Comparative analysis of iterative reconstruction algorithms with resolution recovery and time of flight modeling for 18F-FDG cardiac PET: A multi-center phantom study

Zeitschrift:
Journal of Nuclear Cardiology > Ausgabe 3/2017
Autoren:
PhD Roberta Matheoud, PhD Michela Lecchi, PhD Domenico Lizio, PhD Camilla Scabbio, MD Claudio Marcassa, MD Lucia Leva, MD Angelo del Sole, PhD Carlo Rodella, PhD Luca Indovina, PhD Christian Bracco, PhD Marco Brambilla, PhD Orazio Zoccarato
Wichtige Hinweise
An erratum to this article is available at http://​dx.​doi.​org/​10.​1007/​s12350-016-0415-5.

Abstract

Background

The purpose of this study was to evaluate the image quality in cardiac 18F-FDG PET using the time of flight (TOF) and/or point spread function (PSF) modeling in the iterative reconstruction (IR).

Methods

Three scanners and an anthropomorphic cardiac phantom with an insert simulating a transmural defect (TD) were used. Two sets of scans (with/without TD) were acquired, and four reconstruction schemes were considered: (1) IR; (2) IR + PSF, (3) IR + TOF, and (4) IR + TOF + PSF. LV wall thickness (FWHM), contrast between LV wall and inner chamber (C IC), and TD contrast in LV wall (C TD) were evaluated.

Results

Tests of the reconstruction protocols showed a decrease in FWHM from IR (13 mm) to IR + PSF (11 mm); an increase in the C IC from IR (65%) to IR + PSF (71%) and from IR + TOF (72%) to IR + TOF + PSF (77%); and an increase in the C TD from IR + PSF (72%) to IR + TOF (75%) and to IR + TOF + PSF (77%). Tests of the scanner/software combinations showed a decrease in FWHM from Gemini_TF (13 mm) to Biograph_mCT (12 mm) and to Discovery_690 (11 mm); an increase in the C IC from Gemini_TF (65%) to Biograph_mCT (73%) and to Discovery_690 (75%); and an increase in the C TD from Gemini_TF/Biograph_mCT (72%) to Discovery_690 (77%).

Conclusion

The introduction of TOF and PSF increases image quality in cardiac 18F-FDG PET. The scanner/software combinations exhibit different performances, which should be taken into consideration when making cross comparisons.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 3/2017

Journal of Nuclear Cardiology 3/2017 Zur Ausgabe
  1. Sie können e.Med Innere Medizin 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

  2. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.


 

Neu im Fachgebiet Kardiologie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Kardiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise