRef
20.
Zurück zum Zitat Akamatsu G, Ikari Y, Ohnishi A, Nishida H, Aita K, Sasaki M, et al. Automated PET-only quantification of amyloid deposition with adaptive template and empirically pre-defined ROI. Phys Med Biol. 2016;61:5768–80.CrossRef Akamatsu G, Ikari Y, Ohnishi A, Nishida H, Aita K, Sasaki M, et al. Automated PET-only quantification of amyloid deposition with adaptive template and empirically pre-defined ROI. Phys Med Biol. 2016;61:5768–80.CrossRef
21.
Zurück zum Zitat Ashburner J, Friston KJ. Nonlinear spatial normalization using basis functions. Hum Brain Mapp. 1999;7:254–66.CrossRef Ashburner J, Friston KJ. Nonlinear spatial normalization using basis functions. Hum Brain Mapp. 1999;7:254–66.CrossRef
22.
Zurück zum Zitat Mattoccia S, Tombari F, Di Stefano L. Reliable rejection of mismatching candidates for efficient ZNCC template matching. 15th IEEE International Conference on Image Processing. 2008;849−52. Mattoccia S, Tombari F, Di Stefano L. Reliable rejection of mismatching candidates for efficient ZNCC template matching. 15th IEEE International Conference on Image Processing. 2008;849−52.
23.
Zurück zum Zitat Lancaster JL, Woldorff MG, Parsons LM, Liotti M, Freitas CS, Rainey L, et al. Automated talairach atlas labels for functional brain mapping. Hum Brain Mapp. 2000;10:120–31.CrossRef Lancaster JL, Woldorff MG, Parsons LM, Liotti M, Freitas CS, Rainey L, et al. Automated talairach atlas labels for functional brain mapping. Hum Brain Mapp. 2000;10:120–31.CrossRef
24.
Zurück zum Zitat Kanda Y. Investigation of the freely available easy-to-use software ‘EZR’ for medical statistics. Bone Marrow Transpl. 2013;48:452–8.CrossRef Kanda Y. Investigation of the freely available easy-to-use software ‘EZR’ for medical statistics. Bone Marrow Transpl. 2013;48:452–8.CrossRef
25.
Zurück zum Zitat Bourgeat P, Villemagne VL, Dore V, Brown B, Macaulay SL, Martins R, et al. Comparison of MR-less PiB SUVR quantification methods. Neurobiol Aging. 2015;36(Suppl 1):S159–66.CrossRef Bourgeat P, Villemagne VL, Dore V, Brown B, Macaulay SL, Martins R, et al. Comparison of MR-less PiB SUVR quantification methods. Neurobiol Aging. 2015;36(Suppl 1):S159–66.CrossRef
26.
Zurück zum Zitat Edison P, Carter SF, Rinne JO, Gelosa G, Herholz K, Nordberg A, et al. Comparison of MRI based and PET template based approaches in the quantitative analysis of amyloid imaging with PIB-PET. Neuroimage. 2013;70:423–33.CrossRef Edison P, Carter SF, Rinne JO, Gelosa G, Herholz K, Nordberg A, et al. Comparison of MRI based and PET template based approaches in the quantitative analysis of amyloid imaging with PIB-PET. Neuroimage. 2013;70:423–33.CrossRef
27.
Zurück zum Zitat Amadoru S, Doré V, McLean CA, Hinton F, Shepherd CE, Halliday GM, et al. Comparison of amyloid PET measured in Centiloid units with neuropathological findings in Alzheimer’s disease. Alzheimers Res Ther. 2020;12:22.CrossRef Amadoru S, Doré V, McLean CA, Hinton F, Shepherd CE, Halliday GM, et al. Comparison of amyloid PET measured in Centiloid units with neuropathological findings in Alzheimer’s disease. Alzheimers Res Ther. 2020;12:22.CrossRef
28.
Zurück zum Zitat Su Y, Flores S, Hornbeck RC, Speidel B, Vlassenko AG, Gordon BA, et al. Utilizing the centiloid scale in cross-sectional and longitudinal PiB PET studies. Neuroimage Clin. 2018;19:406–16.CrossRef Su Y, Flores S, Hornbeck RC, Speidel B, Vlassenko AG, Gordon BA, et al. Utilizing the centiloid scale in cross-sectional and longitudinal PiB PET studies. Neuroimage Clin. 2018;19:406–16.CrossRef
Metadaten
Titel
Automated semi-quantitative amyloid PET analysis technique without MR images for Alzheimer’s disease
verfasst von
Etsuko Imabayashi
Naoyuki Tamamura
Yuzuho Yamaguchi
Yuto Kamitaka
Muneyuki Sakata
Kenji Ishii
Publikationsdatum
11.07.2022
Verlag
Springer Nature Singapore
Erschienen in
Annals of Nuclear Medicine / Ausgabe 10/2022
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-022-01769-x