Comparative Cardiac Phantom Study Using Tc-99m/I-123 and Tl-201/I-123 Tracers with Cadmium-Zinc-Telluride Detector-Based Single-Photon Emission Computed Tomography

verfasst von: Takanaga Niimi, Mamoru Nanasato, Mitsuo Sugimoto, Hisatoshi Maeda

Erschienen in: Nuclear Medicine and Molecular Imaging | Ausgabe 1/2019

Einloggen, um Zugang zu erhalten

Abstract

Objective

A recently introduced single-photon emission computed tomography (SPECT), based on cadmium-zinc-telluride (CZT) detectors (D-SPECT), supports high energy resolution for cardiac imaging. Importantly, the high energy resolution may allow simultaneous dual-isotope (SDI) imaging (e.g., using Tc-99m and I-123). We quantitatively evaluated Tc-99m/I-123 SDI imaging by D-SPECT in comparison with conventional T1-201/I-123.

Materials and Methods

Energy resolution was measured as a percentage of the full width at half maximum (FWHM) for Tc-99m, I-123, and Tl-201. The impact of cross-talk and reconstructed image contrast were quantified by measuring the contrast-to-noise ratio (CNR), and the transmural defect contrast in the left ventricle wall (CTD) induced by a difference in energy, for combinations of Tc-99m/I-123 or Tl-201/I-123, using an RH-2 cardiac phantom. Corresponding measurement was also carried out in Anger SPECT (A-SPECT).

Results

The energy resolution of the D-SPECT system was 5.4%/5.1% for Tc-99m/I-123 and 5.4%/5.3% for Tl-201/I-123, which was approximately two times higher than the A-SPECT. No notable difference was confirmed in the CNRs of the two systems, but T1-201/I-123 showed overall higher value than Tc-99m/I-123. Compared to A-SPECT, CTD of D-SPECT significantly increased with both Tc-99m/I-123 and T1-201/I-123 (p < 0.05). In DSPECT, the combination of Tc-99m/I-123 had a slightly better CTD than T1-201/I-123. In addition, CTD of Tc-99m/I-123 was improved with scatter correction at both nuclides (p < 0.05), but in Tl-201/I-123, no significant improvement was confirmed in I-123 (p > 0.05).

Conclusion

D-SPECT was considered to be capable of performing high-quality SDI imaging using Tc-99m/I-123.
Literatur
1.
Zurück zum Zitat Garcia EV, Faber TL, Esteves FP. Cardiac dedicated ultrafast SPECT cameras: new designs and clinical implications. J Nucl Med. 2011;52:210–7.CrossRefPubMed Garcia EV, Faber TL, Esteves FP. Cardiac dedicated ultrafast SPECT cameras: new designs and clinical implications. J Nucl Med. 2011;52:210–7.CrossRefPubMed
2.
Zurück zum Zitat Neill J, Prvulovich EM, Fish MB, Berman DS, Slomka PJ, Sharir T, et al. Initial multicenter experience of high-speed myocardial perfusion imaging: comparison between high-speed and conventional single-photon emission computed tomography with angiographic validation. Eur J Nucl Med Mol Imaging. 2013;40:1084–94.CrossRefPubMed Neill J, Prvulovich EM, Fish MB, Berman DS, Slomka PJ, Sharir T, et al. Initial multicenter experience of high-speed myocardial perfusion imaging: comparison between high-speed and conventional single-photon emission computed tomography with angiographic validation. Eur J Nucl Med Mol Imaging. 2013;40:1084–94.CrossRefPubMed
3.
Zurück zum Zitat Duvall WL, Slomka PJ, Gerlach JR, Sweeny JM, Baber U, Croft LB, et al. High-efficiency SPECT MPI: comparison of automated quantification, visual interpretation, and coronary angiography. J Nucl Cardiol. 2013;20:763–73.CrossRefPubMed Duvall WL, Slomka PJ, Gerlach JR, Sweeny JM, Baber U, Croft LB, et al. High-efficiency SPECT MPI: comparison of automated quantification, visual interpretation, and coronary angiography. J Nucl Cardiol. 2013;20:763–73.CrossRefPubMed
4.
Zurück zum Zitat Ben-Haim S, Kacperski K, Hain S, Van Gramberg D, Hutton BF, Erlandsson K, et al. Simultaneous dual-radionuclide myocardial perfusion imaging with a solid-state dedicated cardiac camera. Eur J Nucl Med Mol Imaging. 2010;37:1710–21.CrossRefPubMedPubMedCentral Ben-Haim S, Kacperski K, Hain S, Van Gramberg D, Hutton BF, Erlandsson K, et al. Simultaneous dual-radionuclide myocardial perfusion imaging with a solid-state dedicated cardiac camera. Eur J Nucl Med Mol Imaging. 2010;37:1710–21.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Ko T, Utanohara Y, Suzuki Y, Kurihara M, Iguchi N, Umemura J, et al. A preliminary feasibility study of simultaneous dual-isotope imaging with a solid-state dedicated cardiac camera for evaluating myocardial perfusion and fatty acid metabolism. Heart Vessel. 2016;31:38–45.CrossRef Ko T, Utanohara Y, Suzuki Y, Kurihara M, Iguchi N, Umemura J, et al. A preliminary feasibility study of simultaneous dual-isotope imaging with a solid-state dedicated cardiac camera for evaluating myocardial perfusion and fatty acid metabolism. Heart Vessel. 2016;31:38–45.CrossRef
6.
Zurück zum Zitat Kobayashi M, Matsunari I, Nishi K, Mizutani A, Miyazaki Y, Ogai K, et al. Simultaneous acquisition of (99m)Tc- and (123)I-labeled radiotracers using a preclinical SPECT scanner with CZT detectors. Ann Nucl Med. 2016;30:263–71.CrossRefPubMed Kobayashi M, Matsunari I, Nishi K, Mizutani A, Miyazaki Y, Ogai K, et al. Simultaneous acquisition of (99m)Tc- and (123)I-labeled radiotracers using a preclinical SPECT scanner with CZT detectors. Ann Nucl Med. 2016;30:263–71.CrossRefPubMed
7.
Zurück zum Zitat Blaire T, Bailliez A, Bouallegue FB, Bellevre D, Agostini D, Manrique A. Left ventricular function assessment using 123I/99mTc dual-isotope acquisition with two semi-conductor cadmium–zinc–telluride (CZT) cameras: a gated cardiac phantom study. EJNMMI Phys. 2016;3:27–37.CrossRefPubMedPubMedCentral Blaire T, Bailliez A, Bouallegue FB, Bellevre D, Agostini D, Manrique A. Left ventricular function assessment using 123I/99mTc dual-isotope acquisition with two semi-conductor cadmium–zinc–telluride (CZT) cameras: a gated cardiac phantom study. EJNMMI Phys. 2016;3:27–37.CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Berger HJ, Gottschalk A, Zaret BL. Dual radionuclide study of acute myocardial infarction: comparison of thallium-201 and technetium-99m stannous prophosphate imaging in man. Ann Intern Med. 1978;88:145–54.CrossRefPubMed Berger HJ, Gottschalk A, Zaret BL. Dual radionuclide study of acute myocardial infarction: comparison of thallium-201 and technetium-99m stannous prophosphate imaging in man. Ann Intern Med. 1978;88:145–54.CrossRefPubMed
9.
Zurück zum Zitat Tsuji A, Kojima A, Matsumoto M, Oyama Y, Tomiguchi S, Kira T, et al. A new method for crosstalk correction in simultaneous dual-isotope myocardial imaging with Tl-201 and I-123. Ann Nucl Med. 1999;13:317–23.CrossRefPubMed Tsuji A, Kojima A, Matsumoto M, Oyama Y, Tomiguchi S, Kira T, et al. A new method for crosstalk correction in simultaneous dual-isotope myocardial imaging with Tl-201 and I-123. Ann Nucl Med. 1999;13:317–23.CrossRefPubMed
10.
Zurück zum Zitat Nakajima K, Shimizu K, Taki J, Uetani Y, Konishi S, Tonami N, et al. Utility of iodine-123-BMIPP in the diagnosis and follow-up of vasospastic angina. J NucI Med. 1995;36:1934–40. Nakajima K, Shimizu K, Taki J, Uetani Y, Konishi S, Tonami N, et al. Utility of iodine-123-BMIPP in the diagnosis and follow-up of vasospastic angina. J NucI Med. 1995;36:1934–40.
11.
Zurück zum Zitat Yamagishi H, Akioka K, Shirai N, Yoshiyama M, Teragaki M, Takeuchi K, et al. Effects of smoking on myocardial injury in patients with conservatively treated acute myocardial infarction—a study with resting 123I-15-iodophenyl 3-methyl pentadecanoic acid/201Tl myocardial single photon emission computed tomography. Jpn Circ J. 2001;65:769–74.CrossRefPubMed Yamagishi H, Akioka K, Shirai N, Yoshiyama M, Teragaki M, Takeuchi K, et al. Effects of smoking on myocardial injury in patients with conservatively treated acute myocardial infarction—a study with resting 123I-15-iodophenyl 3-methyl pentadecanoic acid/201Tl myocardial single photon emission computed tomography. Jpn Circ J. 2001;65:769–74.CrossRefPubMed
12.
Zurück zum Zitat Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ. Radiation dose to patients from cardiac diagnostic imaging. Circulation. 2007;116:1290–305.CrossRefPubMed Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ. Radiation dose to patients from cardiac diagnostic imaging. Circulation. 2007;116:1290–305.CrossRefPubMed
13.
Zurück zum Zitat Einstein AJ, Pascual TNB, Mercuri M, Karthikeyan G, Vitola JV, Mahmarian JJ, et al. Current worldwide nuclear cardiology practices and radiation exposure: results from the 65 country IAEA Nuclear Cardiology Protocols Cross-Sectional Study (INCAPS). Eur Heart J. 2015;36:1689–96.CrossRefPubMedPubMedCentral Einstein AJ, Pascual TNB, Mercuri M, Karthikeyan G, Vitola JV, Mahmarian JJ, et al. Current worldwide nuclear cardiology practices and radiation exposure: results from the 65 country IAEA Nuclear Cardiology Protocols Cross-Sectional Study (INCAPS). Eur Heart J. 2015;36:1689–96.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Gambhir SS, Berman DS, Ziffer J, Nagler M, Sandler M, Patton J, et al. A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera. J Nucl Med. 2009;50:635–43.CrossRefPubMed Gambhir SS, Berman DS, Ziffer J, Nagler M, Sandler M, Patton J, et al. A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera. J Nucl Med. 2009;50:635–43.CrossRefPubMed
15.
Zurück zum Zitat Erlandsson K, Kacperski K, van Gramberg D, Hutton BF. Performance evaluation of D-SPECT: a novel SPECT system for nuclear cardiology. Phys Med Biol. 2009;54:2635–49.CrossRefPubMed Erlandsson K, Kacperski K, van Gramberg D, Hutton BF. Performance evaluation of D-SPECT: a novel SPECT system for nuclear cardiology. Phys Med Biol. 2009;54:2635–49.CrossRefPubMed
16.
Zurück zum Zitat Verger A, Imbert L, Yagdigul Y, Fay R, Djaballah W, Rouzet F, et al. Factors affecting the myocardial activity acquired during exercise SPECT with a high-sensitivity cardiac CZT camera as compared with conventional Anger camera. Eur J Nucl Med Mol Imaging. 2014;41:522–8.CrossRefPubMed Verger A, Imbert L, Yagdigul Y, Fay R, Djaballah W, Rouzet F, et al. Factors affecting the myocardial activity acquired during exercise SPECT with a high-sensitivity cardiac CZT camera as compared with conventional Anger camera. Eur J Nucl Med Mol Imaging. 2014;41:522–8.CrossRefPubMed
17.
Zurück zum Zitat Park MA, Moore SC, Muller SP, McQuaid SJ, Kijewski MF. Performance of a high-sensitivity dedicated cardiac SPECT scanner for striatal uptake quantification in the brain based on analysis of projection data. Med Phys. 2013;40:042504.CrossRefPubMedPubMedCentral Park MA, Moore SC, Muller SP, McQuaid SJ, Kijewski MF. Performance of a high-sensitivity dedicated cardiac SPECT scanner for striatal uptake quantification in the brain based on analysis of projection data. Med Phys. 2013;40:042504.CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Kacperski K, Erlandsson K, Ben-Haim S, Hutton BF. Iterative deconvolution of simultaneous 99mTc and 201Tl projection data measured on a CdZnTe-based cardiac SPECT scanner. Phys Med Biol. 2011;56:1397–414.CrossRefPubMedPubMedCentral Kacperski K, Erlandsson K, Ben-Haim S, Hutton BF. Iterative deconvolution of simultaneous 99mTc and 201Tl projection data measured on a CdZnTe-based cardiac SPECT scanner. Phys Med Biol. 2011;56:1397–414.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Niimi T, Sugimoto M, Nanasato M, Maeda H. Evaluation of cadmium-zinc-telluride detector-based single-photon emission computed tomography for nuclear cardiology: a comparison with conventional Anger single-photon emission computed tomography. Nucl Med Mol Imaging. 2017;51:331–7.CrossRefPubMedPubMedCentral Niimi T, Sugimoto M, Nanasato M, Maeda H. Evaluation of cadmium-zinc-telluride detector-based single-photon emission computed tomography for nuclear cardiology: a comparison with conventional Anger single-photon emission computed tomography. Nucl Med Mol Imaging. 2017;51:331–7.CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Esquerre JP, Coca FJ, Martinez SJ, Guiraud RF. Prone decubitus: a solution to inferior wall attenuation in thallium-201 myocardial tomography. J Nucl Med. 1989;30:398–401.PubMed Esquerre JP, Coca FJ, Martinez SJ, Guiraud RF. Prone decubitus: a solution to inferior wall attenuation in thallium-201 myocardial tomography. J Nucl Med. 1989;30:398–401.PubMed
21.
Zurück zum Zitat Zoccarato O, Lizio D, Savi A, Indovina L, Scabbio C, Leva L, et al. Comparative analysis of cadmium-zincum-telluride cameras dedicated to myocardial perfusion SPECT: a phantom study. J Nucl Cardiol. 2016;23:885–93.CrossRefPubMed Zoccarato O, Lizio D, Savi A, Indovina L, Scabbio C, Leva L, et al. Comparative analysis of cadmium-zincum-telluride cameras dedicated to myocardial perfusion SPECT: a phantom study. J Nucl Cardiol. 2016;23:885–93.CrossRefPubMed
22.
Zurück zum Zitat Kadrmas DJ, Frey EC, Tsui BMW. Simultaneous technetium-99m/thallium-201 SPECT imaging with model-based compensation for cross-contaminating effects. Phys Med Biol. 1999;44:1843–60.CrossRefPubMedPubMedCentral Kadrmas DJ, Frey EC, Tsui BMW. Simultaneous technetium-99m/thallium-201 SPECT imaging with model-based compensation for cross-contaminating effects. Phys Med Biol. 1999;44:1843–60.CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Nakazato R, Berman DS, Hayes SW, Fish M, Padgett R, Xu Y, et al. Myocardial perfusion imaging with a solid-state camera: simulation of a very low dose imaging protocol. J Nucl Med. 2013;54:373–9.CrossRefPubMed Nakazato R, Berman DS, Hayes SW, Fish M, Padgett R, Xu Y, et al. Myocardial perfusion imaging with a solid-state camera: simulation of a very low dose imaging protocol. J Nucl Med. 2013;54:373–9.CrossRefPubMed
Metadaten
Titel
Comparative Cardiac Phantom Study Using Tc-99m/I-123 and Tl-201/I-123 Tracers with Cadmium-Zinc-Telluride Detector-Based Single-Photon Emission Computed Tomography
verfasst von
Takanaga Niimi
Mamoru Nanasato
Mitsuo Sugimoto
Hisatoshi Maeda
Publikationsdatum
12.12.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Nuclear Medicine and Molecular Imaging / Ausgabe 1/2019
Print ISSN: 1869-3474
Elektronische ISSN: 1869-3482
DOI
https://doi.org/10.1007/s13139-018-0559-0

Weitere Artikel der Ausgabe 1/2019

Nuclear Medicine and Molecular Imaging 1/2019 Zur Ausgabe