24.01.2018 | Arthroscopy and Sports Medicine | Ausgabe 4/2018

Archives of Orthopaedic and Trauma Surgery 4/2018

Biomechanical performance of the Actifit® scaffold is significantly improved by selection of irrigation fluid

Zeitschrift:
Archives of Orthopaedic and Trauma Surgery > Ausgabe 4/2018
Autoren:
Arnd Hoburg, Philipp von Roth, Shimon Roy-Ali, Jan-Erik Ode, Dag Wulsten, Tobias M. Jung, Clemens Gwinner

Abstract

Purpose

Clinical reports on meniscal scaffolds seem promising, albeit relatively paucity exists regarding their biomechanical behavior. The aim of the study is to delineate the impact of differing suture materials and the type as well as the temperature of the irrigation fluid on the pull-out strength of a polyurethane meniscal scaffold (Actifit®).

Materials and methods

128 specimens were utilized with horizontal sutures and uniaxial load-to-failure testing was performed. We compared two different suture materials—polydioxanone (PDS) and non-absorbable, braided polyester sutures (NABP)—as well as two common irrigation fluids—lactated Ringer’s and electrolyte-free, hypotonic Mannitol-Sorbitol. All specimens were further evaluated according to two different temperatures [room temperature (20 °C) and near-core body temperature (37 °C)].

Results

Mean load-to-failure was 53.3 ± 6.5 N. There was no significant difference between the NABP and the PDS group. Ringer group showed a significantly higher load-to-failure compared to Purisole (P = .0002). This was equivalent for both PDS (P = .0008) and NABP sutures (P = .0008). Significantly higher failure loads could be established for the 37° group (P = .041); yet, this difference was neither confirmed for the PDS or in in the NABP subgroup. Only the subgroup using Purisole at 37° showed significantly higher failure loads compared to 20° (P = .017).

Conclusions

This study underlines the potential to improve pull-out strength during implantation of an Actifit® scaffold by alteration of the type of irrigation fluid. Lactated Ringer solution provided the highest construct stability in regard to load-to-failure testing and should be considered whenever implantation of a polyurethane meniscal scaffold is conducted.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 4/2018

Archives of Orthopaedic and Trauma Surgery 4/2018 Zur Ausgabe
  1. Das kostenlose Testabonnement läuft nach 14 Tagen automatisch und formlos aus. Dieses Abonnement kann nur einmal getestet werden.

Neu im Fachgebiet Orthopädie und Unfallchirurgie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Orthopädie und Unfallchirurgie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise