Skip to main content
Erschienen in: Oral Radiology 3/2019

Open Access 13.10.2018 | Original Article

Comparability of dental implant site ridge measurements using ultra-low-dose multidetector row computed tomography combined with filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction

verfasst von: Asma’a Abdurrahman Al-Ekrish, Reema Al-Shawaf, Wafa Alfaleh, Romed Hörmann, Wolfgang Puelacher, Gerlig Widmann

Erschienen in: Oral Radiology | Ausgabe 3/2019

Abstract

Objective

To assess the linear measurements of edentulous ridges recorded from multidetector row computed tomography (MDCT) images obtained by a previously untested ultra-low dose in combination with filtered back-projection (FBP), adaptive statistical iterative reconstruction (ASIR), and model-based iterative reconstruction (MBIR).

Methods

Three cadavers were imaged using a reference protocol with a standard dose and FBP (volume CT dose index (CTDIvol): 29.4 mGy) and two ultra-low-dose protocols, LD1 and LD2 (CTDIvol: 0.53 and 0.29 mGy). All test examinations were reconstructed with FBP, ASIR 50, ASIR 100, and MBIR. Linear measurements from the images of the edentulous ridges recorded from the test protocols were compared with those from the reference using a one-sample t test, Bland–Altman plots, and linear regression. Statistical significance was set at a p value of 0.05.

Results

The one-sample t test demonstrated a statistically significant difference between the measurements from the reference protocol and all test protocols. The difference was not clinically significant for the following three test protocols: LD1/FBP, LD1/ASIR 50, and LD2/FBP. Bland–Altman plots with linear regression showed no systematic variation between the measurements obtained with the reference protocol and these three test protocols.

Conclusions

The lowest-dose protocol to demonstrate comparable measurements with a standard MDCT dose was CTDIvol 0.29 mGy with FBP. These results must be considered with caution for areas of the jaws with thin cortication. The results in areas of thin cortication should be verified by studies with larger sample sizes at such areas and comparison with true gold standard measurements.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s11282-018-0350-z) contains supplementary material, which is available to authorized users.

Introduction

The widespread use of computed tomography (CT) in the analysis of prospective dental implant sites, although diagnostically beneficial, may contribute to the increasing collective dose of ionizing radiation to populations [1]. Therefore, dose optimization of multidetector row CT (MDCT) imaging in dental implantology is needed [2].
Although cone beam CT is advocated for replacement of MDCT in implant diagnostics because of the reportedly lower radiation dose and cost, MDCT is still the only option available in some settings. Radiation doses imparted by cone beam CT and MDCT are affected by numerous factors but may be comparable in some situations, especially if ultra-low MDCT doses are utilized [35].
One method that has been advocated because of its allowance of dose reductions in MDCT is the use of iterative reconstruction techniques (IRTs) to replace the filtered back-projection (FBP) technique in reconstruction of MDCT images. Two examples of IRTs are adaptive statistical iterative reconstruction (ASIR) and model-based iterative reconstruction (MBIR). Compared with FBP, the use of ASIR and MBIR have been shown to produce images with reduced noise at ultra-low doses [6]; however, they may also lead to over-smoothening of the images, which may adversely affect identification of anatomical interfaces in the images [7, 8]. Use of ASIR and MBIR, as well as FBP, with ultra-low-dose MDCT at a volume CT dose index (CTDIvol) of 0.53 mGy has been demonstrated to produce linear measurements of dental implant sites comparable with those produced with a standard dose (CTDIvol of 36.71 mGy); however, ASIR and MBIR did not provide an advantage over FBP [9].
Notably, most of the samples in the previous study were mandibular sites, which have thicker cortical boundaries than maxillary sites. It is therefore conceivable that identification of ridge boundaries and recording of ridge dimensions at maxillary sites may be more adversely affected at ultra-low MDCT doses and may thus benefit from reduced noise levels produced by ASIR and MBIR. Furthermore, the lowest dose limit for production of acceptable linear measurements and which reconstruction technique would support the lowest dose reductions remain unknown.
Investigation of more aggressive dose reductions using such techniques has the potential to further reduce the radiation doses imparted by MDCT imaging of dental implant sites. However, because the effect of dose on image quality is directly influenced by the reconstruction technique used, investigations of dose reductions must also take the reconstruction technique into consideration. Therefore, this study was performed to assess the comparability of linear measurements of edentulous ridges recorded from MDCT images obtained by a previously untested ultra-low dose in combination with FBP, ASIR, and MBIR compared with a reference MDCT protocol. Testing the various combinations of doses with reconstruction techniques will help to determine which reconstruction technique allows use of the lowest dose.

Materials and methods

This study was performed using three cadaveric heads. Institutional review board approval was not required because the bodies used in the study were donated by people who had given their informed consent for their use for scientific and educational purposes prior to death and the study fulfilled all requirements necessary for studies on human cadavers according to the regulations of the Division of Clinical and Functional Anatomy, Medical University of Innsbruck [10, 11]. The cadavers had been preserved using an arterial injection of a formaldehyde–phenol/alcohol–glycerin solution and immersion in phenolic acid in water for 1–3 months [12]. Two of the cadavers were completely edentulous, and one was partially dentulous in the mandibular anterior/premolar regions.

Imaging protocols

Each of the cadavers was imaged using a 64-row CT scanner (Discovery CT750 HD; GE Healthcare, Vienna, Austria) using a reference protocol and two ultra-low-dose test protocols, LD1 and LD2. The position of each cadaver within the gantry of the MDCT device was not changed between examinations. The exposure parameters, reconstruction technique, and imparted dose of each examination are detailed in Table 1. The effective doses of the protocols were also calculated from the dose–length product based on a k factor of 0.0021 for adult head examinations [13]. Nine MDCT datasets (1 reference and 8 test datasets) were obtained for each cadaver, for a total of 27 datasets. All datasets were obtained with a bone convolution kernel except for the MBIR datasets, which were obtained with a standard convolution kernel because a bone kernel is not available with MBIR. The datasets were uploaded in Digital Imaging and Communication in Medicine format to an online storage site, Google Drive; they were then downloaded and imported into a three-dimensional image reformatting software program (OnDemand Software, version 1.0; Cybermed Inc., Seoul, South Korea) for reformatting of the sample sites.
Table 1
List of exposure parameters and reconstruction techniques used in the reference and ultra-low-dose test protocols
Exposure protocol
Reconstruction technique
Matrix
In-plane voxel size (mm)
kV
mA
Rotation time (s)
Pitch
CTDIvol (mGy)
Effective dose (µSv)a
Scan length of 5 cm
Scan length of 10 cm
Reference
FBP
512 × 512
0.391
120
100
1.0
0.53
29.4
308.7
617.4
LD1
FBP
512 × 512
0.391
80
10
0.4
0.53
0.53
5.6
11.1
ASIR 50
ASIR 100
MBIR
1024 × 1024
0.195
LD2
FBP
512 × 512
0.391
0.97
0.29
3.0
6.1
ASIR 50
ASIR 100
MBIR
1024 × 1024
0.195
CTDIvol volume CT dose index, LD low-dose protocol, FBP filtered back-projection, ASIR adaptive statistical iterative reconstruction, MBIR model-based iterative reconstruction
aCalculated from the dose–length product based upon a k factor of 0.0021

Image processing

The image processing and analysis method used was similar to that described in a previous study [9]. Processing and reformatting of all sample images were performed by one maxillofacial radiologist (A.A.) with 12 years of experience in MDCT image reformatting. The entire lengths of the maxillary and mandibular edentulous areas were used to obtain the cross-sectional sample images. The exclusion criteria for the sample sites were areas containing foreign objects or artificial defects and areas with buccolingual or occlusoapical dimensions of < 2 mm.
Each sample image was reformatted individually using the three-dimensional module of the OnDemand software. Because the position of each cadaver was fixed during the MDCT examinations, the default position and angulation of the sectional planes were identical in all datasets of each cadaver. Thus, the measured shift and angulation of the image planes allowed the production of sample sites with a standardized location and orientation between different examination protocols. The reformatting protocol for each cadaver is available upon request. The mA and kVp values were turned off, and each sample site was saved by bookmarking in the OnDemand software on the hard drive of the computer. Table 2 shows the images of one sample site obtained by the reference and test protocols.
Table 2
Images of a sample site obtained with the various combinations of exposure protocols and reconstructions techniques
https://static-content.springer.com/image/art%3A10.1007%2Fs11282-018-0350-z/MediaObjects/11282_2018_350_Tab2_HTML.png
LD low-dose protocol, FBP filtered back-projection, ASIR adaptive statistical iterative reconstruction, MBIR model-based iterative reconstruction

Image analysis

Two oral and maxillofacial radiologists (R.S. and W.A.) with 8 and 14 years of experience, respectively, in interpretation of CT images recorded the test measurements. The examiners were blinded to the exposure protocols and reconstruction techniques and were calibrated in identification of the outer boundaries of the edentulous bone on cross-sectional images. The examiners viewed the images at a set window width/window level of 3000/650, but they were permitted to adjust the magnification level for clarity.
The height and width measurements of the ridge were recorded on the sample images along the lines representing the parasagittal and axial planes, respectively. The limits of the measurements were the outer cortical boundaries of the ridge. The apical cortical boundary at the anterior maxillary sites was the floor of the nasal cavity, and that at the posterior maxillary sites was the floor of the maxillary sinus (Fig. 1). The first examiner (R.S.) recorded all measurements from all of the samples once. Next, using a random number generator (http://​stattrek.​com/​statistics/​random-number-generator.​aspx), 50 measurements were selected for reliability testing. These measurements were then repeated by the two examiners independently of each other.

Statistical evaluation

The two measurements recorded by the first examiner were used to calculate intra-examiner reliability. For calculation of the inter-examiner reliability, the mean of the two measurements recorded by the first examiner was compared with the corresponding measurements recorded by the second examiner. Intra-examiner and inter-examiner reliability were analyzed by Cronbach’s alpha and inter-item correlation.
Agreement between the measurements recorded from the reference protocol and each of the test protocols was tested by subtracting the test measurements from the reference measurements and analyzing the difference with a one-sample t test (test value: zero). For each test protocol, if the one-sample t test revealed the absolute mean difference to be < 0.3 mm with a 95% confidence interval within ± 0.5 mm, then it was considered that no clinically significant difference was present; in such cases, a Bland–Altman plot was constructed and linear regression was used to test for proportional bias [14]. Statistical analysis was performed using the Statistical Package for the Social Sciences (SPSS) version 24 (IBM Corp., Armonk, NY, USA). Statistical significance was set at a p value of 0.05.

Results

Thirty sample images were obtained from the three cadavers. The site distribution of the samples is shown in Table 3. Height and width measurements were recorded from each sample for a total of 60 measurements from each imaging protocol. Nine imaging protocols were used for each sample for a total of 540 measurements.
Table 3
Distribution of the sample sites
Cadaver
Maxillary sites
Mandibular sites
Total
Incisor
Canine/premolar
Molar
Incisor
Canine/premolar
Molar
1
1
2
2
2
4
11
2
1
2
2
1
4
10
3
1
2
2
 
1
3
9
Total
3
6
6
3
1
11
30
The intra-examiner and inter-examiner reliability of the measurements were found to be very high, with Cronbach’s alpha and intraclass correlation coefficients of 0.997 for the intra-examiner reliability and 0.998 for the inter-examiner reliability (p < 0.001).
The one-sample t test revealed a statistically significant difference in measurements between the reference protocol and all the test protocols (Table 4). However, the difference was not clinically significant for the following three test protocols: LD1/FBP, LD1/ASIR 50, and LD2/FBP. Bland–Altman plots (Electronic Supplementary Fig. 1) with linear regression (Table 4) showed no systematic variation between the measurements obtained with the reference protocol and these three test protocols.
Table 4
Differences between linear measurements obtained by the reference protocol (reference dose/filtered back-projection) and the various test protocols
 
Mean difference (mm)*
Standard deviation of difference (mm)
95% confidence interval limits (mm)
Significance (linear regression)
Lower limit
Upper limit
LD1/FBP
− 0.2083
0.61128
− 0.3662
− 0.0504
0.605
LD1/ASIR 50
− 0.2817
0.66498
− 0.4534
− 0.1099
0.394
LD1/ASIR 100
− 0.3467
0.61959
− 0.5067
− 0.1866
LD1/MBIR
− 0.4667
1.01508
− 0.7289
− 0.2044
LD2/FBP
− 0.2500
0.76878
− 0.4486
− 0.0514
0.584
LD2/ASIR 50
− 0.6367
1.47269
− 1.0171
− 0.2562
LD2/ASIR 100
− 0.4050
0.55736
− 0.5490
− 0.2610
LD2/MBIR
− 0.4583
1.05899
− 0.7319
− 0.1848
– Bland–Altman plot and linear regression were not performed because the one-sample t test demonstrated a mean difference greater than ± 0.3 mm with 95% confidence limits greater than ± 0.5 mm
LD low-dose protocol, FBP filtered back-projection, ASIR adaptive statistical iterative reconstruction, MBIR model-based iterative reconstruction
*All mean differences were statistically significant (p < 0.05)

Discussion

The present study was performed to test the accuracy of linear measurements of edentulous ridges recorded from previously untested ultra-low-dose MDCT images compared with those from a reference dose protocol. The effective doses achieved in the present study are lower than those reported for many cone beam CT devices, even for some small-field-of-view examinations [4]. The results indicate that the measurements recorded using the lowest dose tested, CTDIvol of 0.29, and FBP were comparable with those from a reference dose protocol. The IRTs of ASIR and MBIR did not allow similarly accurate measurements. As such, the present study demonstrates previously unrealized potentials of ultra-low reduction in doses and FBP in implant site imaging.
The images of some sample sites, especially in the maxillary posterior regions, had very thin cortical boundaries and thus subjectively appeared to be more adversely affected by the amount of noise seen with the LD2/FBP protocol. However, the boundaries of the ridge were still identifiable in the images acquired by LD2/FBP as evidenced by the objective measurements presented in Table 4. This finding, namely that clinically useful information may be obtained with images that are not subjectively pleasing to the eye, was addressed by the National Council on Radiation Protection and Measurements when they modified the ALARA concept of imaging (as low as reasonably achievable) to the ALADA concept (as low as diagnostically achievable) [2]. The modification is meant to emphasize that regardless of whether the images are subjectively pleasing to the eye, the lowest doses should be used provided that the diagnostic information is obtainable. The European Union Council Directive 2013/59/EURATOM also addressed this issue by mandating that for all imaging procedures, the dose of ionizing radiation should be kept as low as reasonably achievable, consistent with obtaining the required medical information [15].
The improved performance of FBP over ASIR and MBIR in the present study occurred despite smaller voxel sizes being used with MBIR. The poorer performance of ASIR and MBIR compared with FBP may be due to their over-smoothening effect, which may increase the area of blurriness at the margins of the bone [7, 8]. For the same reason, previous studies have shown that ASIR and MBIR did not improve spatial resolution or the detection of fractures [8, 16] or identification of the roof of the inferior alveolar canal [17].
In the present study, the clinical significance of the absolute measurement differences was designated as > 0.3 mm with a 95% confidence interval beyond ± 0.5 mm for several reasons. First, this range of error is smaller than the size of the image voxel. Second, the measurement error of the reference protocol images when compared with the gold standard of measurements recorded directly from the bone is expected to be larger than this value; previous studies that compared ridge dimensions from MDCT images acquired with the standard protocol versus the dimensions recorded directly from bone showed that the smallest mean error was 0.36 mm with a standard deviation of ± 0.24 mm [1821]. Third, implant diameters may include 0.3-mm increments, and during interpretation of images, operators commonly round off linear measurements to the nearest 0.5 mm.
The test protocols in the present study utilized the lowest mA and kV settings possible with the device used. Thus, to further reduce the radiation dose, the rotation time was reduced to less than half of that used for the reference protocol. The combined effect of reduced mA and rotation time is a reduction in the amount of X-ray photons, but an increase in the quantum noise in the resultant images. However, a previous study demonstrated accurate ridge dimensions using the same mA and kV and rotation time settings as LD1 [9]. Therefore, to further lower the dose, the present study used a higher pitch value in LD2 than that used in the previous study. The pitch value used in the present study did not exceed the maximum acceptable pitch values for 64-row MDCT scanners, namely 0.9844, beyond which interpolation of data occurs [22]. Although some researchers reported that the use of a pitch factor of 2 did not reduce the subjective quality of MDCT images of implant sites when combined with standard exposure parameters [23], other investigators reported that an increasing pitch factor was negatively associated with the number of successful objective linear measurements [24]. However, considering the high degree of comparability between the reference and lowest-dose protocol in the present study, investigation of further dose optimization using further increases in pitch is recommended.
The results of the present study are in agreement with a previous study that revealed comparable measurements between a reference dose protocol and an ultra-low-dose protocol similar to LD1 of the present study [9]. However, the previous study indicated that ASIR and MBIR produced measurements comparable to FBP, whereas the present study showed that FBP was more accurate at such very low doses. The reason for the differences may be that in the previous study, the sample sites were mostly in the mandibular molar region, which has a thick layer of dense cortical bone that is more clearly depicted in images and does not require image contrast as high as that needed for thin, low-density bone, which was found in other parts of the jaws in the present study. Thus, although ASIR and MBIR are beneficial in terms of noise reduction, FBP images should be used for bone measurements.
A limitation of the present study is that a site-specific analysis of the measurement differences was not conducted because of the small sample size in each specific area of the jaws. Although the overall study sample demonstrated no clinically significant difference with three of the test protocols, larger clinically significant differences might occur at particular sites with thin, low-density margins. As such, further studies with larger sample sizes at each area of the jaws and separate analysis of the measurements from each area are recommended. Additionally, the present study did not evaluate the influence of various sharpness levels of convolution kernels using ultra-low-dose images and IRTs. The use of various sharpness levels might produce different results than those obtained in the present study.
Another limitation of the present study is that the test measurements were not compared with a true gold standard of direct bone measurements. Using the reference protocol measurements as the standard must be considered with caution because such measurements have been shown to have statistically significant error [21]. Whether the measurement differences between the test and reference protocols in the present study would translate to increased or decreased error compared with bone measurements remains unknown. Therefore, further studies are recommended to investigate the absolute error of measurements compared with a true gold standard.
The lowest-dose protocol to demonstrate comparable measurements with a standard MDCT dose was found to be a CTDIvol of 0.29 mGy with FBP. The IRTs of ASIR and MBIR yielded larger measurement differences. These results must be considered with caution when evaluating areas of the jaws with thin cortication. The results in such areas should be verified by studies with larger sample sizes in such areas, and the findings should be compared with true gold standard measurements.

Acknowledgements

The authors extend their appreciation to the Deanship of Scientific Research at King Saud University for funding this work through research group number RGP-1438-037. The authors also thank all individuals who have donated their bodies and tissues for the advancement of education and research.

Compliance with ethical standards

Conflict of interest

Asma’a Abdurrahman Al-Ekrish, Reema Al-Shawaf, Wafa Alfaleh, Romed Hörmann, Wolfgang Puelacher, and Gerlig Widmann declare that they have no conflict of interest.

Human rights statement

This article does not contain any studies with human or animal subjects performed by any of the authors.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Dent – Das Online-Abo der Zahnmedizin

Online-Abonnement

Mit e.Dent erhalten Sie Zugang zu allen zahnmedizinischen Fortbildungen und unseren zahnmedizinischen und ausgesuchten medizinischen Zeitschriften.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Anhänge

Electronic supplementary material

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation. Official Records of the General Assembly, Sixty-third Session, Supplement No. 46; New York: United Nations; 2010. United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation. Official Records of the General Assembly, Sixty-third Session, Supplement No. 46; New York: United Nations; 2010.
2.
Zurück zum Zitat National Council on Radiation Protection and Measurements. Achievements of the past 50 years and addressing the needs of the future. In: NCRP fiftieth annual meeting program; 2014. National Council on Radiation Protection and Measurements. Achievements of the past 50 years and addressing the needs of the future. In: NCRP fiftieth annual meeting program; 2014.
3.
Zurück zum Zitat Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;106(1):106–14.CrossRefPubMed Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;106(1):106–14.CrossRefPubMed
4.
Zurück zum Zitat Ludlow J, Timothy R, Walker C, Hunter R, Benavides E, Samuelson D, et al. Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol. 2015;44(1):20140197.CrossRefPubMed Ludlow J, Timothy R, Walker C, Hunter R, Benavides E, Samuelson D, et al. Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol. 2015;44(1):20140197.CrossRefPubMed
5.
Zurück zum Zitat Al-Ekrish A, Alfadda S, Tamimi D, Alfaleh W, Hörmann R, Puelacher W, et al. Do ultra-low multidetector computed tomography doses and iterative reconstruction techniques affect subjective classification of bone type at dental implant sites? Int J Prosthodont. 2018;31(5):465–70.CrossRefPubMed Al-Ekrish A, Alfadda S, Tamimi D, Alfaleh W, Hörmann R, Puelacher W, et al. Do ultra-low multidetector computed tomography doses and iterative reconstruction techniques affect subjective classification of bone type at dental implant sites? Int J Prosthodont. 2018;31(5):465–70.CrossRefPubMed
6.
Zurück zum Zitat Widmann G, Al-Shawaf R, Schullian P, Al-Sadhan R, Hormann R, Al-Ekrish AA. Effect of ultra-low doses, ASIR and MBIR on density and noise levels of MDCT images of dental implant sites. Eur Radiol. 2017;27(5):2225–34.CrossRefPubMed Widmann G, Al-Shawaf R, Schullian P, Al-Sadhan R, Hormann R, Al-Ekrish AA. Effect of ultra-low doses, ASIR and MBIR on density and noise levels of MDCT images of dental implant sites. Eur Radiol. 2017;27(5):2225–34.CrossRefPubMed
7.
Zurück zum Zitat Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W. Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol. 2010;194(1):191–9.CrossRefPubMed Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W. Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol. 2010;194(1):191–9.CrossRefPubMed
8.
Zurück zum Zitat Widmann G, Dalla Torre D, Hoermann R, Schullian P, Gassner EM, Bale R, et al. Ultralow-dose computed tomography imaging for surgery of midfacial and orbital fractures using ASIR and MBIR. Int J Oral Maxillofac Surg. 2015;44(4):441–6.CrossRefPubMed Widmann G, Dalla Torre D, Hoermann R, Schullian P, Gassner EM, Bale R, et al. Ultralow-dose computed tomography imaging for surgery of midfacial and orbital fractures using ASIR and MBIR. Int J Oral Maxillofac Surg. 2015;44(4):441–6.CrossRefPubMed
9.
Zurück zum Zitat Al-Ekrish A, Al-Shawaf R, Schullian P, Al-Sadhan R, Hörmann R, Widmann G. Validity of linear measurements of the jaws using ultralow-dose MDCT and the iterative techniques of ASIR and MBIR. Int J Comput Assist Radiol Surg. 2016;11(10):1791–801.CrossRefPubMed Al-Ekrish A, Al-Shawaf R, Schullian P, Al-Sadhan R, Hörmann R, Widmann G. Validity of linear measurements of the jaws using ultralow-dose MDCT and the iterative techniques of ASIR and MBIR. Int J Comput Assist Radiol Surg. 2016;11(10):1791–801.CrossRefPubMed
10.
Zurück zum Zitat McHanwell S, Brenner E, Chirculescu ARM, Drukker J, van Mameren H, Mazzotti G, et al. The legal and ethical framework governing body donation in Europe—a review of current practice and recommendations for good practice. Eur J Anat. 2008;12(1):1–24. McHanwell S, Brenner E, Chirculescu ARM, Drukker J, van Mameren H, Mazzotti G, et al. The legal and ethical framework governing body donation in Europe—a review of current practice and recommendations for good practice. Eur J Anat. 2008;12(1):1–24.
11.
Zurück zum Zitat Riederer BM, Bolt S, Brenner E, Bueno-Lopez JL, Cirulescu ARM, Davies DC, et al. The legal and ethical framework governing body donation in Europe—1st update on current practice. Eur J Anat. 2012;16(1):1–21. Riederer BM, Bolt S, Brenner E, Bueno-Lopez JL, Cirulescu ARM, Davies DC, et al. The legal and ethical framework governing body donation in Europe—1st update on current practice. Eur J Anat. 2012;16(1):1–21.
12.
Zurück zum Zitat Platzer W, Putz R, Poisel S. Ein neues Konservierungs- und Aufbewahrungssystem für anatomisches Material. Acta Anat (Basel). 1978;102(1):60–7.CrossRef Platzer W, Putz R, Poisel S. Ein neues Konservierungs- und Aufbewahrungssystem für anatomisches Material. Acta Anat (Basel). 1978;102(1):60–7.CrossRef
13.
Zurück zum Zitat Christner JA, Kofler JM, McCollough CH. Estimating effective dose for CT using dose–length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection Publication 103 or dual-energy scanning. AJR Am J Roentgenol. 2010;194(4):881–9.CrossRefPubMed Christner JA, Kofler JM, McCollough CH. Estimating effective dose for CT using dose–length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection Publication 103 or dual-energy scanning. AJR Am J Roentgenol. 2010;194(4):881–9.CrossRefPubMed
14.
Zurück zum Zitat Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.CrossRef Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327(8476):307–10.CrossRef
15.
Zurück zum Zitat European Union. Council Directive 2013/59/Euratom of 5 December 2013. Laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation, and repealing Directives 89/618/Euratom, 90/641/Euratom, 96/29/Euratom, 97/43/Euratom and 2003/122/Euratom. Off J Eur Union. 2014;17:13/1–13/73. European Union. Council Directive 2013/59/Euratom of 5 December 2013. Laying down basic safety standards for protection against the dangers arising from exposure to ionising radiation, and repealing Directives 89/618/Euratom, 90/641/Euratom, 96/29/Euratom, 97/43/Euratom and 2003/122/Euratom. Off J Eur Union. 2014;17:13/1–13/73.
16.
Zurück zum Zitat Widmann G, Bischel A, Stratis A, Bosmans H, Jacobs R, Gassner EM, et al. Spatial and contrast resolution of ultralow dose dentomaxillofacial CT imaging using iterative reconstruction technology. Dentomaxillofac Radiol. 2017;46(4):20160452.CrossRefPubMedPubMedCentral Widmann G, Bischel A, Stratis A, Bosmans H, Jacobs R, Gassner EM, et al. Spatial and contrast resolution of ultralow dose dentomaxillofacial CT imaging using iterative reconstruction technology. Dentomaxillofac Radiol. 2017;46(4):20160452.CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Al-Ekrish A, Alfaleh W, Hörmann R, Alabdulwahid A, Puelacher W, Widmann G. Localization of the inferior alveolar canal using ultralow dose CT with iterative reconstruction techniques. Dentomaxillofac Radiol. 2018;47:20170477.CrossRefPubMed Al-Ekrish A, Alfaleh W, Hörmann R, Alabdulwahid A, Puelacher W, Widmann G. Localization of the inferior alveolar canal using ultralow dose CT with iterative reconstruction techniques. Dentomaxillofac Radiol. 2018;47:20170477.CrossRefPubMed
18.
Zurück zum Zitat Cavalcanti MG, Haller JW, Vannier MW. Three-dimensional computed tomography landmark measurement in craniofacial surgical planning: experimental validation in vitro. J Oral Maxillofac Surg. 1999;57(6):690–4.CrossRefPubMed Cavalcanti MG, Haller JW, Vannier MW. Three-dimensional computed tomography landmark measurement in craniofacial surgical planning: experimental validation in vitro. J Oral Maxillofac Surg. 1999;57(6):690–4.CrossRefPubMed
19.
Zurück zum Zitat Yang J, Cavalcanti MG, Ruprecht A, Vannier MW. 2-D and 3-D reconstructions of spiral computed tomography in localization of the inferior alveolar canal for dental implants. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1999;87(3):369–74.CrossRefPubMed Yang J, Cavalcanti MG, Ruprecht A, Vannier MW. 2-D and 3-D reconstructions of spiral computed tomography in localization of the inferior alveolar canal for dental implants. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1999;87(3):369–74.CrossRefPubMed
20.
Zurück zum Zitat Kobayashi K, Shimoda S, Nakagawa Y, Yamamoto A. Accuracy in measurement of distance using limited cone-beam computerized tomography. Int J Oral Maxillofac Implants. 2004;19(2):228–31.PubMed Kobayashi K, Shimoda S, Nakagawa Y, Yamamoto A. Accuracy in measurement of distance using limited cone-beam computerized tomography. Int J Oral Maxillofac Implants. 2004;19(2):228–31.PubMed
21.
Zurück zum Zitat Al-Ekrish A, Ekram M. A comparative study of the accuracy and reliability of multidetector computed tomography and cone beam computed tomography in the assessment of dental implant site dimensions. Dentomaxillofac Radiol. 2011;40(2):67–75.CrossRefPubMedPubMedCentral Al-Ekrish A, Ekram M. A comparative study of the accuracy and reliability of multidetector computed tomography and cone beam computed tomography in the assessment of dental implant site dimensions. Dentomaxillofac Radiol. 2011;40(2):67–75.CrossRefPubMedPubMedCentral
22.
23.
Zurück zum Zitat Rustemeyer P, Streubühr U, Suttmoeller J. Low-dose dental computed tomography: significant dose reduction without loss of image quality. Acta Radiol. 2004;45(8):847–53.CrossRefPubMed Rustemeyer P, Streubühr U, Suttmoeller J. Low-dose dental computed tomography: significant dose reduction without loss of image quality. Acta Radiol. 2004;45(8):847–53.CrossRefPubMed
24.
Zurück zum Zitat Suomalainen A, Vehmas T, Kortesniemi M, Robinson S, Peltola J. Accuracy of linear measurements using dental cone beam and conventional multislice computed tomography. Dentomaxillofac Radiol. 2008;37(1):10–7.CrossRefPubMed Suomalainen A, Vehmas T, Kortesniemi M, Robinson S, Peltola J. Accuracy of linear measurements using dental cone beam and conventional multislice computed tomography. Dentomaxillofac Radiol. 2008;37(1):10–7.CrossRefPubMed
Metadaten
Titel
Comparability of dental implant site ridge measurements using ultra-low-dose multidetector row computed tomography combined with filtered back-projection, adaptive statistical iterative reconstruction, and model-based iterative reconstruction
verfasst von
Asma’a Abdurrahman Al-Ekrish
Reema Al-Shawaf
Wafa Alfaleh
Romed Hörmann
Wolfgang Puelacher
Gerlig Widmann
Publikationsdatum
13.10.2018
Verlag
Springer Singapore
Erschienen in
Oral Radiology / Ausgabe 3/2019
Print ISSN: 0911-6028
Elektronische ISSN: 1613-9674
DOI
https://doi.org/10.1007/s11282-018-0350-z

Weitere Artikel der Ausgabe 3/2019

Oral Radiology 3/2019 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.