Biomechanical performance of the Actifit® scaffold is significantly improved by selection of irrigation fluid

verfasst von: Arnd Hoburg, Philipp von Roth, Shimon Roy-Ali, Jan-Erik Ode, Dag Wulsten, Tobias M. Jung, Clemens Gwinner

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 4/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Clinical reports on meniscal scaffolds seem promising, albeit relatively paucity exists regarding their biomechanical behavior. The aim of the study is to delineate the impact of differing suture materials and the type as well as the temperature of the irrigation fluid on the pull-out strength of a polyurethane meniscal scaffold (Actifit®).

Materials and methods

128 specimens were utilized with horizontal sutures and uniaxial load-to-failure testing was performed. We compared two different suture materials—polydioxanone (PDS) and non-absorbable, braided polyester sutures (NABP)—as well as two common irrigation fluids—lactated Ringer’s and electrolyte-free, hypotonic Mannitol-Sorbitol. All specimens were further evaluated according to two different temperatures [room temperature (20 °C) and near-core body temperature (37 °C)].

Results

Mean load-to-failure was 53.3 ± 6.5 N. There was no significant difference between the NABP and the PDS group. Ringer group showed a significantly higher load-to-failure compared to Purisole (P = .0002). This was equivalent for both PDS (P = .0008) and NABP sutures (P = .0008). Significantly higher failure loads could be established for the 37° group (P = .041); yet, this difference was neither confirmed for the PDS or in in the NABP subgroup. Only the subgroup using Purisole at 37° showed significantly higher failure loads compared to 20° (P = .017).

Conclusions

This study underlines the potential to improve pull-out strength during implantation of an Actifit® scaffold by alteration of the type of irrigation fluid. Lactated Ringer solution provided the highest construct stability in regard to load-to-failure testing and should be considered whenever implantation of a polyurethane meniscal scaffold is conducted.
Literatur
4.
Zurück zum Zitat Chatain F, Adeleine P, Chambat P, Neyret P, Societe Francaise dA (2003) A comparative study of medial versus lateral arthroscopic partial meniscectomy on stable knees: 10-year minimum follow-up. Arthroscopy 19 (8):842–849CrossRefPubMed Chatain F, Adeleine P, Chambat P, Neyret P, Societe Francaise dA (2003) A comparative study of medial versus lateral arthroscopic partial meniscectomy on stable knees: 10-year minimum follow-up. Arthroscopy 19 (8):842–849CrossRefPubMed
5.
Zurück zum Zitat Baratz ME, Fu FH, Mengato R (1986) Meniscal tears: the effect of meniscectomy and of repair on intraarticular contact areas and stress in the human knee. A preliminary report. Am J Sports Med 14(4):270–275CrossRefPubMed Baratz ME, Fu FH, Mengato R (1986) Meniscal tears: the effect of meniscectomy and of repair on intraarticular contact areas and stress in the human knee. A preliminary report. Am J Sports Med 14(4):270–275CrossRefPubMed
6.
Zurück zum Zitat Fairbank TJ (1948) Knee joint changes after meniscectomy. J Bone Jt Surg Br 30B(4):664–670CrossRef Fairbank TJ (1948) Knee joint changes after meniscectomy. J Bone Jt Surg Br 30B(4):664–670CrossRef
12.
Zurück zum Zitat Schuttler KF, Haberhauer F, Gesslein M, Heyse TJ, Figiel J, Lorbach O, Efe T, Roessler PP (2016) Midterm follow-up after implantation of a polyurethane meniscal scaffold for segmental medial meniscus loss: maintenance of good clinical and MRI outcome. Knee Surg Sports Traumatol Arthrosc 24(5):1478–1484. https://doi.org/10.1007/s00167-015-3759-5 CrossRefPubMed Schuttler KF, Haberhauer F, Gesslein M, Heyse TJ, Figiel J, Lorbach O, Efe T, Roessler PP (2016) Midterm follow-up after implantation of a polyurethane meniscal scaffold for segmental medial meniscus loss: maintenance of good clinical and MRI outcome. Knee Surg Sports Traumatol Arthrosc 24(5):1478–1484. https://​doi.​org/​10.​1007/​s00167-015-3759-5 CrossRefPubMed
13.
Zurück zum Zitat Schuttler KF, Pottgen S, Getgood A, Rominger MB, Fuchs-Winkelmann S, Roessler PP, Ziring E, Efe T (2015) Improvement in outcomes after implantation of a novel polyurethane meniscal scaffold for the treatment of medial meniscus deficiency. Knee Surg Sports Traumatol Arthrosc 23(7):1929–1935. https://doi.org/10.1007/s00167-014-2977-6 CrossRefPubMed Schuttler KF, Pottgen S, Getgood A, Rominger MB, Fuchs-Winkelmann S, Roessler PP, Ziring E, Efe T (2015) Improvement in outcomes after implantation of a novel polyurethane meniscal scaffold for the treatment of medial meniscus deficiency. Knee Surg Sports Traumatol Arthrosc 23(7):1929–1935. https://​doi.​org/​10.​1007/​s00167-014-2977-6 CrossRefPubMed
16.
Zurück zum Zitat Zaffagnini S, Marcheggiani Muccioli GM, Lopomo N, Bruni D, Giordano G, Ravazzolo G, Molinari M, Marcacci M (2011) Prospective long-term outcomes of the medial collagen meniscus implant versus partial medial meniscectomy: a minimum 10-year follow-up study. Am J Sports Med 39(5):977–985. https://doi.org/10.1177/0363546510391179 CrossRefPubMed Zaffagnini S, Marcheggiani Muccioli GM, Lopomo N, Bruni D, Giordano G, Ravazzolo G, Molinari M, Marcacci M (2011) Prospective long-term outcomes of the medial collagen meniscus implant versus partial medial meniscectomy: a minimum 10-year follow-up study. Am J Sports Med 39(5):977–985. https://​doi.​org/​10.​1177/​0363546510391179​ CrossRefPubMed
17.
Zurück zum Zitat Durselen L, Hebisch A, Claes LE, Bauer G (2003) Gapping phenomenon of longitudinal meniscal tears. Clin Biomech (Bristol Avon) 18(6):505–510CrossRef Durselen L, Hebisch A, Claes LE, Bauer G (2003) Gapping phenomenon of longitudinal meniscal tears. Clin Biomech (Bristol Avon) 18(6):505–510CrossRef
18.
Zurück zum Zitat Shinjo H, Nakata K, Shino K, Hamada M, Nakamura N, Mae T, Miyama T, Horibe S, Yoshikawa H, Ochi T (2002) Effect of irrigation solutions for arthroscopic surgery on intraarticular tissue: comparison in human meniscus-derived primary cell culture between lactate Ringer’s solution and saline solution. J Orthop Res 20(6):1305–1310. https://doi.org/10.1016/S0736-0266(02)00062-1 CrossRefPubMed Shinjo H, Nakata K, Shino K, Hamada M, Nakamura N, Mae T, Miyama T, Horibe S, Yoshikawa H, Ochi T (2002) Effect of irrigation solutions for arthroscopic surgery on intraarticular tissue: comparison in human meniscus-derived primary cell culture between lactate Ringer’s solution and saline solution. J Orthop Res 20(6):1305–1310. https://​doi.​org/​10.​1016/​S0736-0266(02)00062-1 CrossRefPubMed
19.
Zurück zum Zitat Reagan BF, McInerny VK, Treadwell BV, Zarins B, Mankin HJ (1983) Irrigating solutions for arthroscopy. A metabolic study. J Bone Jt Surg Am 65(5):629–631CrossRef Reagan BF, McInerny VK, Treadwell BV, Zarins B, Mankin HJ (1983) Irrigating solutions for arthroscopy. A metabolic study. J Bone Jt Surg Am 65(5):629–631CrossRef
20.
Zurück zum Zitat Jurvelin JS, Jurvelin JA, Kiviranta I, Klauser RJ (1994) Effects of different irrigation liquids and times on articular cartilage: an experimental, biomechanical study. Arthroscopy 10(6):667–672CrossRefPubMed Jurvelin JS, Jurvelin JA, Kiviranta I, Klauser RJ (1994) Effects of different irrigation liquids and times on articular cartilage: an experimental, biomechanical study. Arthroscopy 10(6):667–672CrossRefPubMed
21.
Zurück zum Zitat Mah ET, Lee WK, Southwood RT, Carbone A, Leppard PJ (1991) Effects of irrigation fluid on human menisci: an experimental comparison of water, normal saline, and glycine. Arthroscopy 7(1):24–32CrossRefPubMed Mah ET, Lee WK, Southwood RT, Carbone A, Leppard PJ (1991) Effects of irrigation fluid on human menisci: an experimental comparison of water, normal saline, and glycine. Arthroscopy 7(1):24–32CrossRefPubMed
22.
Zurück zum Zitat Chao GT, Qian ZY, Huang MJ, Kan B, Gu YC, Gong CY, Yang JL, Wang K, Dai M, Li XY, Gou ML, Tu MJ, Wei YQ (2008) Synthesis, characterization, and hydrolytic degradation behavior of a novel biodegradable pH-sensitive hydrogel based on polycaprolactone, methacrylic acid, and poly(ethylene glycol). J Biomed Mater Res A 85(1):36–46. https://doi.org/10.1002/jbm.a.31362 CrossRefPubMed Chao GT, Qian ZY, Huang MJ, Kan B, Gu YC, Gong CY, Yang JL, Wang K, Dai M, Li XY, Gou ML, Tu MJ, Wei YQ (2008) Synthesis, characterization, and hydrolytic degradation behavior of a novel biodegradable pH-sensitive hydrogel based on polycaprolactone, methacrylic acid, and poly(ethylene glycol). J Biomed Mater Res A 85(1):36–46. https://​doi.​org/​10.​1002/​jbm.​a.​31362 CrossRefPubMed
23.
Zurück zum Zitat Marechal Y (2006) The Hydrogen Bond and the Water Molecule - The Physics and Chemistry of Water, Aqueous and Bio-Media. Elsevier, Amsterdam Marechal Y (2006) The Hydrogen Bond and the Water Molecule - The Physics and Chemistry of Water, Aqueous and Bio-Media. Elsevier, Amsterdam
27.
Zurück zum Zitat Dervin GF, Downing KJ, Keene GC, McBride DG (1997) Failure strengths of suture versus biodegradable arrow for meniscal repair: an in vitro study. Arthroscopy 13(3):296–300CrossRefPubMed Dervin GF, Downing KJ, Keene GC, McBride DG (1997) Failure strengths of suture versus biodegradable arrow for meniscal repair: an in vitro study. Arthroscopy 13(3):296–300CrossRefPubMed
28.
Zurück zum Zitat Kohn D, Siebert W (1989) Meniscus suture techniques: a comparative biomechanical cadaver study. Arthroscopy 5(4):324–327CrossRefPubMed Kohn D, Siebert W (1989) Meniscus suture techniques: a comparative biomechanical cadaver study. Arthroscopy 5(4):324–327CrossRefPubMed
29.
Zurück zum Zitat Post WR, Akers SR, Kish V (1997) Load to failure of common meniscal repair techniques: effects of suture technique and suture material. Arthroscopy 13(6):731–736CrossRefPubMed Post WR, Akers SR, Kish V (1997) Load to failure of common meniscal repair techniques: effects of suture technique and suture material. Arthroscopy 13(6):731–736CrossRefPubMed
30.
Zurück zum Zitat Rimmer MG, Nawana NS, Keene GC, Pearcy MJ (1995) Failure strengths of different meniscal suturing techniques. Arthroscopy 11(2):146–150CrossRefPubMed Rimmer MG, Nawana NS, Keene GC, Pearcy MJ (1995) Failure strengths of different meniscal suturing techniques. Arthroscopy 11(2):146–150CrossRefPubMed
Metadaten
Titel
Biomechanical performance of the Actifit® scaffold is significantly improved by selection of irrigation fluid
verfasst von
Arnd Hoburg
Philipp von Roth
Shimon Roy-Ali
Jan-Erik Ode
Dag Wulsten
Tobias M. Jung
Clemens Gwinner
Publikationsdatum
24.01.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 4/2018
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-018-2883-3

Weitere Artikel der Ausgabe 4/2018

Archives of Orthopaedic and Trauma Surgery 4/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.