Erschienen in: BMC Cardiovascular Disorders 1/2019

Open Access 01.12.2019 | Case report

Cardiac resynchronization therapy-defibrillator pocket infection caused by Mycobacterium fortuitum: a case report and review of the literature

verfasst von: Jun Zhu, Qingluan Yang, Junjie Pan, Haiming Shi, Bo Jin, Qiying Chen

Erschienen in: BMC Cardiovascular Disorders | Ausgabe 1/2019

Abstract

Background

With the rising utilization of cardiovascular implantable electronic devices (CIEDs), infections secondary to device implantation are increasingly encountered. Staphylococcus aureus and coagulase-negative staphylococci are usually the predominant causative organisms. A CIED infection due to non-tuberculous mycobacteria (NTM) is extremely rare.

Case presentation

A 68-year-old man was admitted to our hospital with a history of pain and swelling at his cardiac resynchronization therapy-defibrillator (CRT-D) pocket site, for 4 days. The CRT-D had been implanted 2 weeks prior. The exudate smear was positive for acid-fast bacilli and culture results revealed rapidly growing nontuberculous mycobacteria (RGM). After an urgent removal of the device followed by 1 year of antibiotic treatment, the patient was completely cured. A new device was finally implanted, 3 years later.

Conclusions

Infections caused by nontuberculous mycobacteria following the implantation of cardiac devices are very rare. The typical manifestations of post-implantation CIED infections caused by RGMs include an early onset, with local redness, swelling, and spontaneous drainage. Systemic symptoms such as fever, chills, and fatigue are absent. Mycobacterium fortuitum is the most common species of RGM implicated in CIED infections, the manifestations of which usually appear within several weeks of the implantation procedure. An urgent removal of the device and appropriate antibiotic therapy are essential therapeutic measures. This is the first such reported case, in which the patient has been re-implanted with another device at the same site, after achieving a complete cure. We followed-up the patient for an additional 3 years and observed that the patient remained free of infection. Our case report shows that though an RGM infection is rare and difficult to treat, it can be completely cured. In addition, we demonstrated that it is subsequently possible to safely re-implant a CIED for the patient, at the same site.
Abkürzungen
CIED
Cardiovascular implantable electronic devices
CRT-D
Cardiac resynchronization therapy-defibrillator
ICD
Implantable cardioverter defibrillator
NTM
Non-tuberculous mycobacteria
PM
Permanent pacemaker
RGM
Rapid growing nontuberculous mycobacteria

Background

In the last few decades, cardiac implantable electronic devices (CIEDs), which include permanent pacemakers (PMs), implantable cardioverter defibrillators (ICDs), and cardiac resynchronization therapy (CRT) devices have been widely utilized for the burgeoning aged population. As the survival benefits of CIED have been observed in clinical practice, their application is expected to increase. Despite the obvious benefits, the use of these implantable devices is also associated with serious complications. While improvements in implant technology have resulted in high success rates and lower occurrence of complications, an infection due to the device remains a major concern with the rate of secondary infection at about 1–2% [1]. An infection of the CIED significantly increases the morbidity, mortality, and healthcare costs of the procedure by prolonging the hospitalization time and mandating an aggressive management of the patient. An implant-associated infection can occur early or may be delayed for several months after the procedure. The main pathogens implicated in CIED infections are Staphylococcus aureus and coagulase-negative staphylococci [2]. A CIED infection caused by non-tuberculous mycobacteria (NTM) is extremely rare. To our knowledge, an infection has been reported in only 37 post-implantation cases, 28 of which were caused by a rapidly growing mycobacteria (RGM). Here, we present a case of CRT-D pocket infection caused by Mycobacterium fortuitum, which is the first reported case of its kind in China. It is also the first instance in which, after a complete recovery from the infection, a new device was re-implanted in the patient at the same site.

Case presentation

A 68-year-old man had a history of chronic systolic heart failure and dilated cardiomyopathy, with a left ventricular ejection fraction of 32% and a left ventricular internal diastolic diameter of 81 mm. His electrocardiograph had revealed a sinus rhythm with a QRS duration of 140 ms along with a left bundle branch block morphology. He underwent a cardiac resynchronization therapy-defibrillator (CRT-D) device implantation. However, 2 weeks after the procedure, he was re-admitted to our hospital with a 4-day history of pain and swelling at the CRT-D pocket site, associated with scant serous drainage. Prior to this admission, he had undergone a skin incision and drainage at another hospital, but the procedure had failed to relieve his symptoms. The patient had a history of hypertension, a myocardial bridge in the left anterior descending coronary artery and diabetes mellitus (on irregular therapy). He also had a history of pulmonary tuberculosis 40 years previously and had completed the curative treatment course successfully, at the time. His regular medications included metoprolol, perindopril, torsemide, and amiodarone.
During re-admission, the patient did not have complaints of fever, chills, or fatigue. The physical examination was unremarkable, except for a 2-cm long, open incision on the upper left side of the chest, with mild localized edema over the CRT-D pocket. The skin around the incision was erythematous, and a small amount of scant serous discharge was noted on pressing the pocket site. Considering the likelihood of CRT-D infection, we collected the patient’s blood samples to perform cultures for aerobic and anaerobic bacteria. The samples of the exudate were also tested, using microscopic examination and culture, not only for the usual causative bacteria but also for the rarer acid-fast bacilli. On initial microscopic examination, the exudate smear did not reveal any organisms. An echocardiogram showed no evidence of vegetation or thrombosis. Other laboratory test results, including a routine blood count and erythrocyte sedimentation rate, were within normal limits. The patient also tested negative for the human immunodeficiency virus (HIV) and the hepatitis B virus. He was started empirically on intravenous vancomycin and cefepime.
On the 3rd day of hospitalization, we performed a pocket reconstruction. The CRT-D was re-implanted below the pectoralis major muscle in the left pectoral region and connected with the pacing electrodes placed in the earlier pocket site through a subcutaneous tunnel. However, the incision site did not heal, and the localized edema worsened, with copious exudation from the drainage site. After the reoperation, the exudate was collected and cultured. Within 3 days, a smear test of the exudate revealed an acid-fast bacillus, and culture results showed a rapidly growing nontuberculous mycobacterium (RGM). As recommended by our infectious diseases department, the isolate was sent to the microbiology laboratory for a polymerase chain reaction test. An amplification of the bacterial 16S ribosomal RNA gene, helped us identify the bacterial species as Mycobacterium fortuitum (M. fortuitum). The patient was therefore diagnosed with CRT-D infection due to M. fortuitum.
Once the causative organism was definitively identified, the patient’s therapeutic antibiotic regimen was empirically changed to intravenous clarithromycin, moxifloxacin, and amikacin. At the same time, an incision and debridement of the first pocket site was performed, and the entire device along with the leads were extracted, percutaneously. With dedicated wound care, the swab culture tests eventually yielded negative results and the incision healed well. After 2 months of administration of combined intravenous treatment, the patient was discharged. He was further administered oral clarithromycin and moxifloxacin for 1 year. He was continued to follow up for next 2 years. During 3 years, we checked the pocket site every 3 months and evaluated his cardiac function by echocardiography, as well as recorded the QRS complex width by ECG. He remained asymptomatic of incision infection but we found that the patient’s heart failure condition had gradually deteriorated (left ventricular ejection fraction decreased from 32 to 26%) and life-threatening rapid ventricular arrhythmias occurred even under the optimal drug treatment. The patient was then re-implanted with another CRT-D device at nearly the same previous site. The symptoms of heart failure were significantly improved. ECG showed the QRS complex width was decreased from 140 ms to 112 ms. Echocardiography showed the LVEF were increased to 45% after 6 months of the re-implantation. The patient has now remained free of infection over 4 years (Table 1).
Table 1
Timeline
2011.5.30
Implanted the cardiac resynchronization therapy-defibrillator device
2011.6.11
Onset of pain and swelling at his CRT-D pocket site
2011.6.15
Empirically anti-infective treatment by intravenous Vancomycin and Cefepime.
2011.6.18
Pocket reconstruction and CRT-D re-implanted below pectoralis major muscle at the left pectoral region.
2011.6.21
Incision was not healed and exudation was collected for culture
2011.6.23
Exudation smear was positive for acid-fast bacillus and culture results showed rapid growing nontuberculous mycobacteria(RGM)
2011.6.24
PCR amplification of the bacterial 16S ribosomal RNA gene, it was identified as Mycobacterium fortuitum.
2011.6.25
Entire CRT-D device and leads were removed
2011.6.25
Antibacterial treatment: Clarithromycin and Moxifloxacin and Amikacin for 2 months.
2011.8.25
Therapy with oral Clarithromycin and Moxifloxacin for one year.
2012.8
Stop antibacterial treatment
2013.8
Follow- up after one year
2014.8
Follow-up two years and received a new CRT-D device implantation.
2018.8
Follow-up four-years after new device implanted. No symptoms of any infection.

Discussion and conclusions

With growing evidence that a CIED implant is of benefit to patients with chronic cardiac disease, the utilization of these devices has increased over the last few years. The incidence of CIED infection is therefore correspondingly on the rise. The reported rates of infection secondary to CIED implantation, differ across various studies, ranging from 0.13 to 19.9%. The most recent studies report a rate of infection between 1 and 2% [3, 4].
The risk factors for CIED infections include prior immunodeficient states, i.e. patients suffering from HIV infection, diabetes mellitus, and chronic renal insufficiency are at a greater risk. In addition, many studies have found a higher rate of infection associated with ICDs or CRT devices, nearly 2–5 times higher than that associated with PM [5, 6]. This may be due to the larger sizes of the ICD and CRT devices and the longer operative duration required for implanting them. The use of anticoagulants has been associated with a higher risk of pocket hematoma, which also exacerbates the risk of a generator pocket infection.
Gram-positive bacteria are the predominant causative pathogens of CIED infections. Coagulase-negative staphylococci and Staphylococcus aureus account for 70–80% of all infections following CIED implantations [7]. However, RGM infections in implanted cardiac devices are still very rare. We searched the PubMed database for any older studies (no specified start time), case reports, or reviews related to CIED infections, using several search strings that combined the names of the various NTM with the different implantable cardiac devices. To be more specific, we used the following free text terms: (rapid growing nontuberculous mycobacteria OR nontuberculous mycobacteria OR Mycobacterium fortuitum OR Mycobacterium chelonae OR Mycobacterium abscessus OR Mycobacterium smegmatis group OR Mycobacterium mucogenicum) AND (cardiovascular implantable electronic devices OR CIED OR pacemaker OR PM OR cardiac resynchronization therapy OR CRT OR cardiac resynchronization therapy-defibrillator OR CRT-D OR implantable cardioverter defibrillator OR ICD). A review published by Phadke et al. summarized 23 cases with CIED infection caused by RGM up to March 31, 2015 [8]. Including the cases in their study, we found records for a total of 27 older cases that were identified as CIED infections due to RGM [830]. This makes our patient only the 28th recorded case of CIED infection, caused by an RGM species.
RGM comprise approximately half of the currently known mycobacterial species and are divided into five major groups: Mycobacterium fortuitum, M. chelonae/M. abscessus complex, M. smegmatis, M. mucogenicum, and the pigmented RGM [31]. The RGM species are universal organisms that exist in soil, food, natural and tap water, as well as in various plants and animals [32, 33]. They can grow in municipal water systems as well as distilled water and are resistant to sterilizers, antiseptics, and other standard disinfectants [33, 34]. The most commonly encountered RGM species in clinical practice are M. fortuitum and M. abscessus/M. chelonae. Of the recorded 28 CIED infections caused by RGM (Table 2), 17 (61%) were reportedly caused by M. fortuitum. We found 4 reported cases each, of CIED infections caused by the Mycobacterium abscessus/chelonae complex and the Mycobacterium smegmatis group. Only 2 patients were found to be infected by the Mycobacterium mageritense group and the pigmented RGM were identified only in 1 case. While 16 (60%) patients had infections associated with permanent PMs and 9 patients presented with ICD infection, we found no previous reports of CIED following a CRT-D implantation procedure, which makes our case report the first of its kind with respect to the type of device involved. Of the 28 patients, 21 patients (75%) developed an infection in < 6 months after the device implantation, of which, 12 patients were affected within 4 weeks. Therefore, we can surmise that the patients presenting with an early-onset infection are likely to have contracted it primarily during the implantation procedure and not as a secondary infection through the bloodstream. Only 4 of the 28 cases developed clinical manifestations of an infection > 1 year after device implantation. To our knowledge, M. fortuitum is not considered a normal skin commensal. These infections are considered to have been postoperatively contracted from an environmental source. We observed that 13 (46%) of the 28 reported patients with an RGM infection had other associated infections caused by mycobacteria. While 3 of these 13 patients were found to fulfil the clinical criteria for infective endocarditis through echocardiographic findings, in 1 patient, the causative mycobacterium was isolated on culture postoperatively, thus fulfilling the pathologic criteria of diagnosis.
Table 2
Clinical and demographic information for published cases of cardiac device infections due to rapidly growing mycobacteria
Reference
Age (years) /sex
Onseta
Type
Bacteremiab
IEc
Organism
Device Removal
Antibiotics Treatment/Duration
Outcome
Mycobacterium fortuitum group
 Verghese, 1998
74/M
13 d
PPM
NR
NR
Mycobacterium fortuitum / Mycobacterium chelonae
Yes
Fluoroquinolone + Aminoglycoside /1 mo
Cured
 Sharma, 2005
62/F
6mo
PPM
Yes
Yes
Mycobacterium fortuitum
Yes
Clarithromycin + Ciprofloxacin/4wk, Doxycycline+ Ciprofloxacin /24wk
Cured
 Short,2005
74/M
6 wk
ICD
Yes
NR
Mycobacterium fortuitum
Yes
Clarithromycin + Ciprofloxacin /6wk
Cured
 Hemmersbach-Miller,2005
72/M
2wk
PPM
No
No
Mycobacterium fortuitum
Yes
Ciprofloxacin + Aminoglycoside /2wk, Ciprofloxacin /6mo
Cured
 Hemmersbach-Miller,2005
61/M
17mo
ICD
No
No
Mycobacterium fortuitum
Yes
Levofloxacin /≥1 yr
Cured
 Pastor, 2006
80/M
18 d
PPM
Yes
No
Mycobacterium fortuitum
No
Ciprofloxacin + Clarithromycin/6 wk
Cured
 Giannella, 2007
84/F
2 mo
PPM
No
No
Mycobacterium fortuitum
Yes
Levofloxacin/3 mo
Cured
 Siu, 2007
78/F
6 mo
PPM
Yes
No
Mycobacterium fortuitum
Yes
Levofloxacin + Clarithromycin/6 mo
Cured
 Tam, 2007
78/F
4 mo
PPM
Yes
NR
Mycobacterium fortuitum
Yes
Levofloxacin + Clarithromycin/6 mo
Cured
 Al Soub H, 2009
15/F
7 wk
PPM
Yes
No
Mycobacterium fortuitum
Yes
Ciprofloxacin + Clarithromycin/6 mo
Cured
 Van Duin,2010
78/M
NR
PPM
Yes
NR
Mycobacterium fortuitum
Yes
Ciprofloxacin + Clarithromycin/26wk
Cured
 Sharma,2012
43/M
4 yr
ICD
Yes
Yes
Mycobacterium fortuitum
Yes
Ciprofloxacin + Clarithromycin+ Aminoglycoside
Died
 Amraoui,2012
75/M
1 yr
PPM
Yes
No
Mycobacterium peregrinum
Yes
Ciprofloxacin + Clarithromycin/mo
Cured
 Yuhning,2012
56/M
4mo
ICD
Yes
No
Mycobacterium fortuitum
No
Cephalexin, Clarithromycin+ Moxifloxacin/ 2 wk
Died
 Varun,2016
60/M
6wk
ICD
Yes
No
Mycobacterium fortuitum
Yes
Ciprofloxacin + Clarithromycin+ Cefoxitin
Died
 Menfil Orellana-Barriors,2017
59/F
3wk
ICD
No
No
Mycobacterium fortuitum
Yes
Levofloxacin + Clarithromycin/23 wk
Cured
 2018d
68/ F
10 d
CRT
No
No
Mycobacterium fortuitum
Yes
Clarithromycin + Moxifloxacin/12 mo
Cured
Mycobacterium chelonae/abscessus complex
 Cutay, 1998
68/M
19 yr
PPM
NR
NR
Mycobacterium abscessus
Yes
Clarithromycin + Cefoxitin + Amikacin/5 wk
Died
 Kessler, 2004
53/F
2 wk
ICD
NR
NR
Mycobacterium abscessus
Yes
Clarithromycin/6 mo
Cured
 Simmon KE,2007
43/F
11mo
PPM
NR
NR
Mycobacterium massiliense
Yes
Clarithromycin/6 mo
Cured
 Hooda,2014
63/M
NR
PPM
No
Yes
Mycobacterium chelonae
Yes
Clarithromycin + Levofloxacin + Aminoglycoside />2mo
Cured
Mycobacterium smegmatis group
 Toda H,2006
86/M
16d
PPM
Yes
NR
Mycobacterium goodii
Yes
Minocycline + Aminoglycoside /2wk
Cured
 Chrissoheris MP,2008
85/M
<7d
ICD
No
No
Mycobacterium goodii
Yes
Trimethoprim/Sulfamethoxazole /8wk
Cured
 Marchandin H, 2009
23/M
8d
PPM
No
No
Mycobacterium goodii
No
Doxycycline +Fluoroquinolone /6mo
Cured
 David,2016
74/F
1mo
PPM
No
No
Mycobacterium goodii
Yes
Ciprofloxacin + Doxycycline/6mo
Cured
Mycobacterium mageritense group
 Tam, 2007
77/F
3 wk
PPM
NR
NR
Mycobacterium mageritense
No
Levofloxacin/6 mo
Cured
 Masato Fukunaga, 2016
59/F
14d
ICD
Yes
No
Mycobacterium mageritense
Yes
Ciprofloxacin + Clarithromycin/ 1 yr
Cured
The pigmented rapid growing species
 Karnam S, 2011
73/F
1 mo
ICD
No
NR
Mycobacterium phlei
Yes
Trimethoprim/Sulfamethoxazole +Doxycycline /12 mo
Cured
ICD implantable cardioverter-defibrillator, PPM permanent pacemaker, NR not reported, IE infective endocarditis. M Male, F Female, yr. year, mo month, wk. week, d day
aTime since most recent device manipulation
bDefined as positive blood culture, acid fast stain
cTransthoracic or transesophageal echocardiographic findings
dThe patient described in this article
The CIED infections are generally categorized as superficial or deep infections. The superficial infections like a generator pocket infection are commoner than the deep infections affecting the cardiac leads or valves. The most common manifestations of a generator pocket infection are fever, pain at the generator site, and wound discharge. Therefore, when an operated patient is observed to have slow wound healing or has a history of the implant site being re-opened and re-manipulated for any reason, an RGM infection should be strongly suspected (especially in an immunocompromised host). However, while a patient with an RGM may present with localized redness, swelling, and with spontaneous drainage, the typical systemic symptoms of fever, chills and fatigue are absent. The wound exudate is usually thin and watery but may sometimes be thick and purulent [33]. An echocardiography along with cultures of the patient’s blood, device pocket tissue, and leads are essential for the diagnostic evaluation. The tissue cultures of samples obtained during the surgery are more sensitive than swab cultures [35].
While the management of RGM infections must be tailored individually, an urgent device removal followed by antimicrobial therapy forms the mainstay of treatment in all cases. An American Heart Association statement recommended a complete removal of all hardware (Class I evidence) from patients, even in the absence of signs of systemic infection. An incomplete removal of CIED has resulted in higher infection relapse rates. These organisms can form biofilms, which make them difficult to eradicate with antimicrobial drug therapy alone. In our case, the device was urgently extracted from the patient and 3 years later, once he was disease-free, he was re-implanted with another device, which suggests the possibility of re-implantation after cure in other such patients.
After removal of the device, antibiotic treatment is essential. An antimicrobial drug susceptibility test is important for the selection of the appropriate antibiotic therapy. A recommended diagnostic and therapeutic flow chart is shown in Fig. 1. M. fortuitum is usually found to be susceptible to amikacin, cefoxitin, sulfonamides, ciprofloxacin, imipenem and moxifloxacin, with about half of the isolates susceptible to doxycycline [36]. Approximately 80% of M. fortuitum isolates can be inhibited by clarithromycin. Among newer antimicrobials, linezolid has been shown to have in vitro activity against the M. fortuitum group [37]. Unfortunately, most of the effective antimicrobial drugs for RGM have potentially adverse proarrhythmic effects. In our patient, we regretted the fallacy of prescribing antimicrobials empirically, without an initial drug susceptibility test. The treatment duration is dependent on the completeness of device removal, the greater the number of components of a device left in situ, the longer the duration of treatment. A longer duration of antibacterial therapy is necessary due to the propensity of the organisms towards biofilm formation. In 16 of the 28 recorded cases, the antibiotic treatment duration was ≥6 months. The optimal timing of device replacement after removal is not defined, as it may vary individually, depending on the extent of infection in each patient. It is recommended that the device should be replaced on the contralateral side from the older infected site [35]. In our patient, we implanted a new CRT-D, 3 years after the removal of the older infected device. As a generator placed on the right side could decrease the effect of defibrillation, the new device was implanted on the same side (left side) as before. Over 3 years of follow-up, the patient has shown no recurrence of an RGM infection.
Preventive measures are extremely important to minimize the risk of infection. A microbial contamination can occur during implantation, post-implantation or can occur secondary to an erosion of the device through skin. Several observational studies have demonstrated that almost half of the CIED infections can be prevented by prophylactic antibiotic therapy administered in the perioperative period [38]. The risk of infection is proportional to the duration of the implantation procedure. Therefore, we can reduce risk of CIED infections by strict aseptic manipulation, perioperative antibiotic prophylaxis, and by limiting the operative duration. Furthermore, temporary pacing should be avoided, if possible, because temporary transvenous pacing is associated with an increased risk of infection [39]. The utilization of newer devices such as subcutaneous defibrillators, leadless pacemakers and bio-absorbable antibacterial envelopes may also contribute to decreasing the incidence of infection in patients with CIED [4043].
The infections occurring after implantation of cardiac devices and caused by nontuberculous mycobacteria are rare. However, with an increased utilization of cardiac implant devices in patients, the occurrence of consequent infections is also expected to increase. Our patient presented with typical manifestations of a post-implantation CIED infection caused by RGMs, including early onset with localized implant site redness, swelling, spontaneous drainage, and without any systemic symptoms. After identifying M. fortuitum, as the causative organism, the device was completely removed at once, and the appropriate antibiotic therapy was administered. The treatment had a successful outcome as our patient was completely cured. The limitation of our case was that we prescribed antimicrobials empirically, without an initial drug susceptibility test. Furthermore, we implanted another device in the same patient after 3 years, which is an encouraging indication that re-implantation after the resolution of infection in other such patients may be possible. However, further investigation is required to determine the suitable time interval between the successful completion of curative treatment and re-implantation.

Acknowledgements

We gratefully acknowledge Dr. Yong Li for his editorial suggestions.

Funding

This study was supported in part by grants from the National Natural Science Foundation in China (No.81470496 and No.81573710).

Availability of data and materials

All the data supporting our findings is contained within the manuscript.
We identified this patient during routine clinical practice and consented to give venous blood samples after elaborate information. Involvement of the ethical committee of the Huashan Hospital of Fudan University was considered unnecessary, since the project was not based on a study protocol.
Written informed consent was obtained from the patient for publication of this case report and any accompanying images.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Leung S, Prevention DS. Diagnosis, and treatment of cardiac implantable electronic device infections. Curr Cardiol Rep. 2016;18(6):58.PubMed Leung S, Prevention DS. Diagnosis, and treatment of cardiac implantable electronic device infections. Curr Cardiol Rep. 2016;18(6):58.PubMed
2.
Zurück zum Zitat Chua J, Wilkoff B, Lee I, Juratli N, Longworth D, Gordon S. Diagnosis and management of infections involving implantable electrophysiologic cardiac devices. Ann Intern Med. 2000;133(8):604–8.PubMed Chua J, Wilkoff B, Lee I, Juratli N, Longworth D, Gordon S. Diagnosis and management of infections involving implantable electrophysiologic cardiac devices. Ann Intern Med. 2000;133(8):604–8.PubMed
3.
Zurück zum Zitat Polyzos KA, Konstantelias AA, Falagas ME. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace. 2015;17(5):767–77.PubMed Polyzos KA, Konstantelias AA, Falagas ME. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace. 2015;17(5):767–77.PubMed
4.
Zurück zum Zitat Rahman R, Saba S, Bazaz R, Gupta V, Pokrywka M, Shutt K, et al. Infection and readmission rate of cardiac implantable electronic device insertions: an observational single center study. Am J Infect Control. 2016;44(3):278–82.PubMed Rahman R, Saba S, Bazaz R, Gupta V, Pokrywka M, Shutt K, et al. Infection and readmission rate of cardiac implantable electronic device insertions: an observational single center study. Am J Infect Control. 2016;44(3):278–82.PubMed
5.
Zurück zum Zitat Daniel Z, Uslan MRS, Friedman PA, Hayes DL, Wilson WR, Steckelberg JM, Baddour LM. Frequency of permanent pacemaker or implantable cardioverter-defibrillator infection in patients with gram-negative bacteremia. Clin Infect Dis. 2006;43:731–6. Daniel Z, Uslan MRS, Friedman PA, Hayes DL, Wilson WR, Steckelberg JM, Baddour LM. Frequency of permanent pacemaker or implantable cardioverter-defibrillator infection in patients with gram-negative bacteremia. Clin Infect Dis. 2006;43:731–6.
6.
Zurück zum Zitat Johansen JB, Jorgensen OD, Moller M, Arnsbo P, Mortensen PT, Nielsen JC. Infection after pacemaker implantation: infection rates and risk factors associated with infection in a population-based cohort study of 46299 consecutive patients. Eur Heart J. 2011;32(8):991–8.PubMedPubMedCentral Johansen JB, Jorgensen OD, Moller M, Arnsbo P, Mortensen PT, Nielsen JC. Infection after pacemaker implantation: infection rates and risk factors associated with infection in a population-based cohort study of 46299 consecutive patients. Eur Heart J. 2011;32(8):991–8.PubMedPubMedCentral
7.
Zurück zum Zitat Arnold C, Chu V. Cardiovascular Implantable Electronic Device Infections. Infect Dis Clin N Am. 2018;32(4):811-25. Arnold C, Chu V. Cardiovascular Implantable Electronic Device Infections. Infect Dis Clin N Am. 2018;32(4):811-25.
8.
Zurück zum Zitat Phadke VK, DSH NDG. Patient report and review of rapidly growing mycobacterial infection after cardiac device implantation. Emer Infect Dis. 2016;22(3):389–95. Phadke VK, DSH NDG. Patient report and review of rapidly growing mycobacterial infection after cardiac device implantation. Emer Infect Dis. 2016;22(3):389–95.
9.
Zurück zum Zitat Cutay A, Horowitz H, Pooley R, Van Horn K, Wormser G. Infection of epicardial pacemaker wires due to Mycobacterium abscessus. Clin Infect Dis. 1998;26(2):520–1.PubMed Cutay A, Horowitz H, Pooley R, Van Horn K, Wormser G. Infection of epicardial pacemaker wires due to Mycobacterium abscessus. Clin Infect Dis. 1998;26(2):520–1.PubMed
10.
Zurück zum Zitat Verghese S, Mullaseri A, Padmaja P, Subhadra A, Cherian K. Pacemaker implant site infection caused by atypical mycobacteria. Indian Heart J. 1998;50(2):201–2.PubMed Verghese S, Mullaseri A, Padmaja P, Subhadra A, Cherian K. Pacemaker implant site infection caused by atypical mycobacteria. Indian Heart J. 1998;50(2):201–2.PubMed
11.
Zurück zum Zitat Kessler A, Kourtis A. Mycobacterium abscessus as a cause of pacemaker infection. Med Sci Monit. 2004;10(10):CS60–2.PubMed Kessler A, Kourtis A. Mycobacterium abscessus as a cause of pacemaker infection. Med Sci Monit. 2004;10(10):CS60–2.PubMed
12.
Zurück zum Zitat Hemmersbach-Miller M, Cardenes-Santana MA, Conde-Martel A, Bolaños-Guerra JA, Campos-Herrero MI. Cardiac device infections due to Mycobacterium fortuitum. Can J Infect Dis Med Microbiol. 2005;16(3):183–5.PubMedPubMedCentral Hemmersbach-Miller M, Cardenes-Santana MA, Conde-Martel A, Bolaños-Guerra JA, Campos-Herrero MI. Cardiac device infections due to Mycobacterium fortuitum. Can J Infect Dis Med Microbiol. 2005;16(3):183–5.PubMedPubMedCentral
13.
Zurück zum Zitat Sharma S, Tleyjeh I, Espinosa R, Costello B, Baddour L. Pacemaker infection due to Mycobacterium fortuitum. Scand J Infect Dis. 2005;37(1):66–7.PubMed Sharma S, Tleyjeh I, Espinosa R, Costello B, Baddour L. Pacemaker infection due to Mycobacterium fortuitum. Scand J Infect Dis. 2005;37(1):66–7.PubMed
14.
Zurück zum Zitat Short WR, Emery C, Bhandary M, O'Donnell JA. Misidentification of Mycobacterium peregrinum, the causal organism of a case of bacteremia and automatic implantable cardioverter defibrillator-associated infection, due to its unusual acid-fast staining characteristics. J Clin Microbiol. 2005;43(4):2015–7.PubMedPubMedCentral Short WR, Emery C, Bhandary M, O'Donnell JA. Misidentification of Mycobacterium peregrinum, the causal organism of a case of bacteremia and automatic implantable cardioverter defibrillator-associated infection, due to its unusual acid-fast staining characteristics. J Clin Microbiol. 2005;43(4):2015–7.PubMedPubMedCentral
15.
Zurück zum Zitat Pastor E, Luz Andreu A, Llombart M, Chiner E. Mycobacterium fortuitum: a rare cause of pacemaker infection. Enferm Infecc Microbiol Clin. 2006;24(2):136–7.PubMed Pastor E, Luz Andreu A, Llombart M, Chiner E. Mycobacterium fortuitum: a rare cause of pacemaker infection. Enferm Infecc Microbiol Clin. 2006;24(2):136–7.PubMed
16.
Zurück zum Zitat Toda H, Sato K, Iimori M, Yamazumi T, Furuta I, Satoh A, et al. A case of Mycobacterium goodii infection wifh isolation from blood and a pacemaker lead. J Jpn Assoc Infect Dis. 2006;80(3):262–6. Toda H, Sato K, Iimori M, Yamazumi T, Furuta I, Satoh A, et al. A case of Mycobacterium goodii infection wifh isolation from blood and a pacemaker lead. J Jpn Assoc Infect Dis. 2006;80(3):262–6.
17.
Zurück zum Zitat Giannella M, Valerio M, Franco JA, Marin M, Bouza E, Munoz P. Pacemaker infection due to Mycobacterium fortuitum: the role of universal 16S rRNA gene PCR and sequencing. Diagn Microbiol Infect Dis. 2007;57(3):337–9.PubMed Giannella M, Valerio M, Franco JA, Marin M, Bouza E, Munoz P. Pacemaker infection due to Mycobacterium fortuitum: the role of universal 16S rRNA gene PCR and sequencing. Diagn Microbiol Infect Dis. 2007;57(3):337–9.PubMed
18.
Zurück zum Zitat Simmon KE, Pounder JI, Greene JN, Walsh F, Anderson CM, Cohen S, et al. Identification of an emerging pathogen, Mycobacterium massiliense, by rpoB sequencing of clinical isolates collected in the United States. J Clin Microbiol. 2007;45(6):1978–80.PubMedPubMedCentral Simmon KE, Pounder JI, Greene JN, Walsh F, Anderson CM, Cohen S, et al. Identification of an emerging pathogen, Mycobacterium massiliense, by rpoB sequencing of clinical isolates collected in the United States. J Clin Microbiol. 2007;45(6):1978–80.PubMedPubMedCentral
19.
Zurück zum Zitat Siu CW, Cheng LC, Woo PC, Lau CP, Tse HF. A patient with relapsing pacemaker infection due to "gram-positive bacilli". Int J Cardiol. 2007;114(2):E40–1.PubMed Siu CW, Cheng LC, Woo PC, Lau CP, Tse HF. A patient with relapsing pacemaker infection due to "gram-positive bacilli". Int J Cardiol. 2007;114(2):E40–1.PubMed
20.
Zurück zum Zitat Tam W, Yew W, Yam W, Yuen K, Wong P, Tse T. Pacemaker infections due to rapidly growing mycobacteria: further experience. Int J Tuberc Lung Dis. 2007;11(1):118.PubMed Tam W, Yew W, Yam W, Yuen K, Wong P, Tse T. Pacemaker infections due to rapidly growing mycobacteria: further experience. Int J Tuberc Lung Dis. 2007;11(1):118.PubMed
21.
Zurück zum Zitat Chrissoheris M, Kadakia H, Marieb M, Libertin C. Pacemaker pocket infection due to Mycobacterium goodii: case report and review of the literature. Conn Med. 2008;72(2):75–7.PubMed Chrissoheris M, Kadakia H, Marieb M, Libertin C. Pacemaker pocket infection due to Mycobacterium goodii: case report and review of the literature. Conn Med. 2008;72(2):75–7.PubMed
22.
Zurück zum Zitat Al Soub H, Al Maslamani M, Al Khuwaiter J, El Deeb Y, Abu Khattab M. Myocardial abscess and bacteremia complicating Mycobacterium fortuitum pacemaker infection: case report and review of the literature. Pediatr Infect Dis J. 2009;28(11):1032–4.PubMed Al Soub H, Al Maslamani M, Al Khuwaiter J, El Deeb Y, Abu Khattab M. Myocardial abscess and bacteremia complicating Mycobacterium fortuitum pacemaker infection: case report and review of the literature. Pediatr Infect Dis J. 2009;28(11):1032–4.PubMed
23.
Zurück zum Zitat He ́le’ne Marchandin, Battistella, P, Brigitte Calvet,, He ́ le` ne Darbas J-MF, He ́ le` ne Jean-Pierre,Sylvie Parer,Estelle Jumas-Bilak, Philippe Van de Perre and Sylvain Godreuil. Pacemaker surgical site infection caused by Mycobacterium goodii. J Med Microbiol 2009;58:517–520. He ́le’ne Marchandin, Battistella, P, Brigitte Calvet,, He ́ le` ne Darbas J-MF, He ́ le` ne Jean-Pierre,Sylvie Parer,Estelle Jumas-Bilak, Philippe Van de Perre and Sylvain Godreuil. Pacemaker surgical site infection caused by Mycobacterium goodii. J Med Microbiol 2009;58:517–520.
24.
Zurück zum Zitat van Duin D, Goldfarb J, Schmitt SK, Tomford JW, Tuohy MJ, Hall GS. Nontuberculous mycobacterial blood stream and cardiac infections in patients without HIV infection. Diagn Microbiol Infect Dis. 2010;67(3):286–90.PubMed van Duin D, Goldfarb J, Schmitt SK, Tomford JW, Tuohy MJ, Hall GS. Nontuberculous mycobacterial blood stream and cardiac infections in patients without HIV infection. Diagn Microbiol Infect Dis. 2010;67(3):286–90.PubMed
25.
Zurück zum Zitat Karnam S, Alla V, Kwon J, Harbert T, Sharma A, Airey K, et al. Mycobacterium phlei, a previously unreported cause of pacemaker infection: thinking outside the box in cardiac device infections. Cardiol J. 2011;18(6):687–90.PubMed Karnam S, Alla V, Kwon J, Harbert T, Sharma A, Airey K, et al. Mycobacterium phlei, a previously unreported cause of pacemaker infection: thinking outside the box in cardiac device infections. Cardiol J. 2011;18(6):687–90.PubMed
26.
Zurück zum Zitat Amraoui S, Texier-Maugein J, Bordachar P. PET scan in suspected but unproven pacemaker endocarditis. Arch Cardiovasc Dis. 2012;105(2):125–6.PubMed Amraoui S, Texier-Maugein J, Bordachar P. PET scan in suspected but unproven pacemaker endocarditis. Arch Cardiovasc Dis. 2012;105(2):125–6.PubMed
27.
Zurück zum Zitat Sharma H, Keshavan A, Little MA, Cross J, Lipman MC, Talukdar S, et al. Fortuitous vasculitis. Ren Fail. 2012;34(3):378–82.PubMed Sharma H, Keshavan A, Little MA, Cross J, Lipman MC, Talukdar S, et al. Fortuitous vasculitis. Ren Fail. 2012;34(3):378–82.PubMed
29.
Zurück zum Zitat Fukunaga M, Goya M, Ogawa M, Fukuda K, Taniguchi H, Ando K, et al. Implantable cardioverter defibrillator infection due to Mycobacterium mageritense. J Infect Chemother. 2016;22(3):180–3.PubMed Fukunaga M, Goya M, Ogawa M, Fukuda K, Taniguchi H, Ando K, et al. Implantable cardioverter defibrillator infection due to Mycobacterium mageritense. J Infect Chemother. 2016;22(3):180–3.PubMed
32.
Zurück zum Zitat III JOF. The changing pattern of nontuberculous mycobacterial disease. Can J Infect Dis. 2003;14(5):281–6. III JOF. The changing pattern of nontuberculous mycobacterial disease. Can J Infect Dis. 2003;14(5):281–6.
33.
Zurück zum Zitat Brown-Elliott BA, Wallace RJ. Clinical and taxonomic status of pathogenic nonpigmented or late-pigmenting rapidly growing mycobacteria. Clin Microbiol Rev. 2002;15(4):716–46.PubMedPubMedCentral Brown-Elliott BA, Wallace RJ. Clinical and taxonomic status of pathogenic nonpigmented or late-pigmenting rapidly growing mycobacteria. Clin Microbiol Rev. 2002;15(4):716–46.PubMedPubMedCentral
34.
Zurück zum Zitat Gubler JGH, Salfinger M. Graevenitz Av. Pseudoepidemic of nontuberculous mycobacteria due to a contaminated bronchoscope cleaning machine. Chest. 1992;101(5):1245–9.PubMed Gubler JGH, Salfinger M. Graevenitz Av. Pseudoepidemic of nontuberculous mycobacteria due to a contaminated bronchoscope cleaning machine. Chest. 1992;101(5):1245–9.PubMed
35.
Zurück zum Zitat Baddour LM, Epstein AE, Erickson CC, Knight BP, Levison ME, Lockhart PB, et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation. 2010;121(3):458–77.PubMed Baddour LM, Epstein AE, Erickson CC, Knight BP, Levison ME, Lockhart PB, et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation. 2010;121(3):458–77.PubMed
36.
Zurück zum Zitat Wallace RJ, Brown-Elliott BA, Crist CJ, Mann L, Wilson RW. Comparison of the in vitro activity of the Glycylcycline Tigecycline (formerly GAR-936) with those of tetracycline, minocycline, and doxycycline against isolates of nontuberculous mycobacteria. Antimicrob Agents Chemother. 2002;46(10):3164–7.PubMed Wallace RJ, Brown-Elliott BA, Crist CJ, Mann L, Wilson RW. Comparison of the in vitro activity of the Glycylcycline Tigecycline (formerly GAR-936) with those of tetracycline, minocycline, and doxycycline against isolates of nontuberculous mycobacteria. Antimicrob Agents Chemother. 2002;46(10):3164–7.PubMed
37.
Zurück zum Zitat Wallace RJ Jr, Brown-Elliott BA, Ward SC, Crist CJ, Mann LB, Wilson RW. Activities of linezolid against rapidly growing mycobacteria. Antimicrob Agents Chemother. 2001;45(3):764–7.PubMedPubMedCentral Wallace RJ Jr, Brown-Elliott BA, Ward SC, Crist CJ, Mann LB, Wilson RW. Activities of linezolid against rapidly growing mycobacteria. Antimicrob Agents Chemother. 2001;45(3):764–7.PubMedPubMedCentral
38.
Zurück zum Zitat Sohail MR, Uslan DZ, Khan AH, Friedman PA, Hayes DL, Wilson WR, et al. Risk factor analysis of permanent pacemaker infection. Clin Infect Dis. 2007;45(2):166–73.PubMed Sohail MR, Uslan DZ, Khan AH, Friedman PA, Hayes DL, Wilson WR, et al. Risk factor analysis of permanent pacemaker infection. Clin Infect Dis. 2007;45(2):166–73.PubMed
39.
Zurück zum Zitat Klug D, Balde M, Pavin D, Hidden-Lucet F, Clementy J, Sadoul N, et al. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation. 2007;116(12):1349–55.PubMed Klug D, Balde M, Pavin D, Hidden-Lucet F, Clementy J, Sadoul N, et al. Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study. Circulation. 2007;116(12):1349–55.PubMed
40.
Zurück zum Zitat Weiss R, Knight B, Gold M, Leon A, Herre J, Hood M, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. 2013;128(9):944–53.PubMed Weiss R, Knight B, Gold M, Leon A, Herre J, Hood M, et al. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator. Circulation. 2013;128(9):944–53.PubMed
41.
Zurück zum Zitat Mittal S, Shaw R, Michel K, Palekar R, Arshad A, Musat D, et al. Cardiac implantable electronic device infections: incidence, risk factors, and the effect of the AigisRx antibacterial envelope. Heart Rhythm. 2014;11(4):595–601.PubMed Mittal S, Shaw R, Michel K, Palekar R, Arshad A, Musat D, et al. Cardiac implantable electronic device infections: incidence, risk factors, and the effect of the AigisRx antibacterial envelope. Heart Rhythm. 2014;11(4):595–601.PubMed
42.
Zurück zum Zitat Reddy VY, Exner DV, Cantillon DJ, Doshi R, Bunch TJ, Tomassoni GF, et al. Percutaneous implantation of an entirely Intracardiac leadless pacemaker. N Engl J Med. 2015;373(12):1125–35.PubMed Reddy VY, Exner DV, Cantillon DJ, Doshi R, Bunch TJ, Tomassoni GF, et al. Percutaneous implantation of an entirely Intracardiac leadless pacemaker. N Engl J Med. 2015;373(12):1125–35.PubMed
43.
Zurück zum Zitat Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al. A leadless Intracardiac Transcatheter pacing system. N Engl J Med. 2016;374(6):533–41.PubMed Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al. A leadless Intracardiac Transcatheter pacing system. N Engl J Med. 2016;374(6):533–41.PubMed
Metadaten
Titel
Cardiac resynchronization therapy-defibrillator pocket infection caused by Mycobacterium fortuitum: a case report and review of the literature
verfasst von
Jun Zhu
Qingluan Yang
Junjie Pan
Haiming Shi
Bo Jin
Qiying Chen
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
BMC Cardiovascular Disorders / Ausgabe 1/2019
Elektronische ISSN: 1471-2261
DOI
https://doi.org/10.1186/s12872-019-1028-0

Weitere Artikel der Ausgabe 1/2019

BMC Cardiovascular Disorders 1/2019 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.