Biomechanical comparisons of current suspensory fixation devices for anterior cruciate ligament reconstruction

verfasst von: Jin Cheng, Siddhartha Venkata Paluvadi, SungJae Lee, SeungJin Yoo, Eun-Kyoo Song, Jong-Keun Seon

Erschienen in: International Orthopaedics | Ausgabe 6/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Cortical suspensory devices are routinely used for femoral side fixation of soft tissue graft in anterior cruciate ligament (ACL) reconstruction. The purpose of this study was to evaluate the biomechanical properties of a new adjustable loop device (GraftMax®) compared with established devices (EndoButton® and TightRope®) in ACL reconstruction and to investigate whether knotting the free end of TightRope could improve biomechanical properties.

Methods

The three cortical suspensory devices (GraftMax® Button; Conmed, EndoButton® CL; Smith & Nephew, and TightRope® RT; Arthrex) were tested under cyclic load (50–250 N for 1000 cycles) and pull-to-failure conditions at 50 mm/h in a device-only setup using a tensile testing machine. The TightRope was additionally tested with its free suture ends knotted. The statistical analyses were done with one-way analysis of variance (ANOVA) and post hoc Tukey HSD tests.

Results

There are significant differences in the load-to-failure among the devices. The EndoButton showed the highest mean failure load at 1204.7 N compared to other devices (GraftMax (914.2 N), knotted TightRope (868.1 N) and TightRope (800.1 N) (p < 0.001). The mean total displacement after 1000 cycles was 0.76 mm, 2.11 mm, 1.56 mm and 1.38 mm for the EndoButton, GraftMax, TightRope, and knotted TightRope, respectively. The EndoButton showed significantly better properties than both the GraftMax (p = 0.000) and the TightRope (p = 0.020) but not the knotted TightRope (p = n.s.) in total displacement. However, there was no significant difference between the TightRope and GraftMax (p = n.s.).

Conclusion

The fixed loop (EndoButton) showed significantly better mechanical properties in failure load and displacement than TightRope or GraftMax in this biomechanical study. However, the mechanical properties of the GraftMax is comparable to the TightRope. Moreover, the knotting of TightRope improved mechanical properties in total displacement more than TightRope, but not in failure load.

Clinical relevance

The biomechanical properties of the GraftMax are comparable to the TightRope. The TightRope, when knotted, shows an improvement both in load to failure and cyclic displacement, though the differences are not significant.
Literatur
1.
Zurück zum Zitat Kilinc BE, Kara A, Oc Y, Celik H, Camur S, Bilgin E, Erten YT, Sahinkaya T, Eren OT (2016) Transtibial vs anatomical single bundle technique for anterior cruciate ligament reconstruction: a retrospective cohort study. Int J Surg 29:62–69CrossRefPubMed Kilinc BE, Kara A, Oc Y, Celik H, Camur S, Bilgin E, Erten YT, Sahinkaya T, Eren OT (2016) Transtibial vs anatomical single bundle technique for anterior cruciate ligament reconstruction: a retrospective cohort study. Int J Surg 29:62–69CrossRefPubMed
2.
Zurück zum Zitat Nyland J, Mattocks A, Kibbe S, Kalloub A, Greene JW, Caborn DN (2016) Anterior cruciate ligament reconstruction, rehabilitation, and return to play: 2015 update. Open Access J Sports Med 24(7):21–32CrossRef Nyland J, Mattocks A, Kibbe S, Kalloub A, Greene JW, Caborn DN (2016) Anterior cruciate ligament reconstruction, rehabilitation, and return to play: 2015 update. Open Access J Sports Med 24(7):21–32CrossRef
3.
Zurück zum Zitat Ahmad CS, Gardner TR, Groh M, Arnouk J, Levine WN (2004) Mechanical properties of soft tissue femoral fixation devices for anterior cruciate ligament reconstruction. Am J Sports Med 32:635–640CrossRefPubMed Ahmad CS, Gardner TR, Groh M, Arnouk J, Levine WN (2004) Mechanical properties of soft tissue femoral fixation devices for anterior cruciate ligament reconstruction. Am J Sports Med 32:635–640CrossRefPubMed
4.
Zurück zum Zitat Kim KI, Lee SH, Bae C, Bae SH (2017) Three-dimensional reconstruction computed tomography evaluation of the tunnel location and angle in anatomic single-bundle anterior cruciate ligament reconstruction: a comparison of the anteromedial portal and outside-in techniques. Knee Surg Relat Res 29:11–18CrossRefPubMedPubMedCentral Kim KI, Lee SH, Bae C, Bae SH (2017) Three-dimensional reconstruction computed tomography evaluation of the tunnel location and angle in anatomic single-bundle anterior cruciate ligament reconstruction: a comparison of the anteromedial portal and outside-in techniques. Knee Surg Relat Res 29:11–18CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Barrow AE, Pilia M, Guda T, Kadrmas WR, Burns TC (2014) Femoral suspension devices for anterior cruciate ligament reconstruction: do adjustable loops lengthen? Am J Sports Med 42:343–349CrossRefPubMed Barrow AE, Pilia M, Guda T, Kadrmas WR, Burns TC (2014) Femoral suspension devices for anterior cruciate ligament reconstruction: do adjustable loops lengthen? Am J Sports Med 42:343–349CrossRefPubMed
6.
Zurück zum Zitat Petre BM, Smith SD, Jansson KS et al (2013) Femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction: a comparative biomechanical study. Am J Sports Med 41:416–422CrossRefPubMed Petre BM, Smith SD, Jansson KS et al (2013) Femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction: a comparative biomechanical study. Am J Sports Med 41:416–422CrossRefPubMed
7.
Zurück zum Zitat Johnson JS, Smith SD, LaPrade CM, Turnbull TL, LaPrade RF, Wijdicks CA (2015) A biomechanical comparison of femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction under high loads. Am J Sports Med 43:154–160CrossRefPubMed Johnson JS, Smith SD, LaPrade CM, Turnbull TL, LaPrade RF, Wijdicks CA (2015) A biomechanical comparison of femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction under high loads. Am J Sports Med 43:154–160CrossRefPubMed
9.
Zurück zum Zitat Born TR, Biercevicz AM, Koruprolu SC, Paller D, Spenciner D, Fadale PD (2016) Biomechanical and computed tomography analysis of adjustable femoral cortical fixation devices for anterior Cruciate ligament reconstruction in a cadaveric human knee model. Arthroscopy 32:253–261CrossRefPubMed Born TR, Biercevicz AM, Koruprolu SC, Paller D, Spenciner D, Fadale PD (2016) Biomechanical and computed tomography analysis of adjustable femoral cortical fixation devices for anterior Cruciate ligament reconstruction in a cadaveric human knee model. Arthroscopy 32:253–261CrossRefPubMed
10.
Zurück zum Zitat Eguchi A, Ochi M, Adachi N, Deie M, Nakamae A, Usman MA (2014) Mechanical properties of suspensory fixation devices for anterior cruciate ligament reconstruction: comparison of fixed-length loop device versus the adjustable-length loop device. Knee 21:743–748CrossRefPubMed Eguchi A, Ochi M, Adachi N, Deie M, Nakamae A, Usman MA (2014) Mechanical properties of suspensory fixation devices for anterior cruciate ligament reconstruction: comparison of fixed-length loop device versus the adjustable-length loop device. Knee 21:743–748CrossRefPubMed
11.
Zurück zum Zitat Daniel DM, Stone ML, Sachs R, Malcom L (1985) Instrumented measurement of anterior knee laxity in patients with acute anterior cruciate ligament disruption. Am J Sports Med 13:401–407CrossRefPubMed Daniel DM, Stone ML, Sachs R, Malcom L (1985) Instrumented measurement of anterior knee laxity in patients with acute anterior cruciate ligament disruption. Am J Sports Med 13:401–407CrossRefPubMed
Metadaten
Titel
Biomechanical comparisons of current suspensory fixation devices for anterior cruciate ligament reconstruction
verfasst von
Jin Cheng
Siddhartha Venkata Paluvadi
SungJae Lee
SeungJin Yoo
Eun-Kyoo Song
Jong-Keun Seon
Publikationsdatum
29.01.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
International Orthopaedics / Ausgabe 6/2018
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-018-3780-7

Weitere Artikel der Ausgabe 6/2018

International Orthopaedics 6/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.