11.02.2020 | Original Article

Chest CT screening in patients with overweight or obesity using spectral shaping at 150 kVp: compared with 120 kVp protocol and spectral shaping at 100 kVp protocol

verfasst von: Jianwen Li, Zhifeng Mai, Zhihong Zhang, Jiamang Cui, Mingjie Yang, Xia Ma, Yan Wang

Erschienen in: Japanese Journal of Radiology | Ausgabe 5/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the image quality (IQ) and the figure of merit (FoM) of chest CT screening in patients with overweight or obesity using a tin filter for spectral shaping at 150 kVp.

Materials and methods

Patients with overweight or obesity (N = 150, body mass index ≥ 26 kg/m2) with indications for chest CT screening were prospectively enrolled and randomly divided into three groups: 120 kVp group (standard radiation dose/tube voltage, 120 kVp/CT volume does index, 4.68 mGy); Sn100 kVp group (1/10th radiation dose level/100 kVp with a tin filter/0.47 mGy); Sn150 kVp group (1/2th radiation dose level/150 kVp with a tin filter/2.34 mGy). IQ and FoMs were evaluated and compared among the three groups.

Results

Image noise, signal-to-noise ratios and subjective IQ scores were significantly higher in the Sn150 kVp group than those in the Sn100 kVp group (all p < 0.05), but were not significantly different with those in the 120 kVp group. FoMs in the Sn150 kVp group were significantly higher than those in the 120 kVp group (all p < 0.05), but showed no statistical difference with those in the Sn100 kVp group.

Conclusions

Compared with scanning at 120 kVp, chest CT screening performed at 150 kVp with spectral shaping substantially reduces the radiation dose in overweight and obese patients while maintaining IQ.
Literatur
1.
Zurück zum Zitat Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2018;68:394–424.CrossRef Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2018;68:394–424.CrossRef
2.
Zurück zum Zitat Gatsonis CA, Aberle DR, Berg CD, Black WC, Church TR, Fagerstrom RM, et al. The national lung screening trial: overview and study design. Radiology. 2011;258:243–53.CrossRef Gatsonis CA, Aberle DR, Berg CD, Black WC, Church TR, Fagerstrom RM, et al. The national lung screening trial: overview and study design. Radiology. 2011;258:243–53.CrossRef
3.
Zurück zum Zitat De González AB, Darby S. Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet. 2004;363:345–51.CrossRef De González AB, Darby S. Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet. 2004;363:345–51.CrossRef
4.
Zurück zum Zitat Kalra MK, Maher MM, Toth TL, Kamath RS, Halpern EF, Saini S. Comparison of z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis. Radiology. 2004;232:347–53.CrossRef Kalra MK, Maher MM, Toth TL, Kamath RS, Halpern EF, Saini S. Comparison of z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis. Radiology. 2004;232:347–53.CrossRef
5.
Zurück zum Zitat Liu X, Zhang M, Zhang L, Zhao Y, Sun W. Radiation dose reduction using two orthogonal topograms associated with automatic tube voltage selection for lung CT scanning as compared with a single anteroposterior topogram. Jpn J Radiol. 2019;37:292–300.CrossRef Liu X, Zhang M, Zhang L, Zhao Y, Sun W. Radiation dose reduction using two orthogonal topograms associated with automatic tube voltage selection for lung CT scanning as compared with a single anteroposterior topogram. Jpn J Radiol. 2019;37:292–300.CrossRef
6.
Zurück zum Zitat Shi JW, Xu DF, Dai HZ, Shen L, Ji YD. Evaluation of chest CT scan in low-weight children with ultralow tube voltage (70 kVp) combined with Flash scan technique. Br J Radiol. 2016;89:20150184.CrossRef Shi JW, Xu DF, Dai HZ, Shen L, Ji YD. Evaluation of chest CT scan in low-weight children with ultralow tube voltage (70 kVp) combined with Flash scan technique. Br J Radiol. 2016;89:20150184.CrossRef
7.
Zurück zum Zitat Hagelstein C, Henzler T, Haubenreisser H, Meyer M, Sudarski S, Schoenberg SO, et al. Ultra-high-pitch-Thorax-Computertomographie bei 70 kVp Röhrenspannung in einem anthropomorphen Kinderphantom und nicht sedierten pädiatrischen Patienten: Erste Erfahrungen an einem Dual-Source-CT der 3.Generation. Z Med Phys. 2016;26:349–61.CrossRef Hagelstein C, Henzler T, Haubenreisser H, Meyer M, Sudarski S, Schoenberg SO, et al. Ultra-high-pitch-Thorax-Computertomographie bei 70 kVp Röhrenspannung in einem anthropomorphen Kinderphantom und nicht sedierten pädiatrischen Patienten: Erste Erfahrungen an einem Dual-Source-CT der 3.Generation. Z Med Phys. 2016;26:349–61.CrossRef
8.
Zurück zum Zitat Bodelle B, Fischbach C, Booz C, Yel I, Frellesen C, Beeres M, et al. Free-breathing high-pitch 80 kVp dual-source computed tomography of the pediatric chest: image quality, presence of motion artifacts and radiation dose. Eur J Radiol. 2017;89:208–14.CrossRef Bodelle B, Fischbach C, Booz C, Yel I, Frellesen C, Beeres M, et al. Free-breathing high-pitch 80 kVp dual-source computed tomography of the pediatric chest: image quality, presence of motion artifacts and radiation dose. Eur J Radiol. 2017;89:208–14.CrossRef
9.
Zurück zum Zitat Baumueller S, Winklehner A, Karlo C, Goetti R, Flohr T, Russi EW, et al. Low-dose CT of the lung: potential value of iterative reconstructions. Eur Radiol. 2012;22:2597–606.CrossRef Baumueller S, Winklehner A, Karlo C, Goetti R, Flohr T, Russi EW, et al. Low-dose CT of the lung: potential value of iterative reconstructions. Eur Radiol. 2012;22:2597–606.CrossRef
10.
Zurück zum Zitat Prakash P, Kalra MK, Ackman JB, Digumarthy SR, Hsieh J, Do S, et al. Diffuse lung disease: CT of the chest with adaptive statistical iterative reconstruction technique. Radiology. 2010;256:261–9.CrossRef Prakash P, Kalra MK, Ackman JB, Digumarthy SR, Hsieh J, Do S, et al. Diffuse lung disease: CT of the chest with adaptive statistical iterative reconstruction technique. Radiology. 2010;256:261–9.CrossRef
11.
Zurück zum Zitat Weis M, Henzler T, Nance JW, Haubenreisser H, Meyer M, Sudarski S, et al. Radiation dose comparison between 70 kVp and 100 kVp with spectral beam shaping for non-contrast-enhanced pediatric chest computed tomography. Invest Radiol. 2017;52:155–62.CrossRef Weis M, Henzler T, Nance JW, Haubenreisser H, Meyer M, Sudarski S, et al. Radiation dose comparison between 70 kVp and 100 kVp with spectral beam shaping for non-contrast-enhanced pediatric chest computed tomography. Invest Radiol. 2017;52:155–62.CrossRef
12.
Zurück zum Zitat Messerli M, Kluckert T, Knitel M, Wälti S, Desbiolles L, Rengier F, et al. Ultralow dose CT for pulmonary nodule detection with chest X-ray equivalent dose—a prospective intra-individual comparative study. Eur Radiol. 2017;27:3290–9.CrossRef Messerli M, Kluckert T, Knitel M, Wälti S, Desbiolles L, Rengier F, et al. Ultralow dose CT for pulmonary nodule detection with chest X-ray equivalent dose—a prospective intra-individual comparative study. Eur Radiol. 2017;27:3290–9.CrossRef
13.
Zurück zum Zitat Haubenreisser H, Meyer M, Sudarski S, Allmendinger T, Schoenberg SO, Henzler T. Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100 kVp. Eur J Radiol. 2015;84:1608–13.CrossRef Haubenreisser H, Meyer M, Sudarski S, Allmendinger T, Schoenberg SO, Henzler T. Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100 kVp. Eur J Radiol. 2015;84:1608–13.CrossRef
14.
Zurück zum Zitat Gordic S, Morsbach F, Schmidt B, Allmendinger T, Flohr T, Husarik D, et al. Ultralow-dose chest computed tomography for pulmonary nodule detection: first performance evaluation of single energy scanning with spectral shaping. Invest Radiol. 2014;49:465–73.CrossRef Gordic S, Morsbach F, Schmidt B, Allmendinger T, Flohr T, Husarik D, et al. Ultralow-dose chest computed tomography for pulmonary nodule detection: first performance evaluation of single energy scanning with spectral shaping. Invest Radiol. 2014;49:465–73.CrossRef
15.
Zurück zum Zitat Primak AN, Ramirez Giraldo JC, Liu X, Yu L, McCollough CH. Improved dual-energy material discrimination for dual-source CT by means of additional spectral filtration. Med Phys. 2009;36:1359–69.CrossRef Primak AN, Ramirez Giraldo JC, Liu X, Yu L, McCollough CH. Improved dual-energy material discrimination for dual-source CT by means of additional spectral filtration. Med Phys. 2009;36:1359–69.CrossRef
16.
Zurück zum Zitat McCollough CH, Primak AN, Braun N, Kofler J, Yu L, Christner J. Strategies for Reducing Radiation Dose in CT. Radiol Clin N Am. 2009;47:27–40.CrossRef McCollough CH, Primak AN, Braun N, Kofler J, Yu L, Christner J. Strategies for Reducing Radiation Dose in CT. Radiol Clin N Am. 2009;47:27–40.CrossRef
17.
Zurück zum Zitat Deak PD, Smal Y, Kalender WA. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology. 2010;257:158–66.CrossRef Deak PD, Smal Y, Kalender WA. Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology. 2010;257:158–66.CrossRef
18.
Zurück zum Zitat Bone J, Strauss K, Cody D. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Report of AAPM Task Group 204. Bone J, Strauss K, Cody D. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Report of AAPM Task Group 204.
19.
Zurück zum Zitat Nakaura T, Awai K, Maruyama N, Takata N, Yoshinaka I, Harada K, et al. Abdominal dynamic CT in patients with renal dysfunction: contrast agent dose reduction with low tube voltage and high tube current-time product settings at 256-detector row CT. Radiology. 2011;261:467–76.CrossRef Nakaura T, Awai K, Maruyama N, Takata N, Yoshinaka I, Harada K, et al. Abdominal dynamic CT in patients with renal dysfunction: contrast agent dose reduction with low tube voltage and high tube current-time product settings at 256-detector row CT. Radiology. 2011;261:467–76.CrossRef
20.
Zurück zum Zitat Martini K, Barth BK, Higashigaito K, Baumueller S, Alkadhi H, Frauenfelder T. Dose-optimized computed tomography for screening and follow-up of solid pulmonary nodules in obesity: a phantom study. Curr Probl Diagn Radiol. 2017;46:204–9.CrossRef Martini K, Barth BK, Higashigaito K, Baumueller S, Alkadhi H, Frauenfelder T. Dose-optimized computed tomography for screening and follow-up of solid pulmonary nodules in obesity: a phantom study. Curr Probl Diagn Radiol. 2017;46:204–9.CrossRef
22.
Zurück zum Zitat Uppot RN, Sahani DV, Hahn PF, Gervais D, Mueller PR. Impact of obesity on medical imaging and image-guided intervention. Am. J. Roentgenol. 2007;188:433–40.CrossRef Uppot RN, Sahani DV, Hahn PF, Gervais D, Mueller PR. Impact of obesity on medical imaging and image-guided intervention. Am. J. Roentgenol. 2007;188:433–40.CrossRef
23.
Zurück zum Zitat Schindera ST, Nelson RC, Toth TL, Nguyen GT, Toncheva GI, DeLong DM, et al. Effect of patient size on radiation dose for abdominal MDCT with automatic tube current modulation: phantom study. AJR Am J Roentgenol. 2008;190:w100–w105105.CrossRef Schindera ST, Nelson RC, Toth TL, Nguyen GT, Toncheva GI, DeLong DM, et al. Effect of patient size on radiation dose for abdominal MDCT with automatic tube current modulation: phantom study. AJR Am J Roentgenol. 2008;190:w100–w105105.CrossRef
24.
Zurück zum Zitat Wang AJ, Goldsmith ZG, Wang C, Nguyen G, Astroza GM, Neisius A, et al. Obesity triples the radiation dose of stone protocol computerized tomography. J Urol. 2013;189:2142–6.CrossRef Wang AJ, Goldsmith ZG, Wang C, Nguyen G, Astroza GM, Neisius A, et al. Obesity triples the radiation dose of stone protocol computerized tomography. J Urol. 2013;189:2142–6.CrossRef
25.
Zurück zum Zitat Newell JD, Fuld MK, Allmendinger T, Sieren JP, Chan KS, Guo J, et al. Very low-dose (0.15 mGy) chest CT protocols using the COPDGene 2 test object and a third-generation dual-source CT scanner with corresponding third-generation iterative reconstruction software. Invest Radiol. 2015;50:40–5.CrossRef Newell JD, Fuld MK, Allmendinger T, Sieren JP, Chan KS, Guo J, et al. Very low-dose (0.15 mGy) chest CT protocols using the COPDGene 2 test object and a third-generation dual-source CT scanner with corresponding third-generation iterative reconstruction software. Invest Radiol. 2015;50:40–5.CrossRef
26.
Zurück zum Zitat Wuest W, May M, Saake M, Brand M, Uder M, Lell M. Low-dose CT of the paranasal sinuses: minimizing X-ray exposure with spectral shaping. Eur Radiol. 2016;26:4155–61.CrossRef Wuest W, May M, Saake M, Brand M, Uder M, Lell M. Low-dose CT of the paranasal sinuses: minimizing X-ray exposure with spectral shaping. Eur Radiol. 2016;26:4155–61.CrossRef
27.
Zurück zum Zitat Li C, Liu B, Meng H, Lv W, Jia H. Efficacy and radiation exposure of ultra-low-dose chest CT at 100 kVp with tin filtration in CT-guided percutaneous core needle biopsy for small pulmonary lesions using a third-generation dual-source CT scanner. J Vasc Interv Radiol. 2019;30:95–102.CrossRef Li C, Liu B, Meng H, Lv W, Jia H. Efficacy and radiation exposure of ultra-low-dose chest CT at 100 kVp with tin filtration in CT-guided percutaneous core needle biopsy for small pulmonary lesions using a third-generation dual-source CT scanner. J Vasc Interv Radiol. 2019;30:95–102.CrossRef
28.
Zurück zum Zitat Zhang GMY, Shi B, Sun H, Xue HD, Wang Y, Liang JX, et al. High-pitch low-dose abdominopelvic CT with tin-filtration technique for detecting urinary stones. Abdom Radiol. 2017;42:2127–34.CrossRef Zhang GMY, Shi B, Sun H, Xue HD, Wang Y, Liang JX, et al. High-pitch low-dose abdominopelvic CT with tin-filtration technique for detecting urinary stones. Abdom Radiol. 2017;42:2127–34.CrossRef
29.
Zurück zum Zitat Dewes P, Frellesen C, Scholtz JE, Fischer S, Vogl TJ, Bauer RW, et al. Low-dose abdominal computed tomography for detection of urinary stone disease—impact of additional spectral shaping of the X-ray beam on image quality and dose parameters. Eur J Radiol. 2016;85:1058–62.CrossRef Dewes P, Frellesen C, Scholtz JE, Fischer S, Vogl TJ, Bauer RW, et al. Low-dose abdominal computed tomography for detection of urinary stone disease—impact of additional spectral shaping of the X-ray beam on image quality and dose parameters. Eur J Radiol. 2016;85:1058–62.CrossRef
30.
Zurück zum Zitat Su AW, Hillen TJ, Eutsler EP, Bedi A, Ross JR, Larson CM, et al. Low-dose computed tomography reduces radiation exposure by 90% compared with traditional computed tomography among patients undergoing hip-preservation surgery. Arthrosc J Arthrosc Relat Surg. 2019;35:1385–92.CrossRef Su AW, Hillen TJ, Eutsler EP, Bedi A, Ross JR, Larson CM, et al. Low-dose computed tomography reduces radiation exposure by 90% compared with traditional computed tomography among patients undergoing hip-preservation surgery. Arthrosc J Arthrosc Relat Surg. 2019;35:1385–92.CrossRef
31.
Zurück zum Zitat Hell MM, Bittner D, Schuhbaeck A, Muschiol G, Brand M, Lell M, et al. Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: feasibility, image quality, radiation dose, and effect of iterative reconstruction. J Cardiovasc Comput Tomogr. 2014;8:418–25.CrossRef Hell MM, Bittner D, Schuhbaeck A, Muschiol G, Brand M, Lell M, et al. Prospectively ECG-triggered high-pitch coronary angiography with third-generation dual-source CT at 70 kVp tube voltage: feasibility, image quality, radiation dose, and effect of iterative reconstruction. J Cardiovasc Comput Tomogr. 2014;8:418–25.CrossRef
Metadaten
Titel
Chest CT screening in patients with overweight or obesity using spectral shaping at 150 kVp: compared with 120 kVp protocol and spectral shaping at 100 kVp protocol
verfasst von
Jianwen Li
Zhifeng Mai
Zhihong Zhang
Jiamang Cui
Mingjie Yang
Xia Ma
Yan Wang
Publikationsdatum
11.02.2020
Verlag
Springer Singapore
Erschienen in
Japanese Journal of Radiology / Ausgabe 5/2020
Print ISSN: 1867-1071
Elektronische ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-020-00925-5

Weitere Artikel der Ausgabe 5/2020

Japanese Journal of Radiology 5/2020 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.