Classical experiments in whole-body metabolism: open-circuit respirometry—diluted flow chamber, hood, or facemask systems

verfasst von: P. F. M. Schoffelen, G. Plasqui

Erschienen in: European Journal of Applied Physiology | Ausgabe 1/2018

Abstract

For over two centuries, scientists have measured gas exchange in animals and humans and linked this to energy expenditure of the body. The aim of this review is to provide a comprehensive overview of open-circuit diluted flow indirect calorimetry and to help researchers to make the optimal choice for a certain system and its application. A historical perspective shows that ‘open circuit diluted flow’ is a technique first used in the 19th century and applicable today for room calorimeters, ventilated hood systems, and facemasks. Room calorimeters are a classic example of an open-circuit diluted flow system. The broadly applied ventilated hood calorimeters follow the same principle and can be classified as a derivative of these room calorimeters. The basic principle is that the subject breathes freely in a passing airflow that is fully captured and analyzed. Oxygen and CO2 concentrations are measured in inlet ambient air and captured outlet air. The airflow, which is adapted depending on the application (e.g., rest versus exercise), is measured. For a room indirect calorimeter, the dilution in the large room volume is also taken into account, and this is the most complex application of this type of calorimeter. Validity of the systems can be tested by alcohol burns, gas infusions and by performing repeated measurements on subjects. Using the latter, the smallest CV (%) was found for repeated VO2max tests (1.2%) with an SD of approximately 1 kJ min−1. The smallest SD was found for sleeping metabolic rate (0.11 kJ min−1) with a CV (%) of 2.4%.
Hinweise
Communicated by Michael Lindinger.
Abkürzungen
AEE
Activity-related energy expenditure
BMR
Basal metabolic rate
DIT
Diet-induced thermogenesis
EE
Energy expenditure
SMR
Sleeping metabolic rate

General introduction

Indirect calorimetry is the indirect determination of total energy expenditure (EE) of the human body by measuring gas exchange, i.e., O2 consumption and CO2 production, to derive estimates of quantities of substrates oxidized. The combination of known quantity and energy density of substrates then reveal the total amount of EE. In that regard, ‘indirect calorimetry’ is essentially different from ‘direct calorimetry’, which measures the heat dissipated by the body directly. Several aspects of food composition, oxidation, energy released as heat and work and gas exchange had to be discovered and combined before indirect calorimetry became a practical solution for the measurement of human EE.
Even though most current systems are based on indirect calorimetry, one could state that direct calorimetry is in fact the gold standard for measuring EE as no “indirect” conversion factors are needed to calculate EE values from chemical energy stored in substrates. It is important to understand that energy has many forms, and it can be stored, transferred, and converted in at least as many ways. A few examples are energy in the form of heat, radiation (light), and movement. Energy can be transferred and converted from one form into another, which, for instance, will involve chemical reactions or mechanical work. Hence, a second essential difference between direct and indirect calorimetry is that while direct calorimetry measures heat dissipated by the body, indirect calorimetry measures total EE, and this total energy is not always entirely released as heat from the body. When performing exercise, part of the total EE of the body is used to perform this ‘external work’ and is not released as heat. A good example to illustrate this difference is the paper by Webb on “The work of walking: a calorimetric study” (Webb et al. 1988). During walking on a treadmill, subjects’ EE was measured directly using a suit calorimeter and indirectly using gas exchange. It was shown that during walking, part of the energy used by the body was not released as heat, but that ‘work’ was being done. Please note that work converted to heat inside of a direct calorimeter will be included as measured energy, and energy from external sources (lights, fans, airco, tv, computer) must be excluded. The same holds true for indirect calorimetry regarding changes in gas composition not originating from the body (carbonated drinks, burning cigarettes, and flushing toilet).
This review will focus on indirect calorimetry and, more specifically open-circuit, diluted flow systems.

Historical perspective

As early as 360 b.c., Plato described ‘breathing’ in his work “Timaeus”, later discussed by his student Aristotle in “On Youth and Old Age, On Life and Death, On Breathing”, herein combining the terms ‘internal fire’ and ‘breathing’ (Aristotle and Ogle 1897). The mysterious relation of food, energy, and breathing caused Aristotle to write “Moreover, in what sense are we to understand this fantastical notion, that heat is generated out of the breath? For it is out of food rather than out of the air that we see heat developed” (Aristotle and Ogle 1897). This sentence comes as close to describing the basis for both direct and indirect calorimetry as was possible at the time.
In 1614, Santorio Sanctorius found a difference in the weight of all foods consumed and excrements produced (Santorio 1614). He referred to this, at that time unexplained difference, as “insensible perspiration”. In essence, Sanctorius had discovered the weight of carbon leaving the body as CO2, but could not explain it at the time (Santorio 1614).
A next step occurred in 1774 when Priestly found that oxygen in the air was not an unchangeable constant (Priestley 1774), and in the years 1774–1777, discoveries in the field of the respiration of humans and animals by Antoine-Laurent Lavoisier and his wife and coworker followed (Lavoisier 1777). They were able to measure oxygen consumption and carbon dioxide production, yet without knowledge of substrate oxidation and its conversion to EE (Fig. 1) (Grimaux 1888).
Just a few years later, in 1780, Lavoisier and Laplace (Lavoisier 1783) published details on the use of the famous “ice-calorimeter”, enclosing an animal in a sealed and insulated space cooled with ice to reveal its heat production (direct calorimetry) from the quantity of meltwater produced. Thus, Lavoisier probably built both a first direct calorimeter as well as a first indirect calorimeter, even though gas-exchange results of the latter could not yet be converted to EE. Lavoisier concluded “La respiration n’est qu’une combustion lente de carbone et d’hydrogène, qui est semblable en tout á celle qui s’opère dans une lampe ou dans une bougie allumée, et que, sous ce point de vue, les animaux qui respirent sont de véritables corps combustibles qui brûlent et se consument.” or “Breathing is but a slow combustion of carbon and hydrogen, which is similar in all respects to that which takes place in a lamp or in a lighted candle and that from this point of view, animals that breathe are real combustible bodies that burn and consume”, establishing that heat produced by a living being is comparable to the chemical reaction of burning food in fire, though at a slower pace (Fig. 1 ) (Lavoisier 1783).
In the late 19th century, nearly, a 100 years after Lavoisier, the scientific know-how of physics, chemistry, and food composition had progressed far enough to build the first whole-room open-circuit indirect calorimeter, as described by Pettenkofer in 1862 (Fig. 2) (Pettenkofer 1862).
Towards the turn of the century, details on substrate oxidation as published by Magnus-Levy (1893) and Zuntz and Schumburg (1901b); Zuntz and Lehmann (1890) became available. Haldane in 1862 described his apparatus and methods for measuring gas-exchange postulating “it being assumed that no weighable quantities of nitrogen, or any other gaseous or volatile substances, are given off from, or absorbed by, the animal during the experiment” (Haldane 1892). This important postulate regarding conservation of inert gasses in-, or passing through, a calorimeter is known as the “Haldane transformation” or “Haldane correction”. The inert gasses in fresh air, mostly Nitrogen (78%) and small amounts of noble gasses like Argon and Helium (0.9%), are not used in the body. Usually, oxygen consumption is not equal to carbon dioxide production (unless RER = 1), i.e., breathing volume in is not equal to breathing volume out. Haldane states that for inert gasses, volume in remains equal to volume out, and this can be used for calculating changes in breathing volume out from in or vice versa.
Thus, the Haldane correction allows to achieve near-perfect balance regarding in-flow and out-flow, which is still used in indirect calorimeters to this day. Haldane (Haldane 1898) in 1898 also meticulously described a method for chemical analysis of gas composition, and his method as well as Scholander’s from 1947 (Scholander 1947) and modifications were still in use in the 1980s (Fig. 3 ) (McLean and Watts 1976; Webb et al. 1988; Schoffelen 1985).
In 1904, Atwater augmented a whole-room direct calorimeter with closed-circuit gas-exchange measurements (Atwater and Benedict 1905). This combination of direct- and closed-circuit indirect calorimetry “was a landmark in human calorimetry” as Mclean and Tobin remarked (McLean and Tobin 1987). Many more pioneers of the era exist.
Well into the 19th century, scientists were able to start larger scale research into the relation of work, heat, and respiration in both animals and humans (Pettenkofer 1862; Ott and Foster 1891; Haldane 1892; Atwater et al. 1897; Marcet et al. 1898; Zuntz and Schumburg 1901b; Atwater and Benedict 1905; Hagemann 1911; Langworthy and Milner 1911; Hill and Hill 1914). Importantly, the composition of food from three basic substrates fat, protein, and carbohydrates was analyzed and their chemical reaction with oxygen resulting in carbon dioxide and energy was determined (Zuntz and Schumburg 1901b; Magnus-Levy 1893) (Table 1). Note that variation in substrate composition exists (Livesey and Elia 1988).
Table 1
O2 consumption (l), CO2 production (l), and EE (kcal) per gram of substrate used (i.e., carbohydrate, protein, and fat), including correction of EE (kcal) per gram of urinary nitrogen katabolized
Substrate 1 g
O2 uptake (l)
CO2 production (l)
EE (kcal)
EE (kcal)
Carbohydrate
0.8288
0.8288
4.182
Zuntz 1897 (Zuntz and Schumburg 1901b)
Protein
0.967
0.7752
4.316
Loewy-Lusk 1928 (Lusk 1928)
N2-urine katabolized
− 2.17
Weir (1949)
Fat
2.0193
1.4311
9.461
Cathcart 1931 (Cathcart and Cuthbertson 1931)
This allowed calculation of human EE from gas exchange including the variable contribution of carbon dioxide production and the correction for energy loss of N2 compounds in urine (Table 1) (Schoffelen and Plasqui 2016; Livesey and Elia 1988; Consolazio et al. 1963; Brouwer 1957; Weir 1949; Zuntz and Schumburg 1901b). One of the resulting formulae for converting gas-exchange values into human EE was written as
$${\text{EE }}\left[ {{\text{kcal}}} \right]{\text{ }}={\text{ 3}}.{\text{941 }} \cdot {\text{ }}{{\text{O}}_{\text{2}}}{\text{ }}\left[ {\text{l}} \right]{\text{ }}+{\text{ 1}}.{\text{1}}0{\text{6 }} \cdot {\text{ C}}{{\text{O}}_{\text{2}}}{\text{ }}\left[ {\text{l}} \right]{\text{ }} - {\text{2}}.{\text{17 }} \cdot {\text{ }}{{\text{N}}_{\text{2}}}{\text{ }}\left[ {\text{g}} \right].$$
(Weir 1949)
This calculation of EE from gas exchange does not involve measuring heat. It is completely based on measured gas quantities for oxygen consumed and carbon dioxide produced. In the absence of any “direct” measurement of heat, this method based on gas exchange became known as “indirect calorimetry”.
For most of the 20th century, measuring human EE with direct- and indirect calorimeters remained a demanding effort. Computers and electronic equipment were not available or in early development stages. Analysis of gasses required specialized technicians performing chemical procedures to determine values for a single gas sample at a time (Scholander 1947; Haldane 1898), while sensors were manually logged or plotted using chart recorders.
These efforts of reputable research groups resulted in truly excellent work, for the greater part still applicable today, having provided validated methods for heat- (direct) and gas-exchange (indirect) measurements, measured caloric values of the macro-nutrients protein, carbohydrates and fat, as well as formulae for EE calculations from gas-exchange and measured urinary nitrogen (Harris et al. 1919; Lusk 1917; Zuntz and Schumburg 1901b; Haldane 1892; Pettenkofer 1862; Webb 1985).
In the 1960s, developments in electronics started a revolution in equipment, and this also resulted in the first mini- and micro-computers of the 1970s. Research groups focussing on Human EE in Cambridge (Dauncey and Murgatroyd 1978) and Lausanne (Jequier et al. 1978) were among the first to take advantage of these developments, followed within a decade by many others (Seale et al. 1991; Hill and Sun 1991; Rumpler et al. 1990; Charbonnier et al. 1990; Frappell et al. 1989; Webb et al. 1980, 1988; Shetty et al. 1987; de Boer et al. 1986, 1987; Ravussin et al. 1986; Garrow and Webster 1986; Schoffelen 1985; Garby and Lammert 1984; Jequier and Schutz 1983; Aulick et al. 1983; Schutz et al. 1982). This increase in indirect calorimetry accelerated with the advent of commercial metabolic carts, bringing ventilated hood diluted flow indirect calorimetry to the bed side. This proliferation as well as technical developments did not change the basic calorimetric principles applied, yet the field of human indirect calorimetry was vastly expanded.
In recent literature, no operational larger whole-room calorimeters with both direct- and indirect calorimetry could be found; they seem all to have been decommissioned, and the last operational whole-body direct calorimeter (Kenny et al. 2017) may be the re-engineered Snellen air-flow calorimeter in Ottawa, Canada (Reardon et al. 2006). In contrast, whole-room calorimeters with indirect calorimetry have in part remained operational and in part were newly constructed. An estimate of operational whole-room indirect calorimeters is about 12–15 from literature and 10–15 more known to be in existence, effectively steady 20–30 sites worldwide.
A trend in whole-room indirect calorimetry that as yet has to realize its full potential and become a common application is the search for fast dynamic response, providing a high degree of short interval and full-range accuracy. Whole-room indirect calorimeters have a reputation for slow response, as accuracy is dependent on the multiplication of accuracy of gas analysis with a large volume and thus problematic for very short interval evaluation. Technical advances and computers have resulted in a typical 15–30-min interval evaluation (Webb et al. 1988; Nguyen et al. 2003; Henning et al. 1996; Tokuyama et al. 2009), where accuracy for low-level EE remains dependent on the minimum floor-level noise of the individual implementation (Murgatroyd et al. 1987). The individual implementation may cause unfiltered noise levels in gas-exchange values varying from 50 to 1000 ml min−1 peak to peak (personal observation and communication), and the search for fast dynamic response has mostly been focussed on mathematical procedures for re-evaluation of measured results, i.e., noise reduction (Tokuyama et al. 2009; Granato et al. 2004).
It is surmized that a next step in indirect calorimetry will combine improved measured accuracy for gas analysis and volumetric aspects, with mathematical handling of the smallest finite-step evaluation. To achieve this next step, the classic focus on physics must be combined with modern day computer technology, i.e., attention to input (measurements) must match the current attention to mathematics.

Types of open circuit indirect calorimeters

In principle, there are different basic types of indirect calorimeters to choose from recognizable in the way that subjects are connected and how breath is collected and analyzed. Commonly during exercise, and sometimes during rest, a fixed connection to the face is used (Fig. 4d, f, g). This can be either a facemask (Fig. 4d, f) or a mouthpiece with nose clip (Fig. 4g). In this paper, the term facemask implies the possible application of a mouthpiece unless noted otherwise. With a facemask, three different ways of capturing exhaled air can be defined: (1) full capture of non-diluted exhaled breath (Fig. 4g) or (2) full capture of exhaled air, diluted in a bypass flow (Fig. 4d) or (3) breath-by-breath (BxB) analysis for a partial sample of exhaled air (Fig. 4f). For measurements of basal (or resting) metabolic rate, usually, a ventilated hood system is used (Fig. 4c). A small confined space in the form of a clear-plastic hood is placed over the subject’s head, allowing full capture of exhaled air diluted in a bypass flow drawn through the hood. For measurements of EE over longer intervals, a whole-room calorimeter also referred to as respiration chamber is used (Fig. 4b). In this case, exhaled air is fully captured and diluted in both a bypass flow and a large volume (room). Hence, the use of full capture of exhaled air in a diluted flow is applicable to facemasks, ventilated hoods, and respiration chambers (Fig. 4b–d).
With the theoretical exception of breath-by-breath systems, most indirect calorimeters will measure over a longer “aggregation” interval, i.e., over multiple breaths. This is typical for full capture of exhaled air, rendering assumptions for synchronization of breathing and gas analysis unnecessary. This may be considered an advantage due to the fact that synchronization is not easily proven or validated, specifically for the higher breathing frequencies (Proctor and Beck 1996; Yamamoto et al. 1987; Noguchi et al. 1982; Wessel et al. 1979; Perret and Mueller 2006; Larsson et al. 2004; Prieur et al. 2003). A disadvantage is that measurement of a number of breaths over an aggregation interval in a small volume may have an uncertainty of ± 0.5 breaths, unless breathing is measured for synchronization. Hence, slow breathing frequencies at rest may result in visible peaks and valleys of EE.
Since tidal volume of breathing is not constant (Aliverti et al. 2004; Potter et al. 1999; Wessel et al. 1979) even breath-by-breath measurements will show variation and require multiple breaths over an aggregation interval. For larger volume (room) calorimeters, the impact of breathing is supposedly insignificant as the individual breaths average out in the larger volume.

Open-circuit diluted flow calorimeters

Principle

Indirect calorimeters using open-circuit diluted flow respirometry are in essence not different from other open-circuit respirometry applications except for measuring diluted breathing in a larger volume in contrast to measuring undiluted subject exhalation by means of a mixing chamber, Douglas bag or by measuring undiluted breath samples at the mouth (breath-by-breath systems).
The best-known open-circuit diluted flow calorimeters are most whole-room calorimeters or respiration chambers, while the most widely found are the many ventilated hood applications. In all instances, a continuous flow of fresh air is directed through the chamber or hood or facemask and all in- and outgoing air is analyzed for O2− and CO2− concentrations, while the flow through the system is measured (Fig. 4a–c).
A respiration chamber measures EE over longer stay intervals, comfortable and without hindrance during sleep or activities. In contrast, the less comfortable facemask provides measurement of EE for intervals as small as a practical 30 s. A ventilated hood may be considered an intermediary type between a whole room and a facemask, allowing bedside application while providing the minimum comfort required for measurement of BMR.
An essential parameter distinguishing these calorimeters is the volume in which the exhaled air is captured. The volume of a ventilated hood system is about 25–40 l with a typical air flow-through of 60–100 l min−1 for normal weight adults, whereas a whole room has a volume of 5000–30,000 l with a typical air flow-through in the order of 100 l min−1. Using a facemask, the volume of the mask is very small and the flow of air past the mask may vary from < 100 l min−1 for resting or low activity EE up to > 500 l min−1 for maximal exertion testing (VO2max). Volume or volume over time, i.e., flow, is an important parameter for calculating EE using indirect calorimeters. At this instance, it should be noted that any volumetric expression is dependent on temperature, pressure, and often humidity content. Well-known denominators are ATP, BTP, and STP for ambient, body, and standard temperature and pressure, respectively, and these are complemented with saturated (S) or dry (D) regarding water vapour content (no S or D indicates an unspecified percentage of water vapour, possibly measured). Thus, STPD is a dry volume at standard 0 °C and 1013.25 mBar, while BTPS is a saturated “wet” volume at 37 °C body temperature and the prevalent ambient pressure (measured). These volumetric measures are very different and must be correctly defined and measured. Typically, diluted flow calorimeters will correct all volumetric measures to STPD notation, while non-diluted flow calorimeters may use an ATPS assumption for Ve, i.e., expired “wet” air at 37 °C.
Another volumetric aspect is dead space. An advantage of open-circuit diluted flow indirect calorimetry is that a grander stream of the air passes the subject without in-line flowmeter or breathing valve(s), providing the shortest possible open pathway between mouth and fresh air volume or flow. An artificial pathway and/or resistance in-line with subject breathing flow can affect the performance of subjects. The volume of the small pathway or “dead space” artificially lengthens the anatomical dead space, i.e., the part of the airway that does not participate in actual breathing, and effectively decreases alveolar ventilation (the active ventilation of the lung).
In addition, any artificial resistance to flow will be added to the anatomical airway resistance, increasing the required work for breathing. For EE measurement with indirect calorimetry, the best practice is to decrease, or circumvent, breathing dead space, and resistance. In this respect, elimination of flowmeter or breathing valves in-line with subject breathing will decrease, or near enough eliminate, aspects of dead space and resistance.
In Fig. 4, three types of calorimeters that fit the description of open-circuit diluted flow indirect calorimetry are depicted (Fig. 4a–c). For reference, a direct whole-room calorimeter (Fig. 4e), a breath-by-breath indirect calorimeter (Fig. 4f), and a classic non-diluted calorimeter using breathing valves with Douglas bag (Douglas 1911) or flowmeter (Zuntz and Schumburg 1901a; Tissot 1904) (Fig. 4g) are depicted as well. A closed-circuit calorimeter is not drawn.
Placed in historical context, the first indirect room calorimeters captured all exhaled breath and analyzed its chemical composition, this is best represented in Fig. 4b. For non-diluted flow capturing all exhalation (Fig. 4g), this is best comparable to the setup first used by Lavoisier (Fig. 1).

Respiration chambers or whole-room calorimeters

A respiration chamber is used for measuring human EE over longer periods of time, typically between 12 h (e.g., overnight for measuring sleeping metabolic rate) and 36 h (nigh-day-night) and up to several days. The advantage of a chamber compared to a ventilated hood or facemask is that the subject can move freely within the chamber and that 24-h EE can be measured capturing all the different components of EE, i.e., sleeping metabolic rate (SMR), basal metabolic rate (BMR), diet-induced thermogenesis (DIT), and activity-related energy expenditure (AEE). The latter is of course limited due to the confined spaces of the chamber and is not representative of daily life AEE. To better mimic daily life circumstances in terms of total EE, an activity protocol can be introduced (e.g., by adding a bicycle to the room). A typical respiration chamber includes a bed, table, chair, (freeze) toilet, sink, TV, airlocks and sample ports, intercom, phone, and computer (Fig. 5 ). For privacy reasons, windows should have curtains and camera’s (if present) an off-switch.

Capturing the different components of EE in a respiration chamber

Human EE is not constant, and it varies with time of day, level of activity, food digestion, and other parameters. For each point in time, the momentary EE measured equals the sum of EE components. Although these EE components are never measured separately, some of them can be individually identified over a specific time interval, for instance, during sleep, rest, activity, meals, or over 24 h (Fig. 6).
These individual EE components do not simply start and stop. Each component will have an individual response that can be delayed after the start and/or remain active after the end of the interval for a period ranging from minutes up to hours or even days. Given these time-course effects, the protocol used is of significant importance for the analysis of EE components regarding reproducibility, as well as for allowing comparison of results to those found in the literature. For example, overnight sleep EE values for an 8-h interval (23:00–07:00) will be significantly different from EE values of sleep (SMR) taken as the lowest observed EE over a 3-h interval during the night, typically between 3:00–6:00 in the morning (Schoffelen and Westerterp 2008). The same holds true for BMR and resting metabolic rate (RMR), both providing information on EE at rest, but BMR is more strictly defined then RMR. BMR is measured early in the morning after an overnight fast while lying awake in supine position and under thermoneutral conditions. RMR also indicates an EE level of rest, typically when fasted for several non-active hours. However, position, time of day, and other factors may differ. For example, RMR could be measured several hours after lunch in a sitting position with the subject having been active in the morning, and consequently, RMR will be significantly elevated in comparison with BMR.
A typical pattern of EE over the day as measured in respiration chamber is depicted in Fig. 6, showing the different components SMR, BMR, and average EE over 24 h. DIT and AEE are difficult to separate from each other as both occur simultaneously during the course of the day and cannot be determined without measuring additional parameters. This particular experiment shows the subject going to the toilet in the middle of the night, illustrating how SMR can be disturbed if activity is not taken into account; a typical example of biological variation is even found in a strongly controlled environment.
More sources of biological variation exist, for instance, seasonal or climate influences, health issues, abnormal activity levels, diet, travel, age, gender, body composition, genetics, choices for repeated visits, etc. (Claessens-van Ooijen et al. 2006; Schols et al. 1992; Plasqui and Westerterp 2004; Plasqui et al. 2003; Dugas et al. 2009; van Dale et al. 1989; Van Dale et al. 1987; Donahoo et al. 2004; Blaak and Saris 1996; Webb 1993; Fredrix et al. 1990, 1991a, b, c; Brouns et al. 1989).
Knowledge of biological variation is of importance to the design of studies, typically determining the effect of an intervention, i.e., imposed variation, in one group and comparing it to the results of a group without intervention (van Dale et al. 1989; Van Dale et al. 1987; Brouns et al. 1989; Westerterp-Plantenga et al. 2004, 2009; Smeets and Westerterp-Plantenga 2008; Lejeune et al. 2003).
An important aspect of this variety in EE components is the accuracy and measurement interval in relation with the level of EE. An indirect room calorimeter will have a base level of accuracy that is nearly constant and independent of the level of EE, a floor noise level determined by accuracy of analysis and volume of the calorimeter (Murgatroyd et al. 1987). It is a constant that can be divided over the measurement interval i.e., large for small interval and decreasing over time. A second component is a similar floor noise level originating from the throughput flow and its analysis, constant over time, i.e., increasing as a summed total over time. Both are independent of the actual level of EE, except that in practice, flow may be set at different rates for different levels of EE. Given a medium yet constant flow setting, the floor-level accuracy is near constant, and the absolute SD for results is similar for low-level EE and high-level EE. This means that high levels of EE will have a smaller percentage of error than low levels of EE, or can be measured over a shorter interval with the same accuracy as low levels of EE over a longer interval. This aspect must be taken into account when determining the measurement interval for different EE components. For instance, the lowest EE level of sleep typically requires an interval of several hours (Schoffelen and Westerterp 2008), while higher levels may easily be determined over a 20-min interval (Webb et al. 1988; Nguyen et al. 2003).

Push versus pull systems

In Fig. 1b, an indirect whole-room calorimeter was depicted as a pull system, drawing air through the volume. A push system can be drawn near-identical by placing the enforcing fan at the air inside, pushing ambient air into the volume.
Both types exist and have their specific advantages and disadvantages. To explain the pro’s and con’s, let us first envision a system without push or pull flow, simply an enclosure without openings. If the enclosure is 100% airtight, all gas exchanges of a subject inside will integrate over time, showing a decrease in O2 and an increase in CO2 co-linear with the EE of the subject and effectively this creates a closed-circuit calorimeter for a limited time interval. To remove heat and moisture an air conditioning is needed, just as in calorimeters of the push and pull type. The heating and cooling expands and shrinks the air inside, causing pressure inside to increase and decrease accordingly. If the enclosure is not fully airtight, an increase in pressure will cause air to escape and a decrease of pressure will cause air to be drawn in. Effectively, the air conditioning will cause a bi-directional flow, sequential out-flow and in-flow, trough leakages and the co-linearity of the O2 and CO2 curves with EE of the subject is compromised.
Most room-calorimeter enclosures are not fully airtight, and the choice between push and pull is also a choice for how leakage is handled. The push type forces air into the enclosure to eliminate other sources of in-flow, while the pull type draws air out of the enclosure to eliminate other sources of out-flow. Both push and pull operations remove the bi-directional nature of leakage. The push and pull flows may be further modulated by the air conditioning unless a pressure equalization bellow is present. Both push and pull systems seem near-identical with a respective positive or negative pressure in the room, and results may be fully identical. Depending on the individual construction, a significant difference can occur due to the composition of the air that is leaked: a push-type calorimeter must assume that all subject air is fully mixed and sampled before air leaves the enclosure through leaks and output. Effectively, the output and leaks need to have the same gas concentrations. A pull-type calorimeter must assume that air surrounding the room (e.g., air in the laboratory, where the room calorimeter is situated in) is identical to inlet air (e.g., air drawn from outside air on top of the building); effectively, the input and leaks need to have the same gas concentrations. Both types assume that either input or output flow is exactly known, while the other side effectively consists of the sum of flow plus a small amount of leakage. This works as indirect calorimeters use a single main flowmeter at input (push) or output (pull) and the Haldane correction (Haldane 1892). The Haldane correction assumes that Nitrogen in the air (inert gas) is not used passing though the calorimeter: flow of N2in is equal to flow of N2out. For a pull system, measuring the amount of N2out also defines the amount of N2in, and this N2in is the sum of N2 trough inlets, i.e., including N2in through leakage. For a push system the in’s and out’s are exchanged. The Haldane correction thus allows the use of a single flowmeter (or correct a second one) for either in- or out-flow. This is important as any small difference between individual flow measurements of in and out would otherwise be summed over time. In practice, there would not be a perfect balance. Moreover, the Haldane correction allows for a small amount of leakage in one direction to be accounted for as if it was legitimate flow. In this case, a requirement is that gas composition of leakage is identical to the composition of the Haldane-corrected flow.
Since subjects are allowed to move around inside a whole-room calorimeter, the subject exhalation may or may not be fully mixed if the subject is breathing near a leak, this is a matter of design of the calorimeter, and breath leaked before mixing is hence not being measured to calculate EE. However, a pull system will show the identical effect if the surrounding air is not free of human contamination, visitors or personnel breathing near a leak may have part of their breath drawn in and this part will be added to the EE of the subject. A typical place where this might occur for both push- and pull system is the blood-sampling port, where during sampling the subject is facing a possible leak as well as the person taking the sample at the outside.
Regarding the impact of leakage, a pull system may pose less demands on mixing and sampling in comparison with a push system, as the pull system draws all subject air through one outlet. In contrast, a push system will allow flush toilet and other amenities that lose a small amount of the air, all of which must be accounted for or eliminated in a pull system.
Importantly, the above-described effect of personnel surrounding a pull calorimeter may also disturb ventilated hood measurements in the close presence of personnel. These small pull systems draw in and sample surrounding air. If ambient air is infrequently sampled and/or the ambient air sample line is placed too far from hood air inlet, the variation of ambient gas composition will affect the EE result.

Sources of error

In human indirect calorimetry, a vast amount of sources of errors can be identified (McLean and Tobin 1987). A non-exhaustive sample of common sources of errors can be subdivided in theoretical, system (random and systematical), and practical errors.
Theoretical errors originate for the greater part from choices in assumptions and formulae. A few typical assumptions are the absence of leakage, all volumes of gasses are fully mixed before evaluation, all gasses are ideal gasses, and inert gasses pass through the calorimeter unchanged (Haldane correction). A typical choice of a formula is the one used for conversion of gas-exchange results to EE (Weir 1949; Brouwer 1957) that is based on the composition of substrates (Livesey and Elia 1988), which is assumed to be identical to those consumed by the subjects at hand.
In this context, a further requirement is that validation tests are suitable for mimicking the biological application. For instance, it should be realized that when checking a breath-by-breath (BxB) system with alcohol combustion in diluted flow mode, the BxB measurement is a different mode compared to the diluted flow (ventilated hood) mode.
System errors originate from the technical implementation of the calorimeter, and these errors can be random or systematical. The former average out over time, while the latter may create a larger summed error over time.
Some examples are measurement noise and drift, aspects of digitalization, flushing of sample system and clearing of residue, accuracy of calibration, and incorrect validation methods (i.e., not suitable for the biological application). A biological systematical example is the effect of skin diffusion of CO2 (Livesey and Elia 1988; Alkalay et al. 1971; Fitzgerald 1957; Klocke et al. 1963), which accounts for a + 1–2% difference in measured CO2 for whole-room calorimetry versus ventilated hood or facemask calorimetry.
Practical errors origin from the application of the calorimeter, and these can be operator errors as well as deficiencies in locality or supplies. Examples are failure to adhere to protocol, for example, by incorrectly closing the calorimeter or other sources of leakage, or by incorrect setting of flow through the calorimeter causing a too large or too small dilution flow. Localities may impact temperature or ambient air composition, and required supplies such as calibration gasses that may have an incorrect value or certificate accuracy.
A typical source of error has been highlighted above for Push versus pull systems, leakage, and contamination of inlet air. An indirect calorimeter measures air quantities and differences in air composition caused by a single subject. No subject air may go unaccounted for and no air from others may be incorporated at all.
It can be speculated that proliferation of small indirect calorimeters (metabolic carts) and a co-occurring decrease in required effort to perform experiments, also decreased availability of expert know-how at the numerous sites applying indirect calorimetry. In some instances, an indirect calorimeter may be perceived as a magical black box for measuring human EE. However, it remains a sensitive measurement. For example, an acceptable accuracy for BMR measurements may be ± 5%, corresponding to ± 10 ml min−1 O2-consumption, typically measured with a ventilated hood with a flow of ≥ 40 l min−1. However, this minimum demand of only ± 5% biological accuracy requires ≤ 0.025% (i.e., ≤ 10 × 40,000−1) overall technical accuracy. This large difference between biological and required technical accuracy can be confusing and causes an underestimation of required technical effort.
For sites applying whole-room calorimetry, with its further increased sensitivity to errors, the required technical effort will become elevated to a degree, where staff have gained sufficient know-how and experience before achieving operational status at even a basic level.
Note that experience can also have its negative side-effects, as demands on technology change, so must procedures and understanding. It is possible to have proven operating procedures, and simply fail to observe a drift in an apparatus as it constantly remains within the old “approved” error range. A practical example would be a calorimeter supposed to operate within 5% absolute error defined as 1 ± 2% (mean ± SD). Suppose that SD improved from 2 to 0.5%; now, a 3 ± 0.5% (mean ± SD) may pass inspection, while an update in procedures should in fact keep it within 1 ± 0.5%, i.e., within 2% instead of 5% maximum error range.

Methods for system validation

Calibration is not validation

When using a calorimetry system, it is utterly important to perform regular validity testing. Validity testing is not to be confused with calibration. First, the different components inside the calorimeter (e.g., flowmeter, pressure sensor, analyser) may have been calibrated and found suitable for the intended use, but that does not test the functioning of the entire system as a whole. Second, the analysers should be calibrated each day (or even more frequent) before being used to measure a subject. This can be done by consecutively flowing two gasses containing a known concentration of O2 and CO2 through the analysers (dual span), one of which may well be 100% N2 gas (zero for O2 and CO2), or use a single gas and ambient air. The latter assumes a typical 20.9% oxygen in ambient air and commonly filters out the 0.04% ambient CO2 for creating a zero gas for CO2. Besides this analyser calibration, several options are available and required to test validity of the system, which are described below.

Alcohol combustion

A typical validation method for an indirect calorimeter is the combustion of a measured quantity of a high-purity fuel, such as methanol or ethanol. The combustion of 1 g of such a fuel will require a known amount of O2 and produce a known amount of CO2. Measured gas-exchange values from the calorimeter can then be compared to the known values based on the chemical process for the amount of fuel combusted. The most common method uses an alcohol burner, burning high-purity ethanol, or methanol (Fig. 7) inside the calorimeter. For hood systems, the standard hood is replaced by a different one (more heat resistant) to allow for the burning flame and its heat. This method also challenges the capability of the calorimeter to cope with both H2O and heat produced.

Gas infusion

A second method is the infusion of a measured quantity of a high-purity gas, and gas-exchange results of the apparatus must match the amount infused. This process does not produce H2O or heat, though cooling due to adiabatic expansion may occur (Fig. 8).
The match between infused gas and gas-exchange measurements is easily understood for CO2. An amount of CO2 infused into the calorimeter must register exactly as if it originated from a subject, and at the same time, the O2 uptake must remain zero as the total of CO2 infused should be accounted for.
The simulation of O2 consumption is more complex as there cannot be a “negative” infusion of O2 to simulate O2 uptake. Instead, an inert gas, typically N2, is infused, diluting the O2 concentration in the outgoing airstream.
Gas infusion is the method of choice for calorimeters that use a breath-by-breath analysis, measuring inhalation and exhalation separately. In that particular case, the infusion must be combined with a realistic lung simulation for breathing (Fig. 9 ) (Prieur et al. 1998), or else the calorimeter may not even function in this mode.
Unfortunately, a combined gas infusion and realistic lung simulator is not easily available, for the greater part preventing regular on-site validation of breath-by-breath calorimetry. Instead, such calorimeters may be tested only partially in their hood mode.
Gas infusion in general has the advantages of being able to achieve the broadest possible range of gas concentrations and a fast modulation of gas quantities infused. Hence, it is the method of choice for mimicking exercise.

Parallel validation

A third validation method is realized by combining two methods, i.e., one prior validated system and the system under test. The two systems are connected to measure the identical samples for measurements, and results are compared.
In applying this parallel validation, great care must be taken to prevent interaction between systems. This must be proven, as well as showing the intact validity of the prior validated system.
Simply interconnecting two devices, one of which was at some point validated in the literature, will not suffice. A calorimeter used as reference in such a test must be validated on site, and it cannot be assumed to perform as a “gold standard” apparatus without checking its individual performance and including possible interactions.
Note that parallel validation will typically introduce the concept of biological reproducibility, using results of a participant in the parallel measurement.

Using biological reproducibility for validation

Consider a technically validated calorimeter failing to show reproducibility for normal healthy subjects within known limits of biological variability. In that case, the technical validation method may have failed to sufficiently simulate biology, and the calorimeter is not suitable. It cannot be deemed validated. Consider a calorimeter showing reproducibility for normal healthy subjects within known limits of biological variability, though its technical validity or “level” of measured value is unknown. In this instance, the calorimeter may, or may not, be valid. It too cannot be deemed validated. Only a technically validated calorimeter also showing reproducibility for normal healthy subjects within known limits of biological variability may be considered a validated calorimeter. It must have proven to measure exact levels of EE and its ability to reproduce measurements with participants.
Repeated measurements with subjects were performed to test reproducibility (including both technical and biological variation) from the lowest level of EE, i.e., SMR (Schoffelen and Westerterp 2008), BMR (Adriaens et al. 2003), until 24-h EE (Schoffelen et al. 1997) and to the highest possible, i.e., VO2max (Schoffelen et al. 2017) (Table 2). The categories Sleep, OMR, BMR, 24-h EE, and Exercise showed a CV% that was smallest for high-level exercise and largest for outpatient BMR (Table 2). Although exercise had the lowest CV% of 1.2%, it has the largest standard deviation as expressed in kJ min−1. In contrast, SMR with its medium CV% of 2.4% showed the best absolute reproducibility, closely followed by OMR, 24-h EE and even BMR with an outpatient protocol (Table 2).
Table 2
Reproducibility of EE measures using repeated subject testing
Category
EE (kJ min−1)
SD (kJ min−1)
CV (%)
SMR (room)
4.6
0.11
2.4
OMR (room)
4.8
0.13
2.8
BMR (hood)
4.7
0.15
3.3
24-h EE (room)
6.8
0.13
1.9
maximal exercise, VO2max
~ 90a
~ 1a
1.2
Data are expressed as mean EE (kj min−1), standard deviation (kj min−1), and the coefficient of variation (CV%)
EE energy expenditure, SMR sleeping metabolic rate, OMR overnight metabolic rate, BMR basal metabolic rate
aEE values for VO2max were estimated for comparison of size only; they are not exact because during high intensity exercise, RER > 1, and hence, EE cannot be accurately determined due to the contribution of anaerobic metabolism

Fields of application

Since the first built room calorimeter over 150 years ago, calorimeters have evolved over time from very labor intensive instruments with chemical analysis of gasses to modern automated systems with continuous gas analysis and online data processing. These technological advancements have also lead to faster response times and higher sampling resolutions, opening up the field to a wide variety of applications. Room calorimeters have been and, currently, are being, used to study human energy metabolism under all sorts of environmental conditions.
Many studies involved testing the metabolic response to different food products, with manipulation of the amount of food (Wulan et al. 2014; Rynders et al. 2017), the relative contribution of the three main substrates, i.e., carbohydrates, fats, and proteins (Chan et al. 2014; Munsters et al. 2013), or the response to the so-called thermogenic ingredients such as tea extracts (Gregersen et al. 2009) or capsaicin (red pepper extract) (Janssens et al. 2013). Since the discovery of the existence of brown adipose tissue in adult humans, thermal physiology studying human metabolism during cold (or heat) exposure has gained a new interest (Wijers et al. 2010; Chen et al. 2013; Hibi et al. 2016; Peterson et al. 2016). In addition, given our modern global society where people can be in contact 24 h a day all over the world, regularly travel through different time zones, and are exposed to screen light on their electronic devices almost constantly, the study of sleep and circadian rhythms and the effect on human metabolism have become an important field of research (Jung et al. 2011; Gonnissen et al. 2013; McHill et al. 2014). Others have studied the effect of seasons (Plasqui et al. 2003; Minghelli et al. 1991), differences between ethnicities (Dugas et al. 2009; Wulan et al. 2012), or EE during pregnancy (Gilmore et al. 2016).
The controlled environment also provides excellent opportunity for exercise studies, as modern day calorimeters have the resolution to study EE almost in real time, even in a large volume such as a room calorimeter (Robinson et al. 2015; Frost et al. 2014). Recent data have shown that even VO2-max testing is possible in a room calorimeter sized 18 m3 (Kleinloog et al. 2017). In addition, the higher time resolution better allows performing validation studies of different wearable sensors and their ability to estimate EE (Dannecker et al. 2013).
All of the above research topics are just a small fraction of published research and there are many more applications. Room calorimeters provide an ideal platform to study metabolism under strictly controlled conditions, where all food intake, temperature, humidity, lighting, and even O2 pressure can be regulated and the response of human EE and substrate oxidation can be measured.

Conclusion

From the different indirect calorimetry techniques available, this review mainly focused on the open-circuit diluted flow principle, which can be applied with a facemask, ventilated hood or respiration chamber. All exhaled air is captured and diluted in a bypass flow that can be varied depending on the level of gas exchange, i.e., from sleeping metabolic rate until VO2max. One of the advantages is that there are no restrictions at the mouth caused by valves or flowmeters.
For bedside and exercise applications, the open-circuit diluted flow principle can be considered a modern day replacement of the full-capture calorimeters of the past. These applications have become a reality due to developments in electronics and analyser accuracy, exchanging large volumes for fast response while taking advantage of modern day whole-room calorimetric equipment. The application of indirect calorimeters is not as straightforward as the less knowledgeable user may expect. For reliable and accurate results, it requires expert know-how, suitable and frequent validation, and meticulous attention to detail. Old gold standard facts are as applicable as ever, effectively eliminating any calorimeter as being a gold standard “out of the box” and without evaluation. Effectively, there is much to be gained by studying the calorimetric principles published in the distant past.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Allgemeinmedizin

Kombi-Abonnement

Mit e.Med Allgemeinmedizin erhalten Sie Zugang zu allen CME-Fortbildungen und Premium-Inhalten der allgemeinmedizinischen Zeitschriften, inklusive einer gedruckten Allgemeinmedizin-Zeitschrift Ihrer Wahl.

Literatur
Zurück zum Zitat Adriaens MP, Schoffelen PF, Westerterp KR (2003) Intra-individual variation of basal metabolic rate and the influence of daily habitual physical activity before testing. Br J Nutr 90(2):419–423PubMed Adriaens MP, Schoffelen PF, Westerterp KR (2003) Intra-individual variation of basal metabolic rate and the influence of daily habitual physical activity before testing. Br J Nutr 90(2):419–423PubMed
Zurück zum Zitat Alkalay I, Suetsugu S, Constantine H, Stein M (1971) Carbon dioxide elimination across human skin. Am J Physiol 220(5):1434–1436PubMed Alkalay I, Suetsugu S, Constantine H, Stein M (1971) Carbon dioxide elimination across human skin. Am J Physiol 220(5):1434–1436PubMed
Zurück zum Zitat Aristotle, Ogle W (1897) Aristotle on youth & old age, life & death and respiration: translated, with introduction and notes. Longman’s, Green and Co, London Aristotle, Ogle W (1897) Aristotle on youth & old age, life & death and respiration: translated, with introduction and notes. Longman’s, Green and Co, London
Zurück zum Zitat Atwater WO, Benedict FG (1905) The respiration calorimeter : with the appliances for the direct determination of oxygen. Carnegie Institution of Washington publication no. 42. Carnegie Institution of Washington, Washington, DC Atwater WO, Benedict FG (1905) The respiration calorimeter : with the appliances for the direct determination of oxygen. Carnegie Institution of Washington publication no. 42. Carnegie Institution of Washington, Washington, DC
Zurück zum Zitat Atwater WO, Foster MS, Rosa EB, Storrs Experiment S (1897) A respiration calorimeter and experiments on the conservation of energy in the human body. s. n., s. l Atwater WO, Foster MS, Rosa EB, Storrs Experiment S (1897) A respiration calorimeter and experiments on the conservation of energy in the human body. s. n., s. l
Zurück zum Zitat Aulick LH, Arnhold H, Hander EH, Mason AD Jr (1983) A new open and closed respiration chamber. Q J Exp Physiol 68(3):351–357PubMed Aulick LH, Arnhold H, Hander EH, Mason AD Jr (1983) A new open and closed respiration chamber. Q J Exp Physiol 68(3):351–357PubMed
Zurück zum Zitat Blaak EE, Saris WH (1996) Postprandial thermogenesis and substrate utilization after ingestion of different dietary carbohydrates. Metabolism 45(10):1235–1242PubMed Blaak EE, Saris WH (1996) Postprandial thermogenesis and substrate utilization after ingestion of different dietary carbohydrates. Metabolism 45(10):1235–1242PubMed
Zurück zum Zitat Brouns F, Saris WH, Stroecken J, Beckers E, Thijssen R, Rehrer NJ, ten Hoor F (1989) Eating, drinking, and cycling. A controlled Tour de France simulation study, Part I. Int J Sports Med 10(Suppl 1):S32–S40. doi:10.1055/s-2007-1024952 CrossRefPubMed Brouns F, Saris WH, Stroecken J, Beckers E, Thijssen R, Rehrer NJ, ten Hoor F (1989) Eating, drinking, and cycling. A controlled Tour de France simulation study, Part I. Int J Sports Med 10(Suppl 1):S32–S40. doi:10.​1055/​s-2007-1024952 CrossRefPubMed
Zurück zum Zitat Brouwer E (1957) On simple formulae for calculating the heat expenditure and the quantities of carbohydrate and fat oxidized in metabolism of men and animals, from gaseous exchange (Oxygen intake and carbonic acid output) and urine-N. Acta Physiol Pharmacol Neerl 6:795–802PubMed Brouwer E (1957) On simple formulae for calculating the heat expenditure and the quantities of carbohydrate and fat oxidized in metabolism of men and animals, from gaseous exchange (Oxygen intake and carbonic acid output) and urine-N. Acta Physiol Pharmacol Neerl 6:795–802PubMed
Zurück zum Zitat Cathcart EP, Cuthbertson DP (1931) The composition and distribution of the fatty substances of the human subject. J Physiol 72:349–360PubMedPubMedCentral Cathcart EP, Cuthbertson DP (1931) The composition and distribution of the fatty substances of the human subject. J Physiol 72:349–360PubMedPubMedCentral
Zurück zum Zitat Charbonnier A, Jones CD, Schutz Y, Murgatroyd PR, Whitehead RG, Jequier E, Spinnler G (1990) A whole body transportable indirect calorimeter for human use in the tropics. Eur J Clin Nutr 44(10):725–731PubMed Charbonnier A, Jones CD, Schutz Y, Murgatroyd PR, Whitehead RG, Jequier E, Spinnler G (1990) A whole body transportable indirect calorimeter for human use in the tropics. Eur J Clin Nutr 44(10):725–731PubMed
Zurück zum Zitat Chen KY, Brychta RJ, Linderman JD, Smith S, Courville A, Dieckmann W, Herscovitch P, Millo CM, Remaley A, Lee P, Celi FS (2013) Brown fat activation mediates cold-induced thermogenesis in adult humans in response to a mild decrease in ambient temperature. J Clin Endocrinol Metab 98(7):E1218-1223. doi:10.1210/jc.2012-4213 CrossRef Chen KY, Brychta RJ, Linderman JD, Smith S, Courville A, Dieckmann W, Herscovitch P, Millo CM, Remaley A, Lee P, Celi FS (2013) Brown fat activation mediates cold-induced thermogenesis in adult humans in response to a mild decrease in ambient temperature. J Clin Endocrinol Metab 98(7):E1218-1223. doi:10.​1210/​jc.​2012-4213 CrossRef
Zurück zum Zitat Claessens-van Ooijen AM, Westerterp KR, Wouters L, Schoffelen PF, van Steenhoven AA, van Marken Lichtenbelt WD (2006) Heat production and body temperature during cooling and rewarming in overweight and lean men. Obesity (Silver Spring) 14(11):1914–1920. doi:10.1038/oby.2006.223 CrossRef Claessens-van Ooijen AM, Westerterp KR, Wouters L, Schoffelen PF, van Steenhoven AA, van Marken Lichtenbelt WD (2006) Heat production and body temperature during cooling and rewarming in overweight and lean men. Obesity (Silver Spring) 14(11):1914–1920. doi:10.​1038/​oby.​2006.​223 CrossRef
Zurück zum Zitat Consolazio CF, Johnson RE, Pecora LJ (1963) Physiological measurements of metabolic functions in man. In: Consolazio CF, Robert E, Johnson, Louis J. Pecora (eds)) With illustrations and charts.. McGraw-Hill Book Co, New York Consolazio CF, Johnson RE, Pecora LJ (1963) Physiological measurements of metabolic functions in man. In: Consolazio CF, Robert E, Johnson, Louis J. Pecora (eds)) With illustrations and charts.. McGraw-Hill Book Co, New York
Zurück zum Zitat Dauncey MJ, Murgatroyd PR (1978) A direct and indirect calorimeter for studies on energy expenditure in man over 24 h periods [proceedings]. J Physiol 284:7P–8PPubMed Dauncey MJ, Murgatroyd PR (1978) A direct and indirect calorimeter for studies on energy expenditure in man over 24 h periods [proceedings]. J Physiol 284:7P–8PPubMed
Zurück zum Zitat de Boer JO, van Es AJ, Roovers LC, van Raaij JM, Hautvast JG (1986) Adaptation of energy metabolism of overweight women to low-energy intake, studied with whole-body calorimeters. Am J Clin Nutr 44(5):585–595PubMed de Boer JO, van Es AJ, Roovers LC, van Raaij JM, Hautvast JG (1986) Adaptation of energy metabolism of overweight women to low-energy intake, studied with whole-body calorimeters. Am J Clin Nutr 44(5):585–595PubMed
Zurück zum Zitat de Boer JO, van Es AJ, van Raaij JM, Hautvast JG (1987) Energy requirements and energy expenditure of lean and overweight women, measured by indirect calorimetry. Am J Clin Nutr 46(1):13–21PubMed de Boer JO, van Es AJ, van Raaij JM, Hautvast JG (1987) Energy requirements and energy expenditure of lean and overweight women, measured by indirect calorimetry. Am J Clin Nutr 46(1):13–21PubMed
Zurück zum Zitat Donahoo WT, Levine JA, Melanson EL (2004) Variability in energy expenditure and its components. Curr Opin Clin Nutr Metab Care 7(6):599–605PubMed Donahoo WT, Levine JA, Melanson EL (2004) Variability in energy expenditure and its components. Curr Opin Clin Nutr Metab Care 7(6):599–605PubMed
Zurück zum Zitat Douglas CG (1911) A method for determining the total respiratory exchange in man. J Physiol 42:17–18 Douglas CG (1911) A method for determining the total respiratory exchange in man. J Physiol 42:17–18
Zurück zum Zitat Dugas LR, Cohen R, Carstens MT, Schoffelen PF, Luke A, Durazo-Arvizu RA, Goedecke JH, Levitt NS, Lambert EV (2009) Total daily energy expenditure in black and white, lean and obese South African women. Eur J Clin Nutr 63(5):667–673. doi:10.1038/ejcn.2008.8 CrossRefPubMed Dugas LR, Cohen R, Carstens MT, Schoffelen PF, Luke A, Durazo-Arvizu RA, Goedecke JH, Levitt NS, Lambert EV (2009) Total daily energy expenditure in black and white, lean and obese South African women. Eur J Clin Nutr 63(5):667–673. doi:10.​1038/​ejcn.​2008.​8 CrossRefPubMed
Zurück zum Zitat Fitzgerald LR (1957) Cutaneous respiration in man. Physiol Rev 37(3):325–336PubMed Fitzgerald LR (1957) Cutaneous respiration in man. Physiol Rev 37(3):325–336PubMed
Zurück zum Zitat Frappell PB, Blevin HA, Baudinette RV (1989) Understanding respirometry chambers: what goes in must come out. J Theor Biol 138(4):479–494PubMed Frappell PB, Blevin HA, Baudinette RV (1989) Understanding respirometry chambers: what goes in must come out. J Theor Biol 138(4):479–494PubMed
Zurück zum Zitat Fredrix EW, Soeters PB, Deerenberg IM, Kester AD, von Meyenfeldt MF, Saris WH (1990) Resting and sleeping energy expenditure in the elderly. Eur J Clin Nutr 44(10):741–747PubMed Fredrix EW, Soeters PB, Deerenberg IM, Kester AD, von Meyenfeldt MF, Saris WH (1990) Resting and sleeping energy expenditure in the elderly. Eur J Clin Nutr 44(10):741–747PubMed
Zurück zum Zitat Fredrix EW, Soeters PB, von Meyenfeldt MF, Saris WH (1991a) Resting energy expenditure in cancer patients before and after gastrointestinal surgery. JPEN J Parenter Enteral Nutr 15(6):604–607PubMed Fredrix EW, Soeters PB, von Meyenfeldt MF, Saris WH (1991a) Resting energy expenditure in cancer patients before and after gastrointestinal surgery. JPEN J Parenter Enteral Nutr 15(6):604–607PubMed
Zurück zum Zitat Fredrix EW, Soeters PB, Wouters EF, Deerenberg IM, von Meyenfeldt MF, Saris WH (1991b) Effect of different tumor types on resting energy expenditure. Cancer Res 51(22):6138–6141PubMed Fredrix EW, Soeters PB, Wouters EF, Deerenberg IM, von Meyenfeldt MF, Saris WH (1991b) Effect of different tumor types on resting energy expenditure. Cancer Res 51(22):6138–6141PubMed
Zurück zum Zitat Fredrix EW, Wouters EF, Soeters PB, van der Aalst AC, Kester AD, von Meyenfeldt MF, Saris WH (1991c) Resting energy expenditure in patients with non-small cell lung cancer. Cancer 68(7):1616–1621PubMed Fredrix EW, Wouters EF, Soeters PB, van der Aalst AC, Kester AD, von Meyenfeldt MF, Saris WH (1991c) Resting energy expenditure in patients with non-small cell lung cancer. Cancer 68(7):1616–1621PubMed
Zurück zum Zitat Frost EA, Redman LM, de Jonge L, Rood J, Zachwieja JJ, Volaufova J, Bray GA, Smith SR (2014) Interaction between dietary fat and exercise on excess postexercise oxygen consumption. Am J Physiol Endocrinol Metab 306(9):E1093-1098. doi:10.1152/ajpendo.00383.2013 CrossRef Frost EA, Redman LM, de Jonge L, Rood J, Zachwieja JJ, Volaufova J, Bray GA, Smith SR (2014) Interaction between dietary fat and exercise on excess postexercise oxygen consumption. Am J Physiol Endocrinol Metab 306(9):E1093-1098. doi:10.​1152/​ajpendo.​00383.​2013 CrossRef
Zurück zum Zitat Garby L, Lammert O (1984) Within-subjects between-days-and-weeks variation in energy expenditure at rest. Hum Nutr Clin Nutr 38(5):395–397PubMed Garby L, Lammert O (1984) Within-subjects between-days-and-weeks variation in energy expenditure at rest. Hum Nutr Clin Nutr 38(5):395–397PubMed
Zurück zum Zitat Garrow JS, Webster JD (1986) A computer-controlled indirect calorimeter for the measurement of energy expenditure in one or two subjects simultaneously. Hum Nutr Clin Nutr 40(4):315–321PubMed Garrow JS, Webster JD (1986) A computer-controlled indirect calorimeter for the measurement of energy expenditure in one or two subjects simultaneously. Hum Nutr Clin Nutr 40(4):315–321PubMed
Zurück zum Zitat Gorup-Besanez EFV (1867) Lehrbuch der Chemie vol 3. Lehrbuch der Physiologischen Chemie. Friedrich Vieweg und Sohn, Braunschweig Gorup-Besanez EFV (1867) Lehrbuch der Chemie vol 3. Lehrbuch der Physiologischen Chemie. Friedrich Vieweg und Sohn, Braunschweig
Zurück zum Zitat Granato L, Brandes A, Bruni C, Greco AV, Mingrone G (2004) VO2, VCO2, and RQ in a respiratory chamber: accurate estimation based on a new mathematical model using the Kalman-Bucy method. J Appl Physiol (1985) 96(3):1045–1054. doi:10.1152/japplphysiol.00788.2003 CrossRef Granato L, Brandes A, Bruni C, Greco AV, Mingrone G (2004) VO2, VCO2, and RQ in a respiratory chamber: accurate estimation based on a new mathematical model using the Kalman-Bucy method. J Appl Physiol (1985) 96(3):1045–1054. doi:10.​1152/​japplphysiol.​00788.​2003 CrossRef
Zurück zum Zitat Gregersen NT, Bitz C, Krog-Mikkelsen I, Hels O, Kovacs EM, Rycroft JA, Frandsen E, Mela DJ, Astrup A (2009) Effect of moderate intakes of different tea catechins and caffeine on acute measures of energy metabolism under sedentary conditions. Br J Nutr 102(8):1187–1194. doi:10.1017/S0007114509371779 CrossRefPubMed Gregersen NT, Bitz C, Krog-Mikkelsen I, Hels O, Kovacs EM, Rycroft JA, Frandsen E, Mela DJ, Astrup A (2009) Effect of moderate intakes of different tea catechins and caffeine on acute measures of energy metabolism under sedentary conditions. Br J Nutr 102(8):1187–1194. doi:10.​1017/​S000711450937177​9 CrossRefPubMed
Zurück zum Zitat Grimaux É (1888) Lavoisier, 1743–1794: d’après sa correspondance, ses manuscrits, ses papiers de famille et d’autres documents inédits. F. Alcan (Paris) Grimaux É (1888) Lavoisier, 1743–1794: d’après sa correspondance, ses manuscrits, ses papiers de famille et d’autres documents inédits. F. Alcan (Paris)
Zurück zum Zitat Hagemann O (1911) Das Respirations-Calorimeter in Bonn und einige Untersuchungen mit demeselben bei zwei Rindern und einem Pferde. Landw. Jahrb. Bd. 41, Erg. Bd. 1. Berlin Hagemann O (1911) Das Respirations-Calorimeter in Bonn und einige Untersuchungen mit demeselben bei zwei Rindern und einem Pferde. Landw. Jahrb. Bd. 41, Erg. Bd. 1. Berlin
Zurück zum Zitat Haldane J (1892) A new form of apparatus for measuring the respiratory exchange of animals. J Physiol 13(5):419–430PubMedPubMedCentral Haldane J (1892) A new form of apparatus for measuring the respiratory exchange of animals. J Physiol 13(5):419–430PubMedPubMedCentral
Zurück zum Zitat Harris JA, Benedict FG, Pearson K, Galton Laboratory for National E (1919) A biometric study of basal metabolism in man. Carnegie Institution of Washington publication no. 279, vol 279. Carnegie Institution of Washington, Washington Harris JA, Benedict FG, Pearson K, Galton Laboratory for National E (1919) A biometric study of basal metabolism in man. Carnegie Institution of Washington publication no. 279, vol 279. Carnegie Institution of Washington, Washington
Zurück zum Zitat Henning B, Lofgren R, Sjostrom L (1996) Chamber for indirect calorimetry with improved transient response. Med Biol Eng Comput 34(3):207–212PubMed Henning B, Lofgren R, Sjostrom L (1996) Chamber for indirect calorimetry with improved transient response. Med Biol Eng Comput 34(3):207–212PubMed
Zurück zum Zitat Hibi M, Oishi S, Matsushita M, Yoneshiro T, Yamaguchi T, Usui C, Yasunaga K, Katsuragi Y, Kubota K, Tanaka S, Saito M (2016) Brown adipose tissue is involved in diet-induced thermogenesis and whole-body fat utilization in healthy humans. Int J Obes (Lond) 40(11):1655–1661. doi:10.1038/ijo.2016.124 CrossRef Hibi M, Oishi S, Matsushita M, Yoneshiro T, Yamaguchi T, Usui C, Yasunaga K, Katsuragi Y, Kubota K, Tanaka S, Saito M (2016) Brown adipose tissue is involved in diet-induced thermogenesis and whole-body fat utilization in healthy humans. Int J Obes (Lond) 40(11):1655–1661. doi:10.​1038/​ijo.​2016.​124 CrossRef
Zurück zum Zitat Hill AV, Hill AM (1914) A self-recording calorimeter for large animals Hill AV, Hill AM (1914) A self-recording calorimeter for large animals
Zurück zum Zitat Hill JO, Sun M (1991) Measurement of the energy-cost of activity in humans using a whole-room indirect calorimeter equipped with a Mechanical floor. Faseb J 5(6):A1649-A1649 Hill JO, Sun M (1991) Measurement of the energy-cost of activity in humans using a whole-room indirect calorimeter equipped with a Mechanical floor. Faseb J 5(6):A1649-A1649
Zurück zum Zitat Jequier E, Schutz Y (1983) Long-term measurements of energy expenditure in humans using a respiration chamber. Am J Clin Nutr 38(6):989–998PubMed Jequier E, Schutz Y (1983) Long-term measurements of energy expenditure in humans using a respiration chamber. Am J Clin Nutr 38(6):989–998PubMed
Zurück zum Zitat Jequier E, Pittet P, Gygax PH (1978) Thermic effect of glucose and thermal body insulation in lean and obese subjects: a calorimetric approach. Proc Nutr Soc 37(1):45–53PubMed Jequier E, Pittet P, Gygax PH (1978) Thermic effect of glucose and thermal body insulation in lean and obese subjects: a calorimetric approach. Proc Nutr Soc 37(1):45–53PubMed
Zurück zum Zitat Kenny GP, Poirier MP, Metsios GS, Boulay P, Dervis S, Friesen BJ, Malcolm J, Sigal RJ, Seely AJ, Flouris AD (2017) Hyperthermia and cardiovascular strain during an extreme heat exposure in young versus older adults. Temperature (Austin) 4(1):79–88. doi:10.1080/23328940.2016.1230171 CrossRef Kenny GP, Poirier MP, Metsios GS, Boulay P, Dervis S, Friesen BJ, Malcolm J, Sigal RJ, Seely AJ, Flouris AD (2017) Hyperthermia and cardiovascular strain during an extreme heat exposure in young versus older adults. Temperature (Austin) 4(1):79–88. doi:10.​1080/​23328940.​2016.​1230171 CrossRef
Zurück zum Zitat Kleinloog JPD, van Laar SPGA, Schoffelen PFM, Plasqui G (2017) Validity and reproducibility of VO2max assessment in a respiration chamber (submitted for publication) Kleinloog JPD, van Laar SPGA, Schoffelen PFM, Plasqui G (2017) Validity and reproducibility of VO2max assessment in a respiration chamber (submitted for publication)
Zurück zum Zitat Klocke RA, Gurtner GH, Farhi LE (1963) Gas transfer across the skin in man. J Appl Physiol 18:311–316PubMed Klocke RA, Gurtner GH, Farhi LE (1963) Gas transfer across the skin in man. J Appl Physiol 18:311–316PubMed
Zurück zum Zitat Langworthy CF, Milner RD (1911) The respiration calorimeter and the results of experiments with it. From Yearbook, Dept. of agric. Wash Langworthy CF, Milner RD (1911) The respiration calorimeter and the results of experiments with it. From Yearbook, Dept. of agric. Wash
Zurück zum Zitat Lavoisier M (1777) Experiences sur la respiration des animaux, et sur les changemens qui arrivent a l’air en passant par leur poumon. Memoires de l’Academie des Sciences de Paris; 1777, pp 185–194 Lavoisier M (1777) Experiences sur la respiration des animaux, et sur les changemens qui arrivent a l’air en passant par leur poumon. Memoires de l’Academie des Sciences de Paris; 1777, pp 185–194
Zurück zum Zitat Lavoisier A-Ld (1783) Mémoire sur la chaleur, lu à l’Académie royale des sciences, le 28 juin 1783, par MM. Lavoisier et de La Place. Impr. royale, Paris Lavoisier A-Ld (1783) Mémoire sur la chaleur, lu à l’Académie royale des sciences, le 28 juin 1783, par MM. Lavoisier et de La Place. Impr. royale, Paris
Zurück zum Zitat Lejeune MP, Kovacs EM, Westerterp-Plantenga MS (2003) Effect of capsaicin on substrate oxidation and weight maintenance after modest body-weight loss in human subjects. Br J Nutr 90(3):651–659PubMed Lejeune MP, Kovacs EM, Westerterp-Plantenga MS (2003) Effect of capsaicin on substrate oxidation and weight maintenance after modest body-weight loss in human subjects. Br J Nutr 90(3):651–659PubMed
Zurück zum Zitat Livesey G, Elia M (1988) Estimation of energy expenditure, net carbohydrate utilization, and net fat oxidation and synthesis by indirect calorimetry: evaluation of errors with special reference to the detailed composition of fuels. Am J Clin Nutr 47(4):608–628PubMed Livesey G, Elia M (1988) Estimation of energy expenditure, net carbohydrate utilization, and net fat oxidation and synthesis by indirect calorimetry: evaluation of errors with special reference to the detailed composition of fuels. Am J Clin Nutr 47(4):608–628PubMed
Zurück zum Zitat Lusk G (1928) The elements of the science of nutrition, 4th edn. W. B. Saunders Co., Philadelphia Lusk G (1928) The elements of the science of nutrition, 4th edn. W. B. Saunders Co., Philadelphia
Zurück zum Zitat Magnus-Levy A (1893) Ueber die Grösse des respiratorischen Gaswechsels unter dem Einfluss der Nahrungsaufnahme. Archiv für die gesamte Physiologie des Menschen der Tiere 55(1):1–126. doi:10.1007/bf01661783 CrossRef Magnus-Levy A (1893) Ueber die Grösse des respiratorischen Gaswechsels unter dem Einfluss der Nahrungsaufnahme. Archiv für die gesamte Physiologie des Menschen der Tiere 55(1):1–126. doi:10.​1007/​bf01661783 CrossRef
Zurück zum Zitat Marcet W, Marcet W, Foster MS, Royal S, Physiological Laboratory UCL, Marcet W, Royal S (1898) A calorimeter for the human body. s. n., London Marcet W, Marcet W, Foster MS, Royal S, Physiological Laboratory UCL, Marcet W, Royal S (1898) A calorimeter for the human body. s. n., London
Zurück zum Zitat McHill AW, Melanson EL, Higgins J, Connick E, Moehlman TM, Stothard ER, Wright KP Jr (2014) Impact of circadian misalignment on energy metabolism during simulated nightshift work. Proc Natl Acad Sci USA 111(48):17302–17307. doi:10.1073/pnas.1412021111 CrossRefPubMed McHill AW, Melanson EL, Higgins J, Connick E, Moehlman TM, Stothard ER, Wright KP Jr (2014) Impact of circadian misalignment on energy metabolism during simulated nightshift work. Proc Natl Acad Sci USA 111(48):17302–17307. doi:10.​1073/​pnas.​1412021111 CrossRefPubMed
Zurück zum Zitat McLean JA, Tobin G (1987) Animal and human calorimetry. Cambridge University Press, Cambridge McLean JA, Tobin G (1987) Animal and human calorimetry. Cambridge University Press, Cambridge
Zurück zum Zitat McLean JA, Watts PR (1976) Analytical refinements in animal calorimetry. J Appl Physiol 40(5):827–831PubMed McLean JA, Watts PR (1976) Analytical refinements in animal calorimetry. J Appl Physiol 40(5):827–831PubMed
Zurück zum Zitat Minghelli G, Schutz Y, Whitehead R, Jequier E (1991) Seasonal changes in 24-h and basal energy expenditures in rural Gambian men as measured in a respiration chamber. Am J Clin Nutr 53(1):14–20PubMed Minghelli G, Schutz Y, Whitehead R, Jequier E (1991) Seasonal changes in 24-h and basal energy expenditures in rural Gambian men as measured in a respiration chamber. Am J Clin Nutr 53(1):14–20PubMed
Zurück zum Zitat Munsters MJ, Geraedts MC, Saris WH (2013) Effects of different protein and glycemic index diets on metabolic profiles and substrate partitioning in lean healthy males. Appl Physiol Nutr Metab 38(11):1107–1114. doi:10.1139/apnm-2012-0409 CrossRefPubMed Munsters MJ, Geraedts MC, Saris WH (2013) Effects of different protein and glycemic index diets on metabolic profiles and substrate partitioning in lean healthy males. Appl Physiol Nutr Metab 38(11):1107–1114. doi:10.​1139/​apnm-2012-0409 CrossRefPubMed
Zurück zum Zitat Murgatroyd PR, Davies HL, Prentice AM (1987) Intra-individual variability and measurement noise in estimates of energy expenditure by whole body indirect calorimetry. Br J Nutr 58(3):347–356PubMed Murgatroyd PR, Davies HL, Prentice AM (1987) Intra-individual variability and measurement noise in estimates of energy expenditure by whole body indirect calorimetry. Br J Nutr 58(3):347–356PubMed
Zurück zum Zitat Nguyen T, de Jonge L, Smith SR, Bray GA (2003) Chamber for indirect calorimetry with accurate measurement and time discrimination of metabolic plateaus of over 20 min. Med Biol Eng Comput 41(5):572–578PubMed Nguyen T, de Jonge L, Smith SR, Bray GA (2003) Chamber for indirect calorimetry with accurate measurement and time discrimination of metabolic plateaus of over 20 min. Med Biol Eng Comput 41(5):572–578PubMed
Zurück zum Zitat Noguchi H, Ogushi Y, Yoshiya I, Itakura N, Yamabayashi H (1982) Breath-by-breath VCO2 and VO2 required compensation for transport delay and dynamic response. J Appl Physiol Respir Environ Exerc Physiol 52(1):79–84PubMed Noguchi H, Ogushi Y, Yoshiya I, Itakura N, Yamabayashi H (1982) Breath-by-breath VCO2 and VO2 required compensation for transport delay and dynamic response. J Appl Physiol Respir Environ Exerc Physiol 52(1):79–84PubMed
Zurück zum Zitat Ott I, Foster MS (1891) An improved calorimeter. Contributions to the physiology and pathology of the nervous system Part XI. s. n., Easton Ott I, Foster MS (1891) An improved calorimeter. Contributions to the physiology and pathology of the nervous system Part XI. s. n., Easton
Zurück zum Zitat Perret C, Mueller G (2006) Validation of a new portable ergospirometric device (Oxycon Mobile) during exercise. Int J Sports Med 27(5):363–367PubMed Perret C, Mueller G (2006) Validation of a new portable ergospirometric device (Oxycon Mobile) during exercise. Int J Sports Med 27(5):363–367PubMed
Zurück zum Zitat Peterson CM, Lecoultre V, Frost EA, Simmons J, Redman LM, Ravussin E (2016) The thermogenic responses to overfeeding and cold are differentially regulated. Obesity (Silver Spring) 24(1):96–101. doi:10.1002/oby.21233 CrossRef Peterson CM, Lecoultre V, Frost EA, Simmons J, Redman LM, Ravussin E (2016) The thermogenic responses to overfeeding and cold are differentially regulated. Obesity (Silver Spring) 24(1):96–101. doi:10.​1002/​oby.​21233 CrossRef
Zurück zum Zitat Pettenkofer M (1862) Ueber einen neuen Respirations-Apparat. Abhandlungen der Bayerischen Akademie der Wissenschaften, Mathematisch-Physikalische Klasse, vol Bd. 9, Abth. 2, [1]. Verl. der k. Akad., München Pettenkofer M (1862) Ueber einen neuen Respirations-Apparat. Abhandlungen der Bayerischen Akademie der Wissenschaften, Mathematisch-Physikalische Klasse, vol Bd. 9, Abth. 2, [1]. Verl. der k. Akad., München
Zurück zum Zitat Plasqui G, Westerterp KR (2004) Seasonal variation in total energy expenditure and physical activity in Dutch young adults. Obes Res 12(4):688–694PubMed Plasqui G, Westerterp KR (2004) Seasonal variation in total energy expenditure and physical activity in Dutch young adults. Obes Res 12(4):688–694PubMed
Zurück zum Zitat Plasqui G, Kester AD, Westerterp KR (2003) Seasonal variation in sleeping metabolic rate, thyroid activity, and leptin. Am J Physiol Endocrinol Metab 285(2):E338-343 Plasqui G, Kester AD, Westerterp KR (2003) Seasonal variation in sleeping metabolic rate, thyroid activity, and leptin. Am J Physiol Endocrinol Metab 285(2):E338-343
Zurück zum Zitat Potter CR, Childs DJ, Houghton W, Armstrong N (1999) Breath-to-breath “noise” in the ventilatory and gas exchange responses of children to exercise. Eur J Appl Physiol Occup Physiol 80(2):118–124. doi:10.1007/s004210050567 CrossRefPubMed Potter CR, Childs DJ, Houghton W, Armstrong N (1999) Breath-to-breath “noise” in the ventilatory and gas exchange responses of children to exercise. Eur J Appl Physiol Occup Physiol 80(2):118–124. doi:10.​1007/​s004210050567 CrossRefPubMed
Zurück zum Zitat Priestley J (1774) Experiments and observations on different kinds of air, by Joseph Priestley. J. Johnson, London Priestley J (1774) Experiments and observations on different kinds of air, by Joseph Priestley. J. Johnson, London
Zurück zum Zitat Prieur F, Busso T, Castells J, Bonnefoy R, Benoit H, Geyssant A, Denis C (1998) A system to simulate gas exchange in humans to control quality of metabolic measurements. Eur J Appl Physiol Occup Physiol 78(6):549–554. doi:10.1007/s004210050459 CrossRefPubMed Prieur F, Busso T, Castells J, Bonnefoy R, Benoit H, Geyssant A, Denis C (1998) A system to simulate gas exchange in humans to control quality of metabolic measurements. Eur J Appl Physiol Occup Physiol 78(6):549–554. doi:10.​1007/​s004210050459 CrossRefPubMed
Zurück zum Zitat Proctor DN, Beck KC (1996) Delay time adjustments to minimize errors in breath-by-breath measurement of Vo2 during exercise. J Appl Physiol (1985) 81(6):2495–2499PubMed Proctor DN, Beck KC (1996) Delay time adjustments to minimize errors in breath-by-breath measurement of Vo2 during exercise. J Appl Physiol (1985) 81(6):2495–2499PubMed
Zurück zum Zitat Reardon FD, Leppik KE, Wegmann R, Webb P, Ducharme MB, Kenny GP (2006) The Snellen human calorimeter revisited, re-engineered and upgraded: design and performance characteristics. Med Biol Eng Comput 44(8):721–728. doi:10.1007/s11517-006-0086-5 CrossRefPubMed Reardon FD, Leppik KE, Wegmann R, Webb P, Ducharme MB, Kenny GP (2006) The Snellen human calorimeter revisited, re-engineered and upgraded: design and performance characteristics. Med Biol Eng Comput 44(8):721–728. doi:10.​1007/​s11517-006-0086-5 CrossRefPubMed
Zurück zum Zitat Robinson SL, Hattersley J, Frost GS, Chambers ES, Wallis GA (2015) Maximal fat oxidation during exercise is positively associated with 24-hour fat oxidation and insulin sensitivity in young, healthy men. J Appl Physiol (1985) 118(11):1415–1422. doi:10.1152/japplphysiol.00058.2015 CrossRefPubMed Robinson SL, Hattersley J, Frost GS, Chambers ES, Wallis GA (2015) Maximal fat oxidation during exercise is positively associated with 24-hour fat oxidation and insulin sensitivity in young, healthy men. J Appl Physiol (1985) 118(11):1415–1422. doi:10.​1152/​japplphysiol.​00058.​2015 CrossRefPubMed
Zurück zum Zitat Rumpler WV, Seale JL, Conway JM, Moe PW (1990) Repeatability of 24-h energy expenditure measurements in humans by indirect calorimetry. Am J Clin Nutr 51(2):147–152PubMed Rumpler WV, Seale JL, Conway JM, Moe PW (1990) Repeatability of 24-h energy expenditure measurements in humans by indirect calorimetry. Am J Clin Nutr 51(2):147–152PubMed
Zurück zum Zitat Rynders CA, Bergouignan A, Kealey E, Bessesen DH (2017) Ability to adjust nocturnal fat oxidation in response to overfeeding predicts 5-year weight gain in adults. Obesity (Silver Spring) 25(5):873–880. doi:10.1002/oby.21807 CrossRef Rynders CA, Bergouignan A, Kealey E, Bessesen DH (2017) Ability to adjust nocturnal fat oxidation in response to overfeeding predicts 5-year weight gain in adults. Obesity (Silver Spring) 25(5):873–880. doi:10.​1002/​oby.​21807 CrossRef
Zurück zum Zitat Santorio S (1614) Ars … de statica medicina, aphorismorum sectionibus septem comprehensa. Apud Nicolaum Polum, Venetiis Italy Santorio S (1614) Ars … de statica medicina, aphorismorum sectionibus septem comprehensa. Apud Nicolaum Polum, Venetiis Italy
Zurück zum Zitat Schoffelen PFM (1985) Evaluation of an automatic indirect calorimeter for measurement of energy balance in man. In: van Es AJH (ed) European Nutrition Report No 5. Koninklijke Bibliotheek, The Hague, pp 51–54 Schoffelen PFM (1985) Evaluation of an automatic indirect calorimeter for measurement of energy balance in man. In: van Es AJH (ed) European Nutrition Report No 5. Koninklijke Bibliotheek, The Hague, pp 51–54
Zurück zum Zitat Schoffelen PFM, Plasqui G (2016) Utilization of different formulae to calculate substrate oxidation in whole room calorimeters; an updated equation for indirect calorimetry submitted Schoffelen PFM, Plasqui G (2016) Utilization of different formulae to calculate substrate oxidation in whole room calorimeters; an updated equation for indirect calorimetry submitted
Zurück zum Zitat Schoffelen PF, Westerterp KR, Saris WH, Ten Hoor F (1997) A dual-respiration chamber system with automated calibration. J Appl Physiol 83(6):2064–2072PubMed Schoffelen PF, Westerterp KR, Saris WH, Ten Hoor F (1997) A dual-respiration chamber system with automated calibration. J Appl Physiol 83(6):2064–2072PubMed
Zurück zum Zitat Schoffelen PFM, den Hoed M, van Breda E, Plasqui G (2017) Determination of intra-individual variability for VO2max using a standardised outpatient protocol and a validated open circuit calorimeter (submitted for publication) Schoffelen PFM, den Hoed M, van Breda E, Plasqui G (2017) Determination of intra-individual variability for VO2max using a standardised outpatient protocol and a validated open circuit calorimeter (submitted for publication)
Zurück zum Zitat Scholander PF (1947) Accurate analysis of respiratory gases in 0.5 cubic centimeter samples. Fed Proc 6(1 Pt 2):198PubMed Scholander PF (1947) Accurate analysis of respiratory gases in 0.5 cubic centimeter samples. Fed Proc 6(1 Pt 2):198PubMed
Zurück zum Zitat Schols AM, Schoffelen PF, Ceulemans H, Wouters EF, Saris WH (1992) Measurement of resting energy expenditure in patients with chronic obstructive pulmonary disease in a clinical setting. JPEN J Parenter Enteral Nutr 16(4):364–368PubMed Schols AM, Schoffelen PF, Ceulemans H, Wouters EF, Saris WH (1992) Measurement of resting energy expenditure in patients with chronic obstructive pulmonary disease in a clinical setting. JPEN J Parenter Enteral Nutr 16(4):364–368PubMed
Zurück zum Zitat Schutz Y, Ravussin E, Diethelm R, Jequier E (1982) Spontaneous physical activity measured by radar in obese and control subject studied in a respiration chamber. Int J Obes 6(1):23–28PubMed Schutz Y, Ravussin E, Diethelm R, Jequier E (1982) Spontaneous physical activity measured by radar in obese and control subject studied in a respiration chamber. Int J Obes 6(1):23–28PubMed
Zurück zum Zitat Seale JL, Rumpler WV, Moe PW (1991) Description of a direct-indirect room-sized calorimeter. Am J Physiol 260(2 Pt 1):E306-320 Seale JL, Rumpler WV, Moe PW (1991) Description of a direct-indirect room-sized calorimeter. Am J Physiol 260(2 Pt 1):E306-320
Zurück zum Zitat Shetty PS, Sheela ML, Murgatroyd PR, Kurpad AV (1987) An open-circuit indirect whole body calorimeter for the continuous measurement of energy expenditure of man in the tropics. Indian J Med Res 85:453–460PubMed Shetty PS, Sheela ML, Murgatroyd PR, Kurpad AV (1987) An open-circuit indirect whole body calorimeter for the continuous measurement of energy expenditure of man in the tropics. Indian J Med Res 85:453–460PubMed
Zurück zum Zitat Tissot J (1904) Nouvelle méthode de mesure et d’inscription du débit et des mouvements respiratoires de l’homme et des animaux. Journal de physiologie et de pathologie générale 6:688–700 Tissot J (1904) Nouvelle méthode de mesure et d’inscription du débit et des mouvements respiratoires de l’homme et des animaux. Journal de physiologie et de pathologie générale 6:688–700
Zurück zum Zitat Van Dale D, Saris WH, Schoffelen PF, Ten Hoor F (1987) Does exercise give an additional effect in weight reduction regimens? Int J Obes 11(4):367–375PubMed Van Dale D, Saris WH, Schoffelen PF, Ten Hoor F (1987) Does exercise give an additional effect in weight reduction regimens? Int J Obes 11(4):367–375PubMed
Zurück zum Zitat van Dale D, Schoffelen PF, ten Hoor F, Saris WH (1989) Effects of addition of exercise to energy restriction on 24-hour energy expenditure, sleeping metabolic rate and daily physical activity. Eur J Clin Nutr 43(7):441–451PubMed van Dale D, Schoffelen PF, ten Hoor F, Saris WH (1989) Effects of addition of exercise to energy restriction on 24-hour energy expenditure, sleeping metabolic rate and daily physical activity. Eur J Clin Nutr 43(7):441–451PubMed
Zurück zum Zitat Webb P (1985) Human calorimeters. Greenwood Publishing Group Webb P (1985) Human calorimeters. Greenwood Publishing Group
Zurück zum Zitat Webb P (1993) Daily activity and body temperature. Eur J Appl Physiol Occup Physiol 66(2):174–177PubMed Webb P (1993) Daily activity and body temperature. Eur J Appl Physiol Occup Physiol 66(2):174–177PubMed
Zurück zum Zitat Webb P, Annis JF, Troutman SJ Jr (1980) Energy balance in man measured by direct and indirect calorimetry. Am J Clin Nutr 33(6):1287–1298PubMed Webb P, Annis JF, Troutman SJ Jr (1980) Energy balance in man measured by direct and indirect calorimetry. Am J Clin Nutr 33(6):1287–1298PubMed
Zurück zum Zitat Webb P, Saris WH, Schoffelen PF, Van Ingen Schenau GJ, Ten Hoor F (1988) The work of walking: a calorimetric study. Med Sci Sports Exerc 20(4):331–337PubMed Webb P, Saris WH, Schoffelen PF, Van Ingen Schenau GJ, Ten Hoor F (1988) The work of walking: a calorimetric study. Med Sci Sports Exerc 20(4):331–337PubMed
Zurück zum Zitat Weir JB (1949) New methods for calculating metabolic rate with special reference to protein metabolism. J Physiol 109(1–2):1–9PubMedPubMedCentral Weir JB (1949) New methods for calculating metabolic rate with special reference to protein metabolism. J Physiol 109(1–2):1–9PubMedPubMedCentral
Zurück zum Zitat Wessel HU, Stout RL, Bastanier CK, Paul MH (1979) Breath-by-breath variation of FRC: effect on VO2 and VCO2 measured at the mouth. J Appl Physiol Respir Environ Exerc Physiol 46(6):1122–1126PubMed Wessel HU, Stout RL, Bastanier CK, Paul MH (1979) Breath-by-breath variation of FRC: effect on VO2 and VCO2 measured at the mouth. J Appl Physiol Respir Environ Exerc Physiol 46(6):1122–1126PubMed
Zurück zum Zitat Westerterp-Plantenga MS, Lejeune MP, Nijs I, van Ooijen M, Kovacs EM (2004) High protein intake sustains weight maintenance after body weight loss in humans. Int J Obes Relat Metab Disord 28(1):57–64. doi:10.1038/sj.ijo.0802461 CrossRefPubMed Westerterp-Plantenga MS, Lejeune MP, Nijs I, van Ooijen M, Kovacs EM (2004) High protein intake sustains weight maintenance after body weight loss in humans. Int J Obes Relat Metab Disord 28(1):57–64. doi:10.​1038/​sj.​ijo.​0802461 CrossRefPubMed
Zurück zum Zitat Westerterp-Plantenga MS, Lejeune MP, Smeets AJ, Luscombe-Marsh ND (2009) Sex differences in energy homeostatis following a diet relatively high in protein exchanged with carbohydrate, assessed in a respiration chamber in humans. Physiol Behav 97(3–4):414–419. doi:10.1016/j.physbeh.2009.03.010 CrossRefPubMed Westerterp-Plantenga MS, Lejeune MP, Smeets AJ, Luscombe-Marsh ND (2009) Sex differences in energy homeostatis following a diet relatively high in protein exchanged with carbohydrate, assessed in a respiration chamber in humans. Physiol Behav 97(3–4):414–419. doi:10.​1016/​j.​physbeh.​2009.​03.​010 CrossRefPubMed
Zurück zum Zitat Wulan SN, Westerterp KR, Plasqui G (2014) Metabolic profile before and after short-term overfeeding with a high-fat diet: a comparison between South Asian and White men. Br J Nutr 111(10):1853–1861PubMed Wulan SN, Westerterp KR, Plasqui G (2014) Metabolic profile before and after short-term overfeeding with a high-fat diet: a comparison between South Asian and White men. Br J Nutr 111(10):1853–1861PubMed
Zurück zum Zitat Yamamoto Y, Takei Y, Mokushi K, Morita H, Mutoh Y, Miyashita M (1987) Breath-by-breath measurement of alveolar gas exchange with a slow-response gas analyser. Med Biol Eng Comput 25(2):141–146PubMed Yamamoto Y, Takei Y, Mokushi K, Morita H, Mutoh Y, Miyashita M (1987) Breath-by-breath measurement of alveolar gas exchange with a slow-response gas analyser. Med Biol Eng Comput 25(2):141–146PubMed
Zurück zum Zitat Zuntz N, Lehmann C (1890) Remarks on the Chemistry of respiration in the Horse during Rest and Work. J Physiol 11(4–5):396–398PubMedPubMedCentral Zuntz N, Lehmann C (1890) Remarks on the Chemistry of respiration in the Horse during Rest and Work. J Physiol 11(4–5):396–398PubMedPubMedCentral
Zurück zum Zitat Zuntz N, Schumburg W (1901a) Studien zu einer Physiologie des Marsches. Bibliothek von Coler, vol 6. Hirschwald, Berlin Zuntz N, Schumburg W (1901a) Studien zu einer Physiologie des Marsches. Bibliothek von Coler, vol 6. Hirschwald, Berlin
Zurück zum Zitat Zuntz N, Schumburg WAEFD (1901b) Studien zu einer Physiologie des Marsches. Bibliothek von Coler Bd. 6. August Hirschwald, Berlin Zuntz N, Schumburg WAEFD (1901b) Studien zu einer Physiologie des Marsches. Bibliothek von Coler Bd. 6. August Hirschwald, Berlin
Metadaten
Titel
Classical experiments in whole-body metabolism: open-circuit respirometry—diluted flow chamber, hood, or facemask systems
verfasst von
P. F. M. Schoffelen
G. Plasqui
Publikationsdatum
27.10.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Applied Physiology / Ausgabe 1/2018
Print ISSN: 1439-6319
Elektronische ISSN: 1439-6327
DOI
https://doi.org/10.1007/s00421-017-3735-5

Weitere Artikel der Ausgabe 1/2018

European Journal of Applied Physiology 1/2018 Zur Ausgabe