01.06.2005 | Pediatric Original

Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children

verfasst von: Nicole Nagdyman, Thilo Fleck, Stephan Schubert, Peter Ewert, Björn Peters, Peter E. Lange, Hashim Abdul-Khaliq

Erschienen in: Intensive Care Medicine | Ausgabe 6/2005

Einloggen, um Zugang zu erhalten

Abstract

Objective

To compare the cerebral tissue oxygenation index (TOI) measured by near-infrared spectroscopy (NIRS) with venous oxygen saturation in the jugular bulb (SjO2) during elective cardiac catheterization in children.

Design and setting

Prospective observational clinical study in a catheterization laboratory for pediatric cardiology.

Patients

Sixty children with congenital heart defects admitted to the catheterization laboratory.

Measurements and results

TOI measured noninvasively by NIRS was compared to SjO2 measured in the jugular bulb during cardiac catheterization. Patients were divided into two groups regarding body weight: below (n=29) and above 10 kg (n=31). Linear regression analysis and Pearson’s correlation coefficient were calculated. Bland-Altman analysis, sensitivity, and specificity calculation for spatially resolved near-infrared spectroscopy with a cutoff level of 60% were performed. Simultaneously measured values for SjO2 (67.3±9.8%, 40–84.1%) and TOI (65.7±7.2%, 39–80%) showed a significant correlation; the correlation in children weighing under 10 kg was stronger in children weighing over 10 kg. Bland-Altman analysis showed a mean bias of −1.8% with limits of agreement between 11.7% and −15.3% for all children. Sensitivity and specificity of the SRS method were 46% and 91%, respectively, for all children and 53% and 83% respectively in infants weighing under 10 kg.

Conclusions

The results demonstrate that despite a significant correlation, sensitivity of spatially resolved spectroscopy is poor, and it is questionable whether TOI can be used reliably to detect low SjO2.
Literatur
1.
Zurück zum Zitat Jonas RA (1998) Neurological protection during cardiopulmonary bypass/deep hypothermia. Pediatr Cardiol 19:321–330 Jonas RA (1998) Neurological protection during cardiopulmonary bypass/deep hypothermia. Pediatr Cardiol 19:321–330
2.
Zurück zum Zitat Nagdyman N, Fleck TP, Ewert P, Abdul-Khaliq H, Redlin M, Lange PE (2003) Cerebral oxygenation measured by near-infrared spectroscopy during circulatory arrest and cardiopulmonary resuscitation. Br J Anaesth 91:438–442 Nagdyman N, Fleck TP, Ewert P, Abdul-Khaliq H, Redlin M, Lange PE (2003) Cerebral oxygenation measured by near-infrared spectroscopy during circulatory arrest and cardiopulmonary resuscitation. Br J Anaesth 91:438–442
3.
Zurück zum Zitat Wypij D, Newburger JW, Rappaport LA, duPlessis AJ, Jonas RA, Wernovsky G, Lin M, Bellinger DC (2003) The effect of duration of deep hypothermic circulatory arrest in infant heart surgery on late neurodevelopment: the Boston Circulatory Arrest Trial. J Thorac Cardiovasc Surg 126:1397–1403 Wypij D, Newburger JW, Rappaport LA, duPlessis AJ, Jonas RA, Wernovsky G, Lin M, Bellinger DC (2003) The effect of duration of deep hypothermic circulatory arrest in infant heart surgery on late neurodevelopment: the Boston Circulatory Arrest Trial. J Thorac Cardiovasc Surg 126:1397–1403
4.
Zurück zum Zitat Dittrich H, Buhrer C, Grimmer I, Dittrich S, Abdul-Khaliq H, Lange PE (2003) Neurodevelopment at 1 year of age in infants with congenital heart disease. Heart 89:436–441 Dittrich H, Buhrer C, Grimmer I, Dittrich S, Abdul-Khaliq H, Lange PE (2003) Neurodevelopment at 1 year of age in infants with congenital heart disease. Heart 89:436–441
5.
Zurück zum Zitat Cruz J (1998) The first decade of continuous monitoring of jugular bulb oxyhemoglobinsaturation: management strategies and clinical outcome. Crit Care Med 26:344–351 Cruz J (1998) The first decade of continuous monitoring of jugular bulb oxyhemoglobinsaturation: management strategies and clinical outcome. Crit Care Med 26:344–351
6.
Zurück zum Zitat Gayle MO, Frewen TC, Armstrong RF, Gilbert JJ, Kronick JB, Kissoon N, Lee R, Tiffin N, Brown T (1989) Jugular venous bulb catheterization in infants and children. Crit Care Med 17:385–388 Gayle MO, Frewen TC, Armstrong RF, Gilbert JJ, Kronick JB, Kissoon N, Lee R, Tiffin N, Brown T (1989) Jugular venous bulb catheterization in infants and children. Crit Care Med 17:385–388
7.
Zurück zum Zitat Jobsis FF (1977) Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science 198:1264–1267PubMed Jobsis FF (1977) Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science 198:1264–1267PubMed
8.
Zurück zum Zitat Wahr JA, Tremper KK, Samra S, Delpy DT (1996) Near-infrared spectroscopy: theory and applications. J Cardiothorac Vasc Anesth 10:406–418PubMed Wahr JA, Tremper KK, Samra S, Delpy DT (1996) Near-infrared spectroscopy: theory and applications. J Cardiothorac Vasc Anesth 10:406–418PubMed
9.
Zurück zum Zitat Adcock LM, Wafelman LS, Hegemier S, Moise AA, Speer ME, Contant CF, Goddard-Finegold J (1999) Neonatal intensive care applications of near-infrared spectroscopy. Clin Perinatol 26:893–903 Adcock LM, Wafelman LS, Hegemier S, Moise AA, Speer ME, Contant CF, Goddard-Finegold J (1999) Neonatal intensive care applications of near-infrared spectroscopy. Clin Perinatol 26:893–903
10.
Zurück zum Zitat Schulz G, Weiss M, Bauersfeld U, Teller J, Haensse D, Bucher HU, Baenziger O (2002) Liver tissue oxygenation as measured by near-infrared spectroscopy in the critically ill child in correlation with central venous oxygen saturation. Intensive Care Med 28:184–189CrossRefPubMed Schulz G, Weiss M, Bauersfeld U, Teller J, Haensse D, Bucher HU, Baenziger O (2002) Liver tissue oxygenation as measured by near-infrared spectroscopy in the critically ill child in correlation with central venous oxygen saturation. Intensive Care Med 28:184–189CrossRefPubMed
11.
Zurück zum Zitat Fortune PM, Wagstaff M, Petros AJ (2001) Cerebro-splanchnic oxygenation ratio (CSOR) using near infrared spectroscopy may be able to predict splanchnic ischaemia in neonates. Intensive Care Med 27:1401–1407 Fortune PM, Wagstaff M, Petros AJ (2001) Cerebro-splanchnic oxygenation ratio (CSOR) using near infrared spectroscopy may be able to predict splanchnic ischaemia in neonates. Intensive Care Med 27:1401–1407
12.
Zurück zum Zitat Abdul-Khaliq H, Troitzsch D, Schubert S, Wehsack A, Bottcher W, Gutsch E, Hubler M, Hetzer R, Lange PE (2002) Cerebral oxygen monitoring during neonatal cardiopulmonary bypass and deep hypothermic circulatory arrest. Thorac Cardiovasc Surg 50:77–81 Abdul-Khaliq H, Troitzsch D, Schubert S, Wehsack A, Bottcher W, Gutsch E, Hubler M, Hetzer R, Lange PE (2002) Cerebral oxygen monitoring during neonatal cardiopulmonary bypass and deep hypothermic circulatory arrest. Thorac Cardiovasc Surg 50:77–81
13.
Zurück zum Zitat Kurth CD, Steven JM, Nicolson SC, Jacobs ML (1997) cerebral oxygenation during cardiopulmonary bypass in children. J Thorac Cardiovasc Surg 113:71–79 Kurth CD, Steven JM, Nicolson SC, Jacobs ML (1997) cerebral oxygenation during cardiopulmonary bypass in children. J Thorac Cardiovasc Surg 113:71–79
14.
Zurück zum Zitat Nollert G, Jonas JA, Reichart B (2000) Optimizing cerebral oxygenation during cardiac surgery: a review of experimental and clinical investigations with near infrared spectrophometry. Thorac Cardiovasc Surg 48:247–253 Nollert G, Jonas JA, Reichart B (2000) Optimizing cerebral oxygenation during cardiac surgery: a review of experimental and clinical investigations with near infrared spectrophometry. Thorac Cardiovasc Surg 48:247–253
15.
Zurück zum Zitat Fallon P, Rboberts I, Kirkham FJ (1993) Cerebral hemodynamics during cardiopulmonary bypass in children using near-infrared spectroscopy. Ann Thorac Surg 56:1473–1477PubMed Fallon P, Rboberts I, Kirkham FJ (1993) Cerebral hemodynamics during cardiopulmonary bypass in children using near-infrared spectroscopy. Ann Thorac Surg 56:1473–1477PubMed
16.
Zurück zum Zitat Thavasothy M, Broadhead M, Elwell C, Peters M, Smith M (2002) A comparison of cerebral oxygenation as measured by the NIRO 300 and the INVOS 5100 Near-Infrared Spectrophotometers. Anaesthesia 57:999–1006CrossRefPubMed Thavasothy M, Broadhead M, Elwell C, Peters M, Smith M (2002) A comparison of cerebral oxygenation as measured by the NIRO 300 and the INVOS 5100 Near-Infrared Spectrophotometers. Anaesthesia 57:999–1006CrossRefPubMed
17.
Zurück zum Zitat Matcher SJ, Kirkpatrick P, Nahid K, Cope M, Delpy DT (1995) Absolute quantification methods in tissue near infrared spectroscopy. Proc SPIE 2389:486–495 Matcher SJ, Kirkpatrick P, Nahid K, Cope M, Delpy DT (1995) Absolute quantification methods in tissue near infrared spectroscopy. Proc SPIE 2389:486–495
18.
Zurück zum Zitat Suzuki S, Takasaki S, Ozaki T, Kobayashi Y (1999) A tissue oxygenation monitor using NIR spatially resolved spectroscopy. Proc SPIE 3579:486–495 Suzuki S, Takasaki S, Ozaki T, Kobayashi Y (1999) A tissue oxygenation monitor using NIR spatially resolved spectroscopy. Proc SPIE 3579:486–495
19.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310 Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310
20.
Zurück zum Zitat Brown R, Wright G, Royston D (1993) A comparison of two systems for assessing cerebral venous oxyhaemoglobin saturation during cardiopulmonary bypass in humans. Anaesthesia 48:697–700PubMed Brown R, Wright G, Royston D (1993) A comparison of two systems for assessing cerebral venous oxyhaemoglobin saturation during cardiopulmonary bypass in humans. Anaesthesia 48:697–700PubMed
21.
Zurück zum Zitat Daubeney PE, Pilkington SN, Janke E, Charlton GA, Smith DC, Webber SA (1996) Cerebral oxygenation measured by near-infrared spectroscopy: comparison with jugular bulb oximetry. Ann Thorac Surg 61:930–934CrossRefPubMed Daubeney PE, Pilkington SN, Janke E, Charlton GA, Smith DC, Webber SA (1996) Cerebral oxygenation measured by near-infrared spectroscopy: comparison with jugular bulb oximetry. Ann Thorac Surg 61:930–934CrossRefPubMed
22.
Zurück zum Zitat Abdul-Khaliq H, Weipert T, Alexi-Meskhishvili V, Kuppe H, Hetzer R, Lange PE (1998) Perioperative Überwachung der zerebralen Oxygenierung mit Hilfe der Nahinfrarotspektroskopie in der pädiatrischen Herzchirurgie. Z Herz Thorax Gefasschir 12:65–72 Abdul-Khaliq H, Weipert T, Alexi-Meskhishvili V, Kuppe H, Hetzer R, Lange PE (1998) Perioperative Überwachung der zerebralen Oxygenierung mit Hilfe der Nahinfrarotspektroskopie in der pädiatrischen Herzchirurgie. Z Herz Thorax Gefasschir 12:65–72
23.
Zurück zum Zitat Ter Minassian A, Poirier N, Pierrot M, Menei P, Grany JC, Ursino M, Beydon L (1999) Correlation between cerebral oxygen saturation measured by near-infrared spectroscopy and jugular oxygen saturation in patients with severe closed head injury. Anesthesiology 91:985–990 Ter Minassian A, Poirier N, Pierrot M, Menei P, Grany JC, Ursino M, Beydon L (1999) Correlation between cerebral oxygen saturation measured by near-infrared spectroscopy and jugular oxygen saturation in patients with severe closed head injury. Anesthesiology 91:985–990
24.
Zurück zum Zitat Lewis SB, Myburgh JA, Thornton EL, Reilly PL (1996) Cerebral oxygenation monitoring by near-infrared spectroscopy is not clinically useful in patients with severe closed-head injury: a comparison with jugular venous bulb oximetry. Crit Care Med 24:1334–1338CrossRefPubMed Lewis SB, Myburgh JA, Thornton EL, Reilly PL (1996) Cerebral oxygenation monitoring by near-infrared spectroscopy is not clinically useful in patients with severe closed-head injury: a comparison with jugular venous bulb oximetry. Crit Care Med 24:1334–1338CrossRefPubMed
25.
Zurück zum Zitat Ali MS, Harmer M, Vaughan RS, Dunne JA, Latto IP (2001) Spatially resolved spectroscopy (NIRO-300) does not agree with jugular bulb oxygen saturation in patients undergoing warm bypass surgery. Can J Anaesth 48:497–501PubMed Ali MS, Harmer M, Vaughan RS, Dunne JA, Latto IP (2001) Spatially resolved spectroscopy (NIRO-300) does not agree with jugular bulb oxygen saturation in patients undergoing warm bypass surgery. Can J Anaesth 48:497–501PubMed
26.
Zurück zum Zitat Wickramasinghe YA, Palmer KS, Houston R, Spencer SA, Thorniley MS, Oeseburg B, Colier W (1993) Effect of fetal hemoglobin on the determination of neonatal cerebral oxygenation by near-infrared spectroscopy. Pediatr Res 34:15–17 Wickramasinghe YA, Palmer KS, Houston R, Spencer SA, Thorniley MS, Oeseburg B, Colier W (1993) Effect of fetal hemoglobin on the determination of neonatal cerebral oxygenation by near-infrared spectroscopy. Pediatr Res 34:15–17
27.
Zurück zum Zitat Germon TJ, Young AE, Manara AR, Nelson RJ (1995) Extracerebral absorption of near infrared light influences the detection of increased cerebral oxygenation monitored by near infrared spectroscopy. J Neurol Neurosurg Psychiatry 58:477–479PubMed Germon TJ, Young AE, Manara AR, Nelson RJ (1995) Extracerebral absorption of near infrared light influences the detection of increased cerebral oxygenation monitored by near infrared spectroscopy. J Neurol Neurosurg Psychiatry 58:477–479PubMed
28.
Zurück zum Zitat Fearn SJ, Chant HJ, Picton AJ, Mortimer AJ, McCollum CN (1997) The contribution of extracranial blood oxygenation on near-infrared spectroscopy during carotid endarterectomy. Anaesthesia 52:704–705 Fearn SJ, Chant HJ, Picton AJ, Mortimer AJ, McCollum CN (1997) The contribution of extracranial blood oxygenation on near-infrared spectroscopy during carotid endarterectomy. Anaesthesia 52:704–705
29.
Zurück zum Zitat Chan MT, Lam JM, Gomersall CD (1997) The contribution of extracranial blood oxygenation on near infrared spectroscopy during carotid endarterectomy. Anaesthesia 52:705–706 Chan MT, Lam JM, Gomersall CD (1997) The contribution of extracranial blood oxygenation on near infrared spectroscopy during carotid endarterectomy. Anaesthesia 52:705–706
30.
Zurück zum Zitat Al-Rawi PG, Smielewski P, Kirkpatrick PJ (2001) Evaluation of a near-infrared spectrometer (NIRO 300) for the detection of intracranial oxygenation changes in the adult head. Stroke 32:2492–500PubMed Al-Rawi PG, Smielewski P, Kirkpatrick PJ (2001) Evaluation of a near-infrared spectrometer (NIRO 300) for the detection of intracranial oxygenation changes in the adult head. Stroke 32:2492–500PubMed
31.
Zurück zum Zitat Buunk G, van der Hoeven JG, Meinders AE (1998) A comparison of near-infrared spectroscopy and jugular bulb oximetry in comatose patients resuscitated from a cardiac arrest. Anaesthesia 53:13–19 Buunk G, van der Hoeven JG, Meinders AE (1998) A comparison of near-infrared spectroscopy and jugular bulb oximetry in comatose patients resuscitated from a cardiac arrest. Anaesthesia 53:13–19
32.
Zurück zum Zitat Germon TJ, Kane NM, Manara AR, Nelson RJ (1994) Near-infrared spectroscopy in adults: effects of extracranial ischaemia and intracranial hypoxia on estimation of cerebral oxygenation. Br J Anaesth 73:503–506PubMed Germon TJ, Kane NM, Manara AR, Nelson RJ (1994) Near-infrared spectroscopy in adults: effects of extracranial ischaemia and intracranial hypoxia on estimation of cerebral oxygenation. Br J Anaesth 73:503–506PubMed
Metadaten
Titel
Comparison between cerebral tissue oxygenation index measured by near-infrared spectroscopy and venous jugular bulb saturation in children
verfasst von
Nicole Nagdyman
Thilo Fleck
Stephan Schubert
Peter Ewert
Björn Peters
Peter E. Lange
Hashim Abdul-Khaliq
Publikationsdatum
01.06.2005
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 6/2005
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-005-2618-0

Weitere Artikel der Ausgabe 6/2005

Intensive Care Medicine 6/2005 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.