Erschienen in: Virology Journal 1/2022

Open Access 01.12.2022 | Research

Clinical evaluation of a fully automated and high-throughput molecular testing system for detection of influenza virus

verfasst von: Kosuke Kosai, Norihito Kaku, Michiko Horie, Hina Kodama, Norihiko Akamatsu, Yusuke Narita, Yasushi Matsumoto, Tetsuro Matsushita, Yohei Mizuta, Koichi Izumikawa, Hiroshi Mukae, Katsunori Yanagihara

Erschienen in: Virology Journal | Ausgabe 1/2022

Abstract

Introduction

We investigated the performance of the cobas® 6800 system and cobas SARS-CoV-2 & Influenza A/B, a fully automated molecular testing system for influenza viruses and severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). This enabled an assay in a batch of 96 samples in approximately 3 h.

Methods

An assay was performed using the cobas SARS-CoV-2 & Influenza A/B on the cobas 6800 system for samples collected in four facilities between November 2019 and March 2020 in our previous study. The results were compared with those obtained using the reference methods.

Results

Of the 127 samples analyzed, the cobas SARS-CoV-2 & Influenza A/B detected influenza A virus in 75 samples, of which 73 were positive using the reference methods. No false negative results were observed. The overall positive and negative percent agreement for influenza A virus detection were 100.0% and 96.3%, respectively. There were no positive results for the influenza B virus or SARS-CoV-2.

Conclusion

The cobas 6800 system and cobas SARS-CoV-2 & Influenza A/B showed high accuracy for influenza A virus detection and can be useful for clinical laboratories, especially those that routinely assay many samples.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
NAAT
nucleic acid amplification test
POC
point-of-care
SARS-CoV-2
severe acute respiratory syndrome coronavirus
TCID50
50% tissue culture infectious dose
PPA
positive percent agreement
NPA
negative percent agreement

Introduction

Influenza is a major respiratory pathogen that causes seasonal epidemics and outbreaks. Influenza is a self-limiting disease in most patients; however, some patients develop complications such as pneumonia, exacerbation of chronic obstructive pulmonary disease, myocarditis, and encephalopathy [1, 2].
Several methods are available for detecting influenza viruses, including antigen and nucleic acid amplification tests (NAATs). Although antigen testing is rapid and simple, its sensitivity is generally low [3]. NAAT is a reliable method with high sensitivity; however, conventional PCR is time consuming and labor intensive. Recently, automated molecular systems have been developed for the detection of influenza viruses. We recently reported the usefulness of a fully automated molecular point-of-care (POC) testing system, the cobas Influenza A/B & RSV on the cobas® Liat system, for influenza A virus detection. The assay is performed on a single sample within a short time (approximately 20 min) [4]. In contrast, the cobas® 6800 system, a fully automated molecular testing system, provides a batch assay for 96 samples in approximately 3 h. In this study, we evaluated the performance of the cobas 6800 system and cobas SARS-CoV-2 & Influenza A/B, which targeted severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) and influenza A and B viruses, focusing on the detection of influenza A virus.

Methods

Samples

In our previous study [4], we used nasopharyngeal swab samples collected from adult outpatients with one or more symptoms, such as fever (38 °C or an increase of 1 °C from normal body temperature), nasal discharge, nasal congestion, sore throat, cough, headache, chills, fatigue, joint pain, or muscle pain, at Menoto Hospital, Matsumoto Naika, Narita Naika Clinic, and Shinzato Medicare Group Shinzato Clinic between November 2019 and March 2020. Patients were excluded if they were treated with anti-influenza agents within 1 month prior to the outpatient visit. The remaining samples in the UTM (Copan Italia s.p.a.) from our previous study were stored at − 80 °C, and samples with sufficient residual volume for this study were randomly selected and used.

Assay using the cobas SARS-CoV-2 & Influenza A/B on the cobas 6800 system

According to the manufacturer’s instructions, Roche Diagnostics K.K. performed the assay using cobas SARS-CoV-2 & Influenza A/B on a cobas 6800 system. Eight hundred microliters of the sample was transferred to a cobas omni secondary tube and loaded onto the assay machine. Sample preparation, reverse transcription of the target RNA to complementary DNA, and real-time multiplexed PCR were automatically performed on the system. The targets were genes of membrane proteins 1 and 2 for influenza A virus, genes of nonstructural and nuclear export proteins for influenza B virus, and open reading frame 1a/b and envelope protein genes for SARS-CoV-2. Positive, negative, and internal controls were used for each assay. The system displayed the validity of the test and detection results for each target. According to the manufacturer’s instructions, the limits of detection were 0.026–0.14, 0.017–0.053, and 0.0079–0.12 50% tissue culture infectious dose (TCID50)/mL for influenza A virus, influenza B virus, and SARS-CoV-2, respectively.

Reference methods

The detection of influenza viruses was performed in our previous study [4] and the results were used as reference results. Briefly, samples were assayed using the cobas Influenza A/B & RSV on the cobas Liat system, as well as an automated immunochromatographic antigen test (digital immunoassay, DIA test). The concordant results between the two were considered true results. Samples with discrepant results for influenza A virus (positive in the molecular POC test but negative in the DIA test) were confirmed to be true-positive by RT-PCR. None of the samples tested negative in the molecular POC test but positive in the DIA test. Finally, all Liat Flu/RSV results for influenza A virus detection were considered true results.
Roche Diagnostics K.K. performed the assay for SARS-CoV-2 detection according to the Manual for the Detection of Pathogen 2019-nCoV Ver.2.9.1, published by the National Institute of Infectious Diseases in Japan. Briefly, RNA was extracted using the QIAamp Viral RNA Mini (QIAGEN), and one-step real-time reverse transcription-PCR was performed using the QuantiTect Prove RT-PCR kit (QIAGEN) on a LightCycler® 480 Instrument II (Roche Molecular Systems Inc.).

Results

A total of 129 samples were assayed using the cobas SARS-CoV-2 & Influenza A/B. Two samples were excluded from the analysis due to invalid results. Comparisons between the cobas SARS-CoV-2 & Influenza A/B and the reference methods are presented in Table 1. Of the 127 samples tested, the cobas SARS-CoV-2 & Influenza A/B detected influenza A virus in 75 samples, of which 73 were positive using the reference methods. No samples tested negative on the cobas SARS-CoV-2 & Influenza A/B and positive with the reference methods. The overall positive and negative percent agreement (PPA and NPA) for influenza A virus detection were 100.0% and 96.3%, respectively. Influenza B virus was not detected by the cobas SARS-CoV-2 & Influenza A/B or by the reference methods.
Table 1
Comparison of detection results between cobas SARS-CoV-2 & Influenza A/B and the reference methods
Viruses
cobas+, RM+
cobas+, RM-
cobas-, RM+
cobas-, RM-
PPA
NPA
Influenza A virus
73
2
0
52
100.0 (95.1–100.0)
96.3 (87.3–99.5)
Influenza B virus
0
0
0
127
-
100.0 (97.1–100.0)
SARS-CoV-2
0
0
0
47
-
100.0 (92.5–100.0)
RM, reference method; PPA, positive percent agreement; NPA, negative percent agreement
SARS-CoV-2 was not detected by the cobas SARS-CoV-2 & Influenza A/B in 127 samples. Of these samples, 47 were analyzed using the reference method for SARS-CoV-2 detection, and the results were negative.

Discussion

This study demonstrated that the cobas SARS-CoV-2 & Influenza A/B could accurately detect influenza A virus (PPA, 100.0%) using a simple procedure. Unfortunately, we could not evaluate its detection performance for SARS-CoV-2 because the samples used in this study were collected before the SARS-CoV-2 pandemic in our region.
Currently, many molecular POC tests for individual samples are available and provide rapid and accurate diagnosis [46], which might lead to the appropriate use of antiviral agents and the reduction of inappropriate antibiotic use. In contrast, the cobas 6800 system used in this study enables a batch assay and can simultaneously examine 96 samples (up to 384 samples in an 8-hour shift) [7]. This system can be useful in clinical laboratories that routinely assay a large number of samples.
Furthermore, if influenza epidemics and the coronavirus disease 2019 (COVID-19) pandemic occur simultaneously, it may be difficult to differentiate them based on only clinical presentation [8]. This implies the need for accurate diagnostic methods that can distinguish between influenza and COVID-19 and detect their co-infections.

Conclusion

The cobas 6800 system and cobas SARS-CoV-2 & Influenza A/B showed high accuracy for influenza A virus detection and can be useful for clinical laboratories, especially those that assay many samples in daily practice.

Acknowledgements

The cobas SARS-CoV-2 & Influenza A/B test has an Emergency Use Authorization (EUA) in the US, is CE-marked under the In Vitro Diagnostic Medical Device Directive (IVDD) but not under the In Vitro Diagnostic Medical Device Regulation (IVDR), is approved by the Pharmaceuticals and Medical Devices Agency (PMDA), and the status varies internationally; thus, this product is not approved in all regions. COBAS, LIAT, and LIGHTCYCLER are trademarks of Roche.

Declarations

Ethical approval

This study was approved by the Institutional Review Board of Nagasaki University Hospital (approval number:20111620).
Not applicable.

Competing interests

Assays in this study were carried out by Roche Diagnostics K.K. K.Y. received honoraria from Roche Diagnostics K.K. M.H. is an employee of Roche Diagnostics K.K. Funding was provided by Roche Diagnostics K.K.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
2.
Zurück zum Zitat Macias AE, McElhaney JE, Chaves SS, Nealon J, Nunes MC, Samson SI, et al. The disease burden of influenza beyond respiratory illness. Vaccine. 2021;39(Suppl 1):A6–14.CrossRefPubMed Macias AE, McElhaney JE, Chaves SS, Nealon J, Nunes MC, Samson SI, et al. The disease burden of influenza beyond respiratory illness. Vaccine. 2021;39(Suppl 1):A6–14.CrossRefPubMed
3.
Zurück zum Zitat Chartrand C, Leeflang MM, Minion J, Brewer T, Pai M. Accuracy of rapid influenza diagnostic tests: a meta-analysis. Ann Intern Med. 2012;156(7):500–11.CrossRefPubMed Chartrand C, Leeflang MM, Minion J, Brewer T, Pai M. Accuracy of rapid influenza diagnostic tests: a meta-analysis. Ann Intern Med. 2012;156(7):500–11.CrossRefPubMed
4.
Zurück zum Zitat Kaku N, Kodama H, Akamatsu N, Ota K, Kosai K, Morinaga Y, et al. Multicenter evaluation of molecular point-of-care testing and digital immunoassays for influenza virus A/B and respiratory syncytial virus in patients with influenza-like illness. J Infect Chemother. 2021;27(6):820–5.CrossRefPubMed Kaku N, Kodama H, Akamatsu N, Ota K, Kosai K, Morinaga Y, et al. Multicenter evaluation of molecular point-of-care testing and digital immunoassays for influenza virus A/B and respiratory syncytial virus in patients with influenza-like illness. J Infect Chemother. 2021;27(6):820–5.CrossRefPubMed
5.
Zurück zum Zitat Hansen GT. Point-of-care testing in microbiology: A mechanism for improving patient outcomes. Clin Chem. 2020;66(1):124–37.CrossRefPubMed Hansen GT. Point-of-care testing in microbiology: A mechanism for improving patient outcomes. Clin Chem. 2020;66(1):124–37.CrossRefPubMed
6.
Zurück zum Zitat Rezaei M, Razavi Bazaz S, Zhand S, Sayyadi N, Jin D, Stewart MP, et al. Point of care diagnostics in the age of COVID-19. Diagnostics. 2020;11(1). Rezaei M, Razavi Bazaz S, Zhand S, Sayyadi N, Jin D, Stewart MP, et al. Point of care diagnostics in the age of COVID-19. Diagnostics. 2020;11(1).
7.
Zurück zum Zitat Norz D, Hoffmann A, Aepfelbacher M, Pfefferle S, Lutgehetmann M. Clinical evaluation of a fully automated, laboratory-developed multiplex RT-PCR assay integrating dual-target SARS-CoV-2 and influenza A/B detection on a high-throughput platform. J Med Microbiol. 2021;70(2). Norz D, Hoffmann A, Aepfelbacher M, Pfefferle S, Lutgehetmann M. Clinical evaluation of a fully automated, laboratory-developed multiplex RT-PCR assay integrating dual-target SARS-CoV-2 and influenza A/B detection on a high-throughput platform. J Med Microbiol. 2021;70(2).
8.
Zurück zum Zitat Chotpitayasunondh T, Fischer TK, Heraud JM, Hurt AC, Monto AS, Osterhaus A, et al. Influenza and COVID-19: What does co-existence mean? Influenza Other Respir Viruses. 2021;15(3):407–12.CrossRefPubMed Chotpitayasunondh T, Fischer TK, Heraud JM, Hurt AC, Monto AS, Osterhaus A, et al. Influenza and COVID-19: What does co-existence mean? Influenza Other Respir Viruses. 2021;15(3):407–12.CrossRefPubMed
Metadaten
Titel
Clinical evaluation of a fully automated and high-throughput molecular testing system for detection of influenza virus
verfasst von
Kosuke Kosai
Norihito Kaku
Michiko Horie
Hina Kodama
Norihiko Akamatsu
Yusuke Narita
Yasushi Matsumoto
Tetsuro Matsushita
Yohei Mizuta
Koichi Izumikawa
Hiroshi Mukae
Katsunori Yanagihara
Publikationsdatum
01.12.2022
Verlag
BioMed Central
Erschienen in
Virology Journal / Ausgabe 1/2022
Elektronische ISSN: 1743-422X
DOI
https://doi.org/10.1186/s12985-022-01916-w