Objective

The aim of this study was to compare the cyclic fatigue resistance at body temperature and phase transformation behaviors of novel Rotate instrument (25.06) with rotating Mtwo (25.06) and reciprocating Reciproc Blue (25.08) and Reciproc (25.08) instruments.

Materials and methods

The Rotate, Reciproc Blue, Reciproc, and Mtwo instruments free of visible deformations were collected and tested in a static cyclic fatigue test method, which has a ceramic block containing an artificial canal with 60° angle of curvature and a 5-mm radius of curvature at 37 °C (n = 16). All instruments were operated until fracture occurred, and both time to fracture (TF) and the lengths of the fractured fragments were recorded. TF data was analyzed with one-way ANOVA followed by post hoc Tukey tests and Weibull analysis, and fractured fragment length data were subjected to one-way ANOVA and post hoc Tukey tests (P < 0.05). Two unused instruments from each brand were also subjected to differential scanning calorimetry (DSC) analysis to determine their phase transformation temperatures.

Results

The Reciproc Blue instruments showed significantly higher TF values and reliability than the other groups (P < 0.05). Rotate instruments exhibited greater cyclic fatigue resistance than the Reciproc and Mtwo instruments (P < 0.05). No significant difference was detected among the fractured fragment lengths (P > 0.05). The lowest austenite finish temperature was exhibited by the Mtwo, which presented a single transformation peak, followed by the Rotate, Reciproc Blue, and Reciproc instruments which all presented two peaks during transformation.

Conclusions

Cyclic fatigue resistance of instruments manufactured from thermally treated Blue wire instruments was superior to those of the Mtwo and Reciproc, whereas reciprocating the Blue wire showed the highest resistance.

Clinical relevance

The present study compared the cyclic fatigue resistance of the novel Rotate instrument with similar instruments manufactured from conventional nickel–titanium, m-wire, and Blue wire at body temperature and reported that Blue-treated instruments exhibited superior cyclic fatigue resistance.
Literatur
1.
Zurück zum Zitat McGuigan M, Louca C, Duncan H (2013) Endodontic instrument fracture: causes and prevention. Br Dent J 214:341–348CrossRef McGuigan M, Louca C, Duncan H (2013) Endodontic instrument fracture: causes and prevention. Br Dent J 214:341–348CrossRef
2.
Zurück zum Zitat Pedullà E, Lo Savio F, Boninelli S et al (2003) Influence of cyclic torsional preloading on cyclic fatigue resistance of nickel–titanium instruments. Int Endod J 48:1043–1050CrossRef Pedullà E, Lo Savio F, Boninelli S et al (2003) Influence of cyclic torsional preloading on cyclic fatigue resistance of nickel–titanium instruments. Int Endod J 48:1043–1050CrossRef
3.
Zurück zum Zitat Martin B, Zelada G, Varela P, Bahillo JG, Magan F, Ahn S, Rodriguez C (2003) Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J 36:262–266CrossRef Martin B, Zelada G, Varela P, Bahillo JG, Magan F, Ahn S, Rodriguez C (2003) Factors influencing the fracture of nickel-titanium rotary instruments. Int Endod J 36:262–266CrossRef
4.
Zurück zum Zitat Shen Y, Cheung GS, Bian Z, Peng B (2006) Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 32:61–65CrossRef Shen Y, Cheung GS, Bian Z, Peng B (2006) Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 32:61–65CrossRef
5.
Zurück zum Zitat Cheung GS (2007) Instrument fracture: mechanisms, removal of fragments, and clinical outcomes. Endod Top 16:1–26CrossRef Cheung GS (2007) Instrument fracture: mechanisms, removal of fragments, and clinical outcomes. Endod Top 16:1–26CrossRef
6.
Zurück zum Zitat Kitchens GG, Liewehr FR, Moon PC (2007) The effect of operational speed on the fracture of nickel-titanium rotary instruments. J Endod 33:52–54CrossRef Kitchens GG, Liewehr FR, Moon PC (2007) The effect of operational speed on the fracture of nickel-titanium rotary instruments. J Endod 33:52–54CrossRef
7.
Zurück zum Zitat Bui TB, Mitchell JC, Baumgartner JC (2008) Effect of electropolishing ProFile nickel–titanium rotary instruments on cyclic fatigue resistance, torsional resistance, and cutting efficiency. J Endod 34:190–193CrossRef Bui TB, Mitchell JC, Baumgartner JC (2008) Effect of electropolishing ProFile nickel–titanium rotary instruments on cyclic fatigue resistance, torsional resistance, and cutting efficiency. J Endod 34:190–193CrossRef
8.
Zurück zum Zitat Ounsi HF, Al-Shalan T, Salameh Z et al (2008) Quantitative and qualitative elemental analysis of different nickel–titanium rotary instruments by using scanning electron microscopy and energy dispersive spectroscopy. J Endod 34:53–55CrossRef Ounsi HF, Al-Shalan T, Salameh Z et al (2008) Quantitative and qualitative elemental analysis of different nickel–titanium rotary instruments by using scanning electron microscopy and energy dispersive spectroscopy. J Endod 34:53–55CrossRef
9.
Zurück zum Zitat de Vasconcelos RA, Murphy S, Carvalho CAT, Govindjee RG, Govindjee S, Peters OA (2016) Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature. J Endod 42:782–787CrossRef de Vasconcelos RA, Murphy S, Carvalho CAT, Govindjee RG, Govindjee S, Peters OA (2016) Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature. J Endod 42:782–787CrossRef
10.
Zurück zum Zitat Dosanjh A, Paurazas S, Askar M (2017) The effect of temperature on cyclic fatigue of nickel-titanium rotary endodontic instruments. J Endod 43:823–826CrossRef Dosanjh A, Paurazas S, Askar M (2017) The effect of temperature on cyclic fatigue of nickel-titanium rotary endodontic instruments. J Endod 43:823–826CrossRef
11.
Zurück zum Zitat Castelló-Escrivá R, Alegre-Domingo T, Faus-Matoses V, Román-Richon S, Faus-Llácer VJ (2012) In vitro comparison of cyclic fatigue resistance of ProTaper, WaveOne, and Twisted Files. J Endod 38:1521–1524CrossRef Castelló-Escrivá R, Alegre-Domingo T, Faus-Matoses V, Román-Richon S, Faus-Llácer VJ (2012) In vitro comparison of cyclic fatigue resistance of ProTaper, WaveOne, and Twisted Files. J Endod 38:1521–1524CrossRef
12.
Zurück zum Zitat De-Deus G, Moreira E, Lopes H, Elias C (2010) Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J 43:1063–1068CrossRef De-Deus G, Moreira E, Lopes H, Elias C (2010) Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J 43:1063–1068CrossRef
13.
Zurück zum Zitat Yared G (2008) Canal preparation using only one Ni-Ti rotary instrument: preliminary observations. Int Endod J 41:339–344CrossRef Yared G (2008) Canal preparation using only one Ni-Ti rotary instrument: preliminary observations. Int Endod J 41:339–344CrossRef
14.
Zurück zum Zitat Zupanc J, Vahdat-Pajouh N, Schäfer E (2018) New thermomechanically treated NiTi alloys–a review. Int Endod J 51:1088–1103CrossRef Zupanc J, Vahdat-Pajouh N, Schäfer E (2018) New thermomechanically treated NiTi alloys–a review. Int Endod J 51:1088–1103CrossRef
15.
Zurück zum Zitat De-Deus G, Silva EJNL, Vieira VTL et al (2017) Blue thermomechanical treatment optimizes fatigue resistance and flexibility of the Reciproc files. J Endod 43:462–466CrossRef De-Deus G, Silva EJNL, Vieira VTL et al (2017) Blue thermomechanical treatment optimizes fatigue resistance and flexibility of the Reciproc files. J Endod 43:462–466CrossRef
16.
Zurück zum Zitat Inan U, Keskin C, Yilmaz ÖS et al (2019) Cyclic fatigue of Reciproc Blue and Reciproc instruments exposed to intracanal temperature in simulated severe apical curvature. Clin Oral Investig 23:2077–2082CrossRef Inan U, Keskin C, Yilmaz ÖS et al (2019) Cyclic fatigue of Reciproc Blue and Reciproc instruments exposed to intracanal temperature in simulated severe apical curvature. Clin Oral Investig 23:2077–2082CrossRef
17.
Zurück zum Zitat VDW Rotate Brochure, VDW Dental (2019) Last accessed from https://www.vdw-dental.com/en/service/brochure-downloads/ VDW Rotate Brochure, VDW Dental (2019) Last accessed from https://​www.​vdw-dental.​com/​en/​service/​brochure-downloads/​
18.
Zurück zum Zitat Gavini G, Santos MD, Caldeira CL et al (2018) Nickel–titanium instruments in endodontics: a concise review of the state of the art. Braz Oral Res 32:44–65CrossRef Gavini G, Santos MD, Caldeira CL et al (2018) Nickel–titanium instruments in endodontics: a concise review of the state of the art. Braz Oral Res 32:44–65CrossRef
19.
Zurück zum Zitat Ertuğrul IF, Orhan EO (2019) Cyclic fatigue and energy-dispersive X-ray spectroscopy examination of the novel ROTATE instrument. Microsc Res Tech 1–7 Ertuğrul IF, Orhan EO (2019) Cyclic fatigue and energy-dispersive X-ray spectroscopy examination of the novel ROTATE instrument. Microsc Res Tech 1–7
20.
Zurück zum Zitat Weibull W (1951) A statistical distribution function of wide applicability. J Appl Mech 18:293–297 Weibull W (1951) A statistical distribution function of wide applicability. J Appl Mech 18:293–297
21.
Zurück zum Zitat Nguyen HH, Fong H, Paranjpe A, Flake NM, Johnson JD, Peters OA (2014) Evaluation of the resistance to cyclic fatigue among ProTaper Next, ProTaper Universal, and Vortex Blue rotary instruments. J Endod 40:1190–1193CrossRef Nguyen HH, Fong H, Paranjpe A, Flake NM, Johnson JD, Peters OA (2014) Evaluation of the resistance to cyclic fatigue among ProTaper Next, ProTaper Universal, and Vortex Blue rotary instruments. J Endod 40:1190–1193CrossRef
22.
Zurück zum Zitat Arias A, Hejlawy S, Murphy S, de la Macorra JC, Govindjee S, Peters OA (2019) Variable impact by ambient temperature on fatigue resistance of heat-treated nickel titanium instruments. Clin Oral Investig 23:1101–1108CrossRef Arias A, Hejlawy S, Murphy S, de la Macorra JC, Govindjee S, Peters OA (2019) Variable impact by ambient temperature on fatigue resistance of heat-treated nickel titanium instruments. Clin Oral Investig 23:1101–1108CrossRef
23.
Zurück zum Zitat Huang X, Shen Y, Wei X, Haapasalo M (2017) Fatigue resistance of nickel-titanium instruments exposed to high-concentration hypochlorite. J Endod 43:1847–1851CrossRef Huang X, Shen Y, Wei X, Haapasalo M (2017) Fatigue resistance of nickel-titanium instruments exposed to high-concentration hypochlorite. J Endod 43:1847–1851CrossRef
24.
Zurück zum Zitat de Melo MCC, de Azevedo Bahia MG, Buono VTL (2002) Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments. J Endod 28:765–769CrossRef de Melo MCC, de Azevedo Bahia MG, Buono VTL (2002) Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments. J Endod 28:765–769CrossRef
25.
Zurück zum Zitat Cheung G, Darvell B (2007) Low-cycle fatigue of NiTi rotary instruments of various cross-sectional shapes. Int Endod J 40:626–632CrossRef Cheung G, Darvell B (2007) Low-cycle fatigue of NiTi rotary instruments of various cross-sectional shapes. Int Endod J 40:626–632CrossRef
26.
Zurück zum Zitat Scott R, Arias A, Macorra JC, Govindjee S, Peters OA (2019) Resistance to cyclic fatigue of reciprocating instruments determined at body temperature and phase transformation analysis. Aust Endod J 45:400–406CrossRef Scott R, Arias A, Macorra JC, Govindjee S, Peters OA (2019) Resistance to cyclic fatigue of reciprocating instruments determined at body temperature and phase transformation analysis. Aust Endod J 45:400–406CrossRef
27.
Zurück zum Zitat Bouska J, Justman B, Williamson A, DeLong C, Qian F (2012) Resistance to cyclic fatigue failure of a new endodontic rotary file. J Endod 38:667–669CrossRef Bouska J, Justman B, Williamson A, DeLong C, Qian F (2012) Resistance to cyclic fatigue failure of a new endodontic rotary file. J Endod 38:667–669CrossRef
28.
Zurück zum Zitat Elnaghy A (2014) Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files. Int Endod J 47:1034–1039CrossRef Elnaghy A (2014) Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary files. Int Endod J 47:1034–1039CrossRef
29.
Zurück zum Zitat Elnaghy AM, Elsaka SE (2018) Cyclic fatigue resistance of One Curve, 2Shape, ProFile Vortex, Vortex Blue, and RaCe nickel-titanium rotary instruments in single and double curvature canals. J Endod 44:1725–1730CrossRef Elnaghy AM, Elsaka SE (2018) Cyclic fatigue resistance of One Curve, 2Shape, ProFile Vortex, Vortex Blue, and RaCe nickel-titanium rotary instruments in single and double curvature canals. J Endod 44:1725–1730CrossRef
30.
Zurück zum Zitat Lopes HP, Elias CN, Vieira MV et al (2013) Fatigue life of Reciproc and Mtwo instruments subjected to static and dynamic tests. J Endod 39:693–696CrossRef Lopes HP, Elias CN, Vieira MV et al (2013) Fatigue life of Reciproc and Mtwo instruments subjected to static and dynamic tests. J Endod 39:693–696CrossRef
31.
Zurück zum Zitat Pedullà E, Grande NM, Plotino G, Gambarini G, Rapisarda E (2013) Influence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments. J Endod 39:258–261CrossRef Pedullà E, Grande NM, Plotino G, Gambarini G, Rapisarda E (2013) Influence of continuous or reciprocating motion on cyclic fatigue resistance of 4 different nickel-titanium rotary instruments. J Endod 39:258–261CrossRef
32.
Zurück zum Zitat Vadhana S, SaravanaKarthikeyan B, Nandini S, Velmurugan N (2014) Cyclic fatigue resistance of RaCe and Mtwo rotary files in continuous rotation and reciprocating motion. J Endod 40:995–999CrossRef Vadhana S, SaravanaKarthikeyan B, Nandini S, Velmurugan N (2014) Cyclic fatigue resistance of RaCe and Mtwo rotary files in continuous rotation and reciprocating motion. J Endod 40:995–999CrossRef
33.
Zurück zum Zitat Thompson S (2000) An overview of nickel–titanium alloys used in dentistry. Int Endod J 33:297–310CrossRef Thompson S (2000) An overview of nickel–titanium alloys used in dentistry. Int Endod J 33:297–310CrossRef
Metadaten
Titel
Comparison of cyclic fatigue resistance of Rotate instrument with reciprocating and continuous rotary nickel–titanium instruments at body temperature in relation to their transformation temperatures
verfasst von
Cangül Keskin
Özlem Sivas Yilmaz
Ali Keleş
Uğur Inan
Publikationsdatum
15.06.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 1/2021
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-020-03346-w

Weitere Artikel der Ausgabe 1/2021

Clinical Oral Investigations 1/2021 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.