Comparison of full-iodine conventional CT and half-iodine virtual monochromatic imaging: advantages and disadvantages

verfasst von: Haruto Sugawara, Shigeru Suzuki, Yoshiaki Katada, Takuya Ishikawa, Rika Fukui, Yuzo Yamamoto, Osamu Abe

Erschienen in: European Radiology | Ausgabe 3/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare image quality of abdominal arteries between full-iodine-dose conventional CT and half-iodine-dose virtual monochromatic imaging (VMI).

Materials and methods

We retrospectively evaluated images of 21 patients (10 men, 11 women; mean age, 73.9 years) who underwent both full-iodine (600 mg/kg) conventional CT and half-iodine (300 mg/kg) VMI. For each patient, we measured and compared CT attenuation and the contrast-to-noise ratio (CNR) of the aorta, celiac artery, and superior mesenteric artery (SMA). We also compared CT dose index (CTDI). Two board-certified diagnostic radiologists evaluated visualisation of the main trunks and branches of the celiac artery and SMA in maximum-intensity-projection images. We evaluated spatial resolution of the two scans using an acrylic phantom.

Results

The two scans demonstrated no significant difference in CT attenuation of the aorta, celiac artery, and SMA, but CNRs of the aorta and celiac artery were significantly higher in VMI (p = 0.011 and 0.030, respectively). CTDI was significantly higher in VMI (p = 0.024). There was no significant difference in visualisation of the main trunk of the celiac artery and SMA, but visualisation of the gastroduodenal artery, pancreatic arcade, branch of the SMA, marginal arteries, and vasa recta was significantly better in the conventional scan (p < 0.001). The calculated modular transfer function (MTF) suggested decreased spatial resolution of the half-iodine VMI.

Conclusion

Large-vessel depiction and CNRs were comparable between full-iodine conventional CT and half-iodine VMI images, but VMI did not permit clear visualisation of small arteries and required a larger radiation dose.

Key Points

・Reducing the dose of iodine contrast medium is essential for chronic kidney disease patients to prevent contrast-induced nephropathy.
・In virtual monochromatic images at low keV, contrast of relatively large vessels is maintained even with reduced iodine load, but visibility of small vessels is impaired with decreased spatial resolution.
・We should be aware about the advantages and disadvantages associated with virtual monochromatic imaging with reduced iodine dose.
Literatur
1.
Zurück zum Zitat Hong KC, Freeny PC (1999) Pancreaticoduodenal arcades and dorsal pancreatic artery: comparison of CT angiography with three-dimensional volume rendering, maximum intensity projection, and shaded-surface display. AJR Am J Roentgenol 172:925–931CrossRef Hong KC, Freeny PC (1999) Pancreaticoduodenal arcades and dorsal pancreatic artery: comparison of CT angiography with three-dimensional volume rendering, maximum intensity projection, and shaded-surface display. AJR Am J Roentgenol 172:925–931CrossRef
2.
Zurück zum Zitat Winter TC 3rd, Freeny PC, Nghiem HV et al (1995) Hepatic arterial anatomy in transplantation candidates: evaluation with three-dimensional CT arteriography. Radiology 195:363–370CrossRef Winter TC 3rd, Freeny PC, Nghiem HV et al (1995) Hepatic arterial anatomy in transplantation candidates: evaluation with three-dimensional CT arteriography. Radiology 195:363–370CrossRef
3.
Zurück zum Zitat Marti M, Artigas JM, Garzón G, Alvarez-Sala R, Soto JA (2012) Acute lower intestinal bleeding: feasibility and diagnostic performance of CT angiography. Radiology 262:109–116CrossRef Marti M, Artigas JM, Garzón G, Alvarez-Sala R, Soto JA (2012) Acute lower intestinal bleeding: feasibility and diagnostic performance of CT angiography. Radiology 262:109–116CrossRef
4.
Zurück zum Zitat McCullough PA, Wolyn R, Rocher LL, Levin RN, O’Neill WW (1997) Acute renal failure after coronary intervention: incidence, risk factors, and relationship to mortality. Am J Med 103:368–375CrossRef McCullough PA, Wolyn R, Rocher LL, Levin RN, O’Neill WW (1997) Acute renal failure after coronary intervention: incidence, risk factors, and relationship to mortality. Am J Med 103:368–375CrossRef
5.
Zurück zum Zitat McCullough PA, Choi JP, Feghali GA et al (2016) Contrast-induced acute kidney injury. J Am Coll Cardiol 68:1465–1473CrossRef McCullough PA, Choi JP, Feghali GA et al (2016) Contrast-induced acute kidney injury. J Am Coll Cardiol 68:1465–1473CrossRef
6.
Zurück zum Zitat Thomsen HS, Morcos SK (2003) Contrast media and the kidney: European Society of Urogenital Radiology (ESUR) guidelines. Br J Radiol 76:513–518CrossRef Thomsen HS, Morcos SK (2003) Contrast media and the kidney: European Society of Urogenital Radiology (ESUR) guidelines. Br J Radiol 76:513–518CrossRef
7.
Zurück zum Zitat Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259:257–262CrossRef Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259:257–262CrossRef
8.
Zurück zum Zitat Yuan R, Shuman WP, Earls JP et al (2012) Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography–a prospective randomized trial. Radiology 262:290–297CrossRef Yuan R, Shuman WP, Earls JP et al (2012) Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography–a prospective randomized trial. Radiology 262:290–297CrossRef
9.
Zurück zum Zitat Lee JW, Lee G, Lee NK et al (2016) Effectiveness of adaptive statistical iterative reconstruction for 64-slice dual-energy computed tomography pulmonary angiography in patients with a reduced iodine load: comparison with standard computed tomography pulmonary angiography. J Comput Assist Tomogr 40:777–783CrossRef Lee JW, Lee G, Lee NK et al (2016) Effectiveness of adaptive statistical iterative reconstruction for 64-slice dual-energy computed tomography pulmonary angiography in patients with a reduced iodine load: comparison with standard computed tomography pulmonary angiography. J Comput Assist Tomogr 40:777–783CrossRef
10.
Zurück zum Zitat Ma CL, Chen XX, Lei YX et al (2016) Clinical value of dual-energy spectral imaging with adaptive statistical iterative reconstruction for reducing contrast medium dose in CT portal venography: in comparison with standard 120-kVp imaging protocol. Br J Radiol 89:20151022CrossRef Ma CL, Chen XX, Lei YX et al (2016) Clinical value of dual-energy spectral imaging with adaptive statistical iterative reconstruction for reducing contrast medium dose in CT portal venography: in comparison with standard 120-kVp imaging protocol. Br J Radiol 89:20151022CrossRef
11.
Zurück zum Zitat Pontone G, Bertella E, Mushtaq S et al (2014) Coronary artery disease: diagnostic accuracy of CT coronary angiography—a comparison of high and standard spatial resolution scanning. Radiology 271:688–694CrossRef Pontone G, Bertella E, Mushtaq S et al (2014) Coronary artery disease: diagnostic accuracy of CT coronary angiography—a comparison of high and standard spatial resolution scanning. Radiology 271:688–694CrossRef
12.
Zurück zum Zitat Hinson JS, Ehmann MR, Fine DM et al (2017) Risk of acute kidney injury after intravenous contrast media administration. Ann Emerg Med 69:577–586CrossRef Hinson JS, Ehmann MR, Fine DM et al (2017) Risk of acute kidney injury after intravenous contrast media administration. Ann Emerg Med 69:577–586CrossRef
14.
Zurück zum Zitat Richard S, Husarik DB, Yadava G, Murphy SN, Samei E (2012) Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys 39:4115–4122CrossRef Richard S, Husarik DB, Yadava G, Murphy SN, Samei E (2012) Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys 39:4115–4122CrossRef
15.
Zurück zum Zitat Samei E, Richard S (2015) Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 42:314–323CrossRef Samei E, Richard S (2015) Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 42:314–323CrossRef
16.
Zurück zum Zitat Kundel HL, Polansky M (2003) Measurement of observer agreement. Radiology 228:303–308CrossRef Kundel HL, Polansky M (2003) Measurement of observer agreement. Radiology 228:303–308CrossRef
17.
Zurück zum Zitat McCollough CH, Leng S, Yu L, Fletcher JG (2015) Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology 276:637–653CrossRef McCollough CH, Leng S, Yu L, Fletcher JG (2015) Dual- and multi-energy CT: principles, technical approaches, and clinical applications. Radiology 276:637–653CrossRef
18.
Zurück zum Zitat Sellerer T, Noël PB, Patino M et al (2018) Dual-energy CT: a phantom comparison of different platforms for abdominal imaging. Eur Radiol 28:2745–2755CrossRef Sellerer T, Noël PB, Patino M et al (2018) Dual-energy CT: a phantom comparison of different platforms for abdominal imaging. Eur Radiol 28:2745–2755CrossRef
19.
Zurück zum Zitat Rose A (1948) The sensitivity performance of the human eye on an absolute scale. J Opt Soc Am 38:196–208CrossRef Rose A (1948) The sensitivity performance of the human eye on an absolute scale. J Opt Soc Am 38:196–208CrossRef
20.
Zurück zum Zitat Korn A, Fenchel M, Bender B et al (2012) Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection. AJNR Am J Neuroradiol 33:218–224CrossRef Korn A, Fenchel M, Bender B et al (2012) Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection. AJNR Am J Neuroradiol 33:218–224CrossRef
21.
Zurück zum Zitat Gramer BM, Muenzel D, Leber V et al (2012) Impact of iterative reconstruction on CNR and SNR in dynamic myocardial perfusion imaging in an animal model. Eur Radiol 22:2654–2661CrossRef Gramer BM, Muenzel D, Leber V et al (2012) Impact of iterative reconstruction on CNR and SNR in dynamic myocardial perfusion imaging in an animal model. Eur Radiol 22:2654–2661CrossRef
22.
Zurück zum Zitat Christianson O, Chen JJ, Yang Z et al (2015) An improved index of image quality for task-based performance of CT iterative reconstruction across three commercial implementations. Radiology 275:725–734CrossRef Christianson O, Chen JJ, Yang Z et al (2015) An improved index of image quality for task-based performance of CT iterative reconstruction across three commercial implementations. Radiology 275:725–734CrossRef
23.
Zurück zum Zitat Goenka AH, Herts BR, Obuchowski NA et al (2014) Effect of reduced radiation exposure and iterative reconstruction on detection of low-contrast low-attenuation lesions in an anthropomorphic liver phantom: an 18-reader study. Radiology 272:154–163CrossRef Goenka AH, Herts BR, Obuchowski NA et al (2014) Effect of reduced radiation exposure and iterative reconstruction on detection of low-contrast low-attenuation lesions in an anthropomorphic liver phantom: an 18-reader study. Radiology 272:154–163CrossRef
24.
Zurück zum Zitat Jensen K, Andersen HK, Smedby Ö et al (2018) Quantitative measurements versus receiver operating characteristics and visual grading regression in CT images reconstructed with iterative reconstruction: a phantom study. Acad Radiol 25:509–518CrossRef Jensen K, Andersen HK, Smedby Ö et al (2018) Quantitative measurements versus receiver operating characteristics and visual grading regression in CT images reconstructed with iterative reconstruction: a phantom study. Acad Radiol 25:509–518CrossRef
25.
Zurück zum Zitat Jensen K, Martinsen AC, Tingberg A, Aaløkken TM, Fosse E (2014) Comparing five different iterative reconstruction algorithms for computed tomography in an ROC study. Eur Radiol 24:2989–3002CrossRef Jensen K, Martinsen AC, Tingberg A, Aaløkken TM, Fosse E (2014) Comparing five different iterative reconstruction algorithms for computed tomography in an ROC study. Eur Radiol 24:2989–3002CrossRef
26.
Zurück zum Zitat Sahani D, Mehta A, Blake M, Prasad S, Harris G, Saini S (2004) Preoperative hepatic vascular evaluation with CT and MR angiography: implications for surgery. Radiographics 24:1367–1380CrossRef Sahani D, Mehta A, Blake M, Prasad S, Harris G, Saini S (2004) Preoperative hepatic vascular evaluation with CT and MR angiography: implications for surgery. Radiographics 24:1367–1380CrossRef
27.
Zurück zum Zitat Lechel U, Becker C, Langenfeld-Jäger G, Brix G (2009) Dose reduction by automatic exposure control in multidetector computed tomography: comparison between measurement and calculation. Eur Radiol 19:1027–1034CrossRef Lechel U, Becker C, Langenfeld-Jäger G, Brix G (2009) Dose reduction by automatic exposure control in multidetector computed tomography: comparison between measurement and calculation. Eur Radiol 19:1027–1034CrossRef
28.
Zurück zum Zitat Katsura M, Sato J, Akahane M, Misi Y, Sumida K, Abe O (2017) Effects of pure and hybrid iterative reconstruction algorithms on high-resolution computed tomography in the evaluation of interstitial lung disease. Eur J Radiol 93:243–251CrossRef Katsura M, Sato J, Akahane M, Misi Y, Sumida K, Abe O (2017) Effects of pure and hybrid iterative reconstruction algorithms on high-resolution computed tomography in the evaluation of interstitial lung disease. Eur J Radiol 93:243–251CrossRef
29.
Zurück zum Zitat Nishiyama Y, Tada K, Nishiyama Y et al (2016) Effect of the forward-projected model-based iterative reconstruction solution algorithm on image quality and radiation dose in pediatric cardiac computed tomography. Pediatr Radiol 46:1663–1670CrossRef Nishiyama Y, Tada K, Nishiyama Y et al (2016) Effect of the forward-projected model-based iterative reconstruction solution algorithm on image quality and radiation dose in pediatric cardiac computed tomography. Pediatr Radiol 46:1663–1670CrossRef
Metadaten
Titel
Comparison of full-iodine conventional CT and half-iodine virtual monochromatic imaging: advantages and disadvantages
verfasst von
Haruto Sugawara
Shigeru Suzuki
Yoshiaki Katada
Takuya Ishikawa
Rika Fukui
Yuzo Yamamoto
Osamu Abe
Publikationsdatum
12.09.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 3/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5724-4

Weitere Artikel der Ausgabe 3/2019

European Radiology 3/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.